TCS-00-B 概述
Plug-in Signal Conditioners M-UNIT
TCS-00-B 数据手册
通过下载TCS-00-B数据手册来全面了解它。这个PDF文档包含了所有必要的细节,如产品概述、功能特性、引脚定义、引脚排列图等信息。
PDF下载MODEL: TCS
6: B (RH) (Usable range 0 to 1820°C, 32 to 3308°F)
Plug-in Signal Conditioners M-UNIT
7: R (Usable range -50 to +1760°C, -58 to +3200°F)
8: S (Usable range -50 to +1760°C, -58 to +3200°F)
N: N (Usable range -270 to +1300°C, -454 to +2372°F)
0: Specify
THERMOCOUPLE TRANSMITTER
(CE)
Functions & Features
• Accepting direct input from a thermocouple and providing
a standard process signal
[2] OUTPUT
Current
• 7-segment linearization
A: 4 – 20 mA DC (Load resistance 750 Ω max.)
B: 2 – 10 mA DC (Load resistance 1500 Ω max.)
C: 1 – 5 mA DC (Load resistance 3000 Ω max.)
D: 0 – 20 mA DC (Load resistance 750 Ω max.)
E: 0 – 16 mA DC (Load resistance 900 Ω max.)
F: 0 – 10 mA DC (Load resistance 1500 Ω max.)
G: 0 – 1 mA DC (Load resistance 15 kΩ max.)
Z: Specify current (See OUTPUT SPECIFICATIONS)
Voltage
• Burnout protection
• High-accuracy cold junction compensation
• Signal isolation
• Fast response type available
• High-density mounting
Typical Applications
• High-accuracy cold junction compensation benefits narrow
span measurements
1: 0 – 10 mV DC (Load resistance 10 kΩ min.)
2: 0 – 100 mV DC (Load resistance 100 kΩ min.)
3: 0 – 1 V DC (Load resistance 3000 Ω min.)
4: 0 – 10 V DC (Load resistance 10 kΩ min.)
5: 0 – 5 V DC (Load resistance 5000 Ω min.)
6: 1 – 5 V DC (Load resistance 5000 Ω min.)
0: Specify voltage (See OUTPUT SPECIFICATIONS)
• 0.1 μA burnout sensing enables long distance
transmission with minimum offset drifts
• Electric furnace (isolation)
• No burnout type can connect to a single T/C in parallel
with a recorder
50
(1.97)
[3] POWER INPUT
AC Power
80
(3.15)
B: 100 V AC
C: 110 V AC
D: 115 V AC
F: 120 V AC
G: 200 V AC
H: 220 V AC
J: 240 V AC
127
(5)
mm (inch)
MODEL: TCS–[1][2]–[3][4]
DC Power
S: 12 V DC
ORDERING INFORMATION
• Code number: TCS-[1][2]-[3][4]
Specify a code from below for each of [1] through [4].
(e.g. TCS-2A-B/K/BL/CE/Q)
R: 24 V DC
V: 48 V DC
[4] OPTIONS (multiple selections)
Response Time (0 – 90 %)
blank: Standard (≤ 0.5 sec.)
/K: Fast Response (Approx. 25 msec.)
Burnout
• Temperature range (e.g. 0 – 800°C)
• Special output range (For codes Z & 0)
• Specify the specification for option code /Q
(e.g. /C01/S01)
blank: Upscale burnout
[1] INPUT THERMOCOUPLE
/BL: Downscale burnout
/BN: No burnout
1: (PR) (Usable Range 0 to 1760°C, 32 to 3200°F)
2: K (CA) (Usable range -270 to +1370°C, -454 to +2498°F)
3: E (CRC) (Usable range -270 to +1000°C, -454 to +1832°F)
4: J (IC) (Usable range -210 to +1200°C, -346 to +2192°F)
5: T (CC) (Usable range -270 to +400°C, -454 to +752°F)
Standards & Approvals (must be specified)
/CE: CE marking
Other Options
blank: none
TCS SPECIFICATIONS
ES-1331-CE Rev.8 Page 1/4
http://www.m-system.co.jp/
MODEL: TCS
/Q: Option other than the above (specify the specification)
N: min. span 200°F
Note: The described accuracy may be partially not satisfied
when the temperature ranges below 0°C. Consult factory.
SPECIFICATIONS OF OPTION: Q (multiple selections)
COATING (For the detail, refer to M-System's web site.)
/C01: Silicone coating
/C02: Polyurethane coating
OUTPUT SPECIFICATIONS
■ DC Current: 0 – 20 mA DC
/C03: Rubber coating
TERMINAL SCREW MATERIAL
/S01: Stainless steel
Minimum span: 1 mA
Offset: Max. 1.5 times span
Load resistance: Output drive 15 V max.
■ DC Voltage: -10 – +12 V DC
Minimum span: 5 mV
GENERAL SPECIFICATIONS
Construction: Plug-in
Offset: Max. 1.5 times span
Connection: M3.5 screw terminals
Load resistance: Output drive 1 mA max. at ≥ 3 V
Screw terminal: Chromated steel (standard) or stainless
steel
Housing material: Flame-resistant resin (black)
Isolation: Input to output to power
INSTALLATION
Power input
Overrange output: Approx. -10 to +120 % at 1 – 5 V
Zero adjustment: -5 to +5 % (front)
• AC: Operational voltage range: rating ±10 %,
50/60 ±2 Hz, approx. 2 VA
Span adjustment: 95 to 105 % (front)
At burnout: Downscale ≤ -10 %, Upscale ≥ 110 %
Linearization: Standard
•DC: Operational voltage range: rating ±10 %
ripple 10 %p-p max., approx. 2.6 W (110 mA at 24 V)
Operating temperature: -5 to +60°C (23 to 140°F)
Operating humidity: 30 to 90 %RH (non-condensing)
Mounting: Surface or DIN rail
Cold junction compensation: CJC sensor attached to the
input terminals (B thermocouple is without CJC as standard)
Weight: 350 g (0.77 lb)
INPUT SPECIFICATIONS
Minimum span: 3 mV
PERFORMANCE in percentage of span
Accuracy: ±0.3 % (at over 400°C or 750°F for R, S
and PR; over 770°C or 1420°F for B)
Cold junction compensation error
Offset: Max. 1.5 times span
Input resistance: 30 kΩ min.
Burnout sensing: 0.1 μA
(at 20°C ±10°C or 68°F ±18°F)
Minimum span (in °C)
(PR): min. span 370°C
K (CA): min. span 75°C
E (CRC): min. span 50°C
J (IC): min. span 60°C
T (CC): min. span 75°C
B (RH): min. span 780°C
R: min. span 360°C
K, E, J, T & N: ±0.5°C or ±0.9°F
S, R & PR: ±1°C or ±1.8°F
Temp. coefficient: ±0.015 %/°C (±0.008 %/°F)
(at over 400°C or 750°F for R, S and PR; over 770°C or
1420°F for B)
Burnout response: ≤ 10 sec.
Line voltage effect: ±0.1 % over voltage range
Insulation resistance: ≥ 100 MΩ with 500 V DC
Dielectric strength:
S: min. span 380°C
N: min. span 110°C
1350 V AC @1 minute (input to output)
2300 V AC @1 minute (input or output to power to ground)
Minimum span (in °F)
(PR): min. span 670°F
K (CA): min. span 140°F
E (CRC): min. span 90°F
J (IC): min. span 110°F
T (CC): min. span 140°F
B (RH): min. span 1410°F
R: min. span 650°F
STANDARDS & APPROVALS
EU conformity:
EMC Directive
EMI EN 61000-6-4
EMS EN 61000-6-2
Low Voltage Directive
EN 61010-1
S: min. span 690°F
TCS SPECIFICATIONS
ES-1331-CE Rev.8 Page 2/4
http://www.m-system.co.jp/
MODEL: TCS
Installation Category II
Pollution Degree 2
Input or output to power: Reinforced insulation (300 V)
Input to output: Basic insulation (300 V)
RoHS Directive
EN 50581
EXTERNAL DIMENSIONS unit: mm (inch)
CLAMP
(top & bottom)
7.8 (.31)
DIN RAIL
35mm wide
6
5
4
3
2–4.5 (.18) dia.
MTG HOLE
15 (.59) deep
7
8
1
2
8–M3.5
SCREW
50 (1.97)
107 (4.21)
127 (5)
40 (1.57)
50 (1.97)
[3.3 (.13)]
• When mounting, no extra space is needed between units.
TERMINAL ASSIGNMENTS unit: mm (inch)
CJC SENSOR
(model: CJM)
green
red
3
6 5
4
1
7
8
2
TCS SPECIFICATIONS
ES-1331-CE Rev.8 Page 3/4
http://www.m-system.co.jp/
MODEL: TCS
SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM
comp. leadwire
Z
S
Isolation
T / C
+
Low Drift
Amplifier
Output
Driver
+
+
–
3
Linearizer
1
2
7
8
OUTPUT
–
metal leg
–
U(+)
POWER
V(–)
4
5
6
CJC
SENSOR
red
*
CJC Sensing
Volt. Generator
model: CJM
green
Base Socket
*Deleted with B thermocouple input.
Specifications are subject to change without notice.
TCS SPECIFICATIONS
ES-1331-CE Rev.8 Page 4/4
http://www.m-system.co.jp/
TCS-00-B 相关器件
型号 | 制造商 | 描述 | 价格 | 文档 |
TCS-00-B/BL | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
TCS-00-B/BN | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
TCS-00-B/CE | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
TCS-00-B/E | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
TCS-00-B/K | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
TCS-00-B/Q | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
TCS-00-C | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
TCS-00-C/BL | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
TCS-00-C/BN | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 | |
TCS-00-C/CE | MSYSTEM | Plug-in Signal Conditioners M-UNIT | 获取价格 |
TCS-00-B 相关文章
- 2024-10-31
- 6
- 2024-10-31
- 6
- 2024-10-31
- 7
- 2024-10-31
- 8