EDS3-61H [FUJI]

THERMAL CONDUCTIVITY GAS ANALYZER THERMOMAT; 热导式气体分析仪THERMOMAT
EDS3-61H
型号: EDS3-61H
厂家: FUJI ELECTRIC    FUJI ELECTRIC
描述:

THERMAL CONDUCTIVITY GAS ANALYZER THERMOMAT
热导式气体分析仪THERMOMAT

文件: 总4页 (文件大小:72K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
THERMAL CONDUCTIVITY GAS ANALYZER THERMOMAT  
DATA SHEET  
ZAF  
In modern industries, it is indispensable to analyze  
components and measure concentrations of mixture gases  
and combustion gases for strict quality control and to file data  
for establishing production schedules which will allow proper  
management of industrial processes.  
Of the various analytical methods and measuring prin-  
ciples available for analyzing process gases, this gas analyzer  
utilizes the fact that electric resistance of a heated platinum  
wire changes depending on kind and concentration of gas.  
Relative thermal conductivity  
(at 0°C)expressed taking that  
Kind of Gas  
of air(5.572 × 10-5cal/cm.sec.  
deg.)as 1  
Sulfurdioxide SO  
2
Carbondioxide CO  
2
Argon Ar  
Carbon monoxide CO  
Water vapor(100°C)H2O  
Air  
Nitrogen N  
Oxygen O  
Methane CH  
2
2
4
Measuring mode:  
Hydrogen H  
2
1 range or 2 ranges (for measuring the same  
sample gas components with the same  
measuring system)  
Measuring range:  
CO2  
FEATURES  
0 to 10 ....100  
100 to 90 ....  
0 to 3 ....100  
100 to 95 ....  
0 to 10 ....100  
100 to 90 ....  
0 to 20 ....100  
100 to 80 ....  
0 to 5 ....100  
100 to 90 ....  
Output signal: DC 4 to 20 mA  
(load resistance less  
0
1. Continuous measurement possible.  
2. Quick response.  
3. Integral unit incorporating detector, current regulator and  
H2  
0
Ar  
indicator.  
0
4. Simple and robust mechanism assuring high maintainabil-  
CH4  
He  
ity.  
0
5. Minimum influence due to external and operating condi-  
tions.  
6. Stable accuracy.  
0
than 400)  
Simultaneously  
available  
SPECIFICATIONS  
DC 0 to 10mV  
}
(output resistance 100)  
Measuring principle:  
Measurement of thermal conductivity  
Measuring component:  
0 to 100µA (load resistance 100) is  
[
]
further available for special specification.  
Repeatability:  
Two-component gases (from viewpoint of  
thermal conductivity)  
such as CO2, H2, Ar, CH4, He, SF6  
±1% of F.S.  
Within zero point ±2.5% of full scale/week  
Within span ±2% of full scale/week  
Drift:  
EDS3-61h  
Date Jun. 27, 2002  
ZAF  
Response speed:  
CODE SYMBOLS  
Within 30 sec for 90% response (for 20%  
CO2)  
Within 60 sec for 90% response (for 0 to  
10% H2)  
(varies with kind of gas and range)  
Description  
Z A F  
P
2 -  
Reference  
Sample  
gas  
CO2  
CO2  
H2  
Measuring  
range  
gas  
N2  
Sample gas flow rate:  
3A  
4A  
4B  
4C  
4D  
4E  
4F  
4Z  
6A  
6B  
6C  
6D  
6E  
6F  
6Z  
0 to 20%  
Others than  
0 to20%  
0.4 ±0.05 /min, constant  
Standard requirements for sample gas:  
Temperature 0 to 50°C  
N2  
Ar  
Dust  
0.3µm or less  
Pressure  
10kPa or less  
CH4  
He  
As specified  
by customer  
(See the table  
below)  
(flow rate 0.4L/min)  
Materials of gas-contacting parts:  
Lead, stainless steel JIS SUS316, platinum,  
platinum iridium, epoxy resin, silver,  
fluororubber  
Power supply: AC 100V ±10%, 50/60 Hz  
110V, 200V, or 220V usable for special  
specification  
Power consumption:  
SF6  
Others  
CO2  
H2  
Others than  
N2 gas  
As specified  
by customer  
(See the table  
below)  
Ar  
CH4  
He  
Approx. 10VA  
Indoor panel plug-in  
SF6  
Mounting:  
Others  
Power supply  
1
2
3
4
AC 100V 50/60Hz  
AC 110V 50/60Hz  
AC 200V 50/60Hz  
AC 220V 50/60Hz  
Measuring  
α
α = 90°  
Warmup time: At least 30 min.  
Ambient temperature:  
-5 to 45°C  
Ambient humidity:  
Less than 90% RH  
1
2
1 range  
2 ranges (available only for measuring  
the same sample gas)  
Weight:  
Approx. 5kg  
(Note 1) Reference gas components are as listed in the table below.  
(Note 2) Sample gas, herby, means a gas components to be measured,  
and reference gas stands for each component other than the  
sample gas (remaining gas component).  
External dimensions (H x W x D):  
240x197x 212 mm  
Finish color:  
Munsell 7.5BG 3.2/0.8  
STANDARD MEASURING  
MEASURING PRINCIPLE  
Compari-  
Measur- son gas  
ing gas compo-  
nent  
Output  
characteristic  
Measuring range Range ratio  
N2, O2,  
0-10, 20, 50, 100%  
0-10, 20% : Linear  
Others : non-linear  
Sample gas  
1 : max. 5  
Air, He  
100-90%, 100-80%  
Impossible  
CO2  
Ar  
N2, CO2, 0-3, 5, 10, 20, 50,  
100-90% : Linear  
Others : non-linear  
Ar  
80, 100%  
1 : max. 10  
Enclosed  
gas inlet  
H2  
Ar  
100-90%, 100-80%  
Measuring  
chamber  
Air, O2  
Impossible  
0-10, 20, 50, 80,  
100%  
100-90%, 100-80%  
0-10, 20%,  
100-90% : Linear  
Others : non-linear  
N2, O2,  
Air, He  
Reference  
chamber  
1 : max. 5  
CO2  
Impossible  
0-20, 40, 50, 60,  
80, 100%  
100-80%  
N2, CO2,  
Ar, He  
1 : max. 5 Non-linear  
CH4  
He  
V-I  
converter  
Air, O2  
Impossible  
N2, CO2, 0-5, 10, 20, 30, 40,  
Ar, O2,  
Air  
V/I  
100-90% : Linear  
Others : non-linear  
50, 80, 100%  
100-90%, 100-80%  
1 : max. 10  
DC 4 to 20mA  
DC 0 to 10mV  
Fixed  
resistor  
(Note 1) Linear output is ±2.5% FS or less.  
(Note 2) On 2-range type, the zero point is used in common.  
(Note 3) When sample gas consists of more than 3 components, contact  
our office with a table of gas components submitted for confirma-  
tion of whether or not applicable analyzers are available.  
DC  
amplifier  
or 0 to 100  
A
Fixed  
resistor  
2
GAS SAMPLING SYSTEM DIAGRAM (EXAMPLE)  
Flow  
checker  
Membrane  
filter  
Pump  
ZAF  
Atmosphere  
Note 1  
0.4L/min  
Peltier gas  
Note 2  
cooler  
Note 2  
Drain pot  
(Note 1) Dust must be purged adequately (for protection of the pump, flowmeter, etc.).  
(Note 2) Unnecessary when sample gas is dry.  
CONNECTION DIAGRAM  
SCOPE OF DELIVERY  
Analyzer main unit  
Panel mounting brackets  
Fuse 2 pieces  
1
2
3
4
5
6
7
8
9
10  
Items to be prepared separately  
POWER  
G
AC 100V 50/60Hz  
OR  
4 to 20mA  
Sampling equipment  
Receiving instrument  
0 to 10mA  
OR  
AC 200, 110, 220V  
0 to 100  
A
3
ZAF  
OUTLINE DIAGRAM (Unit:mm)  
2
t
10  
Panel  
140  
Indicator  
t
GAS  
INLET  
THERMOMAT  
8
POWER  
RANGE  
GAS  
OUTLET  
ZERO SPAN  
19  
169  
24.2  
192  
197  
Terminal  
cover  
188  
Terminal  
board  
Rear View  
46  
Panel Cutout Dimensions  
184+2  
0
INLET  
OUTLET  
Specification  
card  
Tapping Rc1/4  
182  
Asterisked * items: Non-standard  
Caution on Safety  
*Before using this product, be sure to read its instruction manual in advance.  
Head Office  
6-17, Sanbancho, Chiyoda-ku, Tokyo 102-0075, Japan  
http://www.fesys.co.jp/eng  
Sales Div.  
International Sales Dept.  
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan  
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187  
http: //www.fic-net.jp/eng  
Printed in Japan  
Information in this catalog is subject to change without notice.  

相关型号:

EDS3-62H

PARAMAGNETIC OXYGEN ANALYZER
FUJI

EDS3-69F

PARAMAGNETIC OXYGEN ANALYZER
FUJI

EDS5-101A

FC SERIES PRESSURE TRANSMITTER
FUJI

EDS5-102B

FC SERIES ABSOLUTE PRESSURE TRANSMITTER
FUJI

EDS5-103A

FC SERIES PRESSURE TRANSMITTER WITH REMOTE SEALING
FUJI

EDS5-14D

FCX SERIES ABSOLUTE PRESSURE TRANSMITTER
FUJI

EDS5-15F

FCX SERIES REMOTE SEAL TYPE PRESSURE TRANSMITTER
FUJI

EDS5-16D

FCX SERIES PRESSURE TRANSMITTER
FUJI

EDS5-54A

OPTICAL GAUGE PRESSURE TRANSMITTER
FUJI

EDS5-55A

OPTICAL GAUGE PRESSURE TRANSMITTER WITH REMOTE SEAL DIAPHRAGM
FUJI

EDS5-56A

OPTICAL ABSOLUTE PRESSURE TRANSMITTER
FUJI

EDS5-61B

FCX-A SERIES ABSOLUTE PRESSURE TRANSMITTER
FUJI