KELLER  
INTELLIGENT DIGITAL MANOMETERS  
MANO GAUGE / MANO RECORD  
MANO CALI  
This is a highly accurate microprocessor-controlled pressure measuring instrument. SensoBrainTM, a KELLER-designed low power signal  
processor, transmits the signals of the pressure- and temperature sensor to the microprocessor, which digitally compensates the sensor data.  
This mathematical modelling of the pressure sensor virtually eliminates all errors over a wide temperature range.  
All KELLER- and KAVLICO-sensors are compatible with the electronics of the Intelligent Manometers.  
MANO GAUGE (Basic Version)  
Display of:  
Alternative:  
Pressure, Pressure Unit, Temperature, Elapsed Time  
Pressure, Pressure Unit, Max.- and Min. Value  
All functions are executed with the front key of the instrument. It allows the display  
to be set to zero and to change the measuring cycle for prolonged battery life.  
Two pressure sensors can be connected to the MANO GAUGE (optional).  
Analog outputs can be provided 4…20 mA or 0…10 V (optional).  
MANO RECORD (MANO GAUGE plus data recording)  
Based on the MANO GAUGE, the MANO RECORD records and stores the measured  
values. The front key is used to select which data shall be displayed and stored,  
and to set the recording interval and the number of readings (memory for 960 or  
optionally 8100 readings). The read-out of the stored data is via RS 232 serial  
interface (Logger Software and K101 cable available as option).  
MANO CALI (CALIBRATOR VERSION plus data recording)  
In the basic mode, the MANO CALI displays: Pressure, Pressure Unit, Max.- and Min. Values.  
Units are selected via the front key (bar, PSI, kPa, MPa,…). MANO CALI also incorporates the  
functions TARA (sets a new temporary zero reference), LEAK (measures a pressure change  
over a programmable time) and RES (reduces the standard resolution by factor 10).  
A seperate HOLD-key on the instrument freezes the reading. This instrument has the same  
data recording features as the MANO RECORD.  
HOLD-Key  
(Mano Cali  
only)  
ø 76 mm  
KELLER  
KELLER  
KELLER  
Front Key  
Button  
ø 45  
HEX 27  
HEX 19  
G 1/4"  
G 1/4"  
G 1/4"  
Integrated Capacitive Cell  
(30 mbar…1 bar)  
Series 11 Transducer  
(1…10 bar / -1…2 bar / -1…1000 bar)  
Series 6 Transducer  
(10…600 bar)  
Subject to alterations  
3/98  
KELLER AG für Druckmesstechnik  
KELLER Gesellschaft für druckmesstechnik mbH  
St. Gallerstrasse 119  
Schwarzwaldstrasse 17  
CH-8404 Winterthur  
D-79798 Jestetten  
Tel. 052 - 232 11 26  
Tel. 07745 - 9214 - 0  
Fax 052 - 232 70 67  
Fax 07745 - 9214 - 50  
Companies approved to ISO 9001 / EN 29001  
Internet: http://www.keller-druck.ch  
KELLER  
SPECIFICATIONS  
MANO GAUGE / MANO RECORD  
Vented Gauge or Absolute Sealed Gauge or Absolute  
MANO CALI  
Sealed or Vented Gauge  
In MANO GAUGE and MANO RECORD the  
measuring ranges are divided by 12 bit  
or 3000 points. In MANO CALI over 13  
bit or 6000 points.  
Range  
Resolution  
0,01 mbar  
0,1 mbar  
0,3 mbar  
1 mbar  
Range  
Resolution  
Range  
Resolution  
0,5 mbar  
1,8 mbar  
4,3 mbar  
0,03 bar  
0,05 bar  
0,1 bar  
30 mbar  
300 mbar  
1000 mbar  
-12 bar  
10 bar  
-12 bar  
-110 bar  
-125 bar  
200 bar  
100 bar  
300 bar  
600 bar  
1000 bar  
0,03 bar  
0,1 bar  
0,2 bar  
0,3 bar  
The resolution is determined by the  
measuring range and the number of  
measuring points.  
3 mbar  
350 bar  
30 bar  
10 mbar  
700 bar  
• Specific Characteristics  
Total Accuracy (1)  
• Environmental Characteristics  
MANO GAUGE / RECORD  
MANO CALI  
0,05 % typ.  
0,1% max.  
2
Operating Temperature  
Compensated Range (5)  
Storage Temperature  
Display (5)  
°C  
°C  
°C  
°C  
- 1060  
050  
Range 0…50°C  
± 0,1% typ.  
± 0,2% max.  
1
- 3080  
050  
Measuring Interval (2)  
sec  
(1)  
(5)  
Includes linearity, repeatability, hysteresis, temperature and resolution of the display  
x 2 for two transducers (not for MANO CALI)  
others upon request  
(2)  
• Mechanical Characteristics  
Manometer Housing  
Transducer  
• Electrical Characteristics  
Supply  
Polyamid 12  
Lithium-Battery 3,6 V (included)  
> 200 days at continuous operation (4)  
See data sheet of transducer  
approx. 250 g  
Battery-Life  
Total Weight  
Output Signal (optional)  
- Analog (3) (MANO RECORD / GAUGE) 4..20 mA or 0..10 V  
Pressure Connection  
Protection  
See data sheet of transducer  
IP 65  
- Numerical (MANO RECORD / CALI)  
RS 232  
(3)  
(4)  
The use of an analog signal requires an external supply between 10…28 VDC  
> 100 days for MANO CALI  
• Options (on request)  
- Manometer Housing in Metal*  
- 1 or 2 external Transducers (remote)*  
- Transmitter Output Connector*  
- Special Programs  
- Logger Software and K101 Cable  
- Intrinsically Safe Version (EEx-ia)*  
* only MANO RECORD and MANO GAUGE  
LOGGER  
The Logger-Software allows the read-out of the stored  
data from the Intelligent Manometer. The data transfer  
takes place via the serial interface on the PC.  
The MANO RECORD / MANO CALI is connected through a  
cable K101 (supplied with Logger software).  
The LOGGER-software offers many further processing  
possibilities of the read-out data, and meets all the  
requirements of modern Windows-Software:  
- Graphical Display  
- Tabulated Display  
- Export into other Windows-Programs  
- Storage into Files  
Subject to alterations  
3/98  
KELLER AG für Druckmesstechnik  
KELLER Gesellschaft für druckmesstechnik mbH  
St. Gallerstrasse 119  
Schwarzwaldstrasse 17  
CH-8404 Winterthur  
D-79798 Jestetten  
Tel. 052 - 232 11 26  
Tel. 07745 - 9214 - 0  
Fax 052 - 232 70 67  
Fax 07745 - 9214 - 50  
Companies approved to ISO 9001 / EN 29001  
Internet: http://www.keller-druck.ch  

相关型号:

PA1103NL

HIGH FREQUENCY WIRE WOUND TRANSFORMERS EFD20 Platforms - SMT
PULSE

PA1110

1800-2000 MHz.Ultra Linear Power Amplifier
TE

PA1130NL

HIGH FREQUENCY WIRE WOUND TRANSFORMERS PoE Power Transformers - EP7, EP10 & EP13
PULSE

PA1131

850-900 MHz. Ultra Linear Amplifier
TE

PA1131NL

HIGH FREQUENCY WIRE WOUND TRANSFORMERS PoE Power Transformers - EP7, EP10 & EP13
PULSE

PA1132

1800-2000 MHz. Linear Power Amplifier
TE

PA1132NL

HIGH FREQUENCY WIRE WOUND TRANSFORMERS PoE Power Transformers - EP7, EP10 & EP13
PULSE

PA1133

1850-1910 MHz,Linear Power Amplifier
TE

PA1133NL

HIGH FREQUENCY WIRE WOUND TRANSFORMERS PoE Power Transformers - EP7, EP10 & EP13
PULSE

PA1133NLT

XFRMR 3.3V 7W 253UH POE EP10 SMD
PULSE

PA1134NL

HIGH FREQUENCY WIRE WOUND TRANSFORMERS PoE Power Transformers - EP7, EP10 & EP13
PULSE

PA1135NL

HIGH FREQUENCY WIRE WOUND TRANSFORMERS PoE Power Transformers - EP7, EP10 & EP13
PULSE