GB40S18K01 [SLPOWER]
AC-DC Regulated Power Supply Module,;型号: | GB40S18K01 |
厂家: | SL Power Electronics |
描述: | AC-DC Regulated Power Supply Module, |
文件: | 总4页 (文件大小:14263K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
40W Single Output
Medical & Industrial Grade
GB40 Family
FEATURES AND BENEFITS
Medical
Approved to CSA/EN/IEC/UL60601-1, 3rd Edition
40W Open Frame and PCB-mount Power Supply
2.0” x 4.0” x 1.0” Package
E-cap Life of >8 Years
>1,000,000 Hours MTBF
3 Year Warranty
Universal Input 90-264VAC
<0.1W No Load Input Power
Approved to CSA/EN/IEC/UL62368-1
Test &
Measurement
Meets Class B Radiated & Conducted EMI, with
Margin
Industrial
Meets Heavy Industrial and IEC60601-1-2 4th
Edition Levels of EMC
Note: *Consult Factory for compliance information.
MODEL SELECTION
Output
Power
Rated
Current
Ripple &
Noise1
Line
Regulation
Load
Regulation
Input
Class/Termination
Output
Termination
Model Number2
Volts
Class I (Grounded) input,
3-pin AMP/Molex type
connector
Change “K” to “C” for
class II input
Change “K” to “P” for PCB
mount pins, class I input
GB40S05K01
GB40S09K01
GB40S12K01
GB40S18K01
GB40S24K01
GB40S48K01
75mV pk-pk
90mV pk-pk
120mV pk-pk
120mV pk-pk
240mV pk-pk
480mV pk-pk
5.0V
9.0V
5.0A
4.0A
25W
±1%
±1%
±1%
±1%
±5%
±5%
±5%
±5%
4-pin AMP/Molex type
connector for “K” and “C”
versions
36W
3.4A
2.22A
1.7A
40W
40W
12.0V
18.0V
PCB mount pins for
“P” and “V” versions
24.0V
48.0V
40W
40W
±1%
±1%
±5%
±5%
Change “K” to “PCB mount
pins”, class II input
0.83A
Note: 1. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uF & 47uF parallel capacitor.
2. Other output voltages available, consult factory.
3. All specifications are typical at 230VAC, full load, at 25°C ambient unless noted.
INPUT
OUTPUT
Input Voltage and
Turn On Time
Hold-Up Time
<700ms
100-240VAC, ±10%, 47-63Hz, 1Ø
Frequency
20ms/100VAC at full load
Input Current
Inrush Current
Input Fuses
115VAC: 1.2A, 230VAC: 0.6A
25W-40W continuous – See models chart for specific
voltage model ratings
Output Power
Output Voltage
264VAC, cold start: will not exceed 40A peak
3.15A, 250VAC fuse in both line and neutral
See models chart
Earth Leakage Current
(Input to Ground)
<500μA@264VAC, 60Hz, NC
<1mA@264VAC, 60Hz, SFC
500μs resp. time for return to w/in 0.5% of final value
for any 50% load step from 5% to 100% of rated load,
∆i/∆t<0.2A/μs
Transient Response
Earth Leakage Current
(Output to Ground)
<100μA@264VAC, 60Hz, NC
<500μA@264VAC, 60Hz, SFC
Max voltage deviation is +/-3.5%
Regulation
See models chart
Efficiency
>88%, typical
Note: All specifications are typical at 230VAC input, full load, at 25°C ambient unless noted.
0.9, min., 230VAC, 80-100% load vector, 25°C
ambient
Power Factor
Note: All specifications are typical at 230VAC input, full load, at 25°C ambient unless noted.
GB40 Datasheet v0818
Copyright © 2018 SL Power Electronics Corp. - All rights reserved.
Page 1
40W Single Output
Medical & Industrial Grade
GB40 Family
PROTECTION
SAFETY
Will shutdown upon an overtemperature condition,
Auto-recovery
Overtemperature
ITE/Industrial Safety
Medical Safety
EN/IEC/UL62368-1
Protection
EN/IEC/UL60601-1, 3rd edition
130% to 160% of rated output current value,
Hiccup mode
Overload Protection
Hiccup mode
Short Circuit Protection
Overvoltage Protection
120% to 150% of nominal output voltage,
Hiccup Mode
EMI/EMC COMPLIANCE
RELIABILITY
EN55032, EN55011/CISPR11 Class B, FCC Part
15.107, Class B: 6db margin type, at 115 and 230VAC
>1,000,000 hours, full load, 110 & 220VAC input,
25°C amb., per telcordia 332 issue 6, stress method
MTBF
Conducted Emissions
>8 year life based on calculations at 115VAC/60Hz
& 230VAC/50Hz, ambient 25°C at 24 hrs per day,
365 days/year, 6 power up cycles per day
EN55032, EN55011/CISPR11 Class B, FCC Part
Radiated Emissions
E-cap Life
15.109, Class B: 3db margin type, at 115 and 230VAC
Electro-Static Discharge EN55024/IEC61000-4-2, Level 4: +/- 8kV contact,
(ESD) Immunity on
Power Ports
+/- 15kV air, Criteria A
IEC60601-1-2, 4th edition, Table 4
ISOLATION SPECIFICATIONS
EN55022/EN61000-4-3, 10V/m, 80MHz-2.7GHz,
80% AM at 1kHz
Radiated RF EM Fields
Susceptibility
Input-Output
Input-Ground : 1500VAC (1 MOPP)
Output-Ground : 1500VAC (1 MOPP)
: 4000VAC (2 MOPP)
IEC60601-1-2, 4th edition, Table 4
Isolation
EN55024/IEC61000-4-4, Level 4, +/- 4.4kV,
100Khz rep rate, 40A, Criteria A
Electrical Fast
Transients (EFT)/Bursts
IEC60601-1-2, 4th edition, Table 5
I/P-O/P, I/P-FG, O/P-FG: TBD
Isolation Resistance
Surges, Line to Line
(Diff Mode) and Line to
GND (CMN Mode)
EN55024/IEC61000-4-5, Level 4, +/-2kV DM,
+/-4kV CM, Criteria A
Surpasses IEC60601-1-2, 4th edition requirements
ENVIRONMENT
EN55022/IEC61000-4-6, 3.6V/m – Level 4, (0.15 to
-25 ~ +70°C, see derating curve for operation above
60°C
Conducted Disturbances 80Mhz; and 12V/m) in ISM and amateur radio bands
Induced by RF Fields
Operating Temperature
StorageTemperature
between 0.15Mhz and 80Mhz, 80% AM at 1KHz
IEC60601-1-2, 4th edition, Table 5
-40°C ~ +85°C
Rated Power Frequency EN55024/IEC1000-4-8, Level 4: 30 A/m, 50/60Hz
Convection
Cooling
Magnetic Fields
IEC60601-1-2, 4th edition, Table 4
Relative Humidity
5% to 90%, Non-condensing
EN55024/IECEN61000-4-11:
Operating: 0.003g/Hz, 1.5grms overall, 3 axes,
10 min/axis, 1-500Hz
Non-Oper.: random waveform, 3 minutes per axis, 3
axes and sine waveform, vib. frequency/acceleration:
10-500Hz/1g, sweep rate of 1 octave/minutes,
vibration time of 10 sweeps/axes, 3 axes
--100% dip for 10ms, at 0, 45, 90, 135, 180, 225,
270 and 315 degrees, 100% dip for 20ms, 0 deg.,
Criteria A
Voltage Interruptions,
Dips, Sags & Surges
Vibration
--100% dip for 5000 ms (250/300 cycles), Criteria B
--60% dip for 100ms, Criteria B
--30% dip for 500ms, Criteria A
IEC60601-1-2, 4th edition, Table 5
Operating: Half-sine, 20gpk, 10ms, 3 axes, 6 shocks
total
Non-Operating: Half-sine waveform, impact
acceleration of 50G, pulse duration of 6ms,
Number of shocks: 3 for each of the three axis
Harmonic Current
Emissions
EN55011/EN61000-3-2, Class A
Shock
Flicker Test
EN61000-3-3
48.3 x 101.6 x 25mm
Note: 1. The power supply is considered a component which will be installed into a final equipment.
The final equipment must be re-confirmed that it still meets EMC directives.
2.0 x 4.0 x 0.97 inch (Height is from top of tallest
component to max. component pin protrusion on
the bottom)
Dimensions
Weight
2. All specifications are typical at nominal input, full load, at 25°C ambient unless noted. Consult
factory for information regarding testing for or usage under special environments.
110g
Note: Same dimensions for PCB & Pin Variants.
GB40 Datasheet v0818
Copyright © 2018 SL Power Electronics Corp. - All rights reserved.
Page 2
40W Single Output
Medical & Industrial Grade
GB40 Family
DERATING CURVE
120
100
80
60
40
20
0
-20
-10
0
10
20
30
40
50
60
70
80
Ambient Temperature (°C)
MECHANICAL DRAWING
Mounting hole diameter: 3.2mm (0.125”) 4 places
GB40 Datasheet v0818
Copyright © 2018 SL Power Electronics Corp. - All rights reserved.
Page 3
40W Single Output
Medical & Industrial Grade
GB40 Family
CONNECTOR AND TERMINATION INFORMATION
Input Connections
Output Connections
Version
Connector Pinout
Connector Type/Part No.
Connector Pinout
Connector Type/Part No.
Connector:
Connector:
Pin 1: +Vout
Pin 2: +Vout
Pin 3: -Vout
Pin 4: -Vout
Open Frame:
“K”, “C”
Pin 1: AC LINE
Pin 2: GND
Pin 3: AC NEUTRAL
TE/AMP P/N 640445-3
Mating Connector:
TE/AMP P/N 640250-3,
Pins= 770476-1
TE/AMP P/N 640445-4
Mating Connector:
TE/AMP P/N 640250-4,
Pins= 770476-1
Pin 4: +Vout
Pin 5: +Vout
Pin 6: -Vout
Pin 7: -Vout
Pin 1: AC Line
Pin 2: GND
Pin 3: AC Neutral
PCB Mount:
“P”, “V”
Pencom PI3207 or equivalent
Pencom PI3207 or equivalent
GB40 Datasheet v0818
Copyright © 2018 SL Power Electronics Corp. - All rights reserved.
Page 4
相关型号: