MB1048SMRW-12RI4 [MPD]

EN 60601 Approved Compact, 2:1 Input, 10W DC/DC Con vert ers;
MB1048SMRW-12RI4
型号: MB1048SMRW-12RI4
厂家: MICROPOWER DIRECT, LLC    MICROPOWER DIRECT, LLC
描述:

EN 60601 Approved Compact, 2:1 Input, 10W DC/DC Con vert ers

文件: 总3页 (文件大小:388K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
MB1000MRWRI4  
EN 60601 Approved  
Compact, 2:1 Input, 10W  
DC/DC Converters  
Electrical Specifications  
Key Features:  
Specifications typical @ +25°C, nominal input voltage & rated output current, unless otherwise noted. Specifications subject to change without notice.  
EN 60601 3RD Ed. Approved  
10W Output Power  
Input  
Parameter  
Conditions  
12 VDC Input  
Min.  
7.0  
Typ.  
8.0  
Max.  
9.0  
Units  
4.2 kVrms Isolation  
Input Start Voltage  
24 VDC Input 13.0  
48 VDC Input 30.0  
12 VDC Input  
15.0  
33.0  
18.0  
36.0  
8.5  
VDC  
Reinforced Insulation  
1 x MOPP & 2 x MOOP per  
EN 60601-1 3RD Edition &  
ANSI/AAMI ES 60601-1  
10 µA Max Leakage Current  
Wide 2:1 Input Range  
Compact 1 x 2 In Case  
Single & Dual Outputs  
1.0 MH MTBF  
Under Voltage Shutdown  
24 VDC Input  
48 VDC Input  
16.0  
34.0  
VDC  
mW  
Input Filter  
Short Circuit Input Power  
Output  
Parameter  
Output Voltage Accuracy  
Output Voltage Balance  
Line Regulation  
(Pi) Filter  
3,000  
Conditions  
Min.  
Typ.  
Max.  
1.0  
Units  
%
Dual Output , Balanced Loads  
VIN = Min to Max  
0.5  
0.3  
0.5  
0.6  
2.0  
0.5  
1.0  
1.2  
100  
150  
%
%
IOUT = 15% to 100%  
IOUT = 5% to 100%  
5V Output  
Load Regulation  
%
Ripple & Noise (20 MHz), See Note 2  
mV P - P  
All Other Outputs  
Output Power Protection  
Transient Recovery Time, See Note 3  
Transient Response Deviation  
Temperature Coefficient  
Output Short Circuit  
General  
120  
150  
300  
3.0  
%
µSec  
%
600  
5.0  
0.05  
25% Load Step Change  
0.02  
%/ºC  
Continuous (Autorecovery)  
Parameter  
Conditions  
Min.  
60 Seconds 4,200  
300 Vrms  
Typ.  
Max.  
Units  
Vrms  
Isolation Voltage, Rated  
Reinforced Insulation Working Voltage  
Leakage Current  
240 VAC, 60 Hz  
500 VDC  
10  
µA  
G⍀  
pF  
Isolation Resistance  
Isolation Capacitance  
Switching Frequency  
Environmental  
10  
100 kHz, 1V  
60  
150  
80  
180  
120  
kHz  
Parameter  
Conditions  
Min.  
-40  
Typ.  
Max.  
+75  
+95  
Units  
ºC  
ºC  
Ambient  
Case  
Operating Temperature Range  
Storage Temperature Range  
Cooling  
Humidity  
-50  
+125  
ºC  
Free Air Convection  
RH, Non-condensing  
95  
4,000  
%
m
Altitude  
Physical  
Case Size  
See Mechanical Diagram (Page 3)  
Case Material  
Weight  
Reliability Specifications  
Parameter  
Non-Conductive Black Plastic (UL94-V0)  
0.86 Oz (24.5g)  
MicroPower Direct  
292 Page Street  
Suite D  
Stoughton, MA 02072  
USA  
Conditions  
Min.  
Typ.  
Max.  
Units  
MTBF  
MIL HDBK 217F, 25ºC, Gnd Benign  
1.0  
MHours  
IEC/EN 60601-1, EN 60601-1 3RD Edition, 1xMOPP & 2xMOOP  
ANSI/AAMI ES 60601-1 1xMOPP & 2xMOOP Recognition, (UL Certificate)  
ANSI/AAMI ES 60601-1, CAN/CSA-C22.2 No.60601-1  
Safety Standards  
T: (781) 344-8226  
F: (781) 344-8481  
E: sales@micropowerdirect.com  
W: www.micropowerdirect.com  
Absolute Maximum Ratings  
Parameter  
Conditions  
12 VDC Input  
24 VDC Input  
48 VDC Input  
1.5 mm From Case For 10 Sec  
Min.  
Typ.  
Max.  
25.0  
50.0  
100.0  
260  
Units  
VDC  
ºC  
Input Voltage Surge (1 Sec)  
Lead Temperature  
Caution: Exceeding Absolute Maximum Ratings may damage the module. These are not continuous operating ratings.  
www.micropowerdirect.com  
Model Selection Guide  
www.micropowerdirect.com  
Input  
Output  
Reflected  
Ripple  
Capacitive Fuse Rating  
Efficiency  
Model  
Voltage (VDC)  
Current (mA)  
Load  
Slow-Blow  
(mA)  
Voltage Current  
Current  
Number  
Current  
(mA, Typ)  
(%, Typ)  
(µF Max)  
(VDC)  
(mA, Max) (mA, Min)  
Nominal Range Full-Load No-Load  
MB1012SMRW-05RI4  
MB1012SMRW-12RI4  
MB1012DMRW-12RI4  
MB1012DMRW-15RI4  
MB1024SMRW-05RI4  
MB1024SMRW-12RI4  
MB1024DMRW-12RI4  
MB1024DMRW-15RI4  
MB1048SMRW-05RI4  
MB1048SMRW-12RI4  
MB1048DMRW-12RI4  
MB1048DMRW-15RI4  
12  
12  
12  
12  
24  
24  
24  
24  
48  
48  
48  
48  
9.0 - 18.0  
9.0 - 18.0  
9.0 - 18.0  
9.0 - 18.0  
18.0 - 36.0  
18.0 - 36.0  
18.0 - 36.0  
18.0 - 36.0  
36.0 - 75.0  
36.0 - 75.0  
36.0 - 75.0  
36.0 - 75.0  
907  
1,044  
1,042  
1,028  
559  
30  
30  
30  
30  
20  
20  
20  
20  
10  
10  
10  
10  
100  
100  
100  
100  
50  
5.0  
12.0  
12.0  
15.0  
5.0  
1,600  
835  
0.0  
0.0  
0.0  
0.0  
0.0  
0.0  
0.0  
0.0  
0.0  
0.0  
0.0  
0.0  
74  
80  
80  
81  
75  
81  
81  
82  
75  
81  
81  
82  
1,000  
470  
3,000  
3,000  
3,000  
3,000  
1,500  
1,500  
1,500  
1,500  
750  
417  
220  
333  
220  
2,000  
835  
1,000  
470  
516  
50  
12.0  
12.0  
15.0  
5.0  
516  
50  
417  
220  
508  
50  
333  
220  
280  
25  
2,000  
835  
1,000  
470  
258  
25  
12.0  
12.0  
15.0  
750  
258  
25  
417  
220  
750  
254  
25  
333  
220  
750  
Notes:  
Derating Curve  
1. The specified maximum capacitive load is for each output.  
2. When measuring output ripple, it is recommended that an external 0.47 µF ceramic capacitor  
be placed from the +VOUT pin to the -VOUT pin for single output units and from each output to  
common for dual output units. For noise sensitive applications, the use of 3.3 µF capacitors  
will reduce the output ripple.  
3. Transient recovery is measured to within a 1% error band for a load step change of 75% to  
100%.  
3. Dual output units may be connected to provide a 24 VDC or 30 VDC output. To do this, connect  
the load across the positive (+VOUT) and negative (-VOUT) outputs and float the output common.  
4. The converter should be connected to a low ac-impedance source. An input source with a  
highly inductive impedance may affect the stability of the converter. In applications where the  
converter output loading is high and input power is supplied over long lines, it may be necessary  
to use a capacitor on the input to insure start-up.  
In this case, it is recommended that a low ESR (ESR <1.0at 100 kHz) capacitor be mounted  
close to the converter. For 12V input units a 10.0 µF is recommended, for 24V a 4.7 µF and  
for 48V units a 2.2 µF.  
5. It is recommended that a fuse be used on the input of a power supply for protection. See the  
table above for the correct rating.  
Typical Connection  
These converters are specified for operation without external components.  
However, in some applications the addition of input/output capacitors, as shown  
in the typical connection diagram above, will enhance stability and reduce output  
ripple.This simple connection includes a low ESR (<1at 100 kHz) capacitor  
connected across the input (C1). It is recommended that a 10 µF be used  
for 12V input models, a 4.7 µF for 24V and a 2.2 µF for 48V input units.  
To improve the output ripple performance, a 3.3 µF is connected across  
the output. For dual output units, a 3.3 µF capacitor should be connected from  
each output to common.  
EMI Characteristics  
Parameter  
Standard  
EN 55011, EN 55022  
EN 55024  
Criteria  
Level  
Class A  
Class A  
15 kV Air  
8 kV Contact  
10V/m  
Conducted Emissions  
Radiated Emissions  
ESD  
EN 61000-4-2  
A
RS  
EN 61000-4-3  
EN 61000-4-4  
EN 61000-4-5  
EN 61000-4-6  
EN 61000-4-8  
A
A
A
A
A
EFT, See Note at right  
2 kV  
To meet the specified EN 61000-4-4 and EN 61000-4-5 limits, an external  
capacitor must be connected across the input pins of the module (C1). A  
330 µF/100V capacitor is recommended. This capacitor should be mounted as  
close to the module as possible.  
Surge, See Note at right  
1 kV  
CS  
10 Vrms  
30A/m  
PFMF  
Page 2  
We Power Your Success - For Less  
Mechanical Dimensions  
www.micropowerdirect.com  
Pin Connections  
Pin Single  
Dual  
+VIN  
Pin Single  
Dual  
Comm.  
-VOUT  
1
2
3
+VIN  
-VIN  
+VOUT  
4
No Pin  
-VIN  
+VOUT  
5
-VOUT  
Mechanical Notes:  
All dimensions are typical in inches (mm)  
Tolerance x.xx = 0.01 ( 0.25)  
Related Products  
Medical Approved DC/DC’s  
MA300MRURI4 Series  
3W Output Power  
ML200MRRI4 Series  
2W Output Power  
MB2000MRWRI4 Series  
20W Output Power  
4 kVAC Isolation  
4 kVAC Isolation  
4.2 kVAC Isolation  
Reinforced Insulation  
Meets 1xMOPP & 2xMOOP  
2 µA Leakage Current Max  
Wide 4:1 Input Range  
Compact DIP Case  
Single & Dual Outputs  
1.0 MH MTBF  
Reinforced Insulation  
Meets 1xMOPP & 2xMOOP  
2 µA Leakage Current Max  
Compact SMT Case  
Available on Tape/Reel  
2.0 MH MTBF  
Reinforced Insulation  
Meets 1xMOPP & 2xMOOP  
5 µA Leakage Current Max  
Wide 2:1 Input Range  
Compact 1 x 2 In Case  
Single & Dual Outputs  
1.08 MH MTBF  
EN 60601 Approved  
EN 60601 Approved  
EN 60601 Approved  
Thousands of standard power products ranging from 0.5W to 500W are  
available from MPD in a wide variety of packages and pin-outs. This  
includes many more families with EN 60601 medical approval  
MicroPower Direct  
We Power Your Success - For Less!  
292 Page Street Ste D Stoughton, MA 02072  
TEL: (781) 344-8226  
FAX: (781) 344-8481  
E-Mail: sales@micropowerdirect.com  

相关型号:

MB104G

Bridge Rectifier Diode, 10A, 400V V(RRM),
DIODES

MB104W-G

Rectifier Diode,
COMCHIP

MB105

1.0AMP Single Phase Surface Mount Bridge Rectifiers
SECOS

MB105008TA-220.000MHZ-G1

Oscillator, 0.003MHz Min, 220MHz Max, 220MHz Nom,
ILSI

MB105F

Miniature Glass Passivated Single-Phase Surface Mount Flat Bridge Rectifier
PACELEADER

MB105F

Miniature Glass Passivated Single-Phase Surface Mount Flat Bridge Rectifier
ASEMI

MB106

10 Amp Single Phase Bridge Rectifier 50 to 1000 Volts
MCC

MB106

1.0AMP Single Phase Surface Mount Bridge Rectifiers
SECOS

MB106

SINGLE-PHASE SILICON BRIDGE
CHENG-YI

MB106

Bridge Rectifier Diode, 1 Phase, 10A, 600V V(RRM), Silicon
DIODES

MB106-BP

10 Amp Single Phase Bridge Rectifier 50 to 1000 Volts
MCC

MB106G

Bridge Rectifier Diode, 10A, 600V V(RRM),
DIODES