RPT-75D [MEANWELL]

75W Triple Output Medical Type; 75W三路输出医用型
RPT-75D
型号: RPT-75D
厂家: MEAN WELL ENTERPRISES CO., LTD.    MEAN WELL ENTERPRISES CO., LTD.
描述:

75W Triple Output Medical Type
75W三路输出医用型

文件: 总3页 (文件大小:527K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
75W Triple Output Medical Type  
RPT-75 series  
Features :  
Universal AC input / Full range  
Low leakage current<200uA  
Protections: Short circuit / Overload / Over voltage  
Free air convection for rated power and 23.5CFM forced air  
convection for peak load  
UL60601-1/IEC60601-1/EN60601-1 medical safety approved  
Fixed switching frequency at 65KHz  
3 years warranty  
EN60601-1  
IEC60601-1  
SPECIFICATION  
MODEL  
RPT-75A  
RPT-75B  
RPT-75C  
OUTPUT NUMBER  
CH1  
5V  
CH2  
12V  
CH3  
CH1  
CH2  
12V  
CH3  
CH1  
CH2  
CH3  
DC VOLTAGE  
-5V  
5V  
-12V  
5V  
15V  
-15V  
RATED CURRENT  
CURRENT RANGE  
RATED POWER  
6A  
3A  
0.5A  
6A  
3A  
0.5A  
6A  
2.3A  
0.5A  
0.6 ~ 8A  
68.5W  
93W  
0.2 ~ 4A  
0.1 ~ 1A  
0.6 ~ 8A  
72W  
0.2 ~ 4A  
0.1 ~ 1A  
0.6 ~ 8A  
72W  
0.1 ~ 3A  
0.1 ~ 1A  
PEAK LOAD (23.5CFM)  
100W  
80mVp-p  
100W  
80mVp-p  
RIPPLE & NOISE (max.) Note.2 80mVp-p  
120mVp-p  
120mVp-p  
120mVp-p  
120mVp-p  
150mVp-p  
150mVp-p  
OUTPUT  
VOLTAGE ADJ. RANGE  
CH1:4.75 ~ 5.5V  
VOLTAGE TOLERANCE Note.3 2.0%  
6.0%  
1.0%  
3.0%  
5.0%  
1.0%  
1.0%  
2.0%  
0.5%  
1.5%  
6.0%  
1.0%  
3.0%  
5.0%  
1.0%  
1.0%  
2.0%  
0.5%  
1.5%  
8.0%  
1.0%  
3.0%  
5.0%  
1.0%  
1.0%  
LINE REGULATION  
LOAD REGULATION  
SETUP, RISE TIME  
0.5%  
1.5%  
500ms, 30ms/230VAC  
500ms, 30ms/115VAC at full load  
20ms/115VAC at full load  
127 ~ 370VDC  
HOLD UP TIME (Typ.)  
VOLTAGE RANGE  
80ms/230VAC  
90 ~ 264VAC  
47 ~ 63Hz  
76%  
FREQUENCY RANGE  
EFFICIENCY(Typ.)  
77%  
77%  
INPUT  
AC CURRENT (Typ.)  
INRUSH CURRENT (Typ.)  
LEAKAGE CURRENT  
1.5A/115VAC  
1A/230VAC  
COLD START 25A/115VAC  
50A/230VAC  
Earth leakage current <200uA / 264VAC, Patient leakage current <100uA/264VAC  
140 ~ 180% rated output power  
OVERLOAD  
Protection type : Hiccup mode, recovers automatically after fault condition is removed  
CH1: 5.75 ~ 6.75V  
PROTECTION  
ENVIRONMENT  
SAFETY &  
OVER VOLTAGE  
Protection type : Shut down o/p voltage, re-power on to recover  
-20 ~ +70 (Refer to output load derating curve)  
20 ~ 90% RH non-condensing  
WORKING TEMP.  
WORKING HUMIDITY  
STORAGE TEMP., HUMIDITY  
-40 ~ +85 , 10 ~ 95% RH  
0.03%/ (0 ~ 45  
)
TEMP. COEFFICIENT  
VIBRATION  
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes  
UL60601-1, TUV EN60601-1, IEC60601-1 approved  
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:1.5KVAC  
I/P-O/P, I/P-FG, O/P-FG:100M Ohms/500VDC  
SAFETY STANDARDS  
WITHSTAND VOLTAGE  
ISOLATION RESISTANCE  
EMI CONDUCTION & RADIATION  
HARMONIC CURRENT  
EMS IMMUNITY  
EMC  
(Note 4)  
Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B  
Compliance to EN61000-3-2,-3  
Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2, EN61000-6-2, EN61204-3, heavy industry level, EN61204-3 medical level, criteria A  
MTBF  
521.2K hrs min. MIL-HDBK-217F (25  
127*76.2*31mm (L*W*H)  
)
OTHERS  
NOTE  
DIMENSION  
PACKING  
0.26Kg; 63pcs/17.4Kg/1.35CUFT  
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.  
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.  
3. Tolerance : includes set up tolerance, line regulation and load regulation.  
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets  
EMC directives.  
5. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.  
File Name:RPT-75-SPEC 2008-03-25  
75W Triple Output Medical Type  
RPT-75 series  
Features :  
Universal AC input / Full range  
Low leakage current<200uA  
Protections: Short circuit / Overload / Over voltage  
Free air convection for rated power and 23.5CFM forced air  
convection for peak load  
UL60601-1/IEC60601-1/EN60601-1 medical safety approved  
Fixed switching frequency at 65KHz  
3 years warranty  
EN60601-1  
IEC60601-1  
SPECIFICATION  
MODEL  
RPT-75D  
RPT-7503  
OUTPUT NUMBER  
CH1  
5V  
CH2  
CH3  
12V  
CH1  
CH2  
5V  
CH3  
12V  
DC VOLTAGE  
24V  
3.3V  
RATED CURRENT  
CURRENT RANGE  
RATED POWER  
5A  
1.5A  
1A  
6A  
6A  
1A  
0.6 ~ 7A  
73W  
95W  
0.1 ~ 2A  
0.1 ~ 1A  
0.7 ~ 7A  
61.8W  
81W  
0 ~ 8A  
0 ~ 1.5A  
PEAK LOAD (23.5CFM)  
RIPPLE & NOISE (max.) Note.2 80mVp-p  
VOLTAGE ADJ. RANGE  
VOLTAGE TOLERANCE Note.3 2.0%  
200mVp-p  
120mVp-p  
80mVp-p  
------  
120mVp-p  
120mVp-p  
OUTPUT  
CH1:4.75 ~ 5.5V  
8.0%  
1.0%  
3.0%  
8.0%  
1.0%  
3.0%  
4.0%  
6.0%  
1.0%  
+10,-6%  
1.5%  
LINE REGULATION  
LOAD REGULATION  
SETUP, RISE TIME  
0.5%  
1.0%  
1.5%  
+3,-4%  
+5,-4%  
6.0%  
500ms, 30ms/230VAC  
500ms, 30ms/115VAC at full load  
HOLD UP TIME (Typ.)  
VOLTAGE RANGE  
80ms/230VAC  
90 ~ 264VAC  
47 ~ 63Hz  
79%  
20ms/115VAC at full load  
127 ~ 370VDC  
FREQUENCY RANGE  
EFFICIENCY(Typ.)  
74%  
INPUT  
AC CURRENT (Typ.)  
INRUSH CURRENT (Typ.)  
LEAKAGE CURRENT  
1.5A/115VAC  
1A/230VAC  
COLD START 25A/115VAC  
50A/230VAC  
Earth leakage current <200uA / 264VAC, Patient leakage current <100uA/264VAC  
140 ~ 180% rated output power  
OVERLOAD  
Protection type : Hiccup mode, recovers automatically after fault condition is removed  
PROTECTION  
ENVIRONMENT  
SAFETY &  
CH1: 5.75 ~ 6.75V  
CH1: 3.8 ~ 4.45V  
OVER VOLTAGE  
Protection type : Shut down o/p voltage, re-power on to recover  
-20 ~ +70 (Refer to output load derating curve)  
20 ~ 90% RH non-condensing  
WORKING TEMP.  
WORKING HUMIDITY  
STORAGE TEMP., HUMIDITY  
-40 ~ +85 , 10 ~ 95% RH  
0.03%/ (0 ~ 45  
)
TEMP. COEFFICIENT  
VIBRATION  
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes  
UL60601-1, TUV EN60601-1, IEC60601-1 approved  
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:1.5KVAC  
I/P-O/P, I/P-FG, O/P-FG:100M Ohms/500VDC  
SAFETY STANDARDS  
WITHSTAND VOLTAGE  
ISOLATION RESISTANCE  
EMI CONDUCTION & RADIATION  
HARMONIC CURRENT  
EMS IMMUNITY  
EMC  
(Note 4)  
Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B  
Compliance to EN61000-3-2,-3  
Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2, EN61000-6-2, EN61204-3 heavy industry level, EN61204-3 medical level, criteria A  
MTBF  
521.2K hrs min. MIL-HDBK-217F (25  
127*76.2*31mm (L*W*H)  
)
OTHERS  
NOTE  
DIMENSION  
PACKING  
0.26Kg; 63pcs/17.4Kg/1.35CUFT  
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.  
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.  
3. Tolerance : includes set up tolerance, line regulation and load regulation.  
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets  
EMC directives.  
5. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time.  
File Name:RPT-75-SPEC 2008-03-25  
75W Triple Output Medical Type  
RPT-75 series  
Mechanical Specification  
Unit:mm  
FAN  
23.5CFM min.  
CN2  
1
2
3
4
5
6
7
8
CN1  
3
2
1
AC Input Connector (CN1) : JST B3P-VH or equivalent  
Pin No. Assignment  
Mating Housing  
Terminal  
1
2
3
AC/N  
No Pin  
AC/L  
JST VHR  
or equivalent  
JST SVH-21T-P1.1  
or equivalent  
DC Output Connector (CN2) : JST B8P-VH or equivalent  
Pin No. Assignment  
Mating Housing  
Terminal  
115.8  
127  
5.6  
1,2  
3,4,5  
6,7  
8
V1  
COM  
V2  
JST VHR  
or equivalent  
JST SVH-21T-P1.1  
or equivalent  
V3  
Block Diagram  
fosc : 65KHz  
RECTIFIERS  
&
FILTER  
V3  
RECTIFIERS  
&
FILTER  
+V2  
RECTIFIERS  
&
FILTER  
RECTIFIERS  
&
FILTER  
EMI  
FILTER  
POWER  
SWITCHING  
+V1  
COM  
I/P  
FG  
DETECTION  
CIRCUIT  
CONTROL  
O.L.P.  
O.V.P.  
Derating Curve  
Output Derating VS Input Voltage  
Ta=25  
100  
90  
133  
23.5CFM Forced air Cooling  
Convection Cooling  
100  
80  
70  
80  
50  
60  
50  
40  
20  
-20  
0
10  
20  
30  
40  
50  
60  
70  
(HORIZONTAL)  
90 95 100 115 120 140 160 180 200 220 240 264  
AMBIENT TEMPERATURE (  
)
INPUT VOLTAGE (VAC) 60Hz  
File Name:RPT-75-SPEC 2008-03-25  

相关型号:

RPT-75_11

75W Triple Output Medical Type
MEANWELL

RPT-75_12

75W Triple Output Medical Type
MEANWELL

RPT-82FQ

IC DATACOM, PCM TRANSCEIVER, CDIP16, HERMETIC SEALED, CERAMIC, DIP-16, Digital Transmission Interface
ADI

RPT-82FS

IC DATACOM, PCM TRANSCEIVER, PDSO16, SO-16, Digital Transmission Interface
ADI

RPT-82FSR

IC DATACOM, PCM TRANSCEIVER, PDSO16, SO-16, Digital Transmission Interface
ADI

RPT-83FS

IC DATACOM, PCM TRANSCEIVER, PDSO16, SO-16, Digital Transmission Interface
ADI

RPT-83FSR

IC DATACOM, PCM TRANSCEIVER, PDSO16, SO-16, Digital Transmission Interface
ADI

RPT-85FQ

IC DATACOM, PCM TRANSCEIVER, CDIP16, HERMETIC SEALED, CERDIP-16, Digital Transmission Interface
ADI

RPT-86BIFP

IC DATACOM, PCM TRANSCEIVER, PDIP16, PLASTIC, DIP-16, Digital Transmission Interface
ADI

RPT-86BIFQ

IC DATACOM, PCM TRANSCEIVER, CDIP16, CERDIP-16, Digital Transmission Interface
ADI