User's Guide  
Digital Psychrometer + IR Thermometer  
Models RH401 and RH405 (kit)  
C
DP  
Wb  
umi  
%
H
Introduction  
Congratulations on your purchase of the Extech RH401 Digital Psychrometer + InfraRed  
Thermometer. This device measures Humidity, Air Temperature, Dew Point and Wet Bulb.  
Advanced features include Infrared Non-Contact Surface Temperature Measurement, Data  
Hold and programmable Auto Power Off, and an RS-232 interface for capturing readings to  
a PC using optional 407752 software. The retractable probe protects the sensors when the  
meter is not in use. Careful use of this meter will provide years of reliable service.  
Meter Description  
13  
1. Emissivity adjust indicator  
2. Laser pointer on indicator  
3. Low Battery indicator  
4. IR Surface Temperature button  
1
2
3
12  
11  
10  
C
DP  
Wb  
5. ON/OFF/Setup button  
6. Back Light button  
umi  
H
%
7. Laser/Down button  
8. °F/°C/Up button  
15  
9. Dew Point/Wet Bulb button  
10. % Humidity display  
4
9
14  
5
6
11. Air Temperature, Dew Point or Wet Bulb display  
12. IR Surface Temperature display  
8
7
13. Extended probe with Humidity &  
Air Temperature Sensors  
14. PC interface port  
15. AC adaptor jack  
Note: The battery compartment is located on the rear of the instrument  
Warranty  
EXTECH INSTRUMENTS CORPORATION (A FLIR COMPANY) warrants this instrument to  
be free of defects in parts and workmanship for one year from date of shipment (a six  
month limited warranty applies to sensors and cables). If it should become necessary to  
return the instrument for service during or beyond the warranty period, contact the  
Customer Service Department at (781) 890-7440 ext. 210 for authorization or visit our  
website www.extech.com for contact information. A Return Authorization (RA) number  
must be issued before any product is returned to Extech. The sender is responsible for  
shipping charges, freight, insurance and proper packaging to prevent damage in transit.  
This warranty does not apply to defects resulting from action of the user such as misuse,  
improper wiring, operation outside of specification, improper maintenance or repair, or  
unauthorized modification. Extech specifically disclaims any implied warranties or  
merchantability or fitness for a specific purpose and will not be liable for any direct, indirect,  
incidental or consequential damages. Extech's total liability is limited to repair or  
replacement of the product. The warranty set forth above is inclusive and no other  
warranty, whether written or oral, is expressed or implied.  
2
RH401 V1.5 8/08  
Operation  
Retractable Humidity and Air Temperature Sensor  
Press Here  
Press down on the sensor, located at the top of the meter, to  
release it. Humidity and air temperature will not be displayed if the  
sensor is not extended.  
Sensor  
Extended  
After use, press the sensor into the stored (protected) position.  
Power ON/OFF  
Sensor  
Stored  
Press the  
(or the  
button) to turn power on. The  
meter will perform a short self-test when turned on.  
Press and hold the  
button for three seconds to turn power off.  
Humidity and Air Temperature measurements  
1. Hold the meter in the area to be tested.  
2. Allow adequate time for readings to stabilize.  
3. View the Relative Humidity measurement in the lower display.  
4. View the Air (probe)Temperature measurement in the middle display  
Wet Bulb and Dew Point measurements  
1. To display the Dew Point measurement, press and hold the  
symbol DP appears in the middle display.  
button until the  
2. Dew Point temperature is now displayed.  
3. To display the Wet Bulb measurement, press and hold the  
symbol Wb appears in the middle display .  
button until the  
4. Wet Bulb temperature is now displayed.  
5. To display IR temp –DP temp (IR surface temperature-Dew Point), press and hold the  
button for more than 2 seconds then press the IR Temp button. The reading  
can now be viewed in the top display.  
6. Press and hold the  
to IR temperature.  
button for more than 2 seconds to toggle the display back  
Non-Contact InfraRed Surface Temperature measurements  
1. Point the meter toward the surface to be measured.  
2. Press and hold the  
button.  
3. View the temperature measurement in the upper display  
4. Release the  
taken.  
nd the reading will be held on the display until a new reading is  
5. While holding down the  
button, press the  
button to toggle the laser  
targeting pointer on/off. The symbol will appear in the middle left of the display  
indicating that surface temperature measurement is active and the laser pointer is on.  
WARNING! The Laser button  
enables/disables the Laser pointer. When  
enabled, each press of the  
button turns on the Laser beam. Avoid looking directly  
into the path of the Laser or pointing the Laser toward anyone’s eyes. Mirrored surfaces  
near a measurement object can redirect the Laser, use extreme caution. Do not allow  
the Laser beam to be directed toward explosive gases.  
3
RH401 V1.5 8/08  
IR Emissivity Adjustment  
Emissivity is adjustable from 0.3 to 0.99. The default value is 0.95 which is the typical  
emissivity value of many non-metallic surfaces. If the surface to be measured is highly  
reflective and the emissivity is not known, cover the surface with paint or tape to improve  
measurement accuracy. If the emissivity of the surface is known, follow this procedure to  
set the value.  
1. With the meter on, Press the  
key twice to enter the Emissivity setting procedure.  
will flash on the top left side of the display.  
2. Use the  
up key and the  
down key to adjust the value.  
3. When the desired setting appears in the display, press  
back to normal operation.  
again to save and return  
Emissivity Factors for Common Materials  
Material under test  
Asphalt  
Concrete  
Cement  
Sand  
Emissivity  
Material under test  
Emissivity  
0.98  
0.90 to 0.98 Cloth (black)  
0.94  
0.96  
0.90  
Skin (human)  
Lather  
0.98  
0.75 to 0.80  
0.96  
Charcoal (powder)  
Soil  
0.92 to 0.96 Lacquer  
0.80 to 0.95  
0.97  
Water  
0.92 to 0.96 Lacquer (matt)  
0.96 to 0.98 Rubber (black)  
Ice  
0.94  
Snow  
0.83  
Plastic  
0.85 to 0.95  
0.90  
Glass  
0.90 to 0.95 Timber  
0.90 to 0.94 Paper  
Ceramic  
Marble  
Plaster  
Mortar  
0.70 to 0.94  
0.81  
0.94  
Chromium Oxides  
0.80 to 0.90 Copper Oxides  
0.89 to 0.91 Iron Oxides  
0.93 to 0.96 Textiles  
0.78  
0.78 to 0.82  
0.90  
Brick  
Spot to Distance Ratio  
The 8:1 spot to distance ratio determines the size of the measured surface area with  
respect to the distance the meter is held away from the surface.  
4
RH401 V1.5 8/08  
Selecting temperature units of measure (C/F)  
Press and hold the oC/oF button momentarily to toggle the temperature units.  
Automatic Power OFF  
The meter automatically shuts off after a programmed period of time.  
To disable auto power off, with the meter off, press  
release the keys and the meter is now in Non-Sleep Mode.  
The power off time is adjustable from 5 to 600 seconds. To change the auto power off  
setting, press the key to enter the setting procedure. Use the up key and the  
down key to set the value. When the desired power-off time appears in the display,  
press the key again to save the parameters.  
and  
. When “n” appears,  
Backlight  
Press the BKLT button to turn the display backlight on. The backlight will remain on until  
the BKLT button is pressed again or the meter is turned off.  
Error Messages  
An error message will appear on the display if the meter fails an internal diagnostic test.  
1. E1 (Top display): IR temperature failure. Repair/replacement is necessary.  
2. E2 (Top display): IR temperature >932°F (500°C).  
3. E3 (Top display): IR temperature <-40°F (-40°C).  
4. E9 (Top display): Calibration error.  
5. E1 (Middle display): Temperature failure. Repair/replacement is necessary.  
6. E2 (Middle display): Dew Point is out of range or Repair/replacement is necessary.  
7. E3 (Middle display): Air temperature is out of range or Repair/replacement is  
necessary.  
8. E1 (Bottom display): RH circuit error. Repair/replacement is necessary.  
9. E5 (Bottom display): RH circuit error. Repair/replacement is necessary.  
Setup memory  
The selected display, emissivity value, auto power off time, backlight status and the laser  
pointer status are stored in memory and will remain in memory as the user default setting  
until reprogrammed by the user.  
5
RH401 V1.5 8/08  
Calibration  
The following verification and calibration procedures requires a humidity chamber or the  
33% and 75% RH reference bottles which are supplied with the Model RH405 kit.  
RH Accuracy Verification  
33% or 75%  
Checking the 33% or 75% RH Calibration  
Cal Bottle  
1. Insert meter’s sensor into the 33% or 75% salt reference  
bottle.  
Rotating the bottle will assist in placing it over the sensor.  
2. Check the reading after 10 minutes  
3. Verify that the reading is within the accuracy specification.  
RH Calibration (33% and 75%)  
This two point calibration takes approximately one hour to  
complete.  
33% Calibration  
1. Turn off the meter and insert the meter sensor into the 33%  
salt reference bottle. Rotating the bottle will assist in  
placing it over the sensor.  
2. Hold down the  
and  
keys and then press the  
key to turn the unit on.  
3. “32.8” (reading may vary depending on temperature) will  
flash on the display. After 30 minutes, the flashing will stop  
to indicate that the 33% calibration is complete.  
75% Calibration  
4. Insert the meter sensor into the 75% salt reference bottle. From a steady “32.8”  
display press the  
button to enter the 75% calibration.  
5. “75.2” (reading may vary depending on temperature) will flash on the display. After  
30 minutes, the flashing will stop to indicate that the 75% calibration is complete  
and the calibration data has been saved to memory.  
6. The screen will return to the normal display. Calibration is now complete.  
Note: If the salt at the bottom of the calibration bottles appears dry, the bottles should be  
replaced.  
6
RH401 V1.5 8/08  
RS-232 PC Communications Interface  
The meter is equipped with an RS-232 PC interface jack (3.5mm phono) for connection to  
a PC. The cable to connect the meter to the PC is included in the optional 407752 data  
acquisition kit. The kit includes WindowsTM compatible software that allows the user to  
store readings in a text file and display real-time measurements in a series of selectable  
formats. For more information or specific operating instructions, refer to the User Guide  
included with the 407752 kit or contact Extech Instruments.  
Specifications  
Function  
Range and Resolution  
Accuracy  
Humidity  
0.0 to 100.0% RH  
±3% RH (10 to 90%)  
Temperature (Air)  
-4.0 to 122.0oF (-20 to 50oC) ±1.8oF (±1oC)  
±3% reading or ±3oC/6oF  
whichever is greater  
-4 to 842oF (-20 to 450oC)  
Temperature (InfraRed)  
-40 to -4 oF, 842 to 932oF  
-40 to -20 oC, 450 to 500 oC  
Triple LCD with backlight  
±4% reading or ±4oC/8oF  
whichever is greater  
(@25ºC ±5ºC)  
Display  
Sensor Type  
Dew Point  
Humidity: Capacitance sensor; Temperature (air): Thermistor  
-90.4 to 122.0oF (-68 to 50oC) (calculated from humidity and  
temperature measurements)  
Wet Bulb  
-6.9 to 122.0oF (-21.6 to 50oC) (calculated from humidity and  
temperature measurements)  
IR Emissivity  
Adjustable from 0.3 to 0.99  
IR Spot/Distance Ratio 8:1  
IR Response Time  
Operating Conditions  
Storage Conditions  
Power Supply  
0.5 seconds  
-4 to 122oF (-20 to 50oC); < 99% RH non-condensing  
-40 to 185oF (-40 to 85oC); <99% RH non-condensing  
4 x 1.5V ‘AAA’ batteries or AC adaptor  
Battery Life  
approx. 150 hours (laser inactive); 30 hours (laser active)  
6.8x2.7x1.9” (175x70x50mm); 4.9 oz. (140g)  
Dimensions / Weight  
7
RH401 V1.5 8/08  
Maintenance  
Cleaning and storage  
1. The meter should be cleaned with a damp cloth and mild detergent when necessary.  
Do not use solvents or abrasives.  
2. Store the meter in an area with moderate temperature and humidity (refer to the  
operating and storage range in the specifications chart earlier in this manual).  
Battery Replacement  
When the battery power falls low, the symbol will appear on the lower left of the LCD.  
Replace the four (4) 1.5 ‘AAA’ batteries by removing the rear battery compartment cover  
and accessing the battery compartment. Observe polarity when placing the batteries in the  
compartment. Ensure that the compartment cover is securely fastened when finished.  
Calibration and Repair Services  
Extech offers repair and calibration services for the products we sell. Extech also provides  
NIST certification for most products. Call the Customer Service Department for information  
on calibration services available for this product. Extech recommends that annual  
calibrations be performed to verify meter performance and accuracy.  
Support line (781) 890-7440  
Technical support: Extension 200; E-mail: support@extech.com  
Repair & Returns: Extension 210; E-mail: repair@extech.com  
Product specifications subject to change without notice  
For the latest version of this User’s Guide, Software updates, and other up-to-the-  
minute product information, visit our website: www.extech.com  
Copyright © 2008 Extech Instruments Corporation (a FLIR company)  
All rights reserved including the right of reproduction in whole or in part in any form.  
8
RH401 V1.5 8/08  

相关型号:

RH405

Digital Psychrometer + IR Thermometer
EXTECH

RH40M-4/105553

MOTORIZED IMPELLER
EBMPAPST

RH40M-4/203473

MOTORIZED IMPELLER
EBMPAPST

RH40M-6/111192

MOTORIZED IMPELLER
EBMPAPST

RH40M-6/206522

MOTORIZED IMPELLER
EBMPAPST

RH40M-6206522

MOTORIZED IMPELLER
EBMPAPST

RH40N25FSY1

N-channel 250V - 0.084ohm - TO-254AA Rad-hard low gate charge STripFET TM Power MOSFET
STMICROELECTR

RH40N25FSY3

N-channel 250V - 0.084ohm - TO-254AA Rad-hard low gate charge STripFET TM Power MOSFET
STMICROELECTR

RH41

SMPS Capacitors (RH Style)
KYOCERA AVX

RH411C125KA30A3

Stacked MLCC (RH)
KYOCERA AVX

RH411C125MA30A3

Stacked MLCC (RH)
KYOCERA AVX

RH411C155KA30A3

Stacked MLCC (RH)
KYOCERA AVX