EDS6-141D [FUJI]
ULTRASONIC FLOWMETER(2-Path System type); 超声波流量计( 2路系统型)型号: | EDS6-141D |
厂家: | FUJI ELECTRIC |
描述: | ULTRASONIC FLOWMETER(2-Path System type) |
文件: | 总12页 (文件大小:1019K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
ULTRASONIC FLOWMETER
(2-Path System type)
DATA SHEET
FSH, FSG, FSD
This flowmeter is a fixed type ultrasonic flowmeter based
on transit-time measuring method.
The flowmeter is a compact and light-weight instrument
incorporating the latest electronics and high speed digital
signal processing technologies, realizing high performance
and easy operation. This high performance flowmeter is
used for 2-path system only.
FEATURES
1. Compact and light-weight
Flow transmitter
(FSH)
Thanks to the adoption of the latest electronics the
flow transmitter size and mass are 1/2 of our traditional
instrument.
2. Full variety of sensors
The flowmeter can be used with various types of sensors
applicable for wide range of pipe size (ø13 to ø6000mm)
and fluid temperature (-40 to +200°C).
3. High accuracy
Detector
(FSD)
The flowmeter is designed for high accuracy (± 1.0%).
The adoption of new sound velocity measurement
system permits measurements of fluids of unknown
sound velocity, and also slightly affection from fluid
temperature and pressure.
4. Excellent resistance against aerated flow
Fuji's unique ABM feature improves measurement reli-
ability for different flow like slurries, sludge, raw sewage
and bubble-contained flow (acceptable up to air bubble
of 12% volume at 1m/s velocity).
Detector
(FSG)
5. Quick response
With the use of high-speed micro-processor suited for digi-
tal signal processing, the fast response time is realized.
6. Multi-lingual
The following languages are supported for display:
Japanese (Katakana), English, German French, and
Spanish.
7. Excellent performance and easy operation
LCD and function keys are allowing easy configuration
and trouble shooting.
Applicable fluid:Homogenous liquid where the ultrasonic
signal can be transmitted
Bubble quantity: 0 to 12vol% (for pipe size
50A, water, velocity 1m/s)
Fluid turbidity: 10000mg/L max.
Type of flow: Fully-developed turbulent or
laminar flow in a full-filled pipe
Applicable flow pipe:
Material:
Plastics (PVC, FRP, etc.) or
Metals (carbon steel, SS, cop-
per, aluminum, etc.)
− LCD with back light
− Easy mounting of sensor
− Trouble shooting
Pipe size:
50 to 6000 mm
(inside diameter)
Liner:
Tar epoxy, mortar, etc.
Straight pipe length: Typically 10D for upstream
and 5D for downstream. Refer
SPECIFICATIONS
Operational specifications
to JEMIS-032 in detail.
System configuration:
Velocity:
Power supply:
0 to ±0.3 ········±32 m/s
The system is composed of two detectors
(Model: FSG, FSD) and one Flow transmitter
(Model: FSH), realizing two-path measure-
ment.
100 to 240 Vac+10%/-15%, 50/60Hz or 20 to
30 Vdc
EDS6-141d
Date
Apr. 1, 2011
FSH, FSG, FSD
Signal cable: High-frequency coaxial cable and 3-wire
shielded cable.
Display device:
Graphic LCD (number of pixels: 240x64) with
back light
Display language:
Japanese, English, French, German or Span-
ish selectable
Velocity/Flow rate display:
Maximum cable length: 150m
Allowable temperature limit: 80ºC
Coaxial cable (4 pcs)
Environment: Non-explosive environment without direct
sunlight, corrosive gas and heat radiation
Ambient temperature:
-10 to +50ºC for flow transmitter,
-20 to +80ºC for detector
Ambient humidity:
Display of velocity and/or flow rate with flow
direction
Data: up to 10 digits (decimal point to be
counted as 1 digit)
95%RH or less for flow transmitter, 100%RH
or less for detector
Grounding: Class D (less than 100 ohm)
Unit: Metric/English system selectable
Metric system
English system
ft/s
Velocity
m/s
Arrester:
Surge absorbers for outputs and power sup-
ply incorporated as standard
Flow rate
L/s, L/min, L/h, L/d, kL/h,
ML/d, m3/s, m3/min, m3/h,
m3/d, km3/d, Mm3/d,
BBL/s, BBL/min, BBL/h,
kBBL/d, MBBL/d
ft3/s, ft3/min, ft3/h, Mft3/d,
gal/s, gal/min, gal/h, Mgal/d,
BBL/s, BBL/min, BBL/h,
BBL/d, kBBL/d, MBBL/d
Performance specifications
Accuracy :
Note: ˝gal˝ means US gal.
Detector
Pipe size
Applicable pipe
material
Flow velocity Accuracy
(diameter)
Total display: Display of forward or reverse total
Data: up to 10 digits (decimal point to be
counted as 1 digit)
2 to 32m/s
0 to 2m/s
±1.0% of rate
±0.02m/s
ø50 to below ø300
FSGS32
FSD32
FSGS51
0.75 to 32m/s ±1.0% of rate
0 to 0.75m/s ±0.0075m/s
ø300 to ø6000 mm
ø50 to below ø300
ø300 to ø6000 mm
Unit: Metric/English system selectable
Plastic, metal pipe
2 to 32m/s
0 to 2m/s
±1.5% of rate
±0.03m/s
Metric system
English system
FSGS31
FSGS50
ft3, kft3, Mft3, gal, kgal,
mL, L, m3, km3, Mm3,
Total
0.75 to 32m/s ±1.5% of rate
0 to 0.75m/s ±0.0113m/s
mBBL, BBL, kBBL, ACRf
mBBL, BBL, kBBL
Configuration:
Response time:
Transit time method:
Power consumption:
Fully configurable on keyboard by menu-
driven software
0.5sec
Zero adjustment:
Set zero/Clear available.
20W or less
Damping:
0 to 100s (every 0.1s) configurable for analog
output and display
Functional specifications
Analog output: 4 to 20 mAdc (1 point)
Max. load resistance: 1k ohm
Digital output: +total, -total, alarm, acting range, flow
switch or total switch -- arbitrarily selectable
Mechanical relay contact:
Low flow cut off:
0 to 5 m/s configurable
Alarm:
Hardware fault/process fault can be tied to
digital output
Analog output: Hold/Upper limit/Lower limit/
Zero/Not-used selectable
Total: Hold/Count selectable
Timer: 0 to 900s (every 1s) configurable
Burnout:
1 point with socket (replaceable)
Normally closed/open selectable
Capacity:240 Vac/30 Vdc, 1 A
Bi-directional range:
Forward and reverse ranges configurable in-
Total pulse: less than 1 p/s
(Pulse width: 50, 100 or 200 ms
selectable)
Transistor open collector: 2 points
Capacity: 30 Vdc, 0.1 A
dependently
Hysteresis: 0 to 20% of working range con-
figurable
Working range applicable to digital output
Normally off/on selectable
Auto-2 ranges:
Total pulse: less than 1000 p/s
(Pulse width: 0.5, 1, 2, 5, 10, 20,
50, 100 or 200 ms selectable)
Forward 2 ranges configurable independently
Hysteresis: 0 to 20% of working range con-
figurable
Communication interface:
Working range applicable to digital output
RS-232C equivalent / RS-485 (selectable)
Number of connectable units: one (RS-232C)/
up to 31 (RS-485)
Baud rate: 9600/19200/38400 bps selectable
Parity: none/odd/even selectable
Stop bit: 1 or 2 bits selectable
Distance: up to 15 m (RS-232C)/up to 1k m
(RS-485)
Flow switch: Lower and upper switching points configu-
rable independently
Acting point applicable to digital output
Total switch: +total switching point configurable
Acting point applicable to digital output
Data: velocity, flow rate, +total, -total, status
(standard), velocity profile (option)
2
Signal cable:
FLY8, FLY9 (applicable detector: FSG,
FSD)
•Structure: High frequency coaxial cable
(double shield)
•Sheath: Black flame-resisting PVC
•Outer diameter: ø7.5mm
•Termination: M3.5 amp terminal (flow
transmitter side) and M4 amp terminal
(FLY8).Note, however, that the detecter
side of FSD32 is provided with BNC
connector (FLY9).
