LCS-5A-G/E2 概述
Plug-in Signal Conditioners M-UNIT
LCS-5A-G/E2 数据手册
通过下载LCS-5A-G/E2数据手册来全面了解它。这个PDF文档包含了所有必要的细节,如产品概述、功能特性、引脚定义、引脚排列图等信息。
PDF下载MODEL: LCS
6: 10 mV/V
7: 20 mV/V
Plug-in Signal Conditioners M-UNIT
0: Specify (strain gauge and excitation)
STRAIN GAUGE TRANSMITTER
(isolated)
[2] OUTPUT
Functions & Features
Current
• Providing a DC output signal proportional to a bridge type
strain gauge utilized in load cells, pressure transducers
• Supplying required excitation voltage
• Drives bridges 80 Ω or above
A: 4 – 20 mA DC (Load resistance 750 Ω max.)
B: 2 – 10 mA DC (Load resistance 1500 Ω max.)
C: 1 – 5 mA DC (Load resistance 3000 Ω max.)
D: 0 – 20 mA DC (Load resistance 750 Ω max.)
E: 0 – 16 mA DC (Load resistance 900 Ω max.)
F: 0 – 10 mA DC (Load resistance 1500 Ω max.)
G: 0 – 1 mA DC (Load resistance 15 kΩ max.)
Z: Specify current (See OUTPUT SPECIFICATIONS)
Voltage
• Excitation adjustable from 2 V to 10 V
• Wide-range adjustment: 0 – 80 % for zero, 20 – 100 % for
span
• Fast response type available
• LCD meter (engineering unit display selectable)
• Simple loop test output (0 % and 100 %)
• High-density mounting
1: 0 – 10 mV DC (Load resistance 10 kΩ min.)
2: 0 – 100 mV DC (Load resistance 100 kΩ min.)
3: 0 – 1 V DC (Load resistance 100 Ω min.)
4: 0 – 10 V DC (Load resistance 1000 Ω min.)
5: 0 – 5 V DC (Load resistance 500 Ω min.)
6: 1 – 5 V DC (Load resistance 500 Ω min.)
4W: -10 – +10 V DC (Load resistance 2000 Ω min.)
5W: -5 – +5 V DC (Load resistance 1000 Ω min.)
0: Specify voltage (See OUTPUT SPECIFICATIONS)
Typical Applications
• Weighing system for tanks, hoppers, silos
• Weighing system using cranes
• Float level meter utilizing strain gauges
50
(1.97)
80
(3.15)
[3] POWER INPUT
AC Power
B: 100 V AC
C: 110 V AC
136
(5.35)
D: 115 V AC
mm (inch)
F: 120 V AC
G: 200 V AC
H: 220 V AC
MODEL: LCS–[1][2]–[3][4]
J: 240 V AC
DC Power
ORDERING INFORMATION
• Code number: LCS-[1][2]-[3][4]
Specify a code from below for each of [1] through [4].
(e.g. LCS-2A-B/E2/Q)
R: 24V DC (Not selectable with option /E or /E2)
[4] OPTIONS (multiple selections)
Input Signal Indicator
• Special output range (For codes Z & 0)
• Specify the specification for option code /Q
(e.g. /C01/S01)
blank: Without
/E: With (0.0 - 100.0 % display)
/E2: With (in engineering unit with backlight and the
simple loop test output)
[1] INPUT STRAIN GAUGE
1: 1 mV/V
Response Time (0 – 90 %)
blank: Standard (≤ 0.5 sec.)
12: 1.25 mV/V
15: 1.5 mV/V
2: 2 mV/V
/K: Fast response (Approx. 25 msec.)
(Not selectable with Option /E2)
Other Options
3: 3 mV/V
blank: none
4: 4 mV/V
/Q: Option other than the above (specify the specification)
5: 5 mV/V
LCS SPECIFICATIONS
ES-1360 Rev.3 Page 1/3
http://www.m-system.co.jp/
MODEL: LCS
Minimum span: 5 mV
SPECIFICATIONS OF OPTION: Q (multiple selections)
COATING (For the detail, refer to M-System's web site.)
/C01: Silicone coating
Offset: Max. 1.5 times span
Load resistance: Output drive 10 mA max.; 5 mA for
negative voltage output; at ≥ 0.5 V
/C02: Polyurethane coating
/C03: Rubber coating
TERMINAL SCREW MATERIAL
/S01: Stainless steel
INSTALLATION
Power input
•AC: Operational voltage range: rating ±10 %, 50/60 ±2 Hz,
approx. 3 VA
GENERAL SPECIFICATIONS
Construction: Plug-in
(approx. 4 VA with Option /E2)
•DC: Operational voltage range: 24 V ±10 %,
approx. 150 mA, ripple 10 %p-p max.
