Be Ready for Winter
Features
Operational down to –40 °C
temperature
Increase productivity with Blue-
tooth® wireless
connectivity
– OLED display with 2000:1 contrast
ratio and 160 degrees viewing angle3,4
– 30,000-count resolution
– Measure up to 1000 V AC and DC
– Measure up to 10 A (20 A for 30 s)
– Resistance, diode test, temperature,
capacitance
– Low Impedance mode2,3,4 and Low
Pass Filter
– Peak detection of up to 250 µs
– Continuity test with beeper and back-
light1,2
The U1273AX OLED handheld digital
multimeter, the latest addition to U1270
Series, is capable of operating in winter
weather down to –40°C temperature. Even
in frigid conditions, the U1273AX enables
you to achieve immediate and accurate
results without the need to warm up in
advance.
For wireless connectivity to smartphones
and tablets, the U1270 Series is compat-
ible with the U1177A infrared-to-Bluetooth
adapter for maximum efficiency and
productivity in completing measure-
ments tasks. Adding the optional U1177A
to a U1270 Series you can easily perform
remote monitoring and data
logging via Android devices or Windows-
based PC.
– Seven readings/s measurement rate
for voltage and current
– Smooth function for accurately stable
readings
– Up to 10,000 points internal memory
for data logging
– Bluetooth wireless connectivity with
optional U1177A Bluetooth
– PC connectivity with optional U1173A
IR-USB cable
– IP 54 certified – water and dust resis-
tant
– CAT III 1000 V, CAT IV 600 V safety
rating
– Up to 3000m operating altitude
– –40 to 55 °C operating temperature4
OLED for more display clarity
Designed with OLED display, you can ex-
perience crystal-clear measurement read-
ings with its outstanding 2000:1 contrast
ratio. The display also allows wider viewing
angles up to 160 degrees ensuring you get
the right readings at the first glance even
in poorly lit environments.
1. U1271A
2. U1272A
3. U1273A
4. U1273AX
3 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Key Functions
Water and dust resistance (IP54) High measurement rate at seven Up to 10,000 recording points
readings per second for Voltage
and Current
for manual, auto and event log-
ging
The series’ tightly sealed design helps
protect against water, dust and damage.
Each handheld DMM is certified with IP 54
ratings so that you can carry out tests and
measurements with confidence, even in
harsh working conditions.
You can detect even the slightest change
in your sensitive signals (Voltage and Cur-
rent) with its high measurement rate capa- huge internal memory of up to 10,000
bility. By clicking the resettable smooth
function button, you may customize the
readings’ sensitivity suitable for various
tests.
Record measurements on-the-go and
transfer data to PC conveniently with the
recording points. The GUI Data Logging
software and optional U1173A IR-USB
cable are required to transfer data or per-
form real time data logging on a PC.
Operational up to 3000
meters altitude
For high altitude applications such as
wind farm maintenance, you can measure
with confidence using the U1270 Series,
capable of measuring up to 3000 meters
above sea level.
Visual alert for continuity test
(for U1271A and U1272A only)
Built-in Low Pass Filter
The U1270 Series offers a 1 kHz LPF or
Low Pass Filter to provide accurate Vari-
able Frequency Drive (VFD) output mea-
surements. This function eliminates high
frequency noise and harmonics, ensuring
motor filter efficiency.
Continuity detection in noisy and dark
environments is made easy with U1270
Series’ loud beeper and flashing backlight
that indicates continuity and thus im-
proves safety.
Figure 1. Comparison of voltage output from industrial motor VFD without and with Low Pass Filter functionality.
4 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Key Functions
Low impedance mode
Peak detect at 250 µs
Stray voltages are usually found in non-
energized electrical wiring adjacent to
powered wires due to capacitive or induc-
tive coupling between these wires. The low
impedance mode serves to eliminate false
readings by dissipating these stray voltag-
es thus improves safety and measurement
efficiency during voltage measurement.
The peak detect function allows you to
capture the engine or motor startup tran-
sient as fast as 250 µs.
Figure 2. U1272A helps you identify the presence of stray voltage on a disconnected wire running parallel with the wire powering up the
VFD to an industrial motor. The image on the right shows the U1272A in low impedance mode.
5 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Front and Back Panel Description
OLED display with 2000:1
contrast ratio and 160
degree viewing angle
30,000 counts resolution
Smart Ω1 removes residual
Peak detect records tran-
sients as fast as 250 μs
voltage of up to 1000 mV
Measure up to
1000 V AC and DC
Auto diode1 automatically
determines diode polarity
Low Pass Filter (LPF)
removes unwanted high
frequency signals
Temperature measurement:
J type thermocouple
–210 to 1200 °C1, K type ther-
mocouple –200 to 1732 °C
ZLOW 1 provides both high
and low impedance modes
to eliminate stray voltages
Measure up to 10 A
(20 A for 30 s)
Front panel
1. U1272A, U1273A and U1273AX only
Once connected to any HH DMM via Bluetooth adapter you are able to log and view measurements
graphically from smart phones and tablets.
6 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Front and Back Panel Description
Hang hole
Probe holder/storage
IR-USB connectivity with
optional IR-USB cable
Slim, ergonomic design
for better grip
Easy fuse and battery
access from battery cover
beneath the stand
Back panel
7 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Choose Among These Four Models
U1271A
U1272A
U1273A
U1273AX
Basic Features
Display resolution
Display
30,000 counts
30,000 counts
30,000 counts
30,000 counts
LCD
Yes
AC
LCD
OLED
N/A
OLED
N/A
Backlight
Yes
True RMS
AC + DC
AC + DC
AC + DC
Measurements
Voltage
Up to 1000 V AC, DC
0.05% + 2 counts
Up to 1000 V AC, DC
0.05% + 2 counts
Up to 1000 V AC, DC
0.05% + 2 counts
Up to 1000 V AC, DC
0.05% + 2 counts
Basic dcV accuracy
Up to 10 A
(20 A for 30 s)
Up to 10 A
(20 A for 30 s)
Up to 10 A
(20 A for 30 s)
Up to 10 A
(20 A for 30 s)
Current
Resistance
Up to 100 MΩ
Up to 300 MΩ
Up to 300 MΩ
Up to 300 MΩ
Frequency, capacitance,
temperature, continuity,
diode test
Frequency, capacitance,
temperature, continuity,
diode test
Frequency, capacitance,
temperature, continuity,
diode test
Frequency, capacitance,
temperature, continuity,
diode test
Other measurements
AC bandwidth
20 kHz
Yes
100 kHz
Yes
100 kHz
Yes
100 kHz
Yes
Low pass filter
Low impedance mode
Smart Ohm
—
—
Yes
Yes
Yes
Yes
Yes
Yes
Safety and Regulatory
Over-voltage safety
protection
CAT III 1000 V,
CAT IV 600 V
CAT III 1000 V,
CAT IV 600 V
CAT III 1000 V,
CAT IV 600 V
CAT III 1000 V,
CAT IV 600 V
General
Logging memory
200 points
10,000 points
10,000 points
10,000 points
Optional IR-USB and
Bluetooth
Optional IR-USB and
Bluetooth
Optional IR-USB and
Bluetooth
Optional IR-USB and
Bluetooth
Connectivity
Operating temperature
Altitude
–20 to 55 °C
3000 meters
–20 to 55 °C
3000 meters
–20 to 55 °C
3000 meters
–40 to 55 °C
3000 meters
Water and dust ingress
protection
IP 54
IP 54
IP 54
IP 54
Up to 300 hours 4X AAA
Alkaline
Up to 300 hours 4X AAA
Alkaline
Up to 60 hours 4X AAA
Alkaline
Up to 100 hours 4X AAA
Lithium
Battery life
Warranty
3 years
3 years
3 years
3 years
8 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
General Specifications
Display
U1271A and U1272A: Liquid crystal display (LCD) (with maximum reading of 33,000 counts)
U1273A/U1273AX: Organic LED (OLED) display (with maximum reading of 33,000 counts)
Power consumption
Battery type
U1271A/U1272A: 460 mVA maximum (with backlight enabled)
U1273A/U1273AX: 180 mVA maximum (with maximum brightness)
4 × 1.5 V Alkaline battery (ANSI/NEDA 24A or IEC LR03), or
4 × 1.5 V Zinc Chloride battery (ANSI/NEDA 24D or IEC R03)
4 × 1.5 V Lithium battery (ANSI/NEDA 24LF or IEC FR03)
Battery life
U1271A and U1272A: 300 hours typical (based on new Alkaline batteries for DC voltage measurement)
U1273A/U1273AX:
Based on new Alkaline batteries for DC voltage measurement:
30/45/60 hours typical at High/Medium/Low brightness, respectively
Based on new Lithium batteries for DC voltage measurement:
50/100 hours typical at High/Low brightness, respectively
Low battery indicator will flash when the battery voltage drops:
For non-rechargeable batteries: 4.4 V (approximately)
For rechargeable batteries: 4.5 V (approximately)
Fuse
10 × 35 mm 440 mA/1000 V 30 kA fast-acting fuse
10 × 38 mm 11 A/1000 V 30 kA fast-acting fuse
Input impedance at off mode
Operating environment
1.67 kΩ (protected by positive temperature coefficient resistor)
(U1272A, U1273A and U1273AX only)
Operating temperature:
U1271A/ U1272A/U1273A: –20 to 55 °C, 0% to 80% RH
U1273AX: –40 to 55 °C, 0% to 80% RH (using Lithium batteries)
Full accuracy up to 80% RH for temperatures up to 30 °C, decreasing linearly to
50% RH at 55 °C
Altitude up to 3000 meters
Pollution degree II
Storage compliance
Safety compliance
–40 to 70 °C, 0 to 80% RH
CAN/CSA-C22.2 No. 61010-1-04
EN/IEC 61010-1:2001
ANSI/UL 61010-1:2004
Measurement category
CAT III 1000 V/CAT IV 600 V
Commercial limits compliance with EN61326-1
IP-54
Electromagnetic compatibility (EMC)
Ingress protection rating
Temperature coefficient
U1271A/U1272A/U1273A: 0.05 × (specified accuracy)/°C (from –20 to 18 °C, or 28 to 55 °C)
U1273AX: 0.05 x (specified accuracy/ °C (from –40 to 18 °C, or 28 to 55 °C)
Common Mode Rejection Ratio (CMRR)
Normal Mode Rejection Ration (NMRR)
Dimensions (W x H x D)
> 120 dB at DC, 50/60 Hz ± 0.1% (1 kΩ unbalanced)
> 60 dB at 50/60 Hz ± 0.1%
92 × 207 × 59 mm
Weight
U1271A: 518 grams (with batteries)
U1272A: 520 grams (with batteries)
U1273A: 500 grams (with batteries)
U1273AX: 500 grams (with batteries)
Warranty
Three years for product
Three months for product’s accessories
Calibration cycle
One year
9 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Specification Assumptions
– Accuracy is given as ±(% of reading + counts of least significant digit) at 23 °C ± 5 °C, with relative humidity less than
80% RH.
