REM1-0512S [RECOM]
1 Watt SIP7 Single Output;型号: | REM1-0512S |
厂家: | RECOM INTERNATIONAL POWER |
描述: | 1 Watt SIP7 Single Output |
文件: | 总5页 (文件大小:481K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
• Medical Grade DC/DC Converter
Features
• 250VAC Working, 2MOPP
DC/DC Converter
• 5.2kVDC Isolation for 250VAC working voltage
• -40°C Up To +90°C Operating Temperature
• 3rd Ed. Safety And 4th. Ed. EMC (pending)
Unregulated
Converters
REM1
Description
1 Watt
The REM1 complements the existing board-mount REM3, REM6 and REM10 series by offering
a 1W medical grade unregulated DC/DC converter in a more compact SIP7 package. The REM1
features reinforced 5.2kVDC/1 minute isolation and 2MOPP/250VAC working voltage. The REM1 is
available with 3.3, 5, 12, 15 or 24V inputs and offers 3.3, 5 or 12V outputs with up to 85%
efficiency. The operating temperature range is -40°C up to +90°C. The converter is Class B EMC and
60601-1-2 (4th Ed.) medical EMC certified using a simple external LC filter. The converters are
fully certified to CB, IEC/EN and ANSI/AAMI 60601 third edition safety standards, RoHS2+ (10/10)
and REACH and come with a 5 year warranty.
SIP7
Single Output
Selection Guide
Part
Number
nom. Input
Voltage
[VDC]
Output
Voltage
[VDC]
Output
Current
[mA]
Efficiency Max. Capacitive
typ. (1)
[%]
Load (2)
[µF]
REM1-3.33.3S
REM1-3.305S
REM1-3.312S
REM1-053.3S
REM1-0505S
REM1-0512S
REM1-123.3S
REM1-1205S
REM1-153.3S
REM1-1505S
REM1-243.3S
REM1-2405S
3.3
3.3
3.3
5
3.3
5
303
200
84
78
81
85
79
80
84
78
81
77
81
76
80
2200
2200
470
12
3.3
5
E314885
303
200
84
2200
2200
470
5
5
12
3.3
5
2MOPP
250VAC
12
12
15
15
24
24
303
200
303
200
303
200
2200
2200
2200
2200
2200
2200
IEC/EN60601-1 certified
ANSI/AAMI ES60601-1 certified
EN62368-1 certified
IEC/EN60601-1-2
EN55011
3.3
5
3.3
5
CB Report
Notes:
Note1: Efficiency tested by nominal input and full load at +25°C ambient.
Note2: Max Cap Load tested by nominal input and full resisitive load.
Specifications (measured @ ta= 25°C, nominal input voltage, full load and after warm-up)
BASIC CHARACTERISTICS
Parameter
Condition
Min.
Typ.
Max.
Internal Input Filter
Capacitor
3.3VDC
2.97VDC
4.5VDC
10.8VDC
13.5VDC
21.6VDC
3.63VDC
5.5VDC
13.2VDC
16.5VDC
26.4VDC
5VDC
nom. Vin= 12VDC
15VDC
Input Voltage Range
Maximum Reverse Voltage
Quiescent Current
24VDC
0VDC
3.3VDC
5VDC
nom. Vin= 12VDC
15VDc
40mA
25mA
12mA
10mA
7mA
24VDC
continued on next page
www.recom-power.com
REV.: 2/2017
MED-1
REM1
DC/DC Converter
Specifications (measured @ ta= 25°C, nominal input voltage, full load and after warm-up)
Series
Min.
Typ.
Max.
Parameter
Condition
40kHz
Internal Operating Frequency
Minimum Load
0%
Output Ripple and Noise (3)
20MHz BW
75mVp-p
Notes:
Note3: Measurements are made with a 0.1µF MLCC across output. (low ESR).
Efficiency vs. Load
REM1-xx3.3S
REM1-xx05S
100
90
80
70
60
50
40
30
20
10
0
100
90
80
70
60
50
40
30
20
10
0
3.305
3.33.3
053.3
123.3
153.3
243.3
0505
1205
1505
2405
0
10
20
30
40
50
60
70
80
90 100
0
10
20
30
40
50
60
70
80
90 100
Load Current [A]
Load Current [A]
REM1-xx12S
100
90
80
70
60
50
40
30
20
10
0
3.312
0512
0
10
20
30
40
50
60
70
80
90 100
Load Current [A]
Tolerance Envelope
REM1-3.33.3S
REM1-2405S
+12
+8
+3
+6
+5
+5
V Nom.
V Nom.
-0.5
-5
-0.5
-5
10
50
100
10
50
100
+Vout and -Vout Load Current [%]
+Vout and -Vout Load Current [%]
MED-2
REV.: 2/2017
www.recom-power.com
REM1
DC/DC Converter
Specifications (measured @ ta= 25°C, nominal input voltage, full load and after warm-up)
Series
REGULATIONS
Parameter
Output Accuracy
Line Regulation
Condition
Value
5% max.
low line to high line, full load
1.2% typ. @ 1% of Vin
3.3Vout and 5Vout
12Vout
8% typ. / 12% max.
5% typ. / 8% max.
Load Regulation
10% to 100% load
Accuracy vs. Load
REM1-xx3.3S
REM1-xx05S
30
25
20
15
10
5
30
25
20
15
10
5
3.33.3
053.3
123.3
153.3
243.3
3.305
0505
1205
1505
2405
0
0
-5
-5
-10
-10
0
10
20
30
40
50
60
70
80
90 100
0
10
20
30
40
50
60
70
80
90 100
Load Current [A]
Load Current [A]
REM1-xx12S
30
3.312
0512
25
20
15
10
5
0
-5
-10
0
10
20
30
40
50
60
70
80
90 100
Load Current [A]
PROTECTIONS
Parameter
Type
Value
5.2kVDC
4kVAC
Isolation Voltage (4)
I/P to O/P
tested for 1 minute
Isolation Resistance
Isolation Capacitance
Insulation Grade
10GΩ min.
25pF typ.
reinforced
2MOPP
Means of Protection
250VAC working voltage
Creepage and Clearance
≥8mm
Notes:
Note4: For repeat Hi-Pot testing, reduce the time and/or the test voltage.
Note5: An input fuse is required if the mains supply is not over-current protected. Recommended fuse: T1A slow blow type
www.recom-power.com
REV.: 2/2017
MED-3
REM1
DC/DC Converter
Specifications (measured @ ta= 25°C, nominal input voltage, full load and after warm-up)
Series
ENVIRONMENTAL
Parameter
Condition
without derating (see graph)
Value
-40°C °C to +85°C
105°C
Operating Temperature Range
Maximum Case Temperature
Temperature Coefficient
0.02%/°C typ.
according to IEC/EN60601-1
accroding to IEC/EN62368-1
3000m
5000m
Operating Altitude
Operating Humidity
Pollution Degree
non-condensing
5% - 95% RH max.
PD2
+25°C
+85°C
18200 x 103 hours
7500 x 103 hours
MTBF
according to MIL-HDBK-217F, G.B.
Vibration
according to MIL-STD-202G standard
Derating Graph
(@ chamber and natural convection 0.1m/s)
100
80
60
40
30
20
0
-40 -20
0
20 40
60
80 100 120
85 90
Ambient Temperature [°C]
SAFETY AND CERTIFICATIONS
Certificate Type (Safety)
Report / File Number
Standard
Audio/video, information and communication technology equipment. Safety
requirements (CB Scheme)
L0339m31-A-L
EN62368-1:2014
Medical Electric Equipment, General Requirements for Saftey and Essential
Performance
ANSI/AAMI ES60601-1:2012 + A2 2010/2012
CSA C22.2 No. 60601-1:14, 3rd Edition, 2014
E314885-D1000-1/A0/C0-UL
E314885-D1000-1/A0/C0-CB
Medical Electric Equipment, General Requirements for Safety and Essential
Performance
IEC60601-1:2005 +AM1:2012
EN60601-1:2006 + A12:2014
RoHs2+
RoHS-2011/65/EU + AM-2015/863
EMC Compliance
Condition
Standard / Criterion
Medical electrical equipment - Part 1-2: General requirements for basic safety
and essential performance - Collateral standard: Electromagnetic compatibility
IEC60601-1-2:2014
EN60601-1-2:2015
Industrial, scientific and medical equipment - Radio frequency disturbance
characteristics - Limits and methods of measurement
with external filter
EN55011, 2009+A1:2010, Class B
ESD Electrostatic discharge immunity test
Radiated, radio-frequency, electromagnetic field immunity test
Fast Transient and Burst Immunity
air: 15kV; contact: 8kV
3V/m
IEC61000-4-2:2008 , Criteria A
IEC61000-4-3:2006+A2:2010, Criteria A
IEC61000-4-4:2012, Criteria A
DC port: 2kV
DC port: 1kV
DC port: 6V
Surge Immunity
IEC61000-4-5:2014, Criteria B
Immunity to conducted disturbances, induced by radio-frequency fields
Power Magnetic Field Immunity
IEC61000-4-6:2013, Criteria A
50Hz, 30A/m
IEC61000-4-8:2009, Criteria A
continued on next page
MED-4
REV.: 2/2017
www.recom-power.com
REM1
DC/DC Converter
Specifications (measured @ ta= 25°C, nominal input voltage, full load and after warm-up)
Series
EMC Filtering Suggestions according to EN55011
C2
Component list Class B
Input Voltage
3.3VDC
5VDC
C1
C2
L1
L1
+Vin
-Vin
4.7µF
2.2µF
22µH Choke
47µH Choke
C1
12VDC
470pF/6kVDC
15VDC
24VDC
DIMENSION and PHYSICAL CHARACTERISTICS
Parameter
Type
Case
Potting
PCB
Value
black plastic, (UL94 V-0)
silicone (UL94 V-0)
FR4 (UL94 V-0)
Material
Package Dimension (LxWxH)
Package Weight
19.6 x 6.0 x 10.2mm
2.6g typ.
Dimension Drawing (mm)
embossed logo
19.6
6.0
Pin Connections
Pin #
Single
1
2
6
7
+Vin
-Vin
-Vout
+Vout
Marking
1 2
6 7
Tolerance: xx.x=
xx.xx= 0.25mm
Pin dimension: 0.1mm
0.5mm
0.51
recommended footprint detail
1.00 Ø +0.15/-0
Bottom View
Top View
2.0
6x2.54= 15.24
2.54
PACKAGING INFORMATION
Parameter
Type
Value
Packaging Dimension (LxWxH)
tube
520.0 x 16.0 x 9.3mm
Packaging Quantity
25pcs
Storage Temperature Range
Storage Humidtiy
-55°C to +125°C
TBD
non-condensing
The product information and specifications are subject to change without prior notice. RECOM products are not authorized for use in safety-critical applications (such as life support) without RECOM’s explicit written
consent. A safety-critical application is defined as an application where a failure of a RECOM product may reasonably be expected to endanger or cause loss of life, inflict bodily harm or damage property. The buyer
shall indemnify and hold harmless RECOM, its affiliated companies and its representatives against any damage claims in connection with the unauthorized use of RECOM products in such safety-critical applications.
www.recom-power.com
REV.: 2/2017
MED-5
相关型号:
©2020 ICPDF网 联系我们和版权申明