•Mass: Approx. 90g/m
Flow transmitter FSH:
H240×W247×D134mm
Detector: H46×W410×D50mm (FSGS3)
H67×W78×D84mm (FSGS5)
Physical specifications
Enclosure protection:
Flow Transmitter: IP67,
Detector:
FSG (common type):
IP67 (Silicon compound is filled on the
terminal part when wiring)
FSG (submersible type):
IP68 (submersible in water for 5 days)
FSD (high temperature type): IP52
Flow Transmitter: wall mount
Detector: clamped on pipe surface
Mounting:
Dimensions:
Acoustic coupler: Silicon compound (RTV)
Material:
Flow Transmitter: aluminum alloy
Detector:
Detector Sensor housing Sensor cover Guide rail
H205×W530×D52mm (FSD32)
FSGS3
FSGS5
FSD32
PBT
SUS304
SUS304
–
SUS304 + plastic
–
PBT
Mass:
Flow transmitter : 5kg
Detector: 0.6kg (FSGS3)
1.2kg (FSGS5)
SUS304
SUS304 + aluminum alloy
1.6kg (FSD32)
Conditions on straight pipe
( D : Inside diameter of pipe)
Classification
Upstream side
Downstream side
5D
L
L
10D
90 bend
Detector
L
L
10D
50D
Tee
10D or more
L
5D
L
30D
Diffuser
1.5D
L
5D
L
10D
Reducer
L
30D
L
10D
Various Valve
In case that flow control valve exists on
upstream side.
In case that flow control valve exists on
downstream side.
Stop valve
Check valve
L
50D
Pump
P
(Note) The source : JEMIS-032
3
FSH, FSG, FSD
MEASUREMENT PRINCIPLE
MOUNTING OF DETECTOR
Ultrasonic pulses are propagated aslant between the up-
stream and downstream sensors, detecting the time differ-
ence due to flow for measurement.
Plastic wedge
Pipe
Ultrasonic oscillator
Upstream sensor
Lining
Output signal
Flow
transmitter
D
t2
t1
Velocity V
Downstream sensor
CONFIGURATION DIAGRAM
(1) 2-path system (V method)
Cable
(2) 2-path system (Z method)
Cable
Flow
transmitter
Flow
Detector
Detector
transmitter
4-20mA DC
4-20mA DC
Contact (up to 3 points)
Contact (up to 3 points)
DETECTOR SELECTION GUIDE
Inner diameter of piping ø (mm)
Type
Fluid
Mounting
13 25
50
100
200
250
300
400
1000
3000
6000
temperature [°C] method
FSGS32 Note)
FSGS31
300
V
50
Px, P, M
–40 to 80
V
200
200
3000
Px, P, M
FSGS51 Note)
FSGS50
6000
Z
Px, P, M
50
250
Px, P, M
V
FSD32
–40 to 200
Z
150
Px, P, M
400
Px : PP, PVDF
: Plastic (PVC, etc.)
M : Msetallic piping (steel pipe, copper pipe, aluminum, etc.)
Classification of
piping materials
P
Note: The ultrasonic signal cannot be transmitted easily when the classification of piping material is Px or the turbidity is high.
In such a case, a preliminary check by a portable ultrasonic flowmeter is recommended.
4
<Detector, submersible type>
CODE SYMBOLS
1
2
3
4
5
6
7
8
9
10 11 12 13
Description
F S G S
A 1
1
<Flow transmitter>
Type (5th and 6th digits)
1
2
3
4
5
6
7
8
9 10 11 12
3 2
3 1
5 1
5 0
Small sensor 2MHz (ø50 to ø300)
Small sensor 1MHz (ø50 to ø300)*2
Large sensor 1MHz (ø200 to ø6000)
Large sensor 0.5KHz (ø200 to ø6000)*2
V method
F S H Y A
1
S Y Y
Description
Type (4th digit)
V or Z
method
E
Standard
Use (6th digit)
Dedicated signal cable (9th digit)
Transit time measurement (two path)
B
C
D
E
10m
20m
30m
40m
50m
60m
70m
80m
A
Power supply (7th digit)
100 to 240 Vac, 50/60 Hz
20 to 30 Vdc
1
4
F
Conduit connection (10th digit)
G1/2 and G3/8 (female screw) with
water-proof connection
G
H
J
Y
K
L
M
N
P
Q
R
Z
90m
Parameter setting, Tag Plate (12th digit)
None
With setting
With setting and Tag Plate
With Tag Plate
100m
110m
120m
130m
140m
150m
Specified length
Y
A
B
C
(Contact us if length is more than 150m.
