Keysight Technologies  
E4990A Impedance Analyzer  
20 Hz to 10/20/30/50/120 MHz  
Data Sheet  
02 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
This document provides technical specifications for the E4990A.  
Options  
The following options are available.  
E4990A-120 20 Hz to 120 MHz  
E4990A-050 20 Hz to 50 MHz  
E4990A-030 20 Hz to 30 MHz  
E4990A-020 20 Hz to 20 MHz  
E4990A-010 20 Hz to 10 MHz  
E4990A-001 Enhanced measurement speed (option 010/020/030/050 only).  
Definitions  
Specification (spec.):  
Warranted performance. All specifications apply at 23 °C (± 5 °C), unless otherwise  
stated, and 90 minutes after the instrument has been turned on. Specifications include  
guard bands to account for the expected statistical performance distribution, measure-  
ment uncertainties, and changes in performance due to environmental conditions.  
Typical (typ.):  
Expected performance of an average unit which does not include guardbands. It is not  
covered by the product warranty.  
General characteristics:  
A general, descriptive term that does not imply a level of performance.  
03 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Basic Measurement Characteristic  
Measurement parameters  
Impedance parameters  
|Z|, θz, |Y|, θy, Cp, Cs, Lp, Ls, Rp, Rs (R), D, Q, X, G, B, Complex Z, Complex Y  
Level monitor  
Vac, Iac, Vdc, Idc  
Measurement terminal  
Configuration  
Four-terminal pair configuration  
Connector type  
Four BNC (female) connectors.  
Option 120: Can be converted to one port terminal using the Keysight Technologies, Inc. 42942A Terminal  
Adapter (7-mm port) or 42941A Impedance Probe (SMA (f) port).  
Source Characteristics  
Frequency  
Range  
20 Hz to 120 MHz (Option 120)  
20 Hz to 50 MHz (Option 050)  
20 Hz to 30 MHz (Option 030)  
20 Hz to 20 MHz (Option 020)  
20 Hz to 10 MHz (Option 010)  
Resolution  
1 mHz  
Accuracy  
without Option 1E5  
± 7 ppm ± 1 mHz (at 23 ± 5 °C)  
± 7 ppm ± 1 mHz (at 5 to 40 °C, typ.)  
with Option 1E5  
Stability  
± 1 ppm ± 1 mHz (at 23 ± 5 °C)  
with Option 1E5  
± 0.5 ppm (at 5 to 40 °C, Typical)  
± 0.5 ppm per year (Typical)  
04 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Voltage Signal Level  
Range  
5 mVrms to 1 Vrms  
1 mV  
Resolution  
Accuracy  
at four-terminal pair port of the  
E4990A or 7-mm port of the 42942A  
± [(10 + 0.05 × f )% + 1 mV]  
± [(15 + 0.1 × f )% + 1 mV]  
at measurement port of the 42941A,  
16048G/H  
NOTE  
f : frequency [MHz].  
These characteristics apply when OPEN is connected to each port.  
Test signal level should be ≤ 0.5 Vrms when the measured impedance is ≤ 50 .  
Beyond 23 ± 5 °C of temperature, test signal level setting accuracy is twice as bad as described.  
Current Signal Level  
Range  
200 µArms to 20 mArms  
Resolution  
20 µA  
Accuracy  
at four-terminal pair port of the E4990A  
at ≤ 15 MHz  
+[10% + 50 µA], –[(10 + 0.2 × f2 )% + 50 µA] (typical)  
at > 15 MHz  
± [(10 + 0.3 × f )% + 50 µA] (typical)  
at 7-mm port of the 42942A  
at ≤ 5 MHz  
+[10% + 50 µA], –[(10 + 1 × f2 )% + 50 µA] (typical)  
at > 5 MHz  
± [(10 + 0.3 × f )% + 50 µA] (typical)  
at measurement port of the 42941A, 16048G/H  
at ≤ 5 MHz  
at > 5 MHz  
+[10% + 50 µA], –[(15 + 1.5 × f2 )% + 50 µA] (typical)  
± [(20 + 0.3 × f )% + 50 µA] (typical)  
NOTE  
f : frequency [MHz].  
These characteristics apply when SHORT is connected to each port.  
Test signal level should be ≤ 20 mArms when the measured impedance is ≤ 50 .  
05 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Signal Level Monitor  
Voltage range  
(Same as the voltage signal level setting range)  
Voltage monitor accuracy  
at four-terminal pair port of the E4990A or  
7-mm port of the 42942A  
± (10 + 0.05 × f + 100/Zx)[%] (typical)  
± (10 + 0.15 × f + 100/Zx)[%] (typical)  
(Same as the current signal level setting range)  
at measurement port of the 42941A,  
16048G/H  
Current range  
Current monitor accuracy  
at four-terminal pair port of the E4990A or  
7-mm port of the 42942A  
± (10 + 0.3 × f + Zx/100)[%] (typical)  
± (10 + 0.4 × f + Zx/100)[%] (typical)  
at measurement port of the 42941A,  
16048G/H  
NOTE  
f : frequency [MHz], Zx: impedance measurement value [Ω].  
Beyond 23 ± 5 °C, the test signal level monitor accuracy is twice as bad as described.  
Output Impedance  
Output impedance  
25 Ω (nominal)  
06 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
DC Bias Function  
DC voltage bias  
Range  
0 to ± 40 V (see Figure 1)  
1 mV  
Resolution  
Accuracy  
± [0.1% + (5 + 30 × |Imon|) mV]  
± [0.2% + (10 + 30 × |Imon|) mV] (beyond 23 ± 5 °C)  
DC current bias  
Range  
0 to ± 100 mA (see Figure 1)  
40 µA  
Resolution  
Accuracy  
± [2% + (0.2 + |Vmon|/20) mA]  
± [4% + (0.4 + |Vmon|/20) mA] (beyond 23 ± 5 °C)  
DC voltage bias at constant voltage mode  
Range  
0 to ± 40 V (see Figure 1)  
1 mV  
Resolution  
Accuracy  
± [0.5% + (5 + Zd × |Imon|) mV] (typical)  
± [1.0% + (10 + Zd × |Imon|) mV] (beyond 23 ± 5 °C, typical)  
DC current bias at constant current mode  
Range  
0 to ± 100 mA (see Figure 1)  
40 µA  
Resolution  
Accuracy  
± [1% + (0.5 + |Vmon|/10000) mA] (typical)  
± [2% + (1.0 + |Vmon|/5000) mA] (beyond 23 ± 5 °C, typical)  
DC bias monitor  
DC voltage range  
DC voltage accuracy  
(Same as the dc voltage bias setting range)  
± [0.2% + (5 + Zd × |Imon|) mV]  
± [0.4% + (10 + Zd × |Imon|) mV] (beyond 23 ± 5 °C)  
DC current range  
(Same as the dc voltage bias setting range)  
DC current monitor accuracy  
± [1% + (0.5 + |Vmon|/10000) mA]  
± [2% + (1.0 + |Vmon|/5000) mA] (beyond 23 ± 5 °C)  
Output impedance  
25 Ω (nominal)  
NOTE  
Vmon : DC voltage bias monitor reading value [mV]  
Imon  
Zd  
Zd  
Zd  
Zd  
Zd  
: DC current bias monitor reading value [mA]  
= 0.3 (at four-terminal pair port of the E4990A, adapter setup: NONE)  
= 2.0 (at test port of the 42941A, adapter setup: 42941A Impedance Probe)  
= 0.5 (at 7-mm port of the 42942A, adapter setup: 42942A Terminal Adapter)  
= 1.0 (at measurement port of the 16048G, adapter setup: four-terminal pair 1 m)  
= 1.5 (at measurement port of the 16048H, adapter setup: four-terminal pair 2 m)  
Figure 1. DC bias range  
07 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Sweep Characteristics  
Available sweep parameters  
Sweep type  
Frequency, signal voltage, signal current, DC bias voltage, DC bias current  
Linear frequency, log frequency, OSC level (voltage, current), DC bias (voltage, current), log DC bias (voltage,  
current)  
Sweep direction  
Up sweep, down sweep  
2 to 1601 points  
Number of measurement points  
Segment sweep  
Available setup parameters for  
measurement points, Signal each  
segment  
Sweep frequency range, number of level (voltage or current), DC bias (voltage or current), measurement time,  
point averaging factor, segment time, segment delay  
Number of segments  
Sweep span type  
Delay time  
1 to 201  
Order base or frequency base  
Type  
Point delay, sweep delay, segment delay or DC bias delay  
Range  
0 sec to 30 sec  
1 msec  
Resolution  
Trigger Function  
Trigger type  
Continuous, single, averaging  
Trigger source  
Trigger event type  
Internal (free run), external (BNC connector input), GPIB/USB/LAN, manual (Front key)  
Point trigger, Sweep trigger  
Measurement Time/Averaging  
Measurement time  
Range  
1 (fast) to 5 (precise), 5 steps  
Averaging  
Type  
Sweep-to-sweep averaging, point averaging  
1 to 999 (integer)  
Averaging factor  
08 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Measurement Time  
Measurement frequency [Hz]  
Figure 2. Measurement time (Option 050/ 030/ 020/ 010 with Option 001 or Option 120, adapter: None, 1 m, 2m, Typical)  
Measurement frequency [Hz]  
Figure 3. Measurement time (Option 120, adapter: 7mm 42942A/probe 42941A, Typical)  
09 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Measurement Time (continued)  
Measurement frequency [Hz]  
Figure 4. Measurement time (Option 050/ 030/ 020/ 010 without option 001, Typical)  
Adapter Setup  
Adapter selection  
NONE  
No adapter (the 16047E, etc. direct connection type test fixture is connected)  
4TP 1M  
Four-terminal pair 1 m (16048G)  
Four-terminal pair 2 m (16048H)  
Terminal adapter (42942A)  
4TP 2M  
7-mm 42942A1  
Probe 42941A1  
Impedance probe (42941A)  
1. Option 120 only.  
Calibration  
Calibration  
User calibration  
Port extension  
Calibration performed with user-defined calibration kit (OPEN, SHORT, LOAD)  
Compensation performed when the measurement terminal is expanded from the 7-mm connector of the  
42942A Terminal Adapter or the test port of the 42941A Impedance Probe. Enter electrical length or delay time  
for the extension.  
Fixture compensation  
Calibration points  
Compensation performed at the device contacts of the test fixture using OPEN, SHORT, LOAD.  
Fixed points, or User points determined by sweep setups  
10 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Measurement Accuracy  
Conditions of accuracy  
Temperature  
Four-terminal pair port of the  
E4990A’s front panel  
23 ± 5 °C  
Beyond 23 ± 5 °C, the measurement accuracy is twice as bad as described.  
Measurement terminal of the  
16048G or 16048H  
Within ± 5 °C from the adapter setup temperature. Measurement accuracy applies when the adapter setup is  
performed at 23 ± 5 °C. When the adapter setup is performed beyond 23 ± 5 °C, the measurement accuracy is  
twice as bad as described.  
7-mm port of the 42942A  
terminal adapter  
Within ± 5 °C from the adapter setup temperature. Measurement accuracy applies when the adapter setup is  
performed at 23 ± 5 °C. When the adapter setup is performed beyond 23 ± 5 °C, the measurement accuracy is  
twice as bad as described.  
Test port of the 42941A  
impedance probe  
Within ± 5 °C from the adapter setup temperature. Measurement accuracy applies when the adapter setup is  
performed at 23 ± 5 °C. When the adapter setup is performed beyond 23 ± 5 °C, the measurement accuracy is  
twice as bad as described.  
Measurement accuracy  
|Z|, |Y| accuracy  
θ accuracy  
± E [%] (see equation 1 on page 11, Equation 2 on page 14, equation 3 on page 17)  
± E/100 [rad]  
L, C, X, B accuracy  
at Dx ≤ 0.1  
± E [%]  
at Dx > 0.1  
± E × 1 + D2x [%]  
R accuracy  
at Dx ≤ 0.1 (Qx ≥ 10)  
E
Rp: ±  
[%]  
D
E/100  
x
Rs: ± E/Dx [%]  
at 0.1 < Dx < 10  
(0.1 < Qx < 10)  
1 + D2x  
1 + D2x  
Rp: ± E ×  
Dx  
[%]  
E
1 + D2x  
Rs: ± E ×  
[%]  
×
Dx  
100  
,
at Dx ≥ 10 (Qx ≤ 0.1)  
D accuracy  
± E [%]  
at Dx ≤ 0.1  
± E/100  
at 0.1 < Dx ≤ 1  
± E × (1 + Dx)/100  
Q accuracy (at Qx × Da < 1)  
at Qx ≤ 10 (Dx ≥ 0.1)  
Q2x × E (1 + Dx )/100  
1 Qx × E (1 + Dx )/100  
±
±
Q2x × E/100  
1 Qx × E/100  
at Qx > 10 (Dx < 0.1)  
G accuracy  
at Dx > 0.1  
1 + D2x  
± E ×  
[%]  
Dx  
at Dx ≤ 0.1  
± E/Dx [%]  
NOTE  
Dx : measurement value of D.  
Qx : measurement value of Q.  
Da : measurement accuracy of D.  
Under an AC magnetic field, the following equation is applied to the measurement accuracy.  
E × ( 1 + B × ( 5 + 500/Vmv )) [%] (Typical)  
B
: Magnetic flux density [Gauss]  
11 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Impedance measurement accuracy at four-terminal pair port  
Equation 1 shows the impedance measurement accuracy [%] at four-terminal pair port of the E4990A, or measurement port of the  
16048G/16048H.  
Equation 1. Impedance measurement accuracy [%] at four-terminal pair port (Typical at > 10 MHz)  
Zs'  
Zx  
E = Ep' +  
+ Yo' × Zx × 100  
Where,  
Ep' = Epl + Epbw + Eposc + Ep [%]  
Yo' = Yol + Kbw × Kyosc × (Yodc + Yo) [S]  
Zs' = Zsl + Kbw × Kzosc × Zs [Ω]  
1m: 0.02 + 2 × f/100  
2m: 0.02 + 3 × f/100  
Epl [%]  
Meas Time 5: 0  
Meas Time 4: 0.06 at < 50 kHz, 0.03 at ≥ 50 kHz  
Meas Time 3: 0.2 at < 50 kHz, 0.1 at ≥ 50 kHz  
Meas Time 2: 0.4 at < 50 kHz, 0.2 at ≥ 50 kHz  
Meas Time 1: 0.8 at < 50 kHz, 0.4 at ≥ 50 kHz  
E
E
pbw [%]  
Vosc > 500 mV: 0.018 × (1000/Vmv-1) + f/100  
200 mV < Vosc ≤ 500 mV: 0.018 × (500/Vmv-1)  
100 mV < Vosc ≤ 200 mV: 0.018 × (200/Vmv-1)  
posc [%]  
V
osc ≤ 100 mV: (100/Vmv-1) × (0.018 + Epbw)  
20 Hz ≤ fm < 100 Hz: 0.5  
100 Hz ≤ fm ≤ 800 Hz: 0.3  
800 Hz < fm ≤ 1 MHz: 0.075  
1 MHz < fm ≤ 15 MHz: 0.1 × f  
15 MHz < fm ≤ 120 MHz: 1.5  
Ep [%]  
Yol [S]  
Kbw  
1 m (16048G): 500n × f/100  
2 m (16048H): 1µ × f/100  
Meas Time 5: 1  
Meas Time 4: 1  
Meas Time 3: 3 at ≤ 1 MHz, 4 at > 1 MHz  
Meas Time 2: 4 at ≤ 1 MHz, 5 at > 1 MHz  
Meas Time 1: 6 at ≤ 1 MHz, 10 at > 1 MHz  
Vosc > 500 mV: 1000/Vmv  
Kyosc  
V
osc ≤ 500 mV: 500/Vmv  
DCI range 1 mA: 0  
DCI range 10 mA: 1 µ  
DCI range 100 mA: 10 µ  
Yodc [S]  
12 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Impedance measurement accuracy at four-terminal pair port (continued)  
20 Hz ≤ fm < 100 Hz: 10 n  
100 Hz ≤ fm ≤ 200 kHz: 2.5 n  
Yo [S]  
200 kHz < fm ≤ 1 MHz: 5 n  
1 MHz < fm ≤ 15 MHz: 50 n  
15 MHz < fm ≤ 120 MHz: 500 n  
0 m: 0  
1 m (16048G), 2 m (16048H):  
20 Hz ≤ fm < 500 Hz: 5 m  
500 Hz ≤ fm ≤ 120 MHz: 2 m  
Zsl [Ω]  
Meas Time 5: 1  
Meas Time 4: 1  
Kbw  
Meas Time 3: 3 at ≤ 1 MHz, 4 at > 1 MHz  
Meas Time 2: 4 at ≤ 1 MHz, 5 at > 1 MHz  
Meas Time 1: 6 at ≤ 1 MHz, 10 at > 1 MHz  
Vosc > 500 mV: 2  
200 mV < Vosc ≤ 500 mV: 1  
100 mV < Vosc ≤ 200 mV: 200/Vmv  
Kzosc  
V
osc ≤ 100 mV: 100/Vmv  
20Hz ≤ fm < 100 Hz: 10 m  
100Hz ≤ fm ≤ 120 MHz: 2.5 m  
Zs [Ω]  
NOTE  
fm : measurement frequency  
f
: frequency [MHz]  
Vosc : oscillator level  
Vmv : Vosc [mV]  
13 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Impedance measurement accuracy at four-terminal pair port (continued)  
Figure 5. Impedance measurement accuracy at four-terminal pair port of the E4990A's  
front panel (Oscillator level = 0.5 Vrms), measurement time = 5 (Typical at > 10 MHz)  
14 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Impedance measurement accuracy at 7-mm port of the Keysight 42942A  
Equation 2 shows the impedance measurement accuracy [%] at 7-mm port of the 42942A terminal adapter.  
Equation 2. Impedance Measurement Accuracy [%] at 7-mm port of E4990A  
Zs'  
Zx  
E = Ep' +  
+ Yo' × Zx × 100  
Where,  
Ep' = Epl + Epbw + Eposc + Ep [%]  
Yo' = Yol + Kbw × Kyosc × (Yodc + Yo) [S]  
Zs' = Zsl + Kbw × Kzosc × Zs [Ω]  
Epl [%]  
0
Meas Time 5: 0  
Meas Time 4: 0.06 at < 50 kHz, 0.03 at ≥ 50 kHz  
Meas Time 3: 0.2 at < 50 kHz, 0.1 at ≥ 50 kHz  
Meas Time 2: 0.4 at < 50 kHz, 0.2 at ≥ 50 kHz  
Meas Time 1: 0.8 at < 50 kHz, 0.4 at ≥ 50 kHz  
Epbw [%]  
Vosc > 500 mV: f/100 × (Vmv/500-1)  
E
posc [%]  
100 mV < Vosc ≤ 500 mV: 0  
V
osc ≤ 100 mV: (100/Vmv-1) × (0.05 + Epbw)  
20 Hz ≤ fm ≤ 15 MHz: 0.6  
15 MHz < fm ≤ 120 MHz: 0.95  
Ep [%]  
Yol [S]  
0
Meas Time 5: 1  
Meas Time 4: 1  
Meas Time 3: 3  
Meas Time 2: 4  
Meas Time 1: 6  
Kbw  
Vosc ≥ 500 mV: 1  
Kyosc  
V
osc < 500 mV: 500/Vmv  
DCI range 1 mA: 0  
Yodc [S]  
DCI range 10 mA: 1 µ  
DCI range 100 mA: 10 µ  
15 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Impedance measurement accuracy at 7-mm port of the Keysight 42942A (continued)  
20 Hz ≤ fm < 100 Hz: 100 n  
100 Hz ≤ fm ≤ 200 kHz: 25 n1  
Yo [S]  
200 kHz < fm ≤ 1 MHz: 50 n1  
1 MHz < fm ≤ 120 MHz: 5 µf/100 + 50 n1  
Zsl [Ω]  
0
Meas Time 5: 1  
Meas Time 4: 1  
Meas Time 3: 3  
Meas Time 2: 4  
Meas Time 1: 6  
Kbw  
Vosc > 500 mV: 2 + f/100  
200 mV < Vosc ≤ 500 mV: 1  
100 mV < Vosc ≤ 200 mV: 200/Vmv  
Kzosc  
V
osc ≤ 100 mV: 100/Vmv  
20 Hz ≤ fm < 100 Hz: 20 m  
100 Hz ≤ fm ≤ 120 MHz: 5 m + 50 m×f/100  
Zs [Ω]  
1. The specification might not be met at the following range due to internal spurious response.  
± 10% range of the following frequencies.  
110 kHz, 170 kHz, 220 kHz, 340 kHz, 510 kHz, 600 kHz, 680 kHz, 850 kHz, 1200 kHz.  
± 2% range of the following frequencies.  
109 kHz × N (N = 12 to 89)  
118 kHz × M (M = 11 to 83)  
16 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Impedance measurement accuracy at 7-mm port of the Keysight 42942A (continued)  
Figure 6. Impedance measurement accuracy at 7-mm port of the Keysight 42942A termi-  
nal adapter connected to the E4990A (Oscillator level = 0.5 Vrms), measurement time = 5  
17 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Impedance measurement accuracy at test port of the Keysight 42941A  
Equation 3 shows the impedance measurement accuracy [%] at test port of the 42941A impedance probe.  
Equation 3. Impedance measurement accuracy [%] at test port of the Keysight 42941A  
Zs'  
Zx  
E = Ep' +  
+ Yo' × Zx × 100  
Where,  
Ep' = Epl + Epbw + Eposc + Ep [%]  
Yo' = Yol + Kbw × Kyosc × (Yodc + Yo) [S]  
Zs' = Zsl + Kbw × Kzosc × Zs [Ω]  
Epl [%]  
0
Meas Time 5: 0  
Meas Time 4: 0.06 at < 50 kHz, 0.03 at ≥ 50 kHz  
Meas Time 3: 0.2 at < 50 kHz, 0.1 at ≥ 50 kHz  
Meas Time 2: 0.4 at < 50 kHz, 0.2 at ≥ 50 kHz  
Meas Time 1: 0.8 at < 50 kHz, 0.4 at ≥ 50 kHz  
Epbw [%]  
Vosc > 500 mV: f/1001 (Vmv/500-1)  
E
posc [%]  
100 mV < Vosc ≤ 500 mV: 0  
V
osc ≤ 100 mV: (100/Vmv-1) × (0.05 + Epbw)  
20 Hz < = fm ≤ 15 MHz: 0.8  
15 MHz < fm ≤ 120 MHz: 1.5  
Ep [%]  
Yol [S]  
0
Meas Time 5: 1  
Meas Time 4: 1  
Meas Time 3: 3  
Meas Time 2: 4  
Meas Time 1: 6  
Kbw  
Vosc ≥ 500 mV: 1  
Kyosc  
V
osc < 500 mV: 500/Vmv  
DCI range 1 mA: 0  
Yodc [S]  
DCI range 10 mA: 1 µ  
DCI range 100 mA: 10 µ  
18 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Impedance measurement accuracy at test port of the Keysight 42941A (continued)  
20 Hz ≤ fm < 100 Hz: 100 n  
100 Hz ≤ fm ≤ 200 kHz: 25 n1  
Yo [S]  
200 kHz < fm ≤ 1 MHz: 50 n1  
1 MHz < fm ≤ 120 MHz: 20 µ × f/1001  
Zsl [Ω]  
0
Meas Time 5: 1  
Meas Time 4: 1  
Meas Time 3: 3  
Meas Time 2: 4  
Meas Time 1: 6  
Kbw  
Vosc > 500 mV: 2 + f/100  
200 mV < Vosc ≤ 500 mV: 1  
100 mV < Vosc ≤ 200 mV: 200/Vmv  
Kzosc  
V
osc ≤ 100 mV: 100/Vmv  
20 Hz ≤ fm < 100 Hz: 20 m  
100 Hz ≤ fm ≤ 120 MHz: 5 m + 100 m × f/100  
Zs [Ω]  
1. The specification might not be met at the following range due to internal spurious response.  
± 10% range of the following frequencies.  
110 kHz, 170 kHz, 220 kHz, 340 kHz, 510 kHz, 600 kHz, 680 kHz, 850 kHz, 1200 kHz.  
± 2% range of the following frequencies.  
109 kHz × N (N = 12 to 89)  
118 kHz × M (M = 11 to 83)  
19 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Impedance measurement accuracy at test port of the Keysight 42941A (continued)  
Figure 7. Impedance measurement accuracy at test port of the Keysight 42941A impedance probe  
connected to the E4990A (Oscillator level = 0.5 Vrms), measurement time = 5  
20 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Display Function  
Display  
Size/type  
Number or pixels1  
10.4 inch color LCD (TFT)  
1024 X 768 (XGA)  
Scale type  
X axis scale  
Linear and log  
Y axis scale  
Linear and log (depends on the sweep type)  
Number of traces  
Data trace  
4 traces per channel  
Memory trace  
Data math function  
4 traces per channel  
Data + memory, data - memory, data × memory, data/memory, offset, equation editor  
1. Valid pixels are 99.99% and more. Below 0.01% of fixed points of black, blue, green or red are not regarded as failure.  
Marker function  
Marker type and number  
10 independent markers per trace. Reference marker available for delta marker operation.  
Marker search  
Search type  
Maximum value, minimum value, multi-peak, multi-target, peak, peak left, peak right, target, target left, target  
right, and width parameters with user defined bandwidth values.  
Search track  
Search range  
Marker X-axis display  
Others  
Performs search by each sweep  
User define  
Sweep parameter value, sweep elapsed time, or relaxation time (1/( 2πf))  
Marker continuous mode, ∆ marker mode, marker coupled mode,  
Marker value substitution (Marker→), marker zooming, marker table, marker statistics  
Equivalent circuit analysis  
Circuit model  
3-component model (4 models), 4-component model (3 models)  
Analysis type  
Equivalent circuit parameters calculation, frequency characteristics simulation  
Limit Line Test  
Funcions  
Define the test limit lines that appear on the display for pass/fail testing. Defined limits may be any combina-  
tion of horizontal/sloping lines and discrete data points.  
Other functions  
Beep fail, Limit line offset  
21 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Interface  
GPIB  
24-pin D-Sub (Type D-24), female; compatible with IEEE-488.  
IEEE-488 interface specification is designed to be used in environment where electrical noise is relatively low.  
LAN or USBTMC interface is recommended to use at the higher electrical noise environment.  
USB host port  
Universal serial bus jack, Type A configuration; female; provides connection to mouse, key board, printer or  
USB stick memory.  
USB (USBTMC ) interface port  
Universal serial bus jack, Type B configuration (4 contacts inline); female; provides connection to an external  
PC; compatible with USBTMC-USB488 and USB 2.0.LA  
USB Test and Measurement Class (TMC) interface that communicates over USB, complying with the IEEE 488.1  
and IEEE 488.2 standards.  
LAN  
10/100/1000 Base T Ethernet, 8-pin configuration; auto selects between the two data rates  
15-pin mini D-Sub; female; drives VGA compatible monitors  
Video output  
Handler interface  
Connector type  
Signal type  
36-pin centronics, female  
Negative logic, opto-isolated, open collector output  
See the following figure. Refer to Help for the definition of each pin.  
Pin location  
18 17  
15 14  
13 12 11 10  
16  
9
8
7
6
5
4
3
2
1
36 35 34 33 32 31 30 29 28 27 26 25 24 23 22 21 20 19  
Figure 8. 24 bit I/O port pin assignment  
22 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Table 1. 24 bit I/O port pin assignment  
Pin No.  
1
Signal Name  
Signal Standard  
GND  
0 V  
2
INPUT1  
TTL level, pulse input, pulse width: 1 µs or above  
TTL level, latch output  
3
OUTPUT1  
4
OUTPUT2  
TTL level, latch output  
5
Output port A0  
TTL level, latch output  
6
Output port A1  
TTL level, latch output  
7
Output port A2  
TTL level, latch output  
8
Output port A3  
TTL level, latch output  
9
Output port A4  
TTL level, latch output  
10  
11  
12  
13  
14  
15  
16  
17  
18  
19  
20  
21  
22  
23  
24  
25  
26  
27  
28  
29  
30  
31  
32  
33  
34  
35  
36  
Output port A5  
TTL level, latch output  
Output port A6  
TTL level, latch output  
Output port A7  
TTL level, latch output  
Output port B0  
TTL level, latch output  
Output port B1  
TTL level, latch output  
Output port B2  
TTL level, latch output  
Output port B3  
TTL level, latch output  
Output port B4  
TTL level, latch output  
Output port B5  
TTL level, latch output  
Output port B6, index (selectable)  
Output port B7, ready for trigger (selectable)  
Input/output port C0  
Input/output port C1  
Input/output port C2  
Input/output port C3  
Input/output port D0  
Input/output port D1  
Input/output port D2  
Input/output port D3  
Port C status  
TTL level, latch output  
TTL level, latch output  
TTL level, latch output  
TTL level, latch output  
TTL level, latch output  
TTL level, latch output  
TTL level, latch output  
TTL level, latch output  
TTL level, latch output  
TTL level, latch output  
TTL level, input mode; LOW, output mode: HIGH  
TTL level, input mode; LOW, output mode: HIGH  
TTL level, active low, pulse output (width: 10 µs, Typical)  
Port D status  
Write strobe signal  
+5 V pullup  
SWEEP END signal  
+5 V  
TTL level, active low, pulse output (width: 20 µs, Typical)  
+5 V, 100 mA MAX  
PASS/FAIL signal  
PASS/FAIL write strobe signal  
TTL level, PASS: HIGH, FAIL; LOW, latch output  
TTL level, active low, pulse output (width: 10 µs, Typical)  
23 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Analyzer Environmental Specifications  
Measurement circuit protection  
Max discharge withstand voltage1  
1000 V @ C < 2 µF, (2/C) V @ C ≥ 2 µF  
NOTE: Discharge capacitors before connecting them to the UNKNOWN terminal or a test fixture.  
Operating environment  
Temperature  
Humidity  
Altitude  
+5 °C to +40 °C  
20% to 80% at wet bulb temperature < +29 °C (non-condensation)  
0 to 2,000 m (0 to 6561 feet)  
Vibration  
0.21 Grms maximum, 5 Hz to 500 Hz  
Non-operating environment  
Temperature  
Humidity  
Altitude  
–10 °C to +60 °C  
20% to 90% at wet bulb temperature < +40 °C (non-condensation)  
0 to 4,572 m (0 to 15,000 feet)  
Vibration  
2.09 Grms maximum, 5 to 500 Hz  
1. The maximum discharge withstand voltage, where the internal circuit remains protected if a charged capacitor is connected to the UNKNOWN terminal.  
24 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
General Characteristics  
External reference input  
Frequency  
10 MHz ± 10 ppm (Typical)  
0 dBm to ± 3 dB (Typical)  
50 Ω (nominal)  
Level  
Input impedance  
Connector type  
BNC (female)  
Internal reference output  
Frequency  
10 MHz ± 7 ppm (Typical)  
0 dBm ± 3 dB into 50 Ω (Typical)  
50 Ω (nominal)  
Level  
Output impedance  
Connector type  
BNC (female)  
High stability frequency reference output (Option 1E5)  
Frequency  
10 MHz ± 1 ppm  
0 dBm minimum  
50 Ω (nominal)  
BNC (female)  
Level  
Output impedance  
Connector type  
External trigger input  
Level  
TTL  
Pulse width (Tp)  
Polarity  
≥ 2 µs (Typical); see Figure 9 for the definition of Tp.  
Positive or negative (selective)  
BNC (female)  
Connector type  
Tp  
Tp  
Tp  
Tp  
5V  
0V  
5V  
0V  
Positive Trigger Signal  
Negative Trigger Signal  
Figure 9. Required pulse width (Tp) for external trigger input  
Line power  
Frequency  
47 to 63 Hz  
Voltage  
90-264 VAC (Vpeak > 120 V)  
300 VA max.  
VA max  
Power consumption  
160 W1 (Typical)  
1. At preset condition. No application running other than the E4990A on windows.  
25 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
EMC, safety, environment and compliance  
Description  
Specification  
EMC  
European Council Directive 2004/108/EC  
IEC 61326-1:2012  
EN 61326-1:2013  
CISPR 11:2009 +A1:2010  
EN 55011: 2009 +A1:2010  
Group 1, Class A  
IEC 61000-4-2:2008  
EN 61000-4-2:2009  
4 kV CD / 8 kV AD  
IEC 61000-4-3:2006 +A1:2007 +A2:2010  
EN 61000-4-3:2006 +A1:2008 +A2:2010  
3 V/m, 80-1000 MHz, 1.4 - 2.0 GHz / 1V/m, 2.0 - 2.7 GHz, 80% AM  
IEC 61000-4-4:2004 +A1:2010  
EN 61000-4-4:2004 +A1:2010  
1 kV power lines / 0.5 kV signal lines  
IEC 61000-4-5:2005  
EN 61000-4-5:2006  
0.5 kV line-line / 1 kV line-ground  
IEC 61000-4-6:2008  
EN 61000-4-6:2009  
3 V, 0.15-80 MHz, 80% AM  
IEC 61000-4-8:2009  
EN 61000-4-8:2010  
30A/m, 50/60Hz  
IEC 61000-4-11:2004  
EN 61000-4-11:2004  
0.5-300 cycle, 0% / 70%  
NOTE-1:  
When tested at 3 V/m according to EN61000-4-3, the measurement accuracy will be within specifications  
over the full immunity test frequency range except when the analyzer frequency is identical to the transmitted  
interference signal test frequency.  
NOTE-2:  
When tested at 3 V according to EN61000-4-6, the measurement accuracy will be within specifications over the  
full immunity test frequency range except when the analyzer frequency is identical to the transmitted interfer-  
ence signal test frequency.  
ICES-001:2006 Group 1, Class A  
AS/NZS CISPR11:2004  
Group 1, Class A  
KN11, KN61000-6-1 and KN61000-6-2  
Group 1, Class A  
26 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Safety  
European Council Directive 2006/95/EC  
IEC 61010-1:2010 / EN 61010-1:2010  
Measurement Category I  
Pollution Degree 2  
Indoor Use  
CAN/CSA C22.2 No. 61010-1-12  
Measurement Category I  
Pollution Degree 2  
Indoor Use  
Environment  
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates  
that you must not discard this electrical/electronic product in domestic household waste.  
Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed  
as a “Monitoring and Control instrumentation” product.  
Do not dispose in domestic household waste.  
To return unwanted products, contact your local Keysight office, or see  
http://www.keysight.com/environment/product/ for more information.  
Compliance  
Class C  
27 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Dimensions, weight  
Weight  
14 kg  
Dimensions  
See Figures 10 through 12  
Figure 10. Front view (Millimeters)  
28 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Figure 11. Rear view (Millimeters)  
Figure 12. Side view (Millimeters)  
29 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
Additional Information  
Websites  
Have access to the following website to acquire the latest news, product and support  
information, application literature and more.  
www.keysight.com/find/impedance  
www.keysight.com/find/e4990a  
Literature  
E4990A, Brochure, 5991-3888EN  
E4990A, Configuration Guide, 5991-3891EN  
LCR Meters, Impedance Analyzers and Test Fixtures, Selection Guide, 5952-1430E  
Accessories Catalog for Impedance Measurements, 5965-4792E  
Impedance Measurement Handbook, 5950-3000  
Power of Impedance Analyzer - Comparison to Network Analyzer, 5992-0338EN  
30 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet  
For more information on Keysight  
Evolving  
Technologies’ products, applications or  
services, please contact your local Keysight  
office. The complete list is available at:  
www.keysight.com/find/contactus  
Our unique combination of hardware, software, support, and people can help  
you reach your next breakthrough. We are unlocking the future of technology.  
Americas  
Canada  
Brazil  
Mexico  
United States  
(877) 894 4414  
55 11 3351 7010  
001 800 254 2440  
(800) 829 4444  
Asia Pacific  
Australia  
China  
Hong Kong  
India  
Japan  
Korea  
Malaysia  
Singapore  
Taiwan  
1 800 629 485  
800 810 0189  
800 938 693  
1 800 11 2626  
0120 (421) 345  
080 769 0800  
1 800 888 848  
1 800 375 8100  
0800 047 866  
(65) 6375 8100  
From Hewlett-Packard to Agilent to Keysight  
myKeysight  
www.keysight.com/find/mykeysight  
A personalized view into the information most relevant to you.  
Other AP Countries  
Keysight Infoline  
Europe & Middle East  
Austria  
www.keysight.com/find/Infoline  
0800 001122  
0800 58580  
0800 523252  
0805 980333  
0800 6270999  
1800 832700  
1 809 343051  
800 599100  
+32 800 58580  
0800 0233200  
8800 5009286  
800 000154  
0200 882255  
0800 805353  
Opt. 1 (DE)  
Keysight’s insight to best in class information management. Free access to  
your Keysight equipment company reports and e-library.  
Belgium  
Finland  
France  
Germany  
Ireland  
Israel  
Keysight Services  
www.keysight.com/find/services  
Our deep offering in design, test, and measurement services deploys an  
industry-leading array of people, processes, and tools. The result? We help  
you implement new technologies and engineer improved processes that  
lower costs.  
KEYSIGHT  
SERVICES  
Italy  
Luxembourg  
Netherlands  
Russia  
Spain  
Sweden  
Three-Year Warranty  
www.keysight.com/find/ThreeYearWarranty  
Switzerland  
Keysight’s committed to superior product quality and lower total cost  
of ownership. Keysight is the only test and measurement company with  
three-year warranty standard on all instruments, worldwide. And, we provide  
a one-year warranty on many accessories, calibration devices, systems and  
custom products.  
Opt. 2 (FR)  
Opt. 3 (IT)  
0800 0260637  
United Kingdom  
Keysight Assurance Plans  
For other unlisted countries:  
www.keysight.com/find/AssurancePlans  
Up to ten years of protection and no budgetary surprises to ensure your  
instruments are operating to specification, so you can rely on accurate  
measurements.  
www.keysight.com/find/contactus  
(BP-06-08-16)  
Keysight Channel Partners  
www.keysight.com/find/channelpartners  
DEKRA Certified  
ISO9001 Quality Management System  
Get the best of both worlds: Keysight’s measurement expertise and product  
breadth, combined with channel partner convenience.  
www.keysight.com/go/quality  
Keysight Technologies, Inc.  
DEKRA Certified ISO 9001:2015  
Quality Management System  
www.keysight.com/find/e4990a  
This information is subject to change without notice.  
© Keysight Technologies, 2014 - 2016  
Published in USA, September 6, 2016  
5991-3890EN  
www.keysight.com  

相关型号:

E4990A-001

Impedance Analyzer
KEYSIGHT

E4991A

RF Impedance/Material Analyzer
AGILENT

E4991A-001

RF Impedance/Material Analyzer
HP

E4991A-002

RF Impedance/Material Analyzer
AGILENT

E4991A-007

RF Impedance/Material Analyzer
AGILENT

E4991A-010

RF Impedance/Material Analyzer
AGILENT

E4991A-1D5

RF Impedance/Material Analyzer
AGILENT

E4991B

Impedance Analyzer
KEYSIGHT

E4991B-002

Impedance Analyzer
KEYSIGHT

E4991B-007

Impedance Analyzer
KEYSIGHT

E4991B-010

Impedance Analyzer
KEYSIGHT

E4A-3K

内藏电源超声波开关(反射型)
ETC