GLOBAL PERFORMANCE SWITCHERS  
GLC75 Commercial/GLM75 Medical  
75 Watt Multiple Output Global Performance Switchers  
FEATURES:  
• Cost-effective multiple output power source  
• Universal input 90-264 Vac  
• 7.00”x 4.25”x 1.30” (Meets 1U height)  
• 2-year warranty  
• Conducted EMI exceeds FCC Class B and CISPR 22  
Class B (Commercial models) and CISPR 11 Class B  
(Medical models)  
• Complies with EN61000-3-2 Class A  
• Also available in single output versions  
• Commercial UL1950 3rd Edition, CSA22.2 No. 950 and  
EN60950-1 approvals  
• Medical Approved to UL2601-1,  
CSA22.2 No. 601.1-M90, and IEC/EN60601-1  
• RoHS Compliant Model Available (G suffix)  
SPECIFICATIONS  
Efficiency  
Ac Input  
68ꢀ -78ꢀ depending on model and load distribution.  
90-264 Vac, 47-63 Hz single phase.  
Input Protection  
Input Current  
Internal ac fuse provided on all units. Designed to blow only if a catastrophic  
failure occurs in the unit.  
Maximum input current at 120 Vac, 60 Hz with full rated output load not to  
exceed 2.3 A.  
Inrush Current  
Output Power  
Inrush limited by internal thermistors. Inrush at 240 Vac, averaged over the first ac  
half-cycle under cold start conditions will not exceed 37 A.  
Normal continuous output power is 75 W for unrestricted natural convection  
cooling or 110 W with 26 cfm airflow. During peak load conditions output  
regulation may exceed total regulation and noise limits.  
Hold Up Time  
20 ms minimum from loss of ac input power at full load, nominal line (120 Vac).  
Output Regulation  
Measured by 40ꢀ load change from 60ꢀ rated load with all other outputs  
at 60ꢀ rated load and input voltage change from minimum to maximum rat-  
ings. Output #1 requires 1A minimum load for proper regulation of other  
outputs. Initial set tolerance is measured with all outputs at 60ꢀ of full rated  
load. Output #2 requires 0.5A minimum load for proper regulation.  
Temperature Coefficient  
0.03ꢀ/°C typical on all outputs.  
Power Fail  
A standard TTL or CMOS compatible output goes low (< 0.5 V) 5ms before output  
voltage drops more than 4ꢀ below nominal voltage upon loss of ac power. Signal  
is factory set to trip on 84 to 94 Vac brown-out depending upon incoming line  
impedance and distortion. Other settings are available through adjustment of  
built-in potentiometer (consult factory for assistance). Output will stay low for 20  
ms minimum.  
Overload Protection  
Factory set to begin power limiting at approximately 120 W. Fully protected  
against short circuit and output overload. Short circuit protection is cycling  
type power limit.  
Output Noise  
EMI/EMC Compliance  
All models include built-in EMI filtering to meet the following emissions requirements:  
0.5ꢀ rms, 1ꢀ pk-pk, 20 MHz bandwidth, differential mode. Measured with  
noise probe directly across output terminals of the power supply.  
EMI SPECIFICATIONS  
COMPLIANCE LEVEL  
Transient Response  
Main Output: 500μSec typical response time for return to within 0.5ꢀ of final  
value for a 50ꢀ load step change, ∆i/∆t<0.2A/μSec. Maximum voltage devia-  
tion is 3.5ꢀ.  
Conducted Emissions GLC75  
Conducted Emissions GLM75  
Static Discharge  
RF Field Susceptibility  
