The MBC40 Series of open-frame medical power supplies, with its
wide universal 90-264 VAC input range, is available at 40 W of output
power and a variety of single and multiple output voltages.
The MBC series is designed and approved to the latest Medical
standards (EN/IEC 60601-1), providing 2 x MOPP isolation for Class I
& Class II applications.
These medical power supplies are ideal for, monitoring, home health
equipment as well as surgical devices.
4 x 2 x 1.2 Inch Form Factor
Single to Triple Outputs
•
•
•
•
•
•
•
Ultra High Efficiency > 85%
Approved to EN/IEC 60601-1
µ
Low leakage current < 250 A
No Load Power < 0.3 W
IEC Protection Class Options:
o
Class I: Earth pin J4 (no suffix)
Class II: No Earth pin (-2 suffix)
o
RoHS Compliant
•
•
Cover Kit Accessory Available
Dialysis
Surgical Devices
Home Health
Ultrasound
•
•
•
•
•
•
Monitoring
Pumps
2
MBC40 Series
MODEL1
OUTPUT VOLTAGE (VDC)
MAX LOAD (A)2
MINIMUM LOAD (A)3
RIPPLE & NOISE 4
Specifications are for nominal input voltage, 25°C unless otherwise stated.
PARAMETER
Input Voltage
DESCRIPTION / CONDITION
SPECIFICATION
Universal
90 - 264 VAC
Input Frequency6
47 to 400 Hz
0.85 A max.
0.45 A max.
< 0.3 W
120 VAC:
230 VAC:
Single output models
Multi output models
120 VAC:
230 VAC:
120 VAC:
230 VAC:
Input Current
No Load Power
Inrush Current
< 0.5 W
30 A max.
60 A max.
< 140 µA
<250 µA
Leakage Current
Switching Frequency
Typical
67 Hz
1 For Class II add suffix -2 (e.g.: MBC40-1012G-2).
2 Maximum current per output channel. Do not exceed total output power rating.
3 Minimum load specified to meet cross regulation.
4 Ripple is peak to peak with 20 MHz bandwidth and 10 µF (Tantalum capacitor) in parallel with a 0.1 µF capacitor at rated line voltage
and load ranges.
5 For MBC40-3003G efficiency is 75% typical.
6 Safety Approved: 47 to 63 Hz
tech.support@psbel.com
3
MBC40 Series
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
Efficiency
Typical
85%
Hold Up Time
Output Power
Line Regulation
@ 120 VAC typical
>10 ms
40 W
Derate output power linearly to 80% from 90 VAC to 80 VAC input.
+/-0.3%
V1:
V2 & V3:
+/-0.5%
+/-5%
Load Regulation
< 10%,
recovery time < 5 ms
50% to 100% load change, 50/60 Hz, 50% duty cycle, 0.1 A/μs
Transient Response
Rise Time
< 100 ms
V1:
V2 & V3:
± 3%
± 5%
Set Point Tolerance
Output Voltage Adjustment
Over Current Protection
Over Voltage Protection
Short Circuit Protection
V1
± 10%
130%
130%
Typical above rating
Typical for V1 only
Short term, autorecovery
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
Refer to derating curve, Fig. 1
Start-up is guaranteed
-20 to 70°C
-20 to 0°C
Operating Temperature
Storage Temperature
Relative Humidity
-40 to +85°C
Non Condensing
95%
Operating:
Non-Operating:
10,000 ft.
40,000 ft.
Altitude
Reliability
Cooling
MTBF according to Telcordia -SR332-Issue 3
1.87 million hours
Convection
Figure 1. Derating Curve
North America
+1 408 785 5200
Asia-Pacific
+86 755 298 85888
Europe, Middle East
+353 61 225 977
© 2017 Bel Power Solutions & Protection
BCD.00286_AE
4
MBC40 Series
Conducted Emissions
EN 55011-B, CISPR22-B, FCC PART15-B
EN 55011 B
Pass
Radiated Emissions
Pass
Input Current Harmonics
Voltage Fluctuation and Flicker
ESD Immunity
EN 61000-3-2
Class D
EN 61000-3-3
Pass
EN 61000-4-2
Level 4, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 3, Criterion A
Level 4, Criterion A
Criterion A & B
Radiated Field Immunity
Electrical Fast Transient Immunity
Surge Immunity
EN 61000-4-3
EN 61000-4-4
EN 61000-4-5
Conducted Immunity
EN 61000-4-6
Magnetic Field Immunity
Voltage Dips, Interruptions
EN 61000-4-8
EN 61000-4-11
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
Input to Output, 2M0PP
Input to Ground, 1M0PP
Min. 5700 VDC
1500 VAC
Isolation Voltage
Output to Ground
500 VAC
Approved to the latest edition of the following standards:
CSA/UL60601-1, EN60601-1 and IEC60601-1.
Safety Standards
Agency Approvals
CE mark
Nemko, UL, C-UL
Complies with LVD Directive
CONNECTOR
PIN
DESCRIPTION / CONDITION
MANUFACTURER / PN
Molex: 26-60-4030 or equivalent
Mating: 09-50-3031; Pins: 08-50-0106
Pin 1
Pin 2
Pin 1,2
Pin 3,4
Pin 5
AC Neutral
AC Line
V1
AC Input Connector
J1
RTN
Tyco: 640445-6 or equivalent
Mating: 647402-6; Pins: 3-647409-1
DC Output Connector
J2
V3
Pin 6
Pin 1
Pin 2
V2
+V1 Sense
-V1 Sense
Molex: 22-23-2021 or equivalent
Mating: 22-01-2021
Signal Connector
Earth
J3
J4
Molex: 19705-4301
Mating: 190030001
PARAMETER
Weight
DESCRIPTION / CONDITION
150 g (0.33 lbs.)
Dimensions
101.6 x 50.8 x 30.48 mm (4 x 2 x 1.2 inch)
tech.support@psbel.com
5
MBC40 Series
Figure 2. Mechanical Drawing MBC40-1xxxG
Figure 3. Mechanical Drawing MBC40-3xxxG
NOTES: In case the PCB is mounted in a metal enclosure, using metal hardware ensure the following:
1
2
3
Stand off, used to mount PCB has OD of 5.4 mm max.
Screws, used to fix PCB on stand off, have head dia of 6.0 mm max.
Washer, if used, to have dia of 6.5 mm max.
NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support systems,
equipment used in hazardous environments, or nuclear control systems.
TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the
date manufactured. Specifications are subject to change without notice.
North America
Asia-Pacific
Europe, Middle East
+1 408 785 5200
+86 755 298 85888
+353 61 225 977
© 2017 Bel Power Solutions & Protection
BCD.00286_AE
相关型号:
©2020 ICPDF网 联系我们和版权申明