Embedded Power for  
Business-Critical Continuity  
Rev.02.28.07  
NLP65_3300  
1 of 4  
NLP65 Medical  
Series  
Single, dual  
and triple output  
Total Power:  
65W  
Input Voltage: 85 - 264 VAC  
# of Outputs: Single, dual,  
triple  
The NLP65 Medical series is a 65 W universal input ac-dc power  
supply with input harmonic current correction as standard. This  
compact design, packed in a 5 x x3 inch card with a maximum  
component height of 1.26 inches, is ideal for use in a variety of  
medical, laboratory and dental applications such as centrifuges,  
incubators and infusion pumps. The NLP65 Medical series  
comprises of eight models in single, dual and triple output  
configurations. Providing 65 Watts of continuous output power  
with free air convection, the NLP65 Medical series will deliver 75  
Watts output power with 20 CFM of air. The series, with full  
medical safety approval to EN60601 and UL2601, is CE marked,  
greatly accelerating design-in time and reducing system  
compliance costs.  
Special Features  
• 85 VAC to 264 VAC  
universal input range  
• Harmonic current correction  
as standard  
• Maximum component height  
1.26 inches  
• UL, CSA and VDE  
safety approvals  
• Overvoltage and  
short circuit protection  
• 5 x 3 x 1.26 inch (127.0 x 76.2 x  
32mm) footprint  
• Available RoHS compliant  
• 2 year warranty  
Safety  
VDE0705/EN60601-1/IEC1010  
File No. 10401-3336-0156/32480  
Licence No. 121949  
UL1950 File No. E147937  
CSA C22.2 No. 950  
File No. LR41062C  
Embedded Power for  
Business-Critical Continuity  
Rev.02.28.07  
NLP65_3300  
2 of 4  
Specifications  
All specifications are typical at nominal input, full load at 25°C unless otherwise stated  
OUTPUT SPECIFICATIONS  
Output power  
EMC CHARACTERISTICS  
Natural convection  
At turn-on  
65 W max.  
See table  
Conducted emissions  
Radiated emissions  
ESD air  
EN55022, FCC part 15  
Level A  
Level A  
EN55022, FCC part 15  
EN61000-4-2, level 3  
EN61000-4-2, level 4  
EN61000-4-5, level 3  
EN61000-4-4, level 3  
EN61000-4-3, level 3  
EN61000-4-6, level 3  
Total regulation  
Rise time  
Perf. criteria 1  
Perf. criteria 1  
Perf. criteria 1  
Perf. criteria 1  
Perf. criteria 2  
Perf. criteria 2  
ESD contact  
Surge  
1.0 s, max.  
Fast transients  
Radiated immunity  
Conducted immunity  
Transient response  
Main output  
25% step  
at 0.1 A/μs  
5.0% max. dev.,  
1ms recovery  
to 1.0%  
GENERAL SPECIFICATIONS  
Hold-up time  
Temperature coefficient  
Overvoltage protection  
Short circuit protection  
INPUT SPECIFICATIONS  
0.02%/°C  
125%, 10%  
Yes  
120 Vac, 60 Hz  
120 Vac, 65 W  
16 ms @ 65 W  
72% typ.  
Main outputs  
Efficiency  
Cyclic operation  
Isolation voltage  
Input/output  
Input/chassis  
4000 Vac  
1500 Vac  
Input voltage range  
(See Note 2)  
Universal input  
85-264 Vac  
47-63 Hz  
Switching frequency  
Fixed  
100 kHz, 5 kHz  
Approvals and  
standards  
EN60601, UL2601  
CSA 22.2 No. 125  
Input frequency range  
Input surge current  
(cold start)  
120 Vac  
230 Vac  
17 A max.  
32 A max.  
Weight  
MTBF  
283 g (10 oz)  
MIL-HDBK-217F  
150,000 hours  
Safety ground  
leakage current  
264 Vac, 60 Hz  
95 μA  
ENVIRONMENTAL SPECIFICATIONS  
Thermal performance  
Operating ambient,  
0 °C to +70 °C  
Input current  
Input fuse  
120 Vac  
230 Vac  
1.05 A rms  
0.51 A rms  
(See derating curve)  
Non-operating  
-40 °C to +85 °C  
65 W  
250 Vac F 3.15 A  
0 ¡C to 50 ¡C ambient,  
convection cooled  
50 ¡C to 70 ¡C ambient,  
convection cooled  
Derate to  
50% load  
Relative humidity  
Altitude  
Non-condensing  
5% to 95% RH  
Operating  
Non-operating  
10,000 feet max.  
30,000 feet max.  
Vibration (See Note 5)  
Shock  
5-500 Hz  
2.43 G rms approx.  
516.4 Part IV  
per MIL-STD-810E  
Embedded Power for  
Business-Critical Continuity  
Rev.02.28.07  
NLP65_3300  
3 of 4  
Specifications Contd.  
OUTPUT CURRENT  
OUTPUT  
VOLTAGE  
TOTAL  
REGULATION (6)  
MODEL  
NUMBER (11,12)  
RIPPLE (4)  
MAX (1)  
7 A  
PEAK  
9.1 A  
3.3 A  
0.81 A  
9.1 A  
2.9 A  
0.85 A  
9.1 A  
2.6 A  
9.1 A  
3.3 A  
13 A  
FAN (10)  
8 A  
+5 V  
+12 V  
-12 V  
+5 V  
50 mV  
150 mV  
120 mV  
50 mV  
2.0%  
NLP65-9908J  
2.5 A  
0.5 A  
7 A  
3 A  
5.0%  
1 A  
5.0%  
8 A  
2.0%  
NLP65-9910J  
+15 V  
-15 V  
+5 V  
2.2 A  
0.65 A  
7 A  
2.5 A  
0.8 A  
8 A  
150 mV  
150 mV  
50 mV  
5.0%  
5.0%  
2.0%  
NLP65-9920J  
NLP65-9929J  
+24 V  
+5 V  
2 A  
2 A  
240 mV  
50 mV  
5.0%  
7 A  
8 A  
2.0%  
+12 V  
+5 V  
2.5 A  
10 A  
5.4 A  
4.4 A  
2.7 A  
3 A  
150 mV  
50 mV  
5.0%  
12 A  
6.5 A  
5.3 A  
3.5 A  
2.0%  
NLP65-9905J  
NLP65-9912J  
NLP65-9915J  
NLP65-9924J  
+12 V  
+15 V  
+24 V  
7 A  
120 mV  
150 mV  
240 mV  
2.0%  
5.7 A  
3.5 A  
2.0%  
2.0%  
Notes  
7
8
9
For optimum reliability, no part of the heatsink should exceed 120 °C, and  
no semiconductor case temperature should exceed 130 °C.  
CAUTION: Allow a minimum of 1 second after disconnecting line power when  
making thermal measurements.  
This product is only for inclusion by professional installers within other  
equipment and must not be operated as a stand alone product.  
1
Natural convection cooling. Models NLP65-9929J, NLP65-9908J,  
NLP65-9910J must not exceed 62.5 Watts continuous output power with  
natural convection. Model NLP65-9920J not to exceed 65 Watts continuous  
output power with natural convection.  
0W°hCentoth+e4i0np°Cu.tTvhoeltarigpepilselaenssdtrheagnu9la0tioVnacsptheeciofipcaetriaotninsgmteamy npoetrabteumreerta.nge is  
Peak output current lasting less than 60 seconds with duty cycle less than 5%.  
During peak loading, output voltage may exceed total regulation limits.  
Figure is peak-to-peak for convection power rating. Output noise  
measurements are made across a 20MHz bandwidth using a 6 inch twisted  
pair, terminated with a 10 μF electrolytic capacitor and a 0.1 μF ceramic  
capacitor.  
2
3
4
10 Maximum continuous output power for all multiple output models must not  
exceed 75 Watts with 20 CFM forced air cooling at 50 °C.  
11 The ‘J’ suffix indicates that these parts are Pb-free (RoHS 6/6) compliant.  
TSE RoHS 5/6 (non Pb-free) compliant versions may be available on special  
request, please contact your local sales representative for details.  
12 NOTICE: Some models do not support all options. Please contact your local  
Artesyn representative or use the on-line model number search tool at  
http://www.artesyn.com/powergroup/products.htm to find a suitable  
alternative.  
5
6
Three orthogonal axes, random vibration 10 minutes for each axes, 2.4 G rms 5  
Hz to 500 Hz.  
To maintain stated regulation then:  
for single output units  
I ³ 0.2 A I max.  
for multiple output units  
0.25 ² I(A)/I(B) ² 5, for I(A) ³ 0.2 A I(A) max.  
Embedded Power for  
Business-Critical Continuity  
Rev.02.28.07  
NLP65_3300  
4 of 4  
Americas  
5810 Van Allen Way  
Carlsbad, CA 92008  
USA  
Telephone: +1 760 930 4600  
Facsimile: +1 760 930 0698  
5.00  
(127.00)  
4.550  
(115.57)  
0.225  
(5.715)  
Europe (UK)  
Waterfront Business Park  
Merry Hill, Dudley  
West Midlands, DY5 1LX  
United Kingdom  
J1  
J3  
N
L
3
1
6
1
3.00  
(76.20)  
2.550  
(64.770)  
Telephone: +44 (0) 1384 842 211  
Facsimile: +44 (0) 1384 843 355  
T1  
J2  
Asia (HK)  
C19  
16th - 17th Floors, Lu Plaza  
2 Wing Yip Street, Kwun Tong  
Kowloon, Hong Kong  
Telephone: +852 2176 3333  
Facsimile: +852 2176 3888  
Mounting  
Hole Diameter  
0.156 (4.00)  
ALL DIMENSIONS IN INCHES (mm)  
MAXIMUM COMPONENT  
HEIGHT 1.26 (32.00)  
For global contact, visit:  
TOLERANCE: .xxx  
0.005"  
.xx  
0.02"  
www.powerconversion.com  
technicalsupport@powerconversion.com  
While every precaution has been taken to ensure  
accuracy and completeness in this literature, Emerson  
Network Power assumes no responsibility, and disclaims  
all liability for damages resulting from use of this  
information or for any errors or omissions.  
Emerson Network Power.  
The global leader in enabling  
business-critical continuity.  
AC Power  
Connectivity  
DC Power  
Embedded Power  
Inbound Power  
Integrated Cabinet Solutions  
Outside Plant  
Precision Cooling  
Site Monitoring and Services  
EmersonNetworkPower. com  
Emerson Network Power and the Emerson  
Network Power logo are trademarks and  
service marks of Emerson Electric Co.  
©2007 Emerson Electric Co.  

相关型号:

NLP65-9924J

85 VAC to 264 VAC universal input range
EMERSON-NETWO

NLP65-9929J

65 W universal input ac-dc power supply
EMERSON-NETWO

NLP65-X605G

LOW TO MEDIUM POWER AC/DC POWER SUPPLIES
ARTESYN

NLP65-X605GJ

Universal Input
EMERSON-NETWO

NLP65-X608G

LOW TO MEDIUM POWER AC/DC POWER SUPPLIES
ARTESYN

NLP65-X608GJ

Universal Input
EMERSON-NETWO

NLP65-X610G

LOW TO MEDIUM POWER AC/DC POWER SUPPLIES
ARTESYN

NLP65-X610GJ

Universal Input
EMERSON-NETWO

NLP65-X612G

LOW TO MEDIUM POWER AC/DC POWER SUPPLIES
ARTESYN

NLP65-X612GJ

Universal Input
EMERSON-NETWO

NLP65-X615G

LOW TO MEDIUM POWER AC/DC POWER SUPPLIES
ARTESYN

NLP65-X615GJ

Universal Input
EMERSON-NETWO