Max. length is 300m.)
<Detector, high temperature type>
Acoustic coupler (10th digit)
1
2
3
4
5
6
7
8
A
C
Silicon rubber (KE348)
5
F S D
1
Description
Silicone grease (G40M)
*
Additional specification (11th digit)
None
Tag plate
Type (4th, 5th and 6th digits)
High-temperature sensor *1
(ø50 to ø400), V or Z method
Y
A
3 2 0
Wire rope for mounting (12th digit)
Specify it in the case of FSGS41 or FSGS5.
None
Nominal diameter: up to ø500mm
Nominal diameter: up to ø1000mm
Nominal diameter: up to ø1500mm
Belt, acoustic coupler (7th digit) *3
For fixed type
Y
Y
A
B
C
D
E
<
1
>
Detector, common type
Nominal diameter: up to ø3000mm
Nominal diameter: up to ø6000mm
Can be specified
only for FSGS5
2
3
4
5
6
7
8
9
10 11 12 13
Description
F S G S
Y 1
Y
1
Type (5th and 6th digits)
3 2
3 1
5 1
5 0
Small sensor 2MHz (ø50 to ø300)
Small sensor 1MHz (ø50 to ø300)*2
Large sensor 1MHz (ø200 to ø6000)
Large sensor 0.5KHz (ø200 to ø6000)*2
V method
*1: For turbid fluid or old pipe, cast iron pipe, mortar lining pipe
or others through which the ultrasonic signal could not be
transmitted easily, use an optional guide rail (TK4C6164C1),
and carry out mounting by Z method.
V or Z
method
Acoustic coupler (10th digit)
4
Applicable diameter range
V method: ø50 to ø250 Z method: ø150 to ø400
Y
A
B
C
None
*
Silicon rubber (KE348W)
Silicone-free grease (HIGH-Z)
5
*
*2: For aging pipes, cast iron pipes or mortar-lined pipes that
interrupts the propagation of ultrasonic signals, select
FSGS31 or FSGS50.
5
Silicone grease (G40M)
*
Additional specification (11th digit)
None
Tag plate
Y
A
*3: As for the acoustic coupler, grease (KS62M) for high
temperatures is attached.
Wire rope for mounting (12th digit)
Specify it in the case of FSGS41 or FSGS5.
None
*4: Silicone rubber (KE-348W) is provided as a standard
accessory to fill the wiring mold. (It can also be used as an
acoustic coupler.)
Y
A
B
C
D
E
Nominal diameter: up to ø500mm
Nominal diameter: up to ø1000mm
Nominal diameter: up to ø1500mm
Nominal diameter: up to ø3000mm
Nominal diameter: up to ø6000mm
If an additional acoustic coupler is required, select one
among A, B and C.
Can be specified
only for FSGS5
*5: Normally select silicone rubber as acoustic coupler.
Siliconerubber in tube (100g) is furnished. If you place an
order for several units, 1 tube may suffice for every 2 units.
Select silicone-free grease for semiconductor anufacturing
equipment or the like that is vulnerable to silicone.
The silicone-free grease is water-soluble and, therefore,
cannot be used in environment exposed to water or on
piping subjected to a condensation. Since the grease does
not set, a periodic maintenance (cleaning, refilling every
about 6 months at normal temperature) is necessary.
5
FSH, FSG, FSD
<Signal cable>
Note: Must be procured unless the sensor is a submersible type.
• For detector FSG and FSD
1
2
3
4
5
6
7
8
F L Y
1
Description
Type of sensor (4th digit)
8
9
Small and large sensor (for FSG)
Hight temp sensor (for FSD)
Cable length (5,6 and 7th digit)
0 0 5
0 1 0
0 1 5
0 2 0
0 2 5
0 3 0
0 3 5
0 4 0
0 4 5
0 5 0
0 5 5
0 6 0
0 6 5
0 7 0
0 7 5
0 8 0
0 8 5
0 9 0
0 9 5
1 0 0
1 1 0
1 2 0
1 3 0
1 4 0
1 5 0
Z Z Z
5 m
10 m
15 m
20 m
25 m
30 m
35 m
40 m
45 m
50 m
55 m
60 m
65 m
70 m
75 m
80 m
85 m
90 m
95 m
100 m
110 m
120 m
130 m
140 m
150 m
Others (contact us)
Note: Preapare a detector and dedicated signal cable (2 set).
