ML105SM-12RI4 [MPD]
Medical Approved Compact SMT, 1W DC/DC Con vert ers;型号: | ML105SM-12RI4 |
厂家: | MICROPOWER DIRECT, LLC |
描述: | Medical Approved Compact SMT, 1W DC/DC Con vert ers |
文件: | 总2页 (文件大小:253K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
ML100MRI4
Medical Approved
Compact SMT, 1W
DC/DC Converters
Electrical Specifications
Specifications typical @ +25°C, nominal input voltage & rated output current, unless otherwise noted. Specifications subject to change without notice.
Key Features:
Input
Parameter
• EN 60601 3RD Ed. Approved
• 1W Output Power
Conditions
Min.
Typ.
5.0
Max.
5.5
Units
5 VDC Input
4.5
• 4 kVrms Isolation
Input Voltage Range
12 VDC Input 10.8
24 VDC Input 21.6
Internal Capacitor
12.0
24.0
13.2
26.4
VDC
• Reinforced Insulation
• 2 x MOPP per EN 60601-1
3RD Edition & ANSI/AAMI
ES 60601-1
Input Filter
Output
Parameter
Output Voltage Accuracy
Output Voltage Balance
Line Regulation
Conditions
Min.
Typ.
1.0
0.1
1.2
Max.
3.0
1.0
1.5
10
Units
%
%
%
%
Dual Outputs, Balanced Loads
For VIN Change of 1%
IOUT = 10% to 100%
See Note 2
• 2 µA Max Leakage Current
• Compact SMT Case
• Single & Dual Outputs
Load Regulation
Ripple & Noise (20 MHz)
Temperature Coefficient
Output Short Circuit
General
100
0.02
mV P - P
%/°C
0.01
Continuos (Autorecovery)
Parameter
Conditions
Min.
60 Seconds 4,000
250 Vrms
Typ.
Max.
2.0
Units
Vrms
Isolation Voltage
Reinforced Insulation Working Voltage
Leakage Current
Isolation Resistance
Isolation Capacitance
Switching Frequency
EMI Characteristics, See Note 3
Parameter
240 VAC, 60 Hz
500 VDC
µA
G⍀
pF
10
100 kHz, 1V
20
80
50
100
kHz
Standard
Criteria
Level
EMC
EMS
Complies With EN 55011 4TH Edition
Complies With EN 60601-1-2
Environmental
Parameter
Conditions
Min.
-25
Typ.
Max.
+95
+105
+125
Units
°C
°C
Ambient
Case
Operating Temperature Range
Storage Temperature Range
Cooling
Humidity
-50
°C
Free Air Convection
RH, Non-condensing
95
%
Physical
Case Size
See Mechanical Drawing (Page 2)
Case Material
Weight
Non-Conductive Black Plastic (UL94-V0)
0.14 Oz (4.1g)
Reliability Specifications
Parameter
MTBF
MicroPower Direct
292 Page Street
Suite D
Stoughton, MA 02072
USA
Conditions
Min.
2.0
Typ.
Max.
Units
MHours
MIL HDBK 217F, 25°C, Gnd Benign
IEC/EN 60601-1, EN 60601-1 3RD Edition, 2xMOPP
Safety Standards
ANSI/AAMI ES 60601-1, 2xMOPP Recognition, (UL Certificate)
ANSI/AAMI ES 60601-1, CAN/CSA-C22.2 No.60601-1
IPC/JEDEC J-STD-020D.1 Level 2
T: (781) 344-8226
F: (781) 344-8481
E: sales@micropowerdirect.com
W: www.micropowerdirect.com
Moisture Sensitivity Level, See Note 4
Absolute Maximum Ratings
Parameter
Conditions
Min.
Typ.
Max.
