MPM-05MS-12 [MPD]
5W Output Power;型号: | MPM-05MS-12 |
厂家: | MICROPOWER DIRECT, LLC |
描述: | 5W Output Power |
文件: | 总4页 (文件大小:929K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
Series
MPM-05M
Compact Encapsulated,
Medical Approved, 5W
AC/DC Power Supplies
Key Features:
• 5W Output Power
Electrical Specifications
Specifications typical @ +25°C, 230 VAC input voltage & rated output current, unless otherwise noted. Specifications subject to change without notice.
Input
• EN 60601-1 3RD Edition
• Universal 85-264 VAC Input
• Reinforced Insulation (2xMOPP)
• Meets IEC Safety Class II
• Short Circuit Protection
• Over Current Protection
• Over Voltage Protection
• Low Leakage Current
• -25°C to +70°C Operation
Parameter
Conditions
Min.
85
Typ.
Max.
264
370
63
Units
VAC
VDC
Hz
Input Voltage Range
Input Frequency
100
47
115 VAC
230 VAC
10
20
Inrush Current, See Note 2
A Pk
W
No Load Power Consumption
Output
0.3
Parameter
Conditions
Min.
Typ.
2.0
0.5
1.0
50
Max.
Units
Output Voltage Accuracy
Line Regulation
%
%
VIN = Min to Max
IOUT = 10% to 100%
See Note 3
Load Regulation
%
Ripple & Noise (20 MHz)
100
mV Pk-Pk
115 VAC, 60 Hz
230 VAC, 50 Hz
10
Hold-Up Time
mS
80
Temperature Coefficient
Overload Protection
Short Circuit Protection
General
0.02
%/°C
Autorecovery 110
Continuous (Autorecovery)
280
% INOM
Parameter
Conditions
Min.
Typ.
Max.
Units
VAC rms
µA
Isolation Voltage
Input to Output, 60S 4,000
Leakage Current
Isolation Resistance
Switching Frequency
Environmental
80
500 VDC 1,000
M⍀
140
kHz
Parameter
Conditions
Min.
Typ.
Max.
+70
+95
+95
Units
°C
Operating Temperature Range
Maximum Case Temperature
Storage Temperature Range
Cooling
Ambient
Case
-25
+25
°C
-40
°C
Free Air Convection (See Derating Curve)
Humidity
RH, Non-condensing
95
%RH
°C
260
360
MicroPower Direct
292 Page Street
Suite D
Stoughton, MA 02072
USA
Lead Temperature
See Note 4
Physical
Case Size
See Mechanical Diagram (Page 4)
Case Material
Weight
Non-Conductive Plastic Resin (UL94-V0)
1.52 Oz (43g)
T: (781) 344-8226
F: (781) 344-8481
E: sales@micropowerdirect.com
W: www.micropowerdirect.com
Reliability Specifications
Parameter
Conditions
Min.
Typ.
Max.
Units
MTBF
MIL HDBK 217F, 25°C, Gnd Benign
300
kHours
Safety Standards
Safety Class
Insulation Level
UL 60601, EN 60601
Class II
2XMOPP
www.micropowerdirect.com
Model Selection Guide
www.micropowerdirect.com
Over
Output
Current (mA)
Input Current
Capacitive
Load
(µF, Max)
Voltage
Protection
(VDC, Max)
Model
Number
Efficiency
(%, Typ)
(mA Typ)
Voltage
(VDC)
115 VAC
230 VAC
70
Max.
1,000
420
Min.
0.00
MPM-05MS-05
MPM-05MS-12
MPM-05MS-15
MPM-05MS-24
120
120
120
120
5.0
12.0
15.0
24.0
4,000
820
7.5
76
80
81
81
70
0.00
0.00
0.00
16.0
20.0
30.0
70
333
820
70
230
330
Notes:
6. All units are rated for EN 61000-4-4 ( 2 kV) with the addition
of the MOV and NTC shown in the typical connection below.
They will meet EN 61000-4-4 ( 4 kV) with the additional
input components shown in the typical connection diagram
shown on page 3.
7. All units are rated for EN 61000-4-5 ( 1 kV) with the addition of
the MOV and NTC shown in the typical connection below. They will
meet EN 61000-4-5 ( 2 kV / 4 kV) with the input components
shown in the typical connection diagram on page 3.
1. Operation at no load will not damage the units, however, they
may not meet all specifications.
MPD offers a wide variety of input protection
and filtering modules for use with our AC & DC
power modules. The MACFM-02A may be used
with the MPM-05MS, as shown on page 3. Full
datasheets for all our input modules are available
on our website, or call the factory for a more help.
2. Inrush current is given for a cold start at 25°C.
3. When measuring output ripple, it is recommended that an
external 0.1 µF high frequency ceramic capacitor be placed
in parallel with a 47 µF high frequency electrolytic capacitor
from the +VOUT pin to the -VOUT pin.
8. It is recommended that a fuse be used on the input of a
power supply for protection. For the MPM-05MS series,
a 2.0A/250 VAC slow blow should be used.
4. The max lead temp for wave soldering is 5% for a time
period of 5 to 10S. For hand soldering, it is 10% for 3 to 5S.
5. All units are rated for EN 55022 (CE/RE) class B without
external components.
Typical Connection
The diagram above illustrates a typical application connection of the MPM-05MS series. Notes on this circuit (starting with the input circuit) are:
1. It is recommended that an external fuse and NTC be used. The
recommended fuse is a 2.0A/250 VAC slow blow. The NTC a 5D9.
2. All units are rated for EN 55022 (CE/RE) class B without external
components.
3. All units are rated for EN 61000-4-4 ( 2 kV) with the addition of
the MOV and NTC shown in the typical connection above.They will
meet EN 61000-4-4 ( 4 kV) with the additional input components
shown in the typical connection diagram shown on page 3.
4. All units are rated for EN 61000-4-5 ( 1 kV/ 2 kV) with the addi-
tion of the MOV and NTC shown in the typical connection above.
They will meet EN 61000-4-5 ( 2 kV / 4 kV) with the input
components shown in the typical connection diagram on page 3.
5. If output noise levels lower than the specified limits are required,
the addition of C1 and C2 should be sufficient for most applications.
The recommended values are shown in the table at right. The
output filtering capacitor C2 is a high frequency, low resistance
electrolytic capacitor. Capacitor C1 is ceramic.Voltage derating of
capacitors should be 80% or above.
Model
C1
C2
MPM-05MS-05
MPM-05MS-12
MPM-05MS-15
MPM-05MS-24
1.0 µF/50V
1.0 µF/50V
1.0 µF/50V
1.0 µF/50V
220 µF/10V
100 µF/25V
100 µF/25V
47 µF/35V
6. The TVS is added to protect circuits being powered from damage
if the module fails.
Derating Curve
Input Voltage Vs Load
MicroPower Direct
• 292 Page Street Ste D Stoughton, MA 02072 • TEL: (781) 344-8226 • FAX: (781) 344-8481 • E-Mail: sales@micropowerdirect.com
EMI Characteristics
Parameter
Radiated Emissions
Conducted Emissions
www.micropowerdirect.com
Notes:
Conditions
Criteria
Level
Class B
Class B
8 kV Air
6 kV Contact
10V/m
1. To meet the requirements of EN 61000-
4-4 ( 4 kV), external components are
needed. This can be done discretely (as
shown below), or with the addition of the
MACFM-02A.Contact the factory for more
information.
EN 55022
EN 55022
ESD
EN 61000-4-2
EN 61000-4-3
EN 61000-4-4
B
RS
A
B
B
B
B
A
A
B
2. To meet the requirements of EN 61000-4-5
( 2 kV/ 4 kV ), external components are
needed. This can be done discretely (as
shown below), or with the addition of the
MACFM-02A. Contact the factory for more
information.
2 kV
4 kV
1 kV/ 2 kV
2 kV/ 4 kV
10V rms
10A/m
EFT, See Note 1
Surge, See Note 2
EN 61000-4-5
CS
PFM
EN 61000-4-6
EN 61000-4-8
EN 61000-4-11
Voltage Dips, Short, Interruptions
0% - 70%
Typical Connection: With Input Protection/Filtering Components
The diagram above illustrates a typical connection of the MPM-05SV series.
The input components are required to meet the more stringent EFT/Surge levels
of EN 61000-4 (see notes for EMC Characteristics table above). Some notes on
these components are:
1. It is recommended that an external fuse be used. The recommended fuse
size is a 1.0A/300 VAC slow blow.
2. An external MOV is recommended on the input to protect the unit in the event
of a surge. A recommended value is given in the table at right.
3. The output filtering capacitors (C1 & C2) will reduce the output noise below
specified levels. Recommended values are given in the table on page 2.
4. Input protection and filtering modules are available for a number of MPD
AC/DC power supplies. For use with the MPM-05MS product series, the
MACFM-02A filter module is recommended. A typical connection diagram
and board layout with this module is shown in the figures below. For pricing
or full technical information on the MACFM-02A (or any of our other modules)
please contact the factory.
5. Suggested component values are:
Component
Fuse
MPM-05MS-05 MPM-05MS-12
MPM-05SV-15
2.0A/250 VAC
S14K300
MPM-05MS-24
2.0A/250 VAC
S14K300
0.1 µF/275V
4.7 µH/2A
2.2 mH
2.0A/250 VAC
S14K300
0.1 µF/250V
4.7 µH/2A
2.2 mH
2.0A/250 VAC
S14K300
0.1 µF/250V
4.7 µH/2A
2.2 mH
MOV
0.1 µF/250V
4.7 µH/2A
2.2 mH
CX
LDM
LCM
CY1/CY2
NTC
1.0 nF/400 VAC 1.0 nF/400 VAC 1.0 nF/400 VAC 1.0 nF/400 VAC
5D-9
5D-9
5D-9
5D-9
Typical Board Layout: With External Filter Module
Typical Board Layout: With External Filter Module
1. The width of circuit traces should be a minimum of 3 mm.
2. The distance between traces should be a minimum of 6 mm.
3. The distance between a trace and ground should be a minimum of 6 mm.
MicroPower Direct
•
292 Page Street Ste D Stoughton, MA 02072
•
TEL: (781) 344-8226
•
FAX: (781) 344-8481
•
E-Mail: sales@micropowerdirect.com
www.micropowerdirect.com
Efficiency Vs Input Voltage: MPM-05MS-05
Efficiency Vs Ouput Load: MPM-05MS-05
Mechanical Dimensions
Efficiency Vs Input Voltage: MPM-05MS-24
Efficiency Vs Input Voltage: MPM-05MS-24
Pin Connections
Pin
Function
AC Neutral
AC Line
+VOUT
1
2
3
4
-VOUT
Notes:
• All dimensions are typical in inches (mm)
• Tolerance x.xx = 0.02 ( 0.5)
MicroPower Direct
We Power Your Success - For Less!
292 Page Street Ste D Stoughton, MA 02072
•
TEL: (781) 344-8226
•
FAX: (781) 344-8481
•
E-Mail: sales@micropowerdirect.com
相关型号: