Low Power, Three Electrode  
Electrocardiogram (ECG) Analog Front End  
ADAS1000-3/ADAS1000-4  
Data Sheet  
The ADAS1000-4 is a full-featured, 3-channel ECG including  
respiration and pace detection, while the ADAS1000-3 offers  
only ECG channels with no respiration or pace features.  
FEATURES  
Biopotential signals in; digitized signals out  
3 acquisition (ECG) channels and one driven lead  
Can be ganged for 8 electrode + RLD using master  
ADAS1000 or ADAS1000-1  
The ADAS1000-3/ADAS1000-4 are designed to simplify  
the task of acquiring and ensuring quality ECG signals.  
They provide a low power, small data acquisition system for  
biopotential applications. Auxiliary features that aid in better  
quality ECG signal acquisition include: multichannel averaged  
driven lead, selectable reference drive, fast overload recovery,  
flexible respiration circuitry returning magnitude and phase  
information, internal pace detection algorithm operating  
on three leads, and the option of ac or dc lead-off detection.  
Several digital output options ensure flexibility when monitor-  
ing and analyzing signals. Value-added cardiac post processing  
is executed externally on a DSP, microprocessor, or FPGA.  
AC and DC lead-off detection  
Internal pace detection algorithm on 3 leads  
Support for user’s own pace  
Thoracic impedance measurement (internal/external path)  
Selectable reference lead  
Scalable noise vs. power control, power-down modes  
Low power operation from  
11 mW (1 lead), 15 mW (3 leads)  
Lead or electrode data available  
Supports AAMI EC11:1991/(R)2001/(R)2007, AAMI EC38  
R2007, EC13:2002/(R)2007, IEC60601-1 ed. 3.0 b:2005,  
IEC60601-2-25 ed. 2.0 :2011, IEC60601-2-27 ed. 2.0  
b:2005, IEC60601-2-51 ed. 1.0 b: 2005  
Fast overload recovery  
Because ECG systems span different applications, the  
ADAS1000-3/ADAS1000-4 feature a power/noise scaling  
architecture where the noise can be reduced at the expense  
of increasing power consumption. Signal acquisition channels  
may be shut down to save power. Data rates can be reduced to  
save power.  
Low or high speed data output rates  
Serial interface SPI-/QSPI™-/DSP-compatible  
56-lead LFCSP package (9 mm × 9 mm)  
64-lead LQFP package (10 mm × 10 mm body size)  
To ease manufacturing tests and development as well as offer  
holistic power-up testing, the ADAS1000-3/ADAS1000-4  
offer a suite of features, such as dc and ac test excitation via  
the calibration DAC and CRC redundancy testing in addition  
to readback of all relevant register address space.  
APPLICATIONS  
ECG: monitor and diagnostic  
Bedside patient monitoring, portable telemetry, Holter,  
AED, cardiac defibrillators, ambulatory monitors, pace  
maker programmer, patient transport, stress testing  
The input structure is a differential amplifier input thereby  
allowing users a variety of configuration options to best suit  
their application.  
GENERAL DESCRIPTION  
The ADAS1000-3/ADAS1000-4 measure electro cardiac (ECG)  
signals, thoracic impedance, pacing artifacts, and lead-on/off  
status and output this information in the form of a data frame  
supplying either lead/vector or electrode data at programmable  
data rates. Its low power and small size make it suitable for  
portable, battery-powered applications. The high performance  
also makes it suitable for higher end diagnostic machines.  
The ADAS1000-3/ADAS1000-4 are available in two package  
options: either a 56-lead LFCSP or a 64-lead LQFP package;  
they are specified over −40°C to +85°C temperature range.  
Rev. B  
Document Feedback  
Information furnished by Analog Devices is believed to be accurate and reliable. However, no  
responsibility is assumed by Analog Devices for its use, nor for any infringements of patents or other  
rightsof third parties that may result fromits use. Specifications subject to change without notice. No  
license is granted by implication or otherwise under any patent or patent rights of Analog Devices.  
Trademarks andregisteredtrademarks are the property of their respective owners.  
One Technology Way, P.O. Box 9106, Norwood, MA 02062-9106, U.S.A.  
Tel: 781.329.4700 ©2012–2015 Analog Devices, Inc. All rights reserved.  
Technical Support  
www.analog.com  
 
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
TABLE OF CONTENTS  
Features .............................................................................................. 1  
Evaluating Respiration Performance ....................................... 41  
Pacing Artifact Detection Function (ADAS1000-4 Only).... 41  
Biventricular Pacers ................................................................... 45  
Pace Detection Measurements ................................................. 45  
Evaluating Pace Detection Performance................................. 45  
Pace Width .................................................................................. 45  
Pace Latency................................................................................ 45  
Pace Detection via Secondary Serial Interface....................... 45  
Filtering ....................................................................................... 46  
Voltage Reference ....................................................................... 47  
Gang Mode Operation............................................................... 47  
Interfacing in Gang Mode......................................................... 50  
Serial Interfaces............................................................................... 51  
Standard Serial Interface ........................................................... 51  
Secondary Serial Interface......................................................... 55  
Applications....................................................................................... 1  
General Description......................................................................... 1  
Revision History ............................................................................... 3  
Functional Block Diagram .............................................................. 4  
Specifications..................................................................................... 5  
Noise Performance ..................................................................... 10  
Timing Characteristics .............................................................. 11  
Absolute Maximum Ratings.......................................................... 14  
Thermal Resistance .................................................................... 14  
ESD Caution................................................................................ 14  
Pin Configurations and Function Descriptions ......................... 15  
Typical Performance Characteristics ........................................... 18  
Applications Information .............................................................. 25  
Overview...................................................................................... 25  
ECG Inputs—Electrodes/Leads................................................ 27  
ECG Channel .............................................................................. 28  
Electrode/Lead Formation and Input Stage Configuration.. 29  
Defibrillator Protection ............................................................. 33  
ESIS Filtering............................................................................... 33  
ECG Path Input Multiplexing................................................... 33  
Common-Mode Selection and Averaging .............................. 34  
Wilson Central Terminal (WCT)............................................. 35  
Right Leg Drive/Reference Drive ............................................. 35  
Calibration DAC......................................................................... 36  
Gain Calibration ......................................................................... 36  
Lead-Off Detection .................................................................... 36  
Shield Driver ............................................................................... 37  
Respiration (ADAS1000-4 Model Only)................................. 37  
RESET  
.......................................................................................... 55  
Function................................................................................ 55  
PD  
SPI Output Frame Structure (ECG and Status Data) ................ 56  
SPI Register Definitions and Memory Map................................ 57  
Control Registers Details............................................................... 58  
Interface Examples ..................................................................... 74  
Software Flowchart .................................................................... 77  
Power Supply, Grounding, and Decoupling Strategy............ 78  
AVDD .......................................................................................... 78  
ADCVDD and DVDD Supplies............................................... 78  
Unused Pins/Paths ..................................................................... 78  
Layout Recommendations ........................................................ 78  
Outline Dimensions....................................................................... 79  
Ordering Guide .......................................................................... 80  
Rev. B | Page 2 of 80  
Data Sheet  
ADAS1000-3/ADAS1000-4  
REVISION HISTORY  
1/15—Rev. A to Rev. B  
1/13—Rev. 0 to Rev. A  
Changed Frequency Range from 2.031 kHz (Typ) to 2.039 kHz  
(Typ); Table 2 .....................................................................................7  
Changes to Endnote 3; Table 3 and Changes to Table 4.............10  
Changes to Figure 16 ......................................................................18  
Changes to Figure 39 and Figure 40 .............................................22  
Changes to ECG Channel Section ................................................28  
Changes to Digital Lead Mode and Calculation Section and  
Electrode Mode: Common Electrode A and Common  
Changes to Endnote 2, Table 1 ........................................................3  
Changes to Excitation Current Test Conditions/Comments ......5  
Added Table 3....................................................................................9  
Changes to Figure 36, Figure 37, and Figure 39 .........................21  
Changes to Respiration (ADAS1000-4 Model Only) Section  
and Figure 63 ...................................................................................34  
Changes to Figure 64 ......................................................................35  
Changes to Figure 65 ......................................................................36  
Added Evaluating Pace Detection Performance Section...........40  
Changes to Clocks Section.............................................................49  
Changes to RESPAMP Bits Function Description, Table 29.....55  
Electrode B Configuration Section; Added Figure 55;  
Renumbered Sequentially ..............................................................29  
Deleted Table 11; Renumbered Sequentially...............................29  
Added Figure 56 and Figure 57.....................................................30  
Added Figure 58 and Figure 59.....................................................31  
Added Figure 60 ..............................................................................32  
Changes to Figure 61, Figure 62, and Figure 63..........................33  
Changes to Lead-Off Detection Section ......................................36  
Added Figure 66 and Changes to Respiration (ADAS1000-4  
Model Only) Section.......................................................................37  
Changes to External Respiration Path Section............................39  
Changes to Respiration Carrier Frequency Section; Added  
Table 13 and Table 14......................................................................40  
Changes to Pacing Artifact Detection Function (ADAS1000-4  
Only) Section ...................................................................................41  
Changes to Table 15 ........................................................................42  
Changes to Detection Algorithm Overview Section..................43  
Changes to Pace Edge Threshold, Pace Level Threshold, Pace  
Amplitude Threshold, Pace Validation Filters, and Pace Width  
Filter Sections ..................................................................................44  
Changes to Evaluating Pace Detection Performance Section and  
Added Pace Width Section ............................................................45  
Changes to Voltage Reference Section .........................................57  
11/12—Revision 0: Initial Version  
DRDY  
Changes to Data Ready (  
) Section.....................................53  
Changes to Secondary Serial Interface Section and Table 25....55  
Changes to Bit 3, Table 28 ..............................................................58  
Changes to Table 43 ........................................................................68  
Changes to Table 45 ........................................................................69  
Changes to Table 51 and Table 52 .................................................72  
Changes to Example 1: Initialize the Device for ECG Capture  
and Start Streaming Data Section .................................................74  
Rev. B | Page 3 of 80  
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
FUNCTIONAL BLOCK DIAGRAM  
REFIN REFOUT CAL_DAC_IO RLD_SJ  
RLD_OUT CM_IN  
CM_OUT/WCT  
SHIELD  
AVDD  
IOVDD  
DRIVEN  
LEAD  
AMP  
SHIELD  
DRIVE  
AMP  
VREF  
ADCVDD  
ADCVDD, DVDD  
1.8V  
REGULATORS  
DVDD  
CALIBRATION  
DAC  
+
VCM_REF  
(1.3V)  
RESPIRATION  
DAC  
COMMON-  
MODE AMP  
AC  
LEAD-OFF  
DAC  
AC  
LEAD-OFF  
DETECTION  
10k  
BUFFER  
PACE  
DETECTION  
CS  
MUXES  
SCLK  
3× ECG PATH  
AMP  
ELECTRODES  
×3  
SDI  
FILTERS,  
CONTROL,  
AND  
INTERFACE  
LOGIC  
ADC  
SDO  
DRDY  
GPIO0/MCS  
GPIO1/MSCLK  
GPIO2/MSDO  
GPIO3  
EXT_RESP_LA  
EXT_RESP_LL  
ADC  
AMP  
EXT_RESP_RA  
CLOCK GEN/OSC/  
EXTERNAL CLK  
SOURCE  
CLK_IO  
RESPIRATION PATH  
ADAS1000-4  
XTAL1  
XTAL2  
Figure 1. ADAS1000-4 3-Channel Full Featured Model  
Table 1. Overview of Features Available from ADAS1000 Generics  
Pace  
Detection  
Shield  
Driver  
Master  
Package  
Option  
Generic  
ECG  
Operation  
Master/slave  
Master/slave  
Slave  
Right Leg Drive  
Respiration  
Interface1  
Yes  
ADAS1000  
ADAS1000-1  
ADAS1000-22  
ADAS1000-3  
ADAS1000-4  
5 ECG channels  
5 ECG channels  
5 ECG channels  
3 ECG channels  
3 ECG channels  
Yes  
Yes  
Yes  
Yes  
Yes  
LFCSP, LQFP  
LFCSP  
Yes  
Yes  
LFCSP, LQFP  
LFCSP, LQFP  
LFCSP, LQFP  
Master/slave  
Master/slave  
Yes  
Yes  
Yes  
Yes  
Yes  
Yes  
Yes  
Yes  
1 Master interface is provided for users wishing to utilize their own digital pace algorithm; see the Secondary Serial Interface section.  
2 This is a companion device for increased channel count purposes. It has a subset of features and is not intended for standalone use. It may be used in conjunction with  
any master device.  
Rev. B | Page 4 of 80  
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
SPECIFICATIONS  
AVDD = 3.3 V 5%, IOVDD = 1.65 V to 3.6 V, AGND = DGND = 0 V, REFIN tied to REFOUT, externally supplied crystal/clock =  
8.192 MHz. Decoupling for reference and supplies as noted in the Power Supply, Grounding, and Decoupling Strategy section. TA =  
−40°C to +85°C, unless otherwise noted. Typical specifications are mean values at TA = 25°C.  
For specified performance, internal ADCVDD and DVDD linear regulators have been used. They may be supplied from external  
regulators. ADCVDD = 1.8 V 5%, DVDD = 1.8 V 5%.  
Front-end gain settings: GAIN 0 = ×1.4, GAIN 1 = ×2.1, GAIN 2 = ×2.8, GAIN 3 = ×4.2.  
Table 2.  
Parameter  
Min  
Typ  
Max  
Unit  
Test Conditions/Comments  
ECG CHANNEL  
These specifications apply to the following pins:  
ECG1_LA, ECG2_LL, ECG3_RA, CM_IN (CE mode),  
EXT_RESP_xx pins when used in extend switch mode  
Electrode Input Range  
Independent of supply  
GAIN 0 (gain setting ×1.4)  
GAIN 1 (gain setting ×2.1)  
GAIN 2 (gain setting ×2.8)  
GAIN 3 (gain setting ×4.2)  
Relates to each electrode input; over operating range;  
dc and ac lead-off are disabled  
AGND to AVDD  
0.3  
0.63  
0.8  
0.97  
−40  
1.3  
1.3  
1.3  
1.3  
1
2.3  
1.97  
1.8  
1.63  
+40  
V
V
V
V
Input Bias Current  
Input Offset  
nA  
−200  
+200  
nA  
mV  
−7  
Electrode/vector mode with VCM = VCM_REF  
GAIN 3  
GAIN 2  
GAIN 1  
GAIN 0  
−7  
mV  
mV  
mV  
μV/°C  
GΩ||pF  
−15  
−22  
2
Input Offset Tempco1  
Input Amplifier Input  
1||10  
At 10 Hz  
Impedance2  
CMRR2  
105  
110  
dB  
51 kΩ imbalance, 60 Hz with 300 mV differential dc  
offset; per AAMI/IEC standards; with driven leg loop  
closed  
Crosstalk1  
Resolution2  
80  
19  
18  
dB  
Between channels  
Bits  
Bits  
Electrode/vector mode, 2 kHz data rate, 24-bit data-word  
Electrode/vector mode, 16 kHz data rate, 24-bit data-  
word  
16  
Bits  
Electrode/analog lead mode, 128 kHz data rate, 16-bit  
data-word  
Integral Nonlinearity Error  
Differential Nonlinearity Error  
Gain2  
30  
5
ppm  
ppm  
GAIN 0; all data rates  
GAIN 0  
Referred to input; (2 × VREF)/gain/(2N − 1); Applies  
after factory calibration. User calibration adjusts this  
number.  
GAIN 0 (×1.4)  
GAIN 1 (×2.1)  
GAIN 2 (×2.8)  
GAIN 3 (×4.2)  
4.9  
µV/LSB  
μV/LSB  
μV/LSB  
μV/LSB  
μV/LSB  
μV/LSB  
μV/LSB  
μV/LSB  
μV/LSB  
μV/LSB  
At 19-bit level in 2 kHz data rate  
At 18-bit level in 16 kHz data rate  
At 16-bit level in 128 kHz data rate  
At 19-bit level in 2 kHz data rate  
At 18-bit level in 16 kHz data rate  
At 16-bit level in 128 kHz data rate  
At 19-bit level in 2 kHz data rate  
At 18-bit level in 16 kHz data rate  
At 16-bit level in 128 kHz data rate  
No factory calibration for this gain setting  
At 19-bit level in 2 kHz data rate  
At 18-bit level in 16 kHz data rate  
At 16-bit level in 128 kHz data rate  
9.81  
39.24  
3.27  
6.54  
26.15  
2.45  
4.9  
19.62  
1.63  
3.27  
13.08  
μV/LSB  
μV/LSB  
Rev. B | Page 5 of 80  
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
Parameter  
Min  
Typ  
Max  
Unit  
Test Conditions/Comments  
Gain Error  
−1  
+0.01  
+1  
%
GAIN 0 to GAIN 2, factory calibrated; programmable  
user or factory calibration option enables; factory gain  
calibration applies only to standard ECG interface  
−2  
+0.1  
+0.02  
+0.1  
25  
+2  
%
GAIN 3 setting, no factory calibration for this gain  
Gain Matching  
−0.1  
−0.5  
+0.1  
+0.5  
%
GAIN 0 to GAIN 2  
GAIN 3  
%
Gain Tempco1  
ppm/°C  
Input Referred Noise1  
GAIN 2, 2 kHz data rate, see Table 4  
0.5 Hz to 40 Hz; high performance mode  
0.05 Hz to 150 Hz; high performance mode  
0.05 Hz to 150 Hz; low power mode  
0.05 Hz to 150 Hz; high performance mode  
0.05 Hz to 150 Hz; low power mode  
0.05 Hz to 150 Hz; high performance mode  
0.05 Hz to 150 Hz; low power mode  
At 120 Hz  
Analog Lead Mode  
6
μV p-p  
μV p-p  
μV p-p  
μV p-p  
μV p-p  
μV p-p  
μV p-p  
dB  
10  
12  
11  
12  
14  
16  
100  
65  
104  
100  
Electrode Mode  
Digital Lead Mode  
Power Supply Sensitivity2  
Analog Channel Bandwidth1  
Dynamic Range1  
Signal-to-Noise Ratio1  
COMMON-MODE INPUT  
Input Voltage Range  
Input Impedance2  
kHz  
dB  
GAIN 0, 2 kHz data rate, −0.5 dBFS input signal, 10 Hz  
−0.5 dB FS input signal  
dB  
CM_IN pin  
0.3  
2.3  
V
1||10  
1
GΩ||pF  
nA  
Input Bias Current  
−40  
+40  
Over operating range; dc and ac lead-off disabled  
AGND to AVDD  
−200  
+200  
nA  
COMMON-MODE OUTPUT  
VCM_REF  
CM_OUT pin  
1.28  
0.3  
1.3  
1.3  
0.75  
4
1.32  
2.3  
V
Internal voltage; independent of supply  
No dc load  
Output Voltage, VCM  
Output Impedance1  
Short-Circuit Current1  
V
kΩ  
mA  
%
Not intended to drive current  
Electrode Summation  
Weighting Error2  
1
Resistor matching error  
RESPIRATION FUNCTION  
(ADAS1000-4 ONLY)  
These specifications apply to the following pins:  
EXT_RESP_LA, EXT_RESP_LL, EXT_RESP_RA and selected  
internal respiration paths (Lead I, Lead II, Lead III)  
Input Voltage Range  
0.3  
2.3  
V
AC-coupled, independent of supply  
Programmable gain (10 states)  
Input Voltage Range (Linear  
Operation)  
1.8/gain  
V p-p  
Input Bias Current  
Input Referred Noise1  
Frequency2  
−10  
1
+10  
nA  
Applies to EXT_RESP_xx pins over AGND to AVDD  
0.85  
μV rms  
kHz  
46.5 to 64  
Programmable frequency, see Table 30  
Excitation Current  
Respiration drive current corresponding to differential  
voltage programmed by the RESPAMP bits in the  
RESPCTL register; internal respiration mode, cable  
5 kΩ/200 pF, 1.2 kΩ chest impedance  
64  
32  
16  
8
μA p-p  
μA p-p  
μA p-p  
μA p-p  
Bits  
Drive Range A  
Drive Range B2  
Drive Range C2  
Drive Range D2  
Update rate 125 Hz  
Resolution2  
Measurement Resolution1  
24  
0.2  
Ω
Cable <5 kΩ/200 pF per electrode, body resistance  
modeled as 1.2 kΩ  
0.02  
Ω
No cable impedance, body resistance modeled as 1.2 kΩ  
Digitally programmable in steps of 1  
In-Amp Gain1  
Gain Error  
1 to 10  
1
%
LSB weight for GAIN 0 setting  
Gain Tempco1  
25  
ppm/C  
Rev. B | Page 6 of 80  
Data Sheet  
ADAS1000-3/ADAS1000-4  
Parameter  
Min  
Typ  
Max  
Unit  
Test Conditions/Comments  
RIGHT LEG DRIVE/DRIVEN LEAD  
Output Voltage Range  
RLD_OUT Short-Circuit Current  
0.2  
−5  
AVDD − 0.2  
+5  
V
2
mA  
External protection resistor required to meet  
regulatory patient current limits; output shorted to  
AVDD/AGND  
Closed-Loop Gain Range2  
Slew Rate2  
Input Referred Noise1  
25  
V/V  
200  
8
mV/ms  
μV p-p  
MHz  
0.05 Hz to 150 Hz  
Amplifier GBP2  
1.5  
DC LEAD-OFF  
Internal current source, pulls up open ECG pins;  
programmable in 10 nA steps: 10 nA to 70 nA  
Lead-Off Current Accuracy  
High Threshold Level1  
10  
2.4  
%
V
Of programmed value  
Inputs are compared to threshold levels; if inputs  
exceed levels, lead-off flag is raised  
Low Threshold Level1  
Threshold Accuracy  
AC LEAD-OFF  
0.2  
25  
V
mV  
Programmable in 4 steps: 12.5 nA rms, 25 nA rms,  
50 nA rms, 100 nA rms  
Frequency Range  
Lead-Off Current Accuracy  
REFIN  
2.039  
10  
kHz  
%
Fixed frequency  
Of programmed value, measured into low impedance  
Input Range2  
1.76  
450  
1.8  
1.84  
950  
V
Channel gain scales directly with REFIN  
Per active ADC  
Three ECG channels and respiration enabled  
Input Current  
113  
675  
μA  
μA  
REFOUT  
On-chip reference voltage for ADC; not intended to  
drive other components reference inputs directly,  
must be buffered externally  
Output Voltage, VREF  
Reference Tempco1  
Output Impedance2  
Short-Circuit Current1  
Voltage Noise1  
1.785  
1.8  
10  
1.815  
V
ppm/°C  
Ω
0.1  
4.5  
33  
mA  
Short circuit to ground  
μV p-p  
μV p-p  
0.05 Hz to 150 Hz (ECG band)  
0.05 Hz to 5 Hz (respiration)  
17  
CALIBRATION DAC  
Available on CAL_DAC_IO (output for master, input for  
slave)  
DAC Resolution  
10  
Bits  
V
V
LSB  
kΩ  
Full-Scale Output Voltage  
Zero-Scale Output Voltage  
DNL  
2.64  
0.24  
−1  
2.7  
0.3  
2.76  
0.36  
+1  
No load, nominal FS output is 1.5 × REFOUT  
No load  
Output Series Resistance2  
10  
5
Not intended to drive low impedance load, used for  
slave CAL_DAC_IO configured as an input  
Input Current  
nA  
When used as an input  
CALIBRATION DAC TEST TONE  
Output Voltage  
Square Wave  
Low Frequency Sine Wave  
High Frequency Sine Wave  
SHIELD DRIVER  
Output Voltage Range  
Gain  
0.9  
1
1
10  
150  
1.1  
2.3  
mV p-p  
Hz  
Hz  
Rides on common-mode voltage, VCM_REF = 1.3 V  
Hz  
0.3  
V
Rides on common-mode voltage (VCM)  
1
V/V  
mV  
μA  
nF  
Offset Voltage  
−20  
+20  
25  
10  
Short-Circuit Current  
Stable Capacitive Load2  
CRYSTAL OSCILLATOR  
Frequency2  
15  
Output current limited by internal series resistance  
Applied to XTAL1 and XTAL2  
Internal startup  
8.192  
15  
MHz  
ms  
Start-Up Time2  
Rev. B | Page 7 of 80  
ADAS1000-3/ADAS1000-4  
Data Sheet  
Parameter  
Min  
Typ  
Max  
Unit  
Test Conditions/Comments  
CLOCK_IO  
External clock source supplied to CLK_IO; this pin  
is configured as an input when the device is  
programmed as a slave  
Operating Frequency2  
Input Duty Cycle2  
Output Duty Cycle2  
8.192  
50  
MHz  
%
20  
80  
%
DIGITAL INPUTS  
Applies to all digital inputs  
Input Low Voltage, VIL  
Input High Voltage, VIH  
Input Current, IIH, IIL  
0.3 × IOVDD  
V
V
μA  
μA  
pF  
0.7 × IOVDD  
−1  
−20  
+1  
+20  
RESET  
has an internal pull-up resistor  
Pin Capacitance2  
DIGITAL OUTPUTS  
Output Low Voltage, VOL  
Output High Voltage, VOH  
Output Rise/Fall Time  
DVDD REGULATOR  
Output Voltage  
Available Current1  
Short-Circuit Current Limit  
ADCVDD REGULATOR  
3
4
0.4  
V
V
ns  
ISINK = 1 mA  
ISOURCE = −1 mA  
Capacitive load = 15 pF, 20% to 80%  
Internal 1.8 V regulator for DVDD  
IOVDD − 0.4  
1.75  
1.8  
1
40  
1.85  
V
mA  
mA  
Droop < 10 mV; for external device loading purposes  
Internal 1.8 V regulator for ADCVDD; not  
recommended as a supply for other circuitry  
Output Voltage  
Short-Circuit Current Limit  
POWER SUPPLY RANGES2  
AVDD  
IOVDD  
ADCVDD  
1.75  
1.8  
40  
1.85  
V
mA  
3.15  
1.65  
1.71  
1.71  
3.3  
5.5  
3.6  
1.89  
1.89  
V
V
V
V
1.8  
1.8  
If applied by external 1.8 V regulator  
If applied by external 1.8 V regulator  
DVDD  
POWER SUPPLY CURRENTS  
AVDD Standby Current  
IOVDD Standby Current  
785  
1
975  
60  
μA  
μA  
EXTERNALLY SUPPLIED ADCVDD  
AND DVDD  
All three channels enabled, RLD enabled, pace enabled  
AVDD Current  
ADCVDD Current  
DVDD Current  
2.4  
2.2  
3.2  
4.5  
3.3  
5.4  
2.0  
1.1  
2.0  
4.1  
4.1  
mA  
mA  
mA  
mA  
mA  
mA  
mA  
mA  
mA  
High performance mode  
Low performance mode  
High performance mode, respiration enabled  
High performance mode  
Low performance mode  
6.5  
5.5  
High performance mode, respiration enabled  
High performance mode  
Low performance mode  
High performance mode, respiration enabled  
All three channels enabled, RLD enabled, pace enabled  
4
3
INTERNALLY SUPPLIED ADCVDD  
AND DVDD  
AVDD Current  
9
6.6  
11  
12.6  
9.6  
mA  
mA  
mA  
High performance mode  
Low performance mode  
14.6  
High performance mode, respiration enabled  
All 3 channels enabled, RLD enabled, pace enabled  
POWER DISSIPATION  
Externally Supplied ADCVDD  
and DVDD3  
Three Input Channels and RLD  
19.6  
15.2  
mW  
mW  
High performance (low noise)  
Low power mode  
Internally Supplied ADCVDD  
and DVDD  
All three channels enabled, RLD enabled, pace enabled  
Three Input Channels and RLD  
29.7  
21.8  
mW  
mW  
High performance (low noise)  
Low power mode  
Rev. B | Page 8 of 80  
Data Sheet  
ADAS1000-3/ADAS1000-4  
Parameter  
Min  
Typ  
Max  
Unit  
Test Conditions/Comments  
OTHER FUNCTIONS4  
Power Dissipation  
Respiration  
7.6  
150  
mW  
μW  
Shield Driver  
EXTERNALLY SUPPLIED ADCVDD  
AND DVDD  
Two electrodes enabled for one lead measurement,  
RLD enabled, pace enabled  
AVDD Current  
ADCVDD Current  
DVDD Current  
1.9  
1.7  
3.6  
2.5  
1.7  
0.9  
3.7  
3.7  
5.5  
4.5  
4
mA  
mA  
mA  
mA  
mA  
mA  
High performance mode  
Low performance mode  
High performance mode  
Low performance mode  
High performance mode  
Low performance mode  
3
INTERNALLY SUPPLIED ADCVDD  
AND DVDD  
Two electrodes enabled for one lead measurement,  
RLD enabled, pace enabled  
AVDD Current  
7.3  
5.3  
10.7  
8.2  
mA  
mA  
High performance mode  
Low performance mode  
POWER DISSIPATION  
Two electrodes enabled for one lead measurement,  
RLD enabled, pace enabled  
Externally Supplied ADCVDD  
and DVDD3  
Two Input Channels and RLD  
15.8  
11.7  
mW  
mW  
High performance (low noise)  
Low power mode  
Internally Supplied ADCVDD  
and DVDD  
Two Input Channels and RLD  
24  
17.5  
mW  
mW  
High performance (low noise)  
Low power mode  
1 Guaranteed by characterization, not production tested.  
2 Guaranteed by design, not production tested.  
3 ADCVDD and DVDD can be powered from an internal LDO or, alternatively, can be powered from an external 1.8 V rail, which may result in a lower power solution.  
4 Pace is a digital function and incurs no power penalty.  
Rev. B | Page 9 of 80  
ADAS1000-3/ADAS1000-4  
Data Sheet  
NOISE PERFORMANCE  
Table 3. Typical Input Referred Noise over a 0.5 sec Window (μV p-p)1  
GAIN 0 (×1.4) GAIN 1 (×2.1) GAIN 2 (×2.8) GAIN 3 (×4.2)  
Mode  
Data Rate2  
1 VCM  
0.67 VCM  
0.5 VCM  
0.3 VCM  
Analog Lead Mode3  
High Performance Mode  
2 kHz (0.5 Hz to 40 Hz)  
2 kHz (0.05 Hz to 150 Hz)  
8
14  
6
11  
5
9
4
7.5  
1 Typical values measured at 25°C, not subject to production test.  
2 Data gathered using the 2 kHz packet/frame rate is measured over 0.5 seconds. The ADAS1000-3/ADAS1000-4 internal programmable low-pass filter is configured for  
either 40 Hz or 150 Hz bandwidth. The data is gathered and post processed using a digital filter of either 0.05 Hz or 0.5 Hz to provide data over noted frequency bands.  
3 Analog lead mode as shown in Figure 56.  
Table 4. Typical Input Referred Noise (μV p-p)1  
GAIN 0 (×1.4) GAIN 1 (×2.1) GAIN 2 (×2.8) GAIN 3 (×4.2)  
Mode  
Data Rate2  
1 VCM  
0.67 VCM  
0.5 VCM  
0.3 VCM  
Analog Lead Mode3  
High Performance Mode  
2 kHz (0.5 Hz to 40 Hz)  
2 kHz (0.05 Hz to 150 Hz)  
2 kHz (0.05 Hz to 250 Hz)  
2 kHz (0.05 Hz to 450 Hz)  
16 kHz  
12  
20  
27  
33.5  
95  
8.5  
14.5  
18  
24  
65  
6
10  
14.5  
19  
50  
5
8.5  
10.5  
13.5  
39  
128 kHz  
180  
13  
22  
110  
215  
130  
9.5  
15.5  
75  
105  
7.5  
12  
59  
116  
80  
5.5  
9
45  
85  
Low Power Mode  
2 kHz (0.5 Hz to 40 Hz)  
2 kHz (0.05 Hz to 150 Hz)  
16 kHz  
128 kHz  
145  
Electrode Mode4  
High Performance Mode  
2 kHz (0.5 Hz to 40 Hz)  
2 kHz (0.05 Hz to 150 Hz)  
2 kHz (0.05 Hz to 250 Hz)  
2 kHz (0.05 Hz to 450 Hz)  
16 kHz  
13  
21  
26  
34.5  
100  
190  
14  
22  
110  
218  
9.5  
15  
19  
25  
70  
139  
9.5  
15.5  
75  
8
11  
5.5  
9
11.5  
14.5  
41  
15.5  
20.5  
57  
110  
7.5  
12  
128 kHz  
85  
Low Power Mode  
2 kHz (0.5 Hz to 40 Hz)  
2 kHz (0.05 Hz to 150 Hz)  
16 kHz  
5.5  
9.5  
45  
60  
120  
128 kHz  
145  
88  
Digital Lead Mode5, 6  
High Performance Mode  
2 kHz (0.5 Hz to 40 Hz)  
2 kHz (0.05 Hz to 150 Hz)  
2 kHz (0.05 Hz to 250 Hz)  
2 kHz (0.05 Hz to 450 Hz)  
16 kHz  
2 kHz (0.5 Hz to 40 Hz)  
2 kHz (0.05 Hz to 150 Hz)  
16 kHz  
16  
25  
34  
46  
130  
18  
30  
11  
19  
23  
31  
90  
12.5  
21  
9
6.5  
10  
13  
17.5  
50  
7
15  
18  
24  
70  
10  
16  
80  
Low Power Mode  
11  
58  
145  
100  
1 Typical values measured at 25°C, not subject to production test.  
2 Data gathered using the 2 kHz packet/frame rate is measured over 20 seconds. The ADAS1000-3/ADAS1000-4 internal programmable low-pass filter is configured for  
either 40 Hz or 150 Hz bandwidth. The data is gathered and post processed using a digital filter of either 0.05 Hz or 0.5 Hz to provide data over noted frequency bands.  
3 Analog lead mode as shown in Figure 56.  
4 Single-ended input electrode mode as shown in Figure 59. Electrode mode refers to common electrode A, common electrode B, and single-ended input electrode  
configurations.  
5 Digital lead mode as shown in Figure 57.  
6 Digital lead mode is available in 2 kHz and 16 kHz data rates.  
Rev. B | Page 10 of 80  
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
TIMING CHARACTERISTICS  
Standard Serial Interface  
AVDD = 3.3 V ꢀ5% ꢁOVDD = 1.6ꢀ V to 3.6 V% AGND = DGND = 0 V% REFꢁN tied to REFOUT% externally supplied crystal/clock =  
8.192 MHz. TA = −40°C to +8ꢀ°C% unless otherwise noted. Typical specifications are mean values at TA = 2ꢀ°C.  
Table 5.  
IOVDD  
2.5 V  
Parameter1  
Output Rate2  
3.3 V  
1.8 V  
Unit  
Description  
2
128  
kHz  
Across specified IOVDD supply range; three programmable output data  
rates available as configured in FRMCTL register (see Table 37) 2 kHz,  
16 kHz, 128 kHz; use skip mode for slower rates.  
SCLK Cycle Time  
25  
8.5  
3
40  
9.5  
3
50  
12  
3
ns min  
ns min  
ns min  
ns min  
ns min  
ns typ  
ns max  
ns min  
ns min  
ns min  
ns min  
ns min  
See Table 21 for details on SCLK frequency vs. packet data/frame rates.  
CS valid setup time to rising SCLK.  
tCSSA  
tCSHA  
tCH  
tCL  
tDO  
CS valid hold time to rising SCLK.  
8
8
8.5  
11  
2
8
8
11.5  
19  
2
8
8
20  
24  
2
SCLK high time.  
SCLK low time.  
SCLK falling edge to SDO valid delay; SDO capacitance of 15 pF.  
tDS  
tDH  
tCSSD  
tCSHD  
tCSW  
SDI valid setup time from SCLK rising edge.  
SDI valid hold time from SCLK rising edge.  
CS valid setup time from SCLK rising edge.  
CS valid hold time from SCLK rising edge.  
2
2
2
2
2
2
2
2
2
25  
40  
50  
CS high time between writes (if used). Note that CS is an optional input,  
it may be tied permanently low. See a full description in the Serial  
Interfaces section.  
2
tDRDY_CS  
0
0
0
ns min  
ns typ  
ns min  
DRDY to CS setup time.  
tCSO  
6
7
9
Delay from CS assert to SDO active.  
2
RESET  
RESET  
Low Time  
20  
20  
20  
Minimum pulse width;  
is edge triggered.  
1 Guaranteed by characterization, not production tested.  
2 Guaranteed by design, not production tested.  
SCLK  
tCSSA  
tCSHA  
tCSSD  
tCH tCL  
tCSHD  
CS  
tCSW  
tDS  
tDH  
MSB  
LSB  
DB[31]  
DB[30]  
DB[29]  
DB[25]  
DB[24]  
DB[23]  
DB[1]  
DB[0]  
SDI  
R/W  
ADDRESS  
DATA  
tCSO  
LSB  
MSB  
DO_1  
DO_0  
DRDY  
DO_31  
DO_30  
DO_29  
DO_25  
SDO  
tDO  
Figure 2. Data Read and Write Timing Diagram (CPHA = 1, CPOL = 1)  
Rev. B | Page 11 of 80  
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
tDRDY_CS  
DRDY  
tCH  
SCLK  
tCL  
tCSSA  
tCSSD  
tCSHD  
tCSHA  
CS  
tCSW  
tDS  
tDH  
LSB  
MSB  
LSB  
MSB  
DB[1] DB[0]  
N + 1  
N + 1  
DB[30]  
N + 1  
DB[31]  
N
DB[30]  
N
DB[29]  
N
DB[25]  
N
DB[24]  
N
DB[31]  
N + 1  
SDI  
DB[1]  
DB[23]  
DATA = NOP or 0x40  
MSB  
R/W  
ADDRESS = 0x40 (FRAMES)  
DATA  
tCSO  
MSB  
LSB  
LSB  
DB[24]  
N – 1  
DB[31]  
N – 1  
DB[30]  
N – 1  
DB[25]  
N – 1  
DB[23]  
N – 1  
DB[1]  
N – 1  
DB[0]  
N – 1  
DB[31]  
N
DB[30]  
N
DB[1]  
N
DB[0]  
N
SDO  
DRDY  
tDO  
PREVIOUS DATA  
HEADER (FIRST WORD OF FRAME)  
Figure 3. Starting Read Frame Data (CPHA = 1, CPOL = 1)  
tCH  
SCLK  
tCSSD  
tCL  
t
CSSA  
t
CSHA  
tCSHD  
CS  
tCSW  
tDH  
tDS  
LSB  
DB[0]  
MSB  
DB[31]  
DB[30]  
DB[29]  
DB[28]  
DB[24]  
DB[1]  
DB[2]  
SDI  
R/W  
ADDRESS  
DATA  
tDO  
MSB  
LSB  
DO_0  
LAST  
DO_31  
DO_1  
DO_30  
DO_29  
DO_28  
LAST  
SDO  
LAST  
LAST  
LAST  
tDO  
Figure 4. Data Read and Write Timing Diagram (CPHA = 0, CPOL = 0)  
Rev. B | Page 12 of 80  
Data Sheet  
ADAS1000-3/ADAS1000-4  
Secondary Serial Interface (Master Interface for Customer-Based Digital Pace Algorithm) ADAS1000-4 Only  
AVDD = 3.3 V ꢀ5% ꢁOVDD = 1.6ꢀ V to 3.6 V% AGND = DGND = 0 V% REFꢁN tied to REFOUT% externally supplied crystal/clock =  
8.192 MHz. TA = −40°C to +8ꢀ°C% unless otherwise noted. Typical specifications are mean values at TA = 2ꢀ°C. The following timing  
specifications apply for the master interface when the ECGCTL register is configured for high performance mode (ECGCTL[3] = 1)% see  
Table 28.  
Table 6.  
Parameter1  
Min Typ  
Max Unit Description  
Output Frame Rate2  
128  
2.5 × crystal  
kHz  
All five 16-bit ECG data-words are available at frame rate of 128 kHz only  
2
fSCLK  
MHz Crystal frequency = 8.192 MHz  
frequency  
tMCSSA  
tMDO  
tMCSHD  
tMCSW  
24.4  
ns  
ns  
ns  
ns  
MCS valid setup time  
0
48.8  
MSCLK rising edge to MSDO valid delay  
MCS valid hold time from MSCLK falling edge  
2173  
MCS high time, SPIFW = 0, MCS asserted for entire frame as shown in  
Figure 5, and configured in Table 33  
2026  
ns  
MCS high time, SPIFW = 1, MCS asserted for each word in frame as shown in  
Figure 6 and configured in Table 33  
1 Guaranteed by characterization, not production tested.  
2 Guaranteed by design, not production tested.  
tMSCLK  
2
tMSCLK  
MSCLK  
tMCSSA  
tMCSHD  
MCS  
tMCSW  
SPIFW = 0*  
MSB  
LSB  
D0_0  
MSB  
LSB  
MSB  
LSB  
MSDO  
D0_15  
D5_0  
D0_14  
D0_1  
tMDO  
D1_15  
D1_14  
D6_15  
D6_14  
D6_0  
16-BIT CRC WORD  
HEADER: 0xF AND 12-BIT COUNTER  
5 × 16-BIT ECG DATA  
*SPIFW = 0 PROVIDES MCS FOR EACH FRAME, SCLK STAYS HIGH FOR 1/2 MSCLK CYCLE BETWEEN EACH WORD.  
Figure 5. Data Read and Write Timing Diagram for SPIFW = 0, Showing Entire Packet Of Data (Header, 5 ECG Word = [ECG1, ECG2, ECG3 and 2 Words with Zeros],  
and CRC Word)  
tMSCLK  
MSCLK  
tMCSSA  
tMSCLK  
tMCSHD  
MCS  
tMCSW  
SPIFW = 1*  
MSB  
LSB  
D0_0  
MSB  
D1_15  
LSB  
D5_0  
MSB  
LSB  
MSDO  
D0_15  
D0_14  
D0_1  
tMDO  
D1_14  
D6_15  
D6_14  
D6_0  
HEADER: 0xF AND 12-BIT COUNTER  
5 × 16-BIT ECG DATA  
16-BIT CRC WORD  
*SPIFW = 1 PROVIDES MCS FOR EACH FRAME, SCLK STAYS HIGH FOR 1 MSCLK CYCLE BETWEEN EACH WORD.  
Figure 6. Data Read and Write Timing Diagram for SPIFW = 1, Showing Entire Packet Of Data (Header, 5 ECG Word = [ECG1, ECG2, ECG3 and 2 Words with Zeros],  
and CRC Word)  
Rev. B | Page 13 of 80  
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
ABSOLUTE MAXIMUM RATINGS  
Table 7.  
THERMAL RESISTANCE  
θJA is specified for the worst-case conditions, that is, a device  
soldered in a circuit board for surface-mount packages.  
Parameter  
Rating  
AVDD to AGND  
−0.3 V to +6 V  
IOVDD to DGND  
ADCVDD to AGND  
DVDD to DGND  
REFIN/REFOUT to REFGND  
ECG and Analog Inputs to AGND  
Digital Inputs to DGND  
REFIN to ADCVDD  
AGND to DGND  
REFGND to AGND  
−0.3 V to +6 V  
Table 8. Thermal Resistance1  
Package Type  
56-Lead LFCSP  
64-Lead LQFP  
−0.3 V to +2.5 V  
−0.3 V to +2.5 V  
−0.3 V to +2.1 V  
−0.3 V to AVDD + 0.3 V  
−0.3 V to IOVDD + 0.3 V  
ADCVDD + 0.3 V  
−0.3 V to + 0.3 V  
−0.3 V to + 0.3 V  
10 ꢀA  
θJA  
35  
42.5  
Unit  
°C/W  
°C/W  
1 Based on JEDEC standard 4-layer (2S2P) high effective therꢀal conductivity  
test board (JESD51-7) and natural convection.  
ESD CAUTION  
ECG Input Continuous Current  
Storage Teꢀperature Range  
−65°C to +125°C  
Operating Junction Teꢀperature Range −40°C to +85°C  
Reflow Profile  
Junction Teꢀperature  
J-STD-20 (JEDEC)  
150°C ꢀax  
ESD  
HBM  
FICDM  
2500 V  
1000 V  
Stresses at or above those listed under Absolute Maximum  
Ratings may cause permanent damage to the product. This is a  
stress rating only; functional operation of the product at these  
or any other conditions above those indicated in the operational  
section of this specification is not implied. Operation beyond  
the maximum operating conditions for extended periods may  
affect product reliability.  
Rev. B | Page 14 of 80  
 
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
PIN CONFIGURATIONS AND FUNCTION DESCRIPTIONS  
64 63 62 61 60 59 58 57 56 55 54 53 52 51 50 49  
48  
47  
46  
45  
44  
43  
42  
41  
40  
39  
38  
37  
36  
35  
34  
33  
1
2
NC  
NC  
PIN 1  
AGND  
DGND  
IOVDD  
SDO  
3
NC  
NC  
4
1
2
AGND  
NC  
NC  
ECG3_RA  
ECG2_LL  
ECG1_LA  
REFIN  
REFOUT  
REFGND  
NC  
42  
41  
40  
PIN 1  
DGND  
IOVDD  
GPIO0/MCS  
5
SCLK  
NC  
NC  
INDICATOR  
3
6
SDI  
4
39 GPIO1/MSCLK  
7
DRDY  
REFGND  
REFOUT  
REFIN  
ADAS1000-3  
5
38  
37  
36  
35  
34  
33  
32  
31  
30  
29  
GPIO2/MSDO  
GPIO3  
DGND  
CS  
DRDY  
SDI  
SCLK  
SDO  
IOVDD  
DGND  
64-LEAD LQFP  
ADAS1000-3  
8
6
CS  
56-LEAD LFCSP  
7
9
TOP VIEW  
(Not to Scale)  
DGND  
GPIO3  
GPIO2/MSDO  
GPIO1/MSCLK  
GPIO0/MCS  
IOVDD  
DGND  
NC  
8
TOP VIEW  
(Not to Scale)  
10  
11  
ECG1_LA  
ECG2_LL  
9
10  
11  
12  
13  
14  
NC  
NC  
NC  
AGND  
ECG3_RA 12  
13  
14  
15  
16  
NC  
NC  
AGND  
NC  
17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32  
NOTES  
1. PINS LABELED NC CAN BE ALLOWED TO FLOAT, BUT IT IS BETTER  
TO CONNECT THESE PINS TO GROUND. AVOID ROUTING HIGH SPEED  
SIGNALS THROUGH THESE PINS BECAUSE NOISE COUPLING  
MAY RESULT.  
2. THE EXPOSED PAD IS ON THE TOP OF THE PACKAGE;  
IT IS CONNECTED TO THE MOST NEGATIVE POTENTIAL, AGND.  
NOTES  
1. PINS LABELED NC CAN BE ALLOWED TO FLOAT, BUT IT IS BETTER TO CONNECT  
THESE PINS TO GROUND. AVOID ROUTING HIGH SPEED SIGNALS THROUGH THESE  
PINS BECAUSE NOISE COUPLING MAY RESULT.  
Figure 7. ADAS1000-3, 64-Lead LQFP Pin Configuration  
Figure 8. ADAS1000-3, 56-Lead LFCSP Pin Configuration  
64 63 62 61 60 59 58 57 56 55 54 53 52 51 50 49  
48  
47  
46  
45  
44  
43  
42  
41  
40  
39  
38  
37  
36  
35  
34  
33  
1
2
NC  
NC  
PIN 1  
AGND  
DGND  
IOVDD  
SDO  
3
RESPDAC_RA  
EXT_RESP_RA  
EXT_RESP_LL  
EXT_RESP_LA  
REFGND  
1
2
42  
41  
40  
39  
38  
37  
36  
35  
34  
33  
32  
31  
30  
29  
AGND  
NC  
NC  
ECG3_RA  
ECG2_LL  
ECG1_LA  
REFIN  
DGND  
PIN 1  
INDICATOR  
IOVDD  
GPIO0/MCS  
GPIO1/MSCLK  
GPIO2/MSDO  
GPIO3  
DGND  
CS  
DRDY  
SDI  
SCLK  
SDO  
IOVDD  
DGND  
4
3
5
SCLK  
SDI  
4
6
5
ADAS1000-4  
6
7
DRDY  
CS  
ADAS1000-4  
56-LEAD LFCSP  
7
64-LEAD LQFP  
8
REFOUT  
8
REFOUT  
REFGND  
TOP VIEW  
(Not to Scale)  
9
9
REFIN  
DGND  
GPIO3  
GPIO2/MSDO  
TOP VIEW  
(Not to Scale)  
10  
11  
12  
13  
14  
EXT_RESP_LA  
EXT_RESP_LL  
EXT_RESP_RA  
RESPDAC_RA  
AGND  
10  
ECG1_LA  
ECG2_LL 11  
ECG3_RA  
NC  
12  
13  
14  
15  
16  
GPIO1/MSCLK  
____  
GPIO0/MCS  
IOVDD  
DGND  
NC  
AGND  
NC  
NC  
17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32  
NOTES  
1. PINS LABELED NC CAN BE ALLOWED TO FLOAT, BUT IT IS BETTER TO  
CONNECT THESE PINS TO GROUND. AVOID ROUTING HIGH SPEED  
SIGNALS THROUGH THESE PINS BECAUSE NOISE COUPLING  
MAY RESULT.  
2. THE EXPOSED PAD IS ON THE TOP OF THE PACKAGE;  
IT IS CONNECTED TO THE MOST NEGATIVE POTENTIAL, AGND.  
NOTES  
1. PINS LABELED NC CAN BE ALLOWED TO FLOAT, BUT IT IS BETTER TO  
CONNECT THESE PINS TO GROUND. AVOID ROUTING HIGH SPEED  
SIGNALS THROUGH THESE PINS BECAUSE NOISE COUPLING  
MAY RESULT.  
Figure 9. ADAS1000-4, 64-Lead LQFP Pin Configuration  
Figure 10. ADAS1000-4, 56-Lead LFCSP Pin Configuration  
Rev. B | Page 15 of 80  
 
 
 
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
Table 9. Pin Function Descriptions  
ADAS1000-3 Pin No.  
ADAS1000-4 Pin No.  
LQFP  
LFCSP  
LQFP  
LFCSP  
Mnemonic  
Description  
18, 23,  
58, 63  
15, 20,  
51, 56  
18, 23, 58,  
63  
15, 20, 51, AVDD  
56  
Analog Supply. See recommendations for bypass capacitors in the Power Supply,  
Grounding, and Decoupling Strategy section.  
35, 46  
30, 41  
35, 46  
30, 41  
IOVDD  
Digital Supply for Digital Input/Output Voltage Levels. See recommendations for  
bypass capacitors in the Power Supply, Grounding, and Decoupling Strategy section.  
26, 55  
23, 48  
26, 55  
23, 48  
ADCVDD  
Analog Supply for ADC. There is an on-chip linear regulator providing the supply  
voltage for the ADCs. These pins are primarily provided for decoupling purposes;  
however, the pin may also be supplied by an external 1.8 V supply should the user wish  
to use a more efficient supply to minimize power dissipation. In this case, use the  
VREG_EN pin tied to ground to disable the ADCVDD and DVDD regulators. The  
ADCVDD pin should not be used to supply other functions. See recommendations for  
bypass capacitors in the Power Supply, Grounding, and Decoupling Strategy section.  
30, 51  
27, 44  
30, 51  
27, 44  
DVDD  
Digital Supply. There is an on-chip linear regulator providing the supply voltage for the  
digital core. These pins are primarily provided for decoupling purposes; however, the  
pin may also be overdriven supplied by an external 1.8 V supply should the user wish to  
use a more efficient supply to minimize power dissipation. In this case, use the  
VREG_EN pin tied to ground to disable the ADCVDD and DVDD regulators. See  
recommendations for bypass capacitors in the Power Supply, Grounding, and  
Decoupling Strategy section.  
2, 15, 24,  
1, 14, 21,  
2, 15, 24,  
1, 14, 21,  
22, 49, 50  
AGND  
Analog Ground.  
25, 56, 57 22, 49, 50 25, 56, 57  
31, 34, 28, 29, 31, 34, 40,  
40, 47, 50 36, 42, 43 47, 50  
28, 29, 36, DGND  
42, 43  
Digital Ground.  
59  
19  
59  
19  
VREG_EN  
Enables or disables the internal voltage regulators used for ADCVDD and DVDD. Tie this  
pin to AVDD to enable or tie this pin to ground to disable the internal voltage  
regulators.  
10  
11  
12  
6
5
4
10  
11  
12  
4
6
ECG1_LA  
Analog Input, Left Arm (LA).  
5
ECG2_LL  
Analog Input, Left Leg (LL).  
4
ECG3_RA  
Analog Input, Right Arm (RA).  
Optional External Respiration Input.  
Optional External Respiration Input.  
Optional External Respiration Input.  
12  
11  
10  
16  
EXT_RESP_RA  
EXT_RESP_LL  
EXT_RESP_LA  
RESPDAC_LL  
5
6
62  
Optional Path for Higher Performance Respiration Resolution, Respiration DAC Drive,  
Negative Side 0.  
60  
18  
SHIELD/  
Shared Pin (User-Configured).  
RESPDAC_LA  
Output of Shield Driver (SHIELD).  
Optional Path for Higher Performance Respiration Resolution, Respiration DAC Drive,  
Negative Side 1 (RESPDAC_LA).  
60  
22  
18  
52  
SHIELD  
Output of Shield Driver.  
3
13  
52  
RESPDAC_RA  
Optional Path for Higher Performance Respiration Resolution, Respiration DAC Drive,  
Positive Side.  
22  
CM_OUT/WCT Common-Mode Output Voltage (Average of Selected Electrodes). Not intended to drive  
current.  
19  
21  
20  
61  
55  
53  
54  
17  
19  
21  
20  
61  
55  
53  
54  
17  
CM_IN  
Common-Mode Input.  
RLD_SJ  
Summing Junction for Right Leg Drive Amplifier.  
Output and Feedback Junction for Right Leg Drive Amplifier.  
RLD_OUT  
CAL_DAC_IO  
Calibration DAC Input/Output. Output for a master device, input for a slave. Not  
intended to drive current.  
9
7
9
7
REFIN  
Reference Input. For standalone mode, use REFOUT connected to REFIN. External 10 μF  
capacitors with ESR < 0.2 Ω in parallel with 0.1 μF bypass capacitors to GND are  
required and should be placed as close to the pin as possible. An external reference can  
be connected to REFIN.  
8
8
8
8
REFOUT  
Reference Output.  
7
9
7
9
REFGND  
Reference Ground. Connect to a clean ground.  
27, 28  
47, 46  
27, 28  
47, 46  
XTAL1, XTAL2  
External crystal connects between these two pins; external clock drive should be  
applied to CLK_IO. Each XTAL pin requires a 15 pF capacitor to ground.  
29  
41  
45  
35  
29  
41  
45  
35  
CLK_IO  
CS  
Buffered Clock Input/Output. Output for a master device; input for a slave. Powers up in  
high impedance.  
CS  
Chip Select and Frame Sync, Active Low.  
the entire suite of data in framing mode.  
can be used to frame each word or to frame  
Rev. B | Page 16 of 80  
Data Sheet  
ADAS1000-3/ADAS1000-4  
ADAS1000-3 Pin No.  
ADAS1000-4 Pin No.  
LQFP  
LFCSP  
LQFP  
LFCSP  
Mnemonic  
Description  
44  
32  
44  
32  
SCLK  
Clock Input. Data is clocked into the shift register on a rising edge and clocked out on a  
falling edge.  
43  
53  
45  
33  
25  
31  
43  
53  
45  
33  
25  
31  
SDI  
PD  
Serial Data Input.  
Power-Down, Active Low.  
SDO  
Serial Data Output. This pin is used for reading back register configuration data and for  
the data frames.  
42  
34  
42  
34  
DRDY  
Digital Output. This pin indicates that conversion data is ready to be read back when  
low, busy when high. When reading packet data, the entire packet must be read to  
DRDY  
allow  
to return high.  
54  
52  
24  
26  
54  
52  
24  
26  
RESET  
Digital Input. This pin has an internal pull-up resistor. This pin resets all internal nodes to  
their power-on reset values.  
SYNC_GANG  
Digital Input/Output (Output on Master, Input on Slave). Used for synchronization  
control where multiple devices are connected together. Powers up in high impedance.  
36  
37  
38  
39  
40  
39  
38  
37  
36  
37  
38  
39  
40  
39  
38  
37  
2, 3  
MCS  
CS  
.
GPIO0/  
General-Purpose I/O or Master 128 kHz SPI  
GPIO1/MSCLK General-Purpose I/O or Master 128 kHz SPI SCLK.  
GPIO2/MSDO  
GPIO3  
General-Purpose I/O or Master 128 kHz SPI SDO.  
General-Purpose I/O.  
1, 3, 4, 5,  
6, 13, 14,  
16, 17,  
32, 33,  
48, 49,  
62, 64  
2, 3, 10,  
11, 12,  
13, 16  
1, 13, 14,  
16, 17, 32,  
33, 48, 49,  
64  
NC  
No connect. Do not connect to these pins (see Figure 7, Figure 8, Figure 9, and  
Figure 10).  
57  
57  
EPAD  
Exposed Pad. The exposed pad is on the top of the package; it is connected to the most  
negative potential, AGND.  
Rev. B | Page 17 of 80  
ADAS1000-3/ADAS1000-4  
Data Sheet  
TYPICAL PERFORMANCE CHARACTERISTICS  
8
15  
10  
5
0.5Hz TO 40Hz  
GAIN SETTING 0 = 1.4  
DATA RATE = 2kHz  
6
4
10 SECONDS OF DATA  
2
0
0
–5  
–10  
–15  
–2  
–4  
–6  
0.5Hz TO 150Hz  
GAIN SETTING 3 = 4.2  
DATA RATE = 2kHz  
10 SECONDS OF DATA  
0
1
2
3
4
5
6
7
8
9
10  
0
1
2
3
4
5
6
7
8
9
10  
TIME (Seconds)  
TIME (Seconds)  
Figure 11. Input Referred Noise for 0.5 Hz to 40 Hz Bandwidth, 2 kHz Data  
Rate, GAIN 0 (1.4)  
Figure 14. Input Referred Noise for 0.5 Hz to 150 Hz Bandwidth, 2 kHz Data  
Rate, GAIN 3 (4.2)  
8
25  
0.5Hz TO 40Hz  
LA 150Hz  
LA 40Hz  
GAIN SETTING 3 = 4.2  
DATA RATE = 2kHz  
6
10 SECONDS OF DATA  
20  
15  
10  
5
4
2
0
–2  
–4  
–6  
0
0
1
2
3
4
5
6
7
8
9
10  
GAIN 0  
GAIN 1  
GAIN 2  
GAIN 3  
TIME (Seconds)  
GAIN SETTING  
Figure 12. Input Referred Noise for 0.5 Hz to 40 Hz Bandwidth, 2 kHz Data  
Rate, GAIN 3 (4.2)  
Figure 15. ECG Channel Noise Performance over a 0.5 Hz to 40 Hz or 0.5 Hz to  
150 Hz Bandwidth vs. Gain Setting  
15  
10  
5
0.020  
AVDD = 3.3V  
GAIN SETTING 0 = 1.4  
0.018  
0.016  
0.014  
0.012  
0.010  
0
–5  
–10  
–15  
0.5Hz TO 150Hz  
GAIN SETTING 0 = 1.4  
DATA RATE = 2kHz  
10 SECONDS OF DATA  
0
1
2
3
4
5
6
7
8
9
10  
LA  
LL  
RA  
V1  
V2  
ELECTRODE INPUT  
TIME (Seconds)  
Figure 13. Input Referred Noise for 0.5 Hz to 150 Hz Bandwidth, 2 kHz Data  
Rate, GAIN 0 (1.4)  
Figure 16. Typical Gain Error Across Channels  
Rev. B | Page 18 of 80  
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
0.121  
0.215  
0.210  
0.205  
0.200  
0.195  
0.190  
0.185  
0.180  
AVDD = 3.3V  
AVDD = 3.3V  
0.101  
0.081  
0.061  
0.041  
0.021  
0.001  
ECG DC LEAD-OFF THRESHOLD  
RLD DC LEAD-OFF THRESHOLD  
GAIN 0  
GAIN 1  
GAIN 2  
GAIN 3  
–40  
–20  
0
20  
40  
60  
80  
GAIN SETTING  
TEMPERATURE (°C)  
Figure 17. Typical Gain Error vs. Gain  
Figure 20. DC Lead-Off Comparator Low Threshold vs. Temperature  
2.420  
0.15  
0.10  
AVDD = 3.3V  
AVDD = 3.3V  
GAIN SETTING 0 = 1.4  
GAIN SETTING 1 = 2.1  
GAIN SETTING 2 = 2.8  
GAIN SETTING 3 = 4.2  
2.415  
2.410  
2.405  
2.400  
2.395  
2.390  
2.385  
0.05  
0
–0.05  
–0.10  
–0.15  
–0.20  
–0.25  
–0.30  
–0.35  
GAIN ERROR G0  
GAIN ERROR G1  
GAIN ERROR G2  
GAIN ERROR G3  
2.380  
ECG DC LEAD-OFF THRESHOLD  
RLD DC LEAD-OFF THRESHOLD  
2.375  
–40  
–20  
0
20  
40  
60  
80  
–40  
–20  
0
20  
40  
60  
80  
TEMPERATURE (°C)  
TEMPERATURE (°C)  
Figure 18. Typical Gain Error for All Gain Settings Across Temperature  
Figure 21. DC Lead-Off Comparator High Threshold vs. Temperature  
5
0
AVDD = 3.3V  
+85°C  
+55°C  
+25°C  
–5°C  
AVDD = 3.3V  
GAIN SETTING 0 = 1.4  
4
3
–1  
–2  
–3  
–4  
–5  
–6  
–7  
–8  
–9  
–10  
–40°C  
2
1
0
–1  
–2  
–3  
–4  
–5  
0.3  
0.8  
1.3  
1.8  
2.3  
1
10  
100  
FREQUENCY (Hz)  
1k  
VOLTAGE (V)  
Figure 19. Typical ECG Channel Leakage Current over Input Voltage Range  
vs. Temperature  
Figure 22. Filter Response with 40 Hz Filter Enabled, 2 kHz Data Rate; See  
Figure 72 for Digital Filter Overview  
Rev. B | Page 19 of 80  
ADAS1000-3/ADAS1000-4  
Data Sheet  
0
–1  
–2  
–3  
–4  
–5  
–6  
0
–1  
–2  
–3  
–4  
–5  
–6  
–7  
–8  
–9  
–10  
AVDD = 3.3V  
AVDD = 3.3V  
10  
1
100  
1k  
10k  
100k  
1
10  
100  
1k  
FREQUENCY (Hz)  
FREQUENCY (Hz)  
Figure 23. Filter Response with 150 Hz Filter Enabled, 2 kHz Data Rate; See  
Figure 72 for Digital Filter Overview  
Figure 26. Analog Channel Bandwidth  
0
–1  
–2  
–3  
–4  
–5  
–6  
–7  
–8  
–9  
0
–1  
–2  
–3  
–4  
–5  
–6  
–7  
–8  
–9  
–10  
AVDD = 3.3V  
1
10  
100  
1k  
10k  
100k  
1
10  
100  
1k  
FREQUENCY (Hz)  
FREQUENCY (Hz)  
Figure 24. Filter Response with 250 Hz Filter Enabled, 2 kHz Data Rate; See  
Figure 72 for Digital Filter Overview  
Figure 27. Filter Response Running at 128 kHz Data Rate; See Figure 72 for  
Digital Filter Overview  
0
1.8010  
1.8005  
1.8000  
1.7995  
1.7990  
1.7985  
1.7980  
1.7975  
1.7970  
1.7965  
AVDD = 3.3V  
–1  
–2  
–3  
–4  
–5  
–6  
–7  
–8  
–9  
–10  
1
10  
100  
1k  
–40  
–20  
0
20  
40  
60  
80  
FREQUENCY (Hz)  
TEMPERATURE (°C)  
Figure 25. Filter Response with 450 Hz Filter Enabled, 2 kHz Data Rate; See  
Figure 72 for Digital Filter Overview  
Figure 28. Typical Internal VREF vs. Temperature  
Rev. B | Page 20 of 80  
Data Sheet  
ADAS1000-3/ADAS1000-4  
805  
800  
795  
790  
785  
780  
775  
770  
765  
1.3010  
AVDD = 3.3V  
AVDD = 3.3V  
1.3005  
1.3000  
1.2995  
1.2990  
1.2985  
1.2980  
1.2975  
1.2970  
–40  
–20  
0
20  
40  
60  
80  
–40  
–20  
0
40  
TEMPERATURE (°C)  
60  
80  
20  
TEMPERATURE (°C)  
Figure 29. VCM_REF vs. Temperature  
Figure 32. Typical AVDD Supply Current vs. Temperature in Standby Mode  
12.65  
12.50  
12.45  
12.40  
12.35  
12.30  
12.25  
12.20  
LOW NOISE/HIGH  
AVDD = 3.3V  
PERFORMANCE MODE  
3 ECG CHANNELS ENABLED  
INTERNAL LDO UTILIZED  
HIGH PERFORMANCE/LOW NOISE MODE  
12.60  
12.55  
12.50  
12.45  
12.40  
12.35  
3.0  
3.5  
4.0  
4.5  
5.0  
5.5  
6.0  
–40  
–20  
0
20  
40  
60  
80  
VOLTAGE (V)  
TEMPERATURE (°C)  
Figure 30. Typical AVDD Supply Current vs. Temperature, Using Internal  
ADVCDD/DVDD Supplies  
Figure 33. Typical AVDD Supply Current vs. AVDD Supply Voltage  
3.430  
0.142955  
AVDD = 3.3V  
AVDD = 3.3V  
ECG PATH/DEFIB/CABLE IMPEDANCE = 0 Ω  
PATIENT IMPEDANCE = 1kΩ  
3 ECG CHANNELS ENABLED  
ADCVDD AND DVDD SUPPLIED EXTERNALLY  
HIGH PERFORMANCE/LOW NOISE MODE  
RESPIRATION RATE = 10RESPPM  
RESPAMP = 11 = 60µA p-p  
RESPGAIN = 0011 = 4  
3.425  
0.142950  
0.142945  
0.142940  
0.142935  
0.142930  
0.142925  
3.420  
3.415  
3.410  
3.405  
3.400  
3.395  
–40  
–20  
0
20  
40  
60  
80  
0
5
10  
15  
20  
25  
30  
TEMPERATURE (°C)  
TIME (Seconds)  
Figure 31. Typical AVDD Supply Current vs. Temperature, Using Externally  
Supplied ADVCDD/DVDD  
Figure 34. Respiration with 200 mΩ Impedance Variation, Using Internal  
Respiration Paths and Measured with a 0 Ω Patient Cable  
Rev. B | Page 21 of 80  
ADAS1000-3/ADAS1000-4  
Data Sheet  
0.121145  
0.517390  
AVDD = 3.3V  
AVDD = 3.3V  
ECG PATH/DEFIB/CABLE IMPEDANCE = 5k /250pF  
PATIENTIMPEDANCE = 1kΩ  
EXTCAP= 100pF  
ECG PATH/DEFIB/CABLE IMPEDANCE = 0Ω  
PATIENT IMPEDANCE = 1kΩ  
RESPIRATION RATE = 10RESPPM  
RESPAMP = 11 = 60µA p-p  
RESPGAIN = 0011 = 4  
0.121140  
0.517385 RESPIRATION RATE = 10RESPPM  
0.121135  
0.121130  
0.121125  
0.121120  
0.121115  
0.517380  
0.517375  
0.517370  
0.517365  
RESPAMP = 11 = 60µA p-p  
RESPGAIN = 0011 = 4  
0.517360  
0
5
10  
15  
20  
25  
30  
0
5
10  
15  
20  
25  
30  
TIME (Seconds)  
TIME (Seconds)  
Figure 35. Respiration with 100 mΩ Impedance Variation, Using Internal  
Respiration Paths and Measured with a 0 Ω Patient Cable  
Figure 38. Respiration with 200 mΩ Impedance Variation, Using External  
Respiration DAC Driving a 100 pF External Capacitor and Measured with a  
5 kΩ Patient Cable  
0.663160  
0.159775  
AVDD = 3.3V  
AVDD = 3.3V  
ECG PATH/DEFIB/CABLE IMPEDANCE = 5k Ω  
PATIENT IMPEDANCE = 1kΩ  
ECG PATH/DEFIB/CABLE IMPEDANCE = 1.5kΩ/600pF  
PATIENT IMPEDANCE = 1kΩ  
EXTCAP = 1nF  
RESPIRATION RATE = 10RESPPM  
RESPAMP = 11 = 60µA p-p  
RESPGAIN = 0011 = 4  
0.663155  
0.663150  
0.663145  
0.663140  
0.663135  
0.663130  
0.159770  
0.159765  
0.159760  
0.159755  
0.159750  
0.159745  
RESPIRATION RATE = 10RESPPM  
RESPAMP = 11  
RESPGAIN = 0001 = 1  
0
5
10  
15  
20  
25  
30  
0
5
10  
15  
20  
25  
30  
TIME (Seconds)  
TIME (Seconds)  
Figure 36. Respiration with 200 mΩ Impedance Variation, Using Internal  
Respiration Paths and Measured with a 5 kΩ Patient Cable  
Figure 39. Respiration with 200 mΩ Impedance Variation, Using External  
Respiration DAC Driving a 1 nF External Capacitor and Measured with a  
1.5 kΩ Patient Cable  
0.062365  
0.062360  
0.062355  
0.062350  
0.062345  
0.159126  
AVDD = 3.3V  
ECG PATH/DEFIB/CABLE IMPEDANCE = 1.5kΩ/600pF  
PATIENT IMPEDANCE = 1kΩ  
0.159125  
0.159124  
0.159123  
0.159122  
0.159121  
0.159120  
AVDD = 3.3V  
ECG PATH/DEFIB/CABLE IMPEDANCE = 0Ω  
PATIENT IMPEDANCE = 1kΩ  
0.062340  
EXTCAP = 100pF  
RESPIRATION RATE = 10RESPPM  
RESPAMP = 11 = 60µA p-p  
RESPGAIN = 0011 = 4  
EXTCAP = 1nF  
0.159119  
RESPIRATION RATE = 10RESPPM  
RESPAMP = 11  
RESPGAIN = 0001 = 1  
0.062335  
0.159118  
0
5
10  
15  
20  
25  
30  
0
5
10  
15  
20  
25  
30  
TIME (Seconds)  
TIME (Seconds)  
Figure 37. Respiration with 200 mΩ Impedance Variation, Using External  
Respiration DAC Driving a 100 pF External Capacitor and Measured with a  
0 Ω Patient Cable  
Figure 40. Respiration with 100 mΩ Impedance Variation, Using an External  
Respiration DAC Driving a 1 nF External Capacitor and Measured with a  
1.5 kΩ Patient Cable  
Rev. B | Page 22 of 80  
Data Sheet  
ADAS1000-3/ADAS1000-4  
50  
150  
100  
50  
LA  
LL  
RA  
LA  
LL  
RA  
AVDD = 3.3V  
AVDD = 3.3V  
40  
30  
20  
10  
0
0
–10  
–20  
–30  
–40  
–50  
–50  
–100  
–150  
0.3  
0.5  
0.7  
0.9  
1.1  
1.3  
1.5  
1.7  
1.9  
2.1  
2.3  
0.3  
0.5  
0.7  
0.9  
1.1  
1.3  
1.5  
1.7  
1.9  
2.1  
2.3  
INPUT VOLTAGE (V)  
INPUT VOLTAGE (V)  
Figure 41. DNL Error vs. Input Voltage Range Across Electrodes at 25°C  
Figure 44. INL vs. Input Voltage Across Electrode Channel for 2 kHz Data Rate  
150  
50  
GAIN 0  
GAIN 1  
GAIN 2  
GAIN 3  
AVDD = 3.3V  
–40°C  
–5°C  
+25°C  
+55°C  
+85°C  
AVDD = 3.3V  
40  
30  
100  
50  
20  
10  
0
0
–10  
–20  
–30  
–40  
–50  
–50  
–100  
0.3  
0.5  
0.7  
0.9  
1.1  
1.3  
1.5  
1.7  
1.9  
2.1  
2.3  
0.3  
0.5  
0.7  
0.9  
1.1  
1.3  
1.5  
1.7  
1.9  
2.1  
2.3  
INPUT VOLTAGE (V)  
INPUT VOLTAGE (V)  
Figure 42. DNL Error vs. Input Voltage Range Across Temperature  
Figure 45. INL vs. Input Voltage Across Gain Setting for 16 kHz Data Rate  
150  
150  
GAIN 0  
GAIN 1  
GAIN 2  
GAIN 3  
AVDD = 3.3V  
AVDD = 3.3V  
GAIN 0  
GAIN 1  
GAIN 2  
GAIN 3  
100  
50  
100  
50  
0
0
–50  
–100  
–150  
–50  
–100  
–150  
0.3  
0.5  
0.7  
0.9  
1.1  
1.3  
1.5  
1.7  
1.9  
2.1  
2.3  
0.3  
0.5  
0.7  
0.9  
1.1  
1.3  
1.5  
1.7  
1.9  
2.1  
2.3  
INPUT VOLTAGE (V)  
INPUT VOLTAGE (V)  
Figure 43. INL vs. Input Voltage Across Gain Setting for 2 kHz Data Rate  
Figure 46. INL vs. Input Voltage Across Gain Setting for 128 kHz Data Rate  
Rev. B | Page 23 of 80  
ADAS1000-3/ADAS1000-4  
Data Sheet  
0
120  
100  
80  
AVDD = 3.3V  
GAIN 0  
–20  
–40  
DATA RATE = 2kHz  
FILTER SETTING = 150Hz  
60  
–60  
40  
–80  
20  
–100  
–120  
–140  
–160  
–180  
0
–20  
–40  
–60  
–80  
0
50  
100 150 200 250 300 350 400 450 500  
FREQUENCY (Hz)  
100m  
1
10  
100  
1k  
10k 100k 1M 10M 100M 1G  
FREQUENCY (Hz)  
Figure 47. FFT with 60 Hz Input Signal  
Figure 50. Open-Loop Gain Response of Right Leg Drive Amplifier Without  
Loading  
150  
0
AVDD = 3.3V  
–0.5dBFS  
10Hz INPUT SIGNAL  
–50  
–100  
–150  
–200  
–250  
–300  
–350  
100  
50  
SNR  
0
–50  
–100  
THD  
GAIN 0  
GAIN 1  
GAIN 2  
GAIN 3  
100m  
1
10  
100  
1k  
10k 100k 1M 10M 100M 1G  
GAIN SETTING  
FREQUENCY (Hz)  
Figure 48. SNR and THD Across Gain Settings  
Figure 51. Open-Loop Phase Response of Right Leg Drive Amplifier Without  
Loading  
AVDD = 3.3V  
DRDY  
AVDD  
2
1
CH1 2.00V  
M1.00ms  
22.1%  
A CH1  
2.48V  
CH2 1.00V  
T
DRDY  
Figure 49. Power Up AVDD Line to  
Going Low (Ready)  
Rev. B | Page 24 of 80  
Data Sheet  
ADAS1000-3/ADAS1000-4  
APPLICATIONS INFORMATION  
monitor and diagnostic applications. Value-added cardiac post  
processing may be executed externally on a DSP, microproces-  
sor, or FPGA. The ADAS1000-3/ADAS1000-4 are designed for  
operation in both low power, portable telemetry applications  
and line powered systems; therefore, the parts offer power/noise  
scaling to ensure suitability to these varying requirements.  
OVERVIEW  
The ADAS1000-3/ADAS1000-4 are electro cardiac (ECG)  
front-end solutions targeted at a variety of medical applica-  
tions. In addition to ECG measurements, the ADAS1000-3/  
ADAS1000-4 also measure thoracic impedance (respiration)  
and detect pacing artifacts, providing all the measured  
information to the host controller in the form of a data frame  
supplying either lead/vector or electrode data at programmable  
data rates. The ADAS1000-3/ADAS1000-4 are designed to  
simplify the task of acquiring ECG signals for use in both  
The devices also offer a suite of dc and ac test excitation via  
a calibration DAC feature and CRC redundancy checks in  
addition to readback of all relevant register address space.  
REFIN REFOUT CAL_DAC_IO RLD_SJ  
RLD_OUT CM_IN  
CM_OUT/WCT  
SHIELD  
AVDD  
IOVDD  
DRIVEN  
LEAD  
AMP  
VREF  
SHIELD  
DRIVE  
AMP  
ADCVDD  
DVDD  
ADCVDD, DVDD  
1.8V  
REGULATORS  
CALIBRATION  
DAC  
VCM_REF  
(1.3V)  
ADAS1000-3  
COMMON-  
MODE AMP  
VCM  
AC  
LEAD-OFF  
DAC  
VREF  
AC  
LEAD-OFF  
DETECTION  
ECG PATH  
CS  
ECG1_LA  
ECG2_LL  
ECG3_RA  
SCLK  
SDI  
AMP  
AMP  
AMP  
ADC  
SDO  
DC LEAD-  
OFF/MUXES  
ADC  
ADC  
DRDY  
PD  
FILTERS,  
CONTROL,  
AND  
INTERFACE  
LOGIC  
RESET  
SYNC_GANG  
GPIO0/MCS  
GPIO1/MSCLK  
GPIO2/MSDO  
GPIO3  
CLOCK GEN/OSC/  
EXTERNAL CLK  
SOURCE  
CLK_IO  
REFGND  
AGND  
DGND  
XTAL1 XTAL2  
Figure 52. ADAS1000-3 Simplified Block Diagram  
Rev. B | Page 25 of 80  
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
REFIN REFOUT CAL_DAC_IO RLD_SJ  
RLD_OUT CM_IN  
CM_OUT/WCT  
SHIELD  
AVDD  
IOVDD  
DRIVEN  
LEAD  
SHIELD  
DRIVE  
AMP  
VREF  
AMP  
ADCVDD  
DVDD  
ADCVDD, DVDD  
1.8V  
REGULATORS  
CALIBRATION  
DAC  
+
+
VCM_REF  
(1.3V)  
ADAS1000-4  
RESPIRATION  
DAC  
COMMON-  
MODE AMP  
VCM  
AC  
AC  
LEAD-OFF  
DAC  
LEAD-OFF  
DETECTION  
VREF  
PACE  
DETECTION  
DC LEAD-  
OFF/MUXES  
ECG PATH  
AMP  
CS  
ECG1_LA  
ECG2_LL  
ECG3_RA  
ADC  
ADC  
ADC  
SCLK  
SDI  
SDO  
AMP  
AMP  
FILTERS,  
CONTROL,  
AND  
INTERFACE  
LOGIC  
DRDY  
PD  
RESET  
SYNC_GANG  
EXT_RESP_LA  
EXT_RESP_LL  
GPIO0/MCS  
GPIO1/MSCLK  
AMP  
ADC  
GPIO2/MSDO  
GPIO3  
EXT_RESP_RA  
RESPDAC_LA  
MUX  
RESPIRATION PATH  
RESPDAC_LL  
RESPDAC_RA  
CLOCK GEN/OSC/  
EXTERNAL CLK  
SOURCE  
CLK_IO  
REFGND  
AGND  
DGND  
XTAL1 XTAL2  
Figure 53. ADAS1000-4 Simplified Block Diagram  
Rev. B | Page 26 of 80  
Data Sheet  
ADAS1000-3/ADAS1000-4  
can be achieved using one ADAS1000-3 or ADAS1000-4  
device ganged together with one ADAS1000-2 slave device  
as described in the Gang Mode Operation section. Similarly,  
a 12-lead (10-electrode) system can be achieved using one  
ADAS1000 or ADAS1000-1 device ganged together with one  
ADAS1000-2 slave device as described in the Gang Mode  
Operation section. Here, nine ECG electrodes and one RLD  
electrode achieve the 10 electrode system, again leaving one  
spare ECG channel that could be used for alternate purposes  
as suggested previously. In such a system, having nine dedi-  
cated electrodes benefits the user by delivering lead information  
based on electrode measurements and calculations rather than  
deriving leads from other lead measurements.  
ECG INPUTS—ELECTRODES/LEADS  
The ADAS1000-3/ADAS1000-4 ECG product consists of three  
ECG inputs and a reference drive, RLD (right leg drive). In  
a typical 3-lead/vector application, three of the ECG inputs  
(ECG3_RA, ECG1_LA, ECG2_LL) are used in addition to the  
RLD path.  
In a 3-lead system, the ADAS1000-3/ADAS1000-4 can be  
arranged to provide Lead I, Lead II, and Lead III data or  
electrode data directly via the serial interface at all frame  
rates. Note that in 128 kHz data rate, lead data is only available  
when configured in analog lead mode. Digital lead mode is not  
available for this data rate.  
Should the user have a need for increased electrode counts, then  
there are other products within the ADAS1000 family that may  
be suitable. For example, a derived 12-lead (8-electrode) system  
Table 10 outlines the calculation of the leads (vector) from the  
individual electrode measurements when using either the  
ADAS1000-3 or ADAS1000-4.  
Table 10. Lead Composition  
Device  
Lead Name  
Composition  
LA – RA  
Equivalent  
ADAS1000-3 or ADAS1000-4  
I
II  
LL – RA  
III  
LL – LA  
aVR1  
aVL1  
aVF1  
RA – 0.5 × (LA + LL)  
LA – 0.5 × (LL + RA)  
LL – 0.5 × (LA + RA)  
−0.5 × (I + II)  
0.5 × (I − III)  
0.5 × (II + III)  
1 These augmented leads are not calculated within the ADAS1000-3/ADAS1000-4, but can be derived in the host DSP/microcontroller/FPGA.  
Rev. B | Page 27 of 80  
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
The ADAS1000-3/ADAS1000-4 implementation uses a dc-  
ECG CHANNEL  
coupled approach, which requires that the front end be biased  
to operate within the limited dynamic range imposed by the  
relatively low supply voltage. The right leg drive loop performs  
this function by forcing the electrical average of all selected  
electrodes to the internal 1.3 V level, VCM_REF, maximizing  
each channel’s available signal range.  
The ECG channel consists of a programmable gain, low noise,  
differential preamplifier; a fixed gain anti-aliasing filter; buffers;  
and an ADC (see Figure 54). Each electrode input is routed to  
its PGA noninverting input. Internal switches allow the PGA’s  
inverting inputs to be connected to other electrodes and/or  
the Wilson Central Terminal to provide differential analog  
processing (analog lead mode), to a computed average of some  
or all electrodes, or to the internal 1.3 V common-mode  
reference (VCM_REF). The latter two modes support digital  
lead mode (leads computed on-chip) and electrode mode (leads  
calculated off-chip). In all cases, the internal reference level is  
removed from the final lead data.  
All ECG channel amplifiers use chopping to minimize 1/f noise  
contributions in the ECG band. The chopping frequency of  
~250 kHz is well above the bandwidth of any signals of interest.  
The 2-pole anti-aliasing filter has ~65 kHz bandwidth to support  
digital pace detection while still providing greater than 80 dB  
of attenuation at the ADCs sample rate. The ADC is a 14-bit,  
2 MHz SAR converter; 1024 × oversampling helps achieve the  
required system performance. The ADCs full-scale input range  
is 2 × VREF, or 3.6 V, although the analog portion of the ECG  
channel limits the useful signal swing to about 2.8 V. The  
ADAS1000 contains flags to indicate whether the ADC data is  
out of range, indicating a hard electrode off state. Programmable  
overrange and underrange thresholds are shown in the LOFFUTH  
and LOFFLTH registers (see Table 39 and Table 40, respectively).  
The ADC out of range flag is contained in the header word (see  
Table 53).  
TO COMMON-MODE AMPLIFIER  
FOR DRIVEN LEG AND  
SHIELD DRIVER  
AVDD  
PREAMP  
fS  
VREF  
G = 1, 1.5, 2, 3  
PATIENT  
DIFF AMP  
BUFFER  
G = 1.4  
CABLE  
EXTERNAL  
FILTER  
+
ELECTRODE  
ELECTRODE  
RFI AND DEFIB  
PROTECTION  
ADC  
14  
ELECTRODE  
EXTERNAL  
RFI AND DEFIB  
PROTECTION  
VCM  
ADAS1000-3/  
ADAS1000-4  
SHIELD DRIVER  
Figure 54. Simplified Schematic of a Single ECG Channel  
Rev. B | Page 28 of 80  
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
Digital Lead Mode and Calculation  
ELECTRODE/LEAD FORMATION AND INPUT  
STAGE CONFIGURATION  
When the ADAS1000-3/ADAS1000-4 are configured for digital  
lead mode (see the FRMCTL register, 0x0A[4], Table 37), the  
digital core will calculate each lead from the electrode signals.  
This is straightforward for Lead I/ Lead II/Lead III. Calculating  
V1’ and V2’ requires WCT, which is also computed internally  
for this purpose. This mode ignores the common-mode  
configuration specified in the CMREFCTL register  
The input stage of the ADAS1000-3/ADAS1000-4 can be  
arranged in several different manners. The input amplifiers  
are differential amplifiers and can be configured to generate  
the leads in the analog domain, before the ADCs. In addition  
to this, the digital data can be configured to provide either  
electrode or lead format under user control as described in  
Table 37. This allows maximum flexibility of the input stage  
for a variety of applications.  
(Register 0x05). Digital lead calculation is only available in  
2 kHz and 16 kHz data rates (see Figure 57).  
Electrode Mode: Single-Ended Input Electrode  
Configuration  
Analog Lead Mode and Calculation  
Leads are configured in the analog input stage when CHCONFIG  
= 1, as shown in Figure 56. This uses a traditional in-amp  
structure where lead formation is performed prior to digitiza-  
tion, with WCT created using the common-mode amplifier.  
While this results in the inversion of Lead II in the analog  
domain, this is digitally corrected so output data have the  
proper polarity.  
In this mode, the electrode data are digitized relative to the  
common-mode signal, VCM, which can be arranged to be any  
combination of the contributing ECG electrodes. Common-  
mode generation is controlled by the CMREFCTL register  
as described in Table 32 (see Figure 59).  
Electrode Mode: Common Electrode A and Common  
Electrode B Configurations  
In this mode, all electrodes are digitized relative to a common  
electrode (CE), for example, RA. Standard leads must be calculated  
by post processing the output data of the ADAS1000/  
ADAS1000-1/ADAS1000-2 (see Figure 58 and Figure 60).  
0x0A 0x01 0x05  
WORD3  
MODE  
COMMENT  
WORD1  
WORD2  
[4]1 [10]2  
[8]3  
0
LEAD I  
(LA RA)  
LEAD II  
(LL RA)  
LEAD III  
(LL LA)  
ANALOG LEAD  
ANALOG LEAD  
0
0
1
0
DIGITAL LEAD  
SINGLE-ENDED  
INPUT, DIGITALLY  
CALCULATED LEADS  
LEAD I  
LEAD II  
LEAD III  
0
1
(LA RA)  
(LL RA)  
(LL LA)  
COMMON  
COMMON ELECTRODE (CE)  
LEAD I  
LEAD II  
V1’  
0
0
ELECTRODE A  
LEADS (HERE RA ELECTRODE IS  
CONNECTED TO THE CE ELECTRODE  
(CM_IN) AND V1 IS ON ECG3 INPUT)  
(LA RA)  
(LL RA)  
(V1 – RA) (LA RA) (LL RA)  
3
SINGLE-ENDED  
INPUT  
SINGLE-ENDED  
INPUT ELECTRODE  
RELATIVE TO VCM  
LA VCM  
LA CE  
LL VCM RA VCM  
1
1
0
0
0
1
ELECTRODE  
COMMON  
ELECTRODE B  
LEADS FORMED  
RELATIVE TO A  
COMMON  
LL CE  
V1 CE  
ELECTRODE (CE)  
1REGISTER FRMCTL, BIT DATAFMT: 0 = LEAD/VECTOR MODE; 1 = ELECTRODE MODE.  
2REGISTER ECGCTL, BIT CHCONFIG: 0 = SINGLE ENDED INPUT (DIGITAL LEAD MODE OR ELECTRODE MODE); 1 = DIFFERENTIAL INPUT (ANALOG LEAD MODE).  
3REGISTER CMREFCTL, BIT CEREFEN: 0 = CE DISABLED; 1 = CE ENABLED.  
Figure 55. Electrode and Lead Configurations  
Rev. B | Page 29 of 80  
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
VCM = WCT = (LA + LL + RA)/3  
CM_OUT/WCT  
COMMON-  
MODE AMP  
LEAD I  
ECG1_LA  
ECG2_LL  
ECG3_RA  
+
(LA – RA)  
ADC  
ADC  
AMP  
LEAD III  
(LL – LA)  
+
AMP  
LEAD II  
(LL – RA)*  
+
AMP  
ADC  
*GETS MULITPLED  
BY –1 IN DIGITAL  
CM_IN  
FOR EXAMPLE RA  
COMMON  
ELECTRODE CE IN  
0x0A 0x01 0x05  
MODE  
COMMENT  
WORD1  
WORD2  
WORD3  
[4]1 [10]2  
[8]3  
LEAD I  
(LA − RA)  
LEAD II  
(LL − RA)  
LEAD III  
(LL − LA)  
ANALOG LEAD  
ANALOG LEAD  
0
1
0
1REGISTER FRMCTL, BIT DATAFMT: 0 = LEAD/VECTOR MODE; 1 = ELECTRODE MODE.  
2REGISTER ECGCTL, BIT CHCONFIG: 0 = SINGLE ENDED INPUT (DIGITAL LEAD MODE OR ELECTRODE MODE); 1 = DIFFERENTIAL INPUT (ANALOG LEAD MODE).  
3REGISTER CMREFCTL, BIT CEREFEN: 0 = CE DISABLED; 1 = CE ENABLED.  
Figure 56. Electrode and Lead Configurations, Analog Lead Mode  
VCM = WCT = (LA + LL + RA)/3  
CM_OUT/WCT  
COMMON-  
MODE AMP  
LEAD I  
ECG1_LA  
ECG 2_LL  
ECG3_RA  
+
LA – RA  
ADC  
ADC  
AMP  
LEAD II  
LL – RA  
+
AMP  
DIGITAL DOMAIN  
CALCULATIONS  
LEAD III  
LL – LA  
+
AMP  
ADC  
CM_IN  
FOR EXAMPLE, RA  
COMMON  
ELECTRODE CE IN  
0x0A 0x01 0x05  
MODE  
COMMENT  
WORD1  
WORD2  
WORD3  
[4]1 [10]2  
[8]3  
DIGITAL LEAD  
SINGLE-ENDED  
INPUT, DIGITALLY  
CALCULATED LEADS  
LEAD I  
(LA − RA)  
LEAD II  
(LL − RA)  
LEAD III  
(LL − LA)  
0
0
0
1REGISTER FRMCTL, BIT DATAFMT: 0 = LEAD/VECTOR MODE; 1 = ELECTRODE MODE.  
2REGISTER ECGCTL, BIT CHCONFIG: 0 = SINGLE ENDED INPUT (DIGITAL LEAD MODE OR ELECTRODE MODE); 1 = DIFFERENTIAL INPUT (ANALOG LEAD MODE).  
3REGISTER CMREFCTL, BIT CEREFEN: 0 = CE DISABLED; 1 = CE ENABLED.  
Figure 57. Electrode and Lead Configurations, Digital Lead Mode  
Rev. B | Page 30 of 80  
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
VCM = RA  
CM_OUT/WCT  
COMMON-  
MODE AMP  
ECG1_LA  
ECG2_LL  
+
LEAD I  
ADC  
ADC  
ADC  
AMP  
+
AMP  
LEAD II  
DIGITAL DOMAIN  
CALCULATIONS  
+
AMP  
V1’  
ECG3_RA = V1  
CM_IN = RA  
COMMON  
ELECTRODE CE IN  
0x0A 0x01 0x05  
MODE  
COMMENT  
WORD1  
LEAD I  
WORD2  
LEAD II  
WORD3  
[4]1 [10]2  
[8]3  
COMMON ELECTRODE (CE)  
COMMON  
V3’  
0
0
1
ELECTRODE A  
LEADS (HERE RAELECTRODE IS  
CONNECTED TO THECE ELECTRODE  
(CM_IN) AND V3 IS ON ECG3 INPUT)  
(V3 – RA) − (LA − RA) − (LL − RA)  
(LA − RA)  
(LL − RA)  
3
1REGISTER FRMCTL, BIT DATAFMT: 0 = LEAD/VECTOR MODE; 1 = ELECTRODE MODE.  
2REGISTER ECGCTL, BIT CHCONFIG: 0 = SINGLE ENDED INPUT (DIGITAL LEAD MODE OR ELECTRODE MODE); 1 = DIFFERENTIAL INPUT (ANALOG LEAD MODE).  
3REGISTER CMREFCTL, BIT CEREFEN: 0 = CE DISABLED; 1 = CE ENABLED.  
Figure 58. Electrode and Lead Configurations, Common Electrode A  
VCM = (LA + LL + RA + V1)/  
2 IN THIS CASE  
CM_OUT/WCT  
VCM COMMON MODE  
CAN BE ANY COMBINATION  
OF ELECTRODES  
COMMON-  
MODE AMP  
ECG1_LA  
ECG 2_LL  
ECG3_RA  
+
LA – VCM  
LL – VCM  
ADC  
ADC  
AMP  
+
AMP  
+
AMP  
RA – VCM  
ADC  
CM_IN  
FOR EXAMPLE, RA  
COMMON  
ELECTRODE CE IN  
0x0A 0x01 0x05  
MODE  
COMMENT  
WORD1  
WORD2  
WORD3  
1
2
3
[8]  
[4]  
[10]  
SINGLE-ENDED  
INPUT  
ELECTRODE  
SINGLE-ENDED  
INPUT ELECTRODE  
RELATIVE TO VCM  
LA − VCM  
LL − VCM  
RA − VCM  
1
0
0
1
2
3
REGISTER FRMCTL, BIT DATAFMT: 0 = LEAD/VECTOR MODE; 1 = ELECTRODE MODE.  
REGISTER ECGCTL, BIT CHCONFIG: 0 = SINGLE ENDED INPUT (DIGITAL LEAD MODE OR ELECTRODE MODE); 1 = DIFFERENTIAL INPUT (ANALOG LEAD MODE).  
REGISTER CMREFCTL, BIT CEREFEN: 0 = CE DISABLED; 1 = CE ENABLED.  
Figure 59. Electrode and Lead Configurations, Single-Ended Input Electrode  
Rev. B | Page 31 of 80  
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
VCM = CE = RA  
CM_OUT/WCT  
COMMON-  
MODE AMP  
ECG1_LA  
ECG2_LL  
+
LA – RA  
LL – RA  
ADC  
ADC  
AMP  
+
AMP  
+
AMP  
V3 – RA  
ECG3_RA = V3  
CM_IN = RA  
ADC  
COMMON  
ELECTRODE CE IN  
0x0A 0x01 0x05  
MODE  
COMMENT  
WORD1  
LA CE  
WORD2  
WORD3  
V1 CE  
1
2
3
[8]  
[4]  
[10]  
COMMON  
ELECTRODE B  
LEADS FORMED  
RELATIVE TO A  
COMMON  
LL CE  
1
0
1
ELECTRODE (CE)  
1
2
3
REGISTER FRMCTL, BIT DATAFMT: 0 = LEAD/VECTOR MODE; 1 = ELECTRODE MODE.  
REGISTER ECGCTL, BIT CHCONFIG: 0 = SINGLE ENDED INPUT (DIGITAL LEAD MODE OR ELECTRODE MODE); 1 = DIFFERENTIAL INPUT (ANALOG LEAD MODE).  
REGISTER CMREFCTL, BIT CEREFEN: 0 = CE DISABLED; 1 = CE ENABLED.  
Figure 60. Electrode and Lead Configurations, Common Electrode B  
Rev. B | Page 32 of 80  
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
DEFIBRILLATOR PROTECTION  
ESIS FILTERING  
The ADAS1000-3/ADAS1000-4 do not include defibrillation  
protection on chip. Any defibrillation protection required by  
the application requires external components. Figure 61 and  
Figure 62 show examples of external defibrillator protection,  
which is required on each ECG channel, in the RLD path, and  
in the CM_IN path if using the CE input mode. Note that, in  
both cases, the total ECG path resistance is assumed to be 5 kΩ.  
The 22 MΩ resistors shown connected to RLD are optional and  
used to provide a safe termination voltage for an open ECG  
electrode; they may be larger in value. Note that, if using these  
resistors, the dc lead-off feature works best with the highest  
current setting.  
The ADAS1000-3/ADAS1000-4 do not include electrosurgical  
interference suppression (ESIS) protection on chip. Any ESIS  
protection required by the application requires external  
components.  
ECG PATH INPUT MULTIPLEXING  
As shown in Figure 63, signal paths for numerous functions are  
provided on each ECG channel (except respiration, which only  
connects to the ECG1_LA, ECG2_LL, and ECG3_RA pins).  
Note that the channel enable switch occurs after the RLD  
amplifier connection, thus allowing the RLD to be connected  
(redirected into any one of the ECG paths). The CM_IN path is  
treated the same as the ECG signals.  
PATIENT  
CABLE  
4kΩ  
500Ω  
500Ω  
ECG1  
ELECTRODE  
AVDD  
1
22MΩ  
ARGON/NEON  
BULB  
ADAS1000-3/  
ADAS1000-4  
RLD  
1
SP724  
500Ω  
PATIENT  
22MΩ  
CABLE  
4kΩ  
500Ω  
AVDD  
ELECTRODE  
ECG2  
ARGON/NEON  
BULB  
SP724  
1
OPTIONAL.  
Figure 61. Possible Defibrillation Protection on ECG Paths Using Neon Bulbs  
PATIENT  
CABLE  
4.5kΩ  
AVDD  
500Ω  
ELECTRODE  
ECG1  
1
22MΩ  
2
SP724  
ADAS1000-3/  
ADAS1000-4  
RLD  
1
22MΩ  
PATIENT  
CABLE  
4.5kΩ  
500Ω  
ELECTRODE  
ECG2  
AVDD  
2
SP724  
1
2
OPTIONAL.  
TWO LITTELFUSE SP724 CHANNELS PER ELECTRODE MAY PROVIDE  
BEST PROTECTION.  
Figure 62. Possible Defibrillation Protection on ECG Paths Using Diode Protection  
ACLO  
CURRENT  
RESPIRATION  
INPUT  
DCLO  
CURRENT  
RLD AMP  
CALDAC  
11.3pF  
INPUT AMPLIFIER  
ECG PIN  
+
CHANNEL  
ENABLE  
TO  
FILTERING  
MUX FOR LEAD CONFIG,  
COMMON ELECTRODE  
ANALOG  
LEAD  
(RA/LA/LL)  
+
1.3V VCM_REF  
VCM  
FROM CM  
AVERAGING  
TO CM  
AVERAGING  
ADAS1000  
Figure 63. Typical ECG Channel Input Multiplexing  
Rev. B | Page 33 of 80  
 
 
 
 
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
common-mode block. If the physical connection to each  
electrode is buffered, these buffers are omitted for clarity.  
COMMON-MODE SELECTION AND AVERAGING  
The common-mode signal can be derived from any combina-  
tion of one or more electrode channel inputs, the fixed internal  
common-mode voltage reference, VCM_REF, or an external  
source connected to the CM_IN pin. One use of the latter  
arrangement is in gang mode where the master device creates  
the Wilson Central Terminal for the slave device(s). The fixed  
reference option is useful when measuring the calibration DAC  
test tone signals or while attaching electrodes to the patient,  
where it allows a usable signal to be obtained from just two  
electrodes.  
There are several restrictions on the use of the switches:  
If SW1 is closed, SW7 must be open.  
If SW1 is open, at least one electrode switch (SW2 to SW7)  
must be closed.  
SW7 can be closed only when SW2 to SW6 are open, so  
that the 1.3 V VCM_REF is summed in only when all ECG  
channels are disconnected.  
The CM_OUT output is not intended to supply current or drive  
resistive loads, and its accuracy is degraded if it is used to drive  
anything other than the slave ADAS1000-2 devices. An external  
buffer is required if there is any loading on the CM_OUT pin.  
The flexible common-mode generation allows complete  
user control over the contributing channels. It is similar to,  
but independent of, circuitry that creates the right leg drive  
(RLD) signal. Figure 64 shows a simplified version of the  
ADAS1000-3/  
ADAS1000-4  
CM_IN  
SW1  
SW2  
VCM  
+
CM_OUT  
ECG1_LA  
SW3  
ECG2_LL  
SW4  
ECG3_RA  
SW7  
VCM_REF = 1.3V  
(WHEN SELECTED, VCM_REF IS  
SUMMED IN ON EACH EC CHANNEL)  
Figure 64. Common-Mode Generation Block  
Table 11. Truth Table for Common-Mode Selection  
ECGCTL  
Address 0x011  
CMREFCTL Address 0x052  
PWREN  
DRVCM EXTCM LACM LLCM RACM On Switch  
Description  
0
1
1
1
1
X
X
0
0
0
X
0
0
0
0
X
0
1
1
1
X
0
0
1
1
X
0
0
0
1
Powered down, paths disconnected  
Internal VCM_REF = 1.3 V is selected  
Internal CM selection: LA contributes to VCM  
Internal CM selection: LA and LL contribute to VCM  
Internal CM selection: LA, LL, and RA contribute to VCM (WCT)  
SW7  
SW2  
SW2, SW3  
SW2, SW3,  
SW4  
.
.
.
.
.
.
.
.
1
X
1
X
X
X
SW1  
External VCM selected  
1 See Table 28.  
2 See Table 32.  
Rev. B | Page 34 of 80  
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
In some cases, adding lead compensation will prove necessary,  
while in others lag compensation may be more appropriate. The  
RLD amplifiers summing junction is brought out to a package  
pin (RLD_SJ) to facilitate compensation.  
WILSON CENTRAL TERMINAL (WCT)  
The flexibility of the common-mode selection averaging allows  
the user to achieve a Wilson Central Terminal voltage from the  
ECG1_LA, ECG2_LL, ECG3_RA electrodes.  
The RLD amplifiers short circuit current capability exceeds  
regulatory limits. A patient protection resistor is required to  
achieve compliance.  
RIGHT LEG DRIVE/REFERENCE DRIVE  
The right leg drive amplifier or reference amplifier is used as  
part of a feedback loop to force the patients common-mode  
voltage close to the internal 1.3 V reference level (VCM_REF)  
of the ADAS1000-3/ADAS1000-4. This centers all the electrode  
inputs relative to the input span, providing maximum input  
dynamic range. It also helps to reject noise and interference  
from external sources such as fluorescent lights or other  
patient-connected instruments, and absorbs the dc or ac  
lead-off currents injected on the ECG electrodes.  
Within the RLD block, there is lead-off comparator circuitry  
that monitors the RLD amplifier output to determine whether  
the patient feedback loop is closed. An open-loop condition,  
typically the result of the right leg electrode (RLD_OUT)  
becoming detached, tends to drive the amplifiers output  
low. This type of fault is flagged in the header word (see  
Table 53), allowing the system software to take action by  
notifying the user, redirecting the reference drive to another  
electrode via the internal switches of the ADAS1000-3/  
ADAS1000-4, or both. The detection circuitry is local to the  
RLD amplifier and remains functional with a redirected  
reference drive. Table 32 provides details on reference drive  
redirection.  
The RLD amplifier can be used in a variety of ways as shown  
in Figure 65. Its input can be taken from the CM_OUT signal  
using an external resistor. Alternatively, some or all of the  
electrode signals can be combined using the internal switches.  
The dc gain of the RLD amplifier is set by the ratio of the  
external feedback resistor (RFB) to the effective input resistor,  
which can be set by an external resistor, or alternatively, a  
function of the number of selected electrodes as configured in  
the CMREFCTL register (see Table 32). In a typical case, using  
the internal resistors for RIN, all active electrodes would be used  
to derive the right leg drive, resulting in a 2 kΩ effective input  
resistor. Achieving a typical dc gain of 40 dB would thus require  
a 200 kΩ feedback resistor.  
While reference drive redirection may be useful in the event  
that the right leg electrode cannot be reattached, some pre-  
cautions must be observed. Most important is the need for a  
patient protection resistor. Because this is an external resistor,  
it does not follow the redirected reference drive; some provision  
for continued patient protection is needed external to the  
ADAS1000-3/ADAS1000-4. Any additional resistance in the  
ECG paths will certainly interfere with respiration measure-  
ment and may also result in an increase in noise and decrease  
in CMRR.  
The dynamics and stability of the RLD loop depend on the  
chosen dc gain and the resistance and capacitance of the  
patient cabling. In general, loop compensation using  
external components is required, and must be determined  
experimentally for any given instrument design and cable set.  
The RLD amplifier is designed to stably drive a maximum  
capacitance of 5 nF based on the gain configuration (see  
Figure 65) and assuming a 330 kΩ patient protection resistor.  
EXTERNALLY SUPPLIED COMPONENTS  
TO SET RLD LOOP GAIN  
CZ  
2nF  
100k  
RZ  
40kΩ  
IN  
4MΩ  
RFB*  
R
*
RLD_SJ  
RLD_OUT  
CM_OUT/WCT  
ADAS1000-3/  
ADAS1000-4  
ELECTRODE LA  
ELECTRODE LL  
ELECTRODE RA  
+
SW1  
10kΩ  
10kΩ  
10kΩ  
SW2  
SW3  
SW6  
10kΩ  
CM_IN OR  
CM BUFFER OUT  
VCM_REF  
(1.3V)  
RLD_INT_REDIRECT  
*EXTERNAL RESISTOR R IS OPTIONAL. IF DRIVING RLD FROM  
IN  
THE ELECTRODE PATHS, THEN THE SERIES RESISTANCE WILL  
CONTRIBUTE TO THE R IMPEDANCE. WHERE SW1 TO SW5  
IN  
ARE CLOSED, R = 2k. RFB SHOULD BE CHOSEN  
IN  
ACCORDINGLY FOR DESIRED RLD LOOP GAIN.  
Figure 65. Right Leg Drive—Possible External Component Configuration  
Rev. B | Page 35 of 80  
 
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
degrade. The user has full control over the common-mode  
CALIBRATION DAC  
amplifier and can adjust the common-mode configuration to  
remove that electrode from the common-mode generation. In  
this way, the user can continue to make measurements on the  
remaining connected leads.  
Within the ADAS1000-3/ADAS1000-4, there are a number of  
calibration features.  
The 10-bit calibration DAC can be used to correct channel gain  
errors (to ensure channel matching) or to provide several test  
tones. The options are as follows:  
DC Lead-Off Detection  
This method injects a small programmable dc current into each  
input electrode. When an electrode is properly connected, the  
current flows into the right leg (RLD_OUT) and produces a  
minimal voltage shift. If an electrode is off, the current charges  
that pin’s capacitance, causing the voltage at the pin to float  
positive and create a large voltage change that is detected by  
the comparators in each channel. These comparators use fixed,  
gain-independent upper and lower threshold voltages of 2.4 V  
and 0.2 V, respectively. If the input exceeds either of these levels,  
the lead-off flag is raised. The lower threshold is included in the  
event that something pulls the electrode down to ground.  
DC voltage output (range: 0.3 V to 2.7 V). The DAC  
transfer function for dc voltage output is  
code  
0.3 V+ 2.4 V×  
210 1  
)
(
1 mV p-p sine wave of 10 Hz or 150 Hz  
1 mV 1 Hz square wave  
Internal switching allows the calibration DAC signals to be  
routed to the input of each ECG channel (see Figure 63).  
Alternatively, it can be driven out from the CAL_DAC_IO  
pin, enabling measurement and correction for external error  
sources in the entire ECG signal chain.  
The dc lead-off detection current can be programmed via the  
serial interface. Typical currents range from 10 nA to 70 nA in  
10 nA steps. All input pins (RA, LA, LL, V1, V2, and CM_IN)  
use identical dc lead-off detection circuitry.  
To ensure a successful update of the calibration DAC (see  
Table 36), the host controller must issue four additional SCLK  
cycles after writing the new calibration DAC register word.  
Detecting if the right-leg electrode has fallen off is necessarily  
different as RLD_OUT is a low impedance amplifier output. A  
pair of fixed threshold comparators monitor the output voltage  
to detect amplifier saturation that would indicate a lead-off  
condition. This information is available in the DCLEAD-OFF  
register (Register 0x1E) along with the lead-off status of all the  
input pins.  
GAIN CALIBRATION  
The gain for each ECG channel can be adjusted to correct  
for gain mismatches between channels. Factory trimmed  
gain correction coefficients are stored in nonvolatile memory  
on-chip for GAIN 0, GAIN 1, and GAIN 2; there is no factory  
calibration for GAIN 3. The default gain values can be over-  
written by user gain correction coefficients, which are stored  
in volatile memory and available by addressing the appropriate  
gain control registers (see Table 50). The gain calibration applies  
to the ECG data available on the standard interface and applies  
to all data rates.  
The propagation delay for detecting a dc lead-off event depends  
on the cable capacitance and the programmed current. It is  
approximately  
Delay = Voltage × Cable Capacitance/Programmed Current  
For example:  
LEAD-OFF DETECTION  
Delay = 1.2 V × (200 pF/70 nA) = 3.43 ms  
DC Lead-Off and High Gains  
An ECG system must be able to detect if an electrode is no  
longer connected to the patient. The ADAS1000-3/ADAS1000-4  
support two methods of lead-off detection, ac lead-off detection  
and dc lead-off detection. The two systems are independent and  
can be used singly or together under the control of the serial  
interface (see Table 29).  
Using dc lead-off at high gains can result in failure of the circuit  
to flag a lead-off condition. The chopping nature of the input  
amplifier stage contributes to this situation. When the electrode  
is off, the electrode is pulled up; however, in this gain setting,  
the first stage amplifier goes into saturation before the input  
signal crosses the DCLO upper threshold, resulting in no lead-  
off flag. This affects the gain setting GAIN 3 (4.2) and partially  
GAIN 2 (2.8).  
A lead-off event sets a flag in the frame header word (see Table 53).  
Identification of which electrode is off is available as part of the  
data frame or as a register read from the lead-off status register  
(Register LOFF, see Table 47). In the case of ac lead-off, infor-  
mation about the amplitude of the lead-off signal or signals can  
be read back through the serial interface (see Table 51).  
Increasing the AVDD voltage raises the voltage at which the input  
amplifiers saturate, allowing the off electrode voltage to rise high  
enough to trip the DCLO comparator (fixed upper threshold of  
2.4 V). The ADAS1000 operates over a voltage range of 3.15 V  
to 5.5 V. If using GAIN 2/GAIN 3 and dc lead-off, an increased  
AVDD supply voltage (minimum 3.6 V) allows dc lead-off to  
flag correctly at higher gains.  
In a typical ECG configuration, the electrodes RA, LA, and LL  
are used to generate a common mode of Wilson Central Terminal  
(WCT). If one of these electrodes is off, this affects the WCT  
signal and any lead measurements that it contributes to. As a  
result, the ECG measurements on these signals are expected to  
Rev. B | Page 36 of 80  
 
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
AC Lead-Off Detection  
case, there is no ac signal present, yet the electrode may not be  
connected. The lower threshold checks for a minimum signal level.  
The alternative method of sensing if the electrodes are  
connected to the patient is based on injecting ac currents into  
each channel and measuring the amplitudes of the resulting  
voltages. The system uses a fixed carrier frequency at 2.039 kHz,  
which is high enough to be removed by the ADAS1000-3/  
ADAS1000-4 on-chip digital filters without introducing phase  
or amplitude artifacts into the ECG signal.  
In addition to the lead-off flag, the user can also read back the  
resulting voltage measurement available on a per channel basis.  
The measured amplitude for each of the individual electrodes is  
available in Register 0x31 through Register 0x35 (LOAMxx  
registers, see Table 51).  
The propagation delay for detecting an ac lead-off event is  
<10 ms.  
AC LO  
DAC  
Note that the ac lead-off function is disabled when the  
calibration DAC is enabled.  
ADC Out of Range  
2.039kHz  
11kΩ  
11kΩ  
11kΩ  
12.5nA TO  
100nA rms  
When multiple leads are off, the input amplifiers may run into  
saturation. This results in the ADC outputting out of range data  
with no carrier to the leads off algorithm. The ac lead-off algorithm  
then reports little or no ac amplitude. The ADAS1000 contains  
flags to indicate if the ADC data is out of range, indicating a hard  
electrode off state. There are programmable overrange and under-  
range thresholds that can be seen in the LOFFUTH and LOFFLTH  
registers (see Table 39 and Table 40, respectively). The ADC out  
of range flag is contained in the header word (see Table 53).  
LA  
LL  
RA  
CM  
Figure 66. Simplified AC Lead-Off Configuration  
The amplitude of the signal is nominally 2 V p-p and is centered  
on 1.3 V relative to the chip AGND level. It is ac-coupled into each  
electrode. The polarity of the ac lead-off signal can be configured  
on a per-electrode basis through Bits[23:18] of the LOFFCTL  
register (see Table 29). All electrodes can be driven in phase, and  
some can be driven with reversed polarity to minimize the total  
injected ac current. Drive amplitude is also programmable. AC  
lead-off detection functions only on the input pins (LA, LL, RA,  
and CM_IN) and is not supported for the RLD_OUT pin.  
SHIELD DRIVER  
The shield drive amplifier is a unity-gain amplifier. Its purpose  
is to drive the shield of the ECG cables. For power consumption  
purposes, it can be disabled if not in use. Note that, the SHIELD  
pin is shared with the respiration pin function, where it can be  
muxed to be one of the pins for external capacitor connection.  
If the pin is being used for the respiration feature, the shield  
function is not available. In this case, if the application requires  
a shield drive, an external amplifier connected to the CM_OUT  
pin can be used.  
The resulting analog input signal applied to the ECG channels is  
I/Q demodulated and amplitude detected. The resulting amplitude  
is low pass filtered and sent to the digital threshold detectors.  
AC lead-off detection offers user programmable dedicated  
upper and lower threshold voltages (see Table 39 and Table 40).  
Note that these programmed thresholds voltage vary with the  
ECG channel gain. The threshold voltages are not affected by  
the current level that is programmed. All active channels use  
the same detection thresholds.  
RESPIRATION (ADAS1000-4 MODEL ONLY)  
The respiration measurement is performed by driving a high  
frequency (programmable from 46.5 kHz to 64 kHz) differential  
current into two electrodes; the resulting impedance variation  
caused by breathing causes the differential voltage to vary at the  
respiration rate. The signal is ac-coupled onto the patient. The  
acquired signal is AM, with a carrier at the driving frequency  
and a shallow modulation envelope at the respiration frequency.  
The modulation depth is greatly reduced by the resistance of the  
customer-supplied RFI and ESIS protection filters, in addition  
to the impedance of the cable and the electrode to skin interface  
(see Table 12). The goal is to measure small ohm variation  
to sub ohm resolution in the presence of large series resistance.  
The circuit itself consists of a respiration DAC that drives the  
ac-coupled current at a programmable frequency onto the  
chosen pair of electrodes. The resulting variation in voltage  
is amplified, filtered, and synchronously demodulated in the  
digital domain; what results is a digital signal that represents  
the total thoracic or respiration impedance, including cable and  
electrode contributions. While it is heavily low-pass filtered  
on-chip, the user is required to further process it to extract the  
A properly connected electrode has a very small signal as the  
drive current flows into the right leg (RL), whereas a disconnected  
electrode has a larger signal as determined by a capacitive  
voltage divider (source and cable capacitance).  
If the signal measured is larger than the upper threshold, then  
the impedance is high, so a wire is probably off. Selecting the  
appropriate threshold setting depends on the particular cable/  
electrode/protection scheme, as these parameters are typically  
unique for the specific use case. This can take the form of starting  
with a high threshold and ratcheting it down until a lead-off is  
detected, then increasing the threshold by some safety margin.  
This gives simple dynamic thresholding that automatically  
compensates for many of the circuit variables.  
The lower threshold is added for cases where the only ac lead-off  
is in use and for situations where an electrode cable has been off  
for a long time. In this case, the dc voltage has saturated to a rail,  
or the electrode cable has somehow shorted to a supply. In either  
Rev. B | Page 37 of 80  
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
envelope and perform the peak detection needed to establish  
breathing (or lack thereof).  
Internal Respiration Capacitors  
The internal respiration function uses an internal RC network  
(5 kΩ/100 pF), and this circuit is capable of 200 mΩ resolution  
(with up to 5 kΩ total path and cable impedance). The current  
is ac-coupled onto the same pins that the measurement is  
sensed back on. Figure 67 shows the measurement on Lead I,  
but, similarly, the measurement can be configured to measure  
on either Lead II or Lead III. The internal capacitor mode requires  
no external capacitors and produces currents of ~64 μA p-p  
amplitude when configured for maximum amplitude setting  
( 1 V) through the RESPCTRL register (see Table 30).  
Respiration measurement is available on one of the leads  
(Lead I, Lead II, or Lead III) or on an external path via a  
pair of dedicated pins (EXT_RESP_LA, EXT_RESP_RA, or  
EXT_RESP_LL). Only one lead measurement can be made  
at one time. The respiration measurement path is not suited  
for use as additional ECG measurements because the internal  
configuration and demodulation do not align with an ECG  
measurement. The respiration signal processing path is not  
reconfigurable for ECG measurements, as it is specifically  
designed for the respiration signal measurement.  
Table 12. Maximum Allowable Cable and Thoracic Loading  
Cable Resistance  
Cable Capacitance  
R < 1 kΩ  
C < 1200 pF  
1 kΩ < R < 2.5 kΩ  
2.5 kΩ < R < 5 kΩ  
C < 400 pF  
C < 200 pF  
RTHORACIC < 2 kΩ  
±1V  
46.5kHz TO  
64kHz  
RESPIRATION DAC  
DRIVE +  
ADAS1000-4  
5k  
100pF  
CABLE AND ELECTRODE  
RESPIRATION  
MEASURE  
IMPEDANCE < 5kΩ  
ECG1_LA  
EXT_RESP_LA  
ECG2_LL  
LA CABLE  
LL CABLE  
RA CABLE  
FILTER  
FILTER  
FILTER  
IN-AMP AND  
ANTI-ALIASING  
OVERSAMPLED  
HPF  
MAGNITUDE  
AND  
SAR  
ADC  
EXT_RESP_LL  
ECG3_RA  
PHASE  
LPF  
EXT_RESP_RA  
100pF  
5kΩ  
RESPIRATION DAC  
DRIVE–  
46.5kHz TO  
64kHz  
±1V  
Figure 67. Simplified Respiration Block Diagram  
Rev. B | Page 38 of 80  
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
External Respiration Capacitors  
External Respiration Path  
If necessary, the ADAS1000-4 allows the user to connect  
external capacitors into the respiration circuit to achieve higher  
resolution (<200 mΩ). This level of resolution requires that the  
cable impedance be ≤1 kΩ. The diagram in Figure 68 shows the  
connections at RESPDAC_xx paths for the extended respiration  
configuration. Again, the EXT_RESP_xx paths can be con-  
nected at the patient side of any filtering circuit; however, the  
user must provide protection for these pins. While this external  
capacitor mode requires external components, it can deliver a  
larger signal-to-noise ratio. Note again that respiration can be  
measured on only one lead (at one time); therefore, only one  
pair of external respiration paths (and external capacitors) may  
be required.  
The EXT_RESP_xx pins are provided for use either with the  
ECG electrode cables or, alternatively, with a dedicated external  
sensor independent of the ECG electrode path. Additionally, the  
EXT_RESP_xx pins are provided so the user can measure the  
respiration signal at the patient side of any input filtering on the  
front end. In this case, the user must continue to take precautions  
to protect the EXT_RESP_xx pins from any signals applied that  
are in excess of the operating voltage range (for example, ESIS  
or defibrillator signals).  
±1V  
46.5kHz TO  
64kHz  
RESPIRATION DAC  
ADAS1000-4  
DRIVE +  
1nF TO 10nF  
1kΩ  
RESPDAC_LA  
RESPDAC_LL  
1kΩ  
5kΩ  
100pF  
MUTUALLY  
EXCLUSIVE  
CABLE AND ELECTRODE  
RESPIRATION  
MEASURE  
IMPEDANCE < 1kΩ  
ECG1_LA  
EXT_RESP_LA  
ECG2_LL  
LA CABLE  
LL CABLE  
RA CABLE  
FILTER  
IN-AMP AND  
ANTI-ALIASING  
OVERSAMPLED  
HPF  
MAGNITUDE  
AND  
FILTER  
FILTER  
SAR  
ADC  
EXT_RESP_LL  
ECG3_RA  
PHASE  
LPF  
EXT_RESP_RA  
MUTUALLY  
EXCLUSIVE  
100pF  
5kΩ  
1nF TO 10nF  
1kΩ  
RESPDAC_RA  
RESPIRATION DAC  
DRIVE –  
46.5kHz TO  
64kHz  
±1V  
Figure 68. Respiration Measurement Using External Capacitor  
Rev. B | Page 39 of 80  
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
50kHz TO  
56kHz  
ADAS1000-4  
1nF TO 10nF  
100Ω  
RESPDAC_LA  
1kΩ  
±1V  
RESPIRATION DAC  
DRIVE + ve  
CABLE AND ELECTRODE  
IMPEDANCE < 1kΩ  
RESPIRATION  
MEASURE  
LA CABLE  
EXT_RESP_LA  
IN-AMP AND  
ANTI-ALIASING  
OVERSAMPLED  
10kΩ  
HPF  
RA CABLE  
MAGNITUDE  
AND  
PHASE  
SAR  
ADC  
GAIN  
10kΩ  
LPF  
EXT_RESP_RA  
1/2 OF AD8606  
REFOUT = 1.8V  
RESPDAC_RA  
10kΩ  
10kΩ  
46.5kHz TO  
64kHz  
0.9V  
1nF TO 10nF  
100Ω  
1kΩ  
±1V  
RESPIRATION DAC  
DRIVE – ve  
1/2 OF AD8606  
Figure 69. Respiration Using External Capacitor and External Amplifiers  
If required, further improvements in respiration performance  
may be possible with the use of an instrumentation amplifier  
and op amp external to the ADAS1000-4. The instrumentation  
amplifier must have sufficiently low noise performance to meet  
the target performance levels. This mode uses the external  
capacitor mode configuration and is shown in Figure 69. Bit 14  
of the RESPCTL register (Table 30) allows the user to bypass  
the on-chip amplifier when using an external instrumentation  
amplifier.  
In applications where an external signal generator is used to  
develop a respiration carrier signal, that external signal source  
can be synchronized to the internal carrier using the signal  
available on GPIO3 when Bit 7, RESPEXTSEL, is enabled in  
the respiration control register (see Table 30).  
Table 14. Control of Respiration Carrier Frequency  
Available on GPIO3  
RESPALT-  
RESPEXT-  
RESP- Respiration Carrier  
FREQ1  
SYNC1  
HP2 FREQ1 Frequency on GPIO3  
Respiration Carrier Frequency  
0
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
X3  
1
1
1
1
0
0
0
0
XX3  
00  
01  
10  
11  
00  
01  
10  
11  
64  
64  
56  
51.2  
46.5  
32  
28  
25.5  
23  
The frequency of the respiration carrier is programmable and  
can be varied through the RESPCTL register (Address 0x03, see  
Table 30). The status of the HP bit in the ECGCTL register also  
has an influence on the carrier frequency as shown in Table 13.  
Table 13. Control of Respiration Carrier Frequencies  
RESPALT-  
RESPEXT-  
RESP-  
Respiration Carrier  
Frequency  
FREQ1  
SYNC1  
HP2 FREQ1  
0
0
0
0
0
0
0
0
1
1
1
1
1
1
1
1
0
0
0
0
0
0
0
0
X3  
X3  
X3  
X3  
X3  
X3  
X3  
X3  
1
1
1
1
0
0
0
0
1
1
1
1
0
0
0
0
00  
01  
10  
11  
00  
01  
10  
11  
00  
01  
10  
11  
00  
01  
10  
11  
56  
54  
52  
50  
56  
54  
52  
50  
1 Control bits from RESPCTL (Register 0x03).  
2 Control bit from ECGCTL (Register 0x01).  
3 X = don’t care.  
64  
56.9  
51.2  
46.5  
32  
28  
25.5  
23  
1 Control bits from RESPCTL (Register 0x03).  
2 Control bit from ECGCTL (Register 0x01).  
3 X = don’t care.  
Rev. B | Page 40 of 80  
 
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
ECG data packet/frame as dictated by the frame control register  
(see Table 37). The data available in the PACEDATA register  
is limited to seven bits total for width and height information;  
therefore, if more resolution is required on the pace height  
and width, this is available by issuing read commands of the  
PACExDATA registers (Address 0x3A to Address 0x3C) as  
shown in Table 52.  
EVALUATING RESPIRATION PERFORMANCE  
ECG simulators offer a convenient means of studying the  
ADAS1000-3/ADAS1000-4s performance. While many  
simulators offer a variable-resistance respiration capability,  
care must be taken when using this feature.  
Some simulators use electrically-programmable resistors, often  
referred to as digiPOTs, to create the time-varying resistance  
to be measured by the respiration function. The capacitances  
at the digitPOT's terminals are often unequal and code-  
dependent, and these unbalanced capacitances can give rise  
to unexpectedly large or small results on different leads for  
the same programmed resistance variation. Best results are  
obtained with a purpose-built fixture that carefully balances  
the capacitance presented to each ECG electrode.  
The on-chip filtering contributes some delay to the pace signal  
(see the Pace Latency section).  
Choice of Leads  
Three identical and independent state machines are available  
and can be configured to run on up to three of four possible  
leads (Lead I, Lead II, Lead III, and aVF) for pacing artifact  
detection. Any necessary lead calculations are performed  
internally and are independent of EGG channel settings for  
output data rate, low-pass filter cutoff, and mode (electrode,  
analog lead, common electrode). These calculations take into  
account the available front-end configurations as detailed in  
Table 15.  
PACING ARTIFACT DETECTION FUNCTION  
(ADAS1000-4 ONLY)  
The pacing artifact validation function qualifies potential  
pacing artifacts and measures the width and amplitude of valid  
pulses. These parameters are stored in and available from any of  
the pace data registers (Address 0x1A, Address 0x3A to Address  
0x3C). This function runs in parallel with the ECG channels.  
Digital detection is performed using a state machine operating  
on the 128 kHz 16-bit data from the ECG decimation chain.  
The main ECG signals are further decimated before appearing  
in the 2 kHz output stream so that detected pace signals are not  
perfectly time-aligned with fully-filtered ECG data. This time  
difference is deterministic and may be compensated for.  
The pace detection algorithm searches for pulses by analyzing  
samples in the 128 kHz ECG data stream. The algorithm  
searches for a leading edge, a peak, and a trailing edge as  
defined by values in the PACEEDGETH, PACEAMPTH, and  
PACELVLTH registers, along with fixed width qualifiers. The  
post-reset default register values can be overwritten via the SPI  
bus, and different values can be used for each of the three pace  
detection state machines.  
Some users may not want to use three pace leads for detection.  
In this case, Lead II is the vector of choice, because this lead is  
likely to display the best pacing artifact. The other two pace  
instances can be disabled if not in use.  
The pacing artifact validation function can detect and  
measure pacing artifacts with widths from 100 μs to 2 ms  
and with amplitudes of <400 μV to >1000 mV. Its filters are  
designed to reject heartbeat, noise, and minute ventilation  
pulses. The flowchart for the pace detection algorithm is  
shown in Figure 71.  
The first step in pace detection is to search the data stream for  
a valid leading edge. Once a candidate edge has been detected,  
the algorithm begins searching for a second, opposite-polarity  
edge that meets with pulse width criteria and passes the  
(optional) noise filters. Only those pulses meeting all the  
criteria are flagged as valid pace pulses. Detection of a valid  
pace pulse sets the flag(s) in the frame header register and  
stores amplitude and width information in the PACEDATA  
register (Address 0x1A; see Table 44). The pace algorithm  
looks for a negative or positive pulse  
The ADAS1000-4 pace algorithm can operate with the ac lead-  
off and respiration impedance measurement circuitry enabled.  
Once a valid pace has been detected in the assigned leads, the  
pace-detected flags appear in the header word (see Table 53) at  
the start of the packet of ECG words. These bits indicate that a  
pace was qualified. Further information on height and width of  
pace is available by reading the contents of Address 0x1A (Register  
PACEDATA, see Table 44). This word can be included in the  
Rev. B | Page 41 of 80  
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
Table 15. Pace Lead Calculation  
0x04 [8:3]3  
11  
00  
01  
10  
Lead I  
(LA − RA)  
Lead II  
(LL − RA)  
Lead III  
(LL − LA)  
LL − LA  
aVF  
0x01 [10]1 0x05 [8]2 Configuration  
(Lead II + Lead III)/2  
0
0
Digital leads  
LA − RA  
LL − RA  
LL − (LA + RA)/2  
CH1 − CH3 CH2 − CH3 CH2 − CH1  
CH2 − (CH1 + CH3)/2  
0
1
Common  
Electrode  
Lead A  
Lead I  
CH1  
Lead II  
CH2  
Lead II − 0.5 × Lead I  
CH2 − 0.5 × CH1  
Lead II Lead I  
CH2 − CH1  
1
X
Analog leads  
Lead I  
CH1  
Lead II  
− CH3  
Lead III  
CH2  
Lead II − 0.5 × Lead I − CH3 − 0.5 × CH1  
1 Register ECGCTL, Bit CHCONFIG, see Table 28.  
2 Register CMREFCTL, Bit CEREFEN, see Table 32.  
3 Register PACECTL, Bit PACExSEL [1:0], see Table 31.  
Rev. B | Page 42 of 80  
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
START  
Detection Algorithm Overview  
The pace pulse amplitude and width varies over a wide range,  
while its shape is affected by both the internal filtering arising  
from the decimation process and the low pass nature of the  
electrodes, cabling, and components used for defibrillation and  
ESIS protection. The ADAS1000-4 provides user programmable  
variables to optimize the performance of the algorithm within  
the ECG system, given all these limiting elements. The default  
parameter values are probably not optimal for any particular  
system design; experimentation and evaluation are needed to  
ensure robust performance.  
ENABLE PACE DETECTION  
SELECT LEADS  
START PACE DETECTION  
ALGORITHM  
FIND  
LEADING EDGE  
A > PACEEDGETH?  
NO  
NO  
NO  
YES  
START PULSE  
WIDTH TIMER  
PACE  
PULSE  
FIND  
END OF  
LEADING EDGE  
B < PACELVL  
PACELVLTH  
YES  
START NOISE  
FILTERS (if enabled)  
LEADING EDGE  
YES  
LEADING EDGE STOP  
PACEAMPTH  
TRAILING  
EDGE  
DETECTED?  
PACEEDGETH  
PACE WIDTH  
YES  
NO  
NO  
NO  
PACE AMPLITUDE  
> PACEAMPTH  
RECHARGE  
PULSE  
YES  
Figure 70. Typical Pace Signal  
NOISE FILTER  
PASSED?  
The first step in pace detection is to search the data stream for a  
valid leading edge. Once a candidate edge is detected, the algorithm  
verifies that the signal looks like a pulse and then begins searching  
for a second, opposite polarity edge that meets the pulse width  
and amplitude criteria and passes the optional noise filters. Only  
the pulses meeting all requirements are flagged as valid pace pulses.  
Detection of a valid pace pulse sets the flag or flags in the frame  
header register and stores amplitude and width information in  
the PACEDATA register (Address 0x1A; see Table 34).  
YES  
PULSE WIDTH  
> 100µs AND <2ms  
YES  
FLAG PACE DETECTED  
UPDATE REGISTERS WITH  
WIDTH AND HEIGHT  
The pace algorithm detects pulses of both negative and positive  
polarity using a single set of parameters by tracking the slope of  
the leading edge and making the necessary adjustments to internal  
parameter signs. This frees the user to concentrate on determining  
appropriate threshold values based on pulse shape without  
concern for pulse polarity.  
Figure 71. Overview of Pace Algorithm  
The three user controlled parameters for the pace detection  
algorithm are Pace Amplitude Threshold (PACEAMPTH), Pace  
Level Threshold (PACELVLTH), and Pace Edge Threshold  
(PACEEDGETH).  
Rev. B | Page 43 of 80  
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
Pace Edge Threshold  
Pace Amplitude Threshold  
This programmable level (Address 0x0E, see Table 41) is used to  
find a leading edge, signifying the start of a potential pace pulse. A  
candidate edge is one in which the leading edge crosses a threshold  
PACEEDGETH from the recent baseline. PACEEDGETH can  
be assigned any value between 0 and 255. Setting PACEEDGETH  
to 0 forces it to the value PACEAMPTH/2 (see the following  
equation). Non-zero values give the following:  
This register (Address 0x07, see Table 34) sets the minimum  
valid pace pulse amplitude. PACEAMPTH is an unsigned 8-bit  
number. The programmed height is given by:  
2×N ×VREF  
PACEAMPTH setting =  
,
GAIN ×216  
where:  
N is the 8-bit programmed PACEAMPTH value (1 ≤ N ≤ 255).  
VREF is the ADAS1000-4 reference voltage of 1.8 V.  
GAIN is the programmed gain of the ECG channel.  
N ×VREF  
PACEEDGETH setting =  
GAIN ×216  
where:  
The minimum threshold for ×1.4 gain is 19.6 µV, while the  
maximum for the same gain setting is 5.00 mV. PACEAMPTH  
is typically set to the minimum expected pace amplitude and  
must be larger than the value of PACEEDGETH.  
N is the 8-bit programmed PACEEDGETH value (1 ≤ N ≤ 255).  
VREF is the ADAS1000-4 reference voltage of 1.8 V.  
GAIN is the programmed gain of the ECG channel.  
The minimum threshold for ×1.4 gain is 19.6 µV, while the  
maximum for the same gain setting is 5.00 mV.  
The default register setting of N = 0x24 results in 706 μV for a  
gain = 1 setting. An initial PACEAMPTH setting between  
700 µV and 1 mV provides a good starting point for both  
unipolar and biventricular pacing detection. Values below  
250 µV are not recommended because they greatly increase  
sensitivity to ambient noise from the patient. The amplitude  
may need to be adjusted much higher than 1 mV when other  
medical devices are connected to the patient.  
Pace Level Threshold  
This programmable level (Address 0x0F, see Table 42) is used to  
detect when the leading edge of a candidate pulse ends. In general,  
a pace pulse is not perfectly square, and the top, meaning the  
portion after the leading edge, may continue to increase slightly  
or droop back towards the baseline. PACELVLTH defines an  
allowable slope for this portion of the candidate pulse, where  
the slope is defined as the change in value over an internally-  
fixed interval after the pace edge is qualified.  
Pace Validation Filters  
A candidate pulse that successfully passes the combined tests of  
PACEEDGETH, PACELVLTH, and PACEAMPTH is next passed  
through two optional validation filters. These filters are used to  
reject sub-threshold pulses such as minute ventilation (MV) pulse  
and signals from inductively coupled implantable telemetry  
systems. These filters perform different tests of pulse shape  
using a number of samples. Both filters are enabled by default;  
Filter 1 is controlled by Bit 9 in the PACECTL register (see  
Table 31) and Filter 2 is controlled by Bit 10 in the same register.  
These filters are not available on a lead by lead basis; if enabled,  
they are applied to all leads being used for pace detection.  
PACELVLTH is an 8-bit, twos complement number. Positive  
values represent movement away from the baseline (pulse  
amplitude is still increasing) while negative values represent  
droop back towards the baseline.  
N ×VREF  
PACELVLTH setting =  
GAIN ×216  
where:  
N is the 8-bit programmed PACELVLTH value (−128 ≤ N ≤ 127).  
VREF is the ADAS1000-4 reference voltage of 1.8 V.  
GAIN is the programmed gain of the ECG channel.  
Pace Width Filter  
A candidate pulse that successfully passes the edge, amplitude, and  
noise filters is finally checked for width. When this final filter is  
enabled, it checks that the candidate pulse is between 100 μs  
and 2 ms wide. When a valid pace width is detected, the width  
is stored. Disabling this filter affects only the minimum width  
(100 µs) determination; the maximum width detection portion  
of the filter is always active. This filter is controlled by the  
PACECTL register, Bit 11 (see Table 31).  
The minimum value for ×1.4 gain is 9.8 µV, while the maximum  
for the same gain setting is 2.50 mV.  
An additional qualification step, performed after PACELVLTH  
is satisfied, rejects pulses with a leading edge transition time greater  
than about 156 µs. This filter improves immunity to motion and  
other artifacts and cannot be disabled. Overly aggressive ESIS  
filtering causes this filter to disqualify valid pace pulses. In such  
cases, increasing the value of PACEEDGETH provides more  
robust pace pulse detection. Although counterintuitive, this  
change forces a larger initial deviation from the recent baseline  
before the pace detection algorithm starts, reducing the time  
until PACELVLTH comes into play and shortening the apparent  
leading edge transition. Increasing the value of PACEEDGETH  
may require a reduction in PACEAMPTH.  
Rev. B | Page 44 of 80  
Data Sheet  
ADAS1000-3/ADAS1000-4  
BIVENTRICULAR PACERS  
PACE WIDTH  
As described previously, the pace algorithm expects the pace  
pulse to be less than 2 ms wide. In a pacer where both ventricles  
are paced, they can be paced simultaneously. Where they fall  
within the width and height limits programmed into the algo-  
rithm, a valid pace will be flagged, but only one pace pulse may  
be visible.  
The ADAS1000-4 is capable of measuring pace widths of 100 μs  
to 2.00 ms. The measured pace width is available through the  
PACExDATA registers. These registers have limited resolution.  
The minimum pace width is 101.56 μs and the maximum is  
2.00 ms. The pace detection algorithm always returns a width  
greater than what is measured at the 50% point, ensuring that  
the algorithm is capable of measuring a narrow 100 μs pulse. A  
valid pulse width of 100 μs is reported as 101.56 μs. Any valid  
pace pulses ≥2.00 ms and ≤ 2.25 ms are reported as 2.00 ms.  
With the pace width filter enabled, the pace algorithm seeks  
pace pulse widths within a 100 μs to 2 ms window. Assuming  
that this filter is enabled and in a scenario where two ventricle  
pacer pulses fire at slightly different times, resulting in the pulse  
showing in the lead as one large, wider pulse, a valid pace is  
flagged so long as the total width does not exceed 2 ms.  
PACE LATENCY  
The pace algorithm always examines 128 kHz, 16-bit ECG data,  
regardless of the selected frame rate and ECG filter setting. A  
pace pulse is qualified when a valid trailing edge is detected  
and is flagged in the next available frame header. Pace and ECG  
data is always correctly time-aligned at the 128 kHz frame rate,  
but the additional filtering inherent in the slower frame rates  
delays the frame's ECG data relative to the pace pulse flag.  
These delays are summarized in Table 16 and must be taken  
into account to enable correct positioning of the pace event  
relative to the ECG data.  
PACE DETECTION MEASUREMENTS  
Design verification of the ADAS1000-4 digital pace algorithm  
includes detection of a range of simulated pace signals in  
addition to using the ADAS1000-4 and evaluation board with  
one pacemaker device connected to various simulated loads  
(approximately 200 Ω to over 2 kΩ) and covering the following  
four waveform corners.  
Minimum pulse width (100 μs), minimum height (to  
<300 μV)  
Minimum pulse width (100 μs), maximum height (up to  
1.0 V)  
Maximum pulse width (2 ms), minimum height (to  
<300 μV)  
Maximum pulse width (2 ms), maximum height (up to  
1.0 V)  
There is an inherent one-frame-period uncertainty in the exact  
location of the pace trailing edge.  
PACE DETECTION VIA SECONDARY SERIAL  
INTERFACE  
The ADAS1000-3/ADAS1000-4 provide a second serial  
interface for users to implement their own pace detection  
schemes. This interface is configured as a master interface. It  
provides ECG data at the 128 kHz data rate only. The purpose  
of this interface is to allow the user to access the ECG data at a  
rate sufficient to allow them to run their own pace algorithm,  
while maintaining all the filtering and decimation of the ECG  
data that the ADAS1000-3/ADAS1000-4 offer on the standard  
serial interface (2 kHz and 16 kHz data rates). This dedicated  
pace interface uses three of the four GPIO pins, leaving one  
GPIO pin available even when the secondary serial interface is  
enabled. Note that the on-chip digital calibration to ensure  
channel gain matching does not apply to data that is available  
on this interface. This interface is discussed in more detail in  
the Secondary Serial Interface section.  
These scenarios passed with acceptable results. The use of the  
ac lead-off function had no obvious impact on the recorded  
pace height, width, or the ability of the pace detection algorithm  
to identify a pace pulse. The pace algorithm was also evaluated  
with the respiration carrier enabled; again, no differences in the  
threshold or pacer detect were noted from the carrier.  
While these experiments validate the pace algorithm over a  
confined set of circumstances and conditions, they do not  
replace end system verification of the pacer algorithm. This  
can be performed in only the end system, using the system  
manufacturers specified cables and validation data set.  
EVALUATING PACE DETECTION PERFORMANCE  
ECG simulators offer a convenient means of studying the perfor-  
mance and ability of the ADAS1000-4 to capture pace signals  
over the range of widths and heights defined by the various  
regulatory standards. While the pace detection algorithm of the  
ADAS1000 is designed to conform to medical instrument  
standards (pace widths of 100 μs to 2.00 ms and with  
amplitudes of <400 μV to >1000 mV), some simulators put out  
signals wider or narrower than called for in the standards. The  
pace detection algorithm has been designed to measure a  
maximum pace widths of 2 ms with a margin of 0.25 ms to  
allow for simulator variations.  
Rev. B | Page 45 of 80  
 
 
 
 
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
The amount of decimation depends on the selected data rate,  
with more decimation for the lower data rates.  
FILTERING  
Figure 72 shows the ECG digital signal processing. The ADC  
sample rate is programmable. In high performance mode, it  
is 2.048 MHz; in low power mode, the sampling rate is reduced  
to 1.024 MHz. The user can tap off framing data at one of three  
data rates, 128 kHz, 16 kHz, or 2 kHz. Note that although the  
data-word width is 24 bits for the 2 kHz and 16 kHz data rate,  
the usable bits are 19 and 18, respectively.  
Four selectable low-pass filter corners are available at the 2 kHz  
data rate.  
Filters are cleared by a reset. Table 16 shows the filter latencies  
at the different data rates.  
AC LEAD-OFF  
DETECTION  
2.048MHz  
ADC DATA  
PACE  
14-BITS  
128kHz  
–3dB AT 13kHz  
DETECTION  
2.048MHz  
AVAILABLE DATA RATE  
CHOICE OF 1:  
128kHz DATA RATE  
16-BITS WIDE  
ACLO  
CARRIER  
NOTCH  
2kHz  
128kHz  
16kHz DATA RATE  
24-BITS WIDE  
18 USABLE BITS  
16kHz  
–3dB AT 3.5kHz  
16kHz  
2kHz DATA RATE  
24-BITS WIDE  
2kHz  
–3dB AT 450Hz  
19 USABLE BITS  
40Hz  
150Hz  
CALIBRATION  
31.25Hz DATA RATE  
24-BITS WIDE  
250Hz  
(PROGRAMMABLE BESSEL)  
~22 USABLE BITS  
~7Hz  
Figure 72. ECG Channel Filter Signal Flow  
Table 16. Relationship of ECG Waveform to Pace Indication1, 2, 3  
Data Rate  
Conditions  
Apparent Delay of ECG Data Relative to Pace Event4  
2 kHz  
450 Hz ECG bandwidth  
250 Hz ECG bandwidth  
150 Hz ECG bandwidth  
40 Hz ECG bandwidth  
0.984 ms  
1.915 ms  
2.695 ms  
7.641 ms  
109 μs  
16 kHz  
128 kHz  
0
1 ECG waveform delay is the time required to reach 50% of final value following a step input.  
2 Guaranteed by design, not subject to production test.  
3 There is an unavoidable residual uncertainty of 8 μs in determining the pace pulse trailing edge.  
4 Add 38 μs to obtain the absolute delay for any setting.  
Rev. B | Page 46 of 80  
 
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
setting Bit 4, gang, in the same register. When a device is  
configured as a master, the SYNC_GANG pin is automati-  
cally set as an output.  
VOLTAGE REFERENCE  
The ADAS1000-3/ADAS1000-4 have a high performance,  
low noise, on-chip 1.8 V reference for use in the ADC and  
DAC circuits. The REFOUT of one device is intended to  
drive the REFIN of the same device. The internal reference  
is not intended to drive significant external current; for  
optimum performance in gang operation with multiple  
devices, each device should use its own internal reference.  
When a device is configured as a slave (ADAS1000-2), the  
SYNC_GANG and CLK_IO pins are set as inputs.  
Synchronizing Devices  
The ganged devices need to share a common clock to ensure  
that conversions are synchronized. One approach is to drive  
the slave CLK_IO pins from the master CLK_IO pin. Alter-  
natively, an external 8.192 MHz clock can be used to drive  
the CLK_IO pins of all devices. The CLK_IO powers up high  
impedance until configured in gang mode.  
An external 1.8 V reference can be used to provide the  
required VREF. In such cases, there is an internal buffer  
provided for use with external reference. The REFIN pin is a  
dynamic load with an average input current of approximately  
100 μA per enabled channel, including respiration. When the  
internal reference is used, the REFOUT pin requires  
decoupling with a 10 μF capacitor with low ESR (0.2 Ω  
maximum) in parallel with 0.1 μF capacitor to REFGND,  
these capacitors should be placed as close to the device pins  
as possible and on the same side of the PCB as the device.  
In addition, the SYNC_GANG pin is used to synchronize  
the start of the ADC conversion across multiple devices. The  
SYNC_GANG pin is automatically driven by the master and  
is an input to all the slaves. SYNC_GANG is in high  
impedance until enabled via gang mode.  
When connecting devices in gang mode, the SYNC_GANG  
output is triggered once when the master device starts to  
convert. Therefore, to ensure that the slave device(s) receive  
this synchronization signal, configure the slave device first  
for operation and enable conversions, followed by issuing the  
conversion signal to the ECGCTL register in the master  
device.  
GANG MODE OPERATION  
Increasing the number of ECG channels enables the user to  
measure an increased number of patient electrodes. Typically  
a 12-lead system would require nine electrodes (and one right  
leg drive reference electrode), but a derived arrangement is  
possible by using just eight electrodes (and one right leg drive  
reference electrode). As such, mating a 5-electrode ADAS1000,  
ADAS1000-1, or ADAS1000-2 with either a ADAS1000-3 or  
ADAS1000-4 device delivers the required eight electrodes.  
The approach used is a master slave arrangement, where one  
device is designated as master, and any others are designated  
as slaves. It is important that multiple devices operate well  
together; with this in mind, the pertinent inputs/outputs to  
interface between master and slave devices have been made  
available.  
MASTER  
SLAVE 0  
CLK_IO  
CLK_IO  
SYNC_GANG  
CM_OUT  
SYNC_GANG  
CM_IN  
CAL_DAC_IO  
CAL_DAC_IO  
SLAVE 1  
CLK_IO  
SYNC_GANG  
CM_IN  
Note that when using multiple devices, the user must collect  
the ECG data directly from each device. If using a traditional  
12-lead arrangement where the Vx leads are measured  
relative to WCT, the user should configure the master device  
in lead mode with the slave device configured for electrode  
mode. The LSB size for electrode and lead data differs (see  
Table 43 for details).  
CAL_DAC_IO  
Figure 73. Master/Slave Connections in Gang Mode, Using Multiple  
Devices  
Calibration  
The calibration DAC signal from one device (master) can be  
output on the CAL_DAC_IO pin and used as the calibration  
input for other devices (slaves) when used in the gang mode  
of operation. This ensures that they are all being calibrated  
using the same signal which results in better matching across  
channels. This does not happen automatically in gang mode  
but, rather, must be configured via Table 36.  
In gang mode, all devices must be operated in the same  
power mode (either high performance or low power) and  
the same data rate.  
Master/Slave  
Any of the ADAS1000, ADAS1000-1, ADAS1000-3, or  
ADAS1000-4 can be configured as a master or slave, while  
the ADAS1000-2 can only be configured as a slave. A device  
is selected as a master or slave using Bit 5, master, in the  
ECGCTL register (see Table 28). Gang mode is enabled by  
Rev. B | Page 47 of 80  
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
Common Mode  
Right Leg Drive  
The ADAS1000-3/ADAS1000-4 have a dedicated CM_OUT  
pin serving as an output and a CM_IN pin as an input. In  
gang mode, the master device determines the common-mode  
voltage based on the selected input electrodes. This common-  
mode signal (on CM_OUT) can then be used by subsequent  
slave devices (applied to CM_IN) as the common-mode  
reference. All electrodes within the slave device are then  
measured with respect to the CM_IN signal from the master  
device. See the CMREFCTL register in Table 32 for more  
details on the control via the serial interface. Figure 74 shows  
the connections between a master and slave device using  
multiple devices.  
The right leg drive comes from the master device. If the  
internal RLD resistors of the slave device are to contribute  
to the RLD loop, tie the RLD_SJ pins of master and slave  
together.  
Sequencing Devices into Gang Mode  
When entering gang mode with multiple devices, both  
devices can be configured for operation, but the conver-  
sion enable bit (ECGCTL register, Bit 2, Table 28) of the  
master device should be set after the conversion enable bit  
of the slave device. When the master device conversion  
signal is set, the master device generates one edge on its  
SYNC_GANG pin. This applies to any slave SYNC_GANG  
inputs, allowing the devices to synchronize ADC conversions.  
Rev. B | Page 48 of 80  
Data Sheet  
ADAS1000-3/ADAS1000-4  
CM_OUT/  
WCT  
REFOUT  
VREF  
RLD_OUT  
SHIELD  
AVDD  
IOVDD  
REFIN  
CAL_DAC_IO  
DRIVEN  
LEAD AMP  
ADCVDD  
(optional)  
ADCVDD, DVDD  
1.8V  
REGULATORS  
CALIBRATION  
DAC  
SHIELD  
DRIVE  
AMP  
DVDD  
(optional)  
VCM_REF  
(1.3V)  
RESPIRATION  
DAC  
ADAS1000-4  
COMMON-  
MODE AMP  
AC  
LEAD-OFF  
DAC  
LEAD-OFF  
DETECTION  
10kΩ  
PACE  
DETECTION  
CS  
MUXES  
SCLK  
SDI  
SDO  
DRDY  
3 × ECG PATH  
AMP  
TAKE LEAD  
DATA  
ELECTRODES  
×3  
FILTERS,  
CONTROL,  
AND  
INTERFACE  
LOGIC  
ADC  
SYNC_GANG  
EXT RESP_LA  
EXT RESP LL  
EXT RESP_RA  
AMP  
ADC  
CLK_IO  
CLOCK GEN/OSC/  
EXTERNAL CLK  
SOURCE  
RESPIRATION PATH  
XTAL1 XTAL2  
AVDD  
IOVDD  
REFIN  
REFOUT  
VREF  
CAL_DAC_IN  
RLD_SJ  
CM_IN  
ADCVDD  
(optional)  
ADCVDD, DVDD  
1.8V  
REGULATORS  
DVDD  
(optional)  
VCM_REF  
(1.3V)  
ADAS1000-2  
SLAVE  
AC  
COMMON-  
MODE AMP  
LEAD-OFF  
DAC  
LEAD-OFF  
DETECTION  
PACE  
DETECTION  
CS  
MUXES  
SCLK  
SDI  
SDO  
DRDY  
SYNC_GANG  
5 × ECG PATH  
AMP  
TAKE  
ELECTRODE  
DATA  
FILTERS,  
CONTROL,  
AND  
INTERFACE  
LOGIC  
ELECTRODES  
×5  
ADC  
CLK_IO  
CLOCK GEN/OSC/  
EXTERNAL CLK  
SOURCE  
Figure 74. Configuring Multiple Devices to Extend Number of Electrodes/Leads  
(This example uses ADAS1000-4 as master and ADAS1000-2 as slave; other arrangements possible.)  
Table 17. Some Possible Arrangements for Gang Operation  
Master  
Slave 1  
Slave 2  
Features  
Number of Electrodes  
10 ECG, CM_IN, RLD  
15 ECG, CM_IN, RLD  
8 ECG, CM_IN, RLD  
8 ECG, CM_IN, RLD  
8 ECG, CM_IN, RLD  
Number of Leads  
ADAS1000  
ADAS1000  
ADAS1000  
ADAS1000-3 ADAS1000-2  
ADAS1000-4 ADAS1000-2  
ADAS1000-2  
ECG, respiration, pace  
12-lead + spare ADC channel  
15-lead + 3 spare ADC channels  
12-lead (derived leads)  
12-lead (derived leads)  
12-lead (derived leads)  
ADAS1000-2 ADAS1000-2 ECG, respiration, pace  
ADAS1000-3  
ECG, respiration, pace  
ECG  
ECG, respiration, pace  
Rev. B | Page 49 of 80  
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
have the relevant synchronized data. Alternative methods  
might use individual controllers for each device or separate  
SDO paths.  
INTERFACING IN GANG MODE  
As shown in Figure 74, when using multiple devices, the  
user must collect the ECG data directly from each device.  
The example shown in Figure 75 illustrates one possibility  
of how to approach interfacing to a master and slave device.  
For some applications, digital isolation is required between  
the host and the ADAS1000-3/ADAS1000-4. The example  
shown illustrates a means to ensure that the number of lines  
requiring isolation is minimized.  
Note that SCLK, SDO, and SDI are shared here with  
CS  
individual  
lines. This requires the user to read the data  
on both devices twice as fast to ensure that they can capture  
all the data to maintain the chosen data rate and ensure they  
SCLK  
SDI  
MICROCRONTROLLER/  
CS1  
DSP  
CS2  
SDO  
MASTER  
SCLK  
SLAVE  
SCLK  
SDI  
SDI  
CS  
CS  
DRDY (Optional)  
DRDY (Optional)  
SDO  
SDO  
Figure 75. One Method of Interfacing to Multiple Devices  
Rev. B | Page 50 of 80  
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
SERIAL INTERFACES  
The ADAS1000-3/ADAS1000-4 are controlled via a standard  
serial interface allowing configuration of registers and readback  
of ECG data. This is an SPI-compatible interface that can  
operate at SCLK frequencies up to 40 MHz.  
MICROCONTROLLER/  
DSP  
ADAS1000-3/  
ADAS1000-4  
SCLK  
CS  
SCLK  
CS  
MOSI  
MISO  
GPIO  
SDI  
SDO  
DRDY  
The ADAS1000-3/ADAS1000-4 also provide an optional  
secondary serial interface that is capable of providing ECG  
data at the 128 kHz data rate for users wishing to apply their  
own digital pace detection algorithm. This is a master interface  
that operates with an SCLK of 20.48 MHz.  
Figure 76. Serial Interface  
Write Mode  
The serial word for a write is 32 bits long, MSB first. The serial  
interface works with both a continuous and a burst (gated)  
STANDARD SERIAL INTERFACE  
CS  
serial clock. The falling edge of  
starts the write cycle.  
The standard serial interface is LVTTL-compatible when oper-  
ating from a 2.3 V to 3.6 V IOVDD supply. This is the primary  
interface for controlling the ADAS1000-3/ADAS1000-4 reading  
and writing registers, and reading frame data containing all the  
ECG data-words and other status functions within the device.  
Serial data applied to SDI is clocked into the ADAS1000-3/  
ADAS1000-4 on rising SCLK edges. At least 32 rising clock  
edges must be applied to SCLK to clock in 32 bits of data  
CS  
before  
updated on the rising edge of . For another serial transfer to  
CS  
is taken high again. The addressed input register is  
CS  
must be taken low again. Register writes are used  
The SPI is controlled by the following five pins:  
take place,  
CS  
CS  
(frame synchronization input). Asserting  
low  
to configure the device. Once the device is configured and  
enabled for conversions, frame data can be initiated to start  
clocking  
CS  
selects the device. When  
is high, data on the SDI  
CS  
pin is ignored. If  
disabled so that multiple SPI devices can share a common  
CS  
is inactive, the SDO output driver is  
out ECG data on SDO at the programmed data rate. Normal  
operation for the device is to send out frames of ECG data.  
Typically, register reads and writes should be needed only  
during start-up configuration. However, it is possible to write  
new configuration data to the device while in framing mode.  
A new write command is accepted within the frame and,  
depending on the nature of the command, there may be a  
need to flush out the internal filters (wait periods) before  
seeing usable framing data again.  
SDO pin. The  
pin can be tied low to reduce the number  
CS  
of isolated paths required. When  
is tied low, there is no  
frame around the data-words; therefore, the user must be  
aware of where they are within the frame. All data-words  
with 2 kHz and 16 kHz data rates contain register addresses  
at the start of each word within the frame. Users can  
resynchronize the interface by holding SDI high for 64  
SCLK cycles, followed by a read of any register so that  
SDI is brought low for the first bit of the following word.  
SDI (serial data input pin). Data on SDI is clocked into the  
device on the rising edges of SCLK.  
Write/Read Data Format  
Address, data, and the read/write bits are all in the same word.  
Data is updated on the rising edge of  
following word. For all write commands to the ADAS1000-3/  
ADAS1000-4, the data-word is 32 bits, as shown in Table 18.  
Similarly, when using data rates of 2 kHz and 16 kHz, each  
word is 32 bits (address bits and data bits).  
CS  
or the first cycle of the  
SCLK (clocks data in and out of the device). SCLK should  
CS  
idle high when  
is high.  
SDO (serial data output pin for data readback). Data is  
shifted out on SDO on the falling edges of SCLK. The  
CS  
SDO output driver is high-Z when  
is high.  
DRDY  
(data ready, optional). Data ready when low,  
Table 18. Serial Bit Assignment (Applies to All Register  
Writes, 2 kHz and 16 kHz Reads)  
busy when high. Indicates the internal status of the  
ADAS1000-3/ADAS1000-4 digital logic. It is driven  
high/busy during reset. If data frames are enabled and  
the frame buffer is empty, this pin is driven busy/high.  
If the frame buffer is full, this pin is driven low/ready.  
If data frames are not enabled, this pin is driven low to  
indicate that the device is ready to accept register  
read/write commands. When reading packet data, the  
B31  
[B30:B24]  
[B23:B0]  
R/W  
Address Bits[6:0]  
Data Bits[23:0] (MSB first)  
For register reads, data is shifted out during the next word, as  
shown in Table 19.  
Table 19. Read/Write Data Stream  
Digital  
Pin  
DRDY  
entire packet must be read to allow the  
back high.  
return  
Command 1  
Command 2  
Command 2  
SDI  
SDO  
Read Address 1  
Read Address 2  
Address 1  
Write Address 3  
Address 2  
Read Data 1  
Read Data 2  
Rev. B | Page 51 of 80  
 
 
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
In the 128 kHz data rate, all write words are still 32-bit writes  
but the read words in the data packet are now 16 bits (upper  
16 bits of register). There are no address bits, only data bits.  
Register space that is larger than 16 bits spans across 2 ×  
16-bit words (for example, pace and respiration).  
Read Mode  
Although the primary reading function within the ADAS1000-3/  
ADAS1000-4 is the output of the ECG frame data, the devices  
also allow reading of all configuration registers. To read a  
register, the user must first address the device with a read  
command containing the particular register address. If the  
device is already in data framing mode, the read register  
command can be interleaved between the frames by issuing a  
read register command during the last word of frame data.  
Data shifted out during the next word is the register read data.  
To return to framing mode, the user must re-enable framing  
by issuing a read of the frame header register (Address 0x40)  
(see Table 53). This register write can be used to flush out the  
register contents from the previous read command.  
Data Frames/Packets  
The general data packet structure is shown in Table 18. Data  
can be received in two different frame formats. For the 2 kHz  
and 16 kHz data rates, a 32-bit data format is used (where the  
register address is encapsulated in the upper byte, identifying  
the word within the frame) (see Table 22). For the 128 kHz data  
rate, words are provided in 16-bit data format (see Table 23).  
When the configuration is complete, the user can begin reading  
frames by issuing a read command to the frame header register  
(see Table 53). The ADAS1000-3/ADAS1000-4 continue to  
make frames available until another register address is written  
(read or write command). To continue reading frame data,  
continue to write all zeros on SDI, which is a write of the  
NOP register (Address 0x00). A frame is interrupted only  
when another read or write command is issued.  
Table 20. Example of Reading Registers and Frames  
….. NOP  
Read  
Address  
N
Read  
frames  
NOP  
NOP  
…..  
…..  
SDI  
Frame  
data  
Frame  
CRC  
Register  
Data N  
Frame  
header  
Frame  
data  
SDO …..  
Regular register reads are always 32 bits long and MSB first.  
CS  
Each frame can be a large amount of data plus status words.  
Serial Clock Rate  
can toggle between each word of data within a frame, or it can  
be held constantly low during the entire frame.  
The SCLK can be up to 40 MHz, depending on the IOVDD  
voltage level as shown in Table 5. The minimum SCLK  
frequency is set by the requirement that all frame data be  
clocked out before the next frame becomes available.  
Reading all the data-words creates a frame containing 10 × 32  
bit words when reading at 2 kHz or 16 kHz data rates; similarly,  
a frame contains 13 × 16-bit words when reading at 128 kHz.  
Additionally any words not required can be excluded from  
the frame. To arrange the frame with the words of interest,  
configure the appropriate bits in the frame control register (see  
Table 37). The complete set of words per frame are 10 × 32-bit  
words for the 2 kHz or 16 kHz data rates, or 13 × 16-bit words  
at 128 kHz.  
SCLK (min) = frame_rate × words_per_frame × bits_per_word  
The minimum SCLK for the various frame rates is shown in  
Table 21.  
Table 21. SCLK Clock Frequency vs. Packet Data/Frame Rates  
Frame  
Rate  
Word  
Size  
Maximum  
Minimum  
SCLK  
Words/Frame1  
Any data not available within the frame can be read between  
frames. Reading a register interrupts the frame and requires  
the user to issue a new read command of Address 0x40 (see  
Table 53) to start framing again.  
128 kHz  
16 kHz  
2 kHz  
16 bits  
32 bits  
32 bits  
13 words  
10 words  
10 words  
26.62 MHz  
5.12 MHz  
640 kHz  
1 This is the full set of words that a frame contains. It is programmable and can  
be configured to provide only the words of interest. See Table 37.  
Table 22. Default 2 kHz and 16 kHz Data Rate: 32-Bit Frame Word Format  
Register  
Header  
Lead I/LA  
Lead II/LL  
Lead III/RA  
PACE  
RESPM  
0x1B  
RESPPH  
0x1C  
LOFF  
0x1D  
GPIO  
0x06  
CRC  
Address  
0x40  
0x11  
0x12  
0x13  
0x1A  
0x41  
Table 23. Default 128 kHz Data Rate: 16-Bit Frame Word Format  
Header Lead I/LA Lead II/LL Lead III/RA  
Address 0x40  
0x11 0x12 0x13  
Register  
PACE1 PACE2 RESPM1 RESPM2 RESPH1 RESPH2 LOFF GPIO CRC  
0x1D 0x06 0x41  
0x1A  
0x1B  
0x1C  
Rev. B | Page 52 of 80  
 
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
Internal operations are synchronized to the internal master  
clock at either 2.048 MHz or 1.024 MHz (ECGCTL[3]: HP = 1  
and HP = 0, respectively, see Table 28). Because there is no  
guaranteed relationship between the internal clock and the  
SPI's SCLK signal, an internal handshaking scheme is used to  
ensure safe data transfer between the two clock domains. A full  
handshake requires three internal clock cycles and imposes an  
upper speed limit on the SCLK frequency when reading frames  
with small word counts. This is true for all data frame rates.  
When reading packets of data, the entire data packet must be  
DRDY  
read; otherwise,  
stays low.  
DRDY  
There are three methods of detecting  
status.  
pin. This is an output pin from the ADAS1000-3/  
ADAS1000-4 that indicates the device read or busy status.  
DRDY  
DRDY  
No data is valid while this pin is high. The  
signals  
that data is ready to be read by driving low and remaining  
low until the entire frame has been read. It is cleared when  
the last bit of the last word in the frame is clocked onto  
SDO. The use of this pin is optional.  
SCLK (max) = (1.024 MHz × (1 + HP) × words_per_frame ×  
bits_per_word)/3; or 40 MHz, whichever is lower.  
SDO pin. The user can monitor the voltage level of the  
Exceeding the maximum SCLK frequency for a particular  
CS  
SDO pin by bringing  
low. If SDO is low, data is ready;  
DRDY  
operating mode causes erratic behavior in the  
and results in the loss of data.  
signal  
if high, busy. This does not require clocking the SCLK  
input. (CPHA = CPOL = 1 only).  
Data Rate and Skip Mode  
One of the first bits of valid data in the header word availa-  
ble on SDO is a data ready status bit (see Table 43). Within  
the configuration of the ADAS1000-3/ADAS1000-4, the  
user can set the header to repeat until the data is ready. See  
Bit 6 (RDYRPT) in the frame control register in Table 37.  
Although the standard frame rates available are 2 kHz, 16 kHz,  
and 128 kHz, there is also a provision to skip frames to further  
reduce the data rate. This can be configured in the frame  
control register (see Table 37).  
DRDY  
Data Ready (  
DRDY  
)
DRDY  
as  
The host controller must read the entire frame to ensure  
DRDY  
returns low and ready. If the host controller treats the  
The  
pin is used to indicate that a frame composed of  
decimated data at the selected data rate is available to read. It  
is high when busy and low when ready. Send commands only  
an edge triggered signal and then misses a frame or underruns,  
DRDY  
the  
remains high because there is still data available to  
DRDY  
read. The host controller must treat the  
signal as level  
DRDY  
when the status of  
is low or ready. During power-on, the  
is high (busy) while the device initializes itself.  
DRDY  
triggered, ensuring that whenever it goes low, it generates an  
interrupt which can initiate a SPI frame transfer. On completion  
DRDY  
status of  
When initialization is complete,  
goes low and the user  
DRDY  
of the transfer the  
returns high.  
can start configuring the device for operation. When the device  
is configured and enabled for conversions by writing to the  
conversion bit (CNVEN) in the ECGCTL register, the ADCs  
start to convert and the digital interface starts to make data  
available, loading them into the buffer when ready. If conver-  
sions are enabled and the buffer is empty, the device is not  
Detecting Missed Conversion Data  
To ensure that the current data is valid, the entire frame must  
be read at the selected data rate. If a read of the entire frame  
takes longer than the selected data rate allows, the internal  
buffer is not loaded with the latest conversion data. The frame  
header register (see Table 53) provides four settings to indicate  
an overflow of frame data. The settings of Bits[29:28] report  
how many frames have been missed since the last valid frame  
read. A missed frame may occur as a result of the last read  
taking too long. The data in the current frame is valid data,  
but it is not the current data. It is the calculation made directly  
after the last valid read.  
DRDY  
DRDY  
goes high. Once the buffer is full,  
ready and  
goes low to indicate that data is ready to be read out of the  
DRDY  
device. If the device is not enabled for conversions, the  
ignores the state of the buffer full status.  
To clear such an overflow, the user must read the entire frame.  
Rev. B | Page 53 of 80  
ADAS1000-3/ADAS1000-4  
Data Sheet  
CRC Word  
Framed data integrity is provided by CRCs. For the 128 kHz  
frame rates, the 16-bit CRC-CCITT polynomial is used. For the  
2 kHz and 16 kHz frame rates, the 24-bit CRC polynomial used.  
ADAS1000-3/  
ADAS1000-4  
XTAL2  
In both cases, the CRC residue is preset to all 1s and inverted  
before being transmitted. The CRC parameters are summarized  
in Table 24. To verify that data was correctly received, software  
should compute a CRC on both the data and the received  
checksum. If data and checksum are received correctly, the  
resulting CRC residue should equal the check constant shown  
in Table 24. Note that data is shifted through the generator  
polynomial MSB first, the same order that it is shifted out  
serially. The bit and byte order of the CRC that is appended to  
the frame is such that the MSB of the CRC is shifted through  
the generator polynomial first in the same order as the data so  
that the CRC residue XORd with the inverted CRC at the end  
of the frame is all 1s (which is why the check constant is  
identical for all messages). The CRC is based only on the  
data that is sent out.  
XTAL1  
CLK_IO  
Figure 77. Input Clock  
Clocks  
The ADAS1000-3/ADAS1000-4 run from an external crystal or  
clock input frequency of 8.192 MHz. The external clock input is  
provided for use in gang mode so conversions between the two  
devices are synchronized. In this mode, the CLK_IO pin is an  
output from the master and an input from the slave. To reduce  
power, the CLK_IO is disabled when not in gang mode.  
All features within the ADAS1000-3/ADAS1000-4 are a  
function of the frequency of the externally applied clock. Using  
a frequency other than the 8.192 MHz previously noted causes  
scaling of the data rates, filter corners, ac leads-off frequency,  
respiration frequency, and pace algorithm corners accordingly.  
Table 24. CRC Polynomials  
Polynomial  
Frame Rate  
2 kHz, 16 kHz 24 bits  
128 kHz 16 bits  
CRC Size Polynomial  
x24 + x22 + x20 + x19 + x18 + x16 + x14 + x13 + x11 + x10 + x8 + x7 + x6 + x3 + x1 + x0 0x15D6DCB 0x15A0BA  
x16 + x12 + x5 + x0  
0x11021 0x1D0F  
in Hex  
Check Constant  
Rev. B | Page 54 of 80  
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
frame with the data available on MSDO on the falling edge of  
SECONDARY SERIAL INTERFACE  
MCS  
MSCLK. MSCLK idles high when  
is deasserted.  
This second serial interface is an optional interface that  
can be used for the users own pace detection purposes.  
This interface contains ECG data at 128 kHz data rate only.  
If using this interface, the ECG data is still available on the  
standard interface discussed previously at lower rates with all  
the decimation and filtering applied. If this interface is inactive,  
it draws no power.  
The data format for this interface is fixed and not influenced by  
the FRMCTL register settings. All six words are output, even if  
the individual channels are not enabled.  
The header word consists of four bits of all 1s followed by a 12-  
bit sequence counter. This sequence counter increments after  
every frame is sent, thereby allowing the user to tell if any  
frames have been missed and how many.  
Data is available in 16-bit words, MSB first.  
This interface is a master interface, with the ADAS1000-3/  
RESET  
CS  
ADAS1000-4 providing the SCLK, , SDO. Is it shared  
There are two methods of resetting the ADAS1000-3/ADAS1000-4  
to power-on default. Bringing the  
across some of the existing GPIO pins as follows:  
RESET  
line low or setting  
GPIO1/MSCLK  
MCS  
the SWRST bit in the ECGCTL register (Table 28) resets the  
contents of all internal registers to their power-on reset state.  
GPIO0/  
GPIO2/MSDO  
RESET  
The falling edge of the  
pin initiates the reset  
goes high for the duration, returning low when  
process is complete. This sequence takes 1.5 ms  
DRDY  
process;  
This interface can be enabled via the GPIO register (see  
Table 33).  
RESET  
the  
DRDY  
maximum. Do not write to the serial interface while  
is  
returns low,  
RESET  
MICROCONTROLLER/  
ADAS1000-4  
DSP  
RESET  
DRDY  
command. When  
high handling a  
MASTER SPI  
normal operation resumes and the status of the  
pin is  
MSCLK/GPIO1  
MCS/GPIO0  
SCLK  
CS  
ignored until it goes low again. Software reset using the SWRST  
bit (see Table 28) requires that a NOP (no operation) command  
be issued to complete the reset cycle.  
MISO/GPIO  
MSDO/GPIO2  
Figure 78. Master SPI Interface for External Pace Detection Purposes  
PD FUNCTION  
The data format of the frame starts with a header word, three  
ECG data-words, two words filled with zeros and completes  
with the same CRC word as documented in Table 24 for the  
128 kHz rate. All words are 16 bits. MSCLK runs at approxi-  
PD  
The  
pin powers down all functions in low power mode.  
The digital registers maintain their contents. The power-down  
function is also available via the serial interface (ECG control  
register, see Table 28).  
MCS  
mately 20 MHz and the  
function is asserted for the entire  
Table 25. Master SPI Frame Format; All Words are 16 Bits  
Mode\Word  
1
2
3
4
5
6
7
Electrode mode1  
Analog lead mode1  
Header  
Header  
ECG1_LA  
LEAD I  
ECG2_LL  
LEAD III  
ECG3_RA  
−LEAD II (RA-LL)  
All 0’s  
All 0’s  
All 0’s  
All 0’s  
CRC  
CRC  
1 As set by the FRMCTL register data DATAFMT, Bit [4], see Table 37.  
Rev. B | Page 55 of 80  
 
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
SPI OUTPUT FRAME STRUCTURE (ECG AND STATUS DATA)  
Three data rates are offered for reading ECG data: low speed 2 kHz/16 kHz rates for electrode/lead data (32-bit words) and a high speed  
128 kHz for electrode/lead data (16-bit words).  
DRDY  
1
CS  
EACH SCLK WORD IS 32 CLOCK CYCLES  
1
2
3
4
5
6
7
8
9
SCLK  
SDO  
DRIVEN OUTPUT DATA STREAM  
2
L
F
A
A
N
N
R
O
D
ANOTHER FRAME OF DATA  
I
L
/
F
L
/
E
R
/
E
R
O
I
O
I
P
I
I
I
O
I
D
D
I
I
O
-
T
T
G
U
A
D
D
D
C
A
W
T
E
D
A
I
A
A
E
R
I
C
H
E
A
N
E
T
E
P
L
R
E
L
E
L
G
S
L
C
D
A
E
M
E
R
C
A
P
32-BIT  
DATA WORDS  
1
CS MAY BE USED IN ONE OF THE FOLLOWING WAYS:  
A) HELD LOW ALL THE TIME.  
B) USED TO FRAME THE ENTIRE PACKET OF DATA.  
C) USED TO FRAME EACH INDIVIDUAL 32-BIT WORD.  
FULL WORD COUNT = 10 (RESPIRATION PHASE EXCLUDED HERE). WORDS MAY BE EXCLUDED, SEE THE FRMCTL REGISTER.  
2
Figure 79. Output Frame Structure for 2 kHz and 16 kHz Data Rates with SDO Data Configured for Electrode or Lead Data  
DRDY  
1
CS  
EACH SCLK WORD IS 16 CLOCK CYCLES  
1
2
3
4
5
6
7
8
9
10  
11  
SCLK  
SDO  
DRIVEN OUTPUT DATA STREAM  
2
L
ANOTHER FRAME  
A
A
F
R
E
O
I
D
R
L
E
L
F
R
N
P
C
D
E
O
-
O
O
I
G
A
A
D
D
T
W
P
E
U
A
A
T
H
C
I
E
R
I
R
N
L
P
C
G
S
A
E
M
R
16-BIT  
DATA WORDS  
1
CS MAY BE USED IN ONE OF THE FOLLOWING WAYS:  
A) HELD LOW ALL THE TIME.  
B) USED TO FRAME THE ENTIRE PACKET OF DATA.  
C) USED TO FRAME EACH INDIVIDUAL 16-BIT WORD.  
FULL WORD COUNT = 13 (RESPIRATION PHASE EXCLUDED HERE). WORDS MAY BE EXCLUDED, SEE THE FRMCTL REGISTER.  
2
Figure 80. Output Frame Structure for 128 kHz Data Rate with SDO Data Configured for Electrode Data  
(The 128 kHz Data Rate Can Provide Single-Ended Electrode Data or Analog Lead Mode Data Only. Digital Lead Mode Is Not Available at 128 kHz Data Rate.)  
Rev. B | Page 56 of 80  
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
SPI REGISTER DEFINITIONS AND MEMORY MAP  
In 2 kHz and 16 kHz data rates, data takes the form of 32-bit words. Bit A6 to Bit A0 serve as word identifiers. Each 32-bit word has  
24 bits of data. A third high speed data rate is also offered: 128 kHz with data in the form of 16-bit words (all 16 bits as data).  
Table 26. SPI Register Memory Map  
1
R
/W  
A[6:0]  
0x00  
0x01  
0x02  
0x03  
0x04  
0x05  
0x06  
0x07  
0x08  
0x09  
0x0A  
0x0B  
0x0C  
0x0D  
0x0E  
0x0F  
0x11  
0x12  
0x13  
0x1A  
0x1B  
0x1C  
0x1D  
0x1E  
0x1F  
0x21  
0x22  
0x23  
0x31  
0x32  
0x33  
0x3A  
0x3B  
0x3C  
0x40  
0x41  
Other  
D[23:0]  
XXXXXX  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
XXXXXX  
XXXXXX  
XXXXXX  
XXXXXX  
XXXXXX  
XXXXXX  
XXXXXX  
XXXXXX  
XXXXXX  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
dddddd  
XXXXXX  
XXXXXX  
Register Name  
Table  
Register Description  
NOP (no operation)  
Reset Value  
0x000000  
0x000000  
0x000000  
0x000000  
0x000F88  
0xE00000  
0x000000  
0x242424  
0x000000  
0x002000  
0x079000  
0x000000  
0x00FFFF  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x000000  
0x800000  
0xFFFFFF  
XXXXXX  
R
NOP  
R/W  
R/W  
R/W  
R/W  
R/W  
R/W  
R/W  
R/W  
R/W  
R/W  
R/W  
R/W  
R/W  
R/W  
R/W  
R
ECGCTL  
ECG control  
Table 28  
Table 29  
Table 30  
Table 31  
Table 32  
Table 33  
Table 34  
Table 35  
Table 36  
Table 37  
Table 38  
Table 39  
Table 40  
Table 41  
Table 42  
Table 43  
Table 43  
Table 43  
Table 44  
Table 45  
Table 46  
Table 47  
Table 48  
Table 49  
Table 50  
Table 50  
Table 50  
Table 51  
Table 51  
Table 51  
Table 52  
Table 52  
Table 52  
Table 53  
Table 54  
LOFFCTL  
RESPCTL  
PACECTL  
CMREFCTL  
GPIOCTL  
PACEAMPTH  
TESTTONE  
CALDAC  
FRMCTL  
Lead-off control  
Respiration control2  
Pace detection control  
Common-mode, reference, and shield drive control  
GPIO control  
Pace amplitude threshold2  
Test tone  
Calibration DAC  
Frame control  
FILTCTL  
Filter control  
LOFFUTH  
LOFFLTH  
PACEEDGETH  
PACELVLTH  
LADATA  
AC lead-off upper threshold  
AC lead-off lower threshold  
Pace edge threshold2  
Pace level threshold2  
LA or Lead I data  
R
LLDATA  
LL or Lead II data  
R
RADATA  
PACEDATA  
RESPMAG  
RESPPH  
RA or Lead III data  
Read pace detection data/status2  
Read respiration data—magnitude2  
Read respiration data—phase2  
Lead-off status  
R
R
R
R
LOFF  
R
DCLEAD-OFF  
OPSTAT  
DC lead-off  
R
Operating state  
R/W  
R/W  
R/W  
R
CALLA  
User gain calibration LA  
User gain calibration LL  
User gain calibration RA  
Lead-off amplitude for LA  
Lead-off amplitude for LL  
Lead-off amplitude for RA  
Pace1 width and amplitude2  
Pace2 width and amplitude2  
Pace3 width and amplitude2  
Frame header  
CALLL  
CALRA  
LOAMLA  
LOAMLL  
LOAMRA  
PACE1DATA  
PACE2DATA  
PACE3DATA  
FRAMES  
R
R
R
R
R
R
R
CRC  
Reserved3  
Frame CRC  
x
Reserved  
1 R/W = register both readable and writable; R = read only.  
2 ADAS1000-4 model only, ADAS1000-3 model does not contain these features.  
3 Reserved bits in any register are undefined. In some cases, a physical (but unused) memory bit may be present—in other cases not. Do not issue commands to  
reserved registers/space. Read operations of unassigned bits are undefined.  
Rev. B | Page 57 of 80  
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
CONTROL REGISTERS DETAILS  
For each register address, the default setting is noted in a default column in addition to being noted in the function column by “(default)”;  
this format applies throughout the register map.  
Table 27. Serial Bit Assignment  
B31  
[B30:B24]  
[B23:B0]  
R/W  
Address bits  
Data bits (MSB first)  
Table 28. ECG Control Register (ECGCTL) Address 0x01, Reset Value = 0x000000  
R/W  
Default  
Bit  
23  
22  
21  
Name  
LAEN  
LLEN  
Function  
R/W  
R/W  
R/W  
0
0
0
ECG channel enable; shuts down power to the channel; the input becomes high-Z.  
0 (default) = disables ECG channel. When disabled, the entire ECG channel is shut down and dissipates  
minimal power.  
1 = enables ECG channel.  
RAEN  
0
0
[20:11] Reserved  
Reserved, set to 0.  
R/W  
R/W  
10  
CHCONFIG  
GAIN [1:0]  
Setting this bit selects the differential analog front-end (AFE) input. See Figure 56.  
0 (default) = single-ended input (digital lead mode or electrode mode).  
1 = differential input (analog lead mode).  
R/W  
00  
[9:8]  
Preamplifier and anti-aliasing filter overall gain.  
00 (default) = GAIN 0 = ×1.4.  
01 = GAIN 1 = ×2.1.  
10 = GAIN 2 = ×2.8.  
11 = GAIN 3 = ×4.2 (user gain calibration is required for this gain setting).  
R/W  
R/W  
0
0
7
6
VREFBUF  
CLKEXT  
VREF buffer enable.  
0 (default) = disabled.  
1 = enabled (when using the internal VREF, VREFBUF must be enabled).  
Use external clock instead of crystal oscillator. The crystal oscillator is automatically disabled if  
configured as a slave in gang mode and the slave device should receive the clock from the master  
device.  
0 (default) = XTAL is clock source.  
1 = CLK_IO is clock source.  
R/W  
0
5
Master  
In gang mode, this bit selects the master (SYNC_GANG pin is configured as an output). When in single  
channel mode (gang = 0), this bit is ignored.  
0 (default) = slave.  
1 = master.  
R/W  
R/W  
0
0
4
3
Gang  
HP  
Enable gang mode. Setting this bit causes CLK_IO and SYNC_GANG to be activated.  
0 (default) = single channel mode.  
1 = gang mode.  
Selects the noise/power performance. This bit controls the ADC sampling frequency. See the  
Specifications section for further details. This bit also affects the respiration carrier frequency as  
discussed in the Respiration Carrier Frequency section.  
0 (default) = 1 MSPS, low power.  
1 = 2 MSPS, high performance/low noise.  
R/W  
R/W  
0
0
2
1
CNVEN  
PWREN  
Conversion enable. Setting this bit enables the ADC conversion and filters.  
0 (default) = idle.  
1 = conversion enable.  
Power enable. Clearing this bit powers down the device. All analog blocks are powered down and the  
external crystal is disabled. The register contents are retained during power down as long as DVDD is  
not removed.  
0 (default) = power down.  
1 = power enable.  
R/W  
0
0
SWRST  
Software reset. Setting this bit clears all registers to their reset value. This bit automatically clears itself.  
The software reset requires a NOP command to complete the reset.  
0 (default) = NOP.  
1 = reset.  
Rev. B | Page 58 of 80  
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
Table 29. Lead-Off Control Register (LOFFCTL) Address 0x02, Reset Value = 0x000000  
R/W  
R/W  
R/W  
R/W  
R/W  
R/W  
Default Bit  
Name  
LAPH  
LLPH  
RAPH  
Function  
0
0
0
0
0
23  
22  
21  
AC lead-off phase.  
0 (default) = in phase.  
1 = 180° out of phase.  
[20:19] Reserved  
Reserved, set to 0.  
18  
CEPH  
AC lead-off phase.  
0 (default) = in phase.  
1 = 180° out of phase.  
R/W  
R/W  
R/W  
0
0
0
17  
16  
15  
LAACLOEN  
LLACLOEN  
RAACLOEN  
Individual electrode ac lead-off enable. AC lead-off enables are the OR of ACSEL and the  
individual ac lead-off channel enables.  
0 (default) = ac lead-off disabled.  
1 = ac lead-off enabled.  
R/W  
R/W  
0
0
[14:13] Reserved  
Reserved, set to 0.  
12  
CEACLOEN  
Individual electrode ac lead-off enable. AC lead-off enables are the OR of ACSEL and the  
individual ac lead-off channel enables.  
0 (default) = ac lead-off disabled.  
1 = ac lead-off enabled.  
R/W  
R/W  
0
[11:9]  
[8:7]  
Reserved  
Reserved, set to 0.  
00  
ACCURRENT Set current level for ac lead-off.  
00 (default) = 12.5 nA rms.  
01 = 25 nA rms.  
10 = 50 nA rms.  
11 = 100 nA rms.  
R/W  
R/W  
00  
[6:5]  
[4:2]  
Reserved  
Reserved, set to 0.  
000  
DCCURRENT Set current level for dc lead-off (active only for ACSEL = 0).  
000 (default) = 0 nA.  
001 = 10 nA.  
010 = 20 nA.  
011 = 30 nA.  
100 = 40 nA.  
101 = 50 nA.  
110 = 60 nA.  
111 = 70 nA.  
R/W  
0
1
ACSEL  
DC or AC (out-of-band) lead-off detection.  
ACSEL acts as a global ac lead-off enable for RA, LL, LA, electrodes (CE ac lead-off is not  
enabled using ACSEL). AC lead-off enables are the OR of ACSEL and the individual ac lead-off  
channel enables.  
If LOFFEN = 0, this bit is don’t care.  
If LOFFEN = 1,  
0 (default) = dc lead-off detection enabled. (Individual ac lead-off can be enabled through  
Bits[17:12].)  
1 = dc lead-off detection disabled. AC lead-off detection enabled (all electrodes except CE  
electrode).  
When the calibration DAC is enabled, ac lead-off is disabled.  
R/W  
0
0
LOFFEN  
Enable lead-off detection.  
0 (default) = lead-off disabled.  
1 = lead-off enabled.  
Rev. B | Page 59 of 80  
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
Table 30. Respiration Control Register (RESPCTL) Address 0x03, Reset Value = 0x0000001  
R/W  
Default Bit  
Name  
Function  
[23:17] Reserved  
Reserved, set to 0.  
R/W  
0
0
16  
15  
RESPALTFREQ  
Setting this bit to 1 makes the respiration waveform on the GPIO3 pin periodic every cycle. Use in  
conjunction with RESFREQ to select drive frequency.  
0 (default) = periodic every N cycles (default).  
1 = periodic every cycle.  
R/W  
R/W  
RESPEXTSYNC  
RESPEXTAMP  
Set this bit to 1 to drive the MSB of the respiration DAC out onto the GPIO3 pin. This signal can be  
used to synchronize an external generator to the respiration carrier. It is a constant period only  
when RESPALTFREQ = 1.  
0 (default) = normal GPIO3 function.  
1 = MSB of RESPDAC driven onto the GPIO3 pin.  
0
14  
For use with an external instrumentation amplifier with respiration circuit. Bypasses the on-chip  
amplifier stage and input directly to the ADC. See Figure 69.  
0 (default) = disabled.  
1 = enabled.  
R/W  
R/W  
R/W  
0
13  
RESPOUT  
RESPCAP  
Selects external respiration drive output. RESPDAC_RA is automatically selected when RESPCAP = 1  
0 (default) = RESPDAC_LL and RESPDAC_RA.  
1 = RESPDAC_LA and RESPDAC_RA.  
0
12  
Selects source of respiration capacitors.  
0 (default) = use internal capacitors.  
1 = use external capacitors.  
0000  
[11:8]  
RESPGAIN [3:0] Respiration in amp gain (saturates at 10).  
0000 (default) = ×1 gain.  
0001 = ×2 gain.  
0010 = ×3 gain.  
1000 = ×9 gain.  
1001 = ×10 gain.  
11xx = ×10 gain.  
R/W  
R/W  
0
7
RESPEXTSEL  
RESPSEL [1:0]  
Selects between EXT_RESP _LA or EXT_RESP_LL paths. Applies only if the external respiration is  
selected in RESPSEL. EXT_RESP_RA is automatically enabled.  
0 (default) = EXT_RESP_LL.  
1 = EXT_RESP_LA.  
00  
[6:5]  
Set leads for respiration measurement.  
00 (default) = Lead I.  
01 = Lead II.  
10 = Lead III.  
11 = external respiration path.  
R/W  
R/W  
00  
00  
[4:3]  
[2:1]  
RESPAMP  
RESPFREQ  
Set the test tone amplitude for respiration drive signal.  
00 (default) = amplitude/8.  
01 = amplitude/4.  
10 = amplitude/2.  
11 = amplitude.  
Set frequency for respiration.  
RESPFREQ  
RESPALTFREQ = 0  
56 kHz  
RESPALTFREQ = 1 (periodic)  
00 (default)  
64 kHz  
01  
10  
11  
54 kHz  
52 kHz  
50 kHz  
56.9 kHz  
51.2 kHz  
46.5 kHz  
R/W  
0
0
RESPEN  
Enable respiration.  
0 (default) = respiration disabled.  
1 = respiration enabled.  
1 ADAS1000-4 model only, ADAS1000-3 model does not contain this feature.  
Rev. B | Page 60 of 80  
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
Table 31. Pace Detection Control Register (PACECTL) Address 0x04, Reset Value = 0x000F881  
R/W  
Default  
Bit  
Name  
Function  
[23:12]  
11  
Reserved  
PACEFILT W  
Reserved, set to 0  
R/W  
1
Pace width filter  
0 = filter disabled  
1 (default) = filter enabled  
R/W  
R/W  
1
1
10  
9
PACETFILT2  
PACETFILT1  
Pace Validation Filter 2  
0 = filter disabled  
1 (default) = filter enabled  
Pace Validation Filter 1  
0 = filter disabled  
1 (default) = filter enabled  
R/W  
R/W  
R/W  
11  
00  
01  
[8:7]  
[6:5]  
[4:3]  
PACE3SEL [1:0]  
PACE2SEL [1:0]  
PACE1SEL [1:0]  
Set lead for pace detection measurement  
00 = Lead I  
01 = Lead II  
10 = Lead III  
11 = Lead aVF  
R/W  
R/W  
R/W  
0
0
0
2
1
0
PACE3EN  
PACE2EN  
PACE1EN  
Enable pace detection algorithm  
0 (default) = pace detection disabled  
1 = pace detection enabled  
1 ADAS1000-4 model only, ADAS1000-3 model does not contain this feature.  
Rev. B | Page 61 of 80  
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
Table 32. Common-Mode, Reference, and Shield Drive Control Register (CMREFCTL) Address 0x05, Reset Value = 0xE00000  
R/W  
R/W  
R/W  
R/W  
Default Bit  
Name  
LACM  
LLCM  
RACM  
Function  
1
1
1
23  
22  
21  
Common-mode electrode select.  
Any combination of the five input electrodes can be used to create the common-mode  
signal, VCM. Bits[23:19] are ignored when Bit 2 is selected. Common mode is the average of  
the selected electrodes. When a single electrode is selected, common mode is the signal  
level of that electrode alone. The common-mode signal can be driven from the internal  
VCM_REF (1.3 V) when Bits [23:19] = 0.  
0 = does not contribute to the common mode.  
1 = contributes to the common mode.  
R/W  
R/W  
R/W  
R/W  
R/W  
R/W  
0
0
0
0
0
0
[20:15] Reserved  
Reserved, set to 0.  
14  
13  
12  
LARLD  
LLRLD  
RARLD  
RLD summing junction.  
0 (default) = does not contribute to RLD input.  
1 = contributes to RLD input.  
[11:10] Reserved  
Reserved, set to 0.  
9
CERLD  
RLD summing junction.  
0 (default) = does not contribute to RLD input.  
1 = contributes to RLD input.  
R/W  
R/W  
0
8
CEREFEN  
RLDSEL [3:0]  
Common electrode (CE) reference, see Figure 56.  
0 (default) = common electrode disabled.  
1 = common electrode enabled.  
0000  
[7:4]  
Select electrode for reference drive.  
0000 (default) = RLD_OUT.  
0001 = LA.  
0010 = LL.  
0011 = RA.  
0100 to 1111 = reserved.  
R/W  
0
3
DRVCM  
Common-mode output. When set, the internally derived common-mode signal is driven  
out of the common-mode pin. This bit has no effect if an external common mode is  
selected.  
0 (default) = common mode is not driven out.  
1 = common mode is driven out of the external common-mode pin.  
R/W  
R/W  
R/W  
0
0
0
2
1
0
EXTCM  
RLDSEL  
SHLDEN  
Select the source of common mode (use when operating multiple devices together).  
0 (default) = internal common mode selected.  
1 = external common mode selected (all the internal common-mode switches are off).  
Enable right leg drive reference electrode.  
0 (default) = disabled.  
1 = enabled.  
Enable shield drive.  
0 (default) = shield drive disabled.  
1 = shield drive enabled.  
Rev. B | Page 62 of 80  
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
Table 33. GPIO Control Register (GPIOCTL) Address 0x06, Reset Value = 0x000000  
R/W  
Default  
Bit  
Name  
Function  
0
0
[23:19] Reserved  
Reserved, set to 0  
R/W  
18  
SPIFW  
Frame secondary SPI words with chip select  
0 (default) = MCS asserted for entire frame  
1 = MCS asserted for individual word  
R/W  
R/W  
0
0
17  
16  
Reserved  
SPIEN  
Reserved, set to 0  
Secondary SPI enable; SPI interface providing ECG data at 128 kHz data rate for external digital pace  
algorithm detection, uses GPIO0, GPIO1, GPIO2 pins  
0 (default) = disabled  
1 = enabled; the individual control bits for GPIO0, GPIO1, GPIO2 are ignored; GPIO3 is not affected  
by SPIEN  
R/W  
00  
[15:14] G3CTL [1:0] State of GPIO3 pin  
00 (default) = high impedance  
01 = input  
10 = output  
11 = open drain  
R/W  
R
0
13  
12  
G3OUT  
G3IN  
Output value to be written to GPIO3 when the pin is configured as an output or open drain  
0 (default) = low value  
1 = high value  
0
Read only; input value read from GPIO3 when the pin is configured as an input  
0 (default) = low value  
1 = high value  
R/W  
00  
[11:10] G2CTL [1:0] State of GPIO2 pin  
00 (default) = high impedance  
01 = input  
10 = output  
11 = open drain  
R/W  
R
0
9
G2OUT  
G2IN  
Output value to be written to GPIO2 when the pin is configured as an output or open drain  
0 (default) = low value  
1 = high value  
0
8
Read only Input value read from GPIO2 when the pin is configured as an input  
0 (default) = low value  
1 = high value  
R/W  
00  
[7:6]  
G1CTL [1:0] State of GPIO1 pin  
00 (default) = high impedance  
01 = input  
10 = output  
11 = open drain  
R/W  
R
0
5
G1OUT  
G1IN  
Output value to be written to GPIO1 when the pin is configured as an output or open drain  
0 (default) = low value  
1 = high value  
0
4
Read only; input value read from GPIO1 when the pin is configured as an input  
0 (default) = low value  
1 = high value  
R/W  
00  
[3:2]  
G0CTL [1:0] State of the GPIO0 pin  
00 (default) = high impedance  
01 = input  
10 = output  
11 = open drain  
R/W  
R
0
0
1
0
G0OUT  
G0IN  
Output value to be written to GPIO0 when the pin is configured as an output or open drain  
0 (default) = low value  
1 = high value  
(Read only) input value read from GPIO0 when the pin is configured as an input  
0 (default) = low value  
1 = high value  
Rev. B | Page 63 of 80  
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
Table 34. Pace Amplitude Threshold Register (PACEAMPTH) Address 0x07, Reset Value = 0x2424241  
R/W  
R/W  
R/W  
R/W  
Default  
Bit  
Name  
Function  
0010 0100  
0010 0100  
0010 0100  
[23:16]  
[15:8]  
[7:0]  
PACE3AMPTH  
PACE2AMPTH  
PACE1AMPTH  
Pace amplitude threshold  
Threshold = N × 2 × VREF/GAIN/216  
1 ADAS1000-4 model only, ADAS1000-3 model does not contain these features.  
Table 35. Test Tone Register (TESTTONE) Address 0x08, Reset Value = 0x000000  
R/W  
R/W  
R/W  
R/W  
R/W  
R/W  
Default  
Bit  
Name  
Function  
0
0
0
23  
22  
21  
TONLA  
TONLL  
TONRA  
Reserved  
TONTYPE  
Tone select  
0 (default) = 1.3 V VCM_REF  
1 = 1 mV sine wave or square wave for TONINT = 1, no connect for TONINT = 0  
0
[20:5]  
[4:3]  
Reserved, set to 0  
00  
00 (default) = 10 Hz sine wave  
01 = 150 Hz sine wave  
1x = 1 Hz, 1 mV square wave  
R/W  
0
2
TONINT  
Test tone internal or external  
0 (default) = external test tone; test tone to be sent out through CAL_DAC_IO and  
applied externally to enabled channels  
1 = internal test tone; disconnects external switches for all ECG channels and  
connects the calibration DAC test tone internally to all ECG channels; in gang  
mode, the CAL_DAC_IO is connected, and the slave disables the calibration DAC  
R/W  
R/W  
0
0
1
0
TONOUT  
TONEN  
Test tone out enable  
0 (default) = disconnects test tone from CAL_DAC_IO during internal mode only  
1 = connects CAL_DAC_IO to test tone during internal mode  
Enables an internal test tone to drive entire signal chain, from preamplifier to SPI  
interface; this tone comes from the calibration DAC and goes to the preamplifier  
through the internal mux; when TONEN (calibration DAC) is enabled, ac lead-off is  
disabled  
0 (default) = disable the test tone  
1 = enable the 1 mV sine wave test tone (calibration mode has priority)  
Rev. B | Page 64 of 80  
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
Table 36. Calibration DAC Register (CALDAC) Address 0x09, Reset Value = 0x0020001  
R/W  
Default  
Bit  
Name  
Function  
0
1
[23:14] Reserved  
Reserved, set to 0.  
R/W  
13  
12  
11  
CALCHPEN  
Calibration chop clock enable. The calibration DAC output (CAL_DAC_IO) can be  
chopped to lower 1/f noise. Chopping is performed at 256 kHz.  
0 = disabled.  
1 (default) = enabled.  
R/W  
R/W  
0
0
CALMODEEN Calibration mode enable.  
0 (default) = disable calibration mode.  
1 = enable calibration mode; connect CAL DAC_IO, begin data acquisition on ECG  
channels.  
CALINT  
Calibration internal or external.  
0 (default) = external calibration to be performed externally by looping CAL_DAC_IO  
around into ECG channels.  
1 = internal calibration; disconnects external switches for all ECG channels and  
connects calibration DAC signal internally to all ECG channels.  
R/W  
0
10  
CALDACEN  
Enable 10-bit calibration DAC for calibration mode or external use.  
0 (default) = disable calibration DAC.  
1 = enable calibration DAC. If a master device and not in calibration mode, also  
connects the calibration DAC signal out to the CAL_DAC_IO pin for external use. If in  
slave mode, the calibration DAC is disabled to allow master to drive the slave  
CAL_DAC_IO pin. When the calibration DAC is enabled, ac lead-off is disabled.  
R/W  
0000000000 [9:0]  
CALDATA[9:0] Set the calibration DAC value.  
1 To ensure successful update of the calibration DAC, the serial interface must issue four additional SCLK cycles after writing the new calibration DAC register word.  
Rev. B | Page 65 of 80  
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
Table 37. Frame Control Register (FRMCTL) Address 0x0A, Reset Value = 0x079000  
R/W  
R/W  
R/W  
R/W  
Default  
Bit  
23  
22  
21  
Name  
Function  
0
0
0
LEAD I/LADIS  
LEADII/LLDIS  
LEADIII/RADIS  
Include/exclude word from ECG data frame. If the electrode/lead is included in the data-  
word and the electrode falls off, the data-word is undefined.  
0 (default) = included in frame.  
1 = exclude from frame.  
R/W  
R/W  
1111  
0
[20:15] Reserved  
Reserved, set to 111111.  
14  
13  
12  
11  
10  
9
PACEDIS1  
RESPMDIS1  
RESPPHDIS1  
LOFFDIS  
Pace detection.  
0 (default) = included in frame.  
1 = exclude from frame.  
R/W  
R/W  
R/W  
R/W  
R/W  
0
1
0
0
0
Respiration magnitude.  
0 (default) = included in frame.  
1 = exclude from frame.  
Respiration phase.  
0 = included in frame.  
1 (default) = exclude from frame.  
Lead-off status.  
0 (default) = included in frame.  
1 = exclude from frame.  
GPIODIS  
GPIO word disable.  
0 (default) = included in frame.  
1 = exclude from frame.  
CRCDIS  
CRC word disable.  
0 (default) = included in frame.  
1 = exclude from frame.  
R/W  
R/W  
0
0
8
7
Reserved  
ADIS  
Reserved, set to 0.  
Automatically excludes PACEDIS[14], RESPMDIS[13], LOFFDIS[11] words if their flags are  
not set in the header.  
0 (default) = fixed frame format.  
1 = autodisable words (words per frame changes).  
R/W  
0
6
RDYRPT  
Ready repeat. If this bit is set and the frame header indicates data is not ready, the frame  
header is continuously sent until data is ready.  
0 (default) = always send entire frame.  
1 = repeat frame header until ready.  
R/W  
R/W  
0
0
5
4
Reserved  
DATAFMT  
Reserved, set to 0.  
Sets the output data format, see Figure 56.  
0 (default) = digital lead/vector format (available only in 2 kHz and 16 kHz data rates).  
1 = electrode format.  
R/W  
R/W  
00  
00  
[3:2]  
[1:0]  
SKIP[1:0]  
Skip interval. This field provides a way to decimate the data.  
00 (default) = output every frame.  
01 = output every other frame.  
1× = output every 4th frame.  
FRMRATE[1:0] Sets the output data rate.  
00 (default) = 2 kHz output data rate.  
01 = 16 kHz output data rate.  
10 = 128 kHz output data rate (DATAFMT must be set to 1).  
11 = 31.25 Hz.  
1 ADAS1000-4 model only, ADAS1000-3 model does not contain these features.  
Rev. B | Page 66 of 80  
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
Table 38. Filter Control Register (FILTCTL) Address 0x0B, Reset Value = 0x000000  
R/W  
R/W  
R/W  
Default  
Bit  
[23:6]  
5
Name  
Function  
0
0
Reserved  
MN2K  
Reserved, set to 0  
2 kHz notch bypass for SPI master  
0 (default) = notch filter bypassed  
1 = notch filter present  
R/W  
R/W  
0
4
N2KBP  
2 kHz notch bypass  
0 (default) = notch filter present  
1 = notch filter bypassed  
00  
[3:2]  
LPF[1:0]  
00 (default) = 40 Hz  
01 = 150 Hz  
10 = 250 Hz  
11 = 450 Hz  
R/W  
00  
[1:0]  
Reserved  
Reserved, set to 0  
Table 39. AC Lead-Off Upper Threshold Register (LOFFUTH) Address 0x0C, Reset Value = 0x00FFFF  
/W  
R
Default  
Bit  
Name  
Function  
0
0
[23:20]  
[19:16]  
Reserved  
ADCOVER[3:0]  
Reserved, set to 0  
R/W  
ADC overrange threshold  
An ADC out-of-range error is flagged if the ADC output is greater than the overrange  
threshold; the overrange threshold is offset from the maximum value  
Threshold = max_value ADCOVER[3:0] × 26  
0000 = maximum value (disabled)  
0001 = max_value − 64  
0010 = max_value − 128  
1111 = max_value − 960  
R/W  
0xFFFF  
[15:0]  
LOFFUTH[15:0] Applies to ac lead-off upper threshold only; lead-off is detected if the output is ≥ N ×  
2 × VREF/gain/216  
0 = 0 V  
Table 40. AC Lead-Off Lower Threshold Register (LOFFLTH) Address 0x0D, Reset Value = 0x000000  
/W  
R
Default  
Bit  
Name  
Function  
0
0
[23:20]  
[19:16]  
Reserved  
Reserved, set to 0  
R/W  
ADCUNDR[3:0] ADC underrange threshold  
An ADC out-of-range error is flagged if the ADC output is less than the underrange  
threshold  
Threshold = min_value + ADCUNDR[3:0] × 26  
0000 = minimum value (disabled)  
0001 = min_value + 64  
0010 = min_value + 128  
1111 = min_value + 960  
R/W  
0
[15:0]  
LOFFLTH[15:0]  
Applies to ac lead-off lower threshold only; lead-off is detected if the output is ≤ N ×  
2 × VREF/GAIN/216  
0 = 0 V  
Rev. B | Page 67 of 80  
 
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
Table 41. Pace Edge Threshold Register (PACEEDGETH) Address 0x0E, Reset Value = 0x0000001  
/W  
R
Default  
Bit  
Name  
Function  
R/W  
R/W  
R/W  
0
0
0
[23:16]  
[15:8]  
[7:0]  
PACE3EDGTH  
PACE2EDGTH  
PACE1EDGTH  
Pace edge trigger threshold  
0 = PACEAMPTH/2  
1 = VREF/gain/216  
N = N × VREF/gain/216  
1 ADAS1000-4 model only, ADAS1000-3 model does not contain these features.  
Table 42. Pace Level Threshold Register (PACELVLTH) Address 0x0F, Reset Value = 0x0000001  
/W  
R
Default  
Bit  
Name  
Function  
R/W  
R/W  
R/W  
0
0
0
[23:16]  
[15:8]  
[7:0]  
PACE3LVLTH[7:0]  
PACE2LVLTH[7:0]  
PACE1LVLTH[7:0]  
Pace level threshold; This is a signed value  
−1 = 0xFF = −VREF/gain/216  
0 = 0x00 = 0 V  
+1 = 0x01 = +VREF/gain/216  
N = N × VREF/gain/216  
1 ADAS1000-4 model only, ADAS1000-3 model does not contain these features.  
Table 43. Read Electrode/Lead Data Registers (Electrode/Lead) Address 0x11 to 0x13, Reset Value = 0x0000001  
/W  
R
Default  
Bit  
Name  
Function  
[31:24]  
Address [7:0]  
0x11: LA or Lead I.  
0x12: LL or Lead II.  
0x13: RA or Lead III.  
R
0
[23:0]  
ECG data  
Channel data value. Data left justified (MSB) irrespective of data rate.  
The input stage can be configured into different modes (electrode, analog lead, or digital  
lead). In electrode mode and analog lead mode, the digital result value is an unsigned  
integer.  
In digital lead/vector mode, the value is a signed twos complement integer format and has  
a 2× range compared to electrode format because it can swing from +VREF to –VREF;  
therefore, the LSB size is doubled.  
Electrode mode and analog lead mode:  
Minimum value (000…) = 0 V  
Maximum value (1111….) = VREF/GAIN  
LSB = (2 × VREF/GAIN)/(2N– 1)  
ECG (voltage) = ECG Data × (2 × VREF/GAIN)/(2N– 1)  
Digital lead mode:  
Minimum value (1000…) = −(VREF/GAIN)  
Maximum value (0111….) = +VREF/GAIN  
LSB = (4 × VREF/GAIN)/(2N – 1)  
ECG (voltage) = ECG Data × (4 × VREF/GAIN)/(2N – 1)  
where N = number of data bits: 16 for 128 kHz data rate or 24 for 2 kHz/16 kHz data rate.  
1 If using 128 kHz data rate in frame mode, only the upper 16 bits are sent. If using the 128 kHz data rate in regular read/write mode, all 32 bits are sent.  
Rev. B | Page 68 of 80  
 
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
Table 44. Read Pace Detection Data/Status Register (PACEDATA) Address 0x1A, Reset Value = 0x0000001, 2, 3  
/W  
R
Default Bit  
Name  
Function  
R
0
23  
Pace 3 detected  
Pace 3 detected. This bit is set once a pace pulse is detected. This bit is set on the  
trailing edge of the pace pulse.  
0 = pace pulse not detected in current frame.  
1 = pace pulse detected in this frame.  
R
R
R
000  
0000  
0
[22:20]  
[19:16]  
15  
Pace Channel 3 width  
Pace Channel 3 height  
Pace 2 detected  
This bit is log2 (width) − 1 of the pace pulse.  
Width = 2N + 1/128 kHz.  
This bit is the log2 (height) of the pace pulse.  
Height = 2N × VREF/gain/216.  
Pace 2 detected. This bit is set once a pace pulse is detected. This bit is set on the  
trailing edge of the pace pulse.  
0 = pace pulse not detected in current frame.  
1 = pace pulse detected in this frame.  
R
R
R
000  
0000  
0
[14:12]  
[11:8]  
7
Pace Channel 2 width  
Pace Channel 2 height  
Pace 1 detected  
This bit is log2 (width) − 1 of the pace pulse.  
Width = 2N + 1/128 kHz.  
This bit is the log2 (height) of the pace pulse.  
Height = 2N × VREF/gain/216.  
Pace 1 detected. This bit is set once a pace pulse is detected. This bit is set on the  
trailing edge of the pace pulse.  
0 = pace pulse not detected in current frame.  
1 = pace pulse detected in this frame.  
R
R
000  
[6:4]  
[3:0]  
Pace Channel 1 width  
Pace Channel 1 height  
This bit is log2 (width) − 1 of the pace pulse.  
Width = 2N + 1/128 kHz.  
0000  
This bit is the log2 (height) of the pace pulse.  
Height = 2N × VREF/gain/216.  
1 If using 128 kHz data rate in frame mode, this word is stretched over two 16-bit words. If using the 128 kHz data rate in regular read/write mode, all 32 bits are sent.  
2 Log data for width and height is provided here to ensure that it fits in one full 32-bit data-word. As a result, there may be some amount of error in the resulting value.  
For more accurate reading, read the 0x3A, 0x3B, 0x3C registers (see Table 52).  
3 ADAS1000-4 model only, ADAS1000-3 model does not contain these features.  
Table 45. Read Respiration Data—Magnitude Register (RESPMAG) Address 0x1B, Reset Value = 0x0000001, 2  
/W  
R
Default Bit  
[23:0]  
Name  
Function  
R
0
Respiration Magnitude[23:0]  
Magnitude of respiration signal. This is an unsigned value.  
4 × (VREF/(1.6468 × respiration gain))/(224 – 1).  
1 If using 128 kHz data rate in frame mode, this word is stretched over two 16-bit words. If using the 128 kHz data rate in regular read/write mode, all 32 bits are sent.  
2 ADAS1000-4 model only, ADAS1000-3 model does not contain these features.  
Table 46. Read Respiration Data—Phase Register (RESPPH) Address 0x1C, Reset Value = 0x0000001, 2  
/W  
R
Default Bit  
[23:0]  
Name  
Function  
R
0
Respiration  
Phase[23:0]  
Phase of respiration signal. Can be interpreted as either signed or unsigned value. If  
unsigned, the range is from 0 to 2π. If signed, the range is from – π to +π.  
0x000000 = 0.  
0x000001 = 2π/224.  
0x400000 = π/2.  
0x800000 = +π = − π.  
0xC00000 = +3π/2 = − π/2.  
0xFFFFFF = +2π(1 − 2−24) = −2π/224.  
1 This register is not part of framing data, but may be read by issuing a register read command of this address.  
2 ADAS1000-4 model only, ADAS1000-3 model does not contain these features.  
Rev. B | Page 69 of 80  
 
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
Table 47. Lead-Off Status Register (LOFF) Address 0x1D, Reset Value = 0x000000  
/W  
R
Default  
Bit  
Name  
Function  
R
0
23  
RLD lead-off  
Status  
Electrode connection status.  
If either dc or ac lead-off is enabled, these bits are the corresponding lead-off status. If  
both dc and ac lead-off are enabled, these bits reflect only the ac lead-off status. DC  
lead-off is available in the DCLEAD-OFF register (see Table 48).  
The common electrodes have only dc lead-off detection.  
An ac lead-off signal can be injected into the common electrode, but there is no ADC  
input to measure its amplitude. If the common electrode is off, it affects the ac lead-off  
amplitude of the other electrodes.  
22  
21  
20  
13  
LA lead-off status  
LL lead-off status  
RA lead-off status  
CELO  
These bits accumulate in the frame buffer and are cleared when the frame buffer is  
loaded into the SPI buffer.  
0 = electrode is connected.  
1 = electrode is disconnected.  
R
R
0
0
[19:14] Reserved  
Reserved.  
12  
11  
10  
LAADCOR  
LLADCOR  
RAADCOR  
ADC out of range error.  
These status bits indicate the resulting ADC code is out of range.  
These bits accumulate in the frame buffer and are cleared when the frame buffer is  
loaded into the SPI buffer.  
R
0
[9:0]  
Reserved  
Reserved.  
Table 48. DC Lead-Off Register (DCLEAD-OFF) Address 0x1E, Reset Value = 0x0000001  
/W  
R
Default  
Bit  
Name  
Function  
R
0
23  
RLD input  
overrange  
The dc lead-off detection is comparator based and compares to a fixed level. Individual  
electrode bits flag indicate if the dc lead-off comparator threshold level has been  
exceeded.  
0 = electrode < overrange threshold, 2.4 V.  
1 = electrode > overrange threshold, 2.4 V.  
22  
21  
20  
13  
LA input  
overrange  
LL input  
overrange  
RA input  
overrange  
CE input  
overrange  
R
R
0
0
[19:14] Reserved  
[8:3]  
Reserved.  
12  
11  
10  
9
RLD input  
underrange  
The dc lead-off detection is comparator based and compares to a fixed level. Individual  
electrode bits indicate if the dc lead-off comparator threshold level has been exceeded.  
0 = electrode > underrange threshold, 0.2 V.  
LA input  
underrange  
1 = electrode < underrange threshold, 0.2 V.  
LL input  
underrange  
RA input  
underrange  
2
CE input  
underrange  
R
0
[1:0]  
Reserved  
1 This register is not part of framing data, but can be read by issuing a register read command of this address.  
Rev. B | Page 70 of 80  
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
Table 49. Operating State Register (OPSTAT) Address 0x1F, Reset Value = 0x0000001  
/W  
R
R
R
R
Default Bit  
Name  
Function  
0
0
0
[23:4] Reserved  
Reserved.  
3
2
Internal error  
Internal digital failure. This is set if an error is detected in the digital core.  
Configuration status  
This bit is set after a reset indicating that the configuration has not been read yet.  
Once the configuration is set, this bit is ready.  
0 = ready.  
1 = busy.  
R
R
0
0
1
0
PLL lock  
PLL lock lost. This bit is set if the internal PLL loses lock after it is enabled and  
locked. This bit is cleared once this register is read or the PWREN bit (Address  
0x01[1]) is cleared.  
0 = PLL locked.  
1 = PLL lost lock.  
PLL locked status  
This bit indicates the current state of the PLL locked status.  
0 = PLL not locked.  
1 = PLL locked.  
1 This register is not part of framing data, but can be read by issuing a register read command of this address. This register assists support efforts giving insight into  
potential areas of malfunction within a failing device.  
Table 50. User Gain Calibration Registers (CALxx) Address 0x21 to Address 0x23, Reset Value = 0x000000  
R/W  
Default Bit  
[31:24]  
Name  
Function  
Address [7:0]  
0x21: calibration LA.  
0x22: calibration LL.  
0x23: calibration RA.  
R/W  
0
23  
USRCAL  
User can choose between default calibration values or user calibration values for GAIN 0,  
GAIN 1, GAIN 2.  
Note that for GAIN 3, there is no factory calibration.  
0 = default calibration values (factory calibration).  
1 = user calibration values.  
R/W  
R/W  
0
0
[22:12]  
[11:0]  
Reserved  
Reserved, set to 0.  
CALVALUE  
Gain calibration value.  
Result = data × (1 + gain × 2−17).  
The value read from this register is the current gain calibration value. If the USRCAL bit is set  
to 0, this register returns the default value for the current gain setting.  
0x7FF (+2047) = ×1.00000011111111111b.  
0x001 (+1) = ×1.00000000000000001b.  
0x000 (0) = ×1.00000000000000000b.  
0xFFF (−1) = ×0.11111111111111111b.  
0x800 (−2048) = ×0.11111100000000000b.  
Rev. B | Page 71 of 80  
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
Table 51. Read AC Lead-Off Amplitude Registers (LOAMxx) Address 0x31 to Address 0x33, Reset Value = 0x0000001  
/W  
R
Default Bit  
[31:24]  
Name  
Function  
Address [7:0]  
0x31: LA ac lead-off amplitude.  
0x32: LL ac lead-off amplitude.  
0x33: RA ac lead-off amplitude.  
R/W  
R
0
0
[23:16]  
[15:0]  
Reserved  
LOFFAM  
Reserved.  
Measured amplitude.  
When ac lead-off is selected, the data is the average of the rectified 2 kHz band-pass filter  
with an update rate of 8 Hz and cutoff frequency at 2 Hz. The output is the amplitude of the  
2 kHz signal scaled by 2/π approximately = 0.6 (average of rectified sine wave). To convert to  
rms, scale the output by π/(2√2).  
Lead-off (unsigned).  
Minimum 0x0000 = 0 V.  
LSB 0x0001= VREF/GAIN/216.  
Maximum 0xFFFF = VREF/GAIN.  
RMS = [π/(2√2)] × [(Code × VREF)/(GAIN × 216)]  
Peak-to-peak = π × [(Code × VREF)/(GAIN × 216)]  
1 This register is not part of framing data, but can be read by issuing a register read command of this address.  
Table 52. Pace Width and Amplitude Registers (PACExDATA) Address 0x3A to Address 0x3C, Reset Value = 0x0000001, 2  
/W  
R
Default Bit  
[31:24]  
Name  
Function  
Address [7:0]  
0x3A: PACE1DATA  
0x3B: PACE2DATA  
0x3C: PACE3DATA  
R
R
0
0
[23:8]  
[7:0]  
Pace height  
Pace width  
Measured pace height in signed twos complement value  
0 = 0  
1 = VREF/gain/216  
N = N × VREF/gain/216  
Measured pace width in 128 kHz samples  
N: (N + 1)/128 kHz = width  
12: (12 + 1)/128 kHz = 101.56 µs (minimum when pace width filter enabled)  
255: (255 + 1)/128 kHz = 2.0 ms  
Disabling the pace width filter allows the pace measurement system to  
return values of N < 12, that is, pulses narrower than 101.56 μs.  
1 These registers are not part of framing data but can be read by issuing a register read command of these addresses.  
2 ADAS1000-4 model only, ADAS1000-3 model does not contain these features.  
Rev. B | Page 72 of 80  
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
Table 53. Frame Header (FRAMES) Address 0x40, Reset Value = 0x8000001  
/W  
R
R
R
Default  
Bit  
31  
30  
Name  
Function  
1
0
Marker  
Ready bit  
Header marker, set to 1 for the header.  
Ready bit indicates if ECG frame data is calculated and ready for reading.  
0 = ready, data frame follows.  
1 = busy.  
R
0
[29:28]  
Overflow [1:0]  
Overflow bits indicate that since the last frame read, a number of frames have  
been missed. This field saturates at the maximum count. The data in the frame  
including this header word is valid but old if the overflow bits are >0.  
When using skip mode (FRMCTL register (0x0A), Bits[3:2]), the overflow bit acts as a  
flag, where a nonzero value indicates an overflow.  
00 = 0 missed.  
01 = 1 frame missed.  
10 = 2 frames missed.  
11 = 3 or more frames missed.  
R
R
R
R
0
0
0
0
27  
26  
25  
24  
Fault  
Internal device error detected.  
0 = normal operation.  
1 = error condition.  
Pace 3 detected  
Pace 2 detected  
Pace 1 detected  
Pace 3 indicates pacing artifact was qualified at most recent point.  
0 = no pacing artifact.  
1 = pacing artifact present.  
Pace 2 indicates pacing artifact was qualified at most recent point.  
0 = no pacing artifact.  
1 = pacing artifact present.  
Pace 1 indicates pacing artifact was qualified at most recent point.  
0 = no pacing artifact.  
1 = pacing artifact present.  
R
R
0
0
23  
22  
Respiration  
0 = no new respiration data.  
1 = respiration data updated.  
Lead-off detected  
If both dc and ac lead-off are enabled, this bit is the OR of all the ac lead-off detect  
flags. If only ac or dc lead-off is enabled, this bit reflects the OR of all dc and ac  
lead-off flags.  
0 = all leads connected.  
1 = one or more lead-off detected.  
R
R
0
0
0
21  
DC lead-off detected  
ADC out of range  
Reserved  
0 = all leads connected.  
1 = one or more lead-off detected.  
20  
0 = ADC within range.  
1 = ADC out of range.  
Reserved.  
[19:0]  
1 If using 128 kHz data rate in frame mode, only the upper 16 bits are sent. If using the 128 kHz data rate in regular read/write mode, all 32 bits are sent.  
Table 54. Frame CRC Register (CRC) Address 0x41, Reset Value = 0xFFFFFF1  
/W  
R
Bit  
Name  
Function  
R
[23:0]  
CRC  
Cyclic redundancy check  
1 The CRC register is a 32-bit word for 2 kHz and 16 kHz data rate and a 16-bit word for 128 kHz rate. See Table 24 for more details.  
Rev. B | Page 73 of 80  
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
INTERFACE EXAMPLES  
The following examples show register commands required  
to configure the ADAS1000-3/ADAS1000-4 devices into  
particular modes of operation and to start framing ECG data.  
Example 2: Enable Respiration and Stream Conversion  
Data (Applies to ADAS1000-4 Only)  
1. Write 1 configures the RESPCTL register with a 56 kHz  
respiration drive signal, gain = 1, driving out through  
the respiration capacitors and measuring on Lead I.  
2. Write 2 issues the read command to start putting the  
converted data out on the SDO pin.  
3. Continue to issue SCLK cycles to read the converted data  
at the configured packet data rate.  
4. Note that this example assumes that the FRMCTL register  
has already been configured such that the respiration  
magnitude is available in the data frame, as arranged in  
Write 2 of Example 1.  
Example 1: Initialize the Device for ECG Capture and  
Start Streaming Data  
1. Write 1 configures the CMREFCTL register for CM =  
WCT = (LA + LL + RA)/3; RLD is enabled onto the  
RLD_OUT electrode. The shield amplifier is enabled.  
2. Write 2 configures the FRMCTL register to output seven  
words per frame/packet. The frame/packet of words  
consist of the header, three ECG words, pace, respiration  
magnitude, and lead-off. The frame is configured to  
always send, irrespective of ready status. The device  
is in analog lead mode with a data rate of 2 kHz.  
3. Write 3 addresses the ECGCTL register, enabling all  
channels into a gain of 1.4, low noise mode, and differen-  
tial input, which configures the device for analog lead  
mode. This register also configures the device as a master,  
using the external crystal as the input source to the XTALx  
pins. The device is also put into conversion mode in this  
write.  
Example 3: DC Lead-Off and Stream Conversion Data  
1. Write 1 configures the LOFFCTL register with a dc lead-off  
enabled for a lead-off current of 50 nA.  
2. Write 2 issues the read command to start putting the  
converted data out on the SDO pin.  
3. Continue to issue SCLK cycles to read the converted  
data at the configured packet data rate.  
4. Note that this example assumes that the FRMCTL register  
has already been configured such that the dc lead-off word  
is available in the data frame, as arranged in Write 2 of  
Example 1.  
4. Write 4 issues the read command to start putting the  
converted data out on the SDO pin.  
5. Continue to issue SCLK cycles to read the converted data  
at the configured packet data rate (2 kHz). The SDI input  
should be held low when reading back the conversion  
data because any commands issued to the interface during  
read of frame/packet are understood to be a change of  
configuration data and will stop the ADC conversions  
to allow the interface to process the new command.  
Table 55. Example 1: Initialize the Device for ECG Capture and Start Streaming Data  
Write Command  
Register Addressed  
CMREFCTL  
FRMCTL  
ECGCTL  
FRAMES  
Read/Write Bit  
Register Address  
Data  
32-Bit Write Command  
0x85E0000B  
0x8A1F9600  
0x81E004AE  
0x40000000  
Write 1  
Write 2  
Write 3  
Write 4  
1
1
1
0
000 0101  
000 1010  
000 0001  
100 0000  
1110 0000 0000 0000 0000 1011  
0001 1111 1001 0110 0000 0000  
1110 0000 0000 0100 1010 1110  
0000 0000 0000 0000 0000 0000  
Table 56. Example 2: Enable Respiration and Stream Conversion Data (Applies to ADAS1000-4 Only)  
Write Command  
Register Addressed  
Read/Write Bit  
Register Address  
Data  
32-Bit Write Command  
0x83002099  
0x40000000  
Write 1  
Write 2  
RESPCTL  
FRAMES  
1
0
000 0011  
100 0000  
0000 0000 0010 0000 1001 1001  
0000 0000 0000 0000 0000 0000  
Table 57. Example 3: Enable DC Lead-Off and Stream Conversion Data  
Write Command  
Register addressed  
Read/Write Bit  
Register Address  
Data  
32-Bit Write Command  
0x82000015  
0x40000000  
Write 1  
Write 2  
LOFFCTL  
FRAMES  
1
0
000 0010  
100 0000  
0000 0000 0000 0000 0001 0101  
0000 0000 0000 0000 0000 0000  
Rev. B | Page 74 of 80  
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
Example 5: Enable Pace Detection and Stream  
Conversion Data (Applies to ADAS1000-4 Only)  
Example 4: Configure 150 Hz Test Tone Sine Wave on  
Each ECG Channel and Stream Conversion Data  
1. Write 1 configures the PACECTL register with all three  
pace detection instances enabled, PACE1EN detecting on  
Lead II, PACE2EN detecting on Lead I, and PACE3EN  
detecting on Lead aVF. The pace width filter and validation  
filters are also enabled.  
1. Write 1 configures the CMREFCTL register to VCM_REF  
= 1.3 V (no electrodes contribute to VCM). RLD is enabled  
to RLD_OUT, and the shield amplifier enabled.  
2. Write 2 addresses the TESTTONE register to enable the  
150 Hz sine wave onto all electrode channels.  
3. Write 3 addresses the FILTCTL register to change the  
internal low-pass filter to 250 Hz to ensure that the 150 Hz  
sine wave can pass through.  
4. Write 4 configures the FRMCTL register to output nine  
words per frame/packet. The frame/packet of words  
consists of the header and three ECG words, pace,  
respiration magnitude, and lead-off. The frame is  
configured to always send, irrespective of ready status.  
The device is in electrode format mode with a data rate  
of 2 kHz. Electrode format is required to see the test tone  
signal correctly on each electrode channel.  
5. Write 5 addresses the ECGCTL register, enabling all  
channels into a gain of 1.4, low noise mode. It configures  
the device as a master and driven from the XTAL input  
source. The device is also put into conversion mode  
in this write.  
2. Write 2 issues the read command to start putting the  
converted data out on the SDO pin.  
3. Continue to issue SCLK cycles to read the converted data  
at the configured packet data rate. When a valid pace is  
detected, the detection flags are confirmed in the header  
word and the PACEDATA register contains information  
on the width and height of the measured pulse from each  
measured lead.  
4. Note that the PACEAMPTH register default setting is  
0x242424, setting the amplitude of each of the pace  
instances to 1.98 mV/gain.  
5. Note that this example assumes that the FRMCTL register  
has already been configured such that the PACEDATA  
word is available in the data frame, as arranged in Write 2  
of Example 1.  
6. Write 6 issues the read command to start putting the  
converted data out on the SDO pin.  
7. Continue to issue SCLK cycles to read the converted data  
at the configured packet data rate.  
Table 58. Example 4: Configure 150 Hz Test Tone Sine Wave on Each ECG Channel and Stream Conversion Data  
Write Command  
Register Addressed  
CMREFCTL  
TESTTONE  
FILTCTL  
FRMCTL  
ECGCTL  
Read/Write Bit Register Address  
Data  
32-Bit Write Command  
Write 1  
Write 2  
Write 3  
Write 4  
Write 5  
Write 6  
1
1
1
1
1
0
000 0101  
000 1000  
000 1011  
000 1010  
000 0001  
100 0000  
0000 0000 0000 0000 0000 1011  
1110 0000 0000 0000 0000 1101  
0000 0000 0000 0000 0000 1000  
0001 1111 1001 0110 0001 0000  
1110 0000 0000 0000 1010 1110  
0000 0000 0000 0000 0000 0000  
0x8500000B  
0x88E0000D  
0x8B000008  
0x8A1F9610  
0x81E000AE  
0x40000000  
FRAMES  
Table 59. Example 5: Enable Pace Detection and Stream Conversion Data (Applies to ADAS1000-4 Only)  
Write Command  
Register Addressed  
Read/Write Bit  
Register Address  
Data  
32-Bit Write Command  
0x84000F8F  
0x40000000  
Write 1  
Write 2  
PACECTL  
FRAMES  
1
0
000 0100  
100 0000  
0000 0000 0000 1111 1000 1111  
0000 0000 0000 0000 0000 0000  
Rev. B | Page 75 of 80  
ADAS1000-3/ADAS1000-4  
Data Sheet  
Example 6: Writing to Master and Slave Devices and Streaming Conversion Data  
This example uses the ADAS1000-3 as the slave device and the  
ADAS1000 as the master device to achieve a configuration with  
eight input electrodes and one right leg drive.  
Master Configuration (ADAS1000)  
Write 4 configures the FRMCTL register to output seven  
words per frame/packet (note that this differs from the  
number of words in a frame available from the slave  
device). The frame/packet of words consists of the header,  
five ECG words, pace, respiration magnitude, and lead-off.  
In this example, the frame is configured to always send  
irrespective of ready status. The master, ADAS1000, is in  
vector mode format with a data rate of 2 kHz. Similar  
to the slave device, the master could be configured for  
electrode mode; the host controller would then be required  
to make the lead calculations.  
Slave Configuration (ADAS1000-3)  
1. Write 1 configures the FRMCTL register to output five  
words per frame/packet. The frame/packet of words  
consists of the header, three ECG words, and lead-off. The  
frame is configured to always send, irrespective of ready  
status. The slave ADAS1000-3 is in electrode mode format  
with a data rate of 2 kHz.  
2. Write 2 configures the CMREFCTL register to receive an  
external common mode from the master.  
1. Write 5 configures the CMREFCTL register for CM =  
WCT = (LA + LL + RA)/3; RLD is enabled onto  
RLD_OUT electrode. The shield amplifier is enabled.  
The CM = WCT signal is driven out of the master device  
(CM_OUT) into the slave device (CM_IN).  
2. Write 6 addresses the ECGCTL register, enabling all  
channels into a gain of 1.4, low noise mode. It configures  
the device as a master in gang mode and driven from  
the XTAL input source. The ADAS1000 master is set to  
differential input, which places it in analog lead mode.  
This ECGCTL register write puts the master into  
conversion mode, where the device sends an edge on  
the SYNC_GANG pin to the slave device to trigger the  
simultaneous conversions of both devices.  
3. Write 3 addresses the ECGCTL register, enabling all  
channels into a gain of 1.4, low noise mode. It configures  
the device as a slave, in gang mode and driven from the  
CLK_IN input source (derived from master ADAS1000).  
The ADAS1000-3 slave is also put into conversion mode  
in this write, but waits for the SYNC_GANG signal from  
the master device before it starts converting.  
3. Write 7 issues the read command to both devices to start  
putting the converted and decimated data out on the  
respective SDO pins.  
4. Continue to issue SCLK cycles to read the converted data  
at the configured packet data rate.  
Table 60. Example 6: Writing to Master and Slave Devices and Streaming Conversion Data  
Device Write Command  
Register Addressed  
R/W Register Address  
Data  
32-Bit Write Command  
0x8A1FF610  
0x85000004  
Slave  
Write 1  
Write 2  
Write 3  
Write 4  
Write 5  
Write 6  
Write 7  
FRMCTL  
CMREFCTL  
ECGCTL  
1
1
1
1
1
1
0
000 1010  
000 0101  
000 0001  
000 1010  
000 0101  
000 0001  
100 0000  
0001 1111 1111 0110 0001 0000  
0000 0000 0000 0000 0000 0100  
1110 0000 0000 0000 1101 1110  
0001 1111 1001 0110 0000 0000  
1110 0000 0000 0000 0000 1011  
1110 0000 0000 0100 1011 1110  
0000 0000 0000 0000 0000 0000  
0x81E000DE  
Master  
FRMCTL  
CMREFCTL  
ECGCTL  
0x8A1F9600  
0x85E0000B  
0x81E004BE  
Master  
and  
FRAMES  
0x40000000  
Slave  
Rev. B | Page 76 of 80  
Data Sheet  
ADAS1000-3/ADAS1000-4  
SOFTWARE FLOWCHART  
Figure 81 shows a suggested sequence of steps to be taken to interface to multiple devices.  
POWER UP ADAS1000  
DEVICES  
WAIT FOR POR ROUTINE  
TO COMPLETE, 1.5ms  
INITIALIZE SLAVE  
DEVICES  
INITIALIZE MASTER DEVICE  
ENABLING CONVERSION  
ISSUE READ FRAME  
COMMAND (WRITE TO 0x40)  
NO  
DRDY  
LOW?  
YES  
ISSUE SCLK CYCLES (SDI = 0)  
TO CLOCK FRAME DATA OUT  
AT PROGRAMMED DATA RATE  
NO  
DISCARD  
FRAME DATA  
IS CRC  
CORRECT?  
YES  
NO  
ACTIVITY  
ON  
SDI?  
YES  
ADAS1000 STOPS CONVERTING,  
SDI WORD USED TO  
RECONFIGURE DEVICE  
RETURN  
TO ECG  
CAPTURE?  
NO  
YES  
ISSUE READ FRAME  
COMMAND (WRITE TO 0x40)  
NO  
POWER-DOWN?  
YES  
ECG CAPTURE COMPLETE  
POWER-DOWN ADAS1000  
ECGCTL = 0x0  
ADAS1000 GOES INTO  
POWER-DOWN MODE  
Figure 81. Suggested Software Flowchart for Interfacing to Multiple Devices  
Rev. B | Page 77 of 80  
 
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
POWER SUPPLY, GROUNDING, AND DECOUPLING  
STRATEGY  
ADCVDD AND DVDD SUPPLIES  
The AVDD supply rail powers the analog blocks in addition to  
the internal 1.8 V regulators for the ADC and the digital core.  
If using the internal regulators, connect the VREG_EN pin to  
AVDD and then use the ADCVDD and DVDD pins for  
decoupling purposes.  
The ADAS1000-3/ADAS1000-4 should have ample supply  
decoupling of 0.01 μF on each supply pin located as close to the  
device pin as possible, ideally right up against the device. In  
addition, there should be one 4.7 μF capacitor for each of the  
power domains, AVDD and IOVDD, again located as close to  
the device as possible. IOVDD is best split from AVDD due to  
its noisy nature.  
The DVDD regulator can be used to drive other external digital  
circuitry as required; however, the ADCVDD pin is purely  
provided for bypassing purposes and does not have available  
current for other components.  
Similarly, the ADCVDD and DVDD power domains each  
require one 2.2 μF capacitor with ESR in the range of 0.5 Ω to  
2 Ω. The ideal location for each 2.2 μF capacitor is dependent  
on package type. For the LQFP package and DVDD decoupling,  
the 2.2 μF capacitor is best placed between Pin 30 and Pin 31,  
while for ADCVDD, the 2.2 μF capacitor should be placed  
between Pin 55 and Pin 56. Similarly for the LFCSP package,  
the DVDD 2.2 μF capacitor is ideal between Pin 43 and Pin 44,  
and between Pin 22 and Pin 23 for ADCVDD. A 0.01 μF  
capacitor is recommended for high frequency decoupling at  
each pin. The 0.01 μF capacitors should have low effective series  
resistance (ESR) and effective series inductance (ESL), such as  
the common ceramic capacitors that provide a low impedance  
path to ground at high frequencies to handle transient currents  
due to internal logic switching.  
Where overall power consumption must be minimized, using  
external 1.8 V supply rails for both ADCVDD and DVDD  
would provide a more efficient solution. The ADCVDD and  
DVDD inputs have been designed to be driven externally and  
the internal regulators may be disabled by tying VREG_EN pin  
directly to ground.  
UNUSED PINS/PATHS  
In applications where not all ECG paths or functions might be  
used, the preferred method of biasing the different functions is  
as follows:  
Unused ECG paths power up disabled. For low power  
operation, they should be kept disabled throughout  
operation. Ideally, these pins should be connected to  
RLD_OUT if not being used.  
Digital lines running under the device should be avoided  
because these couple noise onto the device. The analog ground  
plane should be allowed to run under the device to avoid noise  
coupling. The power supply lines should use as large a trace as  
possible to provide low impedance paths and reduce the effects  
of glitches on the power supply line. Fast switching digital  
signals should be shielded with digital ground to avoid radiating  
noise to other parts of the board and should never be run near  
the reference inputs. It is essential to minimize noise on VREF  
lines. Avoid crossover of digital and analog signals. Traces on  
opposite sides of the board should run at right angles to each  
other. This reduces the effects of feedthrough throughout the  
board. As is the case for all thin packages, take care to avoid  
flexing the package and to avoid a point load on the surface of  
this package during the assembly process.  
Unused external respiration inputs can be tied to ground if  
not in use.  
If unused, the shield driver can be disabled and output left  
to float.  
DRDY  
CM_OUT, CAL_DAC_IO,  
SYNC_GANG can be left open.  
, GPIOx, CLK_IO,  
LAYOUT RECOMMENDATIONS  
To maximize CMRR performance, pay careful attention to  
the ECG path layout for each channel. All channels should be  
identical to minimize difference in capacitance across the paths.  
Place all decoupling as close to the ADAS1000-3/ADAS1000-4  
devices as possible, with an emphasis on ensuring that the  
VREF decoupling be prioritized, with VREF decoupling on  
the same side as the ADAS1000-3/ADAS1000-4devices, where  
possible.  
During layout of board, ensure that bypass capacitors are placed  
as close to the relevant pin as possible, with short, wide traces  
ideally on the topside.  
AVDD  
While the ADAS1000-3/ADAS1000-4 are designed to operate  
from a wide supply rail, 3.15 V to 5.5 V, the performance is  
similar over the full range, but overall power increases with  
increasing voltage.  
Rev. B | Page 78 of 80  
 
 
 
 
 
Data Sheet  
ADAS1000-3/ADAS1000-4  
OUTLINE DIMENSIONS  
9.10  
9.00 SQ  
8.90  
0.60  
0.42  
0.24  
0.275  
0.60  
0.42  
0.24  
0.150  
56  
43  
42  
1
PIN 1  
INDICATOR  
8.75  
BSC SQ  
0.50  
BSC  
6.05  
5.95 SQ  
5.85  
*
EXPOSED PAD  
0.75  
0.65  
0.55  
29  
14  
15  
28  
TOP VIEW  
BOTTOM VIEW  
6.50 REF  
0.70 MAX  
0.65 NOM  
12° MAX  
0.90  
0.85  
0.80  
0.05 MAX  
0.01 NOM  
0.30  
0.23  
0.18  
SEATING  
PLANE  
0.20 REF  
*
FOR PROPER CONNECTION OF THE EXPOSED PAD, REFER TO  
THE PIN CONFIGURATION AND FUNCTION DESCRIPTIONS  
SECTION OF THIS DATA SHEET.  
Figure 82. 56-Lead, Lead Frame Chip Scale Package [LFCSP_VQ]  
9 mm × 9 mm Body, Very Thin Quad  
(CP-56-7)  
Dimensions shown in millimeters  
12.20  
12.00 SQ  
11.80  
0.75  
0.60  
0.45  
1.60  
MAX  
64  
49  
48  
1
PIN 1  
10.20  
10.00 SQ  
9.80  
TOP VIEW  
(PINS DOWN)  
1.45  
1.40  
1.35  
0.20  
0.09  
7°  
3.5°  
0°  
0.08  
COPLANARITY  
16  
33  
32  
0.15  
0.05  
SEATING  
PLANE  
17  
VIEW A  
0.27  
0.22  
0.17  
0.50  
BSC  
LEAD PITCH  
VIEW A  
ROTATED 90° CCW  
COMPLIANT TO JEDEC STANDARDS MS-026-BCD  
Figure 83. 64-Lead Low Profile Quad Flat Package [LQFP]  
(ST-64-2)  
Dimensions shown in millimeters  
Rev. B | Page 79 of 80  
 
ADAS1000-3/ADAS1000-4  
Data Sheet  
ORDERING GUIDE  
Temperature  
Range  
Package  
Description  
Package  
Option  
Model1  
Option  
Tray  
Description  
ADAS1000-3BSTZ  
ADAS1000-3BSTZ-RL  
ADAS1000-3BCPZ  
ADAS1000-3BCPZ-RL  
ADAS1000-4BSTZ  
ADAS1000-4BSTZ-RL  
ADAS1000-4BCPZ  
ADAS1000-4BCPZ-RL  
EVAL-ADAS1000SDZ  
EVAL-SDP-CB1Z  
3 ECG Channels  
−40°C to +85°C  
−40°C to +85°C  
−40°C to +85°C  
−40°C to +85°C  
−40°C to +85°C  
−40°C to +85°C  
−40°C to +85°C  
−40°C to +85°C  
64-Lead LQFP  
ST-64-2  
ST-64-2  
CP-56-7  
CP-56-7  
ST-64-2  
ST-64-2  
CP-56-7  
CP-56-7  
Reel, 1000  
Tray  
64-Lead LQFP  
56-Lead LFCSP_VQ  
56-Lead LFCSP_VQ  
64-Lead LQFP  
Reel, 2500  
Tray  
3 ECG Channels, Pace Algorithm, Respiration Circuit  
Reel, 1000  
Tray  
64-Lead LQFP  
56-Lead LFCSP_VQ  
56-Lead LFCSP_VQ  
Evaluation Kit2  
Reel, 2500  
ADAS1000 Evaluation Board  
System Demonstration Board (SDP), used as a controller  
board for data transfer via USB interface to PC  
Controller Board3  
1 Z = RoHS Compliant Part.  
2 This evaluation kit consists of ADAS1000BSTZ × 2 for up to 12-lead configuration. Because the ADAS1000 contains all features, it is the evaluation vehicle for  
all ADAS1000 variants.  
3 This board allows a PC to control and communicate with all Analog Devices evaluation boards ending in the SD designator.  
©2012–2015 Analog Devices, Inc. All rights reserved. Trademarks and  
registered trademarks are the property of their respective owners.  
D10997-0-1/15(B)  
Rev. B | Page 80 of 80  
 

相关型号:

ADAS1000-4BCPZ-RL

Low Power, Three Electrode Electrocardiogram (ECG) Analog Front End
ADI

ADAS1000-4BSTZ

Low Power, Three Electrode Electrocardiogram (ECG) Analog Front End
ADI

ADAS1000-4BSTZ-RL

Low Power, 3-Electrode Electrocardiogram (ECG) Analog Front End with respiration measurement and pace detection
ADI

ADAS1000BCPZ

Low Power, Five Electrode Electrocardiogram (ECG) Analog Front End
ADI

ADAS1000BCPZ-RL

Low Power, Five Electrode Electrocardiogram (ECG) Analog Front End
ADI

ADAS1000BSTZ

Low Power, Five Electrode trocardiogram (ECG) Analog Front End
ADI

ADAS1000BSTZ-RL

Low Power, Five Electrode Electrocardiogram (ECG) Analog Front End
ADI

ADAS1000_17

Low Power, Five Electrode Electrocardiogram (ECG) Analog Front End
ADI

ADAS1127

64-Channel,24-Bit Current-to-Digital ADC
ADI

ADAS1128

128-Channel, 24-Bit Current-to-Digital ADC
ADI

ADAS1135

256-Channel, 24-Bit, Current-to-Digital ADC Module
ADI

ADAS1135_17

256-Channel, 24-Bit, Current-to-Digital ADC Module
ADI