Orthopantomograph® OP200 & OP200 D  
Orthoceph® OC200 & OC200 D  
Service Program Manual  
5140241-100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
Copyright  
Code: 5140241-100 rev 3  
Date: 15 May 2008  
Document code: D500293 rev 3  
Copyright © 05/2008 by PaloDEx Group Oy.  
All rights reserved.  
®
®
Orthopantomograph and Orthoceph are registered trademarks  
of Instrumentarium Dental. U.S. patents 4,641,336; 5,016,264;  
5,425,065, 5,444,754, 6,731,717 and 6,829,326. German patent  
®
4,344,745. Finnish patents 112594 and 114383. Windows is  
trademark of Microsoft Corporation in the United States of  
®
America and other countries. Pentium is a registered trademark  
®
®
of Intel Corporation. Iomega Jaz is a registered trademark of  
Iomega Corp.  
Documentation, trademark and the software are copyrighted  
with all rights reserved. Under the copyright laws the  
documentation may not be copied, photocopied, reproduced,  
translated, or reduced to any electronic medium or machine  
readable form in whole or part, without the prior written  
permission of Instrumentarium Dental.  
The original language of this manual is English.  
Instrumentarium Dental reserves the right to make changes in  
specification and features shown herein, or discontinue the  
product described at any time without notice or obligation.  
Contact your Instrumentarium Dental representative for the most  
current information.  
Manufactured by  
Instrumentarium Dental  
Nahkelantie 160 (P.O. Box 20)  
FI-04300 Tuusula  
FINLAND  
Tel. +358 45 7882 2000  
Fax. +358 9 851 4048  
For service, contact your local distributor.  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
Table of Contents  
1
Introduction..................................................................................................................1  
1.1 General .................................................................................................................1  
1.2 Exhibition mode.....................................................................................................1  
1.3 Service program quick reference ..........................................................................2  
1.4 How to use the service programs..........................................................................4  
2
Service program features ...........................................................................................5  
2.1 Sr 70 Log: Display error log ..................................................................................5  
2.2 Sr 71 PAy: Set lease period..................................................................................7  
2.3 Sr 72 LCA: Adjust lateral TMJ image area  
(film units only)......................................................................................................8  
2.4 Sr 74 IOC: Display CPU input/output signal check .............................................10  
2.5 Sr 76 PUP: X-ray tube warm-up..........................................................................16  
2.6 Sr 77 Prh: Adjust the preheat setting ..................................................................18  
2.7 Sr 78 thA: Set Dose Calibration Constant (DCC)................................................19  
2.8 Sr 79 SUP: Display line voltage ..........................................................................20  
2.9 Sr 80 CrL: Test motor movement........................................................................ 20  
2.10 Sr 81 bPL: Display biteplate properties  
(Volumetric Tomography option only) .................................................................22  
2.11 Sr 82 COL: Set/display collimator information ....................................................23  
2.12 Sr 86 Sto: Store settings .....................................................................................25  
2.13 Sr 87 LAL: Set linear adjustment value for VT....................................................25  
2.14 Sr 88 CAL: Disable automatic cephalostat alignment correction  
(digital units only) ................................................................................................26  
2.15 Sr 88 CAL: Set cephalostat alignment value (film units only) .............................27  
2.16 Sr 89 COP: Set country options..........................................................................27  
2.17 Sr 90 PIN: Panorama installation program.......................................................... 30  
2.18 Sr 91 CIN: Cephalostat installation program (digital units only)..........................31  
2.19 Sr 92 CHE: Set level of self-checks ....................................................................32  
2.20 Sr 93 BAL: Adjust beam alignment board sensitivity ..........................................33  
5140241-100 rev 3  
Instrumentarium Dental  
i
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
ii  
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
1 Introduction  
1 Introduction  
1.1 General  
Instrumentarium Dental Ortopantomograph® OP200 is a  
panoramic X-ray unit to produce images of the dentition,  
TM-joints and skull (hereafter called “OP200”).  
OP200 has two kinds of programs, user programs (Pr) and  
service programs (Sr).  
User programs (Pr) are accessible by the user and allows  
unit configuration for daily use and changing technique  
factors to optimize image quality. See OP200 User  
Manual.  
Maintenance & Service programs (Sr) are intended for  
technical personnel for use during installation and service.  
The Pr and Sr letters in the kV display indicate the unit in  
user or service programming mode. The mA display  
indicate number of the currently selected program. Each  
program has a short name describing its purpose in the  
time/dose display.  
1.2 Exhibition mode  
OP200 has a special exhibition mode in which no X-ray  
radiation is generated. It is activated by setting the I/O  
board jumper X11 to ON.  
5140241-100 rev 3  
Instrumentarium Dental  
1
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
1 Introduction  
1.3 Service program quick reference  
SERVICE PROGRAM  
REFERENCE TABLE  
Sr  
70  
ERROR LOG PROGRAM  
Display failure counters.  
Log  
Sr  
SET TRIAL PERIOD LIMIT  
71  
Set the length of the trial period (lease time).  
PAy  
Sr  
72  
LATERAL CASSETTE ADJUSTMENT  
(Film units only)  
LCA  
Adjust the symmetry of TMJ image areas.  
Sr  
INPUT / OUTPUT SIGNAL CHECK  
74  
Display the states of the CPU's input/output signals.  
IOC  
Sr  
WARMING UP SEQUENCE  
76  
PUP  
Warm up the X-ray tube after installation or when the  
unit has not been used in a while.  
Sr  
PREHEAT ADJUSTMENT  
77  
Perform an automatic preheat adjustment.  
Prh  
Sr  
78  
thA  
DOSE CALIBRATION CONSTANT SELECTION  
Set the dose calibration constant according to the  
tubehead's dose rate.  
Sr  
79  
LINE VOLTAGE DISPLAY  
Display the line voltage.  
SUP  
Sr  
80  
CrL  
INDIVIDUAL MOTOR MOVEMENT TESTS  
Test drive individual motors to ensure proper  
operation.  
Sr  
BITEPLATE TEST PROGRAM  
81  
bPL  
Display the currently attached biteplate and the  
corresponding input voltages (only with VT enabled).  
Sr  
COLLIMATOR INFORMATION PROGRAM  
82  
Display collimator configuration and I/O signal states.  
COL  
Sr  
LINEAR ADJUSTMENT VALUE  
87  
LAL  
Adjust the linear position of the sharp layer in VT  
images (only with VT enabled).  
2
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
1 Introduction  
SERVICE PROGRAM  
REFERENCE TABLE  
Sr  
Digital units: SECONDARY COLLIMATOR  
88  
ALIGNMENT  
CAL  
Film units: CEPHALOSTAT BEAM ALIGNMENT  
Adjust the horizontal position the x-ray beam in  
cephalometric imaging programs.  
Sr  
COUNTRY OPTIONS  
89  
COP  
Configure the unit according to installed options,  
country specific regulations and user preferences.  
Sr  
90  
PIn  
PANORAMA INSTALLATION PROGRAM  
Adjust the AEC sensitivity and collimator alignment  
for panoramic imaging.  
Sr  
CEPH INSTALLATION PROGRAM  
91  
Adjust the alignment for cephalometric imaging.  
CIn  
Sr  
DISABLE/ENABLE CHECKS  
92  
ChE  
Temporarily bypass self-checks performed by the  
unit.  
Sr  
BEAM ALIGNMENT BOARD ADJUSTMENT  
93  
Adjust the sensitivity of the beam alignment board.  
bAL  
5140241-100 rev 3  
Instrumentarium Dental  
3
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
1 Introduction  
1.4 How to use the service programs  
1. Turn the unit on.  
2. Press (hold for five seconds) OK to enter user  
programming mode.  
3. Press the left and right button to enter service  
programming mode in the following order:  
press and hold left,  
press and hold right,  
release left, release right.  
A tone confirms that service mode is entered.  
4. Press the up button to scroll pass the user programs  
and to access the service programs.  
5. Press the up and down buttons to scroll the list of  
service programs. Select the desired Sr program by  
pressing OK.  
6. See chapter 2 Service program features on how to use  
the programs.  
7. Press OK to exit the program.  
NOTE! Changed settings are stored when a program is  
exited if not stated otherwise in the program description. If  
the unit is switched off while a program is still running, the  
settings may not be stored.  
8. Press (and hold for five seconds) OK to enter back to  
normal operating mode.  
WARNING! Always reset the unit by switching power off  
and on again after accessing the service programming  
mode. This prevents unauthorized access to service  
programs and leaves the unit in state that is safe for the  
user and patient.  
4
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
2 Service program features  
2.1 Sr 70 Log: Display error log  
Each time a failure occurs in OP200, a counter  
corresponding to the failure is incremented. This program  
is used to view the readings of these counters.  
NOTE! Only failures that are relevant for troubleshooting  
and failure diagnostics are recorded (see the table below  
for details). Some failure counters have non-zero values at  
the time of installation. This is normal.  
1. Select Sr 70 Log.  
2. Press the up and down buttons to scroll the list of  
counters. The control panel shows the current counter,  
continually switching between its failure code and  
counter value.  
3. Press OK to exit the program.  
FILM UNIT OP200  
DIGITAL UNIT  
OP200 D  
kV-  
display  
mA-  
display  
Second /  
dose -  
mA-  
display  
Second /  
dose -  
display  
display  
Ch / Sy /  
Er  
error  
number  
error  
number  
Error name  
FILM UNIT  
Error name  
DIGITAL  
UNIT  
Ch  
Ch  
5
***  
5
***  
LINE VOLTAGE  
LINE VOLTAGE  
13  
16  
21  
LbL  
IMAGE LABEL  
FAILURE  
Ch  
Sy  
16  
21  
StP  
EMERGENCY  
SWITCH  
StP  
EMERGENCY  
SWITCH  
HHo  
TUBEHEAD  
HOT  
HHo  
TUBEHEAD  
HOT  
5140241-100 rev 3  
Instrumentarium Dental  
5
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
FILM UNIT OP200  
DIGITAL UNIT  
OP200 D  
kV-  
display  
mA-  
display  
Second /  
dose -  
mA-  
display  
Second /  
dose -  
display  
display  
Ch / Sy /  
Er  
error  
number  
error  
number  
Error name  
FILM UNIT  
Error name  
DIGITAL  
UNIT  
Sy  
Sy  
22  
23  
Arc  
22  
23  
Arc  
TUBEHEAD  
ARC  
TUBEHEAD  
ARC  
Inu  
INVERTER  
FAILURE  
Inu  
INVERTER  
FAILURE  
Sy  
Sy  
24  
25  
FIL  
24  
25  
FIL  
FILAMENT  
FAILURE  
FILAMENT  
FAILURE  
AEC  
AEC  
AEC  
AEC  
FREQUENCY  
ERROR  
FREQUENCY  
ERROR  
Sy  
27  
Por  
ROTATION  
MOVEMENT  
FAILURE  
27  
Por  
ROTATION  
MOVEMENT  
FAILURE  
Sy  
Sy  
28  
29  
PoC  
28  
29  
CCd  
CASSETTE  
MOVEMENT  
FAILURE  
CAMERA  
ERROR  
PoL  
LINEAR  
PoL  
LINEAR  
MOVEMENT  
FAILURE  
MOVEMENT  
FAILURE  
Sy  
30  
31  
PoH  
30  
PoC  
CASSETTE  
HOLDER  
CEPH  
MOVEMENT  
FAILURE  
MOVEMENT  
FAILURE  
Sy  
Sy  
PoU  
Z MOVEMENT  
FAILURE  
31  
32  
PoU  
Z MOVEMENT  
FAILURE  
PoA  
CEPH BEAM  
ALIGNMENT  
FAILURE  
Sy  
Sy  
33  
nSy  
NO SYNC  
ERROR  
33  
34  
nSy  
NO SYNC  
ERROR  
6
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
FILM UNIT OP200  
DIGITAL UNIT  
OP200 D  
kV-  
display  
mA-  
display  
Second /  
dose -  
mA-  
display  
Second /  
dose -  
display  
display  
Ch / Sy /  
Er  
error  
number  
error  
number  
Error name  
FILM UNIT  
Error name  
DIGITAL  
UNIT  
Sy  
Sy  
Sy  
Sy  
35  
36  
37  
38  
COL  
AUTO-  
COLLIMATOR  
FAILURE  
Er  
Er  
40  
43  
CPU  
40  
43  
CPU  
CPU ERROR  
CPU ERROR  
***  
***  
MAINS SWITCH  
MAINS SWITCH  
Er  
Er  
44  
45  
FIL  
44  
45  
FIL  
FILAMENT  
REFERENCE  
ERROR  
FILAMENT  
REFERENCE  
ERROR  
InP  
INPUT ERROR  
InP  
INPUT ERROR  
2.2 Sr 71 PAy: Set lease period  
Allows limited quantity of exposures. After this exposure is  
prohibited and the control panel displays Er 46 PAy  
message. For equipment leasing and trial period purposes.  
The lease number can be set within the range 0...2000 or  
to the special value OFF (feature disabled). Number zero  
(0) activates Er 46 PAy, which prevents exposures. Other  
numbers adjust the stored exposure counter into a number  
above the total exposure count.  
5140241-100 rev 3  
Instrumentarium Dental  
7
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
Press OK to test if test limit has been set in normal  
operation. The cumulative exposure counter value is  
shown, followed by the number of allowed exposures (all  
indicators are lit).  
NOTE! The trial period refers to the number of exposures  
allowed in normal operating or QA mode. Exposures made  
while the unit is in service mode are compensated  
automatically.  
Usage:  
1. Select Sr 71 Pay. The time/dose display shows OFF  
or a number from 1 to 2000.  
2. Disable the lease period feature or select the desired  
lease period using the left and right buttons.  
3. Press OK to exit the program.  
2.3 Sr 72 LCA: Adjust lateral TMJ image area  
(film units only)  
For adjusting the horizontal alignment of TMJ image areas  
on the film.  
With imaging program P7 (TMJ Lateral view jaw closed  
and open) , the two first images are shown on the edges of  
the film. The next two are shown in the middle of the film.  
The center image area can be adjusted in respect to the  
other image pair. The adjustment range is -10...+10 mm, in  
steps of 0.5 mm.  
Usage:  
1. Select Sr 72 LCA. Use the customer’s latest TMJ  
Lateral view jaw closed and open X-ray image, or take  
a new image.  
2. Select imaging program P7 and the lowest technique  
factors. Make an exposure and process the film.  
8
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
3. Check the film for center image pair symmetry.  
Calculate the amount (-0.25...+1.00 cm) of correction  
needed.  
4. Select Sr 72 LCA. The control panel shows a number  
from -0.25 to 1.00. The number indicates the  
adjustment of the positioning of the second image pair  
areas (center) compared to the first image pair areas  
(on each side).  
5. Use the left and right buttons to shift the middle image  
pair to the left / right in steps of 0.5 mm.  
6. Press OK to exit the program.  
7. Enter the normal operating mode. Make an exposure  
and process the film. Check the film for symmetry. If  
the result is not satisfactory, return to Sr 72 LCA and  
repeat the steps 1 to 7.  
5140241-100 rev 3  
Instrumentarium Dental  
9
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
2.4 Sr 74 IOC: Display CPU input/output  
signal check  
For checking input and output signal states as seen by the  
CPU. Check whether a signal reaches the CPU correctly,  
when troubleshooting electrical problems related to e.g.  
microswitches and buttons.  
The I/O signals are divided into several pages, named 0L -  
0H - 1L - 1H - 2L - 2H - 3L - 3H - 4L - 4H - 5L - 5H - 6L - 6H.  
The tables below describe the signals shown on each  
page.  
The current page is shown on the kV/mA displays.  
Individual signal states are shown as on/off signals by the  
imaging program LEDs. A LED being lit means that the  
corresponding signal is active. A hexadecimal number  
summarizing the whole page is shown on the time/dose  
display.  
Usage:  
1. Select Sr 74 IOC.  
2. Use the left and right buttons to select page.  
3. Check if the signal state shown on the control panel  
matches the corresponding electrical signal level. If  
not, the signal does not reach the CPU correctly.  
4. Decide signal states according to the tables. For  
trouble shooting electrical connections, use the CPU  
board schematics to trace the signals.  
NOTE! The control panel display is updated with a small  
delay. Remember this when interpreting frequently  
changing signals.  
5. Press OK to exit the program.  
10  
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
Sr 74 IOC, page 0L  
Signal name  
Description  
LED  
(Status, when the LED is lit)  
PROJLIT  
PILLED  
Projector lights (on)  
1
5
Layer adjustment led, closest to  
pillar  
CENLED  
PATLED  
Layer adjustment led, center  
6
7
Layer adjustment led, closest to  
patient  
Sr 74 IOC, page 0H  
Signal name  
Description  
LED  
(Status, when the LED is lit)  
LASLIT  
ZDIR  
Panorama laser lights  
2
5
Vertical movement direction  
(down)  
ZENA  
Vertical movement enable  
Rack movement direction (up)  
Rack movement enable  
6
7
8
RACKDIR  
RACKENA  
Sr 74 IOC, Input page 1L  
Signal name  
CASDIR  
Description  
(Status, when the LED is lit)  
LED  
1
Cassette movement direction  
(left)  
CASENA  
ROTDIR  
Cassette movement enable  
2
3
Rotational movement direction  
(CW)  
ROTENA  
CEPHDIR  
CEPHENA  
LINDIR  
Rotational movement enable  
Ceph movement direction (right)  
Ceph movement enable  
4
5
6
7
Linear movement direction  
(toward pillar)  
LINENA  
Linear movement enable  
8
5140241-100 rev 3  
Instrumentarium Dental  
11  
D500293, 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
2 Service program features  
Sr 74 IOC, page 1H  
Signal name  
Description  
LED  
(Status, when the LED is lit)  
PPOWER  
PIMAGE  
Panoramic camera power enable  
Panoramic camera image enable  
1
2
3
CPOWER  
Cephalostat camera power  
enable  
CIMAGE  
CLASLIT  
Cephalostat camera image  
enable  
4
7
Ceph laser light  
Sr 74 IOC, page 2L  
Signal name  
EXPSW  
Description  
(Status, when the LED is lit)  
LED  
1
Exposure switch, remote or ctrl.  
panel  
INSTRUSW  
Factory mode jumper, digital I/O  
board X14  
2
MAOK  
Tube current monitoring  
3
4
PNLEXPSW  
Exposure switch on control panel,  
also triggers EXPSW  
EXHSW  
MAINS  
Exhibition mode jumper,  
Digital I/O board X11  
5
7
Mains voltage selection (110 v)  
Sr 74 IOC, page 2H  
Signal name  
PANCASSW  
CASLIMSW  
CASMIDSW  
LINMIDSW  
Description  
(Status, when the LED is lit)  
LED  
Panoramic cassette inserted /  
Autocollimator in slot  
2
3
4
5
Cassette limit switch / head  
support vertical position (down)  
Cassette middle switch /  
Autocollimator presence detection  
Linear movement middle switch  
12  
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
Sr 74 IOC, page 2H  
LINLIMSW  
Linear movement limit switch  
6
7
RACKMIDSW  
Cassette rack middle switch /  
Autocollimator Cu filter ON, S61  
RACKLIMSW  
Cassette rack limit switch /  
8
Autocollimator Cu filter OFF, S62  
Sr 74 IOC, page 3L  
Signal name  
WD  
Description  
(Status, when the LED is lit)  
LED  
5
CPU Watchdog timer  
Sr 74 IOC, page 3H  
Signal name  
Description  
LED  
(Status, when the LED is lit)  
No external signals  
Sr 74 IOC, page 4L  
Signal name  
Description  
LED  
(Status, when the LED is lit)  
HOME(RIGHT) Patient positioning button, right  
side  
1
2
3
4
HOME(LEFT)  
PILWARD  
Patient positioning button, left  
side  
Layer adjustment button, towards  
pillar  
CENWARD  
PATWARD  
Layer adjustment button, center  
Layer adjustment button, towards  
patient  
5
6
7
TEMPFAIL  
STOPSW  
Tubehead overheated  
Emergency stop switch  
5140241-100 rev 3  
Instrumentarium Dental  
13  
D500293, 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
2 Service program features  
Sr 74 IOC, page 4H  
Signal name  
Description  
LED  
(Status, when the LED is lit)  
Collimator position signal 1  
Collimator position signal 2  
Collimator position signal 3  
COL1SW  
COL2SW  
COL3SW  
SIDE  
1
2
3
4
Start position button side  
detection (left)  
START  
Start position button trigger,  
activated by either start position  
button  
5
PROJTRIG  
ZDOWN  
ZUP  
Positioning lights button  
6
7
8
Vertical movement down button  
Vertical movement up button  
Sr 74 IOC, page 5L  
Signal name  
CEPHLOK  
CEPHCOK  
CEPHROK  
Description  
(Status, when the LED is lit)  
LED  
Ceph cassette detection /  
Beam alignment board left sensor  
1
2
3
Beam alignment board mid  
sensor  
Beam alignment board right  
sensor  
LAT/PA  
CEPHL  
Earholder position detection (P/A) 4  
Ceph at left side  
5
(Digital I/O board jumper X15: L)  
CEPHR  
Ceph at right side  
6
(Digital I/O board jumper X15: R)  
CEPHDOWN  
Vertical movement up button  
on cephalostat  
7
14  
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
Sr 74 IOC, page 5L  
CEPHUP  
Vertical movement down button  
on cephalostat  
8
Sr 74 IOC, page 5H  
Signal name  
ROT1SW  
ROT2SW  
ROT3SW  
ROT4SW  
Description  
(Status, when the LED is lit)  
LED  
Rotational movement switch 1  
(see fig 2.1)  
1
2
3
4
Rotational movement switch 2  
(see fig 2.1)  
Rotational movement switch 3  
(see fig 2.1)  
Rotational movement switch 4  
(see fig 2.1)  
ZMIDSW  
Vertical movement mid switch  
Vertical movement limit switch  
Cephalostat middle switch  
Cephalostat limit switch  
5
6
7
8
ZLIMSW  
CEPHMIDSW  
CEPHLIMSW  
Fig 2.1. Led procedures concerning movement of the rotating unit.  
5140241-100 rev 3  
Instrumentarium Dental  
15  
D500293, 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
2 Service program features  
Sr 74 IOC, page 6L and 6H  
Signal name  
Description  
LED  
(Status, when the LED is lit)  
No external signals  
2.5 Sr 76 PUP: X-ray tube warm-up  
WARNING! Generates x-rays!  
To facilitate testing of a defective tubehead assembly and  
making a new tubehead, or one that has not been used for  
a long time, ready for normal use.  
NOTE! OP200, that has not been used for three months or  
more, should be warmed up using this program. Always  
run this program after replacing the tubehead assembly.  
Usage:  
1. Select Sr 76 PUP.  
2. Make an exposure.  
OP200 exposes with the lowest kV / mA values.  
3. The exposure values automatically rise to the next kV /  
mA level (see the table below). Keep making  
exposures until the highest values are reached or until  
FAIL is displayed on the kV and mA display.  
A FAIL message indicates that arcing and/or  
breakdown has occurred in the tubehead. Restart the  
test by pressing the OK button twice. If the FAIL  
message  
occurs  
repeatedly,  
check  
OP200  
Troubleshooting Manual, chapter Sy 22 Arc for further  
instructions.  
16  
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
Sr 76 PUP: kV & mA Feedback Reference Voltages  
kV  
20  
30  
40  
50  
54  
57  
60  
63  
66  
70  
73  
77  
81  
85  
85  
85  
85  
85  
85  
85  
85  
85  
85  
mA  
1.0  
1.0  
1.0  
1.0  
1.0  
1.0  
1.0  
1.0  
1.0  
1.0  
1.0  
1.0  
1.0  
1.0  
2.0  
2.5  
3.2  
4.0  
5.0  
6.4  
8.0  
10.0  
12.0  
s
(VkVref)  
(1.00 V)  
(1.50 V)  
(2.00 V)  
(2.50 V)  
(2.70 V)  
(2.85 V)  
(3.00 V)  
(3.15 V)  
(3.30 V)  
(3.50 V)  
(3.65 V)  
(3.85 V)  
(4.05 V)  
(4.25 V)  
(4.25 V)  
(4.25 V)  
(4.25 V)  
(4.25 V)  
(4.25 V)  
(4.25 V)  
(4.25 V)  
(4.25 V)  
(4.25 V)  
(VmAref)  
(0.24 V)  
(0.24 V)  
(0.24 V)  
(0.24 V)  
(0.24 V)  
(0.24 V)  
(0.24 V)  
(0.24 V)  
(0.24 V)  
(0.24 V)  
(0.24 V)  
(0.24 V)  
(0.24 V)  
(0.24 V)  
(0.49 V)  
(0.61 V)  
(0.78 V)  
(0.97 V)  
(1.22 V)  
(1.56 V)  
(1.94 V)  
(2.43 V)  
(2.92 V)  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
3.2  
4. Press OK to exit the program.  
5. Switch the unit off and on again to reset it to normal  
operating mode.  
6. Ensure proper operation by making a panoramic  
exposure (P1) using maximum exposure values.  
5140241-100 rev 3  
Instrumentarium Dental  
17  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
2.6 Sr 77 Prh: Adjust the preheat setting  
WARNING! Generates X-rays!  
To make the unit adjust its filament preheat level. The  
adjustment is done automatically, you only need to press  
the exposure button.  
Usage:  
1. Select Sr 77 Prh. The exposure values that are used  
during the exposure are shown on the kV and mA  
displays. The current preheat reference is shown on  
the time/dose display.  
2. Press and hold the exposure button. OP200 starts  
increasing the preheat reference, continuously  
monitoring the mA feedback from the filament. The  
current reference value is shown on the time/dose  
display.  
3. When a proper mA feedback value is received from  
the filament, the preheat reference feedback reference  
is locked and the exposure is terminated. The final  
preheat reference value (between 50 and 60) is  
displayed on the time/dose display. If correct mA  
feedback was not reached during the exposure, the  
previous preheat reference is displayed.  
4. Press OK to exit the program.  
18  
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
2.7 Sr 78 thA: Set Dose Calibration Constant  
(DCC)  
For setting the Dose Calibration Constant (DCC). The unit  
needs this constant to be able to calculate the dose  
correctly. If the calibration constant is missing, a Ch 12  
dCC failure code is shown when the unit is switched on.  
Set the dose calibration constant according to the dose  
rate of the tubehead, which is measured as part of the  
manufacturing process of the unit. The measured value is  
recorded on a label on the back of the tubehead. Enter the  
value in steps on 0.1 R/min in the range 6.5-30 R/min.  
Since this value is supposed to change only when the  
tubehead is exchanged, the preheat offset (see Pr 77 Prh)  
and tubehead specific counter are set to 0 when a new  
dose calibration constant is entered.  
Usage:  
1. Select Sr 78 thA. By pressing the left and right buttons  
the kV display shows cncL (stands for "cancel") or  
t100.  
2. To exit without making changes, press OK.  
3. To edit the calibration constant for the current  
tubehead, select t100 by pressing the right button  
once. Move down to the calibration constant value  
shown in the time/dose display by pressing the down  
button once. Use the left and right button to modify the  
value.  
4. To store the new dose calibration constant and exit the  
program, press OK. To cancel any changes, select  
cncL by first pressing the up button and thereafter the  
left button.  
5. Press OK to exit the program.  
5140241-100 rev 3  
Instrumentarium Dental  
19  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
2.8 Sr 79 SUP: Display line voltage  
To view power supply line voltage as seen by OP200.  
Used both for checking if the line voltage is sufficient for  
the unit to operate correctly, and troubleshooting problems  
related to the voltage level (e.g. sporadic Ch 5 failures).  
The line voltage is derived from a +25V supply in the  
Filament Control Board. Thus if the +25V voltage is absent,  
for example due to the emergency stop button being  
depressed, the voltage is not correctly displayed.  
The accuracy of the voltage display is ±5%.  
Usage:  
1. Select Sr 79 SUP. The time/dose display shows the  
approximate line voltage value.  
2. Press OK to exit the program.  
2.9 Sr 80 CrL: Test motor movement  
Lets you test individual movements for proper operation.  
The movement type currently selected is shown in the kV  
field. In film units, the available tests are cassette (CA),  
cephalostat (CE), rotation (RO) and linear (LI). For digital  
units these are: automatic collimator (Co), cephalostat  
(CE), rotation (RO) and linear (LI). Only relevant tests are  
accessible, i.e. CE and Co are not shown if the  
corresponding options are not installed.  
The mA field shows the number of accelerations to be  
done in one test. The number is settable in the range 0-4  
for the collimator test (Co), and 1-5 for the others. Setting  
the value to 0 in the collimator test means that only the  
copper filter is moved.  
The speed ratio to be to used is shown in the time/dose  
display. If the ratio is set to 1.00, the movements are done  
using nominal speed (the speed most typically used) of the  
movement in question. Setting it to e.g. 0.50 means the  
movement is done at 50% of the nominal speed.  
20  
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
Some movements can be done with or without exposure.  
In test mode (T) the test is made without X-rays. In manual  
mode (M), a static exposure is active. Edit the exposure  
values in the kV and mA fields.  
Usage:  
1. Select Sr 80 CrL. The current movement type is  
shown on the kV display, the number of accelerations  
on the mA display and the speed ratio on the time/  
dose display.  
2. Move between the fields using the up and down  
buttons, and change the settings using the left and  
right buttons. Test by pressing the exposure button.  
3. A FAIL message is shown on the control panel. See  
OP200 Trouble Shooting Manual, chapter Sy 30 PoC  
(ceph), Sy 28 PoC (cassette), Sy 29 PoL (linear), Sy  
27 Por (rotation) or Sr 38 COL (automatic collimator)  
according to the current type of movement.  
4. If the movement succeeded, the difference between  
the measured and the expected length of the  
movement is displayed on the time/dose display (for  
negative values, the minus sign is shown on the mA  
display). Write down this value. No value is shown in  
the Co test.  
5. Repeat the test in the other direction by re-pressing  
the exposure button. Compare the value shown in the  
time/dose display to the value gained in the other  
direction. The difference should be kept within the  
tolerances listed in the table below for the unit to  
function properly.  
6. Press OK to exit the program.  
Motor  
movement test  
Limit values for  
acceptation  
CA (cassette)  
CE (cephalostat)  
ro (rotation)  
li (linear)  
± 1.0 mm  
± 1.0 mm  
o
± 2.0  
± 1.2 mm  
5140241-100 rev 3  
Instrumentarium Dental  
21  
D500293, 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
2 Service program features  
2.10 Sr 81 bPL: Display biteplate properties  
(Volumetric Tomography option only)  
For checking the electrical operation of the Volumetric  
Tomography patient positioning and calibration devices.  
Two analog signals are used with these devices; one  
senses the device type, and the other the position of the  
device. The signals are named DEV and POS on OP200  
wiring diagram. The information is shown on the control  
panel as follows:  
„ The kV display shows the currently selected side when  
the projection biteplate is attached. The side and  
region are also indicated by the section display.  
Sections  
1. Anterior  
2. Premolar  
3. Molar  
4. Jaw joint (not in use)  
„ The mA display shows the type of the device.  
„ The time and dose indicator LEDs indicate the  
currently selected signal (s for POS and mGycm2 for  
DEV). The signal level of the selected channel is  
shown in the time/dose display.  
POS (only valid for projection bite plate)  
Voltage  
Position  
kV display  
0,65 +/- 0,2V  
1,4 +/- 0,2V  
2,02 +/- 0,2V  
2,73 +/- 0,2V  
3,44 +/- 0,2V  
Left Molar  
LE  
LE  
--  
Left Canine  
Incisor  
Right Canine  
Right Molar  
ri  
ri  
22  
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
DEV  
Voltage  
Device  
mA display  
1,35 +/- 0,25V  
1,8 +/- 0,25V  
2,7 +/- 0,25V  
Panorama  
Calibration  
Projection  
PA  
CA  
Pr  
Usage:  
1. Select Sr 81 bPL.  
2. Press the left / right button to view the voltage level for  
POS or DEV.  
3. Insert different patient positioning devices and/or turn  
the positioning device into different positions. Verify  
that the displayed values match the ones listed above.  
4. Press OK to exit the program.  
2.11 Sr 82 COL: Set/display collimator  
information  
In digital units: displays collimator information, which  
cannot be edited.  
„ The kV and mA displays show the type of collimator.  
“Std” stands for standard (manual) and “Auto” for  
automatic collimator.  
„ With automatic collimator, also some related signals  
are shown.  
– The signals PANCASSW, COL1SW, COL2SW and  
COL3SW are shown as P1 - P4. These LEDs  
directly match LED H2-H5 on the collimator board.  
– The states of the copper filter switches are shown as  
LED P5 (filter used) and P6 (filter not used).  
– Collimator position is shown in the time display. The  
position ranges from 0 to 10, uneven numbers being  
the in-slot positions.  
Usage:  
1. Select Sr 82 COL.  
2. Use the information displayed for e.g. troubleshooting  
purposes.  
3. Press OK to exit the program.  
5140241-100 rev 3  
Instrumentarium Dental  
23  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
In film units: To select the type of collimator aperture  
installed in each of the configurable slots. The slots are  
numbered 1-3 in the program. Slot number one is the one  
closest to the panorama slot, number two is in the middle,  
and three farthest away from the panorama slot. The  
information to be entered in the program is engraved in the  
collimator plate.  
NOTE! Available in digital units starting from software  
version R2.4  
Usage:  
1. Select Sr 82 COL.  
2. Select the slot to be edited using the up / down arrows.  
The current slot is shown blinking in the time/dose  
display a few seconds after the slot has changed.  
3. Select the correct aperture type for the current slot  
using the left / right buttons. The aperture size is  
shown in the kV and mA display. Additional  
information such as is the aperture is of type left (L),  
right (r) or symmetrical (S) is shown in the time/dose  
display. For horizontal apertures, an additional “H” is  
displayed in the time/dose display. Unconfigured or  
empty collimator slots are shown as “----”.  
Example: The installed collimator is named  
"18X24AH". The kV display shows "18" and the mA  
display shows "24". The time display shows "rH" or  
"LH", depending on whether the collimator is of type  
"LEFT" or "RIGHT"  
NOTE! Configuration of the first slot is not possible if  
OrthoTrans is enabled. This slot is shown as “ortr”.  
4. Press OK to exit the program.  
24  
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
2.12 Sr 86 Sto: Store settings  
To store the current configuration for future use as default  
values. The program affects the configuration into which  
the system is restored using the program Pr 53 nor. The  
stored values are listed as follows.  
Configuration item  
Set by user program  
Default procedure and mode  
Contrast levels  
Ortho Trans settings  
VT settings  
Pr 51 PUS  
Pr 52 pCo, Pr 52 gCo  
Pr 50 Lay  
Pr 50 S3d  
Beeper settings  
Pr 60 bEP  
Dose display options  
Horizontal ceph limit  
Pr 65 doS  
Pr 63 CEL  
Usage:  
1. Select Sr 86 Sto.  
2. To store the settings when the program is exited,  
select "on" by pressing the right button. To cancel  
storing the values, select "OFF" by pressing the left  
button.  
3. Press OK to exit the program.  
2.13 Sr 87 LAL: Set linear adjustment value  
for VT  
NOTE! See Volumetric Tomography® Option Installation &  
Adjustments Manual.  
For adjusting the linear position of the imaging layer in VT  
imaging programs.  
Usage:  
1. With Sr 92 Che set to nBp, take a VT panoramic  
image of the pin-ball phantom.  
2. Select Sr 87 LAL. The current adjustment value is  
shown on the time/dose display.  
5140241-100 rev 3  
Instrumentarium Dental  
25  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
3. Use the left/right buttons to enter the width of the ball  
in the kV field. Move the flashing light to the mA field  
using the down button. Enter the height of the ball  
using the left/right buttons. The program calculates a  
new linear adjustment value and displays it on the  
time/dose display.  
4. Press OK to exit the program.  
5. Take a new VT panorama image of the ball-pin  
phantom. If the ball is still not round (within a tolerance  
of 0.1 mm), repeat steps 2-5 using the height/width  
value from the newly taken image.  
The adjustment value can also be edited manually. This  
can be done for example if the correct value is known in  
advance. To edit the value manually, move the flashing  
light to the time/dose display. Modify the value using the  
left/right buttons.  
NOTE! The linear adjustment value is not changed if height  
or width is not entered, or if height and width are set equal  
(for example height=6.0, width=6.0).  
2.14 Sr 88 CAL: Disable automatic  
cephalostat alignment correction  
(digital units only)  
Lets you disable automatic adjustment of the cephalostatic  
beam during exposure. Automatic adjustment should be  
enabled to allow the unit to compensate for small beam  
alignment errors, but may be temporarily disabled for  
troubleshooting purposes.  
Usage:  
1. Select Sr 88 CAL. The current setting is displayed on  
the time/dose display. The value on means that  
automatic adjustment is enabled (normal usage) while  
OFF means that it's disabled (troubleshooting).  
2. Using the left and right buttons, disable or enable  
automatic adjustment as appropriate.  
3. Press OK to exit the program.  
26  
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
2.15 Sr 88 CAL: Set cephalostat alignment  
value (film units only)  
To align the THA with the cephalostat. The adjustment  
value set here affects the rotational angle of the THA, and  
thereby the horizontal position of the x-ray beam on the  
film. A positive adjustment value moves the beam to the  
right and vice versa. The value is shown in centimeters.  
Usage:  
1. Select Sr 88 CAL. The current adjustment value is  
shown on the time/dose display.  
2. Using the left and right buttons, modify the adjustment  
value as appropriate.  
3. Press OK to exit the program.  
2.16 Sr 89 COP: Set country options  
Lets you customize the unit according to regional  
regulations and user expectations.  
Lists a number of options, each having a number (shown in  
the kV field) and a short name (shown in the mA field). The  
value associated with the option is shown in the time/dose  
field. The options available are described below.  
5140241-100 rev 3  
Instrumentarium Dental  
27  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
Sr 89 COP: COUNTRY OPTIONS  
Option  
Value  
Description  
1
rE  
OFF  
Exposure is possible with both control  
panel and remote exposure button.  
on  
Exposure is possible only with the  
remote exposure button in AEC and  
Manual mode. In the test mode  
exposure is possible from both  
remote exposure button and control  
panel.  
2
C1  
(Film  
only)  
OFF  
on  
Image size 24x30 cm is allowed.  
Image size 24x30 is prohibited.  
Initiating an exposure with the 24x30  
collimator results in a collimator error.  
3
nA  
OFF  
on  
AEC is available in normal operating  
mode.  
AEC is disabled; OP200 can be used  
without AEC Board (useful in eg.  
OP200 CR models missing this  
feature).  
4
FE  
OFF  
on  
Exposure values are selected is steps  
as combinations of kV/mA/s. The  
difference in radiation output between  
each step is constant.  
"Free selection" is enabled. Exposure  
values (kV/mA/s) can be selected  
independently of each other.  
5
Er  
OFF  
on  
Error messages are not sent to the  
Ortho ID device.  
Error messages are sent to the Ortho  
ID.  
6
P6  
OFF  
on  
The Lateral TMJ imaging program is  
used as imaging program P6.  
The Ortho (axially corrected) Lateral  
TMJ imaging program is used as  
imaging program P6.  
7
Or  
(FILM  
ONLY)  
OFF  
on  
Ortho Trans is disabled.  
Ortho Trans is enabled.  
28  
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
Sr 89 COP: COUNTRY OPTIONS  
8
3d  
(DIGITA  
L ONLY)  
OFF  
on  
Volumetric Tomography is disabled.  
Volumetric Tomography is enabled.  
NOTE! Setting the 3D option on  
resets the VT exposure values to their  
default values.  
9
qA  
OFF  
on  
Quality assurance images consist of  
15 exposed areas using different kV/  
mA values.  
Quality assurance images are taken  
of special test objects in accordance  
with IEC 61223-3-4 (digital units  
only).  
10  
C2  
OFF  
on  
Only normal panoramic collimator is  
allowed in P1, P3-P5. Only pediatric  
(height limited) collimator is allowed in  
P2.  
Normal and pediatric (height limited)  
collimator can be used  
interchangeably in P1-P5.  
12  
SE  
OFF  
on  
Section selection disabled.  
Section selection enabled. Imaging  
can be restricted to certain regions of  
interest, reducing dose and imaging  
time (not available in all imaging  
programs).  
Usage:  
1. Select Sr 89 COP. One of the options is displayed.  
2. Select the desired option using the up and down  
buttons.  
3. Change the option's value to OFF by pressing the left  
button or on by pressing the right button.  
4. Press OK to exit the program.  
5140241-100 rev 3  
Instrumentarium Dental  
29  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
2.17 Sr 90 PIN: Panorama installation  
program  
NOTE! See OP200 installation manual.  
WARNING! Generates x-rays!  
For taking exposures without movements and can be used  
for troubleshooting and calibration. When this program is  
used, the KVOK and MAOK signals are not monitored.  
Therefore error codes Sy 23 Inu and Sy 24 FIL are not  
enabled.  
Usage:  
1. Select Sr 90 PIn.  
2. Press the up and down button to get the A (automatic  
mode), M (manual mode) or T (test mode) indicator  
light blinking. Select mode using the left and right  
buttons.  
3.  
„ Automatic collimator (selection only possible in  
Manual mode): Select collimator using the left and  
right buttons.  
„ Manual collimator: Move the collimator as  
appropriate. In AEC mode use the standard panoramic  
collimator.  
4. Select exposure values by pressing the up and down  
buttons to select the desired field and left and right  
buttons to edit the value. No exposure values can be  
entered in Test mode.  
5. Prepare the unit for exposure. In Manual mode, attach  
the fluorescent tool or enable imaging on the PC. In  
AEC mode, attach the calibration tool to the tubehead.  
6. Press the exposure switch to start the exposure. In  
AEC and Test mode the AEC frequency is displayed in  
the time/dose field.  
7. Make adjustments and repeat steps 1-7 if needed.  
8. Press OK to exit the program.  
30  
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
2.18 Sr 91 CIN: Cephalostat installation  
program (digital units only)  
For alignment adjustment of the cephalostat.  
NOTE! The cephalostat alignment of film based units is  
performed in normal operating mode. See OP200  
Installation Manual.  
The program has three modes of operation:  
„ Automatic mode (A) produces an adjustment image by  
keeping the x-ray beam still while the cephalostat  
camera and secondary collimator are driven from a  
