EMMA240063-P6RP-IC [CUI]
AC/DC WALL MOUNT ADAPTER 24V 15W;型号: | EMMA240063-P6RP-IC |
厂家: | CUI INC |
描述: | AC/DC WALL MOUNT ADAPTER 24V 15W |
文件: | 总4页 (文件大小:390K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
Additional Resources: Product Page
date 06/20/2017
page 1 of 4
SERIES: EMMA 15W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
FEATURES
• up to 15 W power
• interchangeable AC blades
• universal input (90~264 Vac)
• single regulated output from 5~24 Vdc
• over voltage and short circuit protections
• UL/cUL, CE safety approvals
• medical 60601-1 4th edition safety approvals
• designed for 2 x MOPP applications
• level V efficiency
ꢀEC
ꢁ0ꢁ01ꢂ1
EN
®
ꢁ0ꢁ01ꢂ1
output
voltage
output
current
max
output
power
max
ripple
efficiency
MODEL
and noise1
level
max
(Vdc)
5
(A)
(W)
(mVp-p)
EMMA050200
EMMA120110
EMMA150100
EMMA240063
2
10
13.2
15
50
V
V
V
V
12
15
24
1.1
120
150
240
1.0
0.625
15
1. at full load, 100 ~ 240 Vac input, 20 MHz bandwidth oscilloscope, each output terminated with a 10 µF aluminum electrolytic and 0.1 µF ceramic capacitors.
PART NUMBER KEY
EMMA120110 X - XX - IC - CXX
Base Number
Reserved for Custom
example of 12 Vdc, 1.1 A
Configurations
DC Plug Type:
Factory
See output plug options
on page 3 for more details
Designation
Blades:
“blank” = North American, European, United Kingdom,
Australian
N = North American blade included
E = European blade included
B = United Kingdom blade included
A = Australian blade included
K = No blades included
cui.com
Additional Resources: Product Page
CUI Inc │ SERIES: EMMA 15W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 06/20/2017│ page 2 of 4
INPUT
parameter
conditions/description
min
90
typ
max
264
63
units
Vac
Hz
voltage
frequency
47
inrush current
no load power consumption
at 240 V ac
50
A
0.3
W
OUTPUT
parameter
conditions/description
min
typ
max
units
line regulation1
±1
%
5 Vdc output
all other outputs
±4
±2
%
%
load regulation2
5 Vdc output
all other outputs
±3
±2
%
%
voltage accuracy
hold-up time
at 115 Vac
10
95
ms
kHz
switching frequency
temperature coefficient
±0.05
%/°C
Note:
1. measured from 100 ~ 240 Vac, full load
2. measured from 60% to full load and from 60 ~ 20% load (60% ±40% load)
PROTECTIONS
parameter
conditions/description
TVS component to clamp
continuous, auto restart
over voltage protection
short circuit protection
SAFETY & COMPLIANCE
parameter
conditions/description
min
typ
max
units
isolation voltage
safety approvals
EMI/EMC
input to output
5,656
Vdc
medical IEC 60601-1, EN 60601-1, UL 60601-1 4th edition
EN5 5011 Class B, FCC Part 15 Class B, EN 60601-1-2, EN 61000-3-(2,3)
leakage current
MTBF
0.1
mA
as per MIL-HDBK-217F, 115 Vac, 25 °C
2011/65/EU
200,000
hours
RoHS
ENVIRONMENTAL
parameter
conditions/description
min
0
typ
max
50
units
°C
operating temperature
storage temperature
humidity
-20
85
°C
non-condensing
93
%
cui.com
Additional Resources: Product Page
CUI Inc │ SERIES: EMMA 15W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 06/20/2017│ page 3 of 4
MECHANICAL
parameter
dimensions
input plug
conditions/description
min
typ
max
units
80.60 x 47.90 x 43.30 (3.173 x 1.886 x 1.705 inch)
interchangeable blades (US, Europe, UK, Australia)
mm
weight
130
g
MECHANICAL DRAWING
units: mm [inches]
tolerance: ±0.5[0.02]
47.90 1.886
43.30 1.705
80.60 3.173
EMM-US
EMM-AU
EMM-EU
EMM-UK
DC OUTPUT PLUG OPTIONS / DC CORD
cord length
ń
ń
Standard PX
Right Angle PXR
MODEL NO.
EMMA050200
EMMA120110
EMMA150100
EMMA240063
CABLE GAUGE
18 AWG
CORD LENGTH
1,800 mm ±50
1,800 mm ±50
1,800 mm ±50
1,800 mm ±50
A
B
C
Unit
mm
mm
P5/P5R
5.5
5.5
2.1
2.5
9.5
9.5
18 AWG
P6/P6R
18 AWG
18 AWG
PXX X X
Plug Type
Plug Angle:
“Blank” = Standard “Blank” = N/A
R = Right Angle P = Center Positive
Plug Polarity:
N = Center Negative
cui.com
Additional Resources: Product Page
CUI Inc │ SERIES: EMMA 15W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 06/20/2017│ page 4 of 4
REVISION HISTORY
rev.
description
date
1.0
initial release
updated P7/P7R B dimension
12/16/2011
03/23/2012
08/21/2012
04/29/2013
07/17/2015
06/20/2017
1.01
1.02
1.03
1.04
1.05
V-Infinity branding removed
updated safety approvals
updated datasheet
updated to medical 60601-1 4th edition
The revision history provided is for informational purposes only and is believed to be accurate.
Headquarters
20050 SW 112th Ave.
Tualatin, OR 97062
800.275.4899
Fax 503.612.2383
cui.com
techsupport@cui.com
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause undesired operation.
CUI offers a one (1) year limited warranty. Complete warranty information is listed on our website.
CUI reserves the right to make changes to the product at any time without notice. Information provided by CUI is believed to be accurate and reliable. However, no responsibility is
assumed by CUI for its use, nor for any infringements of patents or other rights of third parties which may result from its use.
CUI products are not authorized or warranted for use as critical components in equipment that requires an extremely high level of reliability. A critical
component is any component of a life support device or system whose failure to perform can be reasonably expected to cause the failure of the life support device or system, or to
affect its safety or effectiveness.