Fast Transients/Bursts  
Surge Susceptibility  
EN55022 Class B; FCC Class B  
EN55011 Class B: FCC Class B  
EN61000-4-2, 6 kV contact, 8 kV air  
EN61000-4-3, 3 V/meter  
EN61000-4-4, 2 kV, 5 kHz  
EN61000-4-5, 1 kV diff., 2 kV com.  
EN61000-3-2 Class A  
Overvoltage Protection  
Built in on V1 with firing point set per table. OVP firing reduces output #1 and  
#2 to less than 50ꢀ of nominal voltage in 50 ms.  
Line Frequency Harmonics  
Voltage Adjust  
Commercial Safety  
All GLC models are approved to UL1950 3rd Edition, CSA22.2 No. 950, and  
EN60950-1.  
Factory set on standard unit; however, optional potentiometer (“-Vsuffix)  
adjusts voltage from 4.7 V to OVP point (6.2 V nominal) on the +5 V output.  
Note: Output #1 must not be more than 1ꢀ below nominal to achieve full  
output voltage range on Output #2. Output regulation limits in some models  
may be exceeded when the main output is adjusted beyond +1ꢀ of nominal  
voltage. High voltage settings may degrade the reliability of the unit due to  
excessive power dissipation in some outputs.  
Medical Leakage Current  
70 μA 264 V @ 50 Hz (normal conditions).  
Medical Safety  
GLM models are approved to UL2601-1, CSA22.2 No. 601.1M90, IEC/EN60601-1. CB  
Report available.  
All specifications are typical at nominal input, full load at 25OC unless otherwise stated  
Commercial Medical  
RoHS  
Suffix*  
Output  
No.  
Output  
Minimum  
Output  
Output  
Output  
Peak  
Model  
Model  
Output  
Maximum (A) Maximum (B)  
V1 OVP Set  
6.2 0.6 V  
Noise P-P  
Regulation  
GLC75A  
GLM75A  
G
G
G
G
G
G
G
G
G
1
2
3
4
+5.1 V  
+12 V  
-12 V  
1 A  
0.5 A  
0 A  
8 A  
2.5 A  
1 A  
10 A  
3 A  
1 A  
3 A  
12 A  
4 A  
1.2 A  
4 A  
50 mV  
120 mV  
120 mV  
120 mV  
2ꢀ  
+10ꢀ,-5ꢀ (D)  
3ꢀ  
+12 V  
0 A  
2.5 A  
2ꢀ  
GLC75B  
GLC75C  
GLC75D  
GLC75E  
GLC75F  
GLC75H  
GLC75J  
GLC75P  
GLM75B  
GLM75C  
GLM75D  
GLM75E  
GLM75F  
GLM75H  
GLM75J  
GLM75P  
1
2
3
4
+5.1 V  
+12 V  
-5 V  
1 A  
0.5 A  
0 A  
8 A  
2.5 A  
1 A  
10 A  
3 A  
1 A  
3 A  
12 A  
4 A  
1.2 A  
4 A  
6.2 0.6 V  
6.2 0.6 V  
6.2 0.6 V  
6.2 0.6 V  
6.2 0.6 V  
6.2 0.6 V  
6.2 0.6 V  
6.2 0.6 V  
50 mV  
120 mV  
50 mV  
2ꢀ  
+10ꢀ,-5ꢀ (D)  
3ꢀ  
+12 V  
0 A  
2.5 A  
120 mV  
2ꢀ  
1
2
3
4
+5.1 V  
+12 V  
-15 V  
1 A  
0.5 A  
0 A  
8 A  
2.5 A  
1 A  
10 A  
3 A  
1 A  
3 A  
12 A  
4 A  
1.2 A  
4 A  
50 mV  
120 mV  
150 mV  
150 mV  
2ꢀ  
+10ꢀ,-5ꢀ (D)  
3ꢀ  
+15 V  
0 A  
2.5 A  
2ꢀ  
1
2
3
4
+5.1 V  
+24 V  
-12 V  
1 A  
0.5 A  
0 A  
8 A  
2.5 A  
1 A  
10 A  
2.5 A  
1 A  
12 A  
3.5 A  
1.2 A  
4 A  
50 mV  
240 mv  
120 mV  
120 mV  
2ꢀ  
+10ꢀ,-5ꢀ (D)  
3ꢀ  
+12 V  
0 A  
2.5 A  
3 A  
2ꢀ  
1
2
3
4
+5.1 V  
+24 V  
-15 V  
1 A  
0.5 A  
0 A  
8 A  
2.5 A  
1 A  
10 A  
2.5 A  
1 A  
12 A  
3.5 A  
1.2 A  
4 A  
50 mV  
240 mV  
150 mV  
150 mV  
2ꢀ  
+10ꢀ,-5ꢀ (D)  
3ꢀ  
+15 V  
0 A  
2.5 A  
3 A  
2ꢀ  
1
2
3
4
+5.1 V  
+15 V  
-5 V  
1 A  
0.5 A  
0
8 A  
2.5 A  
1 A  
10 A  
3 A  
1 A  
3 A  
12 A  
4 A  
1.2 A  
4 A  
50 mV  
150 mV  
50 mV  
2ꢀ  
+10ꢀ,-5ꢀ (D)  
3ꢀ  
-15 V  
0
2.5 A  
150 mV  
2ꢀ  
1
2
3
4
+5.1 V  
+15 V  
-15 V  
1A  
0.5 A  
0
8 A  
2.5 A  
1 A  
10 A  
3 A  
1 A  
3 A  
12 A  
4 A  
1.2 A  
