MQC100-24S [ARCH]
Switching Power Module for PCB Mountable;![MQC100-24S](http://pdffile.icpdf.com/pdf2/p00331/img/icpdf/MQC100_2032822_icpdf.jpg)
型号: | MQC100-24S |
厂家: | ![]() |
描述: | Switching Power Module for PCB Mountable PC |
文件: | 总3页 (文件大小:243K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
![](http://public.icpdf.com/style/img/ads.jpg)
VER : A_4
update : 2018.06.06
AC -DC ITE & Medical Power Module
MQC100 SERIES
100 Watts
KEY FEATURES
Switching Power Module for PCB Mountable
4000VAC Input to Output 2MOPP Insulation
Cooling by Free Air Convection
High Efficiency up to 93.5%
With P.F.C. Function >0.9
<0.5W No Load Input Power
Protections: Over Load / Over Voltage / Over Temperature / Short Circuit
EMI for Both Class I (with PE) and Class II (without PE) Configuration
Suitable for BF Application with Appropriate System Consideration
UL / IEC / EN 60601 3.1 Edition & UL / IEC / EN 60950 AM2 Safety Approvals
3-Year Product Warranty
ELECTRICAL SPECIFICATIONS
All specifications valid at 230VAC input voltage, full load and +25°C after warm-up time unless otherwise stated.
Model No.
MQC100-12S
100 W
MQC100-24S
MQC100-48S
Max Output Wattage (W)
Voltage
(Note 3)
90-264 VAC
47-63 Hz
Frequency (Hz)
Current (Full load)
Inrush Current (<2ms)
Leakage Current
< 2.0 A max. (115 VAC) / < 1.0 A max. (230 VAC)
< 45 A max. (115 VAC) / < 90 A max. (230 VAC)
< 0.1mA / 264 VAC (Touch Current)
PF>0.9 at Full Load
Input
Power Factor
Voltage (V.DC.)
12V
24V
48V
2.1
Voltage Accuracy
Current (A) (max.)
Line Regulation
±2%
8.33
4.2
±1%
Load Regulation (0-100%)
Minimum Load
±1%
Output
0%
Maximum Capacitive Load
Ripple & Nois (max.)
6000µF
2000µF
93%
330µF
93.5%
(Note 1)
(Note 4)
(Note 2)
1% Vout
Efficiency (at 230VAC)
Hold-up Time (at 115 VAC)
Over Power Protection
92.5%
10 ms min.
Auto recovery, Hiccup mode
Auto recovery
Auto recovery
Over Voltage Protection
Overt Temperature Protection
Protection
Isolation
Protection level 1 (nominal):Continuous, Auto recovery
Protection level 2 (instantaneous high current):Latch
4000VAC or 5656VDC
Short Circuit Protection
Input-Output
Input-PE
2000VAC or 2828VDC
Output-PE
1500VAC or 2121VDC
Operating Temperature
Storage Temperature
Temperature Coefficient
Altitude During Operation
Humidity
-30°C…+70°C (with derating)
-30°C…+85°C
±0.05%/°C
5000m
Environment
95% RH
MTBF
>250,000 h @ 25°C (MIL-HDBK-217F, Notice 1)
56 kPa to 106 kPa
Atmospheric Pressure
Vibration
10~500Hz, 5G 10min./1cycle, 60min. each along X, Y, Z axes.
Dimension (L x W x H)
Weight
4.3 x 2.3 x 1.38 Inches ( 109.0 x 58.5 x 35.0 mm ) Tolerance 0.5 mm
Physical
365 g
Cooling Method
Free convection
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
-1-
VER : A_4
update : 2018.06.06
AC -DC ITE & Medical Power Module
MQC100 SERIES
100 Watts
ELECTRICAL SPECIFICATIONS
All specifications valid at 230VAC input voltage, full load and +25°C after warm-up time unless otherwise stated.
Model No.
Safety
MQC100-12S
MQC100-24S
MQC100-48S
Approval
UL / IEC / EN 60601 3.1rd Edition & UL / IEC / EN 60950 AM2
EN55011 Conducted & Radiated Class B
EN55011 Class I class B / Class II class A
EN60601-1-2 4th edition
(Note 5)
(Note 5)
Conducted EMI
Radiated EMI
EMS
EMC
NOTE
1. Ripple & Noise are measured at 20MHz of bandwidth with 0.1uF & 47uF parallel capacitor.
2. Hold-up Time measured at 90% Vout.
3. Please check the derating curve for more details.
4. After 30 minutes of burn-in
5. Please secure the power supply unit to your metal case by using the four screw holes in the corners for
either Class I or Class II equipment
6. Please refer to our PDF file "AC-DC Application" on our website: www.archcorp.com.tw
DERATING
TRIM
12S
~
24S
~
48S
~
Trim
→
+5%
0%
+5%
0%
+5%
0%
34KΩ
10MΩ
37.4KΩ
10MΩ
38KΩ
10MΩ
-V
Trim
→
-5%
-5%
-5%
0%
0%
0%
10MΩ
~
106KΩ
10MΩ
~
270KΩ
10MΩ
~
640KΩ
+V
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
-2-
VER : A_4
update : 2018.06.06
AC -DC ITE & Medical Power Module
MQC100 SERIES
100 Watts
MECHANICAL DIMENSION ( Top View )
Φ
PIN#
Single
AC IN (N)
AC IN (L)
PE
1.2±0.1%mm
1.2±0.1%mm
1.2±0.1%mm
1.2±0.1%mm
1
2
3
4
ON / OFF
(Provide +5Vdc Controlled)
1.8±0.1%mm
5
+DC OUT
-DC OUT
Trim
1.8±0.1%mm
1.2±0.1%mm
6
7
Remark:
Please reserve the pin 4 hole on PCB.
If the remote on/off function is not required, please
connect the pin 4 circuit layout with pin6, or keep pin
4 floating.
BLOCK DIAGRAM
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
-3-
相关型号:
©2020 ICPDF网 联系我们和版权申明