47LPQ-D0B-M2/Q [MSYSTEM]
Digital Panel Meters 47 Series;型号: | 47LPQ-D0B-M2/Q |
厂家: | M-SYSTEM |
描述: | Digital Panel Meters 47 Series |
文件: | 总8页 (文件大小:186K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
MODEL: 47LPQ
Digital Panel Meters 47 Series
[3] ALARM OUTPUT
0: None
1: N.O. relay contact, 4 points
2: SPDT relay contact, 2 points
PULSE INPUT TOTALIZER
(6-digit, LED display type)
Functions & Features
• 6-digit digital panel meter
• 1/8 DIN size
[4] DISPLAY COLOR
R: Red
• Process integration/totalization with scaling factor
adjustment
YR: Orange
G: Green
• Count stored at the loss of power supply
• Alarm tone optional
BG: Bluegreen
B: Blue
• IP66 front panel
W: White
• Separable terminal block
• Rear terminal cover for safety strapped to the meter
[5] POWER INPUT
AC Power
96 (3.78)
M2: 100 – 240 V AC (Operational voltage range 85 – 264 V, 50/60 Hz)
DC Power
R: 24 V DC
H
H
H
48
(1.89)
P
(Operational voltage range 24 V ±10 %, ripple 10 %p-p max.)
P: 110 V DC
L
L
L
Ini
A
lm
D/P
A
dj
In
Fnc
Min
Ma
x
M
ax
/M
in
A
la
rm
/
it/
Sh
ift
U
p
98.5 (3.88)
(Operational voltage range 85 – 150 V, ripple 10 %p-p max.)
mm (inch)
[6] OPTIONS (multiple selections)
Alarm Tone
MODEL: 47LPQ-[1][2][3][4]-[5][6]
blank: Without
/B: With (selectable only with alarm output)
Other Options
ORDERING INFORMATION
• Code number: 47LPQ-[1][2][3][4]-[5][6]
Specify a code from below for each of [1] through [6].
(e.g. 47LPQ-1A1G-M2/B/Q)
blank: none
/Q: Option other than the above (specify the specification)
• Specify the specification for option code /Q
(e.g. /C01/S01/SET)
SPECIFICATIONS OF OPTION: Q (multiple selections)
COATING (For the detail, refer to M-System's web site.)
Moving parts and indicators are not coated.
/C01: Silicone coating
[1] INPUT
1: Open collector or voltage pulse
/C02: Polyurethane coating
/C03: Rubber coating
2: Mechanical contact
TERMINAL SCREW MATERIAL
/S01: Stainless steel
[2] DC OUTPUT
0: Without
EX-FACTORY SETTING
/SET: Preset according to the Ordering Information Sheet
(No. ESU-9516)
Current
A: 4 – 20 mA DC (Load resistance 550 Ω max.)
D: 0 – 20 mA DC (Load resistance 550 Ω max.)
Voltage
GENERAL SPECIFICATIONS
Construction: Panel flush mounting
4: 0 – 10 V DC (Load resistance 10 kΩ min.)
5: 0 – 5 V DC (Load resistance 5000 Ω min.)
6: 1 – 5 V DC (Load resistance 5000 Ω min.)
4W: -10 – +10 V DC (Load resistance 10 kΩ min.)
Degree of protection: IP66; applicable to the front of the
panel meter mounted according to the specified panel
cutout
Connection: M3 separable screw terminal (torque 0.6 N·m)
Screw terminal: Nickel-plated steel (standard) or stainless
steel
47LPQ SPECIFICATIONS
ES-9516 Rev.14 Page 1/8
https://www.m-system.co.jp/
MODEL: 47LPQ
Housing material: Flame-resistant resin (gray)
Isolation: Input to DC output to HH output or H output to
L output or LL output to power
Setting: (Front button)
INPUT SPECIFICATIONS
Default measuring range: 0 – 999999
Sensor excitation: 12 V DC ±10 %, 30 mA
Current limit protection at approx. 60 mA
Open Collector & Voltage Pulse
• Scaled range
• Input type
Minimum pulse width: ≥ 5 μsec.
Maximum frequency: 100 Hz
• Alarm setpoint
• Hysteresis (deadband)
10kHz for Input type setting of 1/100 counting scale
Mechanical Contact
• Others
(Refer to the instruction manual for details)
Lockout setting: Prohibiting certain operations; protecting
settings
Minimum pulse width: ≥ 500 msec.
Maximum frequency: 1 Hz
■ Open Collector & Mechanical Contact
Detecting voltage/current: Approx. 8 V DC @ 1.6 mA
Detecting levels: ≤ 300 Ω / 0.6 V for ON;
≥ 10 kΩ / 4.5 V for OFF
DISPLAY
Display: 16 mm (.63) high, 6 digits, 7-segment LED
■ Voltage Pulse
Display range: -99999 to 999999
Voltage range: 0 – 5 through 26.4 V
±5 – ±26.4 V (equal amplitude at both poles)
Waveform: Square
Decimal point position: 10–1, 10–2, 10–3, 10–4, 10–5 or none
Zero indication: Higher-digit zeros are suppressed.
Overflow count modes
Input impedance: 10 kΩ min.
Reset: Reset and restart at 0 or 1
Low level: -26.4 – 0.6 V DC
(Scaling values at factory setting)
High level: 4.5 – 26.4 V DC
Hold: Hold at 100 %; display blinking
Input indication: ‘D/P’ turns on for 1 second after a valid
pulse train has been detected (applied to 1/100 scaled
signal if the input type is selected).
‘Adj’ turns on with control input.
■ Control Input: Reset count, Hold count or Invert count
(selectable function)
Detecting time: ≥ 200 msec.
Detecting levels: 4.5 – 26.4 V or short circuit across the
terminals 1 and 5 for ON; -26.4 – 0.6 V for OFF
Alarm status indication
LL indicator: Green LED turns on when the LL alarm is
tripped.
OUTPUT SPECIFICATIONS
■ DC Output
L indicator: Green LED turns on when the L alarm is
tripped.
• DC Current
H indicator: Red LED turns on when the H alarm is tripped.
HH indicator: Red LED turns on when the HH alarm is
tripped.
Operational range: -5 – +105 %
• DC Voltage
Operational range: -5 – +105 %
■ Alarm Output: Relay contact
Rated load: 250 V AC @ 3 A (cos ø = 1)
30 V DC @ 3 A (resistive load)
Maximum switching voltage: 250 V AC, 30 V DC
Maximum switching power: 750 VA, 90 W (resistive load)
Minimum load: 5 V DC @ 10 mA
P indicator: Amber LED turns on when none of the other
alarms is tripped.
Only ‘P’ turns on with no-alarm-output type. ‘LL’ or ‘HH’
does not turn on with dual-alarm-output type.
All setpoints can be independently set either for Hi or Lo
alarm trip.
Function indicators:
Mechanical life: ≥ 5 × 10 cycles (rate 180 cycles/min.)
6
Ini, Alm, D/P, Adj, Fnc, Min, Max
Display mode status and operation status, amber ON or
blink
INSTALLATION
Power consumption
•AC: Approx. 6.5VA
•DC: Approx. 3 W
Engineering unit indication: Sticker label attached
DC, AC, mV, V, kV, μA, mA, A, kA, mW, W,
kW, var, kvar, Mvar, VA, Hz, Ω, kΩ, MΩ,
cm, mm, m, m/sec, mm/min, cm/min, m/min,
Operating temperature: -10 to +55°C (14 to 131°F)
Operating humidity: 30 to 90 %RH (non-condensing)
Mounting: Panel flush mounting
m/h, m/s2, inch, ℓ, ℓ/s, ℓ/min, ℓ/h, m3, m3/sec,
m3
/min, m
3
/h, Nm /h, N·m, N/m , g, kg, kg/h,
3
2
Weight: 300 g (0.66 lb)
N, kN, Pa, kPa, MPa, t, t/h, ℃, °F, %RH, J,
kJ, MJ, rpm, sec, min, min-1, pH, %, ppm, kWh, MWh, Wh
47LPQ SPECIFICATIONS
ES-9516 Rev.14 Page 2/8
https://www.m-system.co.jp/
MODEL: 47LPQ
PERFORMANCE
DC output accuracy: ±0.1 %
Temp. coefficient: ±0.015 %/°C (±0.008 %/°F)
Output resolution: Max. 14 bits
Response time: ≤ 0.5 sec.
(alarm output: 0 – 100 % at 90 % setpoint)
≤ 0.5 sec. (DC output: 0 – 90 %)
Line voltage effect: ±0.1 % over voltage range
Insulation resistance: ≥ 100 MΩ with 500 V DC
Dielectric strength: 2000 V AC @ 1 minute (input to DC
output to HH output or H output to L output or LL output to
power to ground)
STANDARDS & APPROVALS
EU conformity:
EMC Directive
EMI EN 61000-6-4
EMS EN 61000-6-2
Low Voltage Directive
EN 61010-1
Measurement Category II (alarm output)
Installation Category II (power)
Pollution degree 2
Input or DC output to alarm output to power: Reinforced
insulation (300 V)
Input to DC output: Basic insulation (300 V)
RoHS Directive
Protection against access to the terminal blocks:
Finger protection (VDE 0660-514)
47LPQ SPECIFICATIONS
ES-9516 Rev.14 Page 3/8
https://www.m-system.co.jp/
MODEL: 47LPQ
EXTERNAL VIEW
(1) Main Display
HH
H
P
(2) Alarm Indicators
L
LL
Ini
Alm D/P
Adj
Fnc Min Max
(3) Function Indicators
Max/Min
Alarm/
Init/
Shift
Up
(8) Up Button
(7) Shift Button
(6) Init/ Button
(4) Max/Min Button
(5) Alarm/ Button
• COMPONENT IDENTIFICATION
No. COMPONENT
(1) Main display
FUNCTION
Indicates present values, setting values and status of the unit.
Indicate alarm status of the input signal.
(2) Alarm indicators
(3) Function indicators
(4) Max/Min button
Indicate the status in each setting mode.
Used to switch the main display to show the present values, maximum values or minimum values etc.
(5)
Used to con!rm alarm setting value and to move on to the alarm and other setting modes; or to shift through
setting items in each setting mode.
Alarm/ button
(6)
Used to move on to the scaling and other setting modes; or to shift through setting items in each setting
mode.
Init/ button
(7) Shift button
(8) Up button
Used to move on to the setting standby status of each setting mode and shift through display digits in each
setting mode.
Used to reset the counter and select setting values in each setting mode.
Note: Refer to the operating manual for details on each function.
47LPQ SPECIFICATIONS
ES-9516 Rev.14 Page 4/8
https://www.m-system.co.jp/
MODEL: 47LPQ
EXTERNAL DIMENSIONS & TERMINAL ASSIGNMENTS unit: mm (inch)
■
TOP VIEW
95 (3.74)
■
FRONT VIEW
■ SIDE VIEW
103 (4.06)
96 (3.78)
98.5 (3.88)
86 (3.39)
12.5 (.49)
2 (.08)
■
REAR VIEW
• Alarm Output
• No Alarm Output
10–M3 SCREW
TERMINAL
20–M3 SCREW
TERMINAL
6.6 (.26)
2–M3 SCREW
4–M3 SCREW
1
2
3
4
5
6
7
8
9
10
1
2
3
4
5
6
7
8
9
10
20
11
12
13
14
15
16
17
18
19
91.5 (3.60)
47LPQ SPECIFICATIONS
ES-9516 Rev.14 Page 5/8
https://www.m-system.co.jp/
MODEL: 47LPQ
MOUNTING REQUIREMENTS unit: mm
92 +– 00.8
min. 120
Panel thickness: 1.6 to 8.0 mm
47LPQ SPECIFICATIONS
ES-9516 Rev.14 Page 6/8
https://www.m-system.co.jp/
MODEL: 47LPQ
SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM
■ ALARM OUTPUT CODE 0: No alarm output
Isolation
Sensor
Excitation
Output
Driver
EXC+
OC JP
INP+
1
2
3
4
5
+
7
8
9
SHORT- CIRCUIT*1
SENSOR EXCITATION
DC OUTPUT
POWER
–
Input
Circuit
1/100
Scaler
Digital
Computation
INPUT
U(+)
COM
*
–
Control
Input
2
CONTROL INPUT
10 V(–)
DI+
Display /
Setting
6
■ ALARM OUTPUT CODE 1: N.O. contact, 4 points
Isolation
Sensor
Excitation
Output
Driver
EXC+
OC JP
INP+
1
2
3
4
5
+
–
7
8
SHORT- CIRCUIT*1
SENSOR EXCITATION
DC OUTPUT
Input
Circuit
1/100
Scaler
INPUT
Ry
Ry
Digital
Computation
COM
*
–
15 HHa
16 Ha
Control
Input
2
CONTROL INPUT
HH OUTPUT
H OUTPUT
DI+
6
17 Hc, HHc
Ry
Ry
11
12
13
14
18 La
L OUTPUT
19 LLa
LL OUTPUT
20 Lc, LLc
U(+)
10 V(–)
9
POWER
■ ALARM OUTPUT CODE 2: SPDT contact, 2 points
Isolation
Sensor
Excitation
Output
Driver
1
2
3
4
5
EXC+
OC JP
INP+
+
–
7
8
SHORT- CIRCUIT*1
SENSOR EXCITATION
DC OUTPUT
Input
Circuit
1/100
Scaler
INPUT
Ry
Digital
Computation
COM
*
–
15 Ha(NO)
16 Hb(NC)
Control
Input
2
CONTROL INPUT
DI+
H OUTPUT
L OUTPUT
6
17 Hc(COM
18 La(NO)
19 Lb(NC)
)
Ry
11
12
13
14
20 Lc(COM
U(+)
10 V(–)
)
9
POWER
*1. Close across the terminals for open collector or mechanical contact. Refer to the figure below.
*2. For detailed connection of CONTROL INPUT, refer to the operating manual (EM-9516-B).
■ Input Connection Example
• Open Collector or Mechanical Contact
■ Relay Protection
• AC Powered
• DC Powered
Inductive
Load (Coil)
Inductive
Load (Coil)
1
Short
8V
Varistor or Spark
Quenching Circuit
Diode, Varistor or
CR Circuit
LOAD
LOAD
2
5kΩ
3
U
Note 1: The section enclosed by broken line is only with DC output option.
Note 2: Chattering input signals cannot be measured accurately.
Add a capacitor (10000 pF) or similar devices to the input terminals to eliminate
the input signals’ instability.
4
5
COM
47LPQ SPECIFICATIONS
ES-9516 Rev.14 Page 7/8
https://www.m-system.co.jp/
MODEL: 47LPQ
Specifications are subject to change without notice.
47LPQ SPECIFICATIONS
ES-9516 Rev.14 Page 8/8
https://www.m-system.co.jp/
相关型号: