USB-6008 [NI]

8 analog inputs at 12 or 14 bits,up to 48 kS/s; 8个模拟输入,12位或14位, 48 kS / s的
USB-6008
型号: USB-6008
厂家: National Instruments    National Instruments
描述:

8 analog inputs at 12 or 14 bits,up to 48 kS/s
8个模拟输入,12位或14位, 48 kS / s的

文件: 总5页 (文件大小:197K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
Low-Cost, Bus-Powered Multifunction DAQ for USB –  
12- or 14-Bit, up to 48 kS/s, 8 Analog Inputs  
NI USB-6008, NI USB-6009  
• 8 analog inputs at 12 or 14 bits,  
up to 48 kS/s  
Recommended Software  
• LabVIEW  
• 2 analog outputs at 12 bits,  
software-timed  
• 12 TTL/CMOS digital I/O lines  
• 32-bit, 5 MHz counter  
• Digital triggering  
• LabVIEW SignalExpress  
• LabWindows/CVI  
• Measurement Studio  
Other Compatible Software  
• C#, Visual Basic .NET  
• ANSI C/C++  
• Bus-powered  
• 1-year warranty  
Measurement Services  
Software (included)  
• NI-DAQmx driver software  
• Measurement & Automation  
Explorer configuration utility  
• LabVIEW SignalExpress LE  
Operating Systems  
• Windows Vista (32- and 64-bit)/XP/2000  
• Mac OS X1  
®1  
• Linux  
• Windows Mobile1  
• Windows CE1  
1You need to download NI-DAQmx  
Base for these operating systems.  
Input  
Resolution  
(bits)  
Max  
Sampling Rate  
(kS/s)  
Input  
Range  
(V)  
Output  
Resolution  
(bits)  
Output  
Rate  
(Hz)  
Output  
Range  
(V)  
Digital  
I/O  
Lines  
Analog  
Inputs1  
Analog  
Outputs  
32-Bit  
Counter Trigger  
Product  
USB-6009  
USB-6008  
Bus  
USB  
USB  
8 SE/4 DI  
8 SE/4 DI  
14  
12  
48  
10  
1 to 20  
1 to 20  
2
2
12  
12  
150  
150  
0 to 5  
0 to 5  
12  
12  
1
1
Digital  
Digital  
1SE = single ended, DI = differential 2Software-timed  
driver software for application development in NI LabVIEW, LabVIEW  
SignalExpress, LabWindows/CVI, and Measurement Studio software.  
To obtain the latest version of NI-DAQmx, visit  
Overview and Applications  
With recent bandwidth improvements and new innovations from  
National Instruments, USB has evolved into a core bus of choice for  
measurement applications. The NI USB-6008 and USB-6009 are low-  
cost entry points to NI flagship data acquisition (DAQ) devices. With  
plug-and-play USB connectivity, these modules are simple enough for  
quick measurements but versatile enough for more complex  
measurement applications.  
ni.com/support/daq/versions.  
NI measurement services software speeds up your development with  
features including:  
• A guide to create fast and accurate measurements with no  
programming using the DAQ Assistant.  
• Automatic code generation to create your application in LabVIEW.  
• LabWindows/CVI; LabVIEW SignalExpress; and C#, Visual Studio .NET,  
ANSI C/C++, or Visual Basic using Measurement Studio.  
• Multithreaded streaming technology for 1,000 times  
performance improvements.  
The USB-6008 and USB-6009 are ideal for a number of applications  
where low cost, small form factor, and simplicity are essential.  
Examples include:  
• Data logging – quick and easy environmental or voltage data logging  
• Academic lab use – student ownership of DAQ hardware for  
completely interactive lab-based courses (Academic pricing available.  
Visit ni.com/academic for details.)  
• Automatic timing, triggering, and synchronization routing  
to make advanced applications easy.  
• More than 3,000 free software downloads available at  
ni.com/zone to jump-start your project.  
• OEM applications as I/O for embedded systems  
• Software configuration of all digital I/O features without  
hardware switches/jumpers.  
Recommended Software  
National Instruments measurement services software, built around  
NI-DAQmx driver software, includes intuitive application programming  
interfaces, configuration tools, I/O assistants, and other tools  
designed to reduce system setup, configuration, and development time.  
National Instruments recommends using the latest version of NI-DAQmx  
• Single programming interface for analog input, analog output,  
digital I/O, and counters on hundreds of multifunction DAQ hardware  
devices. M Series devices are compatible with the following versions  
(or later) of NI application software – LabVIEW, LabWindows/CVI, or  
Measurement Studio versions 7.x; and LabVIEW SignalExpress 2.x.  
Low-Cost, Bus-Powered Multifunction DAQ for USB – 12- or 14-Bit, up to 48 kS/s, 8 Analog Inputs  
Every M Series data acquisition device also includes a copy of  
Information for Student Ownership  
To supplement simulation, measurement, and automation theory courses  
LabVIEW SignalExpress LE data-logging software, so you can quickly  
acquire, analyze, and present data without programming. The NI-DAQmx  
with practical experiments, NI has developed the USB-6008 and USB-6009  
Base driver software is provided for use with Linux, Mac OS X,  
student kits, which include the LabVIEW Student Edition and a ready-to-run  
Windows Mobile, and Windows CE operating systems.  
data logger application. These kits are exclusively for students, giving them  
Recommended Accessories  
a powerful, low-cost, hands-on learning tool. Visit ni.com/academic for  
more details.  
The USB-6008 and USB-6009 have removable screw terminals for easy  
signal connectivity. For extra flexibility when handling multiple wiring  
configurations, NI offers the USB-600x Connectivity Kit, which includes  
two extra sets of screw terminals, extra labels, and a screwdriver.  
In addition, the USB-600x Prototyping Kit provides space for adding  
more circuitry to the inputs of the USB-6008 or USB-6009.  
Information for OEM Customers  
For information on special configurations and pricing, call (800) 813 3693  
(U.S. only) or visit ni.com/oem. Go to the Ordering Information section  
for part numbers.  
NI USB DAQ for OEMs  
Shorten your time to market by integrating world-class National  
Instruments OEM measurement products into your embedded system  
design. Board-only versions of NI USB DAQ devices are available for  
OEM applications, with competitive quantity pricing and available  
software customization. The NI OEM Elite Program offers free 30-day  
trial kits for qualified customers. Visit ni.com/oem for more information.  
Ordering Information  
NI USB-60081 ........................................................................779051-01  
NI USB-60091 ........................................................................779026-01  
NI USB-6008 OEM ................................................................193132-02  
NI USB-6009 OEM ................................................................193132-01  
NI USB-6008 Student Kit1,2 ..................................................779320-22  
NI USB-6009 Student Kit1,2 ..................................................779321-22  
NI USB-600x Connectivity Kit ..............................................779371-01  
NI USB-600x Prototyping Kit ................................................779511-01  
1 Includes NI-DAQmx software, LabVIEW SignalExpress LE, and a USB cable.  
2 Includes LabVIEW Student Edition.  
BUY NOW!  
For complete product specifications, pricing, and accessory  
information, call 800 813 3693 (U.S. only) or go to ni.com/usb.  
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)  
2
Low-Cost, Bus-Powered Multifunction DAQ for USB – 12- or 14-Bit, up to 48 kS/s, 8 Analog Inputs  
Output range ....................................... 0 to +5 V  
Specifications  
Output impedance............................... 50  
Output current drive............................ 5 mA  
Typical at 25 °C unless otherwise noted.  
Analog Input  
Absolute accuracy, single-ended  
Power-on state.................................... 0 V  
Slew rate............................................. 1 V/µs  
Short-circuit current............................ 50 mA  
Range  
Typical at 25 ˚C (mV)  
Maximum (0 to 55 ˚C) (mV)  
10  
14.7  
138  
Digital I/O  
Number of channels............................ 12 total  
Absolute accuracy at full scale, differential1  
8 (P0.<0..7>)  
Range  
Typical at 25 ˚C (mV)  
Maximum (0 to 55 ˚C) (mV)  
4 (P1.<0..3>)  
Direction control ................................. Each channel individually  
programmable as input or output  
20  
10  
5
4
2.5  
2
14.7  
7.73  
4.28  
3.59  
2.56  
2.21  
1.70  
1.53  
138  
84.8  
58.4  
53.1  
45.1  
42.5  
38.9  
37.5  
Output driver type  
USB-6008........................................ Open-drain  
USB-6009........................................ Each channel individually  
programmable as push-pull or  
1.25  
1
open-drain  
Compatibility....................................... CMOS, TTL, LVTTL  
Internal pull-up resistor ...................... 4.7 kto +5 V  
Power-on state.................................... Input (high impedance)  
Absolute maximum voltage range...... -0.5 to +5.8 V  
Number of channels............................ 8 single-ended/4 differential  
Type of ADC ........................................ Successive approximation  
ADC resolution (bits)  
Module  
USB-6008  
USB-6009  
Differential  
Single-Ended  
Digital logic levels  
12  
14  
11  
13  
Level  
Min  
-0.3  
2.0  
2.0  
2.0  
Max  
0.8  
5.8  
50  
0.8  
3.5  
5.0  
Units  
V
V
µA  
V
V
Input low voltage  
Input high voltage  
Input leakage current  
Output low voltage (I = 8.5 mA)  
Output high voltage (push-pull, I = -8.5 mA)  
Output high voltage (open-drain, I = -0.6 mA, nominal)  
Output high voltage (open-drain, I = -8.5 mA,  
with external pull-up resistor)  
Maximum sampling rate (system dependent)  
Module  
Maximum Sampling Rate (kS/s)  
USB-6008  
USB-6009  
10  
48  
V
2.0  
V
Input range, single-ended................... 10 V  
Input range, differential...................... 20, 10, 5, 4, 2.5, 2,  
1.25, 1 V  
Maximum working voltage................. 10 V  
Overvoltage protection ....................... 35 V  
FIFO buffer size ................................... 512 B  
Timing resolution ................................ 41.67 ns (24 MHz timebase)  
Timing accuracy .................................. 100 ppm of actual sample rate  
Input impedance ................................. 144 kΩ  
