MPM-40M [MPD]

40W Output Power;
MPM-40M
型号: MPM-40M
厂家: MICROPOWER DIRECT, LLC    MICROPOWER DIRECT, LLC
描述:

40W Output Power

文件: 总4页 (文件大小:1215K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
Series  
MPM-40M  
Medical Approved, 40W  
Single & Dual Output  
AC/DC Power Supplies  
Electrical Specifications  
Key Features:  
40W Output Power  
Specifications typical @ +25°C, 230 VAC input voltage & rated output current, unless otherwise noted. Specifications subject to change without notice.  
Input  
EN 60601-1 3RD Edition  
Universal 85-264 VAC Input  
EN 60950 Approved  
Reinforced Insulation  
Meets IEC Safety Class II  
Low Leakage Current  
Single and Dual Outputs  
Meets UL 508  
Parameter  
Conditions  
Min.  
85  
120  
47  
Typ.  
Max.  
264  
370  
440  
Units  
VAC  
VDC  
Hz  
Input Voltage Range  
Input Frequency  
115 VAC  
230 VAC  
30  
60  
Inrush Current, See Note 2  
A Pk  
W
No Load Power Consumption  
Output  
Parameter  
Output Voltage Accuracy  
Line Regulation  
0.3  
Conditions  
Min.  
Typ.  
2.0  
0.5  
1.0  
2.0  
1.5  
1.0  
25  
Max.  
Units  
%
%
VIN = Min to Max  
Single Output  
Load Regulation, IO = Min to Max  
Ripple & Noise (20 MHz), See Note 3  
Hold-Up Time  
%
%VP-P of VO  
mS  
Dual Output  
5.0 VDC Output  
All Other Outputs  
115 VAC, 60 Hz  
230 VAC, 50 Hz  
1.8  
1.3  
Chassis Mount Models  
& DIN Rail Mount  
Option Available!  
80  
Temperature Coefficient  
Over Voltage Protection  
Overshoot  
Overload Protection  
Over Temperature Protection  
Short Circuit Protection, See Note 7  
General  
0.02  
120  
%/°C  
% of VO  
%
% INOM  
°C  
See Note 4  
5
See Note 5 105  
See Note 6  
90  
Continuous (Autorecovery)  
Parameter  
Isolation Voltage (Reinforced)  
Leakage Current  
Isolation Resistance  
Switching Frequency  
Environmental  
Conditions Min. Typ.  
Max.  
Units  
VAC rms  
µA  
M  
kHz  
Input to Output 4,000  
500 VDC 1,000  
80  
130  
Parameter  
Radiated Emissions  
Conducted Emissions  
Standard  
EN 55014, EN 55024  
EN 55022  
Criterea  
Level  
Class B  
Class B  
A
A
A
A
A
A
A
A
B
8 kV Air  
4 kV Contact  
10 V/m  
2 kV  
1 kV  
10 Vrms  
30 A/M  
30% 10 mS  
95% 5,000 mS  
ESD  
EN 61000-4-2  
RS  
EFT  
Surge  
CS  
EN 61000-4-3  
EN 61000-4-4  
EN 61000-4-5  
EN 61000-4-6  
EN 61000-4-8  
EN 61000-4-11  
EN 61000-4-11  
PFM  
Voltage Dips  
Interruptions  
Environmental  
Parameter  
Operating Temperature Range  
Storage Temperature Range  
Cooling, See Note 8  
Humidity  
Physical  
Case Size  
Case Material  
Reliability Specifications  
Parameter  
MicroPower Direct  
292 Page Street  
Suite D  
Stoughton, MA 02072  
USA  
Conditions  
Min.  
Typ.  
+25  
Max.  
+80  
+95  
Units  
°C  
°C  
Ambient  
-40  
-40  
Free Air Convection (See Derating Curve)  
RH, Non-condensing 95  
%
T: (781) 344-8226  
F: (781) 344-8481  
E: sales@micropowerdirect.com  
W: www.micropowerdirect.com  
See Mechanical Diagrams (Pages 3 & 4)  
Non-Conductive Plastic Resin (UL94-V0)  
Conditions  
Min.  
Typ.  
Max.  
Units  
MTBF  
MIL HDBK 217F, 25°C, Gnd Benign  
200  
kHours  
Safety Standards  
UL 60950, EN 60950  
Safety Class  
IEC 60536 Class II  
www.micropowerdirect.com  
Model Selection Guide  
www.micropowerdirect.com  
Output  
Current (A)  
Input Current  
Capacitive  
Load  
F, Max)  
Model  
Number  
Efficiency  
(%, Typ)  
(mA Typ)  
Voltage  
(VDC)  
115 VAC  
230 VAC  
429  
Max.  
8.00  
3.33  
2.66  
1.66  
1.66  
1.33  
Min.  
0.00  
Other outputs may be available  
Contact the factory for details at:  
sales@micropowerdirect.com  
MPM-40MS-05xx  
MPM-40MS-12xx  
MPM-40MS-15xx  
MPM-40MS-24xx  
MPM-40MD-12xx  
MPM-40MD-15xx  
716  
689  
680  
687  
687  
680  
5.0  
12.0  
15.0  
24.0  
12.0  
15.0  
8,000  
3,900  
3,900  
680  
81  
84  
85  
84  
84  
85  
414  
0.00  
0.00  
0.00  
0.00  
0.00  
408  
413  
413  
1,500  
1,000  
408  
Notes:  
Model Number  
1. The specified maximum capacitive load is for each output.  
2. Inrush current is given for a cold start at 25°C.  
3. Ripple and noise is measured with a 0.1 µF/50V MLCC and a  
1 µF/50V aluminum electrolytic connected between the +VOUT  
and -VOUT pins for single output models or from each output to  
common for dual output models.  
4. Over voltage protection is provided by a zener diode clamp.  
5. Output overload protection is provided by a “hiccup mode” circuit  
with auto-recovery.A long-term overload could damage the unit.  
6. The unit will shutdown at 90°C. It will recover automatically at  
about 67 °C.  
7. Output short circuit protection is provided by a “hiccup mode”  
circuit. The unit recovers automatically when the fault condition  
is removed.  
8. Free air (natural) convection is about 20 LFM, it is not equal to  
“still” air.  
Derating Curve  
9. Operation at no-load will not damage these units. However, they  
may not meet all specifications.  
10. It is recommended that an external slow blow fuse also be used  
on the input for protection.  
Chassis Mount & DIN Rail  
All models of the MPM-40Mx series  
are available in an encapsulated  
package that can be mounted to a  
chassis or cabinet and connected with  
flying leads. To order the unit in this  
package, add a “CM” to the end of the  
product number.  
This package can also be adapted for  
use with a standard DIN rail by install-  
ing a simple adapter kit. As illustrated  
at right, this DIN rail kit consists of  
a base plate, a DIN rail bracket and  
the necessary hardware. For the  
MPM-40Mx series the DIN rail kit is:  
AC-DIN-02.  
292 Page Street Ste D Stoughton, MA 02072  
TEL: (781) 344-8226  
FAX: (781) 344-8481 E-Mail: sales@micropowerdirect.com  
Mechanical Dimensions  
www.micropowerdirect.com  
Pin Connections  
Pin  
Single  
AC Neutral  
AC Line  
+VOUT  
Dual  
AC Neutral  
AC Line  
+VOUT  
1
2
3
5
-VOUT  
Common  
-VOUT  
7
NC  
NC = No Connection  
Notes:  
All dimensions are typical in inches (mm)  
Tolerance x.xx = 0.02 ( 0.5)  
Mechanical Dimensions, Chassis Mount  
For the “Chassis Mount” package, add suffix “CM”  
to the model number (i.e. MPM-40MS-12CM)  
Pin Connections  
Pin  
Single  
AC Neutral  
AC Line  
+VOUT  
NC  
Dual  
AC Neutral  
AC Line  
+VOUT  
NC  
1
2
3
4
5
-VOUT  
NC  
Common  
NC  
6
7
NC  
-VOUT  
NC = No Connection  
Notes:  
All dimensions are typical in inches (mm)  
Tolerance x.xx = 0.02 ( 0.5)  
MicroPower Direct  
292 Page Street Ste D Stoughton, MA 02072 TEL: (781) 344-8226 FAX: (781) 344-8481 E-Mail: sales@micropowerdirect.com  
www.micropowerdirect.com  
DIN Rail Mounting  
All chassis mount MPM-40Mx models are easily  
adapted for use with a standard 35 mm DIN rail by  
installing a simple adapter kit.  
The DIN rail kit consists of a base plate, a DIN rail  
bracket and the necessary hardware. Mechanical dimen-  
sions for an MPM-40Mx assembly with the DIN rail  
option are shown below.  
For the MPM-40Mx series the DIN rail kit is: AC-DIN-02.  
Mechanical Dimensions, DIN Rail Mount  
Pin Connections  
Pin  
Single  
AC Neutral  
AC Line  
+VOUT  
NC  
Dual  
AC Neutral  
AC Line  
+VOUT  
NC  
1
2
3
4
5
-VOUT  
NC  
Common  
NC  
6
7
NC  
-VOUT  
NC = No Connection  
Notes:  
All dimensions are typical in inches (mm)  
Tolerance x.xx = 0.02 ( 0.5)  
MicroPower Direct  
We Power Your Success - For Less!  
292 Page Street Ste D Stoughton, MA 02072  
TEL: (781) 344-8226  
FAX: (781) 344-8481  
E-Mail: sales@micropowerdirect.com  

相关型号:

MPM-40MD-12xx

40W Output Power
MPD

MPM-40MD-15xx

40W Output Power
MPD

MPM-40MS-05xx

40W Output Power
MPD

MPM-40MS-12xx

40W Output Power
MPD

MPM-40MS-15xx

40W Output Power
MPD

MPM-40MS-24xx

40W Output Power
MPD

MPM-40S-03

Board Mount, 40W Single, Dual & Triple Out AC/DC Power Supplies
MPD

MPM-40S-05

Board Mount, 40W Single, Dual & Triple Out AC/DC Power Supplies
MPD

MPM-40S-09

Board Mount, 40W Single, Dual & Triple Out AC/DC Power Supplies
MPD

MPM-40S-12

Board Mount, 40W Single, Dual & Triple Out AC/DC Power Supplies
MPD

MPM-40S-15

Board Mount, 40W Single, Dual & Triple Out AC/DC Power Supplies
MPD

MPM-40S-24

Board Mount, 40W Single, Dual & Triple Out AC/DC Power Supplies
MPD