MBE1000 [BEL]
Form Factor 5 x 9.51 x 1.61 in (127 x 241.5 x 41 mm);型号: | MBE1000 |
厂家: | BEL FUSE INC. |
描述: | Form Factor 5 x 9.51 x 1.61 in (127 x 241.5 x 41 mm) |
文件: | 总8页 (文件大小:711K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
The MBE1000 Series medical power supplies, with its wide universal
85 - 264 VAC input range, is available at 1000 W of output power and
a variety of single output voltages.
The MBE series is designed and approved to the latest Medical
standards (EN/IEC 60601-1), providing 2 x MOPP isolationfor Class I
& Class II applications.
These medical power supplies are ideal for monitoring, home health
equipment as well as surgical devices.
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Form Factor 5 x 9.51 x 1.61 in (127 x 241.5 x 41 mm)
Approved to EN 60601
Dual Fusing
Current Sharing Option
Peak Power Capability
Standard IEC 60601-1-2: 2014 (4th Edition)
5 VDC Stand by output
In built 12 V fan output
Power Good / Power Fail Signal
Suitable for BF application
Lesser than 1U high
High voltage output range up to 58 VDC
N+1 redundant power supply
Single wire current sharing
Built in OR-ing diode / FET (- R suffix
•
•
•
•
Diagnostic
Drug Pump
Dialysis
•
•
•
•
Home Health Care
Monitoring
Imaging
Hospital Beds
Therapy Devices
2
MBE1000 Series
MBE1000-1T12
MBE1000-1T15
MBE1000-1T24
MBE1000-1T30
MBE1000-1T48
MBE1000-1T58
12 V
15 V
24 V
30 V
48 V
58 V
Fan Cooled
Fan Cooled
Fan Cooled
Fan Cooled
Fan Cooled
Fan Cooled
41.67 A
41.67 A
41.67 A
33.33 A
20.83 A
17.24 A
0.0 A
0.0 A
0.0 A
0.0 A
0.0 A
0.0 A
2%
2%
2%
2%
2%
2%
1
For Ripplemeasurement minimumoutput power requirement is 25 W.
Ripple is peak to peak with 20 MHz bandwidth and10 µF (Electrolyticcapacitor)in parallel with a 0.1 µF capacitor at rated linevoltageand
load ranges.
Specifications are for nominal input voltage, 25°C unless otherwise stated.
Input Voltage
Universal
85 – 264 VAC / 120 – 390 VDC
47 – 63 Hz
Input Frequency
120 VAC:
240 VAC:
11 A max.
5.5 A max.
Input Current
Input Protection
No Load Power
In Live & Neutral both
F16 A / 250 V
Over entire input range with main output kept OFF using
Remote ON/OFF
3 W typ.
Inrush Current
Leakage Current
Touch Current:
240 VAC:
25 A max.
400 µA
240 VAC / 50 Hz
< 100 µA
120 VAC:
240 VAC:
0.98
0.95
Power Factor
PFC converter: Variable
Resonant converter: Variable
85 kHz typical
100 kHz typical
Switching Frequency
tech.support@psbel.com
3
MBE1000 Series
Output Voltage
Refer to Model Selection table for details
12 V to 58 V
5 V
Standby Output Voltage2
Output Power3
Fan Cooled (Peak 1200 W for 1 ms)
1000 W
120 VAC:
240 VAC:
88 % Typical
93 %
Efficiency
Hold-up Time
Line Regulation
Load Regulation
120 VAC / 240 VAC:
8 ms
± 0.5 %
± 1.0 %
< 10 %,
recovery time < 5 ms
Transient Response
50% to 100% load change, 50 Hz, 50% duty cycle, 0.1 A/µs
Voltage Adjustment
Set Point Tolerance
Rise Time
± 3 %
± 1%
<100 ms
120 to 150 %
114%
Over Current Protection
Over Voltage Protection
Short Circuit Protection
Over Temperature Protection
Current Share
Hic-Up Type, autorecovery
Latch Type, AC Power to be recycled for recovery
Short term, autorecovery
Autorecovery
130-140°C primary heat sink
Up to 3 supplies connected in parallel (optional)
Fan cooled
Cooling
1000 W
2
Standby output voltage5 V / 1.5 A with toleranceincluding set point accuracy, lineand load regulation is +/-10%.Rippleandnoiseis less
than 5%.
Combinedoutput power of main output,fan supply and standby supply shall not exceed max. power rating.
3
Conducted Emissions
Radiated Emissions
EN 55011-B, CISPR22-B, FCC PART15-B
Class B
EN 55011
Class A
Class B
With External king core K5BRC 25x12x15-M or equivalent
Input Current Harmonics
Voltage Fluctuation and Flicker
ESD Immunity
EN 61000-3-2
EN 61000-3-3
EN 61000-4-2
EN 61000-4-3
EN 61000-4-4
EN 61000-4-5
EN 61000-4-6
EN 61000-4-8
EN 61000-4-11
Class A
Complies
A
Radiated Field Immunity
Electrical Fast Transient Immunity
Surge Immunity
A
A
A
Conducted Immunity
A
Magnetic Field Immunity
Voltage Dips, Interruptions
A
A & B
North America
Asia-Pacific
Europe, Middle East
+86 755 298 85888
+353 61 225 977
+1 408 785 5200
© 2020 Bel Power Solutions & Protection
BCD.00973_D
4
MBE1000 Series
Input to Output
Input to Earth
Output to Earth
4245 VAC
1625 VAC
1500 VAC
Isolation Voltage
EN 60601-1, IEC 60601-1 (ed.3), ANSI/AAMI ES 60601-1,
CSA C22.2 No. 60601-1
Safety Standard(s)
Agency Approvals
CE mark
Nemko, UL, C-UL
Complies with LVD Directive
Operating Temperature
Storage Temperature
Relative Humidity
Refer to derating curve
Non-condensing
-40 to +70°C
-40 to +85°C
5% to 95%
Operating:
16,000 ft.
40,000 ft.
Altitude
MTBF
Non-operating:
Telcordia -SR332-issue 3
3.37 million hours
Power Good: Is a TTL signal which goes high after main output reaches 90% of its set value.
The delay is 0.1 s to 0.5 s
Power Fail: Thesamesignalgoes lowat least 1ms before main output falls to 90% ofset value at AC Power off
Power Good / Fail Signal
Remote Sense
Compensates for 200 mV drop
Pin 6 & Pin 7 of J3 can be used for Remoteon/off.
Shorting Pin 6 to Pin 7 enables main output while keeping the pins open disables main output
Note: Provision of Inhibit RemoteON/OFF is available. +5 V at Pin 7 will switch offthe main output.
Remote On / Off
OCP Limit Set
Pin 8 & Pin 9 of J3 must be shorted
Figure 1. Power Good / Fail Signal Diagram
tech.support@psbel.com
5
MBE1000 Series
Pin 1
Pin 2
Pin 3
AC Line
Neutral
Earth
AC Input Connector
J1
J2
TE Connectivity: NC6-P107-03
6-32 inches Screw Pan HD
Mating: Designed to accept Ring Tongue Terminal AMP: 8-31886-1,
wherein one 16 AWG (max) wire can be crimped.
Note: One Ring TongueTerminal with 16 AWG is recommended for
current up to 11 A only.
DC Output
Connector
J2-A
J2-B
+VE
-VE
Use multiple tongue terminals with wire for more current.
Pin 1
Pin 2
Pin 3
Pin 4
Pin 5
Pin 6
Pin 7
Pin 8
Pin 9
Pin 10
GND
5V AUX
PGPF
VS –
VS +
GND
RMT
CL2
Molex: 22-23-2101
Mating: 22-01-2107; Pins: 08-50-0113
Signal Connector
J34
CL1
LS
TE Connectivity AMP Connectors
Description: CONN HEADERVERT 2 POS 2.54 mm,
Pin 1
Pin 2
+VE
-VE
Fan Output5
J10, J11
MPN: 640456-2
Mating: 3-641535-2 / TE Connectivity AMP Connectors
or: 0022013027 / MOLEX with crimping 08-50-0114 / MOLEX
4
PSU is suppliedwith J3, pin-6 andpin-7 shortedto enable main output without remoteon/off feature.
Fan supply output voltageis 12 V / 500 mA with regulation band ±30 % andRipple is less than 10%.To get 12 V Fan supply output voltage,
minimum10 % load on Main output voltageis required.
5
Figure 1. Power Derating w.r.t Input
North America
Asia-Pacific
Europe, Middle East
+86 755 298 85888
+353 61 225 977
+1 408 785 5200
© 2020 Bel Power Solutions & Protection
BCD.00973_D
6
MBE1000 Series
De-rate between 30-50 °C @ 0.833% per °C
De-rate above 50 °C @ 2.5% per °C
Figure 2. Power Derating Curve 12 V
De-rate between 30-50 °C @ 0.833% per °C
De-rate above 50 °C @ 2.5% per °C
Figure 3. Power Derating Curve 15 V
De-rate between 30-50 °C @ 0.833% per °C
De-rate above 50 °C @ 2.5% per °C
Figure 4. Power Derating Curve 24 V and above
tech.support@psbel.com
7
MBE1000 Series
Weight
1300 g
Dimensions
127 x 241.5 x 41 mm (5 x 9.51 x 1.61 inches)
Figure 6. Mechanical drawings
North America
Asia-Pacific
Europe, Middle East
+86 755 298 85888
+353 61 225 977
+1 408 785 5200
© 2020 Bel Power Solutions & Protection
BCD.00973_D
8
MBE1000 Series
During the installation and setup of parallel supplies in a system it is important that a single remote sense point be used for all the
supplies. The remote sense voltage between the supplies must be adjusted to within 1% to ensure the supplies are inside the 1%
capture window. If the supplies are not initially adjusted inside the capture window the supplies will not current share satisfactorily.
SET-UP PROCEDURE:
1. Connect load cables to the outputs of each supply.
2. Connect the remote sense lines to the load in twisted style. (A common remote sense point must be used for all the supplies in
parallel).
3. Connect all the “LS” signal (Pin 10) on the J3 connector between the supplies.
4. Adjust remote sense voltage of each supply to within 1% of rated output voltage or readjust to required set point. (Adjustment
to be done with all other parallel supplies off).
5. Current sharing between the supplies can be verified by monitoring the output current of each supply with a hall effect DC
current probe. The supplies should share to within 10% of the total load current.
6. The current share circuit has a capture window voltage of +/- 1% of the rated output voltage. If the output remote sense
voltage of one of the supplies is adjusted outside the 1% window the supplies will not current share satisfactorily.
Figure 7. Current Sharing Block Diagram
NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for useas critical components in life support systems,
equipment used in hazardous environments, or nuclear control systems.
TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the
date manufactured. Specifications are subject to change without notice.
tech.support@psbel.com
相关型号:
MBER10100CT
Rectifier Diode, Schottky, 1 Phase, 2 Element, 10A, 100V V(RRM), Silicon, TO-220AB, ROHS COMPLIANT, PLASTIC PACKAGE-3
DIODES
MBER1080CT
Rectifier Diode, Schottky, 1 Phase, 2 Element, 10A, 80V V(RRM), Silicon, TO-220AB, ROHS COMPLIANT, PLASTIC PACKAGE-3
DIODES
MBER1090CT
Rectifier Diode, Schottky, 1 Phase, 2 Element, 10A, 90V V(RRM), Silicon, TO-220AB, ROHS COMPLIANT, PLASTIC PACKAGE-3
DIODES
©2020 ICPDF网 联系我们和版权申明