Installation  
Instructions  
LC700  
LC750  
Patient Care Unit with  
Standard Laminate or  
Optional Terreon®  
Lavatory Top  
Table of Contents  
Pre-Installation Information . . . . . . . . . . . . . . . .2  
Installation Instructions  
WARNING  
Unpacking the Patient Care Unit . . . . . . . . . .3  
Installing the Anchors . . . . . . . . . . . . . . . . . . .4  
Water Closet and Waste Connections . . . . . . .5  
Installing the Inline Backflow Preventer . . . .6  
Installing the Cabinet . . . . . . . . . . . . . . . . . . .6  
Installing the Terreon® Lav Top . . . . . . . . . . .6  
Installing the Faucet and Water Supply . . . . .7  
Completing Connections . . . . . . . . . . . . . . . . .8  
Installing the Cabinet Floor and Back Panel .8  
Installing the Door Linkage . . . . . . . . . . . . . .9  
Cleaning and Maintenance . . . . . . . . . . . . . . . .10  
Before beginning installation, all supply,  
drain and waste piping to the unit must  
be completed according to the rough-ins  
received prior to receiving the unit. If  
you have not received rough-ins, contact  
your Bradley representative.  
All mounting hardware provided is  
essential for proper anchoring. Bradley  
Corporation cannot assume any  
responsibility for personal injury or  
damage to equipment due to an  
improperly installed unit.  
Flush all water supply lines before  
beginning installation.  
This installation manual covers all  
instructions for the LC700 or LC750  
series including those with Terreon® Lav  
tops. Omit any steps that do not apply to  
the model you are installing.  
Figures shown in this manual illustrate a  
right-hand model. A left-hand model  
would be opposite to what is shown.  
Product warranties may be found under  
“Product Information” on our web site at  
www.bradleycorp.com.  
P.O. Box 309, Menomonee Falls, WI 53052-0309  
Phone: 1-800-BRADLEY FAX: (262) 251-5817  
http://www.bradleycorp.com  
215-1171 Rev. D; EN 04-1037  
© 2003 Bradley Corporation  
Page 1 of 10 8/16/04  
Patient Care Unit with Standard Laminate or Optional Terreon® Top  
LC700, LC750  
Installation Instructions  
Pre-Installation Information  
Overview  
The Bradley LC700 and LC750 Patient Care Units are all-in-one units containing a water closet and  
®
lavatory. The units come with either a laminate or Terreon lavatory top and may include an  
optional flexible bedpan washer, inline backflow preventer and/or dialysis box.  
Check contents  
Separate all parts from packaging and make sure all parts are accounted for before discarding any  
packaging material. If any parts are missing, do not attempt to install your Bradley Patient Care Unit  
until you obtain the missing parts.  
NOTE: The LC700 and LC750 Patient Care Units are shipped assembled (except for faucet-mount  
option) and secured to a false plywood floor like it will be installed at your location. DO NOT  
DISCARD PLYWOOD FLOOR! The plywood floor will serve as your anchor hole location  
template.  
NOTE: Before beginning installation, all supply, drain and waste piping to your Bradley Patient  
Care Unit must be completed according to rough-ins received prior to receiving the unit. If you have  
not received rough-ins, contact your Bradley representative.  
Supplies required for installation:  
• Steel angles and hardware for anchoring the water closet (if concrete is lightweight or less  
than 4" thick)  
• Hardware for waste outlet connections  
• Shims (for water closet or cabinet installation, if necessary)  
• Supply piping (including 1/2" NPT flexible hoses) to water closet, faucet or foot valve,  
and bedpan washer  
• Waste outlet piping to water closet  
• P-trap and drain piping  
• Silicone caulk  
2
8/16/04  
Bradley Corporation • 215-1171 Rev. D; EN 04-1037  
Patient Care Unit with Standard Laminate or Optional Terreon® Top  
LC700, LC750  
Installation Instructions  
Installation Instructions  
LINKAGE  
BRACKET  
Step 1: Unpack the Patient Care Unit  
1. Remove the packaging around the Patient Care Unit.  
2. Remove the door linkage by removing the nut fastened  
on the linkage bracket (see Figure 1). Note the location  
of the rod on the linkage bracket.  
3. Lift out the trim piece around the water closet on the  
cabinet floor (see Figure 2).  
4. Unfasten the two panel clips and slide the sliding back  
panel so that the panel lips unlock (see Figure 2). Bring  
the panel out of the front of the cabinet.  
DOOR  
LINKAGE  
ARM  
Figure 1  
5. Remove the cabinet floor by removing the two screws  
holding the floor in place and sliding the  
cabinet floor out.  
SLIDING BACK  
6. Remove the flexible hose, hose clamp,  
inlet nut and pipe from the water closet  
(see Figure 3).  
PANEL  
PANEL  
CLIP  
STATIONARY  
BACK PLATE  
7. Unbolt the water closet and cabinet  
assembly from the floor/template. Discard  
the shipping hardware for the water closet  
and cabinet, but DO NOT discard the  
plywood floor/template.  
HOSE CLAMP  
PIPE  
TRIM  
PIECE  
FLEXIBLE  
HOSE  
CABINET  
FLOOR  
CABINET FLOOR  
SCREWS  
INLET NUT  
Figure 2  
Figure 3  
Bradley Corporation • 215-1171 Rev. D; EN 04-1037  
8/16/04  
3
Patient Care Unit with Standard Laminate or Optional Terreon® Top  
LC700, LC750  
Installation Instructions  
Installation Instructions continued . . .  
Step 2: Install the anchors  
NOTE FOR OPTIONAL FOOT  
VALVE: It is recommended that  
floor anchors be used to secure the  
foot valve to the floor (anchors and  
appropriate hardware supplied by  
installer).  
1. Select the location where the  
patient care unit is to be  
installed. Using the supplied  
template, lay out the hole  
pattern for the water closet  
and cabinet anchor locations  
(see Figure 4).  
NOTE FOR OPTIONAL FOOT  
VALVE: Use the foot valve as a  
template and lay out the anchor  
locations on the floor. Drill  
Figure 4  
appropriate holes for two anchors.  
2. Using a 5/8" concrete drill bit, drill four 2-1/8"  
deep holes into the water closet support  
anchoring locations using the template (see  
Figure 4). Holes must be clean and free of  
debris.  
3. Using a 3/8" concrete drill bit, drill four 1-1/8"  
deep holes into the cabinet anchoring locations  
using the template (see Figure 4). Holes must be  
clean and free of debris.  
NOTE: If the floor concrete is lightweight or less  
than 4" thick, Bradley recommends through-bolting  
to steel angles under the floor (hardware supplied by  
installer) (see Figure 7 on page 5).  
Figure 5  
NOTE FOR OPTIONAL FOOT VALVE: Install  
appropriate anchors into the floor for foot valve  
anchoring.  
4. Remove the template. Drive the four 2-1/8"  
anchors provided into the drilled holes for the  
water closet using a hammer and the setting tool  
supplied. The anchor should be driven in flush  
with the floor (see Figure 5).  
5. Drive the four 1/8" anchors provided into the  
drilled holes for the cabinet using a hammer and  
the setting tool. Make sure the anchors are  
driven in flush with the floor (see Figure 6).  
Figure 6  
4
8/16/04  
Bradley Corporation • 215-1171 Rev. D; EN 04-1037  
Patient Care Unit with Standard Laminate or Optional Terreon® Top  
LC700, LC750  
Installation Instructions  
Installation Instructions continued . . .  
Step 3: Install the water closet and waste connections  
1. Place the water closet over the water closet support  
anchors installed in Step 2 on page 4.  
2. Using the gasket supplied, connect the water closet to  
the floor waste outlet (gaskets supplied by installer)  
(see Figure 8).  
NOTE: If the waste outlet is through the back or side  
wall, use a no-hub type connector (available as an option)  
but DO NOT tighten the connector (see Figure 9).  
3. Square the water closet to the back wall. Level and  
support the water closet solidly at the four corners  
using shims (supplied by installer) if necessary.  
4. Secure the water closet to the four floor anchors  
using the 1/2" hex screws and washers supplied and  
tighten securely (see Figure 8).  
Figure 7  
5. On patient care units with a back waste connection,  
tighten the connector clamps at this time.  
Figure 8  
Figure 9  
Bradley Corporation • 215-1171 Rev. D; EN 04-1037  
8/16/04  
5
Patient Care Unit with Standard Laminate or Optional Terreon® Top  
LC700, LC750  
Installation Instructions  
Installation Instructions continued . . .  
Step 4: Install the inline backflow preventer (for volume control or foot control)  
1. Install the supplied inline backflow preventer to the wall  
per local code.  
2. Connect the water supply to the inlet of the foot valve (or  
volume control valve, based on option supplied).  
NOTE The foot valve inlet is located on the backside of the  
valve.  
3. Connect the outlet of the foot valve (or volume control  
valve, based on option supplied) to the inlet of the  
backflow preventer.  
NOTE The foot valve outlet is located on the topside of the  
valve.  
4. Connect the outlet of the backflow preventer to the inlet of  
Figure 10  
the flexible hose.  
Step 5: Install the cabinet  
1. Install the supplied mounting studs into the  
cabinet anchors installed in Step 2 on page 4.  
2. Over each mounting stud, screw on a leveling  
nut and place a washer over the nut (see Figure  
10).  
3. With the cabinet floor removed, locate the  
cabinet in place over the mounting studs (see  
Figure 10).  
4. Add a flat washer and securing nut to each  
cabinet mounting stud. Level the cabinet and  
tighten the nuts.  
NOTE FOR OPTIONAL FOOT VALVE: Install the  
foot valve with the screws provided to the foot valve  
anchors (installed in Step 2 on page 4) and tighten.  
Figure 11  
5. Connect the flexible hose, hose clamp, pipe and inlet nut to the water closet as shown in Figure 3  
on page 3. Hand-tighten the inlet nut.  
6. Connect the flexible hose to the flush valve (see Figure 11).  
6
8/16/04  
Bradley Corporation • 215-1171 Rev. D; EN 04-1037  
Patient Care Unit with Standard Laminate or Optional Terreon® Top  
LC700, LC750  
Installation Instructions  
Installation Instructions continued . . .  
BEAD OF  
CAULK  
Step 5a: Install the optional  
Terreon lavatory top  
BEAD OF  
CAULK  
®
1. Apply a bead of silicone caulk around  
the outer top edge of the cabinet and  
between the lavatory top and the wall.  
2. Set the lavatory top in place on the  
cabinet making sure the back and wall  
sides are flush with the edge of the  
cabinet (see Figure 12).  
3. Clean up any excess caulk.  
Figure 12  
Step 6: Install faucet and connect  
water supply  
FOOT VALVE MODELS (see Figures 13 and 14)  
GOOSENECK  
SPOUT  
1. Place the faucet onto the countertop, gently inserting the faucet  
supply shank through the hole provided in the lavatory bowl.  
2. Using the washer and wingnut provided, mount the faucet to the  
countertop by tightening the wingnut on the faucet's supply  
shank.  
LAVATORY  
BOWL  
3. Connect the flexible supply hose provided from the foot pedal  
valve to the supply shank.  
SUPPLY  
SHANK  
4. Connect the hot and cold water supply (piping supplied by  
installer) to the 1/2” NPT inlet connectors on the foot valve (see  
Figure 14).  
SUPPLY  
HOSE  
Figure 13  
FAUCET VALVE MODELS (see Figure 15)  
1. Place the faucet onto the countertop, gently inserting the faucet supply shanks through the  
holes provided in the lavatory bowl.  
HOT  
INLET  
SUPPLY  
2. Using the washers and wingnuts provided, mount  
HOSE  
the faucet to the countertop by tightening the  
FOOT VALVE  
COLD  
INLET  
wingnut on the faucet's supply shanks.  
OUTLET  
CONNECTION  
3. Connect the hot and cold water supply (piping  
supplied by installer) to the 1/2” NPT inlet  
connectors on the foot valve.  
Figure 15  
FOOT VALVE  
INLET  
CONNECTORS  
LAVATORY  
BOWL  
FOOT  
VALVE  
Figure 14  
WASHER  
WINGNUT  
1/2” NPT  
HOT INLET  
1/2” NPT  
COLD INLET  
COUNTERTOP  
Bradley Corporation • 215-1171 Rev. D; EN 04-1037  
8/16/04  
7
Patient Care Unit with Standard Laminate or Optional Terreon® Top  
LC700, LC750  
Installation Instructions  
Installation Instructions continued . . .  
Step 7: Completing Installation  
1. Connect the flush valve water supply pipe to the flush valve 1" NPT inlet.  
2. Install the strainer and tailpiece.  
3. Install the P-trap (supplied by installer) to the sink drain, then connect the P-trap to the drain  
pipe (piping supplied by installer).  
Step 8: Reinstall the cabinet floor and back panel  
1. Reinstall the cabinet floor and secure it to the cabinet frame with the two screws provided (see  
Figure 16).  
NOTE FOR OPTIONAL FOOT VALVE: The outlet hose must be positioned in the cabinet floor cutout  
as shown in Figure 16.  
2. Lay the trim piece around the base of the water closet on the cabinet floor (see Figure 16).  
3. Using silicone caulk (supplied by installer), caulk between the trim piece and cabinet floor, and  
caulk the gap between the pedestal and trim piece (see Figure 16).  
4. Reinstall the back panel by bringing the sliding back panel in from the front of the module and  
sliding the back panel into place.  
5. Secure the back panel by sliding the back panel onto the frame corner (see Figure 17).  
SLIDING BACK  
PEDESTAL  
TO SPOUT  
PANEL  
PANEL  
CLIP  
STATIONARY  
BACK PLATE  
TRIM  
PIECE  
CAULK  
TRIM  
PIECE  
CABINET FLOOR  
CABINET FLOOR  
SCREWS  
CABINET  
FLOOR  
FOOT PEDAL  
OUTLET HOSE  
Figure 16  
Figure 17  
8
8/16/04  
Bradley Corporation • 215-1171 Rev. D; EN 04-1037  
Patient Care Unit with Standard Laminate or Optional Terreon® Top  
LC700, LC750  
Installation Instructions  
Installation Instructions continued . . .  
Step 9: Reinstall and adjust the door linkage  
1. Reinstall the door linkage by loosely fastening the door linkage rod with the supplied nut onto  
the linkage bracket (see Figure 18).  
2. Close the door. If the water closet contacts the back wall before the door closes, adjust the door  
linkage by turning the door linkage rod so that the door closes before the water closet comes in  
contact with the back wall (see Figure 18). If no adjustment is needed, tighten the nut securely  
on the linkage bracket.  
NOTE: The linkage arm has a factory-set length. When the door linkage is fastened to the linkage  
bracket, the door will open 110 degrees. If the PATIENT CARE UNIT is installed against a corner  
wall, make sure the door does not come in contact with the wall, or the water closet does not come in  
contact with the back wall when the door is opened or closed. Make adjustments, if necessary, and  
refer to Figure 18 with inset.  
3. After the adjustments are made, tighten the nut securely on the linkage arm.  
DOOR  
LINKAGE  
BRACKET  
2nd hole for 110 °  
opening  
3rd hole for 90 °  
opening  
DOOR  
LINKAGE  
ARM  
Figure 18  
Bradley Corporation • 215-1171 Rev. D; EN 04-1037  
8/16/04  
9
Patient Care Unit with Standard Laminate or Optional Terreon® Top  
LC700, LC750  
Installation Instructions  
Patient Care Unit Cleaning and Maintenance  
®
Regular cleaning greatly prolongs the service life of Terreon , stainless steel and laminate fixtures and, at  
the same time, helps to maintain a pleasing surface appearance. The amount and frequency of cleaning  
depends on the service conditions involved. For best results, fixtures should be cleaned as often as films or  
deposits become apparent. This will remove built-up deposits which may eventually cause concentration  
cells to set on the surface.  
CAUTION: USING CLEANERS THAT CONTAIN HARSH CHEMICALS, ABRASIVES, ACIDS  
OR DISINFECTANTS, OR CLEANING WITH STEEL WOOL OR BRUSHES, WILL RESULT IN  
DAMAGE TO ALL SURFACES. IF THESE CHEMICALS COME IN CONTACT WITH ANY  
SURFACE, WIPE OFF THE SURFACE IMMEDIATELY AND FLUSH WITH SOAPY WATER.  
ALWAYS READ THE LABEL ON ANY CLEANING PRODUCT BEFORE APPLYING TO ANY  
FIXTURE.  
Terreon® cleaning  
®
Daily cleaning: Wipe the Terreon with a wet cloth, then rinse with clear water and wipe dry.  
®
Weekly cleaning: Give the Terreon surface a thorough cleaning with a damp cloth and a  
household liquid detergent or ammonia-based cleaner. Most stains can be removed by cleaning the  
®
Terreon surface with a mild abrasive cleaner, then rinsing and allowing to dry. Stubborn stains  
can be cleaned as follows:  
1. using a #7448 Scotch-Brite™ pad, scrub with an abrasive cleanser and water  
2. clean thoroughly with clean soap and water and allow to dry  
Scorch marks: Although Terreon will not burn, a lit cigarette in contact with the Terreon  
®
®
material could leave a scorch mark. Scorch marks can be removed with either an abrasive cleaner  
or by buffing with a #7448 Scotch-Brite™ pad.  
®
Repair kit: A repair kit is available from Bradley to repair/patch Terreon . Contact your Modular  
Services representative to order a repair kit and be sure to specify color when ordering.  
Stainless steel and laminate cleaning  
• Give the laminate surface a thorough cleaning with a non-abrasive liquid tub-and-tile cleaner to  
remove soil and maintain the glossy finish.  
You may safely use the following cleaners:  
®
®
®
®
Formula 409  
Spic and Span Powder  
Liquid Comet  
Glass Plus  
Mr. Clean  
®
®
®
Soft Scrub  
Windex  
DO NOT use the following:  
®
®
®
®
Whitecap  
Lysol Disinfectant Spray Lestoil  
Pinesol  
®
Dow Disinfectant Bathroom Cleaner  
acetone, alcohol or steel wool/brushes  
• Remove all material and deposits that tend to adhere to the laminate surface, taking special note of  
crevices and corners  
• Remove all material and deposits that tend to adhere to the stainless steel surface with stainless  
steel polishing powder (always rub in the direction of the polishing lines)  
• Check for rust spots or streaks on the stainless steel surface and look for the source of the rust (a  
steel nail located in some iron or steel not actually a part of the stainless steel fixture may be the  
cause). Rinse the laminate and stainless steel surfaces thoroughly and dry completely after  
washing.  
10  
8/16/04  
Bradley Corporation • 215-1171 Rev. D; EN 04-1037  

相关型号:

LC700Z

GREEN OVAL LAMP LED
SEOUL

LC701

INFRARED LAMP LED
SEOUL

LC7010

Consumer IC
ETC

LC7020AK

Controller for PLL Frequency Synthesizer
SANYO

LC7020AT

Controller for PLL Frequency Synthesizer
SANYO

LC7020BK

Controller for PLL Frequency Synthesizer
SANYO

LC7020BT

Controller for PLL Frequency Synthesizer
SANYO

LC7030

Consumer IC
ETC

LC70301LG/KLG/W

The highest level*1 of denoising performance in the industry
ETC

LC703200AW

Speech Processing IC
ONSEMI

LC703200AW-8C99-H

Speech Processing IC
ONSEMI

LC703831-300

Analog/Digital Clock LCD Display
NXP