– AC V and AC μA/mA/A specifications are AC coupled, true RMS and are valid from 5% of range to 100% of range.
– The crest factor may be up to 3.0 at full scale except for the 1000 V range where it is 1.5 at full scale.
– For non-sinusoidal waveforms, add (2% reading + 2% full scale) typical, for crest factors up to 3.
– After ZLOW voltage measurements, wait at least 20 minutes for thermal impact to cool before proceeding with any
other measurement.
10 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Electrical Specifications
DC specifications for U1271A, U1272A, U1273A and U1273AX
Accuracy ± (% of reading +
counts of least significant digit)
Test current/
Burden voltage
Function
Range
Resolution
U1271A
—
U1272A
0.05 + 20
0.05 + 5
0.05 + 5
0.05 + 2
0.05 + 2
0.05 + 2
U1273A/U1273AX
0.05 + 20
0.05 + 5
30 mV
300 mV
3 V
0.001 mV
0.01 mV
0.0001 V
0.001 V
0.01 V
—
—
—
—
—
—
0.05 + 5
0.05 + 5
0.05 + 2
0.05 + 2
0.05 + 2
0.05 + 5
30 V
0.05 + 2
300 V
1000 V
0.05 + 2
0.1 V
0.05 + 2
Voltage1
ZLOW (low
input
impedance)
enabled, ap-
plicable for
1000 V range
and resolu-
tion only
0.1 V
—
1 + 20
1 + 20
—
30 Ω
300 Ω
3 kΩ
0.001 Ω
0.01 Ω
—
0.2 + 10
0.2 + 5
0.2 + 5
0.2 + 5
0.2 + 5
0.6 + 5
1.2 + 5
—
0.2 + 10
0.2 + 5
0.2 + 5
0.2 + 5
0.2 + 5
0.6 + 5
1.2 + 5
—
0.65 mA
0.65 mA
0.2 + 5
0.2 + 5
0.2 + 5
0.2 + 5
0.6 + 5
1.2 + 5
2.0 +10
0.0001 kΩ
0.001 kΩ
0.01 kΩ
65 µA
30 kΩ
300 kΩ
3 MΩ
6.5 µA
0.65 µA
Resistance2
0.0001 MΩ
0.001 MΩ
0.01 MΩ
93 nA/10 MΩ
93 nA/10 MΩ
93 nA/10 MΩ
30 MΩ
100 MΩ
2.0 + 10 @ < 100 MΩ
8.0 + 10 @ > 100 MΩ
2.0 + 10 @ < 100 MΩ
8.0 + 10 @ > 100 MΩ
300 MΩ
0.01 MΩ
—
93 nA/10 MΩ
300 nS
300 µA
3000 µA
30 mA
300 mA
3 A
0.01 nS
0.01 µA
0.1 µA
1 + 10
0.2 + 5
0.2 + 5
0.2 + 5
0.2 + 5
0.3 +10
0.3 +10
1 + 10
0.2 + 5
0.2 + 5
0.2 + 5
0.2 + 5
0.3 +10
0.3 +10
1 + 10
0.2 + 5
0.2 + 5
0.2 + 5
0.2 + 5
0.3 +10
0.3 +10
93 nA/10 MΩ
< 0.04 V/100 Ω
< 0.4 V/100 Ω
< 0.08 V/1 Ω
0.001 mA
0.01 mA
0.0001 A
0.001 A
Current3
< 1.00 V/1 Ω
< 0.1 V/0.01 Ω
< 0.3 V/0.01 Ω
10 A
Approximately
1 to 2 mA
3 V
0.0001 V
0.0001 V
0.5 + 5
—
0.5 + 5
0.5 + 5
0.5 + 5
0.5 + 5
Diode Test4
Approximately
0.1 to 0.3 mA
Auto
See notes on next page.
11 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Electrical Specifications
Notes for DC specifications (previous page)
1. Notes for voltage specifications:
– The accuracy of the 30 to 300 mV range is specified after the Null function is used to subtract the thermal effect (by shorting the
test leads).
– For ZLOW measurements, autoranging is disabled and the multimeter’s range is set to 1000 volts in the manual ranging mode.
2. Notes for resistance specifications:
– Overload protection: 1000 Vrms for short circuits with < 0.3 A current.
– Maximum open voltage is < +3.3 V.
– Built-in buzzer beeps when the resistance measured is less than 25 Ω ± 10 Ω. The multimeter can capture intermittent measure-
ments longer than 1 ms.
– The accuracy of the 300 Ω to 3 kΩ range is specified after the Null function is used to subtract the test lead resistance and thermal
effect (by shorting the test leads).
– U1273AX only: The accuracy for all resistance ranges is specified after the Null function is used when measuring at temperatures
below –20 °C. The Null function is used to subtract the test lead resistance and thermal effect (by shorting the test leads).
– For the ranges of 30 MΩ and 100 MΩ, the RH is specified for < 60%.
– The accuracy for ranges < 50 nS is specified after the Null function is used on an open test lead.
– The temperature coefficient of the 100 MΩ and 300 MΩ range is 0.1 × (specified accuracy)/°C (from –40 to 18 °C or 28 to 55 °C).
3. Notes for current specifications:
– Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse.
– Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse.
– Specification for 300 mA range: 440 mA continuous.
– Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals > 10 to 20 A for 30
seconds maximum. After measuring currents > 10 A, cool down the multimeter for twice the duration of the measured time before
proceeding with low current measurements.
4. Notes for diode specifications:
– Overload protection: 1000 Vrms for short circuits with < 0.3 A current.
– Built-in buzzer beeps continuously when the voltage measured is less than 50 mV and beeps once for forward-biased diode or
semiconductor junctions measured between 0.3 V and 0.8 V (0.3 V ≤ reading ≤ 0.8 V).
– Open voltage for diode: < +3.3 V DC.
– Open voltage for Auto diode: < +2.5 V DC and > –1.0 V DC.
12 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Electrical Specifications
AC specifications for U1271A
Accuracy ± (% of reading + counts of least significant digit)
Function
Range
Resolution
45 Hz to
65 Hz
30 Hz to
1 kHz
1 kHz to
5 kHz
5 kHz to
20 kHz
300 mV
3 V
0.01 mV
0.0001 V
0.001 V
0.01 V
0.7 + 20
0.7 + 20
0.7 + 20
0.7 + 20
0.7 + 20
1.0 + 25
1.0 + 25
1.0 + 25
1.0 + 25
1.0 + 25
2.0 + 25
2.0 + 25
2.0 + 25
2.0 + 25
—
2.0 + 40
2.0 + 40
2.0 + 40
—
30 V
True RMS AC
Voltage1
300 V
1000 V
0.1 V
—
LPF (low pass filter) enabled,
applicable for all voltage
ranges and resolution
1.0 + 25 @ < 200 Hz
5.0 + 25 @ < 440 Hz
0.7 + 20
—
—
Accuracy ± (% of reading + counts of least
significant digit)
Function
Range
Resolution
Burden voltage/Shunt
45 Hz to 2 kHz
300 µA
3000 µA
30 mA
300 mA
3 A
0.01 µA
0.1 µA
0.9 + 25
0.9 + 25
0.9 + 25
0.9 + 25
1.0 + 25
1.0 + 25
< 0.04 V/100 Ω
< 0.4 V/100 Ω
< 0.08 V/1 Ω
< 1.00 V/1 Ω
0.001 mA
0.01 mA
0.0001 A
0.001 A
True RMS AC
Current2
< 0.1 V/0.01 Ω
< 0.3 V/0.01 Ω
10 A
1. Notes for voltage specifications:
- Overload protection: 1000 Vrms. For millivolt measurements, 1000 Vrms for short circuits with < 0.3 A current.
- Input impedance: 10 MΩ (nominal) in parallel with < 100 pF.
2. Notes for current specifications:
- Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse.
- Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse.
- Specification for 300 mA range: 440 mA continuous.
- Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals > 10 to 20 A for 30 seconds maximum. After
measuring currents > 10 A, cool down the multimeter for twice the duration of the measured time before proceeding with low current measurements.
13 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Electrical Specifications
AC specifications for U1272A/U1273A and U1273AX
Accuracy ± (% of reading + counts of least significant digit)
Function
Range
Resolution
45 Hz to
65 Hz
20 Hz to
1 kHz
1 kHz to
5 kHz
5 kHz to
20 kHz
20 kHz to
100 kHz
30 mV
300 mV
3 V
0.001 mV
0.01 mV
0.0001 V
0.001 V
0.01 V
0.6 + 20
0.6 + 20
0.6 + 20
0.6 + 20
0.6 + 20
0.6 + 20
0.7 + 25
0.7 + 25
1.0 + 25
1.0 + 25
1.0 + 25
1.0 + 25
1.0 + 25
1.0 + 25
1.5 + 25
1.5 + 25
1.5 + 25
1.5 + 25
1.0 + 40
1.0 + 40
2.0 + 40
2.0 + 40
2.0 + 40
—
3.5 + 40
3.5 + 40
3.5 + 40
3.5 + 40
—
30 V
True RMS AC
Voltage1
300 V
1000 V
0.1 V
—
1.0 + 25 @ <
200 Hz
5.0 + 25 @ <
440 Hz
LPF (low pass filter) enabled,
applicable for all voltage
ranges and resolution
0.6 + 20
2.0 + 40
—
—
—
—
—
—
2 + 40 @ < 440
Hz
ZLOW 1000 V
Accuracy ± (% of reading +
counts of least significant digit)
Function
Range
Resolution
Burden voltage/Shunt
45 Hz to 65 Hz
20 Hz to 2 kHz
300 µA
3000 µA
30 mA
300 mA
3 A
0.01 µA
0.1 µA
0.6 + 25
0.6 + 25
0.6 + 25
0.6 + 25
0.8 + 25
0.8 + 25
0.9 + 25
0.9 + 25
0.9 + 25
0.9 + 25
1.0 + 25
1.0 + 25
< 0.04 V/100 Ω
< 0.4 V/100 Ω
< 0.08 V/1 Ω
< 1.00 V/1 Ω
0.001 mA
0.01 mA
0.0001 A
0.001 A
True RMS AC
Current2
< 0.1 V/0.01 Ω
< 0.3 V/0.01 Ω
10 A
1. Notes for voltage specifications:
- Overload protection: 1000 Vrms. For millivolt measurements, 1000 Vrms for short circuits with < 0.3 A current.
