MINT1175A5606K01 [SLPOWER]
the highest power density for a power supply;型号: | MINT1175A5606K01 |
厂家: | SL Power Electronics |
描述: | the highest power density for a power supply |
文件: | 总8页 (文件大小:1036K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
MINT1175
175W Single Output
Medical Series
Features
ꢀ
ꢀ
ꢀ
ꢀ
ꢀ
ꢀ
2” x 4” x 1.3” Package
For 1U Applications
Class I and Class II versions
175W w/air, 120W convection cooled
Universal Input 90-264 VAC
Average Efficiency meets Level V
Requirements
ꢀ
Approved to CSA/EN/IEC/UL60601-1, 3rd
Edition
ꢀ
ꢀ
ꢀ
ꢀ
ꢀ
2 x MOPP Isolation
12V fan output
Efficiency 90% typical at Full Load
3 Year Warranty
RoHS Compliant
Description
The MINT1175 Series models provide the highest power density for a power supply in a 2” x 4” size. Ideal for use in
Home Healthcare and Portable Medical Device Applications and more. With input & output monitoring alarms plus
isolated 12V/0.4A fan, it is one the most feature rich supplies and can easily fit in 1U chassis. Exceeding the medical
device strict requirements, the MINT1175 has the lowest earth leakage current with less than 200µA, satisfies 2MOPP
reinforced isolation requirements of IEC60601-1 3rd edition, and can be purchased as Class I or II configurations.
Model Selection
Model
Number
Output Current
w/200LFM air Convection
Ripple &
Noise**
Total
Regulation
OVP***
Threshold***
Volts
Fan Output
0.5%RMS,
1.2% pk-pk
MINT1175A1206K01
MINT1175A1506K01
MINT1175A2406K01
MINT1175A4806K01
MINT1175A5606K01
12V
14.6A
11.7A
7.3A
10.0A
8.0A
5.0A
2.5 A
2.1A
12Vdc/0.4A
3%
3%
3%
3%
3%
14.0 1.1V
18.5 1.5V
28.0 2.5V
55.0 4.0V
59.0 1.0V
0.5%RMS,
1.0% pk-pk
15V
24V
48V
56V
12Vdc/0.4A
12Vdc/0.4A
12Vdc/0.4A
12Vdc/0.4A
0.5%RMS,
1.2% pk-pk
0.5%RMS,
1.0% pk-pk
3.6A
0.5%RMS,
1.0% pk-pk
3.1A
Notes:
*
Total convection power is 120 Watts.
** Measured with noise probe directly across output terminals, and load terminated with 0.1µF ceramic and 10µF low ESR capacitors.
*** Output adjustment on 56V model will not exceed 56.2 volts.
General Specifications
Less than 2 sec. @115Vac (inversely
proportional to input voltage and thermistor
temperature)
100-240Vac, 10%, 47-63Hz, 1∅
120–370Vdc
AC Input
Turn On Time
Hold-up Time
115Vac: 2A, 230Vac: 1A
16mS, typical (120W output load)
Input Current
www.slpower.com
MINT1175 REV 1.4, 5-AUG -15
1
MINT1175
175W Single Output
Medical Series
General Specifications (continued)
264Vac, cold start: will not exceed 50A
AC Power Fail, DC OK, Inhibit
Inrush Current
Signals
Overload
Protection
F1, F2: 4A, 250VAC fuses provided on all
models
Hiccup Mode
Input Fuses
Earth Leakage
Current
Short Circuit
Protection
Provided - no damage will occur if the output is
shorted.
<200µA@264Vac, 60Hz, NC; <400µA SFC
Overvoltage
Protection
OVP firing reduces output voltage to <50% of
nominal in <50ms. See chart for trip range.
90% typical at 115Vac
Efficiency
175W continuous, with 200 lfm airflow, 120W
convection cooled – See chart for specific
voltage model ratings.
PFC: Fixed, 65kHz.
Main Converter: Variable 35-200kHz, 65-70kHz
at full load.
Switching
Frequency
Output Power
Input-Output: 4000Vac, 2 x MOPP
Input-Ground: 1800Vac, 1 x MOPP
Output-Ground: 1500Vac
Transient
Response
50% load step. ∆i/∆t: <0.2A/µS
Max Volt Deviation = 3%
Isolation
Operating
Temperature
-10°C to +70°C
Start Up at -40°C, full load
See chart
Ripple and Noise
Output Voltage
Temperature
Derating
Derate output power linearly above 50°C to 50%
at 70°C
See chart
Voltage
Adjustability
Storage
Temperature
+/-5% from nominal
Not required
-40°C to +85°C
Operating: -500 to 10,000 ft.
Non-operating: -500 to 40,000 ft.
Minimum Load
Altitude
+/- 3%. See chart
5% to 95%, non-condensing
Total Regulation
Relative Humidity
Operating: 0.003g²/Hz, 1.5grms overall, 3
axes, 10 min/axis
Non-Operating: 0.026g²/Hz, 5.0grms overall,
3 axes, 1 hr/axis
Operating: Half-sine, 20gpk, 10ms, 3 axes, 6
shocks total
Non-Operating: Half-sine, 40 gpk, 10 ms, 3
axes, 6 shocks total
Vibration
Shock
W: 2.0” x L: 4.0” x H: 1.3”
EN/CSA/UL/IEC 60601-1, 3rd Edition
Dimensions
Weight
Safety Standards
MTBF
210g
331,000 hours, 25°C Ambient, 110Vac input
Auxiliary Signals
During normal operation, stays HIGH. Signal
goes LOW with 5mS warning before loss of
DC output from AC failure.
Open collector logic signal goes and stays HIGH
100mS to 500mS after main output reaches
regulation.
AC Power Fail:
DC OK:
Connect to inhibit pin to output common to
inhibit the DC output.
Inhibit:
www.slpower.com
MINT1175 REV 1.4, 5-AUG -15
2
MINT1175
175W Single Output
Medical Series
EMI/EMC Compliance
EN55011/22 Class B, FCC Part 15, Subpart B, Class B
Conducted Emissions
EN55011/22 Class A; FCC Part 15, Subpart A, Class A w/6db margin
EN61000-4-2, Criteria A, 6kV Contact Discharge, 8kV air discharge
EN61000-4-3, 3V/m. Criteria A
Radiated Emissions
Static Discharge Immunity
Radiated RF Immunity
EFT/Burst Immunity
EN61000-4-4, 2kV/5kHz, Criteria A
EN61000-4-5, 1kV differential, 2kV common-mode, Criteria A
EN61000-4-6, 3Vrms, Criteria A
Line Surge Immunity
Conducted RF Immunity
Power Frequency Magnetic Field Immunity
Voltage Dip Immunity
Line Harmonic Emissions
Flicker Test
EN61000-4-8, 3A/m, Criteria A
EN61000-4-11, 0% Vin, 0.5cycle; 40% Vin, 5 cycles; 70% Vin, 25 cycles;
Criteria A
EN61000-3-2, Class A, B, C, & D
EN61000-3-3, Complies (dmax<6%)
Mechanical Drawing
When installing the power supply into the end-product, a non-conductive Insulator should be placed between the unit and
any conductive metal chassis or mounting platform.
www.slpower.com
MINT1175 REV 1.4, 5-AUG -15
3
MINT1175
175W Single Output
Medical Series
Connector Information
Input Connector
J100
Ground
J101
DC Output Connector
J300
Fan Output Connector
J500
Signal Connector
J600
PIN 1) AC LINE
PIN 2) EMPTY
PIN 3) AC NEUTRAL
PIN 1) +Vout PIN 4) -Vout
PIN 2) +Vout PIN 5) -Vout
PIN 3) +Vout PIN 6) -Vout
PIN 1) Inhibit
PIN 2) PF/DC OK
PIN 3, 4) Common
PIN 1) +12Vfan
PIN 2) RTN
.187 x .032 FASTON TAB
Mating Connector:
Molex 09-50-3031
Pins= 08-52-0072
Mating Connector:
AMP 640250-6
Pins = 640252-1
Mating Connector:
Molex 1375820-2
Pins = 1375819
Mating Connector:
Molex 1375820-4
Pins = 1375819
Mating Connector:
Molex 01-90020009
Fan Output – J500
J500 provides a 12V@0.4mA output to support a system cooling fan. The fan output is always available when AC input is
present, so it also can be used for a 12V standby output is so desired.
AC Power Failure/DC OK and Inhibit Signals – J600
J300–Pin 1
J300–
Pin 1
Timing Sequence
www.slpower.com
MINT1175 REV 1.4, 5-AUG -15
4
MINT1175
175W Single Output
Medical Series
Isolation Specifications
Parameter
Conditions/Description
Min
Nom
Max
Units
Input/Ground
Input/Output
Output/Ground
1 MOPP
2 MOPP
Operational
Insulation Safety Rating
Input/Ground
Input/Output
Output/Ground
1800
4000
1500
Vac
Vac
Vac
Electric Strength Test Voltage
-
-
Input Specifications
All specifications apply over specified input voltage, output load, and temperature range, unless otherwise noted.
Parameter
Conditions/Description
Min
Nom
Max
Units
Input Voltage
Turn-On Input Voltage
Turn-Off Input Voltage
Input Frequency
90
115/230
82.7
67.0
50/60
-
264
Vac
Vac
Vac
Hz
Ramping up
Ramping down
47
-
63
50
-
Inrush Current Limitation
Power Factor
264Vac, cold start
A
Vi nom, Io nom
0.9
-
Vi nom, Io nom
MINT1175A1206K01
MINT1175A1506K01
MINT1175A2406K01
MINT1175A4806K01
MINT1175A5606K01
Efficiency
90%
-
%
-
Output Specifications
Parameter
Conditions/Description
Min
Nom
Max
Units
Output Voltage Setpoint Accuracy
MINT1175A1206K01
Vi nom,, Io1 @ ADC, TC = 25 °C
-3
-
3
% Vo nom
Output Voltage Setpoint Accuracy
MINT1175A1506K01
Vi nom,, Io1 @ ADC, TC = 25 °C
Vi nom,, Io1 @ ADC, TC = 25 °C
Vi nom,, Io1 @ ADC, TC = 25 °C
Vi nom,, Io1 @ ADC, TC = 25 °C
MINT1175A1206K01
-3
-3
-3
-3
-
-
-
-
3
3
3
3
% Vo nom
% Vo nom
% Vo nom
% Vo nom
Output Voltage Setpoint Accuracy
MINT1175A2406K01
Output Voltage Setpoint Accuracy
MINT1175A4806K01
Output Voltage Setpoint Accuracy
MINT1175A5606K01
ADC
ADC
Output Current V1
Output Current V2
0
0
10.0
-
14.6
0.4
Output Current V1
Output Current V2
0
0
8.0
-
11.7
0.4
ADC
ADC
MINT1175A1506K01
Output Current V1
Output Current V2
0
0
5.0
-
7.3
0.4
ADC
ADC
MINT1175A2406K01
ADC
ADC
Output Current V1
Output Current V2
0
0
2.5
-
3.6
0.4
MINT1175A4806K01
ADC
ADC
Output Current V1
Output Current V2
0
0
2.1
-
3.1
0.4
MINT1175A5606K01
Static Line Regulation V1
Vi min-Vi max, Vi nom, 0-100% Io nom
Vi min-Vi max, Vi nom, 0-100% Io nom
-1
-3
-
-
1
3
% Vo nom
% Vo nom
Static Load Regulation V1
(Droop Characteristic)
www.slpower.com
MINT1175 REV 1.4, 5-AUG -15
5
MINT1175
175W Single Output
Medical Series
Output Specifications (continued)
Parameter
Conditions/Description
Min
Nom
Max
Units
Hold-Up Time
Starting at Vi = 230 VAC, Po nom
-
16
-
ms
Load change =50%, di/dt = 0.2A/µS
voltage deviation 3%
Dynamic Load Regulation
Start-Up Time
0
0
3
2
% Vo nom
Vi nom, Io nom
-
s
Protection
All specifications apply over specified input voltage, output load, and temperature range, unless otherwise noted.
Parameter
Input Fuse
Conditions/Description
Not user accessible
Min
Nom
Max
Units
Input Transient Protection
2KV(CM) and 1KV(DM) surge
2
KV (CM)
No-load and short circuit proof
short circuit proof
Hiccup or Latch
Hiccup or Latch
Hiccup
Output
overload (latch style)
Latch style
Overvoltage Protection
Latch
Over temperature Protection
Automatic power shutdown at TC =155°C
Characteristic Curves
Output vs. Temperature
120W convection cooled and 175W continuous with 200 LFM airflow. Derate output power to 50% at 70°C.
www.slpower.com
MINT1175 REV 1.4, 5-AUG -15
6
MINT1175
175W Single Output
Medical Series
Efficiency vs. Loading
The high efficiency is achieved by using LLC technology, PFC topology minimizing switching losses. Synchronous SCHOTTKY or ultra-fast diode is
used as rectifier in MINT1175 family because of high output voltage level.
Noise & Ripple
To verify that the output ripple and noise does not exceed the level specified in the product specification. Measured using a scope probe socket with
0.1µF ceramic and a 10µF electrolytic capacitor connected in parallel across it, BW limit with 20MHz.
www.slpower.com
MINT1175 REV 1.4, 5-AUG -15
7
MINT1175
175W Single Output
Medical Series
Output Transient Response
50% load step within the regulation limits of minimum and maximum load, di/dt<0.2A/µSec. Recovery time not specified as there is no laps in regulation
with a 50% Load Step. Maximum voltage deviation is 3%, This test is performed on the MAIN OUTPUT ONLY.
Output Overload Characteristic
Supply shall protect itself against Overload conditions. The Power Supply shall recover from Overload Conditions without operator intervention.
Output Overshoot
When supply is turned on, off or when the load is stepped 100%, the Power Supply Overshoot/Undershoot shall not exceed minimum or maximum of
output voltage regulation.
Data Sheet © 2012 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. However, SL Power accepts no responsibility for
consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
www.slpower.com
MINT1175 REV 1.4, 5-AUG -15
8
相关型号:
©2020 ICPDF网 联系我们和版权申明