8715-01 [ASM-SENSOR]

POWER HiCORDER; 电源记录仪
8715-01
型号: 8715-01
厂家: ASM GMBH    ASM GMBH
描述:

POWER HiCORDER
电源记录仪

文件: 总4页 (文件大小:926K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
Do you ever worry about power failure?  
POWER HiCORDER 8714-01,8715-01  
Recorders  
Introducing easy-to-use power monitoring recorders!  
See what you’ve been missing !  
Capture Spikes, Sags, and Surges  
Monitor and record power anomalies and fluctuations  
easily, even unattended  
Analyze Leakage Currents as well as AC/DC Current up  
to 2000 amps  
True-RMS with Trigger Function  
250µsec Transient Waveform Capture  
2
Record power anomalies accurately with simple operation !  
- Aplications -  
What a Power Recorder Does  
There are many power-quality concerns in today's  
PC-based businesses, and detecting power anomalies  
is important for maintaining factory facilities and  
equipment. The POWER HiCORDERs 8714-01 and  
8715-01 are monitoring recorders that can capture  
instantaneous power drop-outs, brown-outs, spikes and  
sags, and record voltage changes with simple operation,  
and even monitor and record while totally unattended.  
The compact B5 size and 1.5kg (52.9oz.) weight provide  
excellent portability among worksites. (8714-01  
)
Simple Setup  
1. Simple setup. No worries about operability.  
2. Accepts direct input of line voltages up to 400 V, and  
up to 1000 V AC/ 2000 V DC with the use of the  
DIFFERENTIAL PROBE 9322 accessory.  
3. Voltage and current can be simultaneously measured on  
four isolated channels.  
6. High-speed A/D converter stores data in memory,  
providing simple recording of events that cannot be  
recorded by a pen recorder.  
7. Two measurement functions are  
available: Waveform Measurement  
Mode (memory mode), and RMS  
Trend Measurement Mode (RMS  
recorder & memory).  
4. Data immediately before an anomaly occurs can be stored  
and recorded.  
5. High quality data printouts on site.  
8. HIOKI clamp-on probes (voltage-  
output type) can be directly  
connected.  
Example  
To monitor the current waveform on a 220 V  
power line.  
[ Setup ]  
Measurement mode is set to Waveform, time axis to  
2ms. Recording length is set to 20 division.  
[ Channel setup ]  
For current waveforms, after selecting a clamp-on  
probe for the input type from Mode, select the range  
of the clamp. Here, we select 9010 and 100A. Set  
the range of the clamp-on probe in the same way.  
[ Starting Measurement ]  
Clamp the probe around the conductor to be  
measured.  
Press the START button.  
The current waveform is  
displayed on the screen.  
To stop measuring, press  
he STOP button.  
Current Waveform Example  
Two models: the two-channel (8714-01) and four-channel (8715-01)  
- Specifications -  
Basic Specifications  
8714-01  
8715-01  
Waveform Measurement (Power Waveform Observation) Mode  
Waveform Measurement mode (Power Waveform Observation),  
RMS Trend Measurement mode (RMS Observation)  
200 and 400µs, 1, 2, 5, 10, and 20ms/division, time axis zoom  
×2 to ×10; 3 settings, compression 1/2 to 1/50; 5 settings  
Measurement Mode  
Time axis  
1/80 of time axis ranges (minimum sampling period 2.5µs)  
Sampling period  
Recording length  
Input types and  
Input channels fixed, Analog 2 channels Input channels fixed, Analog 4 channels  
Inter-channels and input-frame isolation  
Inter-channels and input-frame isolation  
number of channels  
20, 50, 100, 200, or 400 division (800 division at sequential save OFF)  
400k sample/s (2.5μs period)  
Simultaneous sampling for 2 or 4 analog channels  
Can record data from before the trigger point, 0/10/20/50, or 100%  
of recording length  
Maximum sampling rate  
Memory capacity  
Pre-trigger  
continuous 2-channels measurement  
(12-bit analog) × 64 kilo-words/ ch  
continuous 4-channels measurement  
(12-bit analog) × 64 kilo-words/ ch  
Voltage axis normal (×1/2), magnified (×1), Left-right waveform scrolling,  
Automatic, manual and partial (between A-B cursors) printing  
Other functions  
PC Card TYPE II slot × 1 : Flash ATA card (max.1 GB), MS-DOS  
format. Memory contents : Settings data, Measurement data  
(binary or text format), Screen data (BMP)  
RMS Trend Measurement (RMS Fluctuation Measurement) Mode  
External memory  
Commercial mains power (50/60Hz)  
Measurement objective  
Clock, waveform data and settings, Battery life approx. 5 years (at  
25°C/ 77°F).  
1/2/5 seconds/DIV (cannot realtime print); 10/30 seconds; 1/2/5/10/30  
minutes; or 1hour/DIV, time axis compression 1/2 to 1/50; 5 settings  
Backup function  
External control  
Time axis  
Terminal block : trigger input/output  
250µs fixed (RMS value 800 data/second)  
Continuous measurement only (up to 200 division of data are stored internally)  
±3 % f.s.  
Sampling period  
Recording length  
RMS accuracy  
Operation: +5°C/41°F to +40°C/104°F, 35% to 80% relative humidity.  
Storage: -10°C/14°F to +50°C/122°F, 35% to 80% relative humidity.  
Environment conditions  
(no condensation)  
Safety: EN61010  
EMC: EN61326, EN61000-3-2, EN61000-3-3  
Applicable standards  
Power supplies  
100V Line: 75 to 125V rms (standard)  
200V Line: 150 to 250V rms (standard)  
400V Line: 275 to 525V rms (standard)  
Clamp: 0A rms to f.s.value on clamp-on probe (rms)  
9322 Differential probe: 400V Line; 275 to 525V rms (standard)  
9322 Differential probe: 600V Line; 475 to 725V rms (standard)  
(1) AC Adapter model 9418-15 or 9418-10 (DC 12V ±10%)  
(2) BATTERY PACK 9447 (AC adapter has priority when used in  
combination with battery pack, fast recharge possible with AC adapter)  
Waveform display  
area  
15VA max.  
Power consumption  
Continuous operation time  
Charge time  
Voltage axis normal (×1/2), magnified (×5), Left-right waveform scroll,  
Automatic, manual and partial (between A-B cursors) printing, Hybrid  
recording: prints logging data for each division together with waveform  
Approx. 3 hours (with BATTERY PACK 9447, trigger standby at 23°C/ 73°F)  
Approx. 2 hours fast charge with power switch OFF (at 23°C/ 73°F)  
approx. 280 (11.02) W × 170 (6.69) H × 52 (2.05) approx. 280 (11.02) W × 170 (6.69) H × 52 (2.05)  
Other functions  
Dimensions and  
mass  
RMS Trend Measurement (Instantaneous Waveform Measurement) Mode  
Dmm (inch), 1.5kg/ 52.91oz (without batteries)  
Dmm (inch) , 1.6kg/ 56.44oz (without batteries)  
200 and 400µs, 1, 2, 5, 10, and 20ms/division, time axis zoom  
×2 to ×10; 3 settings, compression 1/2 to 1/50; 5 settings  
Instruction manual ×1, Measurement guide ×1, Recording paper ×  
1, Paper attachment ×2, Blank box ×1, Strap ×1, AC ADAPTER  
9418-15 ×1, CONNECTION CORD 9197 ×2 (8714-01), ×4 (8715-01),  
Application disk ×1  
Time axis  
Supplied  
accessories  
1/80 of time axis ranges (minimum sampling period 2.5µs)  
Sampling period  
Recording length  
20, 50, 100, or 200 division (400 division at sequential save OFF)  
Can record data from before the trigger point, 0/10/20/50, or 100%  
of recording length  
Pre-trigger  
Voltage axis normal (×1/2), magnified (×1), Left-right waveform scroll  
Other functions  
Other Functions  
Printing of settings including input range, trigger time, etc., cursor  
measurement, start condition retention, auto setup, list & gauge  
printing, power-save function, DMM function (voltage shown as numerals  
on the display), auto saving, partial saving (between A-B cursors or all data),  
sequential save.  
General  
280 mm  
(11.02 in)  
Display update rate: 1 second  
Display contents: RMS value (only measuring DC and 50/60Hz)  
Display digit: 4 (last digit 0 to 4 is rounded zero, 5 to 9 is rounded five)  
Voltage ranges: Auto-select only (five ranges from 10 mV to 100 V/ div.)  
Accuracy: ±3% rdg. ±5dgt.  
Mass (8714-01) : approx. 1.5kg/52.91oz  
Mass (8715-01) : approx. 1.6kg/56.44oz  
DMM Function  
Recording and Display Section  
5.7-inch STN color LCD, with English/Japanese language selector  
240 × 320 dots  
Display  
112mm (4.41in) × 18m (59.06ft), thermal paper roll  
100mm in full scale (10 divisions), 1 division = 10mm (0.39in)  
Max. 1sec/division (with AC adapter use)  
Recording paper  
Recording area  
Recording speed  
Input section (accuracy at 23 ±5 ˚C/73 ±9 ˚F after 30 minutes warm-up time; accuracy guaranteed for 1 year)  
Isolated BNC terminal, Input item: Voltage, 9010, 9018, 9132,  
3283, 3284, 3285, 9322 (selectable)  
Input  
Trigger Function(only for anomalous waveform measurement and instantaneous power fluctuation recording)  
Analog input CH1 - CH4 (8714-01: CH1 - CH2), external (either ON or  
Trigger source  
Voltage : 100, 200, 400V line, or 100V DC  
9010 : 10, 20, 50, 100, 200, or 500A  
9018 : 10, 20, 50, 100, 200, or 500A  
OFF for each source), logical AND/OR of sources  
Level trigger: At preset voltage level, on either rising or falling  
edge  
Window trigger: At entry or exit from preset upper and lower limits  
Voltage drop trigger: Especially for 50/60Hz commercial power,  
when peak voltage falls below preset value  
Period trigger: When period of rising or falling edge of measured  
signal exceeds preset period  
9132 : 20, 50, 100, 200, 500, or 1000A  
3283 : 10mA, 100mA, 1, 10, or 200A  
Measurement range  
3284 : 20 or 200A  
3285 : 200 or 2000A  
9322 : 400 or 600V line  
Anomalous  
waveform triggers  
Waveform judgment trigger: Especially for 50/60Hz commercial  
power, to monitor signals outside of judgment range in real time  
(when time axis range 20ms/division is not applicable)  
400kS/s (simultaneous sampling of all channels)  
Max. sampling rate  
Accuracy, frequency  
characteristics  
±0.5% f.s. (applied accuracy of clamp-on probe), DC to 50kHz ±3dB  
Voltage drop trigger: Especially for 50/60Hz commercial power,  
when peak voltage falls below preset value  
RMS level trigger: At preset effective value, on either rising or  
falling edge  
OFF, 500Hz, 5kHz  
Low-pass filter  
Power fluctuation  
Input resistance and  
capacitance  
measurement triggers  
1MΩ, 7pF approx. (at 100kHz)  
(valid only when instantaneous  
waveform measurement is enabled) RMS window trigger: At entry or exit from preset upper and lower  
limit levels  
DC (fixed)  
Input coupling  
0, 10, 20, 50, 100% (for anomalous waveform measurement, instantaneous  
power fluctuation recording)  
450V AC rms, DC CAT II (upper voltage which when applied to between  
input pins does not damage them)  
Pre-Trigger  
Max. allowable input  
0.25% f.s. (f.s.=10 divisions, 0.1% f.s. for use with waveform judgment trigger)  
Level setting resolution  
Max. rated voltage  
to earth  
450V AC rms, DC (upper voltage which when applied to input channel  
casing or between input channels does not damage them)  
Off/On (0.5 divisions fixed filter width)  
Trigger filter  
4
Composition of options  
Note:ꢀProductꢀnamesꢀappearingꢀhereinꢀareꢀtrademarksꢀorꢀregisteredꢀtrademarksꢀofꢀvariousꢀcompanies.ꢀ  
Record Current Waveform  
Record Current Waveform  
CLAMP ON AC/DC HiTESTER  
3285  
200A or 2000AAC/DC, with  
analog output of 1V f.s. DC, and  
waveform monitor output of 1V  
f.s. AC at 10Hz to 1kHz, or DC.  
Requires the AC ADAPTER  
9445-02/-03  
CARRYING CASE 9391  
Soft case type, for storing options  
Holds more options than the 9648  
hard case  
CLAMP ON PROBE 9018-10  
Input from 10 to 500A  
40Hz to 3kHz for 0.2V AC output.  
BNC terminal  
CARRYING CASE 9648  
Hard case type, for storing  
options  
Supplied with PC  
Card adapter  
CLAMP ON AC/DC HiTESTER  
3284  
PC CARD 128M 9726  
(128MB capacity)  
20A or 200AAC/DC, with  
analog output of 1V f.s. DC, and  
waveform monitor output of 1V  
f.s. AC at 10Hz to 2kHz, or DC.  
Requires the AC ADAPTER  
9445-02/-03  
PC CARD 256M 9727  
Not CE certified  
(256MB capacity)  
PC Card Precaution  
CLAMP ON PROBE 9132-10  
Input from 20 to 1000A  
40Hz to 1kHz for 0.2V AC output.  
BNC terminal  
PC CARD 512M 9728  
Useꢀ onlyꢀ PCꢀ Cardsꢀ soldꢀ  
byꢀ HIOKI.ꢀ Compatibilityꢀ  
andꢀ performanceꢀ areꢀ notꢀ  
guaranteedꢀforꢀPCꢀcardsꢀmadeꢀ  
byothermanufacturers.Youꢀ  
mayꢀbeꢀunableꢀtoꢀreadꢀfromꢀorꢀ  
saveꢀdataꢀtoꢀsuchꢀcards.  
WAVE PROCESSOR 9335  
Data conversion, print functions,  
waveform display, compatible with  
Windows 95/98/Me, Windows NT  
4.0/2000/XP  
(512MB capacity)  
PC CARD 1G 9729  
(1GB capacity)  
CLAMP ON LEAK HiTESTER  
3283  
For leakage current measurement,  
includes 10mA to 200A ranges,  
with analog output of 1V f.s. DC,  
and waveform monitor output  
of 1V f.s. AC at 40Hz to 2kHz.  
Requires the AC ADAPTER  
9445-02/-03  
Input from power-concent  
1kV AC, 2kV DC input  
AC ADAPTER 9445-02/-03  
Use for the 3283, 3284,  
3285, for USA/CSA  
DIFFERENTIAL PROBE 9322  
For inputs up to 2kV DC or  
1kV AC, the 9322 requires  
the AC ADAPTER 9418-15  
CONCENT INPUT CORD  
9448  
9445-02, for EU 9445-03  
Use for 100V commercial  
line, 2m length  
+
Not CE certified  
CONVERSION ADAPTER  
9199  
OUTPUT CORD 9094  
Required along with the  
9199 adapter to connect  
the 3283/3284/3285 to the  
8714-01 or 8715-01  
AC ADAPTER 9418-15  
Use for the 9322, 100 to 240  
V AC  
POWER HiCORDER 8714-01 (2ch model)  
POWER HiCORDER 8715-01 (4ch model)  
Banana-to-BNC, use to  
connect to insulation-BNC  
terminal on Input section  
Included accessories: AC ADAPTER 9418-15 ×1, Recording paper ×1, Blank box ×1, Strap ×1,  
CONNECTION CORD 9197 (8714-01: ×2, 8715-01: ×4), Application disk ×1  
Up to 450V AC input  
+
CONVERSION ADAPTER  
9199  
Not CE certified  
CONNECTION CORD 9197  
For up to 500V,  
1.5 m length  
Banana-to-BNC, use to  
connect to insulation-BNC  
terminal on Input section  
Not CE certified  
LINE SPLITTER CT101A  
For 100V/ 15A,  
convenient for measuring  
CONNECTION CORD  
9217  
RECORDING PAPER  
9234  
PAPER WINDER 220H  
Insulation BNC-to-insulation  
Input cords of included accessories are the 9197 only. For current  
measurement, please purchase the optional clamp-on probe, or clamp-on  
tester.  
Paper width: 70 to 220 mm, 100V AC line current with  
BNC, use to connect to  
insulation-BNC terminal on  
Input Module  
18 m/ 59.06 feet length,  
10 rolls/ 1 set  
using special-purpose AC  
adapter  
clamp-on probe  
CHARGE STAND 9643  
AC ADAPTER 9418-15  
BATTERY PACK 9447  
Independent of main unit the  
8714-01/8715-01, use with the AC  
ADAPTER 9418-15 to charge one  
Model BATTERY PACK 9447.  
Universal 100 to 240 V AC,  
12 V DC/ 2.5 A output  
7.2V, 2400mAh  
n Combination example: 2-channels, with current measurement  
Main unit  
Paper  
Battery pack  
9447×1  
Current up to 500A  
9018-10×2  
Carrying case  
9ꢀ91×1  
Model number  
quantity  
×
8714-01×1  
92ꢀ4×1 (10 rolls)  
n Combination example: 4-channels, with current measurement  
Main unit  
Paper  
Battery pack  
9447×1  
Current up to 500A  
9018-10×4  
Carrying case  
9ꢀ91×1  
Model number  
quantity  
×
8715-01×1  
92ꢀ4×1 (10 rolls)  
DISTRIBUTED BY  
HEAD OFFICE :  
Shanghai Representative Office :  
81 Koizumi, Ueda, Nagano, ꢀ86-1192, Japan  
1ꢀ10 Shanghai Times Square Office  
TEL +81-268-28-0562 / FAX +81-268-28-0568 9ꢀ Huaihai Zhong Road  
E-mail: os-com@hioki.co.jp  
Shanghai, 200021, P.R.China  
TEL +86-21-6ꢀ91-0090, 0092  
FAX +86-21-6ꢀ91-0ꢀ60  
E-mail: info@hioki.cn  
HIOKI USA CORPORATION :  
6 Corporate Drive, Cranbury, NJ 08512 USA  
TEL +1-609-409-9109 / FAX +1-609-409-9108  
E-mail: hioki@hiokiusa.com  
All information correct as of Jan. 25, 2007. All specifications are subject to change without notice.  
8714E4-71E-00P Printed in Japan  

