OBN01046 [POWERBOX]
150-200 WATTS SINGLE & MULTIPLE OUTPUT AC/DC MEDICAL; 150-200沃茨单个与多个输出AC / DC医疗型号: | OBN01046 |
厂家: | Powerbox |
描述: | 150-200 WATTS SINGLE & MULTIPLE OUTPUT AC/DC MEDICAL |
文件: | 总2页 (文件大小:1025K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
Medline 200 Series
1 5 0 - 2 0 0 W AT T S S I N G L E & M U LT I P L E O U T P U T A C / D C M E D I C A L
Features
• Low safety ground leakage current
• Wide input range 90 to 260Vac
• Input surge current protection
• Overvoltage protection
• Overcurrent protection
• Power fail detection (PFD) signal
• Optional cover and fan assembly
• 100% burn-in at full rated load
• Remote inhibit - TTL high disables output
• Short curcuit protection
PFD signal
TTL logic high for normal operation and TTL logic low
Specifications
upon loss of input power.This signal appears at least
1ms prior to +5V output dropping 5% below its
nominal value.This signal also provides a minimum
delay of 100ms after +5V is within regulation
INPUT
Voltage range
Frequency
90-260VAC
47-63Hz.
Efficiency
70% min on all models
20ms min at 110VAC
0.5% maꢀ at full load
Input Current
3.2A (rms) at 115VAC
1.6A (rms) at 230VAC
Hold up time
Leakage current
90µA maꢀ at 115VAC, 60Hz
150A maꢀ at 230VAC, 50Hz
Line regulation
ENVIRONMENTAL
Inrush Current
Isolation
20A @ 115V or 40A at 230V at 25°Cand cold start
Input – Output: 4000VAC
Input – Ground: 1500VAC
Output – Ground: 500VAC
Operating Temperature
Storage temperature
Humidity
0°C to 70°C
–40°C to +85°C
5% to 95% non-condensing
OUTPUT
Derating
Derate from 100% at +50°C linearly to 50% at
+70°C.
Power
200W
Voltage
See table
See table
2% p-p maꢀ
Cooling
25CFM forced air, provided on “C” version, to be
provided for “B” version by user
Current
GENERAL
Ripple & Noise
Overvoltage protection
Switching frequency
MTBF
88-112kHz.
Provided on output #1 only, set at 112-132% of it
nominal output voltage
350,000 hours at full load 25°C ambient ,
calculated per MIL-HDBK-217F
Overcurrent Protection
Temperature coefficient
Transient response
The output protected to short circuit conditions
0.04%°Call outputs
STANDARDS AND APPROVALS
Maꢀ eꢀcursion of 4% or better on all models,
recovering to 1% of final value within 500us after
a 25% step load change
Safety standards
C-Tick
IEC60601-1 TUV EN60601-1, EN60601-1
AS/NZS CISPR11 Group 1, Class A
EMC standards
EN55011 Class B, FCC Class B,VCII, EN61000-3-2, -3,
EN60001-4-2, -3, -4, -5, -6, -8, -11
Technical Illustration
MECHANICAL
Dimensions
Weight
See drawing
0.82kg for “B” version
0.96kg for version “C” version
Notes:
1. Dimensions shown in inches (mm)
2. Tolerance 0.02 (0.5) maximum
3. Input connector P1 is Dinkle DT-35-B01W-09 screws are M3, nickle plated
4. Connector P3 mates with Molex housing 22-01-1043 and Molex 40445 series
crimp terminal
5. Connector P4/P5 mates with Molex housing 22-01-1023 and Molex 40445
series crimp terminal
6. P4 is for DC fan rated 24V/0.2A (models PM201/27), or 5V/0.38A (models
PM201-40-3), and 12V/0.2A (other models, Pin1 +V and Pin2 COM
7. Weight:
0.82 Kgs (1.83 Lbs) approx for “B” version
0.96 Kgs (2.14 Lbs) approx for “C” version
345
Your dependable power partner – www.powerbox.com.au
Medline 200 Series
1 5 0 - 2 0 0 W AT T S S I N G L E & M U LT I P L E O U T P U T A C / D C M E D I C A L
Selection Table
MAX.
OUTPUT
POWER
175W
150W
200W
200W
200W
200W
200W
200W
200W
200W
200W
200W
200W
200W
200W
200W
200W
200W
200W
200W
175W
MODEL
OUTPUT #1
OUTPUT #2
OUTPUT #3
OUTPUT #4
NUMBER
OBN01016
OBN01013
OBN01037
OBN01046
OBN01055
OBN01070
OBN01204
OBN01207
OBN01210
OBN012A5
OBN01310
OBN01313
OBN01316
OBN01315
OBN01319
OBN01404
OBN01403
OBN01407
OBN01410
OBN01420
OBN01413
VNOM IMIN IMAX TOL VNOM IMIN IMAX TOL VNOM IMIN IMAX TOL VNOM IMIN IMAX TOL
5.1V
3.3V
12V
15V
24V
30V
48V
0A
0A
0A
0A
0A
0A
0A
35A
46A
2%
3%
16.7A 2%
13.4A 2%
8.4A
6.7A
4.2A
30A
30A
30A
8A
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
2%
+5.1V 3.0A
+5.1V 3.0A
+5.1V 3.0A
+12V
+15V
+24V
+24V
+12V
+12V
+15V
+15V
+12V
+12V
+15V
+12V
+12V
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
8A
6A
4A
4A
8A
8A
6A
6A
8A
8A
6A
8A
8A
8A
4%
4%
4%
4%
4%
4%
4%
4%
4%
4%
4%
4%
4%
4%
+12V
1.0A
+5.1V 3.0A
+5.1V 3.0A
+5.1V 3.0A
+5.1V 3.0A
+5.1V 3.0A
+5.1V 3.0A
+5.1V 3.0A
+5.1V 3.0A
+5.1V 3.0A
+3.3V 3.0A
30A
30A
30A
30A
30A
30A
30A
30A
30A
30A
-5V
0A
0A
0A
0A
0A
0A
0A
0A
0A
0A
6A
4A
4A
4A
4A
4A
4A
4A
4A
4A
4%
4%
4%
4%
4%
4%
4%
4%
4%
4%
-12V
-15V
-12V
-24V
-12V
-15V
-12V
-15V
-12V
5V
0A
0A
0A
0A
0A
6A
4A
4A
4A
4A
4%
4%
4%
4%
4%
24V
12V
15V
12V
3% +5.1V
NOTES: (1) Add suffix “B” for U-bracket format and “C” for enclosed format.
(2) Peak current is 12A on +12V, 9A on +15V and 6A on +24V.
(3) Output #4 is floating. It can be connected externally for positive or negative output.
(4) 200 watts for “C” version with a cover and fan assembly 100 watts for “B” version without moving air (maximum current of ouput #1 & #2 derated to 50%),
or 200 watts with 25 CFM forced air provided by user.
(5) When the remote Sense facility is not used, +Sense must be connected to +V, and -Sense to Return, on P2 connector.
(6) All models may be operated at no-load. At no-load, output voltage tolerance increases to 10%.
Pin Chart
CONNECTOR
MODEL/PIN
P1
P2
3
P3
4
P4
1
1
2
5
6
7
8
9
2
3
4
1
2
OBN01016
OBN01031
OBN01046
OBN01070
OBN01013
OBN0PM201-13 -Sense
OBN01055
RET.
RET.
RET
RET
FAN
RET
Output Output Output +Sense
COM COM. PFD INHIBIT+INHIBIT-
RET.
#1
#1
#1
OBN01204
OBN01210
OBN01207
Output #1
FAN
FAN
FAN
Output COM COM COM Output
#1 RET. RET. RET. #2
COM COM PFD INHIBIT+INHIBIT-
RET. RET.
N.C.
N.C.
N.C.
N.C.
N.C.
OBN01310
OBN01316
OBN01313
Output #1
OBN01315
Output COM COM COM Output Output
#1 RET. RET. RET. #2 #3
COM COM PFD INHIBIT+INHIBIT-
RET. RET.
OBN01319
Output #1 Output COM COM COM Output N.C. Output Output
COM COM PFD INHIBIT+INHIBIT-
RET. RET.
#1
RET.
RET.
RET.
#2
#3
#3
RETURN
OBN01404
OBN01407
OBN01420
OBN01403
OBN01410
OBN14113
FAN
COM COM COM Output Output Output Output
COM COM PFD INHIBIT+INHIBIT-
RET. RET.
Output
#1
Output #1
RET.
RET.
RET.
#2
#2
#3
#4
RETURN
346
相关型号: