M9217A_17 [KEYSIGHT]

PXIe Single-Slot 2-Channel Isolated Digitizer;
M9217A_17
型号: M9217A_17
厂家: Keysight Technologies    Keysight Technologies
描述:

PXIe Single-Slot 2-Channel Isolated Digitizer

文件: 总8页 (文件大小:1163K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
Keysight M9217A  
PXIe Single-Slot 2-Channel Isolated Digitizer  
Data Sheet  
Introduction  
The Keysight Technologies, Inc. M9217A is a PXI Express based, single-slot 2-channel, isolated  
digitizer that supports up to ±256 V input voltage at 20 MSa/s sampling rate with 16-bit resolution.  
With its built-in peak voltage protection of up to 400 V, the M9217A is ideal for analyzing high-voltage  
and transient signals as seen in automotive, mechatronics and aerospace defense applications.  
Key Features  
– ±256 V input voltage  
– 2-channel isolated, with ADCs per channel, simultaneously sampled channels  
– 20 MSa/s sampling rate at 16-bit resolution  
– Built-in 400 V peak voltage protection  
– Single PXI express slot  
– PXI trigger types supported: PXI TTL / trigger and PXI Star trigger  
– IVI-COM, IVI-C, LabVIEW G drivers  
03 | Keysight | M9217A PXIe Single-Slot 2-Channel Isolated Digitizer - Data Sheet  
±256 V input voltage  
Input channels with the ability to measure waveforms up to 256 V are ideal for analyzing  
high-voltage and transient signals as seen in automotive and aerospace defense  
applications. Most oscilloscopes and PXI digitizers have a maximum input range less  
than 40 Vpk. This also reduces the need to add expensive input signal attenuation and  
signal conditioning circuitry, saving test development time and money.  
2-channel isolated, with ADCs per channel  
In addition, since the ±256 V input voltage range is combined with 16-bit analog-to-  
digital converters (ADCs), the isolated front-end and low input offset allows voltages  
from as low as 250 mV to as high as 256 V to be measured simultaneously.  
The M9217A’s individually isolated 2-channel inputs have been designed for high  
performance with an A/D converter per channel to ensure the signals you measure  
are accurately digitized without distortion or additional noise. Channel input range is  
configurable from ±250 mV up to ±256 V, with a floating voltage up to ±40 V to accom-  
modate differential waveform acquisition.  
20 MSa/s sampling rate at 16-Bit resolution  
The M9217A’s ability to support up to 20 MSa/s sampling rate at 16-bit resolution  
enables sampling of transient waveforms like injector waveforms (Figure 1).  
Vpk  
toff  
tpw  
ton ꢀ 0.1 us  
dson ꢁ 600 mV - 800 mV  
V
toff ꢁ 10 us  
Vpk ꢁ 70 V - 146 V  
200 V (peak)  
Ringing  
ton  
Vdson  
Figure 1. Sample injector waveform  
04 | Keysight | M9217A PXIe Single-Slot 2-Channel Isolated Digitizer - Data Sheet  
Built-in 400 V peak voltage protection  
To support high voltage and transient signals in automotive, mechatronics and aerospace  
and defense applications, the M9217A has a built-in peak voltage protection against  
overvoltage spikes of up to 400 V.  
32 MSa/channel with up to 1024 segmented memory  
Coupled with the memory segmentation feature of up to 1024 records, the M9217A’s  
32M sample memory size enables multiple bursts of data to be digitized without the  
need for re-initialization between bursts. The record size is configured by selecting the  
total number of samples, including pre-trigger samples.  
Support for multiple trigger types  
The M9217A supports different trigger types like PXI trigger bus, PXI Star, external input/  
output, channel and software trigger. Configurable trigger delay and trigger hold-off  
allow you to better define where record data is collected relative to the trigger event.  
The configurable trigger delay feature allows precise positioning of acquisition relative to  
the trigger event, while the trigger hold-off feature helps avoid false triggers.  
The PXI TTL trigger output enables the M9217A digitizer to synchronize to other devices.  
Multiple M9217A digitizers can be synchronized for higher channel count.  
IO Libraries  
Keysight input/output or IO Libraries Suite offers fast and easy connection to instru-  
ments. The latest suite version extends that capability to include modular instruments.  
New support for PXI such as the Keysight IO Libraries Suite helps you display all of  
the modules in your system. From here, you can view information about the installed  
software or start the soft front panel of the module. Launch the soft front panel of the  
module directly from Keysight Connection Expert. Figure 2 shows the M9217A soft front  
panel.  
Figure 2. M9217A soft front panel  
05 | Keysight | M9217A PXIe Single-Slot 2-Channel Isolated Digitizer - Data Sheet  
Drivers  
The M9217A digitizer is supplied with a comprehensive portfolio of module drivers, doc-  
umentation, examples, and software tools to help you quickly develop test systems with  
your software platform of choice. The module comes with IVICOM, IVI-C, and LabVIEW  
G software drivers that work in the most popular test and measurement development  
environments, including LabVIEW and LabWindows/CVI from National Instruments,  
Microsoft C/C++, C#, and VB.NET.  
Easy software integration  
The module software support provides context-sensitive help, complete documentation  
and code examples that allow a quick module set up and basic acquisition functionalities.  
These code examples can be easily modified, so that the card can be quickly integrated  
into a test system. Included are application code examples for LabVIEW, LabWindows/  
CVI, Visual Studio C, C++, and C# and Visual Basic.  
Calibration Intervals  
The M9217A is factory-calibrated and shipped with a calibration certificate. Calibration is  
recommended every year in order to verify product performance.  
M9217A specifications  
2-channel isolated digitizer with ADCs per channel and utilizing one PXI express slot.  
Max sample rate  
20 MSa/s  
Sample resolution  
16 bits  
Input configuration  
Isolated inputs ( each channel independently isolated)  
+40 V  
Isolation voltage (low to chassis)  
Maximum input range (high to low) +256 Vpk  
Input impedance  
Input coupling  
1 MW // 40 pF  
DC or AC  
< 10 Hz  
AC cutoff frequency  
Input ranges  
±256 V, ±128 V, ±64 V, ±32 V, ±16 V, ±8 V,  
±4V, ±2 V, ±1 V, ±500 mV, ±250 mV  
Over voltage protection  
Yes  
Maximum overvoltage protection  
Analog bandwidth (-3 dB)  
±400 Vppk  
20 MHz typical  
06 | Keysight | M9217A PXIe Single-Slot 2-Channel Isolated Digitizer - Data Sheet  
Trigger  
Each trigger event causes acquisition of data into a single record at the configured sample rate. The  
number of data records is configurable from 1 to 1024.  
Source  
Description  
Immediate  
Trigger at initiation of acquisition,  
External  
SMD connector, TTL input edge (rising edge)  
PXI connector XJ4 , TTL input edge (rising edge)  
PXIe connector XJ3, TTL input edge (rising edge)  
Instrument command  
PXI_Trig[0:7], PXI_STAR  
DSTARB  
Software  
Channel/edge  
Channel/window  
Selectable level, rising/falling, hysteresis.  
Selectable high and low levels, leaving/entering.  
Sampling  
Programmable sample rates  
1 kSa/s, 2 kSa/s, 5 kSa/s, 10 kSa/s, 50 kSa/s, 100 kSa/s, 200 kSa/s,  
500 kSa/s, 1 MSa/s, 2 MSa/s, 5 MSa/s, 10 MSa/s, 20 MSa/s  
External event output  
Output signal  
Trigger  
TTL ( rising edge )  
25 or 50 Ω  
1 µs  
Impedance  
Pulse width  
Trigger modes  
Record length  
Pre trigger  
2 to 32 million samples  
0 to record length -1  
Record length-pre-trigger  
Continuous running timestamp  
10 ns  
Post trigger  
Timestamp triggered event  
Timestamp resolution  
Trigger delay  
0 – 3600 s with 50 ns resolution  
0 – 10 s with 50 ns resolution  
50 ns  
Trigger hold-off  
Trigger latency1  
1. Latency between edge trigger detection and first trigger sample.  
Accuracy  
DC accuracy  
Total specification (% of reading + % of range)  
23 °C ±5%  
Range  
±% of reading  
±% of range  
250 mV  
0.1  
0.1  
0.1  
0.1  
0.1  
0.1  
0.3  
0.2  
500 mV  
1 V, 2 V  
0.12  
0.3  
4 V, 64 V  
8 V, 128 V  
0.2  
16 V, 32 V, 256 V  
100,000 reading average @ 1 MSa/s  
Integral nonlinearity  
Differential nonlinearity  
0.12  
±5 LSB  
±1 LSB  
07 | Keysight | M9217A PXIe Single-Slot 2-Channel Isolated Digitizer - Data Sheet  
Dynamic characteristic (typical)  
Noise power ( typical)  
Range  
AC flatness (DC-4 MHz) typical  
Source  
250 mV  
500 mV  
1 V  
SFDR dBc  
THD dBc  
Noise voltage1  
Vrms  
Equivalent2  
S/N dB  
250 mV, 500 mV,  
< 0.3 dB relative to 20 kHz  
1 V, 2 V  
71  
77  
81  
85  
70  
79  
250 mV  
0.5 mV  
1 V  
0.5 mV  
0.55 mV  
0.6 mV  
0.9 mV  
4 mV  
51  
55  
60  
63  
56  
58  
61  
65  
56  
58  
62  
4 V, 8 V, 16 V, 32 V < 0.5 dB relative to 20 kHz  
83  
128 V  
< 1.2 dB relative to 20 kHz  
85  
2 V  
82  
2 V  
4 V  
80  
Utilities  
Calibration cycle  
4 V  
1 year  
Measured for input of 1 Mhz  
8 V  
6.5 mV  
8.5 mV  
11 mV  
16 V  
32 V  
64 V  
128 V  
256 V  
60 mV  
94 mV  
120 mV  
1. 50 Ω source resistance.  
2. For -1 dB of full scale range  
Waveform memory  
Data memory standard1  
32 Msa/ch  
Random access to readings in multiple record mode Capture multiple records from single trigger  
source  
1. Nominal values. Specific sample max is 30,000,000  
Waveform measurements  
Voltage  
Peak-to-peak, Minimum, Maximum, High (Top), Low (Base), Amplitude, RMS,  
Cycle RMS, Average, Cycle Average, Overshoot, Preshoot  
Time  
Rise, Fall, Frequency, Period, Positive Width, Negative Width, Positive Duty  
Cycle, Negative Duty Cycle  
08 | Keysight | M9217A PXIe Single-Slot 2-Channel Isolated Digitizer - Data Sheet  
For more information on Keysight  
Evolving Since 1939  
Our unique combination of hardware, software, services, and people can help you  
reach your next breakthrough. We are unlocking the future of technology.  
From Hewlett-Packard to Agilent to Keysight.  
Technologies’ products, applications or  
services, please contact your local Keysight  
office. The complete list is available at:  
www.keysight.com/find/contactus  
Americas  
Canada  
Brazil  
Mexico  
United States  
(877) 894 4414  
55 11 3351 7010  
001 800 254 2440  
(800) 829 4444  
Asia Pacific  
Australia  
China  
Hong Kong  
India  
Japan  
Korea  
Malaysia  
Singapore  
Taiwan  
1 800 629 485  
800 810 0189  
800 938 693  
1 800 11 2626  
0120 (421) 345  
080 769 0800  
1 800 888 848  
1 800 375 8100  
0800 047 866  
(65) 6375 8100  
myKeysight  
www.keysight.com/find/mykeysight  
A personalized view into the information most relevant to you.  
Other AP Countries  
www.keysight.com/find/emt_product_registration  
Register your products to get up-to-date product information and  
find warranty information.  
Europe & Middle East  
Austria  
Belgium  
Finland  
France  
Germany  
Ireland  
Israel  
Italy  
Luxembourg  
Netherlands  
Russia  
0800 001122  
0800 58580  
0800 523252  
0805 980333  
0800 6270999  
1800 832700  
1 809 343051  
800 599100  
+32 800 58580  
0800 0233200  
8800 5009286  
800 000154  
0200 882255  
0800 805353  
Opt. 1 (DE)  
Keysight Services  
www.keysight.com/find/service  
Keysight Services can help from acquisition to renewal across your  
instrument’s lifecycle. Our comprehensive service offerings—one-  
stop calibration, repair, asset management, technology refresh,  
consulting, training and more—helps you improve product quality  
and lower costs.  
Keysight Assurance Plans  
www.keysight.com/find/AssurancePlans  
Up to ten years of protection and no budgetary surprises to ensure  
your instruments are operating to specification, so you can rely on  
accurate measurements.  
Spain  
Sweden  
Switzerland  
Opt. 2 (FR)  
Opt. 3 (IT)  
0800 0260637  
Keysight Channel Partners  
www.keysight.com/find/channelpartners  
United Kingdom  
Get the best of both worlds: Keysight’s measurement expertise and  
product breadth, combined with channel partner convenience.  
For other unlisted countries:  
www.keysight.com/find/contactus  
(BP-9-7-17)  
www.keysight.com/find/automotive  
DEꢀꢁA Certified  
ISOꢂ001 ꢃuality Management System  
www.keysight.com/go/quality  
Keysight Technologies, Inc.  
DEKRA Certified ISO 9001:2015  
Quality Management System  
This information is subject to change without notice.  
© Keysight Technologies, 2017  
Published in USA, December 1, 2017  
5992-0815EN  
www.keysight.com  

相关型号:

M9240AROA

InfiniiVision PXIe Modular Oscilloscopes
KEYSIGHT

M9240ATOA

InfiniiVision PXIe Modular Oscilloscopes
KEYSIGHT

M9240AWGA

InfiniiVision PXIe Modular Oscilloscopes
KEYSIGHT

M9240CMPA

InfiniiVision PXIe Modular Oscilloscopes
KEYSIGHT

M9240CXPA

InfiniiVision PXIe Modular Oscilloscopes
KEYSIGHT

M9240EMBA

InfiniiVision PXIe Modular Oscilloscopes
KEYSIGHT

M9240FRAA

InfiniiVision PXIe Modular Oscilloscopes
KEYSIGHT

M9240MSKA

InfiniiVision PXIe Modular Oscilloscopes
KEYSIGHT

M9240NFCA

InfiniiVision PXIe Modular Oscilloscopes
KEYSIGHT

M9240NRZA

InfiniiVision PXIe Modular Oscilloscopes
KEYSIGHT

M9240PWRA

InfiniiVision PXIe Modular Oscilloscopes
KEYSIGHT

M9240SNSA

InfiniiVision PXIe Modular Oscilloscopes
KEYSIGHT