position 5 cm before the beam center to 5 cm after it.  
„ Manual mode (M) performs an exposure without  
movements. The resulting image can be used for  
adjusting the collimator's alignment.  
„ Test mode (T) positions the unit for exposures in  
Manual mode. Set an adjustment value affecting the x-  
ray beam's horizontal position in the time/dose field. A  
positive value moves the x-ray beam to the right and  
vice versa. The value is given in centimeters.  
Usage:  
1. Select Sr 91 CIn.  
2. Press the up and down button until one of the A, M or  
T indicator lights is blinking. Select mode using the left  
and right buttons.  
3. Automatic collimator: In mode M: select normal or  
pediatric type collimator by pressing the up button until  
the pediatric collimation indicator light is blinking.  
Press the left button to use the standard height  
collimator (indicated as LED 9a or 9b being lit).  
4. Select exposure values by pressing the up/down  
buttons to select the desired field and left/right buttons  
to edit the value. Only the adjustment value can be  
edited in Test mode.  
5140241-100 rev 3  
Instrumentarium Dental  
31  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
5. Make the system ready for image capturing (modes A  
and M).  
6. Press the exposure button to make an exposure  
(modes A and M) or position the unit (mode T).  
7. Make adjustments and repeat steps 1-7 if needed.  
8. Press OK to exit the program.  
2.19 Sr 92 CHE: Set level of self-checks  
For temporarily disabling checks and operations that  
normally are performed before exposure is started.  
WARNING! The checks may only be disabled while the  
unit is operated by service personnel, and only for the  
purpose of installation and maintenance. Make sure to  
power cycle the unit before handling it over to the user!  
The checks/operations are divided into different levels. The  
table below outlines which checks are disabled on each  
level. Some of the levels are only available in digital / film  
based units.  
Disabled check or operation  
Level  
PC link  
Collimator, Cassette  
Earholder,  
Biteplate  
Radiation  
output  
Ceph beam  
alignment1)  
OFF  
nPC  
nCO  
nCA  
nbP  
nCh  
nra  
„
„
„
„
„
„
„
„
„
„
„
„
non  
„
„
32  
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
1)  
With manual collimator, collimator type checking is  
disabled. With automatic collimator, this level disables  
driving at the collimator and earholder position checking.  
Usage:  
1. Select Sr 92 Che.  
2. Using the left and right button, select the desired level.  
Select OFF to enable all checks and operations.  
3. Press OK to exit the program. Checking is set to the  
selected level.  
2.20 Sr 93 BAL: Adjust beam alignment board  
sensitivity  
NOTE! See the OP200 installation manual.  
For adjusting the sensitivity of the sensors on the beam  
alignment board. It assists in driving the cephalostat  
camera and secondary collimator into the correct position  
and selecting appropriate exposure values.  
This program has two modes of operation:  
„ Test mode positions the unit according to the currently  
selected channel.  
„ Manual mode performs an exposure without  
movements.  
The AEC density LEDs indicate the currently selected  
channel. The midmost LED corresponds to the mid  
channel on the Beam alignment board.  
In units equipped with automatic collimator, the selected  
collimator type (normal height / pediatric) is indicated by  
the pediatric collimation LED.  
Usage:  
1. Select Sr 93 BAL.  
2. Select the channel to be adjusted by pressing the up /  
down buttons until one of the AEC density LEDs is  
blinking. Change channel by pressing the left / right  
buttons.  
3. Select correct collimator.  
Manual collimator: Move the cephalometric collimator  
into position. Use copper filtration as appropriate for the  
channel.  
5140241-100 rev 3  
Instrumentarium Dental  
33  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
2 Service program features  
Automatic collimator: Press the up button until the  
pediatric collimation LED is blinking. Press the left button to  
use normal height collimation or right to use pediatric  
collimation.  
4. Select Test mode. Press the exposure button to drive  
the unit into position for exposure.  
5. Select Manual mode. Press the exposure button to  
make an exposure. The AEC density LEDs reflect the  
channel levels during exposure.  
6. Adjust the channel so that it’s barely triggered during  
exposure (the corresponding LED is lit). Repeat steps  
2 to 6 for all channels.  
7. Press OK to exit the program.  
34  
Instrumentarium Dental  
514024100 rev 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  
Instrumentarium Dental reserves the right to make changes in specification and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your  
Instrumentarium Dental representative for the most current information.  
Copyright © 05/2008 by PaloDEx Group Oy. All rights reserved.  
Instrumentarium Dental  
Nahkelantie 160, FI-04300 Tuusula, Finland  
Tel. +358 45 7882 2000  
Fax +358 9 851 4048  
Americas:  
Instrumentarium Dental Inc.  
Milwaukee, Wisconsin, U.S.A.  
Tel. 800 558 6120  
Fax 414 481 8665  
Orthopantomograph® OP200 & OP200 D  
Orthoceph® OC200 & OC200 D  
Service Program Manual, English  
5140241-100 rev 3 Printed in Finland 5/08  
Reviewed: Rintamäki Markus 2008-05-16 15:20  
Approved: Äärynen Teemu 2008-05-19 06:07  
D500293, 3  

相关型号:

OC200D

Service Program Manual
ETC

OC205H12A17-FREQ

HCMOS Output Clock Oscillator, 10MHz Min, 20MHz Max
PLETRONICS

OC205H12A27-10M

HCMOS Output Clock Oscillator, 10MHz Nom
PLETRONICS
PLETRONICS
PLETRONICS

OC205H12B17-10M

HCMOS Output Clock Oscillator, 10MHz Nom
PLETRONICS

OC205H12B17-FREQ

HCMOS Output Clock Oscillator, 10MHz Min, 20MHz Max
PLETRONICS

OC205H12B27-10M

HCMOS Output Clock Oscillator, 10MHz Nom
PLETRONICS

OC205H12B27-FREQ

HCMOS Output Clock Oscillator, 10MHz Min, 20MHz Max
PLETRONICS

OC205H12B28-10M

HCMOS Output Clock Oscillator, 10MHz Nom
PLETRONICS
PLETRONICS

OC205H12B28-FREQ

HCMOS Output Clock Oscillator, 10MHz Min, 20MHz Max
PLETRONICS