TR70MA180 [CINCON]

70W MEDICAL SWITCHING ADAPTER;
TR70MA180
型号: TR70MA180
厂家: Cincon    Cincon
描述:

70W MEDICAL SWITCHING ADAPTER

开关
文件: 总3页 (文件大小:269K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
V12  
TR70M SERIES  
70W MEDICAL SWITCHING  
ADAPTER  
Features  
* Universal Input Range 80~264VAC  
* Meets EN55011 Class B and CISPR/FCC Class B  
* Continuous Short Circuit Protection  
* Over Voltage Protection  
* No Load Power Consumption <150mW  
* IEC/EN 60601-1-11 for TR70MB  
* Meets CoC Tier 2 & DoE Level VI  
* Meets IEC/EN60335-1  
* Meets 2 MOPP  
* Altitude 5000m  
* Class I (TR70MA) & Class II (TR70MB)  
(TR70MA/B120:OutputCable Length 1220mm16AWG) (note7)  
(TR70MA/B150: Output Cable Length 1220mm 16AWG)  
(TR70MA/B180: Output Cable Length 1800mm 16AWG)  
(Output Cable Length 1800mm 18AWG)  
Ordering information  
TR70MXXXX  
XX  
DC Plug Type OVP  
E: With OV  
X
XX  
Model No.  
DCCableLengthandType  
01: 720mm  
X: A or B  
02: 1220mm  
A:CLASS I  
B:CLASS II  
03: 1800mm  
11: 720mm with Ferrite Core  
12:1220mm with Ferrite Core  
13:1800mm with Ferrite Core  
*18AWG/UL1185  
*16AWG/UL1185 for 12V  
RIPPLE&  
NOISE  
VOLTAGE  
SETPOINT  
NOTE1  
LINE  
LOAD  
EFFICENCY  
OUTPUT  
OUTPUT  
MODEL  
REGULATION REGULATION  
(Typ.)  
VOLTAGE  
CURRENT  
NOTE2  
NOTE3  
1%  
1%  
1%  
1%  
1%  
1%  
1%  
1%  
1%  
1%  
1%  
1%  
NOTE4  
5%  
3%  
2%  
2%  
2%  
2%  
5%  
3%  
2%  
2%  
2%  
2%  
NOTE5  
TR70MA120  
TR70MA150  
TR70MA180  
TR70MA240  
TR70MA360  
TR70MA480  
TR70MB120  
TR70MB150  
TR70MB180  
TR70MB240  
TR70MB360  
TR70MB480  
12V  
15V  
18V  
24V  
36V  
48V  
12V  
15V  
18V  
24V  
36V  
48V  
5.8A  
4.65A  
3.9A  
3.0A  
1.9A  
1.5A  
5.8A  
4.65A  
3.9A  
3.0A  
1.9A  
1.5A  
120mVp-p  
150mVp-p  
180mVp-p  
240mVp-p  
360mVp-p  
480mVp-p  
120mVp-p  
150mVp-p  
180mVp-p  
240mVp-p  
360mVp-p  
480mVp-p  
2%  
2%  
2%  
2%  
2%  
2%  
2%  
2%  
2%  
2%  
2%  
2%  
89%  
89%  
89%  
90%  
90%  
91%  
89%.  
89%  
89%  
90%  
90%  
91%  
Specifications are subject to change without notice.  
1
V12  
Specifications  
TR70M Series Derating Curve  
INPUT SPECIFICATIONS:  
Voltage …....………........…............. 80~264Vac(80Vac with derating curve)  
Frequency ………………………………..………………………..... 47 to 63Hz  
Inrush Current ……………….…… Cold Start @25 100A max. @240Vac  
Conducted EMI …………………………….…………… CISPR/FCC Class B  
Leakage Current ……………………..………………….……..…… 90uA max.  
OUTPUT SPECIFICATIONS:  
Hold-up Time ………………………………………………10ms typ. @115Vac  
Short Circuit Protection………..……………..Hiccup Mode (Auto Recovery)  
Over Voltage Protection ……….. IC Component to Clamp (Auto Recovery)  
Temperature Coefficient ………..………..........................… 0.05%/ max.  
GENERAL SPECIFICATIONS:  
Isolation ………………………………………… Input to Output = 4,400VAC  
Operating Temperature ................. -20 ~ 70 (See Derating Curve, note6)  
Storage Temperature ……………………………...….......……........ -40~85  
Humidity …………………...….......……........ 93% RH max. Non Condensing  
Cooling……………………………………………………. Natural Convection  
Switching Frequency ……………………………………….… 65KHz Typical  
MTBF ……… MIL-HDBK-217F, GB, at 25 /115VAC ….…. 500Khrs min.  
Altitude …………………………………………………………………… 5000m  
Life Time …………………. Ambient 40 75% Load ………………. >3years  
AC Inlet ……………………IEC320/C14(TR70MA) / IEC320/C8(TR70MB)  
Dimensions ………. 4.724x2.047x1.220 inches (120.00x52.00x31.00 mm)  
Weight …………………………..…………….…………… 300g(0.66 Pounds)  
NOTE:  
1. Voltage accuracy at 60% full load.  
SAFETY AND EMC:  
Emissions and Immunity (Ed 4.0) …… EN55011 Class B, EN61000-3-2,3  
EN60601-1-2, IEC 61000-4-2, 3, 4, 5, 6, 8, 11  
2. Add a 0.1uF ceramic capacitor and a 10uF E.L. capacitor to output for  
ripple&noise measurement @20MHz BW.  
Safety Approved (Ed.3.1) ...... IEC/EN 60601-1, UL ANSI/AAMI ES60601-1  
IEC/EN 60601-1-11 for TR70MB  
3. Line regulation is measured from 100Vac to 240Vac, full load.  
4. Load regulation is measured from 60% to 100% full load and from 60%  
to 20% full load (60% +/- 40% full load).  
5. Typical efficiency at 230VAC and 75% load at 25  
.
6. -30 can be start-up at full load.  
7. TR70MA120 & TR70MB120 of 115VAC full load meet DoE Level VI  
and 230VAC full load meet CoC Tier 2  
Mechanical Specification  
All Dimensions are in inches(mm)  
Tolerance:Inches:X.XXX 0.02  
TR70MAXXX Series  
Millimeters:X.XX 0.5  
4.724[120.00]  
70.866[1800.00] 1.969[50.00]  
UNIT:inches[mm]  
DC Plug type: V+ ─●)─ V-  
DC Plug :Right Angle(ψ5.5  
18AWG 1800mm  
ψ2.1) L12mm  
IEC320/C14  
2
V12  
Mechanical Specification  
TR70MBXXX Series  
4.724[120.00]  
70.866[1800.00] 1.969[50.00]  
-
+
DC Plug type: V+ ─●)─ V-  
DC Plug :Right Angle(ψ5.5  
18AWG 1800mm  
ψ2.1) L12mm  
IEC320/C8  
Typical at 25˚C, nominal line and 75% load, unless otherwise Specified  
3

相关型号:

TR70MA240

70W MEDICAL SWITCHING ADAPTER

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
CINCON

TR70MA360

70W MEDICAL SWITCHING ADAPTER

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
CINCON

TR70MA480

70W MEDICAL SWITCHING ADAPTER

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
CINCON

TR70MB120

70W MEDICAL SWITCHING ADAPTER

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
CINCON

TR70MB150

70W MEDICAL SWITCHING ADAPTER

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
CINCON

TR70MB180

70W MEDICAL SWITCHING ADAPTER

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
CINCON

TR70MB240

70W MEDICAL SWITCHING ADAPTER

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
CINCON

TR70MB360

70W MEDICAL SWITCHING ADAPTER

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
CINCON

TR70MB480

70W MEDICAL SWITCHING ADAPTER

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
CINCON

TR7215

RF AMPLIFIER

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
SPECTRUM

TR7215

RF AMPLIFIER MODEL

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
APITECH

TR7216

RF AMPLIFIER

Warning: Undefined variable $rtag in /www/wwwroot/website_ic37/www.icpdf.com/pdf/pdf/index.php on line 217
-
SPECTRUM