JARP2-1A-P/Q 概述
Plug-in Signal Conditioners M-UNIT
JARP2-1A-P/Q 数据手册
通过下载JARP2-1A-P/Q数据手册来全面了解它。这个PDF文档包含了所有必要的细节,如产品概述、功能特性、引脚定义、引脚排列图等信息。
PDF下载MODEL: JARP2
B: 2 – 10 mA DC (Input resistance 500 Ω)
Plug-in Signal Conditioners M-UNIT
C: 1 – 5 mA DC (Input resistance 1000 Ω)
D: 0 – 20 mA DC (Input resistance 50 Ω)
E: 0 – 16 mA DC (Input resistance 62.5 Ω)
F: 0 – 10 mA DC (Input resistance 100 Ω)
G: 0 – 1 mA DC (Input resistance 1000 Ω)
H: 10 – 50 mA DC (Input resistance 100 Ω)
Z: Specify current (See INPUT SPECIFICATIONS)
Voltage
DC/2-PHASE PULSE CONVERTER
(field-programmable)
Functions & Features
• Providing pulse rate outputs in 90-degree phase angle in
proportion to DC input signal
• Parameters are field-programmable via hand-held
programmer PU-2x
1: 0 – 10 mV DC (Input resistance 10 kΩ min.)
2: 0 – 100 mV DC (Input resistance 100 kΩ min.)
3: 0 – 1 V DC (Input resistance 1 MΩ min.)
4: 0 – 10 V DC (Input resistance 1 MΩ min.)
5: 0 – 5 V DC (Input resistance 1 MΩ min.)
6: 1 – 5 V DC (Input resistance 1 MΩ min.)
4W: -10 – +10 V DC (Input resistance 1 MΩ min.)
0: Specify voltage (See INPUT SPECIFICATIONS)
• Isolation up to 2000 V AC
• High-density mounting
Typical Applications
• Remote transmission of rotary encoder signal via 4 – 20
mA cable in combination with the JRP2
50
(1.97)
[2] OUTPUT
A: Open collector
M: 5 V pulse
80
(3.15)
N: 12 V pulse
P: 24 V pulse
136
(5.35)
J: RS-422 line driver pulse
mm (inch)
[3] POWER INPUT
AC Power
MODEL: JARP2–[1][2]–[3][4]
M2: 100 – 240 V AC
DC Power
ORDERING INFORMATION
• Code number: JARP2–[1][2]–[3][4]
S: 12 V DC
R: 24 V DC
Specify a code from below for each of [1] through [4].
(e.g. JARP2-6A-S/Q)
P: 110 V DC
• Special input range (For codes Z & 0)
• Use Ordering Information Sheet (No. ESU-1570). Default
setting (table below) will be used if not otherwise specified.
• Specify the specification for option code /Q
(e.g. /C01/S01)
[4] OPTIONS
blank: none
/Q: With options (specify the specification)
SPECIFICATIONS OF OPTION: Q (multiple selections)
COATING (For the detail, refer to M-System's web site.)
/C01: Silicone coating
ITEM
DEFAULT
Output zero frequency
Output span frequency
-1 kHz
1 kHz
/C02: Polyurethane coating
Alarm setpoint
Alarm hysteresis
Alarm mode
100.00%
1.00%
No alarm
/C03: Rubber coating
TERMINAL SCREW MATERIAL
/S01: Stainless steel
Power ON-delay time
3.0 sec.
RELATED PRODUCTS
• Programming Unit (model: PU-2x)
[1] INPUT
Current
• PC configurator software (model: JXCON)
Downloadable at M-System’s web site.
A dedicated cable is required to connect the module to the
A: 4 – 20 mA DC (Input resistance 250 Ω)
A1: 4 – 20 mA DC (Input resistance 50 Ω)
JARP2 SPECIFICATIONS
ES-1570 Rev.6 Page 1/4
http://www.m-system.co.jp/
MODEL: JARP2
PC. Please refer to the internet software download site or
the users manual for the PC configurator for applicable
cable types.
RNG6 100 mHz: -200 – +200 mHz
Min. span: 10 % of available frequency range
Duty ratio: 50 %
■ Open Collector: 50 V DC @50 mA (resistive load)
Saturation voltage: 0.6 V DC
■ Voltage Pulse
GENERAL SPECIFICATIONS
Construction: Plug-in
H level: Rating 5, 12 or 24 V ±10 %
L level: ≤ 0.5 V
Connection: M3.5 screw terminals
Screw terminal: Chromated steel (standard) or stainless
steel
Load resistance: ≥ 1 kΩ (5 V), ≥ 2.4 kΩ (12 V), ≥ 4.8 kΩ (24
V)
Housing material: Flame-resistant resin (black)
Isolation: Input to alarm output to output to power
Zero adjustment: -5 to +5 % (front)
Span adjustment: 95 to 105 % (front)
Alarm output: Turns on when the input goes above (high) or
below (low) setpoint
■ RS-422 Line Driver Pulse
Transmitter: Conforms to RS-422
■ Relay Contact:
100 V AC @ 1 A (cos ø = 1)
120 V AC @ 1 A (cos ø = 1)
240 V AC @ 0.5 A (cos ø = 1)
30 V DC @ 1 A (resistive load)
Maximum switching voltage: 380 V AC or 125 V DC
Maximum switching power: 120 VA or 30 W
Minimum load: 5 V DC @ 10 mA
Alarm setpoint: -5 – +105 %
Alarm deadband (hysteresis): 0 – 20 %
Output monitor LED: Red LED turns on according to the
output (phase A and B).
Power ON delay time: 2.0 to 1000.0 sec.
Software programming: Programming Unit (model: PU-2x);
(Refer to the users manual of JXCON for the adjustments
configurable with JXCON)
Mechanical life: 5 x 10 cycles (300 cycle/minute)
7
INSTALLATION
• Output frequency (zero and span)
• Alarm
Power input
•AC: Operational voltage range 85 – 264 V, 47 - 66 Hz;
Approx. 4.1 VA at 100 V
• Power ON delay time
• Others
Approx. 5.9 VA at 200 V
Approx. 7.7 VA at 264 V
•DC: Operational voltage range: rating ±10 %, or 85 – 150 V
for 110 V rating (ripple 10 % p-p max.)
Approx. 2.8 W (120 mA at 24 V)
INPUT SPECIFICATIONS
■ DC Current:
Shunt resistor attached to the input terminals (0.5 W)
Specify input resistance value for code Z.
■ DC Voltage: -300 – +300 V DC
Minimum span: 10 mV
Operating temperature: -5 to +55°C (23 to 131°F)
Operating humidity: 30 to 90 %RH (non-condensing)
Mounting: Surface or DIN rail
Weight: 280 g (0.62 lb)
Offset: Max. 1.5 times span
Input resistance
Span 10 – 100 mV : ≥ 10 kΩ
Span 0.1 – 1 V : ≥ 100 kΩ
PERFORMANCE in percentage of span
Accuracy
Span ≥ 1 V : ≥ 1 MΩ
Output 10 kHz range: ±0.3 %
Output ≤ 1 kHz range: ±0.1 %
Alarm setpoint accuracy: ±0.1 %
OUTPUT SPECIFICATIONS
Phase angle: ±90°
Temp. coefficient: ±0.015 %/°C (±0.008 %/°F)
Response time: 0.3 sec. + (one cycle at 100 % output)
Time required from a step input (0 – 100 %) to the first
pulse output.
Forward (positive output): Phase A = Phase B + 90°
Reverse (negative output): Phase A = Phase B – 90°
Frequency range: 100 mHz to 10 kHz (RNG1 to RNG6)
[ Range name : Available frequency ]
RNG1 10 kHz: -20 – +20 kHz
Line voltage effect: ±0.1 % over voltage range
Insulation resistance: ≥ 100 MΩ with 500 V DC
Dielectric strength: 2000 V AC @ 1 minute (input to alarm
output to output to power to ground)
RNG2 1 kHz: -2 – +2 kHz
RNG3 100 Hz: -200 – +200 Hz
RNG4 10 Hz: -20 – +20 Hz
RNG5 1 Hz: -2 – +2 Hz
JARP2 SPECIFICATIONS
ES-1570 Rev.6 Page 2/4
http://www.m-system.co.jp/
MODEL: JARP2
EXTERNAL VIEW
DO NOT change the Output Voltage Adjuster.
Modular Jack
Span Adj.
Zero Adj.
Monitor LED (no function)
Output Monitor LED A
Output Monitor LED B
Output Voltage Adj.*
*only for voltage pulse output
OPERATION
fwd. 100 Hz
fwd. 100 Hz
100%
ACCEL.
DECEL.
INPUT
0%
ACCEL.
STOP
rev. -30 Hz
4mA
0%
rev. -100 Hz
-10V
0%
FORWARD
REVERSE
INPUT
20mA
100%
INPUT
+10V
100%
φ
φ
A
B
OUTPUT
OUTPUT
EXTERNAL DIMENSIONS unit: mm (inch)
CLAMP
(top & bottom)
7.8 (.31)
DIN RAIL
35mm wide
8
9
7
6
5
4
2–
4.5 (.18) dia.
MTG HOLE
25 (.98) deep
3
2
10 11
1
11
–M3.5
SCREW
50 (1.97)
107 (4.21)
136 (5.35)
40 (1.57)
50 (1.97)
[3.3 (.13)]
• When mounting, no extra space is needed between units.
JARP2 SPECIFICATIONS
ES-1570 Rev.6 Page 3/4
http://www.m-system.co.jp/
MODEL: JARP2
TERMINAL ASSIGNMENTS unit: mm (inch)
INPUT RESISTOR
(model: REM)
8
7
6
5
4
9
3
2
10 11
1
Input shunt resistor attached
for current input.
SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM
f
A
OUT.
MON.
Z
S
Isolation
Volt. / Freq.
Converter
f A
Out. Driver
+
–
+
5
6
3
4
Computation
1
2
*
INPUT
R
f
f
A OUTPUT
B OUTPUT
POWER
–
f
B
+
–
+
10
11
7
Out. Driver
Ry
ALARM OUTPUT
PU-2x
f
B
–
OUT.
MON.
MODULAR JACK
U(+)
V(–)
SERIAL TRANSMISSION LINE
Base Socket
8
*Input shunt resistor attached for current input.
Output Connection Examples
■ Open Collector
■ Voltage Pulse
+
–
+
–
+
–
+
–
1
2
1
2
■ RS-422 Line Driver Pulse
+
–
+
–
+
–
1
2
Specifications are subject to change without notice.
JARP2 SPECIFICATIONS
ES-1570 Rev.6 Page 4/4
http://www.m-system.co.jp/
JARP2-1A-P/Q 相关器件
型号 | 制造商 | 描述 | 价格 | 文档 |
JARP2-1A-R | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
JARP2-1A-R/Q | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
JARP2-1A-S | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
JARP2-1A-S/Q | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
JARP2-1J-M2 | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
JARP2-1J-M2/Q | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
JARP2-1J-P | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
JARP2-1J-P/Q | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
JARP2-1J-R | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
JARP2-1J-R/Q | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 |
JARP2-1A-P/Q 相关文章
- 2024-09-20
- 5
- 2024-09-20
- 9
- 2024-09-20
- 8
- 2024-09-20
- 6