1EAH1 [MACOM]
For UL544 Health Care Equipment; 对于UL544保健器材型号: | 1EAH1 |
厂家: | Tyco Electronics |
描述: | For UL544 Health Care Equipment |
文件: | 总2页 (文件大小:301K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
CORCOM Product Guide
For UL544 Health Care Equipment
EAH, EBH Series
UL Recognized
CSA Certified
VDE Approved
EAH1/EBH1
EAH Series
Specifications
Maximum leakage current,each line-to-ground
@ 120 VAC 60 Hz:
This is a new generation of compact medical RFI filters with
an IEC socket. A new design in internal structure and
manufacturing process makes this series cost effective.
This health care series supersedes the H series by
providing superior differential mode performance while still
maintaining a maximum leakage current of 2 micro amps.
2 µA
5 µA
@ 250 VAC 50 Hz:
Hipot rating (one minute):
line-to-ground
line-to-line
1500 VAC
1450 VDC
Operating frequency:
Rated voltage (max.):
50/60 Hz
250 VAC
EBH Series
Also a compact medical RFI filter with IEC socket, the EBH
series is a cost effective alternative to the H series.
Electrically the EBH provides enhanced differential mode
performance compared to both the EAH and the H series.
These filters provide RFI common and differential mode
performance while still allowing medical equipment to
meet UL leakage current specifications for medical and
dental equipment.
Minimum insertion loss in dB:
Line-to-ground in 50 ohm circuit
Current
Rating
Frequency-MHz
.01 .05 .1 .15 .5
1
5
10 30
EAH Models
01A
Additional information on UL544 specification is listed
in Appendix C.
8
-
21 29 32 42 45 32 30 19
10 15 25 27 30 27 22
03A
5
06A
10A
-
-
-
-
5
1
6
5
19 21 24 20 15
12 12 12 12
9
Note: When using the models with an IEC connector,
remember that the leakage current of the companion line
cord, GA400, is more significant than that of the filter -
which may disqualify this line cord from use in patient
care applications.
EBH Models
01A
03A
8
-
21 29 32 42 45 32 25 19
10 15 25 27 30 27 22
5
06A
10A
-
-
-
5
-
8
3
17 20 24 23 18
12 12 12 12
-
8
Line-to-line in 50 ohm circuit
Electrical Schematic
Current
Frequency-MHz
Rating
EAH Models
01A
1
1.5
5
10
30
LINE
LOAD
5
4
2
1
13
6
28
20
19
15
32
27
25
22
25
28
27
27
3A
06A
10A
5
5
Resistor location for reference only.
Current
Rating
Frequency-MHz
.15
.5
1
5
10
30
EBH Models
All Ratings
1
10
18
30
31
31
24
Catalog 1654001
Revised 05-07
Dimensions are in inches and
Dimensions are shown for
reference purposes only.
Specifications subject
to change.
USA Cust. Svc.: 1-800-468-2023
CORCOM Prods: 1-847-680-7400
www.corcom.com
www.tycoelectronics.com
millimeters unless otherwise
specified. Values in italics
are metric equivalents.
Germany:
49-89-6089-0
CORCOM Product Guide
For UL544 Health Care Equipment (Continued)
EAH, EBH Series
Case Style
Case Dimensions
A
(max)
B
C
D
±.010
±.25
E
(max)
EAH1/EBH1
Part No.
(max) (max)
2.15
54.61
2.15
54.61
1.12
28.45
1.12
.90
22.86
.90
1.575
40.01
1.575
40.01
1.98
50.29
1.98
EAH1
B
EBH1
28.45
22.86
50.29
A
C
Part Numbers
01EAH1
03EAH1
06EAH1
10EAH1
01EBH1
03EBH1
06EBH1
10EBH1
GA400
FA601
D
E
Typical dimensions:
Line Cord No.
Insulating Boot No.
Terminals: .250 [6.35] (5)
Slot: .07 x .16 [1.8 x 4.1]
Holes: .07 [1.8] Dia.(2)
Mounting Holes: .132 [3.35] Dia. (2) with .236 Dia.
x 90˚ countersunk for #4 Flathead Screw
Recommended Panel Cutouts
.820
20.8
.140
3.56
Dia. (2)
R
1.14
29.96
1.575
40.01
.12
3.0
Panel Cutout (Front mount)
Tolerance .005 [0.13]
.900
22.9
.140
3.56
Dia. (2)
R
1.200
30.48
1.575
40.01
.234
5.94
Panel Cutout (Back mount)
Tolerance .005 [0.13]
25
Catalog 1654001
Revised 05-07
Dimensions are in inches and
Dimensions are shown for
reference purposes only.
Specifications subject
to change.
USA Cust. Svc.: 1-800-468-2023
CORCOM Prods: 1-847-680-7400
www.corcom.com
www.tycoelectronics.com
millimeters unless otherwise
specified. Values in italics
are metric equivalents.
Germany:
49-89-6089-0
相关型号: