F2SN-1H6 [MSYSTEM]
Super-mini Signal Conditioners F2 Series;型号: | F2SN-1H6 |
厂家: | M-SYSTEM |
描述: | Super-mini Signal Conditioners F2 Series |
文件: | 总3页 (文件大小:103K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
MODEL: F2SN
Super-mini Signal Conditioners F2 Series
[4] OPTIONS
blank: none
/Q: With options (specify the specification)
INPUT LOOP POWERED ISOLATOR
Functions & Features
• Loop-powered design eliminates output loop power supply
• Two isolators housed in one enclosure
• 350 Ω output drive with 4 – 20 mA
SPECIFICATIONS OF OPTION: Q
COATING (For the detail, refer to M-System's web site.)
/C01: Silicone coating
/C02: Polyurethane coating
Typical Applications
• Isolation between control room and field instrumentation,
between telemetering system and input device
• Eliminates ground problems in existing systems
thanks to easiness of application without requiring
additional power wiring
GENERAL SPECIFICATIONS
Construction: Stand-alone; terminal access at the front
Connection: Euro type connector terminal
(applicable wire size: 0.2 to 2.5 mm , stripped length 7 mm)
2
Housing material: Flame-resistant resin (black)
Isolation: Input to output; between channels
Zero adjustment (front)
22.5 (.89)
Voltage output: -3 % to +3 %
Current output: -0.5 % to +0.5%
Span adjustment (front)
75
(2.95)
Voltage output: 95 to 105 %
Current output: 98.5 to 101.5 %
107
(4.21)
INPUT & OUTPUT
mm (inch)
■ Input 4 – 20 mA DC / Output 1 – 5 V DC
Equivalent input impedance: Approx. 250 Ω with 20 mA
input
MODEL: F2SN–[1][2][3][4]
Operational range: 3 – 22 mA DC
(Accuracy is assured within 4 – 22 mA)
Load resistance: ≥ 50 kΩ
ORDERING INFORMATION
• Code number: F2SN-[1][2][3][4]
Specify a code from below for each of [1] through [4].
(e.g. F2SN-2AA/Q)
■ Input 10 – 50 mA DC / Output 1 – 5 V DC
Equivalent input impedance: Approx. 100 Ω with 50 mA
input
• Specify the specification for option code /Q
(e.g. /C01)
Operational range: 7 – 55 mA DC
(Accuracy is assured within 8 – 55 mA)
Load resistance: ≥ 50 kΩ
[1] NO. OF CHANNELS
1: 1 channel
2: 2 channels
■ Input 4 – 20 mA DC / Output 4 – 20 mA DC
Equivalent input impedance: 230 Ω plus load resistance with
20 mA input
[2] INPUT
Current
Operational range: 3 – 22 mA DC
(Accuracy is assured within 4 – 22 mA)
Load resistance: 50 – 350 Ω (min. 50 Ω required for
adequate operation)
A: 4 – 20 mA DC
H: 10 – 50 mA DC
[3] OUTPUT
Current
■ Input 10 – 50 mA DC / Output 4 – 20 mA DC
Equivalent input impedance: 90 Ω + [load resistance×0.16]
with 50 mA input
A: 4 – 20 mA DC
Voltage
Operational range: 7 – 55 mA DC
6: 1 – 5 V DC
(Accuracy is assured within 8 – 55 mA)
Load resistance: 50 – 600 Ω (min. 50 Ω required for
F2SN SPECIFICATIONS
ES-5419 Rev.8 Page 1/3
http://www.m-system.co.jp/
MODEL: F2SN
adequate operation)
Load effect (current output)
•INPUT 4 – 20 mA DC / OUTPUT 1 – 5 V DC
4 – 20 mA input: 0.015 %/Ω (50 – 150 Ω)
0.003 %/Ω (150 – 350 Ω)
5.0
approx. 5.0
10 – 50 mA input: 0.015 %/Ω (50 – 100 Ω)
0.003 %/Ω (100 – 600 Ω)
4.0
approx.
3.7
(The unit is calibrated with 250 Ω load at the factory.)
Insulation resistance: ≥ 100 MΩ with 500 V DC
Dielectric strength:
3.0
4
12
20
500 V AC @1 minute (input to output)
2000 V AC @1 minute (between channels)
2000 V AC @1 minute (input or output to ground)
INPUT CURRENT [mA]
•INPUT 10 – 50 mA DC / OUTPUT 1 – 5 V DC
5.0
approx. 5.0
STANDARDS & APPROVALS
EU conformity:
4.0
approx.
3.7
EMC Directive
EMI EN 61000-6-4
3.0
10
30
50
EMS EN 61000-6-2
INPUT CURRENT [mA]
RoHS Directive
•INPUT 4 – 20mA DC / OUTPUT 4 – 20 mA DC
EN 50581
12.0
Approval:
approx. 11.6
(load 350Ω)
UL/C-UL general safety requirements
(UL 3111-1, CAN/CSA-C22.2 No.1010-1)
10.0
8.0
approx. 9.6
(load 250Ω)
6.0
approx. 5.0
approx. 4.6
approx. 3.8
approx. 5.6
(load 50Ω)
3.0
4
12
20
INPUT CURRENT [mA]
•INPUT 10 – 50 mA DC / OUTPUT 4 – 20 mA DC
10.0
approx. 9.3
(load 600Ω)
8.0
approx. 6.5
(load 250Ω)
6.0
approx. 4.7
approx. 4.1
approx. 3.8
3.0
approx. 4.9
(load 50Ω)
10
30
50
INPUT CURRENT [mA]
INSTALLATION
Operating temperature: -5 to +55°C (23 to 131°F)
Operating humidity: 30 to 90 %RH (non-condensing)
Mounting: DIN rail
Weight: 150 g (0.33 lb)
PERFORMANCE in percentage of span
Accuracy: ±0.1 %
Temp. coefficient
Voltage output: ±0.015 %/°C (±0.008 %/°F)
Current output: ±0.02 %/°C (±0.01 %/°F)
Response time
Voltage output: ≤ 0.5 sec. (0 – 90 %)
Current output: Approx. 15 msec. (0 – 90 %)
F2SN SPECIFICATIONS
ES-5419 Rev.8 Page 2/3
http://www.m-system.co.jp/
MODEL: F2SN
EXTERNAL DIMENSIONS & TERMINAL ASSIGNMENTS unit: mm (inch)
DIN RAIL
35mm wide
1
2 3 4
5
6 7 8
22.5 (.89)
107 (4.21)
[1.5 (.06)]
• When mounting, no extra space is needed between units.
SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM
■ VOLTAGE OUTPUT
Z
S
Buffer
Amplifier
+
–
Chopper
Inverter
+
–
1
2
5
6
Ch 1 INPUT
Ch 2 INPUT
Ch 1 OUTPUT
Ch 2 OUTPUT
Rectifier
+
–
+
–
3
4
7
8
Channel 2
■ CURRENT OUTPUT
S
Isolation
Inverter
Load Effect
Compensation
+
–
+
–
1
2
5
6
Ch 1 OUTPUT
Z
Ch 1 INPUT
+
–
+
–
3
4
7
8
Channel 2
Ch 2. INPUT
Ch 2 OUTPUT
Specifications are subject to change without notice.
F2SN SPECIFICATIONS
ES-5419 Rev.8 Page 3/3
http://www.m-system.co.jp/
相关型号:
©2020 ICPDF网 联系我们和版权申明