ME30A2403B01 [SLPOWER]
20 to 30W Single Output External Power;型号: | ME30A2403B01 |
厂家: | SL Power Electronics |
描述: | 20 to 30W Single Output External Power |
文件: | 总5页 (文件大小:614K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
ME30
20 to 30W Single Output External Power
Medical Series
Features
.
Meets UL/EN/IEC60601-1-2, 4th edition for
EMC*
.
.
.
Approved to EN/IEC/UL60601-1, 3rd edition
2 MOPP input-output isolation
Meets DoE Efficiency Level VI Requirements
.
.
No load input power
Average Efficiency
.
.
Up to 30W of AC-DC Power
Universal Input 90-264Vac Input Range
.
Desktop and Wall-Plug versions
.
Meets EN55011/CISPR11, FCC Part 15.109
Class B Conducted & Radiated Emissions, with
>6db margin
.
.
.
.
E-cap life of >8 years
>1,000,000 hours MTBF
IP22 Rated Enclosure
3 Year Warranty
Description
A high performance AC to DC external power supply family designed for medical applications. The ME30A Medical Series
low power external AC-DC power supplies are approved to safety EN/IEC/UL60601-1, 3rd edition with isolation levels
which satisfy the 2 MOPP requirements and designed to UL/EN/IEC60601-1-2, 4th edition for EMC*. The ME30A Series
models will operate at universal input range of 90 to 264Vac over the wide temperature range of -20°C to +50°C,
delivering full rated output power up to +40°C and applicable output power derating at 50°C. These models are available
in desktop and wall-plug versions, include an IP22 rating per IEC60529 for the enclosure, and output cable terminated at
a variety of output connectors.
*Consult Factory for Table 9 compliance information.
Model Selection
Model
Number
Output
Current
4.00A
Output
Power
20W
Ripple &
Noise¹
75mV pk-pk
Line
Regulation
±1%
Load
Regulation
±5%
Overvoltage
Trip Range
Output
Connector
Input
Configuration
Volts
5.0V
ME30A0503F01
5.75V-7.75V
11.7V-16.2V
14.4V-16.8V
18.0V-21.0V
21.6V-25.2V
28.8V-33.6V
55.0V-60.0V
5.75V-7.75V
11.7V-16.2V
14.4V-16.8V
18.0V-21.0V
21.6V-25.2V
28.8V-33.6V
55.0V-60.0V
ME30A0903F01
ME30A1203F01
ME30A1503F01
ME30A1803F01
ME30A2403F01
ME30A4803F01
ME30A0503N01
ME30A0903N01
ME30A1203N01
ME30A1503N01
ME30A1803N01
ME30A2403N01
ME30A4803N01
9.0V
3.00A
2.50A
2.00A
1.67A
1.33A
0.63A
4.00A
3.00A
2.50A
2.00A
1.67A
1.33A
0.63A
27W
30W
30W
30W
30W
30W
20W
27W
30W
30W
30W
30W
30W
90mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
240mV pk-pk
480mV pk-pk
75mV pk-pk
90mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
240mV pk-pk
480mV pk-pk
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
2.5 x 5.5 x
9.5mm
Straight Barrel
Type, center
positive
Class I
Desktop,
IEC60320 C14
Receptacle
12.0V
15.0V
18.0V
24.0V
48.0V
5.0V
9.0V
2.5 x 5.5 x
9.5mm
Straight Barrel
Type, center
positive
Class II
Desktop,
IEC60320 C8
Receptacle
12.0V
15.0V
18.0V
24.0V
48.0V
www.slpower.com
ME30 Rev 1.9
25-APR-16
1
ME30
20 to 30W Single Output External Power
Medical Series
Model Selection (continued)
Model
Number
Output
Current
4.00A
Output
Power
20W
Ripple &
Noise¹
75mV pk-pk
Line
Regulation
±1%
Load
Regulation
±5%
Overvoltage
Trip Range
Output
Connector
Input
Configuration
Volts
5.0V
ME30A0503Q01
5.75V-7.75V
11.7V-16.2V
14.4V-16.8V
18.0V-21.0V
21.6V-25.2V
28.8V-33.6V
55.0V-60.0V
5.75V-7.75V
11.7V-16.2V
14.4V-16.8V
18.0V-21.0V
21.6V-25.2V
28.8V-33.6V
55.0V-60.0V
5.75V-7.75V
11.7V-16.2V
14.4V-16.8V
18.0V-21.0V
21.6V-25.2V
28.8V-33.6V
55.0V-60.0V
ME30A0903Q01
ME30A1203Q01
ME30A1503Q01
ME30A1803Q01
ME30A2403Q01
ME30A4803Q01
ME30A0503B01
ME30A0903B01
ME30A1203B01
ME30A1503B01
