ME40A0503N01 [SLPOWER]
40W Single Output External Power;型号: | ME40A0503N01 |
厂家: | SL Power Electronics |
描述: | 40W Single Output External Power |
文件: | 总4页 (文件大小:483K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
ME40
40W Single Output External Power
Medical Series
Features
Meets UL/EN/IEC60601-1-2, 4th edition for EMC*
Approved to EN/IEC/UL60601-1, 3rd edition
2 MOPP Input to Output Isolation
.
.
.
.
Meets DoE Efficiency Level VI Requirements
.
.
No load input power
Average Efficiency
.
.
Up to 40W of AC-DC Power
Universal Input 90-264Vac Input Range
.
Desktop and Wall-Plug versions
.
Meets EN55011/CISPR11, FCC Part 15.109 Class B
Conducted & Radiated Emissions, with 6db margin
.
.
.
.
IP22 Rated Enclosure
E-cap life of >8 years
>1,000,000 hours MTBF
3 Year Warranty
Description
A high performance AC to DC external power supply family designed for medical applications. The ME40A Medical Series
low power external AC-DC power supplies are approved to safety EN/IEC/UL60601-1, 3rd edition with isolation levels
which satisfy the 2 MOPP requirements and designed to UL/EN/IEC60601-1-2, 4th edition for EMC*. The ME40A Series
models will operate at universal input range of 90 to 264Vac over the wide temperature range of -20°C to +70°C,
delivering full rated output power up to +40°C and applicable output power derating at 70°C. These models are available
in desktop and wall-plug versions, include an IP22 rating per IEC60529 for the enclosure, and output cable terminated at
a variety of output connectors.
*Consult Factory for Table 9 compliance information.
Model Selection
Overvoltage
Trip Range
Output
Connector
Input
Configuration
Model
Output
Current
Output
Power
Ripple &
Noise¹
Line
Regulation
Load
Regulation
Number
Volts
ME40A0503F01
ME40A1203F01
ME40A1803F01
5.0V
5.00A
3.40A
2.22A
25W
40W
40W
100mV pk-pk
120mV pk-pk
180mV pk-pk
120%-150%
120%-150%
120%-150%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class I Desktop,
IEC60320 C14
Receptacle
12.0V
18.0V
ME40A2403F01
ME40A0503N01
ME40A1203N01
ME40A1803N01
ME40A2403N01
ME40A0503Q01
ME40A1203Q01
ME40A1803Q01
ME40A2403Q01
ME40A0503B01
ME40A1203B01
ME40A1803B01
ME40A2403B01
ME40A0503C01
ME40A1203C01
ME40A1803C01
ME40A2403C01
24.0V
5.0V
1.70A
5.00A
3.40A
2.22A
1.70A
5.00A
3.40A
2.22A
1.70A
5.00A
3.40A
2.22A
1.70A
5.00A
3.40A
2.22A
1.70A
40W
25W
40W
40W
40W
25W
40W
40W
40W
25W
40W
40W
40W
25W
40W
40W
40W
240mV pk-pk
100mV pk-pk
120mV pk-pk
180mV pk-pk
240mV pk-pk
100mV pk-pk
120mV pk-pk
180mV pk-pk
240mV pk-pk
100mV pk-pk
120mV pk-pk
180mV pk-pk
240mV pk-pk
100mV pk-pk
120mV pk-pk
180mV pk-pk
240mV pk-pk
120%-140%
120%-150%
120%-150%
120%-150%
120%-140%
120%-150%
120%-150%
120%-150%
120%-140%
120%-150%
120%-150%
120%-150%
120%-140%
120%-150%
120%-150%
120%-150%
120%-140%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C8
Receptacle
12.0V
18.0V
24.0V
5.0V
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C18
Receptacle
12.0V
18.0V
24.0V
5.0V
Class II Wall-Plug,
Interchangeable
Blades²
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
12.0V
18.0V
24.0V
5.0V
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Wall-Plug,
Fixed North
American Blades³
12.0V
18.0V
24.0V
Notes:
1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors.
2. Standard models are fitted with North American blades. Order blade kit KT-1027K for other blades (EU. UK, Australia)
3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”.
4. All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
5. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (TE40B1203F01).
www.slpower.com
ME40 Rev 2.3 30-SEP-16
1
ME40
40W Single Output External Power
Medical Series
General Specifications
Less than 700mS @115Vac, full load
20mS min., at full Load, 100Vac input
AC Input 100-240Vac, ±10%, 47-63Hz, 1∅
Turn On Time
Hold-up Time
115Vac: 1.2A, 230Vac: 01.6A
Input Current
Inrush Current
Input Fuses
Overtemperature
Protection
Will shutdown upon an over-temperature
condition, auto-recovery.
264Vac, cold start: will not exceed 40A
F1, F2: 2.0A, 250Vac fuses (line & neutral
lines) provided on all models
130 to 180% of rating, Hiccup Mode
Hiccup Mode, auto recovery.
Overload Protection
Earth Leakage
Current
Short Circuit
Protection
Input-GND: <500μA@264Vac, 60Hz, NC
Output-GND: <4mA@264Vac, 60Hz, NC
Overvoltage
Protection
Hiccup mode. See models chart for trip
ranges.
>87%, typical
Efficiency
Input-Output: 2 MOPP
Input-Ground: 1 MOPP
Output-Ground: 1 MOPP
40W continuous – See models chart for
specific voltage model ratings.
Output Power
Isolation
No Load Input
Power
<0.1W per DoE Efficiency Level VI
Requirements
EN/IEC/UL60601-1, 3rd edition and
-20°C to +70°C
Safety Standards
Operating
Temperature
See models chart on pg 1.
Ripple and Noise
Output Voltage
Temperature
Derating
See models chart on pg 1.
See derating chart
500µs response time for return to within 0.5%
Transient of final value for any 50% load step over the
-40°C to +85°C
Storage Temperature
range of 5% to 100% of rated load, ∆i/∆t<
0.2A/µs. Max. voltage deviation is +/-3.5%.
Response
Operating: to 5000m.
Non-operating: -500 to 40,000 ft.
See models chart on pg 1.
Regulation
Altitude
1.4m from table top to wooden platform, 6
faces.
5% to 95%, non-condensing
Drop Test
Vibration
Relative Humidity
Operating: 0.003g/Hz, 1.5grms overall, 3
axes, 10 min/axis, 1-500Hz.
Operating: Half-sine, 20gpk, 10mS, 3 axes,
6 shocks total
Non-Operating: Half-sine waveform, impact
acceleration of 100G, Pulse duration of 6
mS, Number of shocks: 3 for each of the
three axis
Non-Oper.: random waveform, 3 minutes per
axis, 3 axes and Sine waveform, Vib.
frequency/acceleration: 10-500Hz/1g, sweep
rate of 1 octave / minutes, Vibration time of 10
sweeps / axes, 3 axes
Shock
>1,000,000 hours, full load, 110 & 220Vac
input, 25°C amb., per Telcordia 332 Issue 6.
See outline drawings
Dimensions
Weight
MTBF
>8 year life based on calculations at
115Vac/60Hz & 230Vac/50Hz, ambient 25°C
at 24 hrs per day, 365 days/year, 6 power up
cycles per day.
250g
E-Cap Life
All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
www.slpower.com
ME40 Rev 2.3 30-SEP-16
2
ME40
40W Single Output External Power
Medical Series
EMI/EMC Compliance
EN55011/CISPR22 Class B, FCC Part 15.107, Class B: 6db margin typ, at 115 and 230Vac
EN55022/CISPR22 Class B, FCC Part 15.109, Class B: 3db margin typ, at 115 and 230Vac
High Frequency (100kHz-20MHz): <40mA pk-pk
Conducted Emissions:
Radiated Emissions:
Common Mode Noise:
Electro-Static Discharge (ESD)
Immunity on Power ports:
EN55024/IEC61000-4-2, Level 4: +/- 8kV contact, +/- 15kV air, Criteria A
IEC60601-1-2, 4th Edition, Table 4
EN55022/EN61000-4-3, 10V/m, 80MHz-2.7GHz, 80% AM at 1kHz
IEC60601-1-2, 4th Edition, Table 4
Radiated RF EM Fields Susceptibility
Electrical Fast Transients (EFT) /Bursts:
EN55024/IEC61000-4-4, Level 4, +/- 4kV, 100Khz rep rate, 40A, Criteria A
IEC60601-1-2, 4th Edition, Table 5
Surges, Line to Line (Diff Mode) and Line
to GND (CMN Mode)
EN55024/IEC61000-4-5, Level 4, +/-2kV DM, +/-4kV CM, Criteria A
Surpasses IEC60601-1-2, 4th Edition requirements.
EN55022/IEC61000-4-6, 3.6V/m – Level 4, 0.15 to 80Mhz; and 12V/m) in ISM and amateur
radio bands between 0.15Mhz and 80Mhz, 80% AM at 1KHz
IEC60601-1-2, 4th Edition, Table 5.
Conducted Disturbances induced by RF
Fields
EN55024/IEC1000-4-8, Level 4: 30 A/m, 50/60 Hz
IEC60601-1-2, 4th Edition, Table 4
Rated Power frequency magnetic fields
EN55024/IECEN61000-4-11: --100% dip for 10 mS, at 0, 45, 90, 135, 180, 225,
270 and 315 degrees, 100% dip for 20mS, 0 deg., Criteria A
--100% dip for 5000mS (250/300 cycles), Criteria B
--60% dip for 100mS, Criteria B
Voltage Interruptions, Dips, Sags &
Surges
--30% dip for 500mS, Criteria A
IEC60601-1-2, 4th Edition, Table 5
EN55011/EN61000-3-2, Class A
EN61000-3-3
Harmonic Current Emissions
Flicker Test
All specifications are typical at nominal input, full load, at 25°C ambient unless noted. Consult factory for information regarding testing for or usage under special
environments.
Mechanical Drawing
54
33.6
55
IEC60320 C14 Receptacle, 2.5 x 5.5 x 9.5mm Barrel Connector
Interchangeable N.A. Blade, 2.5 x 5.5 x 9.5mm barrel connector
Notes:
1. All dimensions in mm.
2. Interchangeable blade models come with North American blade fitted. For other blades (EU, UK, Aust.) order blade kit KT1027K.
ME40 Rev 2.3 30-SEP-16
www.slpower.com
3
ME40
40W Single Output External Power
Medical Series
Derating Chart:
Connector Information
Standard models include a 2.5 x 5.5 x 9.5mm straight barrel type connector (Ault #3), center positive. Other standard options are listed below.
The “03” in the standard model number is replaced by the applicable digits below:
Data Sheet © 2016 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. However, SL Power accepts no responsibility for
consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
www.slpower.com
ME40 Rev 2.3 30-SEP-16
4
相关型号:
©2020 ICPDF网 联系我们和版权申明