RPS-500-27-TF [MEANWELL]
500W Reliable Green Medical Power Supply;型号: | RPS-500-27-TF |
厂家: | MEAN WELL ENTERPRISES CO., LTD. |
描述: | 500W Reliable Green Medical Power Supply |
文件: | 总10页 (文件大小:752K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
500W 5"×3" Reliable Green Medical Power Supply
RPS-500 series
EN 60601-1
IEC 60601-1
+ +
8
ANSI/AAMI ES60601-1
EN60601-1
IEC60601-1
■ Features
ꢀ5"×3" compact size
■ Applications
ꢀOral irrigator
ꢀ320W convection,500W force air
ꢀHemodialysis machine
ꢀMedical computer monitors
ꢀSleep apnea devices
ꢀPump machine
ꢀ550W peak power (3sec.)
ꢀMedical safety approved (2 x MOPP) accroding to
ANSI/AAMI ES60601-1 and IEC/EN60601-1
ꢀEMI for both ClassⅠ& Class Ⅱ configuration
ꢀ-30~+70℃ wide range operating temperature
ꢀNo load power consumption<0.5W by PS_ON control
ꢀHigh efficiency up to 94%
ꢀElectric bed
ꢀProtections: Short circuit / Overload / Over voltage /
Over temperature
ꢀ5Vdc standby output, 12Vdc fan supply, Power Good,
Power Fail and remote sense
ꢀOperating altitude up to 4000 meters (Note.5)
ꢀLED indicator for power on
ꢀ3 years warranty
■ Description
RPS-500 is a 500W highly reliable green PCB type medical power supply with a high power density on
the 5 by 3 footprint. It accepts 80~264VAC input and offers various output voltages between 12V and
48V.The working efficiency is up to 94% and the extremely low no load power consumption is down below
0.5W.RPS-500 (blank type only) is able to be used for both ClassⅠ(with FG) and ClassⅡ(no FG) system
μ
design. The extremely low leakage current is less than 220 A. In addition, it conforms to international m
regulations (2*MOPP) and EMC EN55011, perfectlyfitting all kinds of BF rated patient contact medical
system equipment. RPS-500 series also offers the enclosed style models( -C / TF /SF )
■ Model Encoding
RPS - 500 - 12 -C
Type
Output voltage
Rated wattage
Series name
Type
Description
Note
Blank
-C
In stock
In stock
In stock
In stock
PCB Type
Enclosed casing Type
Enclosed Type with fan on the top
Enclosed Type with fan on the side
-TF
-SF
File Name:RPS-500-SPEC 2020-02-14
500W 5"×3" Reliable Green Medical Power Supply
RPS-500 series
SPECIFICATION
MODEL
RPS-500-12
12V
RPS-500-15
15V
RPS-500-27
27V
RPS-500-36
36V
RPS-500-18
18V
RPS-500-24
24V
RPS-500-48
48V
DC VOLTAGE
25CFM
41.6A
33.3A
27.8A
20.8A
18.5A
13.9A
10.4A
Blank
C
Convection 26.7A
21.3A
17.8A
13.4A
11.9A
8.9A
6.7A
RATED
CURRENT
Note.7
25CFM
33.3A
27.8A
20.8A
18.5A
13.9A
10.4A
41.6A
Convection
Built-in fan
25.8A
20.7A
17.2A
27.8A
12.9A
11.5A
8.6A
6.5A
33.3A
20.8A
18.5A
13.9A
500.4W
10.4A
499.2W
TF/SF
Blank
41.6A
499.2W
499.5W
500.4W
499.2W
499.5W
25CFM
Convection
320.4W
499.2W
309.6W
319.5W
320.4W
321.6W
499.2W
309.6W
499.2W
321.3W
320.4W
321.6W
RATED
POWER
Note.7
25CFM
499.5W
310.5W
499.5W
500.4W
309.6W
500.4W
499.5W
310.5W
499.5W
500.4W
309.6W
500.4W
499.2W
312W
C
Convection
499.2W
Built-in fan
499.2W
TF/SF
PEAK POWER(3sec.)
RIPPLE & NOISE (max.)
550W
Note.2 200mVp-p
200mVp-p
200mVp-p
200mVp-p
200mVp-p
200mVp-p
200mVp-p
OUTPUT
11.4~12.6V
3.0%
Note.3
14.3~15.8V 17.1~18.9V 22.8~25.2V 25.6 ~ 28.4V 34.2 ~37.8V 45.6 ~50.4V
VOLTAGE ADJ. RANGE(main output )
3.0%
0.5%
1.0%
3.0%
0.5%
1.0%
2.0%
0.5%
1.0%
2.0%
0.5%
1.0%
1.0%
0.5%
1.0%
1.0%
0.5%
1.0%
VOLTAGE TOLERANCE
LINE REGULATION
LOAD REGULATION
SETUP, RISE TIME
HOLD UP TIME (Typ.)
0.5%
1.0%
1000ms, 30ms/230VAC
1500ms, 30ms/115VAC at full load
10ms/230VAC 10ms/115VAC at full load
80 ~ 264VAC
47 ~ 63Hz
113 ~ 370VDC
VOLTAGE RANGE
FREQUENCY RANGE
POWER FACTOR
Note.4
PF>0.94/230VAC PF>0.98/115VAC at full load
EFFICIENCY (Typ.)
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
92.5%
2.9A/230VAC
93%
93.5%
94%
94%
91%
92%
INPUT
5.8A/115VAC
COLD START 40A/115VAC
80A/230VAC
LEAKAGE CURRENT (max.)
Note.5
Earth leakage current <220 A/264VAC 50Hz , Touch current < 100 A/264VAC
μ
μ
105 ~ 135% rated output power
OVERLOAD
Protection type : Hiccup mode, recovers automatically after fault condition is removed
19.8 ~23.4V
Protection type : Shut down o/p voltage, re-power on to recover
39.6 ~ 46.8V 52.8 ~ 62.4V
16.5 ~ 19.5V
26.4 ~ 31.2V 29.7 ~ 35.1V
13.2 ~ 15.6V
PROTECTION
OVER VOLTAGE
Protection type : Shut down o/p voltage, recovers automatically after temperature goes down
OVER TEMPERATURE
5Vsb : 5V@0.6A without fan, 1A with fan 25CFM ;
Tolerance 2%, ripple : 120mVp-p(max.)
5V STANDBY
12V@0.5A for driving fan ;
Tolerance 10%
12V FAN SUPPLY
PS-ON INPUT SIGNAL
FUNCTION
Power ON: PS-ON = "Hi" or " > 2 ~ 5V" ;
Power OFF: PS-ON = "Low" or " < 0 ~ 0.5V"
500ms>PG>10ms ; The TTL signal goes high with 10ms to 500ms delay after power set up ;
The TTL signal goes low at least 1ms before Vo below 90% of rated value
POWER GOOD / POWER FAIL
-30 ~ +70℃ (Refer to "Derating Curve")
20 ~ 90% RH non-condensing
-40 ~ +85℃
WORKING TEMP.
WORKING HUMIDITY
STORAGE TEMP.
TEMP. COEFFICIENT
VIBRATION
ENVIRONMENT
0.03%/℃ (0 ~ 50℃)
10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes
4000 meters
OPERATING ALTITUDE
Note.6
File Name:RPS-500-SPEC 2020-02-14
500W 5"×3" Reliable Green Medical Power Supply
RPS-500 series
SPECIFICATION
IEC60601-1, TUV EN60601-1,
UL ANSI / AAMI ES60601-1 (3.1 version),EAC TP TC 020.
SAFETY STANDARDS
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved;
Design refer to EN60335-1 (By request)
Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP, Secondary-Earth:1xMOPP
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
ISOLATION LEVEL
WITHSTAND VOLTAGE
ISOLATION RESISTANCE I/P-O/P, I/P-FG:100M Ohms / 500VDC / 25℃/ 70% RH
Parameter
Standard
Test Level / Note
EN55011 (CISPR11)
Conducted emission
Radiated emission
Harmonic current
Voltage flicker
ClassⅠ: Class B , Class Ⅱ: Class A
EMC EMISSION
EN55011 (CISPR11)
EN61000-3-2
Class A
Class A
-----
SAFETY &
EMC
EN61000-3-3
(Note 8)
EN55024 , EN60601-1-2, EN61204-3
Parameter
Standard
Test Level / Note
EN61000-4-2
ESD
Level 4, 15KV air ; Level 4, 8KV contact
Level 3, 10V/m( 80MHz~2.7GHz )
EN61000-4-3
RF field susceptibility
Table 9, 9~28V/m( 385MHz~5.78GHz )
EMC IMMUNITY
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
Level 3, 2KV
EFT bursts
Surge susceptibility
Conducted susceptibility
Magnetic field immunity
Level 4, 4KV/Line-FG ; 2KV/Line-Line
Level 3, 10V
Level 4, 30A/m
100% dip 1 periods, 30% dip 25 periods,
100% interruptions 250 periods
Voltage dip, interruption
EN61000-4-11
MTBF
194.1Khrs min. MIL-HDBK-217F (25℃)
RPS-500-SF
RPS-500-C
RPS-500-TF
130x86x66.5mm
Type
RPS-500
DIMENSION
130x86x43mm
127x76.2x41mm
160x86x43mm
L*W*H
OTHERS
5″x3″x1.61″inch
0.46Kg
5.11″x3.39″x1.69″inch 5.11″x3.39″x2.62″inch 6.3″x3.39″x1.69″inch
P.W.
0.58Kg
0.64Kg
0.54Kg
24pcs
24pcs
Q’TY
30pcs
24pcs
PACKING
16.4Kg
14Kg
G.W.
14.8Kg
14.9Kg
M’MENT
0.77CUFT
0.86CUFT
0.91CUFT
1.03CUFT
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
μ
μ
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1 f & 47 f parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. Derating may be needed under low input voltages. Please check the derating curve for more details.
5. Touch current was measured from primary input to DC output.
6. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher
than 2000m(6500ft).
NOTE
7. Please refer to Derating curve.
8. The power supply is considered a component which will be installed into a final equipment. All EMC tests are
executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still
meets EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies.
(as available on http://www.meanwell.com).
Conducted
Class B
Radiated
Class A
Class A
EMI Performance
(with FG)
ClassⅠ
Class Ⅱ
Class A
(no FG)
File Name:RPS-500-SPEC 2020-02-14
500W 5"×3" Reliable Green Medical Power Supply
RPS-500 series
Block Diagram
PFC fosc : 90KHz
PWM fosc : 100KHz
+5Vsb
DETECTION
CIRCUIT
RECTIFIERS
&
FILTER
POWER SWITCHING
PWM CONTROL
+5Vsb
RECTIFIERS
&
FILTER
+12V Fan Supply
+S
RECTIFIERS
RECTIFIERS
&
FILTER
EMI
FILTER
+V
-V
POWER
SWITCHING
I/P
FG
&
PFC
-S
PG
PG
DETECTION
CIRCUIT
CONTROL
O.L.P.
O.V.P.
PS/ON
CONTROL
PS-ON
Output Derating vs Input Voltage
100
90
80
70
60
50
40
80 90 100 115 120 140 160 180 200 220 240 264
INPUT VOLTAGE (VAC) 60Hz
File Name:RPS-500-SPEC 2020-02-14
500W 5"×3" Reliable Green Medical Power Supply
RPS-500 series
Derating Curve
RPS-500
◎
RPS-500-C
◎
With 25CFM fan
With 25CFM fan
500
500
400
320
400
310
300
310
300
250
240
250
240
220
RPS-500:230VAC Converction
RPS-500:115VAC Converction
200
100
200
160
RPS-500-C:230VAC Converction
RPS-500-C:115VAC Converction
100
-30
0
10
20
30
40
50
60
70
(HORIZONTAL)
-30
0
10
20
30
40
50
60
70
(HORIZONTAL)
AMBIENT TEMPERATURE (℃)
AMBIENT TEMPERATURE (℃)
◎
RPS-500-TF/SF
500
400
300
190
100
(HORIZONTAL)
70
-30
0
10
20
30
40
50
60
AMBIENT TEMPERATURE (℃)
RPS-500-SF
Order No.
RPS-500
RPS-500-C
RPS-500-TF
Products
---
500W
Convection
Force Air
320W
500W
310W
500W
---
500W
File Name:RPS-500-SPEC 2020-02-14
500W 5"×3" Reliable Green Medical Power Supply
RPS-500 series
Mechanical Specification
RPS-500 (PCB Type)
127
5.7
115.6
CN95
HS1
HS4
3 1
4 2
21
43
HS3
-V
+V
HS2
CN1
1
3
1
2
F8A
250V
FS1
FG
Top View
Side View
Air flow
direction
25CFM FAN
52
115.6
5.7
1
1
Bottom View
Side View
File Name:RPS-500-SPEC 2020-02-14
500W 5"×3" Reliable Green Medical Power Supply
RPS-500 series
RPS-500-C (Enclosed type)
Case No. 247C-T 269A-D Unit:mm
126
120
5.0
2
2
110.5
15.5
R1.75
Side View
Side View
Top View
Side View
Air flow
direction
25CFM FAN
53.5
Side View
130
5
120
1
1
1
1
Bottom View
File Name:RPS-500-SPEC 2020-02-14
500W 5"×3" Reliable Green Medical Power Supply
RPS-500 series
RPS-500-TF (Enclosed type with fan on the top)
Case No. 269A-D 247D-T Unit:mm
126
5.0
120
2
2
15.5
Side View
Side View
Side View
Top View
Air flow
direction
Side View
130
5
120
1
1
1
1
Bottom View
File Name:RPS-500-SPEC 2020-02-14
500W 5"×3" Reliable Green Medical Power Supply
RPS-500 series
RPS-500-SF (Enclosed type with fan on the side)
Case No. 248B Unit:mm
160
Side View
Air flow
direction
Side View
Side View
Top View
Side View
130
5.0
1
1
120
Bottom View
File Name:RPS-500-SPEC 2020-02-14
500W 5"×3" Reliable Green Medical Power Supply
RPS-500 series
※ Mounting Instruction for -C/-TF/-SF Type
Hole No. Recommended Screw Size
MAX. Penetration Depth L
Recommended mounting torque
1
2
M3
M3
2mm
4mm
4~6Kgf-cm
4~6Kgf-cm
Chassis of RPS-500-C/TF/SF
Mounting Surface
Mounting Screw
L
CONNECTION
※
AC Input Connector (CN1) : JST B3P-VH or equivalent
Function Connector(CN11): TKP DH2I-2X2 or equivalent
Pin No. Assignment
Mating Housing
Terminal
Pin No. Assignment
Mating Housing
Terminal
1
2
3
AC/N
No Pin
AC/L
1
2
3
4
-S
+S
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
TKP DH2
or equivalent
TKP
or equivalent
DC COM
PG
DC Output Connector (CN2,CN3)
Function Connector(CN95): TKP DH2L-2X2 or equivalent
Pin No. Assignment
Output Terminals
M3.5 Pan HD screw in 2 positions
Torque to 8 lbs-in(90cNm)max.
Pin No. Assignment
Mating Housing
Terminal
CN2
CN3
-V
1
2,4
3
5Vsb
DC COM
PS-ON
+V
TKP DH2
or equivalent
TKP
or equivalent
FAN Connector(CN12) : TKP 8812-2 or equivalent
(Except for RPS-500-TF/SF)
HS1,HS2,HS3,HS4 can not be shorted
Pin No. Assignment
Mating Housing
Terminal
1
2
DC COM
+12V
TKP 2502
or equivalent
TKP 8811
or equivalent
※
Note:
1. The enclosed type (-C/TF/SF type) models are not suitable for configuration within a ClassⅡ (without FG) system, but
suggested within a ClassⅠ(with FG) system.
2. Mounting Instruction for enclosed type.
Installation Manual
Please refer to : http://www.meanwell.com/manual.html
File Name:RPS-500-SPEC 2020-02-14
相关型号:
RPS-501786-001
0.50mm (.020") Pitch FFC-to-Board Connector, Plug Jacket Cover, with Ground type, 30 Circuits
MOLEX
RPS-501864-001
0.50mm (.020") Pitch FFC-to-Board Connector, Plug Jacket Cover, with Ground type, 30 Circuits
MOLEX
©2020 ICPDF网 联系我们和版权申明