- Input impedance: 10 MΩ (nominal) in parallel with < 100 pF.
- ZLOW impedance: 2 kΩ (nominal).
- The input signal is lower than the product of 20,000,000 V×Hz.
- For 20 to 100 kHz accuracy: Three counts of the LSD per kHz of additional error is to be added for frequencies > 20 kHz and signal inputs < 10% of range.
- U1273AX only: For all AC voltage ranges, the accuracy is specified at 2.5% + 25 counts when measuring below –20 °C for 20 to 45 Hz AC signals.
2. Notes for current specifications:
- Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse.
- Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse.
- Specification for 300 mA range: 440 mA continuous.
- Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals > 10 to 20 A for 30 seconds maximum. After
measuring currents > 10 A, cool down the multimeter for twice the duration of the measured time before proceeding with low current measurements.
- U1273AX only: The accuracy for the 300 μA range, 3000 μA range, and 30 mA is specified after the Null function is used when measuring at temperatures
below –20 °C. The Null function is used to subtract the test lead resistance and thermal effect (by shorting the test leads).
- U1273AX only: For all AC current ranges, the accuracy is specified at 2.5% + 25 counts when measuring below –20 °C for 20 to 45 Hz AC signals.
14 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Electrical Specifications
AC + DC specifications for U1272A/U1273A and U1273AX
Accuracy ± (% of reading + counts of least significant digit)
Function
Range
Resolution
45 Hz to
65 Hz
20 Hz to
1 kHz
1 kHz to
5 kHz
5 kHz to
20 kHz
20 kHz to
100 kHz
30 mV
300 mV
3 V
0.001 mV
0.01 mV
0.0001 V
0.001 V
0.01 V
0.7 + 40
0.7 + 25
0.7 + 25
0.7 + 25
0.7 + 25
0.7 + 25
0.8 + 45
0.8 + 30
1.1 + 30
1.1 + 30
1.1 + 30
1.1 + 30
1.1 + 45
1.1 + 30
1.6 + 30
1.6 + 30
1.6 + 30
1.6 + 30
1.1 + 60
1.1 + 45
2.1 + 45
2.1 + 45
2.1 + 45
—
3.6 + 60
3.6 + 45
3.6 + 45
3.6 + 45
—
True RMS
AC + DC Voltage1
30 V
300 V
1000 V
0.1 V
—
Accuracy ± (% of reading +
counts of least significant digit)
Function
Range
Resolution
Burden voltage/Shunt
45 Hz to 65 Hz
20 Hz to 2 kHz
300 µA
3000 µA
30 mA
300 mA
3 A
0.01 µA
0.1 µA
0.8 + 30
0.8 + 30
0.8 + 30
0.8 + 30
0.9 + 35
0.9 + 35
1.1 + 30
1.1 + 30
1.1 + 30
1.1 + 30
1.3 + 35
1.3 + 35
< 0.04 V/100 Ω
< 0.4 V/100 Ω
< 0.08 V/1 Ω
< 1.00 V/1 Ω
0.001 mA
0.01 mA
0.0001 A
0.001 A
True RMS
AC + DC Current2
< 0.1 V/0.01 Ω
< 0.3 V/0.01 Ω
10 A
1. Notes for voltage specifications:
- Overload protection: 1000 Vrms. For millivolt measurements, 1000 Vrms for short circuits with < 0.3 A current.
- Input impedance: 10 MΩ (nominal) in parallel with < 100 pF.
- For 20 to 100 kHz accuracy: Three counts of the LSD per kHz of additional error is to be added for frequencies > 20 kHz and signal inputs < 10% of range
- U1273AX only: For all AC+DC voltage ranges, the accuracy is specified at 2.5% + 30 counts when measuring below –20 °C for 20 to 45 Hz AC+DC signals.
2. Notes for current specifications:
- Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm 30 kA fast-acting fuse.
- Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm 30 kA fast-acting fuse.
- Specification for 300 mA range: 440 mA continuous.
- Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals > 10 to 20 A for 30 seconds maximum. After
measuring currents > 10 A, cool down the multimeter for twice the duration of the measured time before proceeding with low current measurements.
- U1273AX only: The accuracy for the 300 μA range, 3000 μA range, and 30 mA is specified after the Null function is used when measuring at temperatures
below –20 °C. The Null function is used to subtract the test lead resistance and thermal effect (by shorting the test leads).
- U1273AX only: For all AC+DC current ranges, the accuracy is specified at 2.5% + 30 counts when measuring below –20 °C for 20 to 45 Hz AC+DC signals.
15 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Electrical Specifications
Temperature specifications1 - 4
Accuracy ± (% of reading +
Thermocouple
as specified below)
Range
Resolution
type
U1271A
U1272A
U1273A/U1273AX
1% of reading
+ 1 °C
1% of reading
+ 1 °C
1% of reading
+ 1 °C
–200 to 1372 °C
–328 to 2502 °F
0.1 °C
0.1 °F
K
J
1% of reading
+ 1.8 °F
1% of reading
+ 1.8 °F
1% of reading
+ 1.8 °F
—
1% of reading
+ 1 °C
1% of reading
+ 1 °C
–210 to 1200 °C
–346 to 2192 °F
0.1 °C
0.1 °F
1% of reading
+ 1.8 °F
1% of reading
+ 1.8 °F
—
1. The specifications above is specified after 60 minutes of warm-up time.
2. The accuracy does not include the tolerance of the thermocouple probe.
3. Do not allow the temperature sensor to contact a surface that is energized above 30 Vrms or 60 V DC. Such voltages pose a shock hazard.
4. The temperature calculation is specified according to the safety standards of EN/IEC-60548-1 and NIST175.
Capacitance specifications5, 6
Accuracy ± (% of reading +
counts of least significant digit)
Range
Resolution
U1271A
1 + 5
1 + 2
1 + 2
1 + 2
1 + 2
1 + 2
1 + 2
U1272A
1 + 5
1 + 2
1 + 2
1 + 2
1 + 2
1 + 2
1 + 2
U1273A/U1273AX
10 nF
100 nF
1000 nF
10 μF
0.001 nF
0.01 nF
0.1 nF
1 + 5
1 + 2
1 + 2
1 + 2
1 + 2
1 + 2
1 + 2
0.001 μF
0.01 μF
0.1 μF
100 μF
1000 μF
10 mF
0.001 mF
5. Overload protection: 1000 Vrms for short circuits with < 0.3 A current.
6. The accuracy for all ranges is specified based on a film capacitor or better, and after the Null function is used to subtract the test lead resistance and ther-
mal effect (by opening the test leads).
16 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Electrical Specifications
Frequency specifications1, 2
Accuracy ± (% of reading +
counts of least significant digit)
Range
Resolution
Minimum input frequency
99.999 Hz
999.99 Hz
9.9999 kHz
99.999 kHz
999.99 kHz
> 1 MHz
0.001 Hz
0.01 Hz
0.1 Hz
0.02 + 5
0.005 + 5
0.005 + 5
0.5 Hz
1 Hz
0.005 + 5
0.01 kHz
0.1 kHz
0.005 + 5
0.005 + 5 @ < 1 MHz
1. Overload protection: 1000 V; input signal is < 20,000,000 V × Hz (product of voltage and frequency).
2. The frequency measurement is susceptible to error when measuring low-voltage, low-frequency signals. Shielding inputs from external noise pickup is criti-
cal for minimizing measurement errors. Turning on the low pass filter may help you to filter out the noise and achieve a stable reading.
Duty Cycle3
Mode
Range
99.99%
99.99%
Accuracy at full scale
0.3 % per kHz + 0.3 %
0.3 % per kHz + 0.3 %
DC coupling
AC coupling
3. Notes for duty cycle specifications:
- The accuracy for duty cycle and pulse width measurements is based on a 3 V square wave input to the DC 3 V range. For AC couplings, the duty cycle
range can be measured within the range of 10% to 90% for signal frequencies > 20 Hz.
- The range of the duty cycle is determined by the frequency of the signal: {10 μs × frequency × 100%} to {[1 – (10 μs × frequency)] × 100%}.
- The pulse width (positive or negative) must be > 10 μs. The range of the pulse width is determined by the frequency of the signal.
Pulse Width4
Range
Resolution
0.01 ms
0.1 ms
Accuracy at full scale
999.99 ms
2000.0 ms
(duty cycle accuracy/frequency) + 0.01 ms
(duty cycle accuracy/frequency) + 0.1 ms
4. Notes for pulse width specifications:
- The accuracy for duty cycle and pulse width measurements is based on a 3 V square wave input to the DC 3 V range.
- The pulse width (positive or negative) must be > 10 μs. The range of the pulse width is determined by the frequency of the signal.
17 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Electrical Specifications
U1271A and U1272A frequency sensitivity for voltage measurements1, 2, 3
Minimum sensitivity (RMS sine wave)
Trigger level for DC coupling
0.5 Hz to 200 kHz
Input range
15 Hz to
100 kHz
0.5 Hz to
200 kHz
Up to
1 MHz
U1271A
U1272A
5 mV
30 mV
300 mV
3 V
3 mV
6 mV
0.12 V
0.6 V
6 V
3 mV
8 mV
—
40 mV
0.4 V
2.6 V
—
—
10 mV
0.15 V
15 mV
0.15 V
0.2 V
30 V
0.8 V
1.5 V
1.5 V
300 V
1000 V
8 V @ < 100 kHz
50 V @ < 100 kHz
9 V @ < 100 kHz
90 V @ < 100 kHz
9 V @ < 100 kHz
90 V @ < 100 kHz
50 V
—
1. Maximum input for specified accuracy, refer to “AC specifications” on page 12.
2. 30 mV range applicable for U1272A only.
3. 200 kHz to 1 MHz range applicable for U1272A only.
U1273A/U1273AX sensitivity for voltage measurements4
Frequency sensitivity and trigger level
Minimum sensitivity (RMS sine wave) Trigger level for DC coupling
Input range
Maximum input for specified
accuracy, refer to AC voltage
15 Hz to
100 kHz
0.5 Hz to
200 kHz
Up to
1 MHz
0.5 Hz to 200 kHz
30 mV
300 mV
3 V
3 mV
7 mV
0.12 V
0.8 V
6.7 V
67 V
3 mV
8 mV
—
38 mV
0.48 V
3.5 V
—
5 mV
15 mV
0.12 V
0.15 V
30 V
0.8 V
1.5 V
300 V
1000 V
8 V < 100 kHz
67 V < 100 kHz
11 V < 100 kHz
110 V < 100 kHz
—
4. Maximum input for specified accuracy, refer to “AC specifications” on page 13.
Frequency sensitivity for current measurements5
Minimum sensitivity (RMS sine wave)
Input range
2 Hz to 30 kHz
U1271A/U1272A
100 µA
70 µA
U1273A/U1273AX
70 μA
300 µA
3000 µA
30 mA
300 mA
3 A
120 μA
1.2 mA
1.2 mA
12 mA
12 mA
0.12 A
0.12 A
10 A
1.2 A
1.2 A
5. Maximum input for specified accuracy, refer to “AC specifications” on page 12 and 13.
18 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Electrical Specifications
Peak hold
Signal width
Accuracy for DC Voltage and Current
Specified accuracy + 400
Single event >1 ms
Repetitive >250 µs
Specified accuracy + 1000
Decibel (dB) for U1272A and U1273A1, 2, 3
dB base
Reference
Default reference
1 mW (dBm)
1 V (dBV)
1 to 9999 Ω
50 Ω
1 V
1 V
1. The reading of dBm is indicated in decibels of power above or below 1 mW, or decibels of voltage above or below 1 V. The formula is calculated according
to the voltage measurement and specified reference impedance. Its accuracy is depended on the accuracy of the voltage measurement. See Decibel (dBV)
accuracy table below.
2. Auto-ranging mode is used.
3. The bandwidth is according to voltage measurement.
Decibel (dBV) accuracy
dBV range
Accuracy
Range
45 Hz to
65 Hz
20 Hz to
1 kHz
1 Hz to
5 kHz
5 kHz to
20 kHz
20 Hz to
100 kHz
Minimum
Maximum
30 mV
300 mV
3 V
–56.48
–36.48
–16.48
+3.52
–30.46
–10.46
+9.54
+29.54
+49.54
+60
0.06
0.06
0.06
0.06
0.06
0.06
0.07
0.07
0.09
0.09
0.09
0.09
0.09
0.09
0.14
0.14
0.14
0.14
0.1
0.1
0.32
0.32
0.32
0.32
—
0.19
0.19
0.19
—
30 V
300 V
1000 V
+23.52
+33.98
—
Measurement rate (approximate)
Times/second
Function
U1271A
U1272A/U1273A/U1273AX
ACV
DCV
7
7
7
7
Ω
14
14
Ω with offset compensation
Diode
-
3
14
14
Auto Diode
Capacitance
DCA
-
3
4 (< 100 µF)
4 (< 100 µF)
7
7
ACA
7
7
Temperature
Frequency
Duty cycle
Pulse width
7
7
2 (> 10 Hz)
1 (> 10 Hz)
1 (> 10 Hz)
2 (> 10 Hz)
1 (> 10 Hz)
1 (> 10 Hz)
19 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Ordering Information
U1271A
U1272A
U1273A
U1273AX
Standard shipped accessories
Standard test leads, test probes with 4-mm tips, K-type thermocouple and adapter, 4x AAA alkaline batteries (4x AAA lithium batteries for U1273AX only),
Certificate of Calibration, UK 6 (test report), Quick Start Guide
Optional accessories
Measuring accessories (non-temperature)
U1161A
Extended test
lead kit
Includes two test leads (red and black), two test probes, medium-sized alligator clips
and 4-mm banana plugs.
– Test leads: CAT III 1000 V, CAT IV 600 V, 15 A
– Test probes (4-mm tips): CAT III 1000 V, CAT IV 600 V, 15 A
– Medium-sized alligator clips: CAT III 1000 V/CAT IV 600 V, 15 A
– 4-mm banana plugs: CAT II 600 V, 10 A
U1162A
Alligator clips
– One pair of insulated alligator clips (red and black). Recommended for use with
Keysight standard test leads.
– CAT III 1000 V, CAT IV 600 V, 15 A
U1163A
– One pair of SMT grabbers (red and black). Recommended for use with Keysight
SMT grabbers
standard test leads.
– Rated CAT II 300 V, 3 A
U1164A
Fine-tip test
probes
– One pair of fine-tip test probes (red and black). Recommended for use with
Keysight standard test leads.
– Rated CAT II 300 V, 3 A
U1168A
Standard test
lead kit
Includes two test leads (red and black), 4-mm test probes, alligator clips, fine-tip test
probes, SMT grabbers and mini grabber (black).
– Test leads: CAT III 1000 V, CAT IV 600 V, 15 A
– Test probe (19-mm tips): CAT II 1000 V, 15 A
– Test probe (4-mm tips): CAT III 1000 V, CAT IV 600 V, 15 A
(highly recommended for CAT IV environment)
– Alligator clips: CAT III 1000 V, CAT IV 600 V, 15 A
– Fine-tip test probes: CAT II 300 V, 3 A
– SMT grabber: CAT II 300 V, 3 A
– Mini grabber: CAT II 300 V, 3 A
U1583B
– Dual range: 40 A and 400 A
AC current clamp
– Rated CAT III 600 V
– BNC-to-banana-plug adapter provided for use with DMMs
– -40 to 55 °C operating temperature
20 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Ordering Information
Optional accessories
Measuring accessories (temperature)
U1180A
Includes thermocouple adapter, thermocouple bead J-type and thermocouple bead
Thermocouple
adapter+lead kit,
J and K types
K-type.
– T/C adapter J/K-type
– T/C bead J-type: –20 to 200 °C
– T/C bead K-type: –20 to 200 °C
U1181A
Immersion tempera-
ture probe
– Type-K T/C for use in oil and other liquids
– Measurement range: –50 to 700 ºC
– Includes adapter U1184A for connection to DMM
U1182A
– Type-K T/C for use on still surfaces
Industrial surface
temperature probe
– Measurement range: –50 to 400 ºC
– Includes adapter U1184A for connection to DMM
U1183A
Air temperature
probe
– Type-K T/C for use in air and non-caustic gas
– Measurement range: –50 to 800 ºC
– Includes adapter U1184A for connection to DMM
U1184A
– Mini-connector-to-banana-plug adapter for use with DMM
Temperature probe
adapter
U1185A
– T/C adapter J/K-type
J-type thermocouple
and adapter
– T/C bead J-type: –20 to 200 °C
U1186A
– T/C adapter J/K-type
K-type thermocou-
ple and adapter
– T/C bead J-type: –20 to 200 °C
21 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
Ordering Information
Optional accessories
U1171A
For fastening the DMM to a steel surface so both hands are free
Magnetic hanging kit
U1173A
IR-to-USB cable
– For remote control and data logging to PC
– Maximum baud rate: 19,200 bits per second
U1174A
The convenient way to carry your DMM and essential accessories
Soft carrying case
– Dimension: 9 inches (H) x 5 inches (W) x 3 inches (D)
U1177A
Bluetooth Adapter
– Enables Bluetooth connection to Keysight handheld digital multimeters
– Support the U1230, U1240, U1250 and U1270 Series handheld multimeters
– Remote monitoring and data logging capabilities via Android devices or
Windows-based PC
– Recommended to use Lithium battery in lower than -20°C in temperature for
long hours of operation
22 | Keysight | U1270 Series Handheld Digital Multimeters - Data Sheet
For more information on Keysight
From Hewlett-Packard through Agilent to Keysight
For more than 75 years, we‘ve been helping you unlock measurement insights. Our
unique combination of hardware, software and people can help you reach your next
breakthrough. Unlocking measurement insights since 1939.
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
55 11 3351 7010
001 800 254 2440
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
1 800 629 485
800 810 0189
800 938 693
1 800 11 2626
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 6375 8100
1939
THE FUTURE
Other AP Countries
myKeysight
Europe & Middle East
Austria
Belgium
Finland
0800 001122
0800 58580
0800 523252
0805 980333
0800 6270999
1800 832700
1 809 343051
800 599100
+32 800 58580
0800 0233200
8800 5009286
800 000154
0200 882255
0800 805353
Opt. 1 (DE)
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
France
Germany
Ireland
Israel
Italy
Luxembourg
Netherlands
Russia
Spain
Sweden
Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty
Keysight’s committed to superior product quality and lower total cost
of ownership. Keysight is the only test and measurement company with
three-year warranty standard on all instruments, worldwide. And, we provide
a one-year warranty on many accessories, calibration devices, systems and
custom products.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to ten years of protection and no budgetary surprises to ensure your
instruments are operating to specification, so you can rely on accurate
measurements.
Switzerland
Opt. 2 (FR)
Opt. 3 (IT)
United Kingdom
0800 0260637
Keysight Infoline
www.keysight.com/find/service
Keysight’s insight to best in class information management. Free access to
your Keysight equipment company reports and e-library.
For other unlisted countries:
www.keysight.com/find/contactus
(BP-02-10-16)
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product
breadth, combined with channel partner convenience.
DEKRA Certified
ISO9001 Quality Management System
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2015
Quality Management System
www.keysight.com/find/handhelddmm
This information is subject to change without notice.
© Keysight Technologies, 2013 - 2016
Published in USA, May 24, 2016
5990-6425EN
www.keysight.com
相关型号:
©2020 ICPDF网 联系我们和版权申明