ME30A1803B01
ME30A2403B01
ME30A4803B01
ME30A0503C01
ME30A0903C01
ME30A1203C01
ME30A1503C01
ME30A1803C01
ME30A2403C01
ME30A4803C01
9.0V
3.00A
2.50A
2.00A
1.67A
1.33A
0.63A
4.00A
3.00A
2.50A
2.00A
1.67A
1.33A
0.63A
4.00A
3.00A
2.50A
2.00A
1.67A
1.33A
0.63A
27W
30W
30W
30W
30W
30W
20W
27W
30W
30W
30W
30W
30W
20W
27W
30W
30W
30W
30W
30W
90mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
240mV pk-pk
480mV pk-pk
75mV pk-pk
90mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
240mV pk-pk
480mV pk-pk
75mV pk-pk
90mV pk-pk
120mV pk-pk
150mV pk-pk
180mV pk-pk
240mV pk-pk
480mV pk-pk
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
2.5 x 5.5 x
9.5mm
Straight Barrel
Type, center
positive
Class II
Desktop,
IEC60320 C18
Receptacle
12.0V
15.0V
18.0V
24.0V
48.0V
5.0V
9.0V
Class II Wall-
Plug,
Interchangeabl
e Blades (North
American Blade
included)²
2.5 x 5.5 x
9.5mm
Straight Barrel
Type, center
positive
12.0V
15.0V
18.0V
24.0V
48.0V
5.0V
9.0V
2.5 x 5.5 x
9.5mm
Straight Barrel
Type, center
positive
Class II Wall-
Plug, Fixed
North American
Blades³
12.0V
15.0V
18.0V
24.0V
48.0V
Notes:
1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors. For 5V and
6V models, values listed are typical, 100mV pk-pk maximum with 0.1μF ceramic and 47μF low ESR capacitors used at measurement point.
2. Order blade kit KT-1027K for other blades (EU. UK, Australia)
3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”.
4. All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
5. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (ME30B1203F01).
General Specifications
Less than 700mS @115Vac, full load
20mS min., at full Load, 100Vac input
AC Input 100-240Vac, ±10%, 47-63Hz, 1∅
Turn On Time
Hold-up Time
115Vac: 1.2A, 230Vac:0.6A
Input Current
Inrush Current
Input Fuses
Overtemperature
Protection
Will shutdown upon an overtemperature
condition, auto-recovery.
264Vac, cold start: will not exceed 40A
F1, F2: 2.0A, 250Vac fuses (line & neutral
lines) provided on all models
130 to 180% of rating, Hiccup Mode
Hiccup Mode, auto recovery.
Overload Protection
Earth Leakage
Current
Short Circuit
Protection
Overvoltage
Protection
Input-GND: <500μA@264Vac, 60Hz, NC
Output-GND: <4mA@264Vac, 60Hz, NC
Hiccup mode, see models chart for trip
ranges.
>87%, typical
Efficiency
Input-Output: 2 MOPP
Input-Ground: 1 MOPP
Output-Ground: 1 MOPP
20 to 30W continuous – See models chart for
specific voltage model ratings.
Output Power
Isolation
No Load Input
Power
<0.1W per DoE Efficiency Level VI
Requirements
EN/IEC/UL60601-1-1, 3rd edition
Safety Standards
-20°C to +50°C
Start Up at -40°C, full load, (warmup period before
all parameters are within published specifications).
Operating
Temperature
See models chart on pg 1.
Ripple and Noise
Output Voltage
Temperature
Derating
Derate output power above 40°C to TBD at
50°C
See models chart on pg 1.
500µs response time for return to within 0.5%
Transient of final value for any 50% load step over the
-40°C to +85°C
Storage Temperature
range of 5% to 100% of rated load, ∆i/∆t<
0.2A/µs. Max. voltage deviation is +/-3.5%.
Response
Operating: to 5000m.
Non-operating: -500 to 40,000 ft.
See models chart on pg 1.
Regulation
Drop Test
Altitude
1.4m from table top to wooden platform, 6
faces.
5% to 95%, non-condensing
Relative Humidity
www.slpower.com
ME30 Rev 1.9
25-APR-16
2
ME30
20 to 30W Single Output External Power
Medical Series
General Specifications (continued)
Operating: 0.003g/Hz, 1.5grms overall, 3
axes, 10 min/axis, 1-500Hz.