4 A  
50 mV  
150 mV  
150 mV  
150 mV  
2ꢀ  
+10ꢀ,-5ꢀ (D)  
3ꢀ  
+15 V  
0
2.5 A  
2ꢀ  
1
2
3
+5.1 V  
+12 V  
-12 V  
5 V  
1 A  
0.5 A  
0
8 A  
2.5 A  
1 A  
10 A  
3 A  
1 A  
3 A  
12 A  
4 A  
1.2 A  
4 A  
50 mV  
120 mV  
120 mV  
50 mV  
2ꢀ  
+10ꢀ,-5ꢀ (D)  
3ꢀ  
4(C)  
0
2.0 A  
2ꢀ  
1
2
3
4
+5.1 V  
+24 V  
-12 V  
1 A  
0.5 A  
0 A  
8 A  
4 A  
1 A  
10 A  
4 A  
1 A  
3 A  
12 A  
4.5 A  
1.2 A  
4 A  
50 mV  
240 mv  
120 mV  
120 mV  
2ꢀ  
+10ꢀ,-5ꢀ (D)  
3ꢀ  
+12 V  
0 A  
2.5 A  
2ꢀ  
* Add “Gsuffix to part number for RoHS compliant model. Contact factory for availability.  
A. Rating with unrestricted convection cooling. Total power not to exceed 75 W.  
B. Rating with 26 cfm forced-air cooling. Total power not to exceed 110 W.  
C. Floating fourth output can be referenced as either positive or negative. Connect pin 12 to Return to provide a positive voltage at Pin 13. Connect pin 13 to Return to  
provide a negative voltage at Pin 12.  
D. To maintain these regulations conditions, the +5V current must be at least 1/5 of V2 and not greater than 5 times the V2 current. Requires +5V to be adjusted to within  
1ꢀ with at least a 1 A load to maintain regulation on this output.  
GLC75/GLM75 - MULTIPLE OUTPUT - MECHANICAL SPECIFICATIONS  
ENVIRONMENTAL SPECIFICATIONS OPERATING  
NON-OPERATING  
-40 to +85°C  
0 to 95ꢀ RH  
40 gpk  
A. Units should be allowed to warm up/operate under non-condensing  
conditions before application of power. Derate output current and total  
output power by 2.5ꢀ per °C above 50°C.  
Temperature (A)  
Humidity (A)  
Shock (B)  
0 to 50°  
0 to 95ꢀ RH  
20 gpk  
B. Shock testing—half-sinusoidal, 10 3 ms duration, direction, 3  
orthogonal axes, total 6 shocks.  
Altitude  
-500 to 10,000 ft  
1.5 g rms0.003 g2/Hz 5 g rms0.026 g2/Hz  
-500 to 40,000 ft  
C. Random vibration—10 to 2000Hz, 6dB/octave roll-off from 350 to  
2000Hz, 3 orthogonal axes. Tested for 10 min./axis operating and 1  
hr./axis non-operating.  
Vibration (C)  
SL Power Electronics, Inc., 6050 King Drive, Bldg. A, Ventura, CA 93003, USA. Phone:(805) 486 4565 Fax:(805) 487 8911 www.slpower.com Rev. 9-23-10.  
Data Sheet © 2010 SL Power Electronics, Inc. The information and specifications contained in this data sheet are believed to be correct at time of publication.  
However, SL Power accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.  

相关型号:

GLM75C

75 Watt Multiple Output Global Performance Switchers
SLPOWER

GLM75D

AC-DC Regulated Power Supply Module, 75W
SLPOWER

GLM75E

75 Watt Multiple Output Global Performance Switchers
SLPOWER

GLM75F

75 Watt Multiple Output Global Performance Switchers
SLPOWER

GLM75H

75 Watt Multiple Output Global Performance Switchers
SLPOWER

GLM75J

75 Watt Multiple Output Global Performance Switchers
SLPOWER

GLM75P

75 Watt Multiple Output Global Performance Switchers
SLPOWER

GLMC1R003A

10 A Synchronous DC-DC Step down Regulator
PANASONIC

GLMHR1103A

General Purpose Inductor, 0.11uH, 20%, 1 Element, SMD, CHIP, 2626
ALPS

GLMHR2303A

General Purpose Inductor, 0.23uH, 20%, 1 Element, SMD, CHIP, 2626
ALPS

GLMR47KAT1A

Ultra-Broadband SMT Inductor
KYOCERA AVX

GLMR47KAT1A_13

Ultra-Broadband SMT Inductor
KYOCERA AVX