相关型号:

87150-1000

Board Connector, 50 Contact(s), 2 Row(s), Female, 0.1 inch Pitch, IDC Terminal, Gray Insulator, Plug
3M

87150-1650

HEADER CONNECTOR,CABLE MNT,PLUG,50 CONTACTS,SKT,0.1 PITCH,IDC TERMINAL,GRAY
3M

87150-1900

HEADER CONNECTOR,CABLE MNT,PLUG,50 CONTACTS,SKT,0.1 PITCH,IDC TERMINAL,GRAY
3M

87150-3000

HEADER CONNECTOR,CABLE MNT,PLUG,50 CONTACTS,SKT,0.1 PITCH,IDC TERMINAL,GRAY
3M

87150-3050

HEADER CONNECTOR,CABLE MNT,PLUG,50 CONTACTS,SKT,0.1 PITCH,IDC TERMINAL,GRAY
3M

87150-3600

Board Connector, 50 Contact(s), 2 Row(s), Female, 0.1 inch Pitch, IDC Terminal, Gray Insulator, Plug
3M

87150-3650

HEADER CONNECTOR,CABLE MNT,PLUG,50 CONTACTS,SKT,0.1 PITCH,IDC TERMINAL,GRAY
3M

87150-3900

HEADER CONNECTOR,CABLE MNT,PLUG,50 CONTACTS,SKT,0.1 PITCH,IDC TERMINAL,GRAY
3M

87150-3950

HEADER CONNECTOR,CABLE MNT,PLUG,50 CONTACTS,SKT,0.1 PITCH,IDC TERMINAL,GRAY
3M

87150-5

HANDRAD VOLL MIT KURBEL DM100150MM
ETC

871503

Board Connector, 3 Contact(s), 1 Row(s), Male, Straight, Solder Terminal,
MOLEX

871503-10A

Board Connector, 3 Contact(s), 1 Row(s), Male, Straight, Solder Terminal
MOLEX