N5465B-1TP [KEYSIGHT]

Infiniium S-Series High-Definition Oscilloscopes;
N5465B-1TP
型号: N5465B-1TP
厂家: Keysight Technologies    Keysight Technologies
描述:

Infiniium S-Series High-Definition Oscilloscopes

文件: 总42页 (文件大小:4513K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
Keysight Technologies  
Infiniium S-Series High-Definition Oscilloscopes  
Data Sheet  
The New Standard for Superior Measurements  
02 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Welcome to the Next Generation of Oscilloscope Technology  
Infiniium S-Series oscilloscopes incorporate innovative technology designed to deliver  
superior measurements.  
There is no better way to experience  
that superiority of the Infiniium  
S-Series oscilloscopes than to use one.  
Our new 10-bit ADC and low-noise front-end technology work together to provide  
up to 8 GHz performance with the industry’s best signal integrity. We put these in an  
advanced frame with a solid state drive for fast boot-up, capacitive 15” display for easy  
touch capability, and a high-powered motherboard for fast processing. It’s all compatible  
with a myriad of Keysight Technologies, Inc. probes and Infiniium applications.  
Contact Keysight today to request a  
demo.  
Or visit,  
www.keysight.com/find/S-Series  
The Infiniium S-Series offers bandwidths from 500 MHz up to 8 GHz. Each model, equipped with a large 15” XGA capacitive touch screen, comes in a  
quiet package that is just 9” (23 cm) deep.  
Scope channels  
DSO models  
4 scope channels  
MSO models  
4 scope channels +  
16 digital channels  
MSOS054A  
MSOS104A  
MSOS204A  
MSOS254A  
MSOS404A  
MSOS604A  
MSOS804A  
Analog  
bandwidth  
Max sample rate  
ADC  
bits  
Standard memory  
depth  
User-installed  
bandwidth upgrades  
DSOS054A  
DSOS104A  
DSOS204A  
DSOS254A  
DSOS404A  
DSOS604A  
DSOS804A  
500 MHz  
1 GHz  
2 GHz  
2.5 GHz  
4 GHz  
6 GHz 1  
8 GHz 1  
20 GSa/s (2 channels)  
10 GSa/s (4 channels)  
100 Mpts (2 channels)  
50 Mpts (4 channels)  
10  
Yes  
1. 6 GHz and 8 GHz bandwidth supported in 2-channel mode. If all four channels are on, a maximum bandwidth of 4 GHz is supported.  
03 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
The New Standard for Superior Measurements  
Industry’s best signal integrity  
– 10-bit ADC up to 8 GHz for additional vertical resolution  
– Low-noise front end for precision signal viewing  
– Only 90 uV noise at 1 mV/div and 1 GHz bandwidth  
– System ENOB values in excess of 8  
– SFDR values down to –73 dBC  
– 2 mV/div vertical scaling supported in hardware  
HW bandwidth limit filters on both 50 Ω and 1 MΩ paths  
– Correction filters ensure flat frequency magnitude and phase response  
– Low intrinsic jitter (100 fs (typical) for excellent jitter characterization)  
– Precision BNCs with > 8 GHz bandwidth  
Most advanced platform  
– Powerful, flexible Infiniium user interface  
– Capacitive touch screen with multi-touch, easy-grab handles, and re-sizing touch  
fields  
– Powerful Intel i5 motherboard with 8 GB RAM for fast processing  
– Removable solid state drive (SSD) for fast boot-up and increased reliability and  
security  
– Fast data offload (up to 200 MB/s) via USB 3.0  
Broadest range of capability  
– 16 digital channels on MSO models  
– Standard feature rich software with > 50 automated measurements, 16 math  
functions, gating, and spectral viewer  
– Expandable with optional software applications and flexible licensing:  
– Add protocol decode and triggering for a wide variety of serial buses  
– Choose from a large selection of analysis applications including eye diagrams  
and measurements with SDA, jitter, InfiniiScan, and de-embedding  
– Test to ensure adherence to industry standards with compliance apps  
Support for > 100 probes – current and voltage, active and passive, 1 MΩ and  
50 Ω inputs  
04 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Industry’s Best Signal Integrity  
S-Series next-generation technology blocks enable superior measurements  
The heart of the oscilloscope is a 20-layer acquisition board with 16 custom ASICs and FPGAs. New technology blocks deliver superior  
signal integrity. You’ll get superior measurements that you won’t get with any other portable scope on the market.  
10-bit ADC  
Each model incorporates the industry’s fastest 10-bit ADC with a sample rate of  
40 GSa/s. This yields 2 channels at 20 GSa/s or 4 channels at 10 GSa/s.  
– 4X more vertical resolution than 8-bit oscilloscopes  
– ADC ENOB up to 8.7 contributes to high system ENOB values  
– Up to 12 bits of resolution with high-res mode  
– SNR are better than historical 8-bit ADC architectures  
– Vertical scaling as low as 2 mV/div supported in hardware  
Keysight’s new 10-bit ADC  
– 65 nm CMOS (9 mm x 14 mm)  
– 130 nm BiCMOS buffer IC  
– Custom 33 mm BGA package  
Superior low-noise front end  
– 10-bit ADC’s usefulness is dependent on the low-noise front end that supports the  
additional quantization levels. Each S-Series oscilloscope incorporates the industry’s  
lowest noise front end for portable oscilloscopes with bandwidth up to 8 GHz  
50 Ω and 1 MΩ input support, each path with bandwidth limit filter support  
– Analog and DSP bandwidth limit filters to reduce unwanted noise  
– 90 uV noise at 1 GHz bandwidth allows viewing of small signal detail  
– 2 mV/div vertical scaling in hardware (in combination with ADC)  
HW bandwidth limit filters on both 50 Ω and 1 MΩ paths  
– Gold-plated precision BNCs rated in excess of 8 GHz bandwidth  
– Electronic attenuators for decreased noise and increased reliability  
– Lower bandwidth models are upgradable to any higher bandwidth model with an  
instant user-installed software license  
S-Series front-end includes three new  
custom ICs including a 130 nm BiiCMOS IC  
that incorporates user-selectable analog  
filters and bandwidth upgrades via a  
software license.  
05 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Industry’s Best Signal Integrity (Continued)  
S-Series next-generation technology blocks enable superior measurements (Continued)  
Superior time base  
Time scale accuracy is critical, especially for deep-memory applications.  
Measurement of jitter is necessary for ensuring high-speed system  
reliability. Intrinsic jitter associated with an oscilloscope includes the jitter  
measurement internal to the scope. The lower the value, the better you’ll  
be able to characterize your device. S-Series scopes achieve precise time  
accuracy with a next-generation time base architecture.  
– Best-in-industry time scale accuracy of 12 parts per billion after  
calibration for accurate deep-memory measurements  
– Low jitter measurement floor with 100 fs (typical) of intrinsic jitter  
Signal processing in hardware  
S-Series oscilloscopes are built with an advanced FPGA dedicated for fast  
and precise signal processing. The technology produces the fastest deep  
memory responsiveness in the industry and provides additional hardware  
filtering for superior measurements.  
– Hardware-based algorithms for accelerated drawing to display (pixel  
placement) enable fast pan and zoom even with deep memory  
– Frequency-response correction filters produce flat responses for both  
magnitude and phase for more accurate waveforms  
– User-selectable hardware bandwidth-limiting correction filters from  
500 MHz up to the oscilloscope’s bandwidth reduce unwanted noise,  
plus additional front end filters for even more bandwidth limiting  
options  
– Supports cabled 2-channel differential inputs (channels 1 to 3 or  
channels 2 to 4) without requiring a differential probe  
– The DSP technology block supports rapid optional de-embedding  
technologies such as InfiniiSim, Precision Probe, and equalization  
Responsive deep memory  
S-Series oscilloscopes come with the industry’s most responsive  
deep memory. With standard 50 Mpts/channel on all four channels  
simultaneously, capture long time periods while retaining fast sample rates.  
Fast update rates mean your oscilloscope will stay responsive with deep  
memory on to ensure precise representation of analog signals.  
06 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Most Advanced Platform  
S-Series next-generation technology enables superior measurements  
S-Series platform standard features  
240 GB removable SSD  
– Fast boot up  
– Increased reliability  
– Easy to remove for secure environments  
Powerful motherboard  
– 3 GHz Intel i5 quad core processor with 8 GB RAM for fast computations — even with  
advanced math and deep memory  
– IO  
– Ethernet 10/100/1000bT  
– 6 USB device ports (2 in front, 4 on side two of which are USB 3.0)  
– DisplayPort and VGA video out. Drivers support up to two simultaneous displays  
Fast data offload  
Have an application that requires programmatically accessing oscilloscope data?  
Fast offload  
– USB 3.0 for up to 200 MB/s offload  
– 1000bT Ethernet for up to 80 MB/s offload  
Touch-screen innovation  
S-Series oscilloscopes incorporate a capacitive touch screen. In  
addition, Infiniium software includes a large number of touch-  
friendly enhancements including handles, enlarged touch fields  
when touch is enabled, and gestures (multi-touch).  
07 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Oscilloscope Overview  
With up to 8 GHz bandwidth, 20 GSa/s  
sample rate, a 10-bit ADC, and a low-noise  
front-end, see a precise representation of  
the analog characteristics of the signals  
you’re testing.  
Noise directly impacts vertical placement of each signal point.  
Oscillocopes with lower noise have more accurate vertical  
placement. S-Series oscilloscopes incorporate next-generation  
technology to deliver the lowest noise measurements in the  
industry. Here’s a compare at 4 GHz with 5 mV/div scaling.  
Superior signal integrity allows for more accurate measurements  
such as rise time values. Extremely low intrinsic jitter (100 fs)  
ensures the lowest possible contribution to jitter measurements  
from the scope itself so you’re using your jitter budget on your  
design.  
08 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Oscilloscope Overview – Time Domain  
Display windows and scale annotations  
Both horizontal and vertical axis values are annotated on the  
scales, leading to fast interpretation. Infiniium oscilloscopes  
uniquely offer 16 grids per waveform area, with up to 8 waveform  
areas.  
Results window  
See up to 20 measurement results simultaneously with statistics.  
Each result is color-coded to the source. Turn on measurement  
annotations for additional documentation capability.  
Math and measurements  
With > 50 standard automated measurements with statistics and  
16 independent math functions, you’ll be able to analyze a wide  
variety of tests. Use any of the industry-best 16 independent  
gates to narrow measurements or math to a specific time window.  
09 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Oscilloscope Overview – Frequency Domain  
With a built-in spectral viewer, controls, gated FFTs, 10-bit ADCs, and excellent SFDR values, the S-Series oscilloscopes provide an  
excellent scope platform for FFT measurements.  
FFTs  
Need to see frequency domain in additional to time domain? The  
standard spectral viewer includes FFT controls like start/stop  
RBW and CF/span. Readout includes power and frequency axis  
annotation and a peak table.  
Gated FFTs  
Infiniium supports gated math and analysis including FFTs.  
Use any of the standard 16 independent gates to narrow FFT  
computations to a specific time window. Drag the gate in the time  
domain, and see time correlated FFT measurements for specified  
time periods. The example at the right shows two simultaneously  
FFTs.  
Envelope measurements  
Need to see the rise time of a burst? Add a rise time measurement  
to an envelope function that provides an AM demodulation of a  
burst.  
10 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Oscilloscope Overview – Frequency Domain (Continued)  
Wideband and multi-channel FFTs  
Need to see > 500 MHz signal spectral bandwidth and/or multiple FFTs simultaneously? Oscilloscopes offer wider bandwidth than  
spectrum analyzers and come standard with four ports (channels) per instrument.  
Infiniium S-Series enables users to make wideband measurements up to 8 GHz and up to 16 simultaneous FFTs. Analyze even higher  
bandwidth signals by combining with a down converter.  
Up  
to 4  
S-Series  
8 GHz  
90000A  
Series  
13 GHz  
V-Series  
33 GHz  
Z-Series  
63 GHz  
PXA  
160 MHz  
UXA  
510 MHz  
1
0 MHz BW  
160 MHz BW 500 MHz BW  
8 GHz BW  
13 GHz BW 33 GHz BW  
... 63 GHz  
RF signal spectral bandwidth  
Use Infiniium capture and analysis of radar bursts, as shown in this  
OFDM example.  
Use Infiniium capture and analysis of radar bursts, as shown in this  
OFDM example.  
11 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Oscilloscope Overview – Frequency Domain (Continued)  
Trying to interpret traditional oscilloscope time-domain specifications can be challenging in determining if a specific scope can  
be recommended for RF/uW/mmW measurements. With correction filters, low-noise front end, and the 10-bit ADC, S-Series  
oscilloscopes can be used for wideband RF applications. Typical RF characteristics for the S-Series are listed below with graphs  
showing characterization results shown at the bottom of the page.  
Typical RF characteristic values from measured results on an 8-GHz S-Series oscilloscope  
Sensitivity/noise density  
—160 dBm/Hz  
1 mV/div; —38 dBm range  
Power spectral density measurement at 1.0001 GHz, 1.0001 GHz center frequency, 500 kHz span, and 3 kHz RBW  
Noise figure  
14 dB  
Derived from measurement above  
Signal-to-noise ratio/dynamic range  
108 dB  
0 dBm 1 GHz input carrier, 0 dBm scope input range  
1 GHz center frequency, 100 MHz span, 1 kHz RBW, measurement at +20 MHz from center  
Absolute amplitude accuracy  
± 1 dB  
0 to 7.5 GHz  
Deviation from linear phase  
± 7 deg  
0 to 7.5 GHz  
Phase noise (at 1 GHz)  
10 kHz offset  
100 kHz offset  
EVM  
—121 dBc/Hz  
—122 dBc/Hz  
—47 dB (0.47%)  
802.121 2.4 GHz carrier, 20 MHz wide, 64 QAM  
Spurious responses (0 dBm signal, 0 dBm input range)  
Spur Free Dynamic Range (SFDR)  
1 GHz, 0 dBm signal present at input, FFT =5 GHz span, 3 GHz center, 100 kHz RBW  
2nd harmonic distortion  
72 dB  
1 GHz input, 0 dBm, 5 GHz span, 3 GHz center, 100 KHz RBW  
3rd harmonic distortion  
—64 dBc  
1 GHz input, 0 dBm, 5 GHz span, 3 GHz center, 100 KHz RBW  
Two-tone Third-Order Intermodulation (TOI) distortion  
0 dBm input tones, 2.435 GHz and 2.439 GHz, 2 MHz separation, 2.437 GHz center frequency, 10 MHz span, 100 kHz  
RBW, 6 dBm input range  
—46 dBc  
+21.5 dB  
Input match  
< 50 mV/div, 0-7 GHz  
>= 50 mV/div, 0-7 GHz  
—15 dB; 1.4 VSWR  
—19 dB; 1.25 VSWR  
Using data acquired from S-Series, VSA shows an excellent TOI value of  
21.5 dB.  
Using data acquired from S-Series, VSA shows an EVM for IEEE 802.11  
QAM 64 of 0.47%.  
12 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Oscilloscope Overview – User Interface  
Infiniium oscilloscopes have been consistently recognized for high usability. The next-generation Infiniium user interface that ships  
standard on S-Series oscilloscopes delivers even more capability.  
Personalized viewing  
– Determine how much space to give to results versus wave-  
forms using sliders  
– Undock and move a window to an external monitor using the  
tabbed layout  
– Easily scale and see independent waveforms with user-  
selectable 1, 2, 4, or even 16 simultaneous grids in each  
waveform window  
Faster and better documentation  
– Quickly determine horizontal and vertical values as they are  
prominently displayed  
– Add annotations using bookmarks, measurement callouts,  
and dynamic delta marker readouts  
– Right-click to copy image without ever having to save to a file  
– Use the multi-purpose button to save screen images easily  
with auto-incrementing file names  
– Quickly save all waveforms, memories, functions, and setups  
in a single .osc file for later recall on an oscilloscope or PC  
– Save screen images as .jpg, .png, .gif, or .tiff  
Best usability, including touch screen  
Extensive research testing led to several touch-screen  
innovations not found in other oscilloscopes. All the following are  
industry firsts.  
– Click once to see handles that enable touch manipulation of  
markers, trigger level, and waveform-tasks that previously  
required a mouse  
– Multi-touch support for multi-touch (gestures) such as  
zooming and panning  
– Auto-sizing when touch button is turned on/off optimizes  
fields for fingers or a mouse  
13 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Oscilloscope Overview – User Interface (Continued)  
Custom mask editor  
Drag and drop up to 15 points on screen to create mask files in  
seconds.  
Analysis gallery  
Easily find and run the test you need from the list of analysis/  
measurement options represented graphically in the analysis  
gallery.  
14 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Using an External Monitor  
Undock and move a window to an external monitor using tabbed  
layout.  
S-Series supports both VGA and DisplayPort IO.  
Infiniium Offline Application  
View and analyze results at your desk. Save your oscilloscope  
file, then view and analyze on your PC without needing additional  
access to your scope.  
Use waveform math, filtering, and FFT spectral analysis and to  
get more insight. Need to see protocol decode, analyze jitter, or  
view eye diagrams? Infiniium Offline helps you get insight into all  
of these areas.  
15 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
MSO Overview  
MSO models add 16 high-speed timing channels with standard 128 Mpts digital memory, allowing you to retain fast 2 GSa/s sample  
rates over long periods of time. All DSOs are user-upgradable to MSOs. The required upgrade time is less than 5 minutes.  
Use the digital channels to evaluate control signal relationships  
and data buses up to 16 bits wide. Use symbols to quickly  
interpret waveforms.  
MSO measurement applications  
– Trigger on and view specific control or data flow events  
– Use digital channels for protocol trigger and decode (I²C, SPI,  
RS-232, JTAG, USB and more)  
– Capture data from FPGA debug ports  
– Trigger on and monitor power rail sequencing timing  
relationships  
– Combine with oscilloscope channels to trigger across up to  
20 channels simultaneously  
16 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Protocol Applications  
Does your design include a serial bus that is a key point for debug or test? Use one of Keysight’s protocol decoding and triggering  
application packages for increased productivity. The software converts DSO or MSO physical layer acquisitions into packets for  
specific protocols. Specify trigger conditions at the packet level.  
Quickly move between physical and protocol layer information  
using the time-correlated tracking marker. Display protocol  
content in waveforms and/or listing.  
License type  
Fixed  
Floating  
Factory-installed on new  
scope purchase or user-in-  
stalled on existing scope  
User-installed  
transportable license  
Server-based license  
(N5435A option)  
Optional application description  
Protocol  
8B/10B  
N5384A-1FP  
N/A  
N5384A-1TP  
N8842A-1TP  
N8803B-1TP  
N5384A-1TP  
N5384A-1TP  
N8800B-1TP  
N5391B-1TP  
N8817B-1TP  
N8820B-1TP  
N8807B-1TP  
N8802B-1TP  
N8809B-1TP  
N8824B-1TP  
N8808B-1TP  
N5463B-1TP  
N5462B-1TP  
N8801A-1TP  
N8812B-1TP  
N5464B-1TP  
N8805B-1TP  
N8819B-1TP  
N8825B-1TP  
N8821A-1TP  
N8837A-1TP  
N8818B-1TP  
N8835A-1TP  
N8803C-1TP  
N8803C-2TP  
003  
106  
033  
003  
003  
N/A  
006  
038  
N/A  
047  
036  
049  
072  
048  
032  
001  
035  
054  
034  
N/A  
N/A  
088  
092  
096  
063  
091  
103  
N/A  
ARINC 429 and MIL-STD-1553 protocol  
CAN, LIN, FlexRay protocol decode and triggering  
DVI  
N8803B-1FP  
N5384A-1FP  
N5384A-1FP  
N8800B-1FP  
N5391B-1FP  
N8817B-1FP  
N8820B-1FP  
N8807B-1FP  
N8802B-1FP  
N8809B-1FP  
N8824B-1FP  
N8808B-1FP  
N5463B-1FP  
N5462B-1FP  
N8801A-1FP  
N8812B-1FP  
N5464B-1FP  
N8805B-1FP  
N8819B-1FP  
N8825B-1FP  
N8821A-1FP  
N8837A-1FP  
N8818B-1FP  
N8835A-1FP  
N8803C-1FP  
N8803C-2FP  
HDMI  
I²C, SPI, and RS-232/UART protocol decode and triggering  
I²C/SPI protocol decode  
JTAG protocol decode  
MIPI® CSI-3 (M-PHY®) protocol decode  
MIPI DigRF v4  
MIPI D-PHY™ protocol decode  
MIPI LLI protocol decode  
MIPI RFFE protocol decode  
MIPI UniPro protocol decode  
PCI Express® Gen1 and Gen2 protocol decode and triggering  
RS-232/UART protocol decode and triggering  
SATA/SAS protocol decode  
SVID protocol decode  
USB 2.0 protocol decode and triggering  
USB 3.0 protocol decode  
USB 3.0 SuperSpeed Inter-Chip (SSIC) protocol decode 1  
10/100 Ethernet protocol triggering and decode  
USB 3.1 protocol triggering and decode  
USB-PD protocol triggering and decode  
Universal Flash Storage (UFS) protocol decode  
eSPI and Quad eSPI protocol decode and trigger  
CAN, LIN, FlexRay and CAN-FD protocol triggering and decode  
Upgrade CAN, LIN, FlexRay with CAN-FD  
1. The server license runs on an external PC and can be checked out for 90 days by connecting to the server from the GUI. The floating or transportable license  
is moved from scope to scope using KLM and can be done online or with USB thumb drive.  
17 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Analysis Applications  
A variety of analysis capabilities enable additional rapid insight  
InfiniiScan  
Rapidly trigger on complex events that you can see but are  
impossible to specify using hardware triggers. This innovative  
software quickly scans through thousands of acquired waveform  
cycles and isolated anomalous signal behavior. Select up to eight  
zones across scope channels are available.  
SDA (Serial Data Analysis)  
Quickly validate signal integrity for high-speed serial interfaces  
with embedded clocks. Recover embedded clocks, and build and  
validate eye diagrams. SDA also includes software-base trigger  
and decode for 8B/10B data.  
EZJIT, EZJIT Plus, and EZJIT Complete  
Characterize and evaluate most commonly needed jitter  
measurements. EZJIT Plus automates Rj/Dj separation, and EZJIT  
Complete additionally characterizes vertical noise.  
See jitter data sheets for additional information.  
18 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Analysis Applications (Continued)  
A variety of analysis capabilities enable additional rapid insight (Continued)  
Crosstalk analysis  
Quickly solve difficult crosstalk problems with the industry’s  
first application software designed to measure both near end  
crosstalk (NEXT) and far end crosstalk (FEXT) on up to four  
signals (one victim and three aggressors) at once.  
– Identify the aggressor and victim  
– Report the amount of crosstalk  
– Remove crosstalk and analyze  
PAM-4 analysis  
Easily make eye and level based measurements with the PAM-4  
analysis tool. Measure eye width, eye height, eye skew, level  
mean, RMS, and thickness for each level, plus:  
– Data time interval error for each threshold  
– Rise/Fall times for each of the six PAM-4 transition types  
– Support for CTLE, FFE, and DFE equalization  
– Bit error rate measurements – cumulative and per  
acquisition  
19 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Analysis Applications (Continued)  
A variety of analysis capabilities enable additional rapid insight (Continued)  
PAM-4 analysis (Continued)  
With the EZJIT and InfiniiScan applications, you can apply limit tests to find burst errors  
and view where errors happen on the waveform.  
20 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Analysis Applications (Continued)  
A variety of analysis capabilities enable additional rapid insight (Continued)  
MultiScope  
View up to 40 analog channels simultaneously by linking multiple  
scopes together. Connect 2, 5, or 10 scopes together and  
calibrate to eliminate skew.  
21 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Other Popular Applications  
National Instrument Drivers  
LabVIEW Plug and Play and IVI-C drivers for S-Series are available on ni.com/idnet  
Vector Signal Analysis  
Expand your oscilloscope with the 89601B vector signal analysis  
software. This application takes data from the scope and provides  
spectrum analysis and digital modulation analysis for wireless  
communication.  
MATLAB integration with User Defined Functions  
Install MATLAB on your oscilloscope, and add your favorite  
.m scripts as math function operators. Export and analyze  
oscilloscope data directly with MATLAB. The FFT waterfall  
spectrogram on the right is an example of processing that can  
be done in MATLAB. Order N8831A with option 001 for MATLAB  
Basic or option 002 for MATLAB advanced.  
License type  
Fixed  
Floating  
Optional application description  
Factory-installed on new scope purchase User-installed  
Server-based license  
or user-installed on existing scope  
transportable license  
(N5435A option)  
Analysis  
DSA (bundle with EZJIT Complete and SDA bundle)  
Equalization emulation  
EZJIT  
EZJIT Plus  
EZJIT Complete vertical noise analysis  
InfiniiScan  
DSOS000-DSA (new purchase only)  
N5461B-1FP  
E2681B-1FP  
N5400B-1FP  
N8823B-1FP  
N5415B-1FP  
N5465B-3FP  
N5465B-1FP  
W2650A  
U1882B-1FP  
N/A  
003 and 055  
026  
002  
001  
067  
004  
026  
027  
N/A  
N/A  
044  
003  
005  
084  
084  
094  
N5461B-1TP  
E2681B-1TP  
N5400B-1TP  
N8823B-1TP  
N5415B-1TP  
N5465B-3TP  
N5465B-1TP  
N/A  
U1882B-1TP  
N2808A-1TP  
N5384A-1TP  
N5430B-1TP  
N8833B-1FP  
N8833B-1FP  
N8836A-1TP  
Infiniisim Basic  
InfiniiSim Advanced  
OSA (Oscilloscope Signal Analyzer)  
Power  
PrecisionProbe  
Serial data analysis  
N2808A-1FP  
N5384A-1FP  
UDF (user-defined function for MATLAB integration) N5430B-1FP  
User Defined Functions  
Crosstalk analysis  
PAM-4  
N8833B-1FP  
N8833B-1FP  
N8836A-1FP  
MultiScope  
2 scopes  
5 scopes  
10 scopes  
N8834A-AFP  
N8834A-BFP  
N8834A-CFP  
N8834A-ATP  
N8834A-BTP  
N8834A-CTP  
085  
086  
090  
22 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Infiniium S-Series Standard Features  
15” XGA display makes it easier to view analog and digital  
signals as well as spectral and protocol views  
Capacitive display makes for responsive touch control and  
supports multi-touch gestures  
View: Up to 8 waveform windows. Each area can have up  
to 16 grids  
Drag and drop enables rapid measurements  
Measurements  
– Over 50 automated measurements  
– View up to 20 simultaneously  
– User-customizable result window  
– (size, position, and information)  
– X and Y markers with dynamic delta values  
Analyze  
– 20 math operators including FFT and filters  
– Up to 16 independent/cascaded math functions  
View windows: Analog, math, spectral, and measurement  
results (simultaneous, tabbed, or undocked)  
Built-in 10-MHz reference in/out ports  
synchronizes multiple measurement  
instruments in a system  
Mixed-signal oscilloscope (MSO)  
models seamlessly integrate four  
analog scope channels with 16 digital  
channels  
XGA and DisplayPort video output ports  
let you add an external monitor  
Standard USB 2.0 and 3.0 and LAN ports  
provide fast data offload, device support,  
and connectivity  
Trig in/out ports provide an easy way to  
synchronize your oscilloscope to other  
instruments  
23 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Dedicated single acquisition button provides better control  
to capture a unique event  
Pressing horizontal delay knob sets the delay to zero  
Zoom button provides quick access to two screen-zoom  
modes  
Responsive deep memory allows you to pan and zoom  
quickly  
Autoscale lets you rapidly display any analog or digital  
active signals, automatically setting the vertical, horizontal  
and trigger controls for the best display, while optimizing  
memory  
Digital channel button provides quick setup access  
Protocol decode button enables quick setup access  
Dedicated per-channel front panel controls make it easy to  
access the vertical and horizontal scaling and offset  
Quick access to fine/course control by pressing the  
horizontal and vertical sensitivity knobs  
Built-in USB ports makes it easy to save your work and  
update your system software quickly  
Gold-plated precision BNCs  
deliver > 8 GHz bandwidth  
through the connector  
AutoProbe interface automatically  
configures the attenuation ratio of the  
probe and provides probe power for  
Keysight’s active probes  
24 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Compliance Applications  
Need to test adherence to an  
industry-standard serial bus?  
The S-Series supports many compliance  
applications.  
License type  
Fixed  
Floating  
Optional application description  
Factory-installed on  
new scope purchase  
or user-installed on  
existing scope  
User-installed  
transportable license  
Server-based license  
(N5435A option)  
Compliance (see individual application data sheets for additional information)  
Minimum bandwidth (GHz)  
BroadR-Reach  
DDR bundle - includes DDR1, 2, 3 and 4  
DDR1 (200 to 1067 MT/s)  
DDR2 + LPDDR2 (400 to 1067 MT/s)  
DDR3 + LPDDR3 (800 to 1067 MT/s)  
eMMC  
Ethernet  
Ethernet + EEE  
10 GBase-T  
HDMI 1.4  
N6467B-1FP  
N5459B-1FP  
U7233B-1FP  
N5413C-1FP  
U7231C-1FP  
N6465B-1FP  
N5392C-3FP  
N5392C-1FP  
U7236B-1FP  
N5399D-3FP  
N6460C-3FP  
U7238D-1FP  
U7249D-1FP  
N6466B-1FP  
N5393E-3FP  
N5467C-1FP  
N7246B-1FP  
N6461B-1FP  
N5416B-1FP  
U7248C-1FP  
N5431B-1FP  
N6467B-1TP  
N/A  
N5435A-062  
N/A  
1
U7233B-2FP  
N5413C-1TP  
U7231C-1TP  
N6465B-1TP  
N5392C-3TP  
N5392C-1TP  
U7236B-1TP  
N5399D-3TP  
N6460C-3TP  
U7238D-1TP  
U7249D-1TP  
N6466B-1TP  
N5393E-3TP  
N5467C-1TP  
N7246B-1TP  
N6461B-1TP  
N5416B-1TP  
U7248C-1TP  
N5431B-1TP  
N5435A-021  
N5435A-037  
N5435A-053  
N5435A-061  
N5435A-008  
N5435A-060  
N5435A-023  
N5435A-011  
N5435A-050  
N5435A-022  
N5435A-043  
N5435A-068  
N5435A-040  
N5435A-058  
N/A  
1 to 4  
2 to 4  
4
1
1
1
2.5  
8
8
4
6
1.5  
6
0.5  
1
6
2
2
6
MHL 2.0  
MIPI D-PHY  
MIPI M-PHY®  
MOST  
PCI Express Gen1  
UDA (User-Defined Application)  
UHS-I  
UHS-II  
USB 2.0  
HSIC  
XAUI  
N5435A-052  
N5435A-017  
N5435A-042  
N5435A-018  
Need to additionally test even higher-speed serial buses?  
Consider Keysight’s Z-Series, 90000A Series, or V-Series.  
25 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Probes Overview  
S-Series oscilloscopes include both 1 MΩ and 50 Ω paths. This  
expands their flexibility by making them compatible with a wider  
range of probes than high-performance oscilloscopes that only  
support a 50 Ω path.  
All S-Series oscilloscopes ship standard with four passive probes  
and support a wide range of about 100 compatible current and  
voltage probes. The table below highlights probes commonly  
used with the S-Series.  
See  
http://literature.cdn.keysight.com/litweb/pdf/5968-7141EN.pdf  
for additional info on probes andaccessories.  
Probe type  
Passive probe  
InfiniiMax active probe  
Ideal for measuring  
Recommended probe model numbers  
N2873A (qty. 4 ship standard)  
1130A/31A/32A/34A, 1168A/69A  
N2830A/31A/32A (InfiniiMode)  
N2750A-52A  
N2795A, N2796A, N2797A (extreme temperature)  
N2790A, N2791A, N2818A, N2819A, N2891A  
1146B, 1147B, N2780B-83B, N2893A  
N2820A, N2821A  
Single-ended voltage up to 500 MHz  
Differential or single-ended voltage  
Differential, single-ended or common mode voltage  
Single-ended, differential, or common mode voltage  
Single-ended voltage up to 2 GHz  
InfiniiMode active probe  
Single-ended active probe  
General purpose differential probes High voltage differential signal up to 7 kV with high CMRR  
Current probe  
Passive probes  
Power rail probe  
High current, AC/DC (mA – 100’s of A)  
High sensitivity current, AC/DC (10’s of uA – A)  
High voltage up to 4 kV peak pulse  
General purpose up to 1.5 GHz  
Single-ended up to 2 GHz with up to ± 24 V offset  
10076B  
N2870A – N2876A  
N7020A  
26 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Probes and Accessories Overview  
N7020A power rail probe  
Do you need more offset than is available in your oscilloscope  
frame? Do you need to zoom in to view and analyze small  
signals like ripple on top of power rails? Do you need more  
input impedance than 50 Ω at DC to make more precise  
measurements? If so, the N7020A power rail probe has the  
feature set that will help you test more efficiently and precisely.  
Developed specifically for power rail testing needs, the probe  
delivers 850 mV dynamic range with an impressive offset range  
of ± 24 volts. Low noise with a 1:1 attenuation ratio and 2 GHz  
bandwidth and low DC loading complements the probes ability to  
deliver superior power rail measurements.  
Go to www.keysight.com/find/N7020A for additional information.  
N2750A Series InfiniiMode active probes  
These active probes offer 1.5/3.5/6 GHz bandwidth and  
InfiniiMode operation modes and provide convenient and quick  
access to various functions on the oscilloscope.  
With wide dynamic range (10 Vpp at 10:1) and offset range  
(± 15 V), these probes can be used for a vast variety of  
measurements.  
Go to www.keysight.com/find/InfiniiMode for additional  
information.  
N2820A Series high-sensitivity current probes  
Need to measure small currents? The N2820A Series high-  
sensitivity current probes can measure down to 50 uA and up to  
5 amps. Combine with the S-Series for lower noise and 10-bit ADC  
high-definition current measurements. This probe can also be used  
to measure sensitive low voltages down to 3 uV.  
Go to www.keysight.com/find/N2820A for additional information.  
27 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Keysight Portfolio Compare  
InfiniiVision  
Infiniium  
Family  
Optimized for  
6000 X-Series  
Fastest update rate and  
lowest price up to 6 GHz  
500 MHz to 6 GHz  
4 Mpts  
8
9000 Series  
Measurements up to  
4 GHz  
600 MHz to 4 GHz  
40 Mpts  
8
S-Series  
90000A Series  
Superior signal integrity  
up to 13 GHz  
2.5 to 13 GHz  
40 Mpts  
Superior signal integrity  
up to 8 GHz  
500 MHz to 8 GHz  
100 Mpts  
Available bandwidths  
Standard memory depth/ch (2-ch)  
ADC bits  
10  
8
Bandwidth filters  
Yes  
20 MHz  
Yes. Extensive  
Yes. Extensive  
(only on 1 MΩ input)  
Bandwidth correction filters  
Probe inputs  
Motherboard and OS  
Standard internal drive  
BNC inputs  
No  
No  
Yes  
Yes  
50 Ω  
50 Ω and 1 MΩ  
None. Embedded  
None  
Traditional  
Yes  
50 Ω and 1 MΩ  
Intel Core 2 Duo. Win7  
HDD  
Traditional  
Yes  
X
50 Ω and 1 MΩ  
Intel i5 Quad-Core. Win7  
Removable SSD  
Precision BNC  
Yes  
Intel Core 2 Duo. Win7  
HDD  
Precision BNC  
No  
2X  
MSO models  
Frame volume comparison  
1/2 X  
X
Infiniium S-Series blends high-performance oscilloscope capability with  
a wide range general-purpose features.  
28 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Accessories  
Quickly remove your solid state drive for additional  
security  
All S-Series models ship standard with a solid state drive. This  
gives you faster boot time, enhanced reliability, and the ability  
to quickly remove the solid state drive for safekeeping in secure  
environments. Need additional solid state drives for secure  
environments? N2746S provides an additional solid state drive  
with Windows 7.  
Need to ship your oscilloscope or take it on an  
airplane?  
Cruzer has created a custom case for Keysight’s Infiniium  
S-Series oscilloscopes. Order the rugged hard-shell transit  
case directly from Cruzer (http://www.casecruzer.com/  
oscilloscope/3f1312-0411j.html) using the product number  
3F1312-0411J. Shipping weight with an S-Series frame and no  
accessories is 51.5 lbs (23.4 kg).  
Rack mount your oscilloscope  
Mount your S-Series oscilloscope in an 8U high, 19” (487 mm)  
wide rack with the N2902B rack mount kit. The sturdy kit provides  
additional BNCs on the rack mount front panel for BNC in/out  
signals on the side of S-Series models.  
29 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Take a Quick Look Using Demo Wizard and USB Demo Board  
Evaluate using demo wizard waveforms  
We’ve captured a variety of signals and measurements to speed  
your evaluation of S-Series oscilloscopes. Use the demo wizard to  
quickly evaluate the scope’s capability without needing to create  
an array of live signals.  
Measurement examples  
For example, pull up an FFT and note the following: separate  
spectral viewer with annotated axis scales; dynamically updated  
X and Y marker delta values; a customizable results window that  
includes an FFT peak table.  
N2867A USB demo board  
Plug a mouse, USB drive, or any other device in one side and  
connect the other to your oscilloscope USB port. You now have  
easy access to a target for quick evaluation or training others on  
how to use the scope.  
The board routes out D+, D-, 5 V, and ground, so you can use it to  
make voltage, current, power, or other measurements.  
30 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Investment Protection  
Expand your oscilloscope’s capability with additional memory, bandwidth, digital channels, or software applications.  
Capture more time and keep faster sample  
Acquisition memory depth upgrade  
Option # on new scope N2113A Option #  
rates as horizontal settings are slowed.  
Additional acquisition memory can be  
added to the initial purchase or after  
purchase.  
(for previously  
purchased scopes)  
100  
200  
400  
Increase to 100 Mpts/200 Mpts (4-ch/2-ch)  
Increase to 200 Mpts/400 Mpts (4-ch/2-ch)  
Increase to 400 Mpts/800 Mpts (4-ch/2-ch)  
DSOS000-100  
DSOS000-200  
DSOS000-400  
All S-Series oscilloscopes ship with the  
same hardware and are calibrated to  
8 GHz on the production line. In less than  
1 minute, any lower-bandwidth S-Series  
scope can be user-upgraded to any higher  
bandwidth.  
Bandwidth upgrades  
Upgrade to 8 GHz bandwidth  
Upgrade to 6 GHz bandwidth  
Upgrade to 4 GHz bandwidth  
Upgrade to 2.5 GHz bandwidth  
Upgrade to 2 GHz bandwidth  
Upgrade to 1 GHz bandwidth  
DSOS8GBW  
DSOS6GBW  
DSOS4GBW  
DSOS2G5BW  
DSOS2GBW  
DSOS1GBW  
Note: Customer-installable calibration is recommended and incurs additional charges.  
Upgrade any DSO model to an MSO in  
less than 1 minute with the N2901E MSO  
upgrade kit.  
31 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
S-Series Ordering Configuration  
S-Series ordering configuration  
Calibration and accessory options  
Precision BNC (M) to SMA (F) adapters (qty 2)  
(standalone PN: 54855-67604)  
170225 compliant calibration  
17025 compliant calibration with accreditation  
ANSI Z540 complaint calibration  
S-Series rackmount kit (8U high)  
Additional SSDs with Win7  
DSOS000-821  
1. Choose needed bandwidth  
2. Choose MSO or DSO  
3. Choose desired software applications  
4. Choose memory depth upgrade  
5. Choose any additional probes and accessories  
DSOS000-1A7  
DSOS000-AMG  
DSOS000-A6J  
N2902B  
N2746S  
GPIB to LAN adapter  
N4865A  
Scope channels  
DSO models  
4 scope channels 4 scope channels +  
16 digital channels  
MSO models  
Analog  
bandwidth  
Max sample rate  
ADC bits Standard memory  
depth  
User-installed  
bandwidth upgrades  
DSOS054A  
DSOS104A  
DSOS204A  
DSOS254A  
DSOS404A  
DSOS604A  
DSOS804A  
MSOS054A  
MSOS104A  
MSOS204A  
MSOS254A  
MSOS404A  
MSOS604A  
MSOS804A  
500 MHz  
1 GHz  
2 GHz  
2.5 GHz  
4 GHz  
6 GHz 1  
8 GHz 1  
20 GSa/s (2 channels)  
10 GSa/s (4 channels)  
100 Mpts (2 channels)  
50 Mpts (4 channels)  
10  
Yes  
1. 6 GHz and 8 GHz bandwidth supported in 2-channel mode. If all four channels are on, a maximum bandwidth of 4 GHz is supported.  
2. For MSO upgrades, customer-installable calibration is recommended and incurs additional charges.  
Standard accessories Included  
All models ship standard with: 3-year warranty, four N2873A 500-MHz passive probes, probe accessory pouch, Keysight I/O libraries  
suite, localized power cord, front panel cover, 8 GHz BNC calibration cable, keyboard, and mouse.  
User guide and programmer’s guide ship on oscilloscope drive. Service guide available on Keysight.com. MSO models additionally ship  
with 16-channel flying lead set logic probe, MSO cable, and MSO calibration fixture.  
32 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Infiniium S-Series Performance Characteristics  
Vertical  
DSO/MSO models  
Vertical - scope channels  
Input channels  
S-054A  
S-104A  
S-204A  
S-254A  
S-404A  
S-604A  
S-804A  
DSO models - 4 analog  
MSO models - 4 analog + 16 digital  
50 Ω 1 500 MHz  
1 MΩ 500 MHz  
Analog bandwidth (–3 dB)  
1 GHz  
500 MHz  
2 GHz  
500 MHz  
2.5 GHz  
500 MHz  
4 GHz  
500 MHz  
6 GHz 6  
500 MHz  
8 GHz 6  
500 MHz  
Vertical resolution 2, 3  
Typical rise time/fall time  
10 to 90% 4  
Typical rise time/fall time  
20 to 80% 5  
10 bits, up to 12 bits with high-resolution mode  
860 ps  
620 ps  
8.1  
430 ps  
310 ps  
7.8  
215 ps  
155 ps  
7.5  
172 ps  
124 ps  
7.4  
107.5 ps  
77.5 ps  
7.2  
71.7 ps  
51.7 ps  
6.8  
53.8 ps  
33.8 ps  
6.4  
ENOB (typical)  
Input impedance 1  
50 Ω: ± 3.5% (typical ± 1% at 25 °C)  
1 MΩ: ± 1% (14 pF typical)  
50 Ω: 1 mV/div to 1 V/div  
1 MΩ: 1 mV/div to 5 V/div  
50 Ω: DC  
Input sensitivity 3  
Input coupling  
1 MΩ: AC (> 11 Hz), DC  
Bandwidth limit filters (analog)  
DSP bandwidth limit filters  
Channel-to-channel isolation  
20 MHz, 200 MHz for both 50 Ω and 1 MΩ paths  
Increments of 500 MHz up to rated scope bandwidth  
DC to 100 MHz: 50 dB  
100 MHz to 1 GHz: 40 dB  
> 1 GHz: 30 dB  
DC gain accuracy 1, 2, 3  
Max input voltage 1  
Offset range  
± 1% of full scale at full resolution (typical 5 mV to 1 V per division)  
± 2% of full scale at full resolution 1  
50 Ω: ± 5 V  
1 MΩ: 300 Vrms or DC and ± 400 Vpp (DC+AC)  
1 MΩ  
50 Ω  
Vertical sensitivity  
All vertical ranges  
Available offset  
± 12 divisions or ± 4 V,  
whichever is smallest  
Vertical sensitivity  
< 10 mV/div  
≥ 10 mV/div  
≥ 20 mV/div  
≥ 100 mV/div  
≥ 1 V/div  
Available offset  
± 2 V  
± 5 V  
± 10 V  
± 20 V  
± 100 V  
Offset accuracy 1, 3  
Dynamic range 7  
± 0.1 div ± 2 mV ± 1% of offset setting for offsets < 2 V  
± 0.1 div ± 2 mV ± 1.5% of offset setting for offsets ≥ 2 V  
50 Ω: ± 4 divisions from center screen  
1 mV to 100 mV; 2nd harmonic distortion of –30 dBc  
101 mV to 5 V; 2nd harmonic distortion of –20 dBc  
1 MΩ: ± 4 divisions from center screen  
DC voltage measurement  
accuracy 2  
Dual cursor: ± [(DC gain accuracy) + (resolution)]  
Single cursor: ± [(DC gain accuracy) + (offset accuracy) + (resolution/2)]  
1. Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 5 °C from firmware calibration  
temperature. Input impedance is valid when V/div scaling is adjusted to show all waveform vertical values within the oscilloscope display.  
2. Vertical resolution = 0.4% of full scale(≤ 5 GSa/s) or 0.1% of full scale(10 GSa/s or 20 GSa/s).  
3. 50 Ω input: Full scale is defined as 8 vertical divisions. Magnification is used below 2 mV/div, full-scale is defined as 16 mV.  
The major scale settings are 5 mV, 10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V. Testing is at maximum sample rate.  
1 MΩ input: Full scale is defined as 8 vertical divisions. Magnification is used below 2 mV/div, full-scale is defined as 16 mV.  
The major scale settings are 5 mV, 10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V, 5 V. Testing is at maximum sample rate.  
4. Calculation based on Tr = 0.43/BW.  
5. Calculation based on Tr = 0.31/BW.  
6. 6 GHz and 8 GHz bandwidth supported in 2-channel mode. If all four channels are on, a maximum bandwidth of 4 GHz is supported.  
7. For a 10:1 probe on the 1 MΩ input, vertical scaling is multiplied by 10.  
33 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Infiniium S-Series Performance Characteristics (Continued)  
RMS noise floor (Vrms ac) on 50 Ω inputs  
Vertical setting (Volts/div)  
1 mV/div  
2 mV/div  
5 mV/div  
10 mV/div  
20 mV/div  
50 mV/div  
100 mV/div  
200 mV/div  
S-054A  
74 uV  
74 uV  
77 uV  
87 uV  
125 uV  
372 uV  
0.78 mV  
1.6 mV  
3.5 mV  
5.1 mV  
S-104A  
90 uV  
90 uV  
S-204A  
120 uV  
120 uV  
129 uV  
163 uV  
233 uV  
610 uV  
1.2 mV  
2.6 mV  
5.5 mV  
9.2 mV  
S-254A  
130 uV  
130 uV  
135 uV  
172 uV  
254 uV  
650 uV  
1.3 mV  
2.8 mV  
6 mV  
S-404A  
153 uV  
153 uV  
173 uV  
220 uV  
330 uV  
768 uV  
1.6 mV  
3.4 mV  
7.3 mV  
12.5 mV  
S-604A  
195 uV  
195 uV  
205 uV  
256 uV  
446 uV  
1.3 mV  
2.3 mV  
4.9 mV  
10.0 mV  
17.6 mV  
S-804A  
260 uV  
260 uV  
320 uV  
390 uV  
620 uV  
1.4 mV  
3.1 mV  
6.4 mV  
13.3 mV  
24.1 mV  
94 uV  
110 uV  
163 uV  
456 uV  
0.96 mV  
2.0 mV  
4.2 mV  
6.8 mV  
500 mV/div  
1 V/div  
10.1 mV  
Vertical - digital channels  
Input channels  
16 channels of timing  
400 MHz  
User-defined, TTL, 5.0 V CMOS, 3.3 V CMOS, 2.5 V CMOS, ECL, PECL  
± 8.00 V in 10 mV increments  
± 40 V peak  
± (100 mV + 3% of threshold setting)  
± 10 V about threshold  
500 mV peak-to-peak  
100 kΩ ± 2% (~ 8 pF) at probe tip  
500 ps (typical)  
Analog bandwidth  
Threshold selections  
User-defined threshold range  
Maximum input voltage  
Threshold accuracy  
Input dynamic range  
Minimum input voltage swing  
Input impedance  
Channel-to-channel skew  
Resolution  
1 bit  
34 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Infiniium S-Series Performance Characteristics (Continued)  
Horizontal  
Horizontal system: Oscilloscope channels  
Main timebase range  
Resolution  
5 ps/div to 50 s/div  
1 ps  
Modes  
Main, Delayed, Roll (200 ms/div to 200 s/div)  
Reference position  
Horizontal position range  
Delayed sweep range  
Time scale accuracy 1, 8  
Oscilloscope channel de-skew range  
Intrinsic jitter 6  
Continuously adjustable across horizontal display range  
0 to ± 500 sec  
1 ps/div to current main time scale setting  
± (12 ppb initial + 75 ppb/year aging)  
–1 ms to +1 ms  
Acquired time range or  
delta-time interval  
< 1 us (100 ns/div)  
10 us (1 us/div)  
100 us (10 us/div)  
1 ms (100 us/div)  
10 ms ( 1 ms/div)  
100 fs rms  
Internal reference  
External reference  
(Sample clock jitter)  
100 fs rms  
123 fs rms  
138 fs rms  
145 fs rms  
200 fs rms  
100 fs rms  
123 fs rms  
138 fs rms  
145 fs rms  
145 fs rms  
Inter-channel intrinsic jitter 3  
Inter-channel skew drift 3, 7  
Jitter measurement floor 2  
Time interval error  
< 500 fs rms  
2
NoiseFloor  
SlewRate  
2
+ Intrinsic Jitter  
(
)
Period jitter  
2
NoiseFloor  
SlewRate  
2
2  
3  
+ Intrinsic Jitter  
(
)
Cycle-cycle/N-cycle jitter  
Inter-channel jitter measurement floor 2, 3, 5  
2
NoiseFloor  
SlewRate  
2
+ Intrinsic Jitter  
(
)
2
2
2
Time Interval  
Time Interval  
Interchannel  
+
+
Error Edge1  
Error Edge2  
(
)
(
)
Intrinsic Jitter  
1. Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 5 °C from firmware calibration  
temperature.  
2. Sample rate at maximum. Noise and slew rate determined at fixed-voltage measurement threshold, near middle of signal. Displayed signal not vertically  
clipped. Slew rate of sine wave = (peak signal amplitude)  
2 π f, slew rate of fast step ~= (10 to 90% rise time).  
⦁ ⦁  
3. Intra-channel = both edges on the same channel, Inter-channel = two edges on different channels. Time Interval Error(Edge1) = time-interval error  
measurement floor of first edge, Time Interval Error(Edge2) = time-interval error measurement floor of second edge.  
4. Reading is the displayed Delta Time Measurement Accuracy measurement value. Do not double the listed Time Scale Accuracy value in Delta Time  
Measurement Accuracy formula.  
5. Scope channels and signal interconnect de-skewed prior to measurement.  
6. External timebase reference values measured using a Wenzel 501-04608A 10 MHz reference. Intrinsic jitter value depends on acquisition time range for  
Time Interval Error formula and depends on delta-time between edges for all two-edge formulas.  
7. Skew between channels caused by ± 5 degrees C temperature change.  
8. Initial = immediately after factory or user calibration.  
35 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Infiniium S-Series Performance Characteristics (Continued)  
Horizontal (Continued)  
Horizontal system: Oscilloscope channels (Continued)  
Delta time measurement accuracy 2, 3, 4, 5  
Intra-channel  
No averaging  
2
2
Time Interval  
Time Interval  
TimeScale Delta  
   
± 5  
+
+
   
Error Edge1  
Error Edge2  
Accuracy  
Time  
(
)
(
)
   
256 averages  
2
2
Time Interval  
Time Interval  
TimeScale Delta  
   
5
±
+
+
   
Error Edge1  
Error Edge2  
Accuracy  
Time  
16  
(
)
(
)
   
Inter-channel  
No averaging  
2
2
2
Time Interval  
Time Interval  
Interchannel  
5  
+
+
Error Edge1  
Error Edge2  
Intrinsic Jitter  
(
)
(
)
±
TimeScale Delta  
Interchannel  
SkewDrift  
   
+
+
   
Accuracy  
Time  
   
256 averages  
2
2
2
Time Interval  
Time Interval  
Interchannel  
5
16  
+
+
Error Edge1  
Error Edge2  
Intrinsic Jitter  
(
)
(
)
±
TimeScale Delta  
Interchannel  
SkewDrift  
   
+
+
   
Accuracy  
Time  
   
2. Sample rate at maximum. Noise and slew rate determined at fixed-voltage measurement threshold, near middle of signal. Displayed signal not vertically  
clipped. Slew rate of sine wave = (peak signal amplitude) f, slew rate of fast step ~= (10 to 90% rise time).  
2
π
3. Intra-channel = both edges on the same channel, Inter-channel = two edges on different channels. Time Interval Error(Edge1) = time-interval error  
measurement floor of first edge, Time Interval Error(Edge2) = time-interval error measurement floor of second edge.  
4. Reading is the displayed Delta Time Measurement Accuracy measurement value. Do not double the listed Time Scale Accuracy value in Delta Time  
Measurement Accuracy formula.  
5. Scope channels and signal interconnect de-skewed prior to measurement.  
36 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Infiniium S-Series Performance Characteristics (Continued)  
Acquisition and trigger  
Acquisition - scope channels  
Maximum real time sample rate  
Standard memory depth  
Memory options  
4 x 10 GSa/s or 2 x 20 GSa/s  
50 Mpts x 4 channels, 100 Mpts x 2 channels  
Option 100: 100 Mpts x 4 channels, 200 Mpts x 2 channels  
Option 200: 200 Mpts x 4 channels, 400 Mpts x 2 channels  
Option 400 (single/run mode)  
Sample rate  
20 GSa/s  
10 GSa/s  
2 channels  
4 channels  
800/400 Mpts  
400/200 Mpts  
536/268 Mpts  
Not available  
400/200 Mpts  
400/200 Mpts  
≤ 5 GSa/s  
Sampling modes  
Real time  
Real time with peak detect  
High resolution (11 or 12 bit user selection)  
Equivalent time  
Segmented memory (minimum time between segments = 3.3 us)  
Memory depth  
Maximum number of segments  
Standard (50 Mpts/ch)  
100 Mpts/ch  
200 Mpts/ch  
16,384  
32,768  
65,536  
65,536  
400 Mpts/ch  
Roll (200 ms to 1 ks per division)  
Sin(x)/x interpolation  
Filters  
Acquisition - digital channels  
Maximum real time sample rate  
Max memory depth per channel  
Minimum width glitch detect  
DSO/MSO models  
2 GSa/s  
128/64 Mpts with 2 GSa/s, 64/32 Mpts with < 2 GSa/s (single/run mode)  
2 ns  
S-054A  
S-104A  
S-204A  
S-254A  
S-404A  
S-604A  
S-804A  
Trigger – scope channels  
Ch 1, 2, 3, 4, aux, and line  
Full bandwidth  
20 MHz  
Max trigger freq on 50 Ω path  
Sensitivity (edge trigger)  
1 MΩ  
3 GHz bandwidth  
200 MHz  
1 GHz  
2.5 GHz  
> 2.5 GHz  
< 5 mV/div  
> 5 mV/div  
< 5 mV/div  
> 5 mV/div  
< 0.7 div  
< 0.3 div  
< 0.15 div  
0 div  
< 1.4 div (up to 500 MHz)  
< 0.8 div (up to 500 MHz)  
< 0.5 div  
< 0.2 div  
0 div  
50 Ω  
< 0.3 div  
0 div  
< 0.45 div  
< 0.1 div  
< 1.6 div  
< 0.6 div  
Trigger level range - any channel  
Ch 1, 2, 3, 4: 50 Ω: ± 4 divisions from center screen  
Ch 1, 2, 3, 4: 1 MΩ: ± 4 divisions from center screen  
Auxiliary: ± 5 V (50 Ω) (maximum input: 5 Vpp)  
Auto, triggered, single  
Sweep modes  
Display jitter 2, 3, 4  
(Trigger jitter)  
520 fs rms  
Trigger hold off range  
Trigger actions  
100 ns to 10 s  
Specify an action to occur (and the frequency of the action) when a trigger condition occurs  
Actions include: email on trigger and execute “multipurpose” user settings  
Trigger coupling  
50 Ω and 1 MΩ: DC, or AC: (10 Hz) low frequency reject (50 kHz high pass filter), high frequency reject (50 kHz low  
pass filter)  
1. Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 5 °C from firmware calibration  
temperature.  
2. Internal edge trigger mode with JitterFree correction. Value depends on scope settings and trigger signal characteristics, and is equal to Time Interval Error  
value expressed in the formula above using the minimum Time Scale Accuracy value.  
3. Value shown represents typical Display jitter for DSOS404A at 100 mV/div triggering on 500 mVpp 2 GHz sin wave signal.  
4. Sample rate at maximum. Noise and slew rate determined at fixed-voltage trigger threshold, near middle of signal. Displayed signal not vertically clipped.  
37 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Infiniium S-Series Performance Characteristics (Continued)  
Trigger  
Trigger: digital channels MSO models  
Threshold range (user defined)  
Threshold accuracy  
± 8.0 V in 10-mV increments  
± (100 mV + 3% of threshold setting)  
Trigger modes  
Edge (analog and digital)  
Triggers on a specified slope (rising, falling or alternating between rising and falling) and voltage level on  
any channel  
Edge transition (analog)  
Trigger on rising or falling edges that cross two voltage levels in > or < the amount of time specified.  
Edge transition setting from 250 ps  
Edge then edge (time) (analog and digital)  
Edge then edge (event) (analog and digital)  
Glitch (analog and digital)  
Pulse width (analog and digital)  
1 GHz and 500 MHz model  
2 GHz and 2.5 GHz model  
4 GHz, 6 GHz, 8 GHz model  
Runt (analog)  
The trigger is qualified by an edge. After a specified time delay between 10 ns to 10 s, a rising or falling  
edge on any one selected input will generate the trigger  
The trigger is qualified by an edge. After a specified delay between 1 to 16,000,000 rising or falling  
edges, another rising or falling edge on any one selected input will generate the trigger  
Triggers on glitches narrower than the other pulses in your waveform by specifying a width less than  
your narrowest pulse and a polarity. Glitch range settings equal pulse width settings  
Trigger on a pulse that is wider or narrower than specified. Pulse width range setting is 250 ps to 10 s for  
analog channels and 2 ns to 10 s for digital channels  
Minimum detectable pulse width  
Minimum detectable pulse width  
Minimum detectable pulse width  
150 ps for analog channels  
1 ns for digital channels  
100 ps for analog channels  
1 ns for digital channels  
50 ps for analog channels  
1 ns for digital channels  
Triggers on a pulse that crosses one threshold but fails to cross a second threshold before crossing the  
first again. Runt settings equal pulse width settings  
Timeout (analog and digital)  
Pattern/pulse range  
Trigger when a channel stays high, low, or unchanged for too long  
Timeout settings equal pulse width settings  
Triggers when a specified logical combination of the channels is entered, exited, present for a (analog  
and digital)  
Specified period of time or is within a specified time range or times out. Each channel can have a value of  
High (H), Low (L) or Don’t care (X)  
State (analog and digital)  
Setup/hold (analog)  
Pattern trigger clocked by the rising, falling or alternating between rising and falling edge of one  
channel.  
Triggers on setup, hold, or setup and hold violations in your circuit. Requires a clock and data signal on  
any two inputs (except aux or line) channels as trigger sources. Setup and/or hold time must then be  
specified.  
Window (analog)  
Protocol packets (hardware serial trigger)  
Zone-qualified  
Trigger on entering, exiting, or inside specified voltage range  
Requires specified protocol option, I2C, SPI, CAN, LIN, RS-232/UART, SVID, USB, PCIe® Gen 1  
Requires InfiniiScan software option. SW-based triggering across up to 8 user-drawn zones. For each  
zone, user specifies “must intersect” or “must not intersect.” Zones can be drawn on multiple channels  
and combined using Boolean expressions  
38 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Infiniium S-Series Performance Characteristics (Continued)  
Measurements and math  
Measurements and math  
Math functions can operate on any combination of channels, memories, or other functions  
Math functions  
Up to 16 independent functions  
Gates  
Up to 16 (any function can be used as a gate)  
Waveform measurements  
Waveform memories  
Voltage (scope channels)  
Can be made on either main, zoom, or gated region with up to 20 simultaneous measurements  
Four for scope waveforms, one memory that includes all digital channels simultaneously  
Peak-to-peak, minimum, maximum, average, RMS, amplitude, base, top, overshoot, V overshoot, preshoot, V  
preshoot, upper, middle, lower, crossing point voltage , pulse top, pulse base, pulse amplitude  
Period, frequency, positive width, negative width, duty cycle, delta time  
Rise time, fall time, period, frequency, positive width, negative width, duty cycle, Tmin, Tmax, Tvolt, channel-to-  
channel delta time, channel-to-channel phase, count pulses, burst width, burst period, burst interval, setup time,  
hold time  
Time (digital channels)  
Time (scope channels)  
Mixed (scope channels only)  
Frequency domain  
Level qualification  
Eye-diagram measurements  
Measurement modes  
Statistics  
Area, slew rate  
FFT frequency, FFT magnitude, FFT delta frequency, FFT delta magnitude  
Any channels that are not involved in a measurement can be used to level-qualify all timing measurements  
Eye height, eye width, eye jitter, crossing percentage, Q factor, and duty-cycle distortion  
Displays the mean, standard deviation, minimum, maximum range, and number of measurement values for the  
displayed automatic measurements  
Histograms  
Source  
Waveform or measurement  
Orientation  
Vertical (for timing and jitter measurements) or horizontal (noise and amplitude change) modes, regions are defined  
using waveform markers  
Measurements  
Mean, standard deviation, mean ± 1, 2, and 3 sigma, median, mode, peak-to-peak, min, max, total hits, peak (area of  
most hits), X scale hits, and X offset hits  
Marker modes  
Waveform math  
Operators  
Manual markers, track waveform data, track measurements, and delta marker values can be displayed  
Absolute value, add, AM demodulation, average, gating, Butterworth 1, common mode, differentiate, divide,  
envelope, FFT magnitude, FFT phase, FIR 1, high pass filter, integrate, invert, LFE 1, low pass filter (4th-order Bessel  
Thompson filter), magnify, max, min, multiply, RT Eye 1, smoothing, SqrtSumOfSquare 1, square, square root,  
subtract, versus, chart (MSO models)  
Automatic measurements  
Multipurpose  
Measure menu access to all measurements, up to 20 measurements can be displayed simultaneously  
User-selectable for saving images when pressed or taking a series of measurements defined by the user, or other  
actions  
Drag-and-drop measurement  
toolbar  
Measurement toolbar with common measurement icons that can be dragged and dropped onto the displayed  
waveforms  
FFT (FFT viewer is standard)  
Frequency range  
Frequency resolution  
Window modes  
DC to 10 GHz (at 20 GSa/s) or 5 GHz (at 10 GSa/s)  
Resolution = sample rate/memory depth  
Hanning, flattop, rectangular, Blackman Harris, Hamming  
1. Requires MATLAB software and license.  
39 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Infiniium S-Series Performance Characteristics (Continued)  
Display, computer system, and IO ports  
Display  
Display  
XGA 15” capacitive touch screen  
Resolution  
Annotation  
Grids  
Infiniium application runs with 1024 pixels horizontally x 768 pixels vertically  
Up to 100 bookmarks can be inserted into the waveform area. Each can float or be tied to a specific waveform  
Up to 16 grids per waveform area  
Waveform windows  
Waveform styles  
Up to 8  
Connected dots, dots, variable persistence, infinite persistence, color graded infinite persistence  
Computer system and peripherals, I/O ports  
Computer system and peripherals  
Operating system  
CPU  
Windows 7 Embedded Standard  
3 GHz Intel i5 quad core  
PC system memory  
Drives  
8 GB RAM  
≥ 240 GB removable SSD (solid state drive)  
Peripherals  
Optical USB mouse and compact keyboard supplied. All Infiniium models support any Windows-compatible input device  
with a USB interface  
LXI class C  
LXI compliance  
IO Ports  
LAN  
RJ-45 connector, supports 10Base-T, 100Base-T, and 1000Base-T. Enables Web-enabled remote control, email on  
trigger, data/file transfers and network printing  
USB  
7 total ports:  
Two USB 2.0 ports on the front  
Four USB ports on the side (two are USB 3.0 and two are USB 2.0)  
One USB 3.0 port on side supporting up to 200 MB/s data offload  
Drivers support up to two simultaneous displays  
External display  
DisplayPort and VGA video out  
Trigger out  
TTL levels into high impedance load  
Auxiliary output  
DC (± 2.4 V), square wave  
Time base reference output  
Amplitude into 50 Ω: 1.65 Vpp ± 50 mV sine wave (8.3 dBm ± 0.3 dB) regardless of whether it is derived from the internal  
or the external reference  
Frequency from internal reference: 10 MHz ± 100 ppb if it has been calibrated within the last year  
Frequency: 10 MHz ± 20 ppm  
Amplitude: 356 mVpp (–5 dBm) minimum to 5 Vpp (+18 dBm) maximum  
Time base external reference  
input (impedance into 50 Ω)  
File types  
Oscilloscope waveforms:  
Compressed  
*.wfm, *.bin, *.h5, and *.osc composite (setup and waveforms)  
Internal formats  
Larger formats  
*.csv, *.tsv, and*.txt  
Digital waveform  
Images  
Support in .osc or .h5 formats  
.bmp, .tiff, .gif, .png or .jpg  
40 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
Infiniium S-Series Performance Characteristics (Continued)  
Environmental and general  
Temperature  
Humidity  
Altitude  
Operating 5 to 40°C  
Non-operating –40 to +65°C  
Operating up to 90% relative humidity (non-condensing) at +40 °C  
Non-operating up to 90% relative humidity (non-condensing) at +65 °C  
Operating up to 4,000 meters (12,000 feet)  
Non-operating up to 15,300 meters (50,000 feet)  
Operating random vibration 5 to 500 Hz, 10 minutes per axis, 0.3 g (rms)  
Non-operating random vibration 5 to 500 Hz, 10 minutes per axis, 2.41 g (rms); resonant search 5 to 500 Hz,  
swept sine, 1 octave/minute sweep rate, (0.75 g), 5 minute resonant dwell at 4 resonances per axis  
100 to 120 V, ± 10% 50/60/400 Hz  
Vibration  
Power  
100 to 240 V, ± 10% 50/60 Hz  
Max power dissipated: 380 W  
Typical operator noise  
Weight  
Dimensions (with feet retracted)  
Safety  
Electromagnetic compatibility  
standards  
35 dB at front of instrument  
Frame: 12 kg (26.4 lbs). Shipping: 20 kg (44.1 lbs)  
Height: 12.9 in (33 cm); width: 16.8 in (43 cm); depth: 9 in (23 cm)  
CAN/CSA22.2 No. 61010-1-12 UL Std. 61010-1 (3rd Edition)  
IEC 61326-1:2005/EN 61326-1:2006  
CISPR 11/EN 55011  
IEC 61000-4-2/EN 61000-4-2  
IEC 61000-4-3/EN 61000-4-3  
IEC 61000-4-4/EN 61000-4-4  
MTBF (mean time before failure)  
110,000 hours (typical)  
Keysight Oscilloscopes  
Multiple form factors from 20 MHz to > 90 GHz | Industry leading specs | Powerful applications  
41 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
www.axiestandard.org  
AdvancedTCA® Extensions for Instrumentation and Test (AXIe) is an  
open standard that extends the AdvancedTCA for general purpose and  
semiconductor test. The business that became Keysight was a founding  
member of the AXIe consortium. ATCA®, AdvancedTCA®, and the ATCA logo  
are registered US trademarks of the PCI Industrial Computer Manufacturers  
Group.  
www.lxistandard.org  
LAN eXtensions for Instruments puts the power of Ethernet and the Web  
inside your test systems. The business that became Keysight was a founding  
member of the LXI consortium.  
www.pxisa.org  
PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a  
rugged, PC-based high-performance measurement and automation system.  
42 | Keysight | Infiniium S-Series High-Definition Oscilloscopes - Data Sheet  
For more information on Keysight  
Evolving  
Technologies’ products, applications or  
services, please contact your local Keysight  
office. The complete list is available at:  
www.keysight.com/find/contactus  
Our unique combination of hardware, software, support, and people can help  
you reach your next breakthrough. We are unlocking the future of technology.  
Americas  
Canada  
Brazil  
Mexico  
United States  
(877) 894 4414  
55 11 3351 7010  
001 800 254 2440  
(800) 829 4444  
Asia Pacific  
Australia  
China  
Hong Kong  
India  
Japan  
Korea  
Malaysia  
Singapore  
Taiwan  
1 800 629 485  
800 810 0189  
800 938 693  
1 800 11 2626  
0120 (421) 345  
080 769 0800  
1 800 888 848  
1 800 375 8100  
0800 047 866  
(65) 6375 8100  
From Hewlett-Packard to Agilent to Keysight  
myKeysight  
www.keysight.com/find/mykeysight  
A personalized view into the information most relevant to you.  
Keysight Services  
www.keysight.com/find/service  
Other AP Countries  
Our deep offering in design, test, and measurement services deploys an  
industry-leading array of people, processes, and tools. The result? We help  
you implement new technologies and engineer improved processes that  
lower costs.  
Europe & Middle East  
Austria  
Belgium  
Finland  
France  
Germany  
Ireland  
Israel  
Italy  
Luxembourg  
Netherlands  
Russia  
0800 001122  
0800 58580  
0800 523252  
0805 980333  
0800 6270999  
1800 832700  
1 809 343051  
800 599100  
+32 800 58580  
0800 0233200  
8800 5009286  
800 000154  
0200 882255  
0800 805353  
Opt. 1 (DE)  
Three-Year Warranty  
www.keysight.com/find/ThreeYearWarranty  
Keysight’s committed to superior product quality and lower total cost  
of ownership. Keysight is the only test and measurement company with  
three-year warranty standard on all instruments, worldwide. And, we provide  
a one-year warranty on many accessories, calibration devices, systems and  
custom products.  
Spain  
Sweden  
Switzerland  
Keysight Assurance Plans  
www.keysight.com/find/AssurancePlans  
Up to ten years of protection and no budgetary surprises to ensure your  
instruments are operating to specification, so you can rely on accurate  
measurements.  
Opt. 2 (FR)  
Opt. 3 (IT)  
United Kingdom  
0800 0260637  
Keysight Channel Partners  
www.keysight.com/find/channelpartners  
Get the best of both worlds: Keysight’s measurement expertise and product  
breadth, combined with channel partner convenience.  
For other unlisted countries:  
www.keysight.com/find/contactus  
(BP-09-23-16)  
PCI-SIG®, PCIe® and the PCI Express® are US registered trademarks and/or  
service marks of PCI-SIG.  
MIPI® service marks and logo marks are owned by MIPI Alliance, Inc. and any  
use of such marks by Keysight Technologies is under license. Other service  
marks and trade names are those of their respective owners.  
DEKRA Certified  
ISO9001 Quality Management System  
www.keysight.com/find/S-Series  
www.keysight.com/go/quality  
Keysight Technologies, Inc.  
DEKRA Certified ISO 9001:2015  
Quality Management System  
This information is subject to change without notice.  
© Keysight Technologies, 2014 - 2016  
Published in USA, October 11, 2016  
5991-3904EN  
www.keysight.com  

相关型号:

N5465B-3FP

Infiniium S-Series High-Definition Oscilloscopes
KEYSIGHT

N5465B-3TP

Infiniium S-Series High-Definition Oscilloscopes
KEYSIGHT

N5468A

Triggering and Hardware-based
KEYSIGHT

N5471A

Infiniium 90000 Series Oscilloscopes
KEYSIGHT

N5471B

Infiniium 90000 Series Oscilloscopes
KEYSIGHT

N5471C

Infiniium 90000 Series Oscilloscopes
KEYSIGHT

N5471D

Infiniium 90000 Series Oscilloscopes
KEYSIGHT

N5471E

Infiniium 90000 Series Oscilloscopes
KEYSIGHT

N5472A

Infiniium 90000 Series Oscilloscopes
KEYSIGHT

N5472B

Infiniium 90000 Series Oscilloscopes
KEYSIGHT

N5472C

Infiniium 90000 Series Oscilloscopes
KEYSIGHT

N5472D

Infiniium 90000 Series Oscilloscopes
KEYSIGHT