REM15-4812SW [RECOM]
15 Watt 4:1 Input 1.6 x 1 Single and Dual Output;型号: | REM15-4812SW |
厂家: | RECOM INTERNATIONAL POWER |
描述: | 15 Watt 4:1 Input 1.6 x 1 Single and Dual Output |
文件: | 总8页 (文件大小:2031K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
• Reinforced insulation for 250VAC working voltage
• Clearance and creepage distance >8.0mm
• 5kVAC I/P to O/P 2MOPP isolation
• 2µA patient leakage current
Features
DC/DC Converter
Regulated
Converter
• Industry standard pinout
• 4:1 wide input range
REM15-W
Description
15 Watt
4:1 Input
1.6“ x 1“
Single and Dual
Output
The REM15-W series of medical grade regulated DC/DC converters features reinforced 5kVAC/1 minute
isolation with low 2µA leakage (B, BF and CF compatible) and are 60601-1 3rd Ed. certified for 250VAC
continuous working voltage isolation. The industry standard 1.6”x1” package offers tightly regulated
single and dual outputs, with low output ripple and zero-load operation.The outputs are also short circuit
and overload protected. The converters are certified to CB, IEC/EN and ANSI/AAMI standards and carry
the UL mark.
Selection Guide
Part
Number
Input
Voltage Range
[VDC]
Output
Voltage
[VDC]
Output
Current
[mA]
Efficiency
typ. (1)
[%]
Max. Capacitive
Load (2)
[µF]
REM15-2405SW (3)
REM15-2412SW (3)
REM15-2415SW (3)
REM15-2424SW (3)
REM15-2405DW (3)
REM15-2412DW (3)
REM15-2415DW (3)
REM15-4805SW (3)
REM15-4812SW (3)
REM15-4815SW (3)
REM15-4824SW (3)
REM15-4805DW (3)
REM15-4812DW (3)
REM15-4815DW (3)
9-36
9-36
5
3000
1250
1000
625
87
88.5
88
3800
650
530
190
1900
380
270
3800
650
530
190
1900
380
270
12
15
24
5
9-36
9-36
88
9-36
1500
625
86
9-36
12
15
5
88
9-36
500
88
18-75
18-75
18-75
18-75
18-75
18-75
18-75
3000
1250
1000
625
89.5
88
12
15
24
5
E314885
2MOPP
88
250VAC
88.5
86
UL60950-1 certified
CSA/CAN C22.2 60950-1-07 certified
UL62368-1 certified
CSA/CAN C22.2 62368-1 certified
CSA/CAN C22.2 60601-01 certified
ANSI/AAMI ES60601-1 certified
EN55011 certified
1500
625
12
15
88.5
88
500
Notes:
Note1: Efficiency is tested at nominal input and full load at +25°C ambient
Note2: Max Cap Load is tested at nominal input and full resisitive load
CB report
Model Numbering
REM15- __ __ _ W/P
Output Power
nom. Input Voltage
nom. Output Voltage
CTRL pin option (positive logic)
Wide Input Voltage Range (4:1)
Single or Dual
Notes:
Note3: standard is with suffix „/P“ (CTRL pin with positive logic)
without suffix is without CTRL pin (no pin) please refer to dimensions drawing
Ordering Examples:
REM15-2412SW/P = 4:1 Input, 9-36Vin, 12Vout, with control pin positive logic
REM15-4815DW = 4:1 Input, 18-75Vin, 15Vout, without control pin
www.recom-power.com
REV.: 0/2018
MED-1
REM15-W
DC/DC Converter
Series
Specifications (measured @ Ta= 25°C, nominal input voltage, full load and after warm-up)
BASIC CHARACTERISTICS
Parameter
Condition
Min.
Typ.
Max.
Internal Input Filter
Pi type
nom. Vin= 24VDC
nom. Vin= 48VDC
9VDC
24VDC
48VDC
36VDC
75VDC
50VDC
100VDC
8.6VDC
Input Voltage Range
Input Surge Voltage
18VDC
nom. Vin= 24VDC
nom. Vin= 48VDC
3 second max.
DC-DC ON
DC-DC OFF
nom. Vin= 24VDC
nom. Vin= 48VDC
7.8VDC
Under Voltage Lockout
Quiescent Current
DC-DC ON
DC-DC OFF
17.4VDC
15.8VDC
nom. Vin= 24VDC
13mA
10mA
nom. Vin= 48VDC
5Vout, 12Vout
-10%
-10%
0%
+10%
+20%
Output Voltage Trimming
Minimum Load
15Vout, 24Vout
ON/OFF CTRL
Power up
10ms
30ms
60ms
60ms
Start-up Time
DC-DC ON
DC-DC OFF
DC-DC ON
DC-DC OFF
Open or 3.5VDC - 12VDC
Short or 0VDC - 1.2VDC
1mA
ON/OFF Control (4)
Input Current oft CTRL Pin
Standby Current
-0.5mA
220kHz
2.5mA
Internal Operating Frequency
250kHz
50mVp-p
75mVp-p
100mVp-p
300kHz
5Vout, single and dual
witth a 10µF X7R/MLCC
with a 4.7µF X7R/MLCC
Output Ripple and Noise (20MHz BW)
12 & 15Vout, single and dual
24Vout single
Notes:
Note4: The ON/OFF control function is positive logic. The pin voltage is referenced to -Vin pin
Efficiency vs. Load
REM15-2405SW
REM15-2412SW
100
90
80
70
60
50
40
30
20
10
0
100
90
80
70
60
50
40
30
20
10
0
24Vin
18Vin
12Vin
9Vin
24Vin
18Vin
12Vin
9Vin
0
10 20 30 40 50 60 70 80 90 100
0
10 20 30 40 50 60 70 80 90 100
Output Load [%]
Output Load [%]
continued on next page
www.recom-power.com
REV.: 0/2018
MED-2
REM15-W
DC/DC Converter
Series
Specifications (measured @ Ta= 25°C, nominal input voltage, full load and after warm-up)
Efficiency vs. Load
REM15-4805SW
REM15-4824SW
100
90
80
70
60
50
40
30
20
10
0
100
90
80
70
60
50
40
30
20
10
0
48Vin
36Vin
24Vin
18Vin
48Vin
36Vin
24Vin
18Vin
0
10 20 30 40 50 60 70 80 90 100
0
10 20 30 40 50 60 70 80 90 100
Output Load [%]
Output Load [%]
Power Dissipation vs. Load
REM15-24xxSW
REM15-48xxSW
3
3
2.5
2.5
2
2
1.5
1.5
1
1
24Vin
12Vin
48Vin
24Vin
0.5
0.5
0
0
0
10 20 30 40 50 60 70 80 90 100
0
10 20 30 40 50 60 70 80 90 100
Output Load [%]
Output Load [%]
Output Voltage Trimming
It allows the user to increase or decrease the output voltage of the module. This is accomplished by connecting an external resistor between the Trim pin and
either the +Vout or -Vout pins. With an external resistor between the Trim and -Vout pin, the output voltage increases. With an external resistor between the
Trim and +Vout pin, the output voltage decreases. The external Trim resistor needs to be at least 1/16W of rated. The values for trim resistors shown in trim
tables below are according to standard E96 values; therefore, the specified voltage may slightly vary.
Trim up
Trim down
-V
+V
out
out
R
R
down
up
Trim
Trim
continued on next page
www.recom-power.com
REV.: 0/2018
MED-3
REM15-W
DC/DC Converter
Series
Specifications (measured @ Ta= 25°C, nominal input voltage, full load and after warm-up)
Trim Calculation
Voutnom = nominal output voltage [VDC]
Voutnom
R1
R2
ku
Vref
Voutset
Vref
= trimmed output voltage [VDC]
5VDC
5k1W
10kW
10kW
56kW
2kW
5k1W
5k1W
13kW
2.5
9.5
= reference voltage
= trim up resistor
= trim down resistor
[VDC]
[W]
12VDC
15VDC
24VDC
Rup
2.5VDC
12.5
21.5
Rdown
[W]
R1 & R2 = internal resistors
[W]
ku
= trim up factor
[ ]
Calculation:
R1 x Vref
(Voutset - Vref ) x R1
- R2
- R2
Rup =
Rdown
=
=
Voutset - Vref - ku
Voutnom - Voutset
Practical Example REM15-1205SW +10% / -10%:
(4.5 - 2.5) x 5k1
5k1 x 2.5
5.5 - 2.5 - 2.5
- 2k = 18k4W
- 2k = 23k5W
Rdown
Rup =
5 - 4.5
Rup according to E96 ≈ 23k7W
Rdown according to E96 ≈ 18k2W
REM15-xx05SW(/P)
Trim up
1
2
3
4
5
6
7
8
9
10
[%]
Vout
Rup (E96) ≈
=
5.05
255k
5.10
127k
5.15
82k5
5.20
61k9
5.25
48k7
5.30
40k2
5.35
34k8
5.40
301
5.45
26k1
5.50
23k7
[VDC]
[W]
set
Trim down
1
2
3
4
5
6
7
8
9
10
[%]
Vout
=
4.95
249k
4.90
121k
4.85
78k7
4.80
56k2
4.75
44k2
4.70
35k7
4.65
29k4
4.60
24k9
4.55
21k
4.50
18k2
[VDC]
[W]
set
R
down (E96) ≈
REM15-xx12SW(/P)
Trim up
1
2
3
4
5
6
7
8
9
10
[%]
Vout
Rup (E96) ≈
=
12.12
205k
12.24
10k
12.36
64k9
12.48
47k5
12.60
36k5
12.72
29k4
12.84
24k9
12.96
20k1
13.08
17k9
13.20
15k8
[VDC]
[W]
set
Trim down
1
2
3
4
5
6
7
8
9
10
[%]
Vout
=
11.88
768k
11.76
383k
11.64
249k
11.52
182k
11.40
143k
11.28
118k
11.16
97k6
11.04
84k5
10.92
73k2
10.8
63k4
[VDC]
[W]
set
R
down (E96) ≈
REM15-xx15SW(/P)
Trim up
1
2
3
4
5
6
7
8
9
10
[%]
Vout
Rup (E96) ≈
=
15.15
162k
15.30
78k7
15.45
49k9
15.60
36k5
15.75
28k
15.90
22k6
16.05
18k7
16.20
15k8
16.35
13k3
16.50
11k5
[VDC]
[W]
set
Trim up
11
16.65
10k
12
16.80
8k8
13
16.95
7k6
14
17.10
6k8
15
17.25
6k
16
17.40
5k3
17
17.55
4k6
18
17.70
4k1
19
17.85
3k6
20
18.00
3k2
[%]
Vout
=
[VDC]
[W]
set
Rup (E96) ≈
Trim down
1
2
3
4
5
6
7
8
9
10
[%]
Vout
=
14.85
825K
14.70
402k
14.55
261k
14.40
191k
14.25
150k
14.10
124k
13.95
105k
13.80
88k7
13.65
76k8
13.50
68k1
[VDC]
[W]
set
R
down (E96) ≈
continued on next page
www.recom-power.com
REV.: 0/2018
MED-4
REM15-W
DC/DC Converter
Series
Specifications (measured @ Ta= 25°C, nominal input voltage, full load and after warm-up)
REM15-xx24SW(/P)
Trim up
Vout
Rup (E96) ≈
1
2
3
4
5
6
7
8
9
10
[%]
=
24.24
576k
24.48
280k
24.72
182k
24.96
133k
25.2
105k
25.44
84k5
25.68
69k8
25.92
95k3
26.16
52k3
26.40
45k3
[VDC]
[W]
set
Trim up
11
12
13
14
15
16
17
18
19
20
[%]
Vout
Rup (E96) ≈
=
26.64
40k2
26.88
35k7
27.12
31k6
27.36
28k7
27.6
26k1
27.84
23k7
28.08
21k5
28.32
19k6
28.56
17k9
28.80
16k2
[VDC]
[W]
set
Trim down
1
2
3
4
5
6
7
8
9
10
[%]
Vout
=
23.76
4M99
23.52
2M43
23.28
1M62
23.04
1M18
22.8
931k
22.56
768k
22.32
649k
22.08
562k
21.84
487k
21.60
432k
[VDC]
[W]
set
R
down (E96) ≈
REGULATIONS
Parameter
Condition
Value
Output Accuracy
1.0% max.
0.2% max.
0.5% max.
0.2% max.
1.0% max.
5.0% max.
Single Output
Line Regulation
Load Regulation
low line to high line
no load to full load
Dual Output
Single Output
Dual Output
Cross Regulation
assymetrical load 25% / 100% full load
only Dual Output
Transient Response
recovery time 25% load step change
100µs typ. / 250µs max.
PROTECTIONS
Parameter
Short Circuit Protection (SCP) (6)
Condition
Value
continuous, auto-recovery
Over Load Protection (OLP)
% of Iout rated
Hiccup mode, 150% typ. / 185% max.
5Vout
12Vout
15Vout
24Vout
6.2VDC typ.
15VDC typ.
20VDC typ.
30VDC typ.
Output Over Voltage Protection (OVP)
Zener diode clamp
Over Temperature Protection (OTP)
Isolation Voltage(5)
at tc point (refer to dimension drawing)
+115°C max.
I/P to O/P
tested for 1 minute
continuous
5kVAC
250VAC
working voltage
Isolation Resistance
Isolation Capacitance
Leakage Current
2GW min.
20pF typ.
2µA typ. / 2.5µA max.
reinforced
240VAC, 60Hz
Insulation Grade
Means of Protection
Medical Device Classification
Clearance/Creepage
2MOPP
built-in power supply
>8.0mm
Notes:
Note5: For repeat Hi-Pot testing, reduce the time and/or the test voltage
Note6: Refer to local safety regulations if input over-current protection is also required. Recommended fuse: slow blow type
www.recom-power.com
REV.: 0/2018
MED-5
REM15-W
DC/DC Converter
Series
Specifications (measured @ Ta= 25°C, nominal input voltage, full load and after warm-up)
ENVIRONMENTAL
Parameter
Condition
with derating @ natural convection 0.1m/s
Value
-40°C to +105°C
+110°C typ.
Operating Temperature Range (7)
Max. Case Temperature
Operating Humidity
5% to 95% RH
0.02%/K
Temperature Coefficient
natural convection (0.1m/s)
9K/W
Thermal Impedance (7)
horizontal direction
0.5m/s
6.4K/W
Operating Altitude
Pollution Degree
Thermal Shock
Vibration
5000m
PD2
according to MIL-STD-810F
according to MIL-STD-810F
2000 x 103 hours
MTBF
according to MIL-HDBK-217F, full load
+25°C
Notes:
Note7: Measured with test PCB: Eurocard 160x100mm 105µm copper, double layer
Derating Graph (7)
(@ Chamber and nominal Vin)
REM15-24xxSW
REM15-48xxSW
100
90
80
70
60
50
40
30
20
100
90
80
70
60
50
40
30
20
10
0
natural convection (0.1m/s)
0.5m/s
natural convection (0.1m/s)
0.5m/s
10
0
-40 -30 -20 -10 0 10 20 30 40 50 60 70 80 90 100 110
-40 -30 -20 -10 0 10 20 30 40 50 60 70 80 90 100 110
89 95 105
96 105
Ambient Temperature [°C]
Ambient Temperature [°C]
SAFETY AND CERTIFICATIONS
Certificate Type (Safety)
Report / File Number
Standard
Information Technology Equipment, General Requirements for
Safety
UL60950-1, 2nd Edition, 2014
CAN/CSA-C22.2 No. 60950-1-07, 2nd Edition, 2014
E196683
Audio/video, information and communication technology equip-
ment. Safety requirements
UL62368-1
CAN/CSA-C22.2 No. 62368-1
Medical Electric Equipment, General Requirements for Safety and
Essential Performance
ANSI/AAMI ES60601-1 (2005/R2012 + A1:2012), 2012
CAN/CSA-C22.2 No. 60601-1:14, 3rd Edition, 2014-03
E314885
Medical Electric Equipment, General Requirements for Safety and
Essential Performance (CB Scheme)
180505201
IEC60601-1:2005, 3rd Edition + AM1:2012
RoHs 2011/65/EU + AM2015/863
RoHS2+
continued on next page
REV.: 0/2018
www.recom-power.com
MED-6
REM15-W
DC/DC Converter
Series
Specifications (measured @ Ta= 25°C, nominal input voltage, full load and after warm-up)
EMC Compliance
Conditions
Standard / Criterion
Medical electrical equipment Part 1-2: Electromagnetic distur-
bances - Requirements and tests
EN60601-1-2:2015
without external filter
with external filter
EN55011, Class A
EN55011, Class B
Industrial, scientific and medical equipment - Radio frequency
disturbance characteristics - Limits and methods of measurement
AS/NZS CISPR11:2011 Class A, B
Air: 2, 4, 8, 15kV and
Contact: 4, 6, 8kV
EN61000-4-2:2009, Criteria A
IEC61000-4-2:2008, Criteria A
ESD Electrostatic discharge immunity test
10V/m (80-2700MHz) and 9V/m,
27V/m, 28V/m (several frequencies)
EN61000-4-3:2006 + A2:2010, Criteria A
IEC61000-4-3:2006 + A2:2010, Criteria A
Radiated, radio-frequency, electromagnetic field immunity test
Fast Transient and Burst Immunity(8)
Surge Immunity(8)
DC Power Port: 0.5, 1, 2kV
DC Power Port: 0.5, 1, 2kV
IEC/EN61000-4-4:2012, Criteria A
IEC/EN61000-4-5:2014, Criteria A
Immunity to conducted disturbances, induced by radio-frequency DC Power Port 10V (0.15-80MHz)
EN61000-4-6:2014 + AC:2015, Criteria A
IEC61000-4-6:2013, Criteria A
fields
DC Power Port 10V (ISM bands)
Power Frequency Magnetic Field
50Hz, 100A/m (1 min)
1000A/m (1 sec)
EN61000-4-8:2010, Criteria A
IEC61000-4-8:2009, Criteria A
Notes:
Note8: 24Vin models tested with 2pcs of Nippon chemi-con KY series (220µF/100V) MLCC and a TVS (SMDJ58A, 58V, 3000W peak pulse power) in parallel
48Vin models tested with 2pcs of Nippon chemi-con KY series (220µF/100V) MLCC and a TVS (SMDJ120A, 120V, 3000W peak pulse power) in parallel
EMC Filtering Suggestions according to EN55011
Single output
Dual output
ꢀ3
ꢀ4
ꢀ3
ꢀ4
ꢁ1
ꢁ1
+VIN
-VIN
+VOUT
+VIN
-VIN
+VOUT
COM
-VOUT
ꢀ1
ꢀ2
ꢀ1
ꢀ2
-VOUT
Component List Class B
nom. Vin
C1
C2
C3, C4
L1
10µF/50V
1210 MLCC
145µH 5.2A
CMC
24VDC
N/A
100pF/3kV
2.2µF/100V
1210 MLCC
2.2µF/100V
1210 MLCC
373µH 3A
CMC
48VDC
100pF/3kV
DIMENSION and PHYSICAL CHARACTERISTICS
Parameter
Type
Value
case
baseplate
potting
non-conductive black plastic (UL94-V0)
non-conductive black plastic (UL94-V0)
silicone (UL94-V0)
Material
Dimension (LxWxH)
Weight
40.6 x 25.4 x 10.2mm
24g
continued on next page
REV.: 0/2018
www.recom-power.com
MED-7
REM15-W
DC/DC Converter
Series
Specifications (measured @ Ta= 25°C, nominal input voltage, full load and after warm-up)
Dimension Drawing (mm)
tc
1
40.6
25.4
0.1
Ø1.0
Recommended Footprint Details
14 x 2.54= 35.56
2.54
6
5
4
4
5
6
1
(3)
2
Bottom View
Top View
2
1
(3)
Pin Connections with suffix /P
Pin Connections without suffix /P
Pin #
Single
+Vin
-Vin
CTRL
Trim
Dual
+Vin
-Vin
CTRL
-Vout
Com
+Vout
Pin #
Single
+Vin
-Vin
Trim
-Vout
+Vout
Dual
+Vin
-Vin
-Vout
Com
+Vout
1
2
3
4
5
6
1
2
4
5
6
-Vout
+Vout
Tolerance: xx.x= 0.5mm
xx.xx= 0.25mm
Tolerance: xx.x= 0.5mm
xx.xx= 0.25mm
PACKAGING INFORMATION
Parameter
Type
Value
Packaging Dimension (LxWxH)
Packaging Quantity
tube
290.0 x 43.5 x 19.7mm
10 pcs
Storage Temperature Range
Stoarge Humidity
-55°C to +125°C
5% to 95% RH max.
non-condensing
The product information and specifications may be subject to changes even without prior written notice.The product has been designed for various applications; its suitability lies in the responsibility of each customer. The
products are not authorized for use in safety-critical applications without RECOM’s explicit written consent. A safety-critical application is an application where a failure may reasonably be expected to endanger or cause
loss of life, inflict bodily harm or damage property. The applicant shall indemnify and hold harmless RECOM, its affiliated companies and its representatives against any damage claims in connection with the unauthorized
use of RECOM products in such safety-critical applications.
www.recom-power.com
REV.: 0/2018
MED-8
相关型号:
©2020 ICPDF网 联系我们和版权申明