SNP-AF87-3 [SKYNET]
Medical & ITE Switching Adapter;型号: | SNP-AF87-3 |
厂家: | Skynet Electronic Copyright |
描述: | Medical & ITE Switching Adapter |
文件: | 总4页 (文件大小:602K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
Rated
Peak
SNP-AF8 Series
72W
100W
Medical & ITE
Switching Adapter
Features:
●ꢀ
Peak load (1.3 ~ 1.5 x rated current, Vo=rated for 5 sec)
Design for BF application
DoE Level VI Eco-design
●ꢀ
●ꢀ
●ꢀ
●ꢀ
●ꢀ
●ꢀ
With ITE and Medical safety
Safety class I / II and EMI class B
-20oC to +60oC operating temperature
5,000m operation altitude
Applications:
●ꢀ
For DoE EPS level VI energy saving application, such
as electronics and office equipment.
For EMI class B application, such as home healthcare
●ꢀ
device, monitors and other medical devices.
For peak load application, such as motor drive, coffee
72 x 145 x 42 (mm)
●ꢀ
machine, vending machine, gaming machine, monitors
and other industrials.
General Specifications:
Input voltage ......................................... 90 VAC to 264 VAC
Input frequency ............................................. 47 Hz to 63 Hz
Inrush current ............................................ < 80A at 230VAC
Average efficiency ............ > 89% at 25%, 50%, 75%, 100%
of rated load and 115Vac/230Vac input
No-load input power ................................................ < 0.15W
Hold up time ......................................................16ms typical
Over load/Short citcuit protection ................... auto recovery
Over voltage protection ............................................ latch of
Operating temperature ..................................... -20oC to 60oC
derating: 2.5% / oC > 40oC
Storage temperature ..................................... -20oC to +85oC
EMI ................................................................ FCC class "B"
CISPR 22 level "B"
Harmonics............................................EN61000-3-2, class A
EMS...................................... EN61000-4-2,-3,-4,-5,-6,-8,-11
Safety ............................................. UL60601-1, UL60950-1
CSA C22.2 No. 60950-1(cUL)
TUV EN60601-1, EN60950-1
EC Code of Conduct
Energy Saving ........................................
for EPS Version (Tier 2)
Mechanical Specifications:
Notes:
1.
Size:
72 x 145 x 42 (mm)
Output Connectors:
2.
145.0 (5.71")
PIN VOLTAGE
1
3
4
2
+V
RTN
RTN
+V
EARTH
72.0
(2.83")
3.
Input Socket:
C14
4.
5.
6.
7.
8.
Cable length is 1.5m approx.
Power ON indication LED is on top of Box.
Box Color is Black.
This product is RoHS compliant.
Packing:
Net weight: 470 g approx. / unit
42.0
(1.65")
Gross weight: 12 kg approx. / carton, 20 units / carton
Carton size (mm): 503 (L) x 362 (W) x 300 (H) (mm)
-Jim-
10 years Warranty (contact Skynet's Distributors for details)
** This data sheet is only for models selection. For business, engineering specification by model must be used.
NOV. 2016
4-1
Rated
Peak
SNP-AF8 Series
72W
100W
Medical & ITE
Switching Adapter
Output Specifications:
MODEL
NO.
OUTPUT
RAIL
LOAD
INITIAL
STEP EFFICIENCY
AVERAGE
ACCURACY
EFFICIENCY.
MIN. RATED MAX. PEAK
@ 25% LOAD @ 50% LOAD @75% LOAD @ 100% LOAD
SNP-AF87-3
SNP-AF87-H
+12V
+24V
0A
0A
6A
8.4A
4.6A
+11.4V~+12.6V
89%
89%
89%
89%
89%
89%
89%
89%
> 89%
> 89%
SNP-AF89-3
SNP-AF89-H
3.2A
+22.8V~+25.2V
Note:
1. Output Load:
Rated 72W~77W for convection cooling.
2. At peak load, the output can last for 5 seconds without shut down.
3. At factory, in 60% rated load condition, each output is checked to be within voltage accuracy.
4. Line regulation is defined by changing ±10% of input voltage from nominal line at rated load.
5. Load regulation is defined by changing ±40% of measured output load from 60% rated load.
6. Ripple & noise is measured by using 20MHz bandwidth limited oscilloscope and terminated each output with a 0.47uF capacitor at rated load and nominal
line.
7. Hold up time is measured from the end of the last charging pulse to the time which the main output drops down to low limit of main output at rated load
and nominal line.
8. Efficiency is measured at rated load, and nominal line.
9. SNP-AF8x-H is for Home Healthcare.
-Jim-
** This data sheet is only for models selection. For business, engineering specification by model must be used.
4-2
NOV. 2016
Rated
Peak
SNP-AF8 Series
72W
100W
Medical & ITE
Switching Adapter
Performance for SNP-AF87-3:
1. Switching frequency ripple
2. Line frequency ripple
+12V/6A
+12V/6A
100mV/div, 5us/div
100mV/div, 5ms/div
3. Output turn on wave form
4. Output turn off wave form
+12V/6A
+12V/6A
5.0V/div, 10ms/div
5.0V/div, 10ms/div
5. Hold-up time
6. Over voltage protection
Trip point around 14.5V (60% Rated load)
5.0V/div
100V/div
Primary DC
10ms/div
5.0V/div, 10ms/div
-Jim-
** This data sheet is only for models selection. For business, engineering specification by model must be used.
NOV. 2016
4-3
Rated
Peak
SNP-AF8 Series
72W
100W
Medical & ITE
Switching Adapter
7. +12V step response
8. FCC B
+12V
0.5V/div, 2ms/div
+12V step from 1.2A to 6A
9. EN 55022 B
10. Power derating curve
60oC
-20oC
40oC
-Jim-
** This data sheet is only for models selection. For business, engineering specification by model must be used.
4-4
NOV. 2016
相关型号: