C9404CA_15 [HAMAMATSU]
Mini-spectrometers;型号: | C9404CA_15 |
厂家: | HAMAMATSU CORPORATION |
描述: | Mini-spectrometers |
文件: | 总8页 (文件大小:423K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
Mini-spectrometers
TG series
C9404CA
C9404CAH
High sensitivity type (integrated with back-
thinned type CCD image sensor)
TG series mini-spectrometers are polychromators integrated with optical elements, an image sensor and a driver circuit. Light
to be measured is guided into the entrance port of TM series through an optical fiber and the spectrum measured with the
built-in image sensor is output from the USB port to a PC for data acquisition. The C9404CA and C9404CAH are high sensitiv-
ity mini-spectrometers employing a back-thinned type CCD image sensor. Their sensitivity is about two orders of magnitude
higher than CMOS types making them even more ideal for low-light-level measurement. The C9404CAH is high resolution type
(resolution: 1 nm Typ.). Their products come supplied with evaluation software that allows setting measurement conditions,
acquiring and saving data, and displaying graphs. Original measurement software can be designed on an end-user’s side as
DLL’s function specification is disclosed.
Features
Applications
Integrated with back-thinned type CCD image sensor:
Sensitivity is about two orders of magnitude higher than
CMOS types.
Low-light-level measurement such as fluorescence measurement
Evaluation of light source characteristics such as UV light source
High resolution 1 nm (C9404CAH)
High throughput due to transmission grating made of quartz
Easy to install into equipment
Wavelength conversion factor*1 is recorded in internal memory
Supprts external trigger input*2
*1: A conversion factor for converting the image sensor pixel number into a wavelength is recorded in the module. A calculation factor
for converting the A/D converted count into the input light intensity is not provided.
*2: Coaxial cable for external trigger input is sold separately. Refer to the “Mini-spectrometers Selection Guide” for details on external
triggers.
Optical characteristics
TG-UV-CCD
Parameter
Spectral response range
Spectral resolution (FWHM)*3
Wavelength reproducibility*4
Wavelength temperature dependence
Spectral stray light*3 *5
Unit
C9404CA
3 max.
C9404CAH
1 typ.
200 to 400
nm
nm
nm
nm/°C
dB
-0.1 to +0.1
-0.02 to +0.02
-35 max.
*3: Depends on the slit opening. Values were measured with the slit listed in the table “ Structure”.
*4: Measured under constant light input conditions
*5: When monochromatic light of 300 nm is input, spectral stray light is defined as the ratio of the count measured at the input
wavelength, to the count measured in a region of the input wavelength ±20 nm.
Electrical characteristics
Parameter
Specification
16
10 to 10000
USB 1.1
100 max.
5
Unit
bit
ms
-
mA
V
A/D conversion
Integration time
Interface
USB bus power current consumption
External power supply
1
www.hamamatsu.com
Mini-spectrometers
TG series
C9404CA, C9404CAH
Structure
Parameter
C9404CA
C9404CAH
Unit
Dimensions (W × D × H)
Weight
125.7 × 115.7 × 75
670
mm
g
Image sensor
Back-thinned type CCD image sensor (S10420-1006-01)
1024
-
Number of pixels
Slit*6 (H × V)
NA*7
pixels
μm
-
-
140 × 500
10 × 1000
0.11
SMA905D
Connector for optical fiber
*6: Entrance slit aperture size
*7: Numeric aperture (solid angle)
Absolute maximum ratings
Parameter
Operating temperature*8
Storage temperature*8
C9404CA
C9404CAH
Unit
°C
°C
+5 to +40
-20 to +70
*8: No dew condensation
When there is a temperature difference between a product and the surrounding area in high humidity environment, dew condensation
may occur on the product surface. Dew condensation on the product may cause deterioration in characteristics and reliability.
Note: Exceeding the absolute maximum ratings even momentarily may cause a drop in product quality. Always be sure to use the
product within the absolute maximum ratings.
Spectral resolution vs. wavelength
Linearity (typical example)
(Typ. Ta=25 °C)
100000
10
8
5
4
3
2
Typical example of A/D output
Ideal value of A/D output
Difference between ideal value
and typical example
6
10000
1000
4
C9404CA
2
(Slit width 140 μm, NA 0.11)
0
-2
-4
-6
100
10
1
0
C9404CAH
-8
(Slit width 10 μm, NA 0.11)
-10
10000
10
100
1000
200
300
400
Wavelength (nm)
Integration time (ms)
KACCB0160EC
KACCB0236EB
A/D output is the output with dark output is
subtracted when light is input. The difference
between the ideal value and typical example
contains a measurement error. The smaller the
A/D output, the larger the measurement error.
2
Mini-spectrometers
TG series
C9404CA, C9404CAH
Dark output vs. ambient temperature (typical example
)
(Integration time 1 s)
70000
60000
50000
40000
30000
20000
10000
0
0
10
20
30
40
50
Ambient temperature (°C)
KACCB0238EB
A/D output is the sum of the sensor and circuit
offset outputs and the sensor dark output.
Optical component layout
TM series mini-spectrometers use a transmission holographic grating made of quartz and precision optical components arranged on a rugged
optical base, making it possible to deliver high throughput and highly accurate optical characteristics.
Focus lens
Transmission grating
Image sensor
Collimating lens
Input slit
KACCC0256EA
3
Mini-spectrometers
TG series
C9404CA, C9404CAH
Connection example (transmission light measurement)
Light to be measured is guided into the entrance port of TM series through an optical fiber and the spectrum measured with the built-in
image sensor is output through the USB port to a PC for data acquisition. There are no moving parts inside the unit so stable measure-
ments are obtained at all times. An optical fiber that guides light input from external sources allows a flexible measurement setup.
AC adapter
Mini-
spectrometer
UV-VIS fiber light source
(Deuterium lamp and Halogen lamp)
L10290
PC
USB cable
Fiber
Quartz cell
(For holding liquid sample)
KACCC0355EA
Measurement example
Line spectra from low-pressure mercuy lamp
100000
10000
1000
100
C9404CA
C9404CAH
10
200
250
300
350
400
Wavelength (nm)
KACCB0166EA
4
Mini-spectrometers
TG series
C9404CA, C9404CAH
Evaluation software package (supplied with unit)
Installing the evaluation software package (Spec Evaluation.exe)*9
into your PC allows running the following basic tasks:
· Measurement data acquisition and save
· Measurement condition setup
· Module information acquisition
(wavelength conversion factor, polychromator type, etc.)
· Graphic display
· Arithmetic operation
Pixel number to wavelength conversion
Comparison calculation with reference data
(transmittance, reflectance)
Dark subtraction
Gaussian approximation
(peak position and count, FWHM)
Note:
· Two or more mini-spectrometers can be connected and used with one PC simultaneously.
· The external trigger input function does not work with the evaluation software. If using an external trigger input or designing original
application software, the user software must be configured to support that function.
*9: Compatible OS: Microsoft® Windows® 7 Professional SP1 (32-bit, 64-bit)
Microsoft® Windows® 8 Professional (32-bit, 64-bit)
DLL for controlling hardware is also provided.
You can develop your own measurement programs by using a following software development environment.
Microsoft® Visual Studio® 2008 (SP1) Visual C++®
Microsoft® Visual Studio® 2008 (SP1) Visual Basic®
Note: Microsoft, Windows, Visual Studio, Visual C++ and Visual Basic are either registerd trademarks or trademarks of Microsoft Corpo-
ration in the United States and other countries.
5
Mini-spectrometers
TG series
C9404CA, C9404CAH
Dimensional outline (unit: mm)
125.7
35
20 0.2
(2 ×) M3 tap depth 5
(from backside)
Tolerance unless otherwise noted: 0.5
Weight: 670 g
KACCA0202EC
Accessories
· USB cable
· Dedicated software (evaluation software, sample software, DLL)
· AC adapter (for power supply)
Options (sold separately)
· Optical fiber for light input
Type no.
Product name
Applicable mini-spectrometer
Core diameter (μm)
Specification
NA=0.22, length 1.5 m,
connectorized SMA905D
at both ends
Fiber for UV/visible range
(resistance to UV)
C9404CA (TG-UV-CCD)
C9404CAH (TG-UV-CCD)
A9762-01
600
· Coaxial cable for external trigger input A10670
Dimensional outline (unit: mm)
Cable length: 1.5 m
Cable specification: 1.5QEV
31.7
26.0
LEMO connector
FFA.00.250 (made by LEMO)
BNC connector
BNC-P-1.5 (40) (made by HIROSE or equivalent)
KACCA0220EB
6
Mini-spectrometers
TG series
C9404CA, C9404CAH
Mini-spectrometer lineup
Spectral resolution
max.
Spectral response range (nm)
Type no.
C10082CA
C10082CAH
C10082MD
C10083CA
Type
Image sensor
200 400 600 800 1000 1200 1400 1600 1800 2000 2200 2400 2600
(nm)
TM-UV/VIS-CCD
High sensitivity
6
Back-thinned CCD
image sensor
TM-UV/VIS-CCD
High resolution
200 to 800
1*
TM-UV/VIS-MOS
Wide dynamic range
CMOS linear
image sensor
6
TM-VIS/NIR-CCD
High sensitivity
8
(λ=320 to 900 nm)
Back-thinned CCD
image sensor
TM-VIS/NIR-CCD
High resolution
(λ=3201to*900 nm)
C10083CAH
C10083MD
C11697MB
320 to 1000
TM-VIS/NIR-MOS
Wide dynamic range
CMOS linear
image sensor
High-sensitivity CMOS
linear image sensor
8
8
TM-VIS/NIR-MOS-II
Trigger-compatible
TG-UV-CCD
High sensitivity
3
C9404CA
C9404CAH
C9405CB
C11713CA
C11714CB
Back-thinned CCD
image sensor
200 to 400
TG-UV-CCD
High resolution
1*
IR-enhanced
back-thinned CCD
image sensor
TG-SWNIR-CCD-II
IR-enhanced
5
500 to 1100
500 to 600
790 to 920
(λ=550 to 900 nm)
TG-RAMAN-I
High resolution
Back-thinned CCD
image sensor
0.3*
0.3*
IR-enhanced
back-thinned CCD
image sensor
TG-RAMAN-II
High resolution
TG2-NIR
Non-cooled type
C11482GA
C9913GC
7
7
900 to 1700
TG-cooled NIR-I
Low noise (cooled type)
InGaAs linear
image sensor
TG-cooled NIR-II
Low noise (cooled type)
C9914GB
1100 to 2200
8
TG-cooled NIR-III
Low noise (cooled type)
C11118GA
C13053MA
C13054MA
C11007MA
C11008MA
20
3.5
0.4*
9
900 to 2550
FT-SWIR-MOS-II
Compact, thin case
500 to 1100
790 to 920
High-sensitivity
CMOS linear
image sensor
FT2-RAMAN
Compact, thin case
RC-VIS-MOS
Spectrometer module
CMOS linear
image sensor
IR-enhanced
CMOS linear
image sensor
340 to 780
RC-SWNIR-MOS
640 to 1050
8
Spectrometer module
* Typ.
For installation into mobile measuring equipment
Spectral resolution
Spectral response range (nm)
max.
(nm)
Type no.
Type
Image sensor
200 400 600 800 1000 1200 1400 1600 1800 2000 2200 2400 2600
RC-VIS-MOS
Spectrometer head
CMOS linear
image sensor
IR-enhanced
CMOS linear
image sensor
340 to 780
640 to 1050
9
8
C11009MA
C11010MA
RC-SWNIR-MOS
Spectrometer head
For installation into mobile measuring equipment (ultra-compact)
Spectral resolution
Spectral response range (nm)
200 400 600 800 1000 1200 1400 1600 1800 2000 2200 2400 2600
max.
(nm)
Image sensor
Type no.
Type
MS-SWNIR-MOS
CMOS linear
image sensor
640 to 1050
C11708MA
20
Spectrometer head
CMOS linear
image sensor
Spectrometer head
Spectrometer head
C12666MA
C12880MA
340 to 780
340 to 850
15
15
High-sensitivity
CMOS linear
image sensor
7
Mini-spectrometers
TG series
C9404CA, C9404CAH
Related information
http://www.hamamatsu.com/sp/ssd/doc_en.html
Precautions
∙ Disclaimer
∙ Mini-spectrometers
Technical information
∙ Mini-spectrometers
Information described in this material is current as of October, 2015.
Product specifications are subject to change without prior notice due to improvements or other reasons. This document has been carefully prepared and the
information contained is believed to be accurate. In rare cases, however, there may be inaccuracies such as text errors. Before using these products, always
contact us for the delivery specification sheet to check the latest specifications.
The product warranty is valid for one year after delivery and is limited to product repair or replacement for defects discovered and reported to us within that
one year period. However, even if within the warranty period we accept absolutely no liability for any loss caused by natural disasters or improper product use.
Copying or reprinting the contents described in this material in whole or in part is prohibited without our prior permission.
www.hamamatsu.com
HAMAMATSU PHOTONICS K.K., Solid State Division
1126-1 Ichino-cho, Higashi-ku, Hamamatsu City, 435-8558 Japan, Telephone: (81) 53-434-3311, Fax: (81) 53-434-5184
U.S.A.: Hamamatsu Corporation: 360 Foothill Road, Bridgewater, N.J. 08807, U.S.A., Telephone: (1) 908-231-0960, Fax: (1) 908-231-1218
Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49) 8152-375-0, Fax: (49) 8152-265-8
France: Hamamatsu Photonics France S.A.R.L.: 19, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: 33-(1) 69 53 71 00, Fax: 33-(1) 69 53 71 10
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road, Welwyn Garden City, Hertfordshire AL7 1BW, United Kingdom, Telephone: (44) 1707-294888, Fax: (44) 1707-325777
North Europe: Hamamatsu Photonics Norden AB: Torshamnsgatan 35 16440 Kista, Sweden, Telephone: (46) 8-509-031-00, Fax: (46) 8-509-031-01
Italy: Hamamatsu Photonics Italia S.r.l.: Strada della Moia, 1 int. 6, 20020 Arese (Milano), Italy, Telephone: (39) 02-93581733, Fax: (39) 02-93581741
China: Hamamatsu Photonics (China) Co., Ltd.: B1201, Jiaming Center, No.27 Dongsanhuan Beilu, Chaoyang District, Beijing 100020, China, Telephone: (86) 10-6586-6006, Fax: (86) 10-6586-2866
8
Cat. No. KACC1131E10 Oct. 2015 DN
相关型号:
C941U101JVSDAA7317
Safety Standard Recognized, 900 Series, Encapsulated, AH Type, X1 400 VAC/Y1 400 VAC (Industrial Grade)
KEMET
C941U101JVSDBA7317
Safety Standard Recognized, 900 Series, Encapsulated, AH Type, X1 400 VAC/Y1 400 VAC (Industrial Grade)
KEMET
C941U101JVSDCA7317
Safety Standard Recognized, 900 Series, Encapsulated, AH Type, X1 400 VAC/Y1 400 VAC (Industrial Grade)
KEMET
C941U101JVVDAA7317
Safety Standard Recognized, 900 Series, Encapsulated, AH Type, X1 400 VAC/Y1 400 VAC (Industrial Grade)
KEMET
C941U101JVVDBA7317
Safety Standard Recognized, 900 Series, Encapsulated, AH Type, X1 400 VAC/Y1 400 VAC (Industrial Grade)
KEMET
C941U101JVVDCA7317
Safety Standard Recognized, 900 Series, Encapsulated, AH Type, X1 400 VAC/Y1 400 VAC (Industrial Grade)
KEMET
©2020 ICPDF网 联系我们和版权申明