RPS-120S-27 [MEANWELL]

120W 3"×2" Reliable Green Medical Power Supply;
RPS-120S-27
型号: RPS-120S-27
厂家: MEAN WELL ENTERPRISES CO., LTD.    MEAN WELL ENTERPRISES CO., LTD.
描述:

120W 3"×2" Reliable Green Medical Power Supply

文件: 总4页 (文件大小:116K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
120W 3"×2" Reliable Green Medical Power Supply  
RPS-120S series  
EN 60601-1  
IEC 60601-1  
+ +  
8
ANSI/AAMI ES60601-1  
EN60601-1  
IEC60601-1  
Features  
Applications  
Oral irrigator  
3"×2" compact size  
120W convention, 150W peak (10sec.)  
Hemodialysis machine  
Medical monitors  
Sleep apnea devices  
Pumps machine  
Medical safety approved (2 x MOPP) accroding to  
ANSI/AAMI ES60601-1 and IEC/EN60601-1  
Suitable for BF application with appropriate system consideration  
EMI for both Class& Class configuration  
-30~+85wide range operating temperature  
No load power consumption<0.3W  
Extremely low leakage current  
Protections: Short circuit / Overload / Over voltage / Over temperature  
Operating altitude up to 4000 meters (Note.6)  
3 years warranty  
Description  
RPS-120S is a 120W highly reliable green PCB type medical power supply with a high power density on  
a 3” by 2” footprint. It accepts 80~264VAC input and offers various models with the output voltages between  
12V and 48V. The working efficiency is up to 94% and the extremely low no load power consumption is down  
below 0.3W. RPS-120S is able to be used for both Class(with FG) & Class(no FG) system design. The  
extremely low leakage current is less than 150 A. In addition, it conforms to the international medical regulations  
μ
(2*MOPP)  
equipment.  
and EMC EN55011, perfectly fitting all kinds of BF rated “patient contact” medical system  
Model Encoding  
RPS -  
- 12  
120S  
Output voltage  
Rated wattage  
Series name  
File Name:RPS-120S-SPEC 2019-09-02  
120W 3"×2" Reliable Green Medical Power Supply  
RPS-120S series  
SPECIFICATION  
MODEL  
RPS-120S-12  
12V  
RPS-120S-15  
15V  
RPS-120S-24  
24V  
RPS-120S-27  
27V  
RPS-120S-48  
48V  
DC VOLTAGE  
Peak(10 sec.)  
Convection  
11.8A  
9.5A  
6.25A  
5.55A  
3.125A  
2.5A  
CURRENT  
9.5A  
7.6A  
5A  
4.44A  
Peak(10 sec.) 141.6W  
114W  
142.5W  
114W  
150W  
149.8W  
119.9W  
150mVp-p  
25.6 ~ 28.4V  
1.0%  
150W  
RATED  
POWER  
120W  
120W  
Convection  
RIPPLE & NOISE (max.) Note.2 100mVp-p  
VOLTAGE ADJ. RANGE 11.4~12.6V  
120mVp-p  
14.3~15.8V  
2.0%  
150mVp-p  
22.8~25.2V  
1.0%  
200mVp-p  
45.6 ~50.4V  
1.0%  
OUTPUT  
VOLTAGE TOLERANCE Note.3  
LINE REGULATION  
LOAD REGULATION  
SETUP, RISE TIME  
2.0%  
0.5%  
0.5%  
0.5%  
0.5%  
0.5%  
1.0%  
1.0%  
1.0%  
1.0%  
1.0%  
600ms, 30ms/230VAC  
600ms, 30ms/115VAC at full load  
HOLD UP TIME (Typ.)  
15ms/230VAC  
15ms/115VAC at full load  
113 ~ 370VDC  
VOLTAGE RANGE Note.4 80 ~ 264VAC  
FREQUENCY RANGE  
POWER FACTOR  
47 ~ 63Hz  
PF>0.94/230VAC  
91%  
PF>0.98/115VAC at full load  
EFFICIENCY (Typ.)  
92%  
93%  
94%  
93.5%  
INPUT  
2.3A/115VAC  
1.1A/230VAC  
AC CURRENT (Typ.)  
INRUSH CURRENT (Typ.)  
LEAKAGE CURRENT(max.) Note.5  
COLD START 30A/115VAC  
60A/230VAC  
Earth leakage current < 150 A/264VAC , touch current < 80 A/264VAC  
μ μ  
130~160% rated output power  
OVERLOAD  
Protection type : Hiccup mode, recovers automatically after fault condition is removed  
16.5 ~ 19.5V 26.4 ~ 31.2V 29.7 ~ 35V  
Protection type : Shut down o/p voltage, re-power on to recover  
52.8 ~ 62.4V  
13.2 ~ 15.6V  
PROTECTION  
OVER VOLTAGE  
OVER TEMPERATURE  
WORKING TEMP.  
Protection type : Shut down o/p voltage, recovers automatically after temperature goes down  
-30 ~ +85(Refer to "Derating Curve")  
20 ~ 90% RH non-condensing  
WORKING HUMIDITY  
STORAGE TEMP.  
-40 ~ +85℃  
ENVIRONMENT  
TEMP. COEFFICIENT  
VIBRATION  
0.03%/(0 ~ 50)  
10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes  
4000 meters  
OPERATING ALTITUDE Note.6  
IEC60601-1, TUV EN60601-1, EAC TP TC 004, UL ANSI / AAMI ES60601-1 (3.1 version),  
SAFETY STANDARDS  
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved; Design refer to EN60335-1 (By request)  
Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP, Secondary-Earth:1xMOPP  
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC  
ISOLATION RESISTANCE  
WITHSTAND VOLTAGE  
ISOLATION RESISTANCE I/P-O/P, I/P-FG, O/P-FG: 100M Ohms / 500VDC / 25/ 70% RH  
Parameter  
Standard  
Test Level / Note  
EN55011 (CISPR11)  
EN55011 (CISPR11)  
EN61000-3-2  
Class B  
Conducted emission  
Radiated emission  
Harmonic current  
Voltage flicker  
EN60601-1-2  
Parameter  
EMC EMISSION  
Class: Class B , Class : Class A  
Class A  
SAFETY &  
EN61000-3-3  
-----  
EMC  
(Note 7)  
Standard  
Test Level / Note  
ESD  
EN61000-4-2  
Level 4, 15KV air ; Level 4, 8KV contact  
Level 3, 10V/m( 80MHz~2.7GHz )  
Table 9, 9~28V/m( 385MHz~5.78GHz )  
Level 3, 2KV  
RF field susceptibility  
EN61000-4-3  
EN61000-4-4  
EN61000-4-5  
EN61000-4-6  
EN61000-4-8  
EFT bursts  
EMC IMMUNITY  
Surge susceptibility  
Conducted susceptibility  
Magnetic field immunity  
Level 4, 4KV/Line-FG; 2KV/Line-Line  
Level 3, 10V  
Level 4, 30A/m  
95% dip 1 periods, 30% dip 25 periods,  
95% interruptions 250 periods  
Voltage dip, interruption  
EN61000-4-11  
MTBF  
468Khrs min. MIL-HDBK-217F (25)  
76.2*50.8*28mm or 3" * 2" *1.1" inch  
0.13Kg; 100pcs/14Kg/1.13CUFT  
OTHERS DIMENSION (L*W*H)  
PACKING  
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.  
μ
μ
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1 f & 47 f parallel capacitor.  
3. Tolerance : includes set up tolerance, line regulation and load regulation.  
4. Derating may be needed under low input voltages. Please check the derating curve for more details.  
5. Touch current was measured from primary input to DC output.  
NOTE  
6. The ambient temperature derating of 3.5/1000m with fanless models and of 5/1000m with fan models for operating altitude higher than 2000m(6500ft).  
7. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by  
mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets  
EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”  
(as available on http://www.meanwell.com)  
File Name:RPS-120S-SPEC 2019-09-02  
120W 3"×2" Reliable Green Medical Power Supply  
RPS-120S series  
Block Diagram  
fosc : 85KHz  
RECTIFIERS  
RECTIFIERS  
POWER  
EMI  
FILTER  
+V  
-V  
I/P  
FG  
&
PFC  
&
SWITCHING  
FILTER  
DETECTION  
CIRCUIT  
O.T.P.  
O.L.P.  
CONTROL  
O.V.P.  
Derating Curve  
Convection  
100  
80  
70  
60  
50  
30  
(HORIZONTAL)  
70 85  
-30  
0
10  
20  
30  
40  
50  
AMBIENT TEMPERATURE ()  
Output Derating VS Input Voltage  
(for 24V/27V/48V)  
120W  
114W  
(for 12V/15V)  
100W  
80W  
80 90 110 120 130 140 160 180 200 220 240 264  
INPUT VOLTAGE (VAC) 60Hz  
File Name:RPS-120S-SPEC 2019-09-02  
120W 3"×2" Reliable Green Medical Power Supply  
RPS-120S series  
Mechanical Specification  
Unit:mm  
SVR1  
C100  
HS1  
1
2
3
4
T1  
FS1  
CN100  
C5  
3
2
1
C105  
C119  
C1  
T3.15A  
/250V  
FS2  
69.9  
3.175  
76.2  
AC Input Connector (CN1) : JST B3P-VH or equivalent  
DC Output Connector (CN2) : JST B4P-VH or equivalent  
Pin No. Assignment  
Mating Housing  
Terminal  
Pin No. Assignment  
Mating Housing  
Terminal  
1
2
3
AC/L  
No Pin  
AC/N  
1,2  
3,4  
+V  
-V  
JST VHR  
or equivalent  
JST SVH-21T-P1.1  
or equivalent  
JST VHR  
or equivalent  
JST SVH-21T-P1.1  
or equivalent  
1.HS1 must have safety isolation distance  
with system case.  
Note :  
1.RPS-120S model delivers EMI Class B for both conducted emission and radiated emission for the power supply,  
when configured into Class (with FG) system.  
2.RPS-120S model delivers EMI Class B conducted emission and Class A radiated emission with King Core K5B RC (12*15*7)  
in output cable for the power supply when configured into Class (no FG) system.  
Installation Manual  
Please refer to : http://www.meanwell.com/manual.html  
File Name:RPS-120S-SPEC 2019-09-02  

相关型号:

RPS-120S-48

120W 3"×2" Reliable Green Medical Power Supply
MEANWELL

RPS-16-10/2.0-4

SLEEVE YELLOW BX250 Inhalt pro Packung: 250 Stk.
ETC

RPS-16-10/2.0-9

SLEEVE WHITE BX250 Inhalt pro Packung: 250 Stk.
ETC

RPS-160

160W Single Output Medical Type
MEANWELL

RPS-160-12

160W Single Output Medical Type
MEANWELL

RPS-160-15

160W Single Output Medical Type
MEANWELL

RPS-160-24

160W Single Output Medical Type
MEANWELL

RPS-160-48

160W Single Output Medical Type
MEANWELL

RPS-160-5

160W Single Output Medical Type
MEANWELL

RPS-160_11

160W Single Output Medical Type
MEANWELL

RPS-160_12

160W Single Output Medical Type
MEANWELL

RPS-18-12/2.0-4

SLEEVE YELLOW BX250 Inhalt pro Packung: 250 Stk.
ETC