Operating: Half-sine, 20gpk, 10mS, 3 axes,
6 shocks total
Non-Operating: Half-sine waveform, impact
acceleration of 100G, Pulse duration of 6
mS, Number of shocks: 3 for each of the
three axis
Non-Oper.: random waveform, 3 minutes per
axis, 3 axes and Sine waveform, Vib.
frequency/acceleration: 10-500Hz/1g, sweep
rate of 1 octave / minutes, Vibration time of 10
sweeps / axes, 3 axes
Vibration
Shock
>1,000,000 hours, full load, 110 & 220Vac
MTBF
See outline drawings
250g
Dimensions
Weight
input, 25°C amb., per Telcordia 332 Issue 6.
>8 year life based on calculations at
115Vac/60Hz & 230Vac/50Hz, ambient 25°C
at 24 hrs per day, 365 days/year, 6 power up
E-Cap Life
cycles per day.
All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
EMI/EMC Compliance
Conducted Emissions:
Radiated Emissions:
Common Mode Noise:
EN55011/CISPR11 Class B, FCC Part 15.107, Class B: >6db margin typ, at 115 and 230Vac
EN55022/CISPR11 Class B, FCC Part 15.109, Class B: >3db margin typ, at 115 and 230Vac
High Frequency (100kHz-20MHz): <40mA pk-pk
Electro-Static Discharge (ESD)
Immunity on Power ports:
EN55024/IEC61000-4-2, Level 4: +/- 8kV contact, +/- 15kV air, Criteria A
IEC60601-1-2, 4th Edition, Table 4
EN55022/EN61000-4-3, 10V/m, 80MHz-2.7GHz, 80% AM at 1kHz
IEC60601-1-2, 4th Edition, Table 4
Radiated RF EM Fields Susceptibility
Electrical Fast Transients (EFT) /Bursts:
EN55024/IEC61000-4-4, Level 4, +/- 4kV, 100Khz rep rate, 40A, Criteria A
IEC60601-1-2, 4th Edition, Table 5
Surges, Line to Line (Diff Mode) and Line
to GND (CMN Mode)
EN55024/IEC61000-4-5, Level 4, +/-2kV DM, +/-4kV CM, Criteria A
Surpasses IEC60601-1-2, 4th Edition requirements.
EN55022/IEC61000-4-6, 3.6V/m – Level 4, 0.15 to 80Mhz; and 12V/m) in ISM and amateur
radio bands between 0.15Mhz and 80Mhz, 80% AM at 1KHz
IEC60601-1-2, 4th Edition, Table 5.
Conducted Disturbances induced by RF
Fields
EN55024/IEC1000-4-8, Level 4: 30A/m, 50/60 Hz
IEC60601-1-2, 4th Edition, Table 4
Rated Power frequency magnetic fields
EN55024/IECEN61000-4-11: --100% dip for 10 mS, at 0, 45, 90, 135, 180, 225,
270 and 315 degrees, 100% dip for 20mS, 0 deg., Criteria A
--100% dip for 5000mS (250/300 cycles), Criteria B
-- 60% dip for 100mS, Criteria B
Voltage Interruptions, Dips, Sags &
Surges
-- 30% dip for 500mS, Criteria A
IEC60601-1-2, 4th Edition, Table 5
EN55011/EN61000-3-2, Class A
EN61000-3-3
Harmonic Current Emissions
Flicker Test
All specifications are typical at nominal input, full load, at 25°C ambient unless noted. Consult factory for information regarding testing for or usage under special
environments.
www.slpower.com
ME30 Rev 1.9
25-APR-16
3
ME30
20 to 30W Single Output External Power
Medical Series
Output Power Derating
Output power is derated above 40°C as follows, for operation over the entire AC input range (90-264Vac).
ͦ
Mechanical Drawing
54
33.6
55
IEC60320 C14 Receptacle, 2.5 x 5.5 x 9.5mm Barrel Connector
Interchangeable N.A. Blade, 2.5 x 5.5 x 9.5mm barrel connector
Notes:
1. All dimensions in mm.
2. Interchangeable blade models come with North American blade fitted. For other blades (EU, UK, Aust.) order blade kit KT1027K.
3. Cable length for 5V models is 1200 +/- 50mm.
www.slpower.com
ME30 Rev 1.9
25-APR-16
4
ME30
20 to 30W Single Output External Power
Medical Series
Connector Information
Standard models include a 2.5 x 5.5 x 9.5mm straight barrel type connector (Ault #3), center positive. Other standard options are listed below.
The “03” in the standard model number is replaced by the applicable digits below:
Data Sheet © 2016 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. However, SL Power accepts no responsibility for
consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
www.slpower.com
ME30 Rev 1.9
25-APR-16
5
相关型号: