REM2-xx12D [RECOM]

2 Watt SIP8 Single and Dual Output;
REM2-xx12D
型号: REM2-xx12D
厂家: RECOM INTERNATIONAL POWER    RECOM INTERNATIONAL POWER
描述:

2 Watt SIP8 Single and Dual Output

文件: 总7页 (文件大小:700K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
• Medical grade DC/DC converter  
Features  
• 250VAC working, 2MOPP up to 5000m altitude  
• 5.2kVDC/1 minute isolation  
DC/DC Converter  
• Single or dual outputs  
Unregulated  
Converter  
• -40°C up to +95°C operating temperature  
• Medical certified (3rd Ed. safety, 4th Ed. EMC)  
REM2  
Description  
2 Watt  
SIP8  
Single and Dual  
Output  
The board-mount REM2 series complements the REM1 series by offering a 2W medical grade DC/DC  
converter in a compact SIP8 package. The REM2 features reinforced 5.2kVDC/1 minute isolation and  
2MOPP/250VAC working voltage at 5000m. It offers single and dual outputs with up to 85% efficiency.  
The operating temperature range is -40°C up to +80°C without derating, and up to +95°C with 50%  
load.The converter is compliant to Class A/B EMC and 60601-1-2 (4th Ed.) medical EMC using a simple  
external LC filter.The REM2 is certified to CB, IEC/EN and ANSI/AAMI 60601 third edition medical safety  
standards (pending) and comes with a 5 year warranty.  
Selection Guide  
Part  
nom. Input  
Voltage  
[VDC]  
Output  
Voltage  
[VDC]  
Output  
Current  
[mA]  
Efficiency Max. Capacitive  
Number  
typ.(1)  
[%]  
Load (2)  
[µF]  
REM2-xx3.3S  
REM2-xx05S  
REM2-xx09S  
REM2-xx12S  
REM2-xx3.3D  
REM2-xx05D  
REM2-xx12D  
3.3 / 5 / 12 / 15 / 24  
3.3 / 5 / 12 / 15 / 24  
3.3 / 5 / 12 / 15 / 24  
3.3 / 5 / 12 / 15 / 24  
5 / 15  
3.3  
5
606  
400  
222  
167  
303  
200  
84  
77  
79  
84  
82  
79  
82  
85  
1000  
1000  
470  
330  
680  
680  
150  
E314885  
9
12  
3.3  
5
3.3 / 5 / 12 / 15 / 24  
3.3 / 5 / 12 / 15 / 24  
2MOPP  
250VAC  
12  
Notes:  
CAN/CSA-C22.2 No. 60601-1:14 certified  
ANSI/AAMI ES60601-1 certified  
IEC/EN60601-1 certified  
IEC/EN62368-1 certified  
EN60601-1-2 compliant  
EN55011 compliant  
Note1: Efficiency is tested at nominal input and full load at +25°C ambient  
Note2: Max Cap Load is tested at minimum input and full resistive load  
CB report  
Model Numbering  
REM2- __ __ _  
nom. Output Power  
Input Voltage  
Single or Dual  
Output Voltage  
Ordering Examples  
REM2-0505S =  
REM2-2412D =  
5Vin  
24Vin  
5Vout  
12Vout  
Single  
Dual  
www.recom-power.com  
REV.: 1/2019  
MED-1  
REM2  
DC/DC Converter  
Specifications (measured @ Ta= 25°C, nom. Vin, full load unless otherwise stated)  
Series  
BASIC CHARACTERISTICS  
Parameter  
Condition  
Min.  
Typ.  
Max.  
Internal Input Filter  
Input Voltage Range  
capacitor  
10%  
750mA  
520mA  
210mA  
175mA  
110mA  
3.3VDC  
5VDC  
nom. Vin = 12VDC  
15VDC  
Input Current  
24VDC  
55mA  
46mA  
24mA  
18mA  
10mA  
3.3VDC  
5VDC  
nom. Vin = 12VDC  
15VDC  
Quiescent Current  
24VDC  
Minimum Load  
0%  
Internal Operating Frequency  
Output Ripple and Noise (3)  
20kHz  
20MHz BW  
150mVp-p  
Notes:  
Note3: Measurements are made with a 0.1µF MLCC across output (low ESR)  
Efficiency vs. Load  
REM2-xx3.3S(D)  
REM2-xx05S(D)  
100  
90  
80  
70  
60  
50  
40  
30  
20  
10  
0
100  
90  
80  
70  
60  
50  
40  
30  
20  
10  
0
0505S  
053.3S  
053.3D  
153.3S  
153.3D  
1205S  
1205D  
2405S  
2405D  
0
10 20 30 40 50 60 70 80 90 100  
0
10 20 30 40 50 60 70 80 90 100  
Output Load [%]  
Output Load [%]  
REM2-3.312S(D)  
100  
90  
80  
70  
60  
50  
40  
30  
20  
10  
0
3.312S  
3.312D  
0
10 20 30 40 50 60 70 80 90 100  
Output Load [%]  
www.recom-power.com  
REV.: 1/2019  
MED-2  
REM2  
DC/DC Converter  
Specifications (measured @ Ta= 25°C, nom. Vin, full load unless otherwise stated)  
Series  
REGULATIONS  
Parameter  
Output Accuracy  
Line Regulation  
Condition  
Value  
5.0% max.  
low line to high line, full load  
1.2% typ. @ 1.0% Vin  
3.3, 5Vout  
9, 12Vout  
15.0% max.  
10.0% max.  
Load Regulation (4)  
Cross Regulation (4)  
10% to 100% load  
10% to 100% load  
dual output only  
5.0% typ.  
Notes:  
Note4: Operation below 10% load will not harm the converter, but specifications may not be met  
Tolerance Envelope  
REM2-3.312S  
REM2-2405S  
+15%  
+10%  
+6%  
+7%  
+5%  
0%  
+5%  
Vnom  
Vnom  
-0.5%  
-5%  
-5%  
10  
50  
100  
10  
50  
100  
Output Load [%]  
Output Load [%]  
Deviation vs. Load  
REM2-xx3.3S(D)  
REM2-xx05S(D)  
20  
15  
10  
5
20  
15  
10  
5
053.3S  
053.3D  
153.3S  
153.3D  
0505S  
1205S  
1205D  
2405S  
2405D  
0
0
-5  
-5  
-10  
-10  
0
10 20 30 40 50 60 70 80 90 100  
0
10 20 30 40 50 60 70 80 90 100  
Output Load [%]  
Output Load [%]  
continued on next page  
REV.: 1/2019  
www.recom-power.com  
MED-3  
REM2  
DC/DC Converter  
Specifications (measured @ Ta= 25°C, nom. Vin, full load unless otherwise stated)  
Series  
Deviation vs. Load  
REM2-3.312S(D)  
20  
15  
10  
5
3.312S  
3.312D  
0
-5  
-10  
0
10 20 30 40 50 60 70 80 90 100  
Output Load [%]  
PROTECTIONS  
Parameter  
Type  
Value  
5.2kVDC  
tested for 1 minute  
I/P to O/P  
Isolation Voltage(5)  
rated for 1 minute  
4kVAC  
Isolation Resistance  
Isolation Capacitance  
Insulation Grade  
Means of Protection  
Medical Device Classification  
Internal  
10Gmin.  
25pF typ.  
reinforced  
250VAC working voltage  
2MOPP  
built-in power supply  
>8.1mm  
clearance / creepage  
clearance / creepage  
>9.0mm  
External  
Notes:  
Note5: For repeat Hi-Pot testing, reduce the time and/or the test voltage  
Note6: Refer to local safety regulations if input over-current protection is also required. Recommended fuse: slow blow type  
ENVIRONMENTAL  
Parameter  
Condition  
Value  
Operating Temperature Range  
Maximum Case Temperature  
Temperature Coefficient  
Thermal Impedance  
Operating Altitude  
full load @ natural convection 0.1m/s (see graph)  
-40°C to +80°C  
+105°C  
0.02%/ꢀ  
0.1m/s, horizontal  
non-condensing  
40ꢀ/W  
5000m  
5% - 95% RH max.  
Operating Humidity  
Pollution Degree  
PD2  
Vibration  
according to MIL-STD-202G standard  
+25°C  
according to MIL-HDBꢀ-217F, G.B.  
+80°C  
12900 x 103 hours  
5300 x 103 hours  
MTBF  
continued on next page  
REV.: 1/2019  
www.recom-power.com  
MED-4  
REM2  
DC/DC Converter  
Specifications (measured @ Ta= 25°C, nom. Vin, full load unless otherwise stated)  
Series  
Derating Graph  
(@ Chamber and natural convection 0.1m/s)  
100  
90  
80  
70  
60  
50  
40  
30  
20  
10  
0
-40 -30 -20 -10  
0
10 20 30 40 50 60 70 80 90 100 110  
95  
Ambient Temperature [°C]  
SAFETY AND CERTIFICATIONS  
Certificate Type (Safety)  
Report / File Number  
Standard  
Medical Electric Equipment, General Requirements for Safety and Essential  
Performance  
CAN/CSA-C22.2 No. 60601-1:14, 3rd Edition, 2014  
ANSI/AAMI ES60601-1 + A2:2010/®2012  
E314885  
Medical Electric Equipment, General Requirements for Safety and Essential  
Performance (CB scheme)  
CB190124C12 001  
WD-SE-R-190048-A0  
WD-SE-R-180528-B0  
IEC60601-1:2005, 3rd Edition + AM1:2012  
EN60601-1:2006 + A1:2013  
Medical Electric Equipment, General Requirements for Safety and Essential  
Performance  
Audio/Video, Information and Communication Technology Equipment -  
Part1: Safety Requirements  
IEC62368-1:2014, 2nd Edition  
EN62368-1:2014 + AC:2015  
RoHS2+  
RoHS-2011/65/EU + AM-2015/863  
EMC Compliance  
Condition  
Standard / Criterion  
Medical Electrical Equipment Part 1-2: Electromagnetic Disturbances –  
Requirements and Tests  
EN60601-1-2:2015  
WH-CE-E18083002  
Industrial, Scientific and Medical Equipment - Radio Frequency  
Disturbance Characteristics - Limits and Methods of Measurement  
EN55011:2016 + A1:2017, Class B  
EN55032, Class B  
Information Technology Equipment - Radio Disturbance  
Characteristics - Limits and Methods of Measurement  
with external filter  
Air 2, 4, 8, 15kV  
Contact 2, 4, 6, 8kV  
ESD Electrostatic Discharge Immunity Test  
IEC61000-4-2, Criteria A  
IEC61000-4-3, Criteria A  
10V/m (80-2700MHz)  
Radiated, Radio-Frequency, Electromagnetic Field Immunity Test  
9, 27, 28V/m (several frequencies)  
DC Power Port: 0.5, 1, 2kV  
DC Power Port: 0.5, 1kV  
DC Power Port: 3, 6V  
Fast Transient and Burst Immunity  
IEC61000-4-4, Criteria A  
IEC61000-4-5, Criteria B  
IEC61000-4-6, Criteria A  
IEC61000-4-8, Criteria A  
Surge Immunity  
Immunity to Conducted Disturbances, Induced by Radio-Frequency Fields  
Power Magnetic Field Immunity  
50Hz, 30A/m  
continued on next page  
www.recom-power.com  
REV.: 1/2019  
MED-5  
REM2  
DC/DC Converter  
Specifications (measured @ Ta= 25°C, nom. Vin, full load unless otherwise stated)  
Series  
EMC Filtering Suggestions according to EN55032  
Single Output  
Dual Output  
C2  
C2  
L1  
L1  
+VIN  
-VIN  
+VOUT  
+Vin  
-Vin  
+Vout  
+Vin  
+Vin  
+Vout  
+Vout  
C1  
C1  
Com  
Com  
-Vin  
-Vout  
-Vin  
-Vout  
-VOUT  
-Vout  
Component List Class B  
Input Voltage  
3.3VDC  
5VDC  
C1  
C2  
L1  
4.7µF  
10µF  
4.7µF  
47µH  
22µH  
47µH  
12VDC  
470pF / 6kVDC  
15VDC  
10µF  
24VDC  
DIMENSION AND PHYSICAL CHARACTERISTICS  
Parameter  
Type  
Value  
black plastic, (UL94V-0)  
silicone, (UL94V-0)  
FR4, (UL94V-0)  
case  
potting  
PCB  
Material  
Dimension (LxWxH)  
Weight  
23.0 x 8.0 x 12.2mm  
4.4g typ.  
Dimension Drawing (mm)  
embossed logo  
23.0  
Pinning information  
Pin #  
Single  
Dual  
+Vin  
-Vin  
-Vout  
Com  
+Vout  
8.0  
1
2
6
7
8
+Vin  
-Vin  
-Vout  
+Vout  
No Pin  
Marking  
Tolerance: xx.x= ±±.ꢀmm  
xx.xx= ±±.2ꢀmm  
0.10  
0.10  
0.75  
0.25  
Recommended Footprint Details  
Bottom View  
1 2  
6 7 8  
Top View  
1 2  
6 7 8  
2.60  
4x2.54= 10.16  
2.54  
www.recom-power.com  
REV.: 1/2019  
MED-6  
REM2  
DC/DC Converter  
Specifications (measured @ Ta= 25°C, nom. Vin, full load unless otherwise stated)  
Series  
PACKAGING INFORMATION  
Parameter  
Packaging Dimension (LxWxH)  
Packaging Quantity  
Type  
tube  
tube  
Value  
520.0 x 22.1 x 10.2mm  
20pcs  
Storage Temperature Range  
Storage Humidity  
-55°C to +125°C  
95% RH max.  
non-condensing  
The product information and specifications may be subject to changes even without prior written notice.The product has been designed for various applications; its suitability lies in the responsibility of each customer. The  
products are not authorized for use in safety-critical applications without RECOM’s explicit written consent. A safety-critical application is an application where a failure may reasonably be expected to endanger or cause  
loss of life, inflict bodily harm or damage property. The applicant shall indemnify and hold harmless RECOM, its affiliated companies and its representatives against any damage claims in connection with the unauthorized  
use of RECOM products in such safety-critical applications.  
www.recom-power.com  
REV.: 1/2019  
MED-7  

相关型号:

REM2-xx12S

2 Watt SIP8 Single and Dual Output
RECOM

REM2-xx3.3D

2 Watt SIP8 Single and Dual Output
RECOM

REM2-xx3.3S

2 Watt SIP8 Single and Dual Output
RECOM

REM20-2405DW

20 Watt 4:1 Input 1.6 x 1 Single and Dual Output
RECOM

REM20-2405SW

20 Watt 4:1 Input 1.6 x 1 Single and Dual Output
RECOM

REM20-2412DW

20 Watt 4:1 Input 1.6 x 1 Single and Dual Output
RECOM

REM20-2412SW

20 Watt 4:1 Input 1.6 x 1 Single and Dual Output
RECOM

REM20-2415DW

20 Watt 4:1 Input 1.6 x 1 Single and Dual Output
RECOM

REM20-2415SW

20 Watt 4:1 Input 1.6 x 1 Single and Dual Output
RECOM

REM20-2424SW

20 Watt 4:1 Input 1.6 x 1 Single and Dual Output
RECOM

REM20-4805DW

20 Watt 4:1 Input 1.6 x 1 Single and Dual Output
RECOM

REM20-4805SW

20 Watt 4:1 Input 1.6 x 1 Single and Dual Output
RECOM