EDSX5-74 [FUJI]

FCX – AX SERIES SMALL FLANGE REMOTE SEAL TYPE PRESSURE TRANSMITTER; FCX â ???? AX系列小法兰远传密封压力变送器
EDSX5-74
型号: EDSX5-74
厂家: FUJI ELECTRIC    FUJI ELECTRIC
描述:

FCX – AX SERIES SMALL FLANGE REMOTE SEAL TYPE PRESSURE TRANSMITTER
FCX â ???? AX系列小法兰远传密封压力变送器

局域网
文件: 总10页 (文件大小:204K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
FCX – AX SERIES  
SMALL FLANGE REMOTE SEAL TYPE  
PRESSURE TRANSMITTER  
...  
DATA SHEET  
FHW, FKW 3  
The FCX –AX Series small flange remote seal type pres-  
sure transmitter accurately measures gauge pressure and  
transmits a proportional 4 to 20mA signal.  
The transmitter utilizes a unique micromachined capaci-  
tance silicon sensor with state-of -the-art microprocessor  
technology to provide exceptional performance and func-  
tionality.  
Totally welded construction of the seals assures excellent  
reliability in high temperature and highly corrosive process  
conditions.  
7. Burnout current flexibility (Under Scale: 3.2 to 3.8mA,  
Over Scale: 20.8 to 21.6mA)  
Burnout signal level is adjustable using Model FXW hand  
Held Communicator (HHC) to comply with NAMUR NE43.  
(Available for amplifier unit from version 24 and FXW(HHC) version  
5.3.)  
FEATURES  
1. Directly connectable to 1-1/2in and 2in flanges  
The transmitter is connectable to 1-1/2in and 2in pipes  
without a reducer.  
2. Connectable to 1/2in and 3/4in pipes  
Use of direct mounting adapter allows the transmitter to  
be connected to the following process.  
1/2in and 3/4in flanges  
Screw connection 1/2-14NPT, 3/4-14NPT, Rc1/2, Rc3/4  
3. Minimum environmental influence  
8. Dry calibration without reference pressure  
Thanks to the best combination of unique construction  
of mechanical parts (Sensor unit) and high performance  
electronics circuit (Electronics unit), reliability of dry  
calibration without reference pressure is at equal level  
as wet calibration.  
The "Floating Cell" design which protects the pressure  
sensor against changes in temperature, and overpressure  
substantially reduces total measurement error in actual  
field applications.  
4. Replaceable Communication Module  
Fuji micro-electronics manufacturing technology offers  
replaceable communication module that makes FCX–AX  
transmitter very unique in design. In case of change in  
communication protocl, all that needs to be done is just to  
replace the module and the transmitter gets upgraded to  
the new version.  
5. Fuji/HART bilingual communication module  
The communication module is "bilingual" to speak both Fuji  
proprietary protocol and HART. Any HART compatible  
devices can communicate with FCX–AX series transmit-  
ters.  
SPECIFICATIONS  
Functional specifications  
Type:  
Model FHW: 4 to 20mA  
Model FKW: 4 to 20mA with digital signal  
Service:  
Liquid, gas, or vapour  
Span, range, and overrange limit:  
Span limit [kPa]{bar}  
Overrange  
limit  
[MPa] {bar}  
Range limit  
[kPa]{bar}  
Type  
Min.  
Max.  
FHW/FKW  
FHW  
300  
{3}  
1000  
{10}  
FKW  
50  
{0.5}  
250  
F
F
W
3
4
3000  
{30}  
10000  
{100}  
–100 to +3000  
{–1 to +30}  
–100 to 10000  
{–1 to 100}  
4.5  
{45}  
15  
6. Application flexibility  
Various options that render the FCX–AX suitable for  
almost any process applications include:  
W
{2.5}  
{150}  
Lower range limit (vacuum limit) ;  
Silicone fill sensor: See Fig. 1  
Fluorinated fill sensor: Atmospheric pressure  
Conversion factors to different units;  
1MPa=103kPa=10bar=10.19716kgf/cm2=145.0377psi  
1kPa=10mbar=101.9716mmH2O=4.01463inH2O  
Analog indicator at either the electronics side or ter-  
minal side  
Full range of hazardous area approvals  
Built-in RFI filter and lightning arrester  
1
/
2
4
-digits LCD meter  
Stainless steel electronics housing  
Wide selection of materials  
High temperature, vacuum seals  
Output signal:  
Model FHW: 4 to 20mA DC 2-wire  
Model FKW: 4 to 20mA DC with digital signal super-  
imposed on the 4 to 20mA signal.  
Power supply:  
Transmitter operates on 10.5V to 45V DC  
at transmitter terminals.  
10.5V to 32V DC for the units with op-  
tional arrester.  
EDSX5-74  
Date  
Apr. 10, 1999  
...  
FHW, FKW 3  
Load limitations: see figure below  
[]  
“Output Underscale”:  
Approx. 3.8mA  
(Adjustable within the range 3.2mA to  
3.8mA from HHC)  
2000  
1533Ω  
4
3.8  
20 20.8  
3.2  
1500  
Over scale  
Burnout  
0
1000  
21.6  
Normal operating range  
Probable under range  
600Ω  
Under scale  
Burnout  
Probable over range  
Operating  
area  
500  
250  
24V  
45V  
50  
Loop-check output:  
Model FHW: Transmitter can output constant signal of  
4mA, 12mA, or 20mA if MODE SWITCH  
is set to the loop check mode.  
10 16.1  
10.5V  
[ V ]  
Power voltage  
Model FKW: Transmitter can be configured to provide  
constant signal 3.8mA through 21.6mA by  
HHC.  
Note: For communication with HHC (Model: FXW), min. of 250is required.  
Hazardous locations: (Approval pending)  
Temperature limit:  
Ambient: 15 to +65°C  
Type N  
Nonincendive  
Authorities  
Flameproof  
Intrinsic safety  
(– 15 to +60°C for arrester option)  
(–10 to +60°C for fluorinated oil fill transmitter)  
(–10 to +60°C for silicone oil "H", "S")  
For explosionproof units (flameproof or intrinsic safety),  
ambient temperature must be within the limits speci-  
fied by each standard.  
BASEEFA  
Factory  
Mutual  
Ex ds IIC T5, T6  
Class I II III  
Div. 1  
EEx ia IIC T4, T5  
Class I II III  
Div. 1  
Ex N II T5  
Class I II III  
Div. 2  
Groups B thru. G Groups A thru. F  
Groups A thru. G  
RIIS  
Ex ds IIB+H2 T4  
Zero/span adjustment:  
Process:  
Model FHW: Zero is adjustable from the external ad-  
justment screw.  
Lower limit  
13th digit of  
"Code symbols"  
Process  
temperature  
of  
Fill fluid  
The adjustment screw can also function to  
adjust span when MODE SWITCH (lo-  
cated on the electronics unit) is in the span  
mode. INHIBIT mode to disable the ad-  
justment screw is also available.  
static press.  
Fluorinated oil W, A and D  
–20 to 120°C Atmospheric  
pressure  
Silicone oil  
H
0 to 250°C  
Y and G  
S
40 to 120°C  
0 to 250°C  
2.7kPa abs  
{20mmHg  
abs}  
Model FKW: Zero and span are adjustable from the  
HHC. Zero is also adjustable externally  
from the adjustment screw.  
Storage: 40 to +70°C  
Humidity limit: 0 to 100% RH  
Communication: (Model FKW only)  
Damping:  
Adjustable electrical damping.  
Model FHW: The time constant is adjustable to 0, 0.3,  
1.2, 4.8, or 19.2 seconds.  
With HHC (Model FXW, consult Data  
Sheet No. EDS8-47), following informa-  
tion can be remotely displayed or recon-  
figured.  
Model FKW: The time constant is adjustable between 0  
to 38.4 seconds.  
Zero elevation/suppression:  
Zero can be elevated or suppressed within  
the specified range limit of each sensor  
model.  
Items  
Display  
Set  
v
v
v
Tag No.  
v
Model No.  
Normal/reverse action:  
v
v
Serial No.  
Selectable by moving a jumper pin located  
Engineering unit  
Range limit  
v
on the electronics unit.  
/
v
v
Indication:  
Analog indicator or 4 1 2-digit LCD meter,  
as specified.  
Measuring range  
Damping  
v
Burnout direction:If self-diagnostic detect transmitter fail-  
ure, the analog signal will be driven to ei-  
ther “Output Hold”, “Output Overscale”  
or “Output Underscale” modes.  
Model FHW: Unless otherwise specified in the order,  
the transmitter will be shipped in “Output  
Hold” mode.  
v
v
v
v
Output mode  
Burnout direction  
Adjustment  
Output adjust  
Data  
v
v
v
v
v
v
v
v
Self diagnoses  
Printer  
(Output signal just before failure happens  
is maintained.)  
Model FKW: Selectable from HHC  
“Output Hold”:  
v
External switch lock  
Transmitter display(*)  
v
v
Output signal is hold as the value just be-  
fore failure happens.  
“Output Overscale”:  
Note: (*) HHC's version must be more than 5.0 (or  
FXW 1– 2), to use this function.  
Approx. 21.6mA  
(Adjustable within the range 20.8mA to  
21.6mA from HHC)  
2
Performance specifications  
Physical specifications  
Accuracy rating: (including linearity, hysteresis, and re-  
Electrical connections:  
(Standard)  
peatability)  
G1/2, 1/2-14 NPT, Pg13.5, or M20 x 1.5  
conduit, as specified.  
Process connections:  
For spans greater than 1/10 of URL: ±0.25% of span  
For spans below 1/10 of URL (Model FKW only):  
JIS;  
0.1 x URL  
Span  
±
0.17+0.08  
% of span  
(
)
10K, 20K, 30K, 63K –40, 50A  
10K, 20K, 30K, 63K –15, 20A (with  
Adapter)  
(Option)  
For spans greater than 1/10 of URL: ±0.1% of span  
For spans below 1/10 of URL (Model FKW only):  
0.1 x URL  
ANSI/JPI;  
150LB, 300LB, 600LB, –1 1 2", 2"  
/
±
0.05+0.05  
% of span  
(
)
Span  
0.1% of calibrated span  
150LB, 300LB, 600LB, –1/2", 3/4" (with  
Adapter)  
Screw connection (with Adapter) ;  
Rc1/2, Rc3/4, 1/2-14NPT, 3/4-14NPT  
Linearity:  
Stability:  
±0.2% of upper range limit (URL) for 24  
months  
Temperature effect:  
Effect per 28°C change between the lim-  
Diaphragm extension:  
0, 50, 100, 150, or 200mm as specified.  
(See model code. Extended diaphragm is  
available only with 316L stainless steel dia-  
phragm)  
its of 15°C and +65°C  
Zero shift: ±0.5%/28°C  
(x equal to 1/6.5 URL or more)  
URL  
Process-wetted parts material:  
Zero shift; ± (0.5  
)%/28°C  
6.5 x x  
(x less than 1/6.5 URL)  
Diaphragm: 316L stainless steel, Hastelloy-C  
Monel or Tantalum  
Flange face: 316 stainless steel, Hastelloy-C  
lining  
Total shift; ±0.75%/28°C  
(x less than 1/6.5 URL or more)  
URL  
Monel lining or Tantalum lin-  
ing  
Total shift; ±(0.25 + 0.5  
)%/28°C  
6.5 x x  
(x less than 1/6.5 URL)  
Extension: 316 stainless steel  
(Refer to "Code symbols")  
Non-wetted parts material:  
Where, x:  
Calibrated span  
URL: Maximum span (Upper Range  
Limit)  
Note 1: Condition:  
Electronics housing: Low copper die-cast  
aluminum alloy (standard), finished  
with polyester coating, or 316 stainless  
steel (SCSI4 per JIS G5121), as speci-  
fied.  
Capillary: In case of 13th code "Y. W. G. A.  
D", PVC armored stainless steel.  
In case of 13th code "H. S",  
stainless steel armored stainless steel.  
Mounting flange: (option) 304 stainless  
steel or carbon steel  
Fill fluid: Silicone oil (standard) or fluori-  
nated oil (Daifloil)  
Mounting bracket: Carbon steel with ep-  
oxy coating or 304 stainless steel, as  
specified  
Capillary length: 3m max.  
In case the capillary length is 5m, the per-  
formance becomes 1.5 times worse than  
above.  
Note 2: In case the 7th code (material code) is  
other than W, A, B, C or D, the perfor-  
mance becomes 2.5 times worse than  
above.  
Note 3: Above specifications are based on the con-  
ditions that flange and sensor unit are at  
the same temperature and in the same  
level. If temperature is different at flange,  
capillary or sensor unit, output variation  
may increase.  
Overrange effect: Zero shift; 0.2% of URL/(1.5 x URL)  
Supply voltage effect:  
Less than 0.05% of calibrated span per  
10V  
Less than 0.2% of URL for the frequen-  
cies of 20 to 1000MHz and field strength  
30 V/m when electronics covers on.  
Environmental protection:  
RFI effect:  
IEC IP67 and NEMA 4X  
Mounting:  
On 60.5mm (JIS 50A) pipe using mount-  
ing bracket, direct wall mounting  
Transmitter approximately 10kg without  
options.  
Add; 0.5kg for mounting bracket  
0.8kg for indicator option  
4.5kg for stainless steel housing  
option  
1.5kg per 50mm extension of diaphragm  
Step response: Time constant: 0.3s (with 1.5m capillary)  
Dead time: approximately 0.3s  
(without electrical damping)  
Dielectric strength:  
500V AC, 50/60Hz 1 min., between circuit  
and earth.  
Insulation resistance:  
Mass {weight}:  
More than 100M/500V DC.  
Turn-on time:  
4 sec.  
Internal resistance for external field indicator:  
12or less  
3
...  
FHW, FKW 3  
Optional features  
ACCESSORIES  
Indicator:  
A plug-in analog indicator (1.5% accuracy)  
Hand-held communicator:  
can be housed in the electronics compart-  
ment or in the terminal box of the hous-  
(Model FXW, refer to Data Sheet No.  
EDS8-47)  
ing.  
Communication module: (Standard for FKW)  
1
/
An optional 4 digits LCD meter is also  
By adding communication module, re-  
2
mote setting function becomes available  
available.  
for model FHW.  
Arrester:  
A built-in arrester protects the electronics  
from lightning surges.  
Lightning surge immunity is 4kV (1.2 x  
50µs).  
Remark: When the communication module is con-  
nected, the operation mode of external  
zero/span adjustment screw is limited to  
zero adjustment only.  
Oxygen service: Special cleaning procedures are followed  
throughout the process to maintain all pro-  
cess wetted parts oil-free.  
The fill fluid is fluorinated oil.  
Chlorine service: Oil-free procedures as above. Includes  
fluorinated oil for fill.  
Not available with material code “W”.  
Degreasing:  
Process-wetted parts are cleaned, but the  
fill fluid is standard silicone oil. Not for use  
on oxygen or chlorine measurement.  
Vacuum and high temperature service:  
Special silicone oil and filling procedure  
are applied.  
See below figure.  
Silicone (Code Y,G)  
[kPa abs]  
{m bar abs}  
Silicone (Code S)  
101  
{1010}  
53.3  
{533}  
Operating area  
13.3  
{133}  
The product conforms to the requirements of the  
Electromagnetic compatibility Directive 89/336/EEC as  
detailed within the technical construction file number  
TN510412. The applicable standards used to demonstrate  
compliance are :-  
Non-operating  
area  
2.7  
{27}  
EMI (Emission)  
EN50081-1 : 1992  
–40  
0
60 85 120  
Process temperature  
200 230 250  
[°C ]  
Test item  
Frequency range  
Basic standard  
Applicable Electro-  
magnetic Radiation  
Disturbance  
EN55022  
Class B  
30-1000MHz  
Fig. 1 Relation between process temperature and  
operating pressure  
EMS (Immunity) EN50082-1 : 1992  
Customer tag:  
A stainless steel tag for customer tag data  
is wired to the transmitter.  
Test  
Basic  
standard  
Performance  
criteria  
No. Test item  
specification  
Coating of cell: Cell's surface is finished with epoxy/poly-  
urethane double coating. Specify if envi-  
ronment is extremely corrosive.  
Electrostatic  
discharge  
1
8kV (Air)  
IEC 801-2:1984  
IEC 801-3:1984  
B
A
Radio-frequency 27-500MHz  
electromagnetic 3V/m  
field.  
2
3
(Unmodulated)  
0.5kV, 5/50  
(Tr/Th) ns  
5kHz Rep.  
Fast transients  
common mode  
IEC 801-4:1988  
B
"LVD - The transmitter is not covered by the requirements  
of the LVD standard."  
4
CODE SYMBOLS  
1
2
3
4
5
6
7
8
9
10 11 12 13  
14 15 16 17 18 19 20  
F
W
3
0
Description  
Type  
4 to 20mA, Output type  
4 to 20mA with digital signal, Output type  
FHW  
FKW  
Conduit connection <4th digit>  
G 1/2  
1/2 - 14NPT  
Pg 13.5  
M20 X 1.5  
S
T
V
W
Flange <5th digit>  
Material  
Size and rating  
0
1
2
3
4
5
6
7
A
B
C
D
E
F
304 stainless  
steel  
JIS 10K 40A  
JIS 10K 50A  
JIS 20K 40A  
JIS 20K 50A  
JIS 30K 40A  
JIS 30K 50A  
JIS 63K 40A  
JIS 63K 50A  
ANSI/JPI 150LB 1 1/2"  
ANSI/JPI 150LB 2"  
ANSI/JPI 300LB 1 1/2"  
ANSI/JPI 300LB 2"  
ANSI/JPI 600LB 1 1/2"  
ANSI/JPI 600LB 2"  
Carbon steel  
JIS 10K 40A  
JIS 10K 50A  
JIS 20K 40A  
JIS 20K 50A  
JIS 30K 40A  
JIS 30K 50A  
JIS 63K 40A  
JIS 63K 50A  
ANSI/JPI 150LB 1 1/2"  
ANSI/JPI 150LB 2"  
ANSI/JPI 300LB 1 1/2"  
ANSI/JPI 300LB 2"  
ANSI/JPI 600LB 1 1/2"  
ANSI/JPI 600LB 2"  
G
H
J
K
L
M
N
P
Q
R
S
T
U
V
None  
(Wafer type)  
40A, 1 1/2B  
50A, 2B  
W
X
Direct mounting adapter connection (* 1)  
Y
Span limit [kPa] {m bar} <6th digit>  
FHW FKW  
...  
..  
...  
...  
...  
...  
...  
3
4
300 3000 (3 30) 50 3000 ( 0.5 30)  
..  
1000 10000 (10 100) 250 10000 (2.5 100)  
Material/diapharagm extension <7th digit>  
Diaphragm extension (mm)  
Diaphragm  
Flange face  
0
50  
100  
150  
316L stainless steel 316 stainless steel  
W
A
B
C
D
(*2)  
200  
0
0
0
Hastelloy-C  
Monel  
Tantalum  
Hastelloy-C  
Monel  
Tantalum  
H
M
T
Note 1: Direct mounting adapter type is specified at 16th to 20th digit.  
Direct mounting adapter is available only for 7th digit code "W".  
2: Diaphragm extension is available only for 2" (50A) flanges.  
5
...  
FHW, FKW 3  
20  
1
2
3
4
5
6
7
8
9
10 11 12 13  
14 15 16 17 18 19  
F
W
3
Description  
Indicator and arrester <9th digit>  
Indicator  
Arrester  
A
B
D
J
E
F
H
K
L
P
Q
S
None  
None  
None  
None  
None  
Yes  
Yes  
Yes  
Yes  
Analog, 0 to 100% linear scale  
Analog, custom scale  
Analog, double scale  
None  
Analog, 0 to 100% linear scale  
Analog, custom scale  
Analog, double scale  
Digital, 0 to 100%  
Digital, custom scale  
Digital, 0 to 100%  
Digital, custom scale  
None  
None (Model FKW only)  
Yes  
Yes (Model FKW only)  
Approvals for hazardous locations <10th digit> (Approval pending)  
None (for ordinary locations)  
JIS, Flameproof (Conduit seal) (Available for 4th digit code "S")  
JIS, Flameproof (Cable gland seal) (Available for 4th digit code "S")  
FM, Flameproof (or explosionproof) (Available for 4th digit code "T")  
BASEEFA, Flameproof (Conduit seal)  
BASEEFA, Flameproof (Cable gland seal) (Conduit connection G 1/2 only)  
FM, Intrinsic safety and Nonincendive  
CENELEC, Intrinsic safety  
A
B
C
D
M
N
H
K
P
CENELEC, Intrinsic safety and BASEEFA, Type N  
Capillary and mounting bracket <11th digit>  
mounting bracket  
Carbon steel  
Capillary  
A
B
G
1.5m  
3m  
5m  
D
E
L
Stainless steel  
1.5m  
3m  
5m  
Stainless steel parts <12th digit>  
Stainless steel tag plate Stainless steel elec. housing  
Coating of cell  
None  
None  
None  
None  
Yes  
Yes  
Yes  
Yes  
Y
B
C
E
M
N
P
None  
Yes  
None  
Yes  
None  
None  
Yes  
Yes  
None  
Yes  
None  
Yes  
None  
None  
Yes  
Q
Yes  
Treatment/Fill fluid <13th digit>  
Treatment  
Fill fluid  
None  
None  
Degreasing  
Oxygen service  
Chlorine service  
High temp. 250°C  
Silicone oil (for general use)  
Fluorinated oil  
Silicone oil  
Y
W
G
A
D
H
S
Fluorinated oil (7th digit code "W", "A", "B", "C" and "D")  
Fluorinated oil (7th digit code "H" and "T")  
Silicone oil  
(7th digit code "W", "A", "B", "C" and "D")  
}
High temp. and vacuum (250°C) Silicone oil  
Teflon membrane <14th digit>  
None  
Yes (7th digit code "W", "H", "M", and "T")  
Y
C
Bolt/nut <15th digit>  
None  
6th digit code "3"  
6th digit code "4"  
Y
A
Standard (Cr-Mo alloy hexagon socket head cap bolt/  
carbon steel nut)  
Cr-Mo alloy hexagon bolt/carbon steel nut  
304 stainless steel bolt/304 stainless steel nut  
B
E
Odering information  
1. When odering this instrument, specify the output orientation (burnout direction) in case of abnormality in the transmitter.  
Unless otherwise specified, the output hold function is supplied.  
6
Specifications of Direct Mounting Adapter {for 15, 20A (1/2, 3/4") connection} and others  
Note 1. When odering the instrument with direct mounting adapter, specify "Y" in the 5th digit of Code Symbol, and specify  
16th digit to 20th digits.  
When odering the instrument without direct mounting adapter, nothing should be filled in the 16th to 20th digits.  
2. Unless otherwise described in the specifications, leave the 21st digit blank.  
1
2
3
4
5
6
7
8
9
10 11 12 13  
14 15 16 17 18 19 20  
F
W
3
Description  
Process connection (direct mounting adapter) <16th to 17 digit>  
JIS 10K 15A  
JIS 10K 20A  
JIS 20K 15A  
JIS 20K 20A  
JIS 30K 15A  
JIS 30K 20A  
JIS 63K 15A  
JIS 63K 20A  
ANSI/JPI 150LB 1/2"  
ANSI/JPI 150LB 3/4"  
ANSI/JPI 300LB 1/2"  
ANSI/JPI 300LB 3/4"  
ANSI/JPI 600LB 1/2"  
ANSI/JPI 600LB 3/4"  
Screw connection Rc1/2  
Screw connection Rc3/4  
Screw connection Rc1/2 - 14NPT  
Screw connection Rc3/4 - 14NPT  
1 1  
1 2  
2 1  
2 2  
3 1  
3 2  
6 1  
6 2  
1 H  
1 T  
2 H  
2 T  
4 H  
4 T  
S R  
S 2  
S N  
S T  
Material (direct mounting adapter) <18th digit>  
Adapter  
Bolts/nuts  
(* 1)  
W
316 Stainless Steel Cr-Mo steel/carbon steel  
Vent/drain (for direct mounting adapter) <19 th digit>  
Standard  
Long type  
G
N
Gasket (for direct mounting adapter) <20th digit>  
Standard (Teflon) (Only Y, W, G, A and D can be specified on 13th digit).  
For high temperature (spiral gasket) (Only H and S can be specified on  
13th digit).  
1
2
*1 For connection of transmitter receiving pressure unit and direct mounting adapter  
7
...  
FHW, FKW 3  
Fastener  
Zero/span adj. screw  
(In case of 10th  
digit code “B,C,M,N”)  
In case of 4th  
digit code “S,V,W,X”  
MTG. pipe  
JIS 50A (2")  
U - bolt M8  
Up  
Down  
30  
38  
“A”  
4 holes  
9 dia.  
Conduit conn.  
224  
58  
Cable gland  
(option)  
See note  
“B”  
Conduit conn.  
N holes  
h dia.  
Inner of gasket  
Detail “A”  
Conduit conn.)  
Detail “B”  
φ 13  
(
φ 15  
Earth terminal  
(both side)  
See table 1  
51  
71  
191  
(27)  
20  
J
G1/2  
t
L
See table 1  
With field ind.  
(option)  
11th digit of Capillary  
code symbols length S [mm]  
7th digit of  
Code symbols  
L
Mass approx.  
[kg]  
W, H, M, T  
0
A,D  
B, E  
G, L  
1500  
3000  
5000  
9.5  
10.5  
11.5  
12  
A
B
C
D
50  
100  
150  
200  
S
2-M8 × 12  
12.5  
Flange  
5th digit of code symbols φ D φ E  
t
N-φ h  
0, G  
1, H  
2, J  
140 105  
155 120  
140 105  
155 120  
160 120  
165 130  
175 130  
185 145  
36 4-19  
36 4-19  
38 4-19  
38 8-19  
42 4-23  
42 8-19  
52 4-25  
54 8.23  
JIS-10K-40A  
50  
70  
JIS-10K-50A  
JIS-20K-40A  
120  
3, K  
4, L  
5, M  
6, N  
7, P  
A, Q  
B, R  
C, S  
D, T  
E, U  
F, V  
JIS-20K-50A  
JIS-30K-40A  
JIS-30K-50A  
JIS-63K-40A  
JIS-63K-50A  
Conduit conn.  
K
4th digit of  
Code symbols  
Earth terminal  
127 98.4 37.5 4-16  
152 120.6 39.5 4-20  
156 114.3 41 4-23  
165 127 42.5 8-20  
156 114.3 42.5 4-23  
165 127 45.5 8-20  
ANSI/JPI-150LB-1 1/2"  
ANSI/JPI-150LB-2"  
ANSI/JPI-300LB-1 1/2"  
ANSI/JPI-300LB-2"  
ANSI/JPI-600LB-1 1/2"  
ANSI/JPI-600LB-2"  
J
M
S
T
No. 8-32UNC  
G1/2  
17  
8
M4  
M4  
M4  
1/2-14NPT 16  
5
V
Pg13.5  
8
4.5  
5
M20x1.5  
16  
W
Note *: Cable gland is supplied in case of flameproof packing type.  
ø11 cable is suitable.  
Table 1  
CONNECTION DIAGRAM  
<Optional stainless steel tag>  
2-M6 screw depth 9  
Inner of gasket  
+
CK+ CK–  
+
φ
φ
0
1
2
3
4
5
6
20  
S
H
K
T
V
20  
3
F
W
Note: In case of wafer type, flange is excluded  
from the scope of supply.  
W
<Wafer type>  
Mount flange, referring to the view.  
8
OUTLINE DIAGRAM (Unit:mm)  
Zero/span adj. screw  
Fastener  
(In case of 10th  
digit code “B,C,M,N”)  
In case of 4th  
digit code “S,V,W,X”  
MTG. pipe  
JIS 50A (2")  
U - bolt M8  
Up  
Down  
30  
38  
“A”  
Conduit conn.  
4 holes  
9 dia.  
224  
58  
Cable gland  
(option)  
See note  
“B”  
Conduit conn.  
Specifications for instrument with  
direct mounting adapter  
80  
20  
N holes H dia.  
t
Detail “A”  
Conduit conn.)  
Detail “B”  
φ 13  
S
(
φ 15  
J
G1/2  
(6.4) In case of  
ANSI/JPI-600LB flange  
See table  
1
(T) In case of  
ANSI/JPI-600LB flange  
With field ind.  
(option)  
Earth terminal  
(both side)  
See table 1  
Vent/drain plug  
51  
71  
191  
(27)  
Long vent/drain plug  
(added specification)  
Table 1  
Conduit conn.  
K
4th digit of  
Code symbols  
S
11th digit of  
code symbols  
Earth terminal  
S
J
M
A,D  
B, E  
G, L  
1500  
3000  
5000  
M4  
No. 8-32UNC  
M4  
G1/2  
17  
8
T
V
1/2-14NPT 16  
5
Pg13.5  
8
4.5  
5
2-M8 × 12  
M4  
W
M20x1.5  
16  
16th digit of 17th digit of  
code symbols code symbols  
øD  
95  
øE  
70  
øF  
t
N-øh  
Flange  
50  
70  
1
1
2
2
3
3
6
6
1
1
2
2
4
4
1
2
1
2
1
2
1
2
H
T
H
T
H
T
51  
56  
51  
56  
55  
60  
55  
60  
12  
14  
14  
16  
18  
18  
23  
25  
4-15 JIS-10K-15A  
4-15 JIS-10K-20A  
4-15 JIS-20K-15A  
4-15 JIS-20K-20A  
4-19 JIS-30K-15A  
4-19 JIS-30K-20A  
4-19 JIS-63K-15A  
4-23 JIS-63K-20A  
120  
100 75  
95 70  
100 75  
115 80  
120 85  
120 85  
135 95  
CONNECTION DIAGRAM  
+
CK+ CK–  
+
89  
98  
95  
60.3 34.9 11.5 4-16 ANSI/JPI-150LB-1/2"  
69.9 42.9 13 4-16 ANSI/JPI-150LB-3/4"  
66.7 34.9 14.5 4-16 ANSI/JPI-300LB-1/2"  
4-20 ANSI/JPI-300LB-3/4"  
66.7 34.9 14.5 4-16 ANSI/JPI-600LB-1/2"  
117 82.5 42.9 16 4-20 ANSI/JPI-600LB-3/4"  
1
2
3
4
6
S
T
V
W
H
K
3
F
W
Y
117 82.5 42.9 16  
95  
Note *: Cable gland is supplied in case of flameproof packing type.  
ø11 cable is suitable.  
9
...  
FHW, FKW 3  
OUTLINE DIAGRAM (Unit:mm)  
Fastener  
(In case of 10th  
digit code “B,C,M,N”)  
Zero/span adj. screw  
In case of 4th  
digit code “S,V,W,X”  
MTG. pipe  
JIS 50A (2")  
U - bolt M8  
UP  
DOWN  
30  
38  
“A”  
4 holes  
9 dia.  
Conduit conn.  
224  
58  
Cable gland  
(option)  
See note  
Detail “A”  
(Contuit conn.)  
Detail “B”  
“B”  
Conduit  
conn.  
φ 13  
φ 15  
23  
20  
J
G1/2  
See table 1  
S
With field ind.  
(option)  
Earth terminal  
(both side)  
See table 1  
Press. conn. "D"  
51  
71  
191  
(27)  
Long vent/drain plug  
Long vent/drain plug  
(option)  
Specifications for instrument  
with direct mounting adapter  
(screw connection type)  
2-M8 × 12  
Table 1  
Conduit conn.  
11th digit of  
code symbols  
4th digit of  
Code symbols  
S
Earth terminal  
S
J
K
M
50  
70  
A,D  
B, E  
G, L  
1500  
3000  
5000  
M4  
No. 8-32UNC  
M4  
G1/2  
17  
8
120  
T
V
1/2-14NPT 16  
5
Pg13.5  
8
4.5  
5
M4  
W
M20x1.5  
16  
CONNECTION DIAGRAM  
Press. conn. "D"  
16th digit of code symbols 17th digit of code symbols  
+
CK+ CK–  
Rc 1/2  
S
S
S
S
R
N
2
1/2 - 14NPT  
Rc 3/4  
+
S
T
H
F
3/4 - 14NPT  
T
Y
3
S
W
K
V
Note *: Cable gland is supplied in case of flameproof packing type.  
ø11 cable is suitable.  
W
Head office  
Sales Div.  
International Sales Dept.  
No.1, Fuji-machi, Hino-city, Tokyo 191-8502 Japan  
Phone: 81-42-585-6201, 6202  
11-2 Osaki 1-chome, Shinagawa-ku, Tokyo, 141-0032 Japan  
Phone: 81-3-5435-7111  
http://www. fujielectric.co.jp/eng/sg/KEISOKU/welcome.htm  
Fax: 81-42-585-6187, 6189  
Printed in Japan  
Information in this catalog is subject to change without notice.  

相关型号:

EDSX5-75

FCX – C SERIES ABSOLUTE PRESSURE TRANSMITTER
FUJI

EDSX5-76A

FCX – C SERIES PRESSURE TRANSMITTER
FUJI

EDSX5-77

FCX –AX SERIES PRESSURE TRANSMITTER Hydroseal® Diaphragm Version
FUJI

EDSX5-79J

ABSOLUTE PRESSURE TRANSMITTER
FUJI

EDSX5-80K

PRESSURE TRANSMITTER
FUJI

EDSX5-82K

SMALL FLANGE REMOTE SEAL TYPE PRESSURE TRANSMITTER
FUJI

EDSX5-83J

REMOTE SEAL TYPE PRESSURE TRANSMITTER
FUJI

EDSX5-85D

ELECTRONIC TRANSMITTER with FIELDBUS
FUJI

EDSX5-86D

PRESSURE TRANSMITTER
FUJI

EDSX5-87A

FCX – C2 PRESSURE TRANSMITTER
FUJI

EDSX5-91H

ABSOLUTE PRESSURE TRANSMITTER
FUJI

EDSX5-92H

PRESSURE TRANSMITTER
FUJI