Trigger source...................................... Software or external digital trigger  
System noise....................................... 5 m Vrms ( 10 V range)  
Counter  
Number of counters............................  
Resolution ........................................... 32 bits  
Counter measurements....................... Edge counting (falling edge)  
Pull-up resistor.................................... 4.7 kto 5 V  
Maximum input frequency.................. 5 MHz  
Minimum high pulse width................. 100 ns  
Minimum low pulse width.................. 100 ns  
Input high voltage............................... 2.0 V  
1
Input low voltage................................ 0.8 V  
Analog Output  
Absolute accuracy (no load) ............... 7 mV typical, 36.4 mV maximum  
Power available at I/O connector  
+5 V output (200 mA maximum)......... +5 V typical  
+4.85 V minimum  
+2.5 V output (1 mA maximum).......... +2.5 V typical  
+2.5 V output accuracy ....................... 0.25% max  
Voltage reference temperature drift... 50 ppm/°C max  
at full scale  
2
Number of channels............................  
Type of DAC ........................................ Successive approximation  
DAC resolution.................................... 12 bits  
Maximum update rate ........................ 150 Hz, software-timed  
1Input voltages may not exceed the working voltage range.  
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)  
3
Low-Cost, Bus-Powered Multifunction DAQ for USB – 12- or 14-Bit, up to 48 kS/s, 8 Analog Inputs  
Physical Characteristics  
Safety and Compliance  
If you need to clean the module, wipe it with a dry towel.  
Dimensions (without connectors)....... 6.35 by 8.51 by 2.31 cm  
(2.50 by 3.35 by 0.91 in.)  
Dimensions (with connectors) ............ 8.18 by 8.51 by 2.31 cm  
(3.22 by 3.35 by 0.91 in.)  
Safety  
This product is designed to meet the requirements of the following  
standards of safety for electrical equipment for measurement, control,  
and laboratory use:  
Weight (without connectors) .............. 59 g (2.1 oz)  
Weight (with connectors) ................... 84 g (3 oz)  
I/O connectors..................................... USB series B receptacle  
(2) 16-position (screw-terminal)  
• IEC 61010-1, EN 61010-1  
• UL 61010-1, CSA 61010-1  
Note: For UL and other safety certifications, refer to the product label  
or visit ni.com/certification, search by model number or product line,  
and click the appropriate link in the Certification column.  
plug headers  
Screw-terminal wiring ........................ 16 to 28 AWG  
Screw-terminal torque........................ 0.22 to 0.25 N•m  
(2.0 to 2.2 lb•in.)  
Electromagnetic Compatibility  
This product is designed to meet the requirements of the following  
standards of EMC for electrical equipment for measurement, control,  
and laboratory use:  
Power Requirement  
USB (4.10 to 5.25 VDC)....................... 80 mA typical  
500 mA maximum  
USB suspend....................................... 300 µA typical  
500 µA maximum  
• EN 61326 EMC requirements; Minimum Immunity  
• EN 55011 Emissions; Group 1, Class A  
• CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A  
Note: For EMC compliance, operate this device according to  
product documentation.  
Environmental  
The USB-6008 and USB-6009 are intended for indoor use only.  
Operating environment  
Ambient temperature range........... 0 to 55 °C (tested in accordance  
with IEC-60068-2-1  
CE Compliance  
This product meets the essential requirements of applicable European  
Directives, as amended for CE marking, as follows:  
and IEC-60068-2-2)  
Relative humidity range ................. 10 to 90%, noncondensing  
(tested in accordance  
• 2006/95/EC; Low-Voltage Directive (safety)  
• 2004/108/EC; Electromagnetic Compatibility Directive (EMC)  
with IEC-60068-2-56)  
Note: Refer to the Declaration of Conformity (DoC) for this product for  
any additional regulatory compliance information. To obtain the DoC for  
this product, visit ni.com/certification, search by model number or  
product line, and click the appropriate link in the Certification column.  
Storage environment  
Ambient temperature range........... -40 to 85 °C (tested in  
accordance with IEC-60068-2-1  
and IEC-60068-2-2)  
Waste Electrical and Electronic Equipment (WEEE)  
Relative humidity range ................. 5 to 90%, noncondensing  
(tested in accordance  
with IEC-60068-2-56)  
Maximum altitude............................... 2,000 m  
(at 25 °C ambient temperature)  
EU Customers: At the end of their life cycle, all products must be  
sent to a WEEE recycling center. For more information about WEEE  
recycling centers and National Instruments WEEE initiatives, visit  
ni.com/environment/weee.htm.  
Pollution degree..................................  
2
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)  
4
NI Services and Support  
NI has the services and support to meet  
Local Sales and Technical Support  
your needs around the globe and through  
the application life cycle – from planning  
and development through deployment  
and ongoing maintenance. We offer  
services and service levels to meet  
customer requirements in research,  
design, validation, and manufacturing.  
Visit ni.com/services.  
In offices worldwide, our staff is local to the country, giving you access  
to engineers who speak your language. NI delivers industry-leading  
technical support through online knowledge bases, our applications  
engineers, and access to 14,000 measurement and automation  
professionals within NI Developer Exchange forums. Find immediate  
answers to your questions at ni.com/support.  
SERVICE  
NEEDS  
We also offer service programs that provide automatic upgrades to  
your application development environment and higher levels of technical  
support. Visit ni.com/ssp.  
Training and Certification  
NI training is the fastest, most certain route to productivity with our  
products. NI training can shorten your learning curve, save development  
time, and reduce maintenance costs over the application life cycle. We  
schedule instructor-led courses in cities worldwide, or we can hold a  
course at your facility. We also offer a professional certification program  
Hardware Services  
NI Factory Installation Services  
NI Factory Installation Services (FIS) is the fastest and easiest way to  
use your PXI or PXI/SCXI combination systems right out of the box.  
that identifies individuals who have high levels of skill and knowledge on Trained NI technicians install the software and hardware and configure  
using NI products. Visit ni.com/training.  
the system to your specifications. NI extends the standard warranty by  
one year on hardware components (controllers, chassis, modules)  
purchased with FIS. To use FIS, simply configure your system online with  
Professional Services  
Our Professional Services Team is comprised of NI applications engineers, ni.com/pxiadvisor.  
NI Consulting Services, and a worldwide National Instruments Alliance  
Calibration Services  
Partner program of more than 600 independent consultants and  
integrators. Services range from  
start-up assistance to turnkey  
NI recognizes the need to maintain properly calibrated devices for  
high-accuracy measurements. We provide manual calibration  
procedures, services to recalibrate your products, and automated  
calibration software specifically designed for use by metrology  
laboratories. Visit ni.com/calibration.  
system integration.  
Visit ni.com/alliance.  
OEM Support  
We offer design-in consulting and product integration assistance if you  
want to use our products for OEM applications. For information about  
special pricing and services for OEM customers, visit ni.com/oem.  
Repair and Extended Warranty  
NI provides complete repair services for our products. Express repair  
and advance replacement services are also available. We offer  
extended warranties to help you meet project life-cycle requirements.  
Visit ni.com/services.  
ni.com • (800) 813 3693  
*351378C-01*  
351378C-01  
2008-9621-301-101-D  
National Instruments • info@ni.com  
©2008 National Instruments. All rights reserved. CVI, LabVIEW, Measurement Studio, National Instruments, National Instruments Alliance Partner, NI, ni.com, SCXI, and SignalExpress  
are trademarks of National Instruments. The mark LabWindows is used under a license from Microsoft Corporation. Windows is a registered trademark of Microsoft Corporation in the  
®
United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries. Other product and company names listed are trademarks or trade  
names of their respective companies. A National Instruments Alliance Partner is a business entity independent from NI and has no agency, partnership, or joint-venture relationship with NI.  

相关型号:

USB-6211

This document provides information about the dimensions, mounting options, connectors
NI

USB-6212

This document provides information about the dimensions, mounting options, connectors
NI

USB-6216

This document provides information about the dimensions, mounting options, connectors
NI

USB-6218

This document provides information about the dimensions, mounting options, connectors
NI

USB-621X

This document provides information about the dimensions, mounting options, connectors
NI

USB-6221

This document provides dimensions, pinouts, and information about the connectors, switch, LEDs
NI

USB-6225

This document provides dimensions, pinouts, and information about the connectors, switch, LEDs
NI

USB-6229

This document provides dimensions, pinouts, and information about the connectors, switch, LEDs
NI

USB-622X

This document provides dimensions, pinouts, and information about the connectors, switch, LEDs
NI

USB-6251

This document provides dimensions, pinouts, and information about the connectors, switch, LEDs
NI

USB-6255

This document provides dimensions, pinouts, and information about the connectors, switch, LEDs
NI

USB-6259

This document provides dimensions, pinouts, and information about the connectors, switch, LEDs
NI