9.0
18.0
30.0
245
Units
VDC
°C
5 VDC Input
Input Voltage Surge (1 Sec)
12 VDC Input
24 VDC Input
See Note 5
Peak Reflow Temperature
Caution: Exceeding Absolute Maximum Ratings may damage the module. These are not continuous operating ratings.
www.micropowerdirect.com
Model Selection Guide
www.micropowerdirect.com
Input
Output
Capacitive
Fuse Rating
Slow-Blow
(mA)
Model
Number
Efficiency
(%, Typ)
Voltage (VDC)
Current (mA)
Full-Load No-Load
Load
(µF, Max)
Voltage Current
Current
(VDC)
(mA, Max) (mA, Min)
Nominal
Range
ML105SM-05RI4
ML105SM-12RI4
ML105SM-15RI4
ML105DM-12RI4
ML105DM-15RI4
ML112SM-05RI4
ML112SM-12RI4
ML112SM-15RI4
ML112DM-12RI4
ML112DM-15RI4
ML124SM-05RI4
ML124SM-12RI4
ML124SM-15RI4
ML124DM-12RI4
ML124DM-15RI4
Notes:
5
5
5
5
4.50 - 5.50
4.50 - 5.50
4.50 - 5.50
4.50 - 5.50
4.50 - 5.50
10.8 - 13.2
10.8 - 13.2
10.8 - 13.2
10.8 - 13.2
10.8 - 13.2
21.6 - 26.4
21.6 - 26.4
21.6 - 26.4
21.6 - 26.4
21.6 - 26.4
263
252
246
252
236
110
106
106
106
103
55
50
50
50
50
50
35
35
35
35
35
20
20
20
20
20
5.0
12.0
15.0
12.0
15.0
5.0
12.0
15.0
12.0
15.0
5.0
200.0
84.0
68.0
42.0
33.0
200.0
84.0
68.0
42.0
33.0
200.0
84.0
68.0
42.0
33.0
4.0
1.68
1.36
0.84
0.66
4.0
1.68
1.36
0.84
0.66
4.0
220
220
220
100
100
220
220
220
100
100
220
220
220
100
100
76
80
83
80
84
76
79
80
79
80
76
80
80
80
80
600
600
600
600
600
250
250
250
250
250
125
125
125
125
125
5
12
12
12
12
12
24
24
24
24
24
53
53
53
52
12.0
15.0
12.0
15.0
1.68
1.36
0.84
0.66
Derating Curve
0. The specified maximum capacitive load is for each output.
2. When measuring output ripple, it is recommended that an external ±.47 µF ceramic
capacitor be placed from the +VOUT pin to the -VOUT pin for single output units and
from each output to common for dual output units. For noise sensitive applications,
the use of 3.3 µF capacitors will reduce the output ripple.
3
Contact the factory for required filter components.
4. Any units that are not packaged in a vacuum sealed container should be stored in a
controlled environment. Contact the factory for more information.
5. The recommended reflow settings are a peak temperature of 245 °C for a maxi-
mum period (TPK) of 0±S and a time above liquidous (TL) of <6± seconds at 207 °C.
For more information, please contact the factory.
6. Operation at no-load will not damage these units. However, they may not meet
all specifications.
7. Dual output units may be connected to provide a 24 VDC or 3± VDC output. To do
this, connect the load across the positive (+VOUT) and negative (-VOUT) outputs and
float the output common.
8. The converter should be connected to a low ac-impedance source. An input source
with a highly inductive impedance may affect the stability of the converter. In appli-
cations where the converter output loading is high and input power is supplied over
long lines, it may be necessary to use a capacitor on the input to insure start-up.
In this case, it is recommended that a low ESR (ESR <0.± at 0±± kHz) capacitor
be mounted close to the converter. For 5V input units a 2.2 µF is recommended,
for 02V input units, a 0.± µF; and for 24V units a ±.47 µF.
9. It is recommended that a fuse be used on the input of a power supply for protection.
See the table above for the correct rating.
Mechanical Dimensions
Pin Connections
Pin
Single
1
-VIN
NC
6
Solder Layout
7
NC
8
+VOUT
-VOUT
+VIN
9
14
Pin
1
Dual
-VIN
6
Common
-VOUT
7
8
+VOUT
Common
+VIN
9
14
NC = No Connection
Mechanical Notes:
• All dimensions are typical in inches (mm)
• Tolerance x.xx = ±±.±0 (±±.25)
MicroPower Direct
We Power Your Success - For Less!
292 Page Street Ste D Stoughton, MA 02072
•
TEL: (781) 344-8226
•
FAX: (781) 344-8481
•
E-Mail: sales@micropowerdirect.com