UFT-1A-3 [ETC]

UNIVERSAL FLOW TRANSMITTER; UNIVERSAL流量变送器
UFT-1A-3
型号: UFT-1A-3
厂家: ETC    ETC
描述:

UNIVERSAL FLOW TRANSMITTER
UNIVERSAL流量变送器

文件: 总2页 (文件大小:69K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
FLOW  
UNIVERSAL FLOW TRANSMITTER  
MODEL UFT-1  
DESCRIPTION  
The Model UFT-1 Universal Flow Transmitter is a  
solid-state, digital signal converter designed to operate  
with Data Industrial 200 Series flow sensors. Both ana-  
log (4-20 mA) and pulse outputs are available. The  
Model UFT-1 may be mounted in an optional NEMA 4X  
enclosure or with digital display of gpm or totalized flow  
(in a non-watertight enclosure).  
UFT-1  
FEATURES  
Analog and pulse outputs  
Optional watertight (NEMA 4X) enclosure  
Optional displays for flow rate and totalization  
Low cost  
Excitation voltage for flow sensors  
LED indication of pulse activity  
UFT-1E-1  
APPLICATION  
Analog Output (rate)  
DIMENSIONS  
The Model UFT-1 analog transmitter converts a Data  
Industrial digital flow signal into a precalibrated 4-20 mA  
signal. It must be calibrated for each Data Industrial flow  
sensor installation. The pipe type, size, and maximum  
flow rate must be specified at the time of order if 4-20  
mA output is to be used.  
in  
(cm)  
Pulse  
Out +  
Pulse Output (totalization)  
Pulse  
Out –  
The Model UFT-1 pulse output divides the Data  
Industrial digital flow signal by a jumper-selectable 10 or  
100 position to provide a more usable digital pulse. The  
pulse output is normally used where flow totalization is  
required. A simple conversion formula, using the flow  
factors for Data Industrial flow sensors on the next page,  
can convert the digital pulses to totalized gallons.  
mA Signal  
Out  
2.19  
(5.56)  
XDCR  
Signal In  
24 VDC  
Power  
Common  
The pulse output is an optoisolated transistor switch that  
can be wired to source or sink pulses to totalizer equip-  
ment.  
3.25  
(8.25)  
Note: This unit is not intended for field setup or  
field calibration.  
SPECIFICATIONS  
Supply voltage  
Operating temp  
Operating humidity  
Output  
24 VDC @ 60 mA  
Max loop resistance  
4-20 mA output 750@  
24 VDC  
40 VDC @ 200 mA  
±0.5%  
32° to 140°F (0° to 60°C)  
5% to 95% noncondensing  
4-20 mA and solid-state switch  
15-150 Hz full scale  
Output switch rating  
Accuracy  
Step response  
Input  
5 sec from 10% to 90%  
(dry or electronic contact)  
2004 KELE CATALOG • www.kele.com • USA 888-397-5353 • International 901-382-6084  
223  
FLOW  
UNIVERSAL FLOW TRANSMITTER  
MODEL UFT-1  
WIRING  
Pulse  
Output Pulse Divider  
Selection Jumper  
Typical  
Totalizer  
Hookup  
Totalize  
Out +  
+
Pulse  
Out  
Pulse  
Out –  
(10 or 100)  
100  
Power  
mA Signal  
Out  
10  
(red) +  
Flow  
Sensor  
XDCR  
Signal In  
+
4-20 mA  
Output  
24 VDC  
Power  
(black)  
XDCR  
Sig In  
Common  
PWR  
+
24 VDC  
Transducer Power Option Jumper  
(Leave on for Data Industrial flow sensors.  
Remove for devices with their own power supply.)  
Power Supply  
INSTALLATION AND CALIBRATION  
TABLE 1. FLOW FACTORS FOR DATA INDUSTRIAL  
FLOW SENSORS  
The UFT-1 transmitter can be mounted in any position. The  
optional watertight enclosure provides a watertight seal  
against contaminants. While field calibration is not required  
with the UFT-1, flow conversion must be accomplished at the  
monitoring computer. The information below is provided for  
making the conversion calculations.  
MODEL  
PIPE SIZE  
in (cm)  
FLOW  
FACTOR JUMPER IN  
10 POSITION 100 POSITION  
GALLONS/PULSE  
JUMPER IN  
228PV-1.5 1-1/2 (3.81)  
228PV-2  
228B-2  
0.03118  
0.04611  
0.04579  
0.04731  
0.005646  
0.007514  
0.007015  
0.01280  
0.01780  
0.03800  
0.07280  
0.1396  
0.2457  
0.3611  
0.6710  
1.080  
0.3118  
0.4611  
0.4579  
0.4731  
0.05646  
0.07514  
0.07015  
0.1280  
0.1780  
0.3800  
0.7280  
1.396  
3.118  
4.611  
4.579  
4.731  
0.5646  
0.7514  
0.7015  
1.280  
1.780  
3.800  
7.280  
13.96  
24.57  
36.11  
67.10  
108.0  
163.0  
194.4  
250.2  
315.8  
2 (5.08)  
2 (5.08)  
2 (5.08)  
1/2 (1.27)  
3/4 (1.91)  
1 (2.54)  
228C-2  
250B-0.5  
250B-0.75  
250B-1  
FLOW RATE  
250B-1.25 1-1/4 (3.18)  
Flow (gpm) = (mA measured - 4 mA) x Maximum gpm  
16  
Maximum gpm is the flow rate at 20 mA output on the trans-  
mitter and must be specified at the time the UFT-1 is ordered  
for proper calibration.  
250B-1.5  
220B-2.5  
220B  
220B  
220B  
220B  
220B  
220B  
220B  
1-1/2 (3.81)  
2-1/2 (6.35)  
3 (7.62)  
4 (10.16)  
5 (12.7)  
6 (15.24)  
8 (20.32)  
10 (25.40)  
12 (30.48)  
14 (35.56)  
16 (40.64)  
18 (45.72)  
2.457  
3.611  
6.710  
10.80  
TOTALIZED FLOW  
1.630  
16.30  
Totalized Gallons = (Flow Factor) x (Output Divider) x (Total  
Pulses)  
220B  
220B  
220B  
1.944  
19.44  
2.502  
25.02  
3
For totalized m , multiply the above by 0.00379.  
3.158  
31.58  
Notes  
1. Flow factors for a Model 225 and 226 are the same as Model 220.  
*2. Flow factor for Model 228S is the same as 228C.  
3. PV Series is sized for SCH 80 PVC pipe.  
Output Divider = 10 or 100 depending on jumper-selection.  
Flow factors per pulse are shown in the table at the right.  
All other series are sized for SCH 40 black iron pipe.  
ORDERING INFORMATION  
*
Pipe size and schedule and  
maximum flow rate must be  
specified at time of order.  
MODEL DESCRIPTION  
UFT-1  
Universal Flow Transmitter Pulse Output Only  
UFT-1A  
UFT-1E  
Universal Flow Transmitter with Pulse and Calibrated 4-20 mA Output*  
Universal Flow Transmitter Pulse Output in NEMA 4X Enclosure  
UFT-1AE Universal Flow Transmitter with Pulse and Calibrated 4-20 mA Output* in NEMA 4X Enclosure  
**  
When a UFT-1AE3 is ordered  
DISPLAY OPTION (enclosed models only)  
the UFT-1A will be in one enclosure  
and the Totalizer and Rate Display  
will be in a separate enclosure.  
1
2
3
Flow totalization only  
Flow rate only  
Flow totalization and flow rate **  
UFT-1  
1
Example: UFT-1A-E-2 Basic transmitter with calibrated 4-20 mA flow rate  
output (4 mA = no flow 20 mA = max flow), enclosed with LCD  
flow rate indication  
2004 KELE CATALOG • www.kele.com • USA 888-397-5353 • International 901-382-6084  
224  

相关型号:

UFT-1AE-1

UNIVERSAL FLOW TRANSMITTER
ETC

UFT-1AE-2

UNIVERSAL FLOW TRANSMITTER
ETC

UFT-1AE-3

UNIVERSAL FLOW TRANSMITTER
ETC

UFT-1E-1

UNIVERSAL FLOW TRANSMITTER
ETC

UFT-1E-2

UNIVERSAL FLOW TRANSMITTER
ETC

UFT-1E-3

UNIVERSAL FLOW TRANSMITTER
ETC

UFT100

ULTRA FAST RECOVERY MODULES
MICROSEMI

UFT100-200HL

Aluminum Electrolytic Capacitor, Polarized, Aluminum (wet), 200V, 75% +Tol, 10% -Tol, 100uF, Through Hole Mount, AXIAL LEADED
CDE

UFT1000

Rectifier Diode, 1 Phase, 1 Element, 10A, 50V V(RRM), Silicon, TO-220, PLASTIC, TO-220A, 2 PIN
WTE

UFT1000-50HP

Aluminum Electrolytic Capacitor, Polarized, Aluminum (wet), 50V, 75% +Tol, 10% -Tol, 1000uF, Through Hole Mount, AXIAL LEADED
CDE

UFT10000-16JT

Aluminum Electrolytic Capacitor, Polarized, Aluminum (wet), 16V, 75% +Tol, 10% -Tol, 10000uF, Through Hole Mount, AXIAL LEADED
CDE

UFT10005

100 Amp Supre Fast Recovery Rectifier 50 to 600 Volts
MCC