MDS-040APS24B [DELTA]

MDS Medical AC-DC Open Frame;
MDS-040APS24B
型号: MDS-040APS24B
厂家: DELTA ELECTRONICS, INC.    DELTA ELECTRONICS, INC.
描述:

MDS Medical AC-DC Open Frame

文件: 总10页 (文件大小:523K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
TECHNICAL DATASHEET  
```  
MDS Medical AC-DC Open Frame  
24Volt, 40Watt / MDS-040APS24 B  
040APS24 B  
Highlights & Features  
– Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1  
– Compliant with IEC 60601-1-2 4th Ed. Requirements  
– Low touch current (<0.1mA Normal & ,0.3mA single fault)  
– Over-Voltage/Load/Temperature & Short Circuit protections  
– 3 Million Hours MTBF  
– 2 x MOPP (means of patient protection)  
– 3 years warranty  
Safety Standards  
CB Certified for worldwide use  
Model Number:  
Unit Weight:  
MDS-040APS24 B  
80grams (2.88 ounces)  
Dimensions (W x L x H): 50.8 x 76.2 x 23 mm  
(2 x 3 x 0.91 inch)  
General Description  
The MDS series of embedded power supply come with universal AC input at 90Vac to 264Vac. Other features include low  
touch current, risk management report available and the electric shock protection comply with 2 x MOPP. The MDS series  
is certified for EMC standards according to EN 55011 for industrial, scientific and medical (ISM) radio-frequency  
equipment and EN 55022 for Industrial Technology Equipment (ITE) radio-frequency equipment. In addition, only  
recognized Japanese capacitors are used.  
The MDS series of embedded power supply come with both medical and ITE safety approvals including UL/cUL/CQC/CE  
and CB certification and are fully compliant with RoHS Directive 2011/65/EU for environmental protection.  
Model Information  
Medical AC-DC Open Frame  
Model Number  
Input Voltage Range  
Output  
Voltage  
Output Current  
MDS-040APS24 B  
90-264Vac  
24Vdc  
1.67A  
Model Numbering  
-
MDS  
Delta Medical  
power Supply  
040  
APS  
24 B  
Output Voltage  
Single Output:  
24 for 24V  
Max wattage in the product series.  
Maybe lower at some voltage.  
040 40W  
Family Code  
All parameters are specified at 25°C ambient unless otherwise noted.  
www.DeltaPSU.com (November 2016, Rev. 02)  
1
TECHNICAL DATASHEET  
```  
MDS Medical AC-DC Open Frame  
24Volt, 40Watt / MDS-040APS24 B  
Specifications  
Input Ratings / Characteristics  
Nominal Input Voltage  
Input Voltage Range  
Nominal Input Frequency  
Input Frequency Range  
Input Current (max.)  
Efficiency (typ.)  
100-240Vac  
90-264Vac  
50-60Hz  
47-63Hz  
0.85A @ 115Vac,0.45A@230Vac  
86.3%, Reference Fig.1  
0.3W  
Standby Power (max.)  
Inrush Current (typ.)  
Touch Current (max.)  
30A @ 115Vac, 60A @ 230Vac  
0.1mA @ 264Vac NC1), 0.3mA @ 264Vac SFC2)  
1) NC: normal condition  
2) SFC: single fault condition  
Fig.1 Efficiency versus output load  
Output Ratings / Characteristics  
Nominal Output Voltage  
Total regulation  
24Vdc  
± 1.5%  
Output Current  
1.67A  
Output Power  
40W  
Line Regulation (max)  
Load Regulation (max)  
Ripple & Noise (typ.)  
±0.5%  
±1%  
34mV pk-pk @ Full load, Reference Fig. 2,  
3000ms @ 115Vac  
Start-up Time (max)  
Hold-up Time (min)  
15ms @ 115Vac  
Dynamic Response (Overshoot & Undershoot O/P Voltage)  
± 3% @ 50-100% load  
All parameters are specified at 25°C ambient unless otherwise noted.  
www.DeltaPSU.com (November 2016, Rev. 02)  
2
TECHNICAL DATASHEET  
```  
MDS Medical AC-DC Open Frame  
24Volt, 40Watt / MDS-040APS24 B  
(a) 115V  
(b) 230V  
Fig. 2 Ripple & Noise example, 20MHz BW  
Ripple & Noise measurement circuit  
PSU  
Load  
0.47uF Ceramic  
10uF Electrolytic  
Scope  
5cm  
Test Fixture  
Mechanical  
Dimensions(W x L x H)  
Weight (typ.)  
50.8 x 76.2 x 23 mm (2 x 3 x 0.91 inch)  
80grams (2.88ounces)  
Input  
JST 2P Connector: B2P3-VH(LF)(SN)  
JST 4P Connector: B4P-VH(LF)(SN)  
Terminal  
Output  
Environment  
Operating -10°C to +70°C  
Storage -40°C to +85°C  
Surrounding Air Temperature  
Power De-rating  
-10°C to +50°C 100% load  
50°C to 70°C de-rate power by 2.5% / °C, See Fig. 3  
10-95% RH (Non-Condensing)  
Operating Humidity  
Operating Altitude  
3,000 meters (9842.52 feet)  
Shock Test (Non-Operating)  
Vibration (Operating)  
50G, 11ms, 3 shocks for each direction  
5-500Hz, 2.09Grms, 20 minute for each three axis  
All parameters are specified at 25°C ambient unless otherwise noted.  
www.DeltaPSU.com (November 2016, Rev. 02)  
3
TECHNICAL DATASHEET  
```  
MDS Medical AC-DC Open Frame  
24Volt, 40Watt / MDS-040APS24 B  
Protections  
Overvoltage (max)  
150%, Latch Mode  
Over load / Over current (max)  
150%~210% of rated load current, Hiccup Mode,  
(Non-Latching, Auto-Recovery)  
Hiccup Mode,  
Over Temperature  
Short Circuit  
(Non-Latching, Auto-Recovery)  
Hiccup Mode,  
(Non-Latching, Auto-Recovery)  
Class I with PE* connection  
Protection Against Shock  
*PE: Protective Earth  
Reliability Data  
MTBF (typ.)  
3 Million Hours based on Telecordia SR-332  
IEC60601-1 3rd and 3.1rd edition CB report  
Safety Standards / Directives  
Medical Safety  
IEC60601-1 edition 3.1rd (2012), EN60601-1 (2006) +  
A11 + A1 + A12, CAN/CSA-C22.2 NO. 60601-1:14,  
ANSI/AAMI ES60601-1:2005/(R)2012  
ITE Safety  
IEC60950-1 (Ed.2,2005), GB4943.1-2011, GB9254-  
2008, GB17625.1-2003  
CE  
MDD Directive 93/42/EEC  
Material and Parts  
RoHS Directive 2011/65/EU Compliant  
Galvanic Isolation  
Input to Output 4000 Vac  
Input to Ground 1500 Vac  
Output to Ground 500 Vac  
All parameters are specified at 25°C ambient unless otherwise noted.  
www.DeltaPSU.com (November 2016, Rev. 02)  
4
TECHNICAL DATASHEET  
```  
MDS Medical AC-DC Open Frame  
24Volt, 40Watt / MDS-040APS24 B  
EMC (Compliant with IEC 60601-1-2 4th Ed. Requirements)  
EMC / Emissions  
EN55011/EN55022,FCC Title 47:Class B  
Harmonic Current Emissions  
Immunity to  
IEC61000-3-2  
Voltage Flicker  
IEC61000-3-3  
Electrostatic Discharge  
IEC61000-4-2 Level 4 Criteria A1)  
Air Discharge: 15kV  
Contact Discharge: 8kV  
Radiated Field  
IEC61000-4-3 Criteria A1)  
80MHz-2700MHz, 10V/m AM modulation  
385MHz-5785MHz, 28V/m Pulse mode and other  
modulation  
Electrical Fast Transient / Burst  
Surge  
IEC61000-4-4 Level 3 Criteria A1) :2kV  
IEC61000-4-5 Level 3 Criteria A1)  
Common Mode3): 2kV  
Differential Mode4): 1kV  
Conducted  
IEC61000-4-6 Level 2 Criteria A1)  
150kHz-80MHz, 3Vrms, 6Vrms at ISM bands and  
Amateur radio bands  
IEC61000-4-8 Criteria A1)  
Power Frequency Magnetic Fields  
Voltage Dips  
Magnetic field strength 30A/m  
Criteria A1)  
IEC61000-4-11  
0% UT, 0.5 cycle (10ms)  
, 0o/45o/90o/135o/180o/225o/270o/315o/360o  
Criteria B2)  
0% UT, 1 cycle (20ms), 0o  
Criteria A1)  
70% UT, 25 cycle (500ms), 0o  
Criteria B2)  
0% UT, 250 cycle (5000ms), 0o  
1) Criteria A: Normal performance within the specification limits  
2) Criteria B: Output out of regulation, or shuts down during test. Automatically restore to normal operation after test.  
3) Asymmetrical: Common mode (Line to earth)  
4) Symmetrical: Differential mode (Line to line)  
All parameters are specified at 25°C ambient unless otherwise noted.  
www.DeltaPSU.com (November 2016, Rev. 02)  
5
TECHNICAL DATASHEET  
```  
MDS Medical AC-DC Open Frame  
24Volt, 40Watt / MDS-040APS24 B  
Block Diagram  
Dimensions  
W x L x H: 50.8 x 76.2 x 23 mm (2 x 3 x 0.91 inch)  
All parameters are specified at 25°C ambient unless otherwise noted.  
www.DeltaPSU.com (November 2016, Rev. 02)  
6
TECHNICAL DATASHEET  
```  
MDS Medical AC-DC Open Frame  
24Volt, 40Watt / MDS-040APS24 B  
Notes  
Dimensions are in mm  
For Open Frame type: There are 4 normal mounting holes. To ensure EMC performance, all mounting holes shall be connected  
to metal case.  
All parameters are specified at 25°C ambient unless otherwise noted.  
www.DeltaPSU.com (November 2016, Rev. 02)  
7
TECHNICAL DATASHEET  
```  
MDS Medical AC-DC Open Frame  
24Volt, 40Watt / MDS-040APS24 B  
Power De-rating  
Output Power  
100%  
50%  
0%  
Ambient  
Temperature  
-10°C  
50°C  
70°C  
Fig. 3  
Functions  
Start-up Time  
The time required for the output voltage (Vo) to reach 90% of its set value, after the input AC voltage is applied.  
Rise Time  
The time required for the output voltage (Vo) to change from 10% to 90% of its steady state value.  
Hold-up Time  
Hold up time is the time when the AC input collapses and output voltage retains regulation for a certain period of time. The time  
required for the output to reach 95% of its set value, after the input voltage is removed.  
Graph illustrating the Start-up Time, Rise Time, and Hold-up Time  
All parameters are specified at 25°C ambient unless otherwise noted.  
www.DeltaPSU.com (November 2016, Rev. 02)  
8
TECHNICAL DATASHEET  
```  
MDS Medical AC-DC Open Frame  
24Volt, 40Watt / MDS-040APS24 B  
Dynamic Response  
Overload & Over current Protections  
The power supply’s Overload (OLP) and Over current (OCP)  
Protections will be activated when output current is between  
150% and 210% of IO (Max load). Upon such an occurrence, VO  
will start to drop. Once the power supply has reached its  
maximum power limit, the protection will be activated. and the  
power supply will go into “Hiccup mode” (Auto-Recovery). The  
power supply will recover once the fault condition causing the  
OLP and OCP is removed and IO is back within the specified limit.  
The power supply output voltage will remain within ±3% of its  
steady state value, when subjected to a dynamic load change  
from 50 to 100% of its rated current.  
100%  
Load  
Output Voltage (Vo)  
Load > 150%  
(Output Hiccup)  
50% Load  
T (sec)  
50ms  
50ms  
50ms  
50ms  
0% Load  
150% Load  
T (sec)  
Inrush Current  
Inrush current is the input current that occurs when the input  
voltage is first applied. For AC input voltages, the maximum peak  
value of inrush current will occur during the first half cycle of the  
applied AC voltage. This peak value decreases exponentially  
during subsequent cycles of AC voltage.  
Additionally, if the IO is <210% but >150% for a prolong period of  
time (depending on the load), the Over Temperature Protection  
(OTP) will be activated due to high temperature on critical  
components. The power supply will then go into hiccup mode until  
the fault is removed; and, the input voltage is removed, then  
reapplied.  
Over Temperature Protection  
As mentioned above, the power supply also has Over  
Temperature Protection (OTP). This is activated when the  
overload condition persists for an extended duration and the  
output current is below the overload trigger point but >100% load.  
In the event of a higher operating condition at 100% load, the  
power supply will run into OTP when the surrounding air  
temperature is higher than the operating temperature. When  
activated, the output voltage will go into hiccup mode until the  
input voltage is removed; then, reapplied, and the surrounding air  
temperature drops to its normal operating temperature.  
Overvoltage Protection  
The power supply’s overvoltage circuit will be activated when its  
internal feedback circuit fails. The output voltage shall not exceed  
its specifications defined on Page 3 under “Protections”. Power  
supply will latch off, and require removal/re-application of input  
AC voltage in order to restart.  
Short Circuit Protection  
The power supply’s output OLP/OCP function also provides  
protection against short circuits. When a short circuit is applied,  
the output current will operate in “Hiccup mode”, as shown in the  
illustration in the OLP/OCP section on this page. The power  
supply will return to normal operation after the short circuit is  
removed.  
All parameters are specified at 25°C ambient unless otherwise noted.  
www.DeltaPSU.com (November 2016, Rev. 02)  
9
TECHNICAL DATASHEET  
```  
MDS Medical AC-DC Open Frame  
24Volt, 40Watt / MDS-040APS24 B  
Certificate  
.
All Delta Medical Power products conform to the European directive 2011/65/EU.  
RoHS is the abbreviation for “Restriction of the use of certain hazardous  
substances.  
Delta has been certified as meeting the requirement of ISO 13485: 2003 and EN  
ISO 13485:2012 for the design and manufacture of switching power supply and  
adaptor for medical device.  
In addition to a UL Total Certification Program (TCP) approved client laboratory for  
IEC60950 and IEC60065. Delta also has participated UL Client Test Data Program  
(CDTP) for IEC 60601  
All parameters are specified at 25°C ambient unless otherwise noted.  
www.DeltaPSU.com (November 2016, Rev. 02)  
10  

相关型号:

MDS-060AAS15-B

MDS Medical AC-DC Adapter
DELTA

MDS-060AAS15B

Medical AC-DC Adapter
DELTA

MDS-060AAS19-B

MDS Medical AC-DC Adapter
DELTA

MDS-060AAS19B

Medical AC-DC Adapter
DELTA

MDS-060AAS24-B

Meet DOE Level VI and ErP Lot 7
DELTA

MDS-060AAS24B

Medical AC-DC Adapter
DELTA

MDS-065APS15-B

MDS Medical AC-DC Open Frame
DELTA

MDS-065APS15B

MDS Medical AC-DC Open Frame
DELTA

MDS-065APS18-B

MDS Medical AC-DC Open Frame
DELTA

MDS-065APS18B

Medical AC-DC Open Frame
DELTA

MDS-065APS24-B

MDS Medical AC-DC Open Frame
DELTA

MDS-065APS24B

MDS Medical AC-DC Open Frame
DELTA