Operating temperature: -5 to +60°C (23 to 140°F)
Operating humidity: 30 to 90 %RH (non-condensing)
Mounting: Surface or DIN rail
Connection: M3.5 screw terminals
Screw terminal: Chromated steel (standard) or stainless
steel
Housing material: Flame-resistant resin (black)
Isolation: Input to output to power
Weight: 430 g (0.95 lb)
Overrange output: Approx. -10 to +120 % at 1 – 5 V
Excitation adjustment: 2 – 10 V (front)
Zero adjustments (tare): 0 – 80 % (front)
(Excitation voltage: factory default)
Span adjustment: 100 – 20 % (front)
(Excitation voltage: factory default)
Simple loop test output: 0 % and 100 % signal simulated by
selecting the front switch positions. (Only for option code
/E2)
PERFORMANCE in percentage of span
Accuracy: ±0.1 % (input ≥ 3 mV)
Display accuracy: ± (0.1 % of FS + 1 digit)
(input ≥ 3 mV)
Simple loop test output setting accuracy: ±0.5 %
Temp. coefficient: ±0.015 %/°C (±0.008 %/°F)
Line voltage effect: ±0.1 % over voltage range
Insulation resistance: ≥ 100 MΩ with 500 V DC
Dielectric strength: 2000 V AC @1 minute for AC power
1000 V AC @1 minute for DC power
■ DISPLAY (Input indicator)
• Option code: /E
LCD digital display: 0.0 - 100.0 % (min. digit 0.1 %)
(No scaling)
(input or output to power)
• Option code: /E2
1000 V AC @1 minute (input to output)
2000 V AC @1 minute
LCD digital display: Engineering unit
Display scaling: -10000 – +10000
(input or output or power to ground)
Decimal position: 10-1 - 10-4 or no decimal point
Engineering unit: %, μV, mV, V, mA, A, °C, °F, Ω, DEG K,
mHz, Hz, kHz, VAC, AAC, mg, g, kg, t, rpm or rps selectable
Back light: Green at normal, red at loop test output enable
Factory setting: scaling 0.00 - 100.00, unit: %
INPUT SPECIFICATIONS
■ Input: Bridge voltage from load cells
• Strain Gauge
Rated output from strain gauge: 1 – 20 mV/V;
Input to the transmitter must be over 3 mV.
• Excitation: 2 – 10 V adjustable (5 V standard)
Maximum current: 35 mA at 10 V, 65 mA at ≤ 7.5 V
OUTPUT SPECIFICATIONS
■ DC Current: 0 – 20 mA DC
Minimum span: 1 mA
Offset: Max. 1.5 times span
Load resistance: Output drive 15 V max.
■ DC Voltage: -10 – +12 V DC
LCS SPECIFICATIONS
ES-1360 Rev.3 Page 2/3
http://www.m-system.co.jp/
MODEL: LCS
EXTERNAL VIEW
■ Option /E
■ Option /E2
LCD Meter
LCD Meter
NEXT Key
Excitation Adj.
Zero Adj.
ENTER Key
Excitation Adj.
Zero Adj.
Loop Test Selector
Span Adj.
Span Adj.
Top
: 0% fixed
Middle : 100% fixed
Bottom : Cancel
(normal operating mode)
EXTERNAL DIMENSIONS & TERMINAL ASSIGNMENTS unit: mm (inch)
CLAMP
(top & bottom)
7.8 (.31)
DIN RAIL
35mm wide
8
7
6
5
4
2–
4.5 (.18) dia.
MTG HOLE
25 (.98) deep
9
3
2
10 11
1
11
–M3.5
SCREW
50 (1.97)
107 (4.21)
136 (5.35)
40 (1.57)
50 (1.97)
[3.3 (.13)]
• When mounting, no extra space is needed between units.
SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM
*
Z
S
Isolation
STRAIN
GAUGE
Low Drift
Amplifier
Output
Driver
+
–
+
–
+
–
10
11
3
1
2
OUTPUT
POWER
**
LOOP TEST
SELECTOR
Low Drift
DC Volt.
Stabilizer
EXC.
U(+)
V(–)
7
8
9
6
Base Socket
*
Option /E, E2
** Option /E2
Specifications are subject to change without notice.
LCS SPECIFICATIONS
ES-1360 Rev.3 Page 3/3
http://www.m-system.co.jp/
LCS-5A-G/E2 相关器件
型号 | 制造商 | 描述 | 价格 | 文档 |
LCS-5A-G/K | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
LCS-5A-G/Q | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
LCS-5A-H | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
LCS-5A-H/E | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
LCS-5A-H/E2 | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
LCS-5A-H/K | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
LCS-5A-H/Q | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
LCS-5A-J | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
LCS-5A-J/E | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
LCS-5A-J/E2 | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 |
LCS-5A-G/E2 相关文章
- 2024-10-31
- 6
- 2024-10-31
- 6
- 2024-10-31
- 7
- 2024-10-31
- 8