MPS-120-3.3 [ASTRODYNE]
120W 1~4 Output for Medical Type; 120W 1 〜 4输出的医用型型号: | MPS-120-3.3 |
厂家: | ASTRODYNE CORPORATION |
描述: | 120W 1~4 Output for Medical Type |
文件: | 总3页 (文件大小:391K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
mps/d/t/q-120 ꢀerieꢀ
120W 1~4 Output for Medical Type
Features
Universal AC�input�/�Full�range
Low�leakage�current�<180 A
Protections:Short�circuit/Over�load/Over�voltage
UL2601-1�medical�safety approved
With�power�good�and�fail�signal�output�(Optional)
100%�full�load�burn-in�test
Fixed�switching�frequency�at�45KHz
3�years�warranty
Specification
AC�INPUT�VOLTAGE
90~264�VAC,�47~440Hz�/�127~370VDC.
TEMP.�COEFFICIENT
0.04%/ on�all�outputs�at�full�load�between�0~50 of�ambient�temperature.
AC�INPUT�CURRENT (Typ.)
Maximum�input�current�2.9A at�115VAC,�60Hz�or
1.7A at�230VAC,�60Hz�with�100%�output�load.
VIBRATION
2G�of�acceleration,�vibrating�frequency�adjust�from�10Hz�~500Hz�within�a
10-minute�cycle.�6�testing�cycles�(60�minutes)�each�along�X, Y,�Z�axes.
INRUSH�CURRENT (Typ.)
SAFETY STANDARDS
Medical : UL2601-1, TUV�EN60601-1�approved
Commercial�: Also�design�refer�to�UL60950-1, TUV�EN60950-1
Inrush�current�is�less�than�22A at�115VAC�or�less�than
45A at�230VAC�under�cold�start�conditions.�Limiting
provided�by�internal�thermistors.
WITHSTAND�VOLTAGE
SETUP,�RISE�TIME
MPS-120:
4000VAC�between�input�and�output
1500VAC�between�input�and�F.G.
500VAC�between�output�and�F.G.
800ms,�20ms�/�230VAC�at�full�load
2000ms,�50ms�/�115VAC�at�full�load
MPD,T,Q-120:500ms,�20ms�/�230VAC�at�full�load
1200ms,�50ms�/�115VAC�at�full�load
ISOLATION�RESISTANCE
>100M�Ohms�for�I/P-O/P,�I/P-FG,�O/P-FG�by�using�500VDC�test�voltage.
HOLD-UP TIME (Typ.)
80ms�/�230VAC�at�full�load
16ms�/�115VAC�at�full�load
EMI�COMPLIANCE
EMI�Specifications
Conducted�&�Radiation
Compliance�Level
EN55011,�Class�B
EN55022,�Class�B
EN61000-3-2
LEAKAGE�CURRENT
Leakage�current�is�less�than�180 A at�264VAC
Harmonic�distortion
Voltage�flicker
EN61000-3-3
DC�OUTPUT ADJ.�RANGE
DC�output�voltage�(or�Ch1�of�multiple�output�models)
can�be�adjusted�between�-5%~+10%�rated�output
voltage�by�potential�meter.
EMS�COMPLIANCE
EMS�Specification
ESD�air
ESD�contact
RF�field�susceptibility
Compliance�Level
EN61000-4-2, Level�3,�8KV
EN61000-4-2, Level�2,�4KV
EN61000-4-3, Level�2,�3V/m
Level�3,�10V/m
EN61000-4-4, Level�2,�1KV/5KHz
Level�3,�2KV/5KHz
OVER�LOAD�PROTECTION
Fully�protected�against�short�circuit�and�output�overload.
The�hiccup�type�protection will�be�activated�at�110~150%
(For�MPD,T,Q-120),120~160%(For�MPS-120)�rated�load
and�recovers�automatically�after�fault�condition�is removed.
EFT(Electrical�Fast Transient)/Burst
Lightning/Surge
EN61000-4-5, Level�4,�2KV/Line-Line
4KV/Line-Earth
OVER�VOLTAGE�PROTECTION
EN61000-4-6, Level�2,�3Vrms/m
Level�3,�10Vrms/m
EN61000-4-8, Level�2,�3A/m
Level�3,�10A/m
Conducted�RF�susceptibility
Magnetic�field�immunity
Provided�on�output�channel�1�only�at�115%~135%
rated�output�voltage.�(120%~140%�for�MPS-120-15/24/48).
Output�will�be�shut�down�when�this�protection�is�activated.
EN61000-4-11,Compliance
ENV50204, Level�2,�3V/m,�900MHz
Level�3,�10A/m,�900MHz
Voltage�dip,�interruption
Digital�phone�carrier�immunity
POWER�GOOD�/�FAIL SIGNAL (OPTIONAL)
TTL logic�high�for�power�good�and TTL low�for
power�fail.�When�the�output�voltage�reaches�90%�of
rated�value,�a�+5V TTL signal�will�be�sent�out�with�a
10~500ms�delay; At�least�1ms�before�the�output
voltage�goes�below�90%�of�the�rated�value,�the TTL
signal�will�be�turned�off.
MTBF
262,100�hours�min.�at�full�load�and�25 of�ambient��temperature,
calculated�per�MIL-HDBK-217F.
DIMENSION�(L*W*H)
177.8x107.95x35.5mm�or�7"x4.25"x1.4"
*�MPS-120-3.3�does�not�have�this�optional�function.
300 myleꢀ sꢁanꢂiꢀh Blvꢂ., taunꢁon, mA 02780
ph: 1-800-823-8082 Fax: 508-823-8181
WWW.AstROdYNE.COm
mps/d/t/q-120 ꢀerieꢀ
120W 1~4 Output for Medical Type
Output�Chart
OUTPUT�CURRENT
RIPPLE�&�NOISE
(Max.)
VOLTAGE
TOLERANCE
(Note�3)
OUTPUT RATED
VOLTAGE CURRENT
LINE
LOAD
PEAK�LOAD
WITH�25CFM�FAN
(Note�4)
MODEL
EFFICIENCY
MINIMUM
LOAD
WITH�FAN
REGULATION REGULATION
CONVECTION
(Note�2)
(25CFM)
(max.)
MPS-120-3.3 3.3V
24A
22A
10A
8A
0A
0A
0A
0A
0A
0A
16A
24A
26A
26A
11A
80mVp-p
80mVp-p
100mVp-p
100mVp-p
120mVp-p
120mVp-p
3.0%
3.0%
2.0%
2.0%
2.0%
2.0%
1.0%
1.0%
1.0%
1.0%
1.0%
1.0%
3.0%
3.0%
2.0%
2.0%
2.0%
2.0%
68%
73%
77%
79%
81%
82%
MPS-120-5
MPS-120-12
MPS-120-15
MPS-120-24
MPS-120-48
5V
14.7A
6.7A
5.3A
3.3A
1.7A
22A
12V
15V
24V
48V
10A
8A
8.8A
5.5A
2.8A
5A
5A
2.5A
2.5A
MPD-120A
MPD-120B
5V
10A
5A
2A
7.3A
3.6A
7A
10A
5A
12A
6A
80mVp-p
120mVp-p
80mVp-p
250mVp-p
2.0%
7.0%
2.0%
8.0%
0.5%
2.0%
0.5%
2.0%
0.5%
3.5%
0.5%
4.0%
75%
76%
12V
5V
0.5A
2A
10A
2.9A
10A
2.9A
12A
3.2A
24V
0.3A
1.9A
MPT-120A
MPT-120B
MPT-120C
MPT-120D
5V
10A
4.8A
0.6A
10A
4.4A
0.6A
10A
4A
2A
7.3A
3.5A
0.3A
7.3A
3.2A
0.4A
7A
10A
4.8A
0.6A
10A
4.4A
0.6A
10A
4A
12A
5.8A
1A
80mVp-p
120mVp-p
80mVp-p
80mVp-p
120mVp-p
80mVp-p
80mVp-p
150mVp-p
80mVp-p
80mVp-p
300mVp-p
120mVp-p
2.0%
+8,-6%
5.0%
0.5%
1.5%
0.5%
0.5%
1.5%
0.5%
0.5%
2.0%
0.5%
0.5%
3.0%
0.5%
1.0%
3.5%
1.0%
1.0%
3.5%
1.0%
1.0%
3.5%
1.0%
1.0%
+4,-3%
1.0%
12V
-5V
5V
0.4A
0A
72%
73%
72%
74%
2A
12A
5.3A
1A
2.0%
12V
-12V
5V
0.4A
0A
6.0%
5.0%
2A
11A
4.4A
1A
2.0%
15V
-15V
5V
0.4A
0A
2.6A
0.4A
7.3A
1.6A
0.4A
+6,-7%
5.0%
0.6A
10A
2.2A
0.6A
0.6A
10A
2.2A
0.6A
2A
12A
2.64A
1A
2.0%
24V
12V
0.4A
0A
+8,-6%
5.0%
MPQ-120B
MPQ-120C
MPQ-120D
MPQ-120E
5V
10A
4.2A
0.6A
0.6A
10A
3.2A
0.6A
0.6A
10A
1A
2A
7.3A
3.1A
0.4A
0.4A
7.3A
2.4A
0.4A
0.4A
7A
10A
4.2A
0.6A
0.6A
10A
3.2A
0.6A
0.6A
10A
1A
11A
5A
80mVp-p
120mVp-p
80mVp-p
80mVp-p
80mVp-p
150mVp-p
80mVp-p
80mVp-p
80mVp-p
150mVp-p
300mVp-p
80mVp-p
80mVp-p
120mVp-p
80mVp-p
80mVp-p
2.0%
6.0%
5.0%
5.0%
2.0%
+6,-7%
5.0%
5.0%
2.0%
+8,-6%
8.0%
5.0%
2.0%
6.0%
8.0%
5.0%
0.5%
1.5%
0.5%
0.5%
0.5%
2.0%
0.5%
0.5%
0.5%
2.0%
2.0%
0.5%
0.5%
2.0%
2.0%
0.5%
0.5%
3.5%
1.0%
1.0%
0.5%
3.5%
1.0%
1.0%
0.5%
3.5%
3.5%
1.0%
0.5%
3.0%
3.0%
1.0%
12V
-5V
-12V
5V
0.5A
0A
71%
71%
74%
73%
1A
0A
1A
2A
11A
3.8A
1A
15V
-5V
-15V
5V
0.5A
0A
0A
1A
2A
11A
1.1A
2.3A
1A
12V
24V
-12V
5V
0.2A
0.3A
0A
0.7A
1.4A
0.3A
7.3A
2.3A
0.4A
0.4A
2.1A
0.6A
10A
3A
2.1A
0.6A
10A
3A
2A
11A
3.3A
1A
12V
15V
24V
0.5A
0A
0.6A
0.6A
0.6A
0.6A
0A
1A
300 myleꢀ sꢁanꢂiꢀh Blvꢂ., taunꢁon, mA 02780
ph: 1-800-823-8082 Fax: 508-823-8181
WWW.AstROdYNE.COm
mps/d/t/q-120 ꢀerieꢀ
120W 1~4 Output for Medical Type
Mechanical�Specification
Unit:mm
177.8[7"]
165.1�[6.5"]
CN1
5
4
3
2
1
13
12
11
10
9
8
7
6
5
4
3
2
1
+V ADJ
CN2
4- 3.96[0.156"]
FAN
Pin Assignment
AC�Intput�Connector�(CN1)�:�Molex�5273-05�or�equivalent
Pin�No.
MPS-120
FG
MPD-120
FG
MPT-120
FG
MPQ-120
FG
Mating�Housing
Terminal
1
2,4
3
No�Pin
AC/N
No�Pin
AC/N
No�Pin
AC/N
AC/L
No�Pin
AC/N
Molex�5195
or�equivalent
Molex�5194
or�equivalent
5
AC/L
AC/L
AC/L
DC�Output�Connector�(CN2)�:�Molex�5273-13�or�equivalent
Pin�No.
MPS-120
+V
MPD-120
V1
MPT-120
V1
MPQ-120
V1
Mating�Housing
Terminal
1
2
+V
V1
V1
V1
3
+V
V1
V1
V1
4
-V
COM
COM
COM
COM
V2
COM
COM
COM
COM
V2
COM
COM
COM
COM
V2
5
-V
6
-V
Molex�5195
or�equivalent
Molex�5194
or�equivalent
7
-V
8
+V
9
+V
V2
V2
V2
10
11
12
13
P.F.D.
NC
P.F.D.
NC
P.F.D.
V3
P.F.D.
V3
No�Pin
NC
No�Pin
NC
No�Pin
NC
No�Pin
V4
Derating�Curve
Static�Characteristics
100
90
80
70
60
50
40
100
25CFM�forced�air�cooling
80
MPS/D-120
Convection�cooling
53W�MPS-120-3.3�only
73.7W�MPS-120-5�only
80W�others
60
40
MPT/Q-120
20
-20
0
10
20
30
40
50 55 60
70
90
100
115
135
155
230 264
AMBIENT�TEMPERATURE�(
)
INPUT�VOLTAGE�(V)�60Hz
300 myleꢀ sꢁanꢂiꢀh Blvꢂ., taunꢁon, mA 02780
ph: 1-800-823-8082 Fax: 508-823-8181
WWW.AstROdYNE.COm
相关型号:
MPS-172208-82
Narrow Band Medium Power Amplifier, 1900MHz Min, 2000MHz Max, 1 Func, GAAS, 10 PIN
IXYS
MPS-1820A9-82
Narrow Band Medium Power Amplifier, 1850MHz Min, 1910MHz Max, 1 Func, GAAS, 10 PIN
IXYS
©2020 ICPDF网 联系我们和版权申明