SERIES: ETMA 200W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
FEATURES
• up to 200 W power
• universal input (90~264 Vac)
• single regulated output from 12~48 V
• over voltage, overload and short circuit protections
• full medical safety approvals
• active PFC
• level V efficiency
• custom designs available
IEC
60601-1
EN
60601-1
RoHS
®
output
voltage
output
current
max
output
power
max
ripple1
efficiency
MODEL
level
max
(Vdc)
12
(A)
(W)
(mVp-p)
ETMA1201666U
ETMA1601250U
ETMA1901053U
ETMA2001000U
ETMA240833U
ETMA360555U
ETMA480416U
16.66
12.5
10.53
10
200
200
200
200
200
200
200
240
300
300
300
300
300
300
V
V
V
V
V
V
V
16
19
20
24
8.33
5.55
4.16
36
48
Notes:
1. At full load, 100 ~ 240 Vac input, 20 MHz bandwidth oscilloscope, each output terminated with 10 μF aluminum electrolytic and 0.1 μF ceramic capacitors.
PART NUMBER KEY
ETMA1201666U C - P51 - WP - CXX
Base Number
Reserved for Custom
example of 12 Vdc, 16.66 A
Configurations
DC Plug Type
Input Cable:
C = included
Factory
Designation
blank = no cable
cui.com
CUI Inc │ SERIES: ETMA 200W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 08/13/2012 │ page 2 of 5
INPUT
parameter
conditions/description
min
90
typ
max
264
63
units
Vac
Hz
A
voltage
frequency
47
current
at 90 Vac, 60 Hz
3
inrush current
no load power consumption
at 230 Vac, cold start
150
0.5
A
W
OUTPUT
parameter
conditions/description
min
typ
±1
±5
max
units
%
line regulation
load regulation
hold-up time
%
at 115 Vac
16
ms
temperature coefficient
±0.05
%/°C
PROTECTIONS
parameter
conditions/description
12 V output
16, 20 V outputs
19 V output
24, 36 V outputs
48 V output
110
110
111
110
110
133
130
137
125
120
%
%
%
%
%
over voltage protection
overload protection
shutdown and auto restart
continuous, auto restart
110
160
%
short circuit protection
SAFETY & COMPLIANCE
parameter
conditions/description
min
typ
max
units
input to output for 2 seconds
input to frame ground for 2 seconds
5,656
2,121
Vdc
Vdc
isolation voltage
input to output, at 500 Vdc
input to frame ground, at 500 Vdc
20
20
MΩ
MΩ
insulation resistance
safety approvals
EMI/EMC
UL/cUL (UL 60601-1), EN 60601-1, IEC 60601-1
FCC 47 CFR Part 18, CE, EN 61000-3-(2, 3), EN 55011, IEC 61000-4-(2, 3, 4, 5, 6, 8, 11)
earth, at 264 Vac
enclosure, at 264 Vac
190
10
300
100
μA
μA
leakage current
MTBF
at 25°C, max. load
yes
140,000
hours
RoHS compliant
ENVIRONMENTAL
parameter
conditions/description
min
0
typ
max
50
units
°C
operating temperature
storage temperature
humidity
-10
20
70
°C
non-condensing
90
%
cui.com
CUI Inc │ SERIES: ETMA 200W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 08/13/2012 │ page 3 of 5
MECHANICAL
parameter
dimensions
input plug
weight
conditions/description
min
typ
max
units
7.42 x 4.11 x 2.36 (188.5 x 104.5 x 60 mm)
IEC320 / C14
inch
1.06
kg
MECHANICAL DRAWING
units: mm
104.5 0.5
60.0 0.5
188.5 0.5
cui.com
CUI Inc │ SERIES: ETMA 200W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 08/13/2012 │ page 4 of 5
DC CORD
PIN 2
Blue (GND)
PIN 1
Red (+Vo)
PIN 4
Black (GND)
PIN 3
Yellow (+Vo)
[2464#16AWG 4C+1(16AWG)]
D.W. (GND)
SECTIONAL DRAWING
4P POWER DIN
FRONT VIEW
Core
14.2*8.0*28.5mm
TERMINAL
ø3.2*2
UL Heat Shrinkable Tube
Black+Blue+D.W.
Yellow+Red
4PIN POWER DIN
COMPONENT LIST
1. Cable: color black, UL/CSA 2464 80/90°C 300V
16 AWG X 4C+Aluminum Shield (16AWG)
2. PVC Wire: color: blue/black/yellow/red
3. Connector: 4PIN power DIN or EQ
4. 4PIN power DIN (female) MDP-002-4PIN (P530600004F0001)
AC CORD
(IEC 320 C13)
(NEMA 5-15P Hospital Grade)
183 5cm
SPECIFICATION:
1)
2)
3)
4)
5)
Length: 1.83m
Color: black
Plug: SL-22 (A)10 (V)125 Approval: UL/cUL
Connector: SL-3(A)10 (V)125 Approval: UL/cUL
Cable: SJT 18AWG/3C Approval: UL/CSA
20 2cm
cui.com
CUI Inc │ SERIES: ETMA 200W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 08/13/2012 │ page 5 of 5
REVISION HISTORY
rev.
description
date
1.0
initial release
03/05/2012
08/13/2012
1.01
V-Infinity branding removed, safety and EMI/EMC data updated
The revision history provided is for informational purposes only and is believed to be accurate.
Headquarters
20050 SW 112th Ave.
Tualatin, OR 97062
800.275.4899
Fax 503.612.2383
cui.com
techsupport@cui.com
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause undesired operation.
CUI offers a two (2) year limited warranty. Complete warranty information is listed on our website.
CUI reserves the right to make changes to the product at any time without notice. Information provided by CUI is believed to be accurate and reliable. However, no responsibility is
assumed by CUI for its use, nor for any infringements of patents or other rights of third parties which may result from its use.
CUI products are not authorized or warranted for use as critical components in equipment that requires an extremely high level of reliability. A critical
component is any component of a life support device or system whose failure to perform can be reasonably expected to cause the failure of the life support device or system, or to
affect its safety or effectiveness.
相关型号:
©2020 ICPDF网 联系我们和版权申明