Loader software for PCs
Equipped as standard
• Works on PC/AT compatible machines.
• Operation on PC98-series machines (NEC) cannot be
guaranteed.
• Operation on self-made PCs or shop-brand PCs cannot
be guaranteed.
• Major functions: Setting/changing of various parameters
for the main unit
•
O/S: Windows2000/XP
• Memory requirement: 128MB or more
• Disk unit: Windows2000/XP-compatible CD-ROM drive
• Hard disk drive capacity: Free space of 52MB or more
Note: PC loader communication cable (type ZZP*TK4H6253,
Specifications: D-sub 9 pin receptacle, cable length
3m) is separately required.
SCOPE OF DELIVERY
ITEMS DESIGNATED ORDERING
•Flow transmitter (Type: FSH): Flow transmitter
CD-ROM (Instruction manual,
1.Detector code symbols
2.Flow transmitter code symbols
3.Signal cable code symbols
Loader software)
•Detector (Type: FSG, FSD): Sensor unit, Mounting belt,
Silicon compound
4.For large sensor: Mounting pipe size
5.Tag No. as necessary
•Signal cable (Type: FLY): Cable (2 wires)
6.If parameter setting is specified, send back the attached
parameter specification table duly filled.
6
(Unit:mm)
OUTLINE DIAGRAM
<Flow transmitter (type: FSH)>
Mtg. plate
Duosonics
72
3
134
14
233
5-G3/8
Female screw
4-G1/2
Female screw
30 30 30
48 48
24
48
CONNECTION DIAGRAM
Power supply terminals (AC)
1
L
2
N
3
4
1
2
3
4
5
6
7
8
9
B
B
A
+
–
100 to 240V AC
50/60Hz
SERIAL
Iout
SERIAL
No.
Power supply terminals (DC)
Communication cable
(option)
RS485
TRX2 (
RS232C
1
+
2
–
3
4
4
5
6
–
)
TXD
RXD
COM
Cap (red)
Cap (yellow)
Cap (black)
TRX1 (+)
20 to 30V DC
SHILD
1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
HF GND HF GND HF GND HF GND
+
–
+
–
+
–
UP STR DOWN STR UP STR DOWN STR
DO1
DO2
DO3
Relay contact
LINE1
LINE2
Open collector
7
FSH, FSG, FSD
Saddle
Connector
Scale(mm)
(Unit:mm)
OUTLINE DIAGRAM
Scale(inch)
Name plate
Lock nut
Element holder
Cursor
φ26
33±0.5
90max.
205max.
44±±
Detector FSD32 (High-temperature sensor)
Frame
Sensor
Frame
410±2
400
Sensor
410±2
400
5
5
5
5
P
P
Name plate
47
Spacing : 0 to 270mm
Name plate
47
Spacing : 0 to 270mm
50±1
33
Chain
50±1
33
Chain
Mounting spring
Mounting spring
Frame guide
VIEW FROM P
Frame guide
VIEW FROM P
(Common type)
(Submersible type)
Detector FSGS3 (Small sensor)
Name plate
Name plate
114
114
78±1
Sensor
Wire rope
Mounting spring
78±1
L(cable length (m) : specified by order)
Wire rope
Sensor
Mounting spring
(Common type)
(Submersible type)
Detector FSGS5 (Large sensor)
8
Signal cable : FLY8 (For FSG)
8
To Detector
ø11
ø7.3
To Flow transmitter
L : By specification
when ordered.
5.5
Signal cable : FLY9 (For FSD)
Loader cable : ZZP*TK4H6253
34
Connector
ø14.5
BNC connector
ø11
To Personal
computer
ø7.3
To Detector
ø6.9
To Flow transmitter
Terminal cap (Yellow)
Terminal cap (Black)
To Flow transmitter
Terminal cap
(Red)
L : By specification
when ordered.
Black
COM
Red
Yellow
RXD
3-ø3.7
TXD
6.6
5.5
RS-232C
9
FSH, FSG, FSD
<Parameter specification table>
No.
1
Setting item
Settable range
Initial value
60.00mm
Settable value
Outer diameter
10.00 to 6200.00mm
[mm, inch]
(0.393 to 244.100 inch)
(2.362 inch)
2
Pipe material
12 menus
PVC
Carbon steel, Stainless steel, PVC, Copper,
Castiron, Aluminum, FRP, Ductileiron, PEEK,
PVDF, Acrylic
Pipe S.V. : 1000 to 3700m/s
(3280 to 12140 ft/s)
Others (Sound velocity :
[m/s, ft/s])
3
4
Wall thickness
Lining material
0.10 to 100.00mm
(0.003 to 3.940 inch)
4.00mm
(0.157 inch)
[mm, inch]
8 menus
No lining
No lining, Tar epoxy, Mortar, Rubber,
Teflon, Pyrex glass, PVC,
Lining S.V. : 1000 to 3700m/s
(3280 to 12140 ft/s)
Others (Sound velocity :
[m/s, ft/s])
5
6
Lining thickness
Kind of Fluid
0.01 to 100.00mm
(0.000 to 3.940 inch)
[mm, inch]
17 menus
Water
Water, Seawater, DIST. water, Ammonia,
Alcohol, Benzene, Bromide, Ethanol, Glycol,
Kerosene, Milk, Methanol, Toluol,
Lube oil,
Fluid S.V. : 500 to 2500m/s
(1641 to 8203 ft/s)
Fuel oil, Petrol,
Others (Sound velocity :
(Kinematic viscosity
[m/s, ft/s])
Kinematic viscosity :
[x10-6m2/s, ft2/s])
0.001 to 999.9999 x 10-6m2/s
(0.0107 to 10763.9088 x 10-6ft2/s)
19 menus
m/s
(ft/s)
m/s, L/s, L/min, L/h, L/d, kL/d, ML/d, m3/s,
m3/min, m3/h, m3/d, km3/d, Mm3/d, BBL/s,
BBL/min, BBL/h, BBL/d, KBBL/d, MBBL/d,
(ft/s, ft3/s, ft3/min, ft3/h, ft3/d, kft3/d, Mft3/d,
gal/s, gal/min, gal/h, gal/d, kgal/d, Mgal/d,
BBL/s, BBL/min, BBL/h, BBL/d, kBBL/d,
MBBL/d)
7
Range unit
8
9
Range type
4 menus
Single
Single, Auto 2, Bi-dir, Bi-dir Auto 2
[
Full scale or
Full scale 1
In terms of flow velocity 0.00 ···
0.30 to 32.00m/s
2.00m/s
(6.56 ft/s)
]
]
(0.98 to 104.98 ft/s)
10
Full scale 2
In terms of flow velocity 0.00 ···
0.30 to 32.00m/s
4.00m/s
(13.12 ft/s)
[
(0.98 to 104.98 ft/s)
11
12
13
14
15
16
Range HYS.
0.00 to 20.0%
-20 to 0%
10.00%
-20%
%
%
%
Output limit LO.
Output limit HI.
Output burnout
Burnout timer
Rate limit
100 to 120%
5 menus
120%
Hold
Not use, Hold, Upper, Lower, Zero
0 to 900sec
10sec
sec
]
0.00 to 5.00m/s
0.00m/s
[
(0.00 to 16.40 ft/s)
in terms of flow velocity
(0.00 ft/s)
17
Rate limit timer
0 to 900sec
0.0 to 100.0sec
7 menus
0sec
sec
sec
18 Damping
5.0sec
19
20
1 : Display kind
Flowrate (m3/s)
Flow velocity, Flowrate, Total forward,
Total reverse,
F : Total pulse, R : Total pulse, Flow rate (%)
2 : Display kind
7 menus
Flow velocity (m/s) Flow velocity, Flowrate, Total forward,
Total reverse,
F : Total pulse, R : Total pulse, Flow rate (%)
21 Low flow cut
0.00 to 5.00m/s
0.01m/s
[
]
(0.00 to 16.40 ft/s)
in terms of flow velocity
(0.03 ft/s)
10
No.
22
Setting item
Total mode
Settable range
Initial value
Total stop
Settable value
3 menus
8 menus
Total stop, Total run, Total reset
23
Total unit
mL
mL, L, m3, km3, Mm3, mBBL, BBL, kBBL,
ft3, kft3, Mft3, kgal, gal, mBBL, BBL, kBBL,
ACRF
24
25
26
27
28
29
30
31
32
Total rate
0.000 to 999999.999
0.000 to 9999999999.999
0.000 to 9999999999.999
0.000 to 9999999999.999
0.000 to 9999999999.999
2 menus
0.000
0.000
0.000
0.000
0.000
Hold
[
]
]
]
]
]
[
F : Total preset
F : Total SW
R : Total preset
R : Total SW
Output burnout
Burnout timer
Pulse width 1
Pulse width 2
[
[
[
Not use, Hold
0 to 900sec
10sec
50ms
50ms
sec
3 menus
50, 100, 200
9 menus
0.5, 1.0, 2.0, 5.0, 10.0, 20.0, 50.0, 100.0,
200.0
33
34
Flow sw high
Flow sw low
In terms of flow velocity
0.00 to 32.00m/s
(0.00 to 104.98 ft/s)
0.00m/s
4.00m/s
[
]
]
In terms of flow velocity
0.00 to 32.00m/s
[
(0.00 to 104.98 ft/s)
35
36
Flow sw HYS.
Output DO1
0 to 20%
15 menus
10%
%
Not use
Not use, Signal error, F : Total pulse,
R : Total pulse, F : Total alarm,
R : Total alarm, F : Total overflow,
R : Total overflow, Flow SW high,
Flow SW Low, Full scale2, AO range over,
Pulse range over, R : Flow direction,
Device error
37
38
Mode DO1
2 menus
Normal
Not use
Normal, Reverse
Output DO2
15 menus
Not use, Signal error, F : Total pulse,
R : Total pulse, F : Total alarm,
R : Total alarm, F : Total overflow,
R : Total overflow, Flow SW high,
Flow SW Low, Full scale2, AO range over,
Pulse range over, R : Flow direction,
Device error
39
40
Mode DO2
2 menus
Normal
Not use
Normal, Reverse
Output DO3
15 menus
Not use, Signal error, F : Total pulse,
R : Total pulse, F : Total alarm,
R : Total alarm, F : Total overflow,
R : Total overflow, Flow SW high,
Flow SW Low, Full scale2, AO range over,
Pulse range over, R : Flow direction,
Device error
41
Mode DO3
2 menus
Normal
Normal, Reverse
11
FSH, FSG, FSD
No.
42
Setting item
Settable range
Initial value
Metric
Settable value
System unit
Language
2 menus
5 menus
Metric, English
43
English
Japanese, English, German, French,
spanish
44
45
46
47
48
49
COM. speed
3 menus
3 menus
2 menus
2 menus
38400BPS
None
1 bit
9600BPS, 19200BPS, 38400BPS
None, Even, Odd
1 bit, 2bits
COM. parity
COM. stop bit
Serial method
Station No.
RS232C
1
RS232C, RS485
1 to 31
31 menus
2 menus
Measurement
mode
1 Path
(2 path)
2 Path
Note2
50
51
AO Definition
3 menus
4 menus
Line 1
(Average) Note2
Average, Line 1, Line 2
FSW12, FSW21, FSW40, FSW50,
FSW12
Sensor Type
FLW11/FSG31, FLW41/FSG41, FLW12/FSG32,
FLD12/FSD12, FLD22/FSD22, FLW32/FSD32,
FLS50/FSG50, FLW51/FSG51
(FLW51/FSG51)
Note2
Note1: When total pulse output has been selected for DO1, DO2 or DO3 specify total pulse value and total pulse width so that conditions 1
and 2 shown below are satisfies.
Flow span-1*[m3/s]
total pulse value*[m3]
Condition 1 :
Condition 2 :
1000 [In the case of DO1 and DO2]
1 [In the case of DO3]
Flow span-1*[m3/s]
total pulse value*[m3]
1000
2 × total pulse width [ms]
* In the case of 2 ranges, perform calculations using either flow span-1 or flow span-2, whichever is greater.
Note2: Settable value at the time of shipment.
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fjielectric.com/products/instruments/
Information in this catalog is subject to change without notice.
Printed in Japan
相关型号: