DTM36CN135-C8 [TDK]
AC-DC电源;型号: | DTM36CN135-C8 |
厂家: | TDK ELECTRONICS |
描述: | AC-DC电源 |
文件: | 总5页 (文件大小:620K) |
中文: | 中文翻译 | 下载: | 下载PDF数据表文档文件 |
DTM36-C8 Series
https://product.tdk.com/en/power/dtm
www.emea.lambda.tdk.com/dtm
25 to 36W Medical External Power Supplies
The DTM36-C8 Class II 25 to 36W external power supplies are certified to IEC/ES/EN60601-1 and EN60601-1-11, are compliant to EN60601-1-2 (Ed4)
and designed to meet IEC60335-1 (Household Appliances). Requiring no earth ground, the DTM series is ideally suited for Medical, Clinical and Home
Healthcare applications. Alternative connectors and cable assemblies are available on request.
Features
Benefits
• Medical Safety Certifications and Immunity
• Class II Input
• Suitable for Medical, Clinical and Home Healthcare Use
• No Earth Ground Connection Needed
• Meets US and EU Energy Saving Legislation
• Fully Isolated For Use With B and BF Rated Equipment
• Flexible System Integration
• Meets DoE Level VI and EU Tier 2 Efficiency, <75mW Off-load Power Consumption
• 4kV (2xMOPP) Input to Output Isolation
• Alternative Connectors and Cable Assemblies
Model Selector
Output
Voltage
(V)
Maximum
Current
(A)
Maximum
Power
(W)
Overvoltage
Protection (V)
(Typical)
Maximum
Load
Capacitance (uF)
Efficiency (%)
(230Vac input,
75% Load)
Average
Efficiency
(%) (1)
Model
DTM36CN050-C8
DTM36CN090-C8
DTM36CN120-C8
DTM36CN135-C8
DTM36CN150-C8
DTM36CN180-C8
DTM36CN240-C8
DTM36CN360-C8
DTM36CN480-C8
5
9
5
3.3
2.5
2.4
2.4
2
1.5
1
0.75
25
29.7
30
32.4
36
36
36
36
36
7.44
12.6
15.9
18.5
21.5
24.8
31.5
45.2
59.6
5,000
3,300
2,500
2,400
2,400
2,000
1,500
1,000
750
85
88
89
89
89
89
89
89
89
85.0
87.7
87.7
87.9
88.3
88.3
88.3
88.3
88.3
12
13.5
15
18
24
36
48
DTM36CN
120
C8
-
xxx
Connector/Cable Options*
(When applicable)
Output voltage 5, 9, 12, 13.5, 15, 18, 24, 36, 48
IEC C8 input connector
* Options
When no option code is stated:
5V output model has a straight 5.5 x 2.1 x 12mm (OD x ID x L) connector, positive center pin and a 16 AWG 1220mm long cable
9V, 12V and 13.5V output models have a straight 5.5 x 2.1 x 12mm (OD x ID x L) connector, positive center pin and an 18 AWG 1800mm long cable
>15 models have a straight 5.5 x 2.1 x 12mm (OD x ID x L) connector, positive center pin and an 20 AWG 1800mm long cable
1
DTM36-C8 Series
Specifications
Model
DTM36-C8
Input
Input Voltage range (2)
Input Frequency
Vac
80 - 264
Hz
47 - 63
Input Current (110/230Vac)
Inrush Current at 240Vac (Cold Start)
Leakage Current
A
< 0.9
A
< 100
uA
< 80
Harmonic Compliance
No Load Power Consumption
Hold Up Time (typ) at 115Vac Input
Efficiency
-
Meets IEC61000-3-2
mW
< 75
ms
10
-
-
-
-
-
-
See model selector table
Average Efficiency
See model selector table
Conducted & Radiated EMI
Immunity
EN55011-B Conducted, EN55011-B Radiated, EN60601-1-2
See immunity table
Insulation Class
Class II
Safety Agency Certifications
IEC/ES/EN60601-1, EN60601-1-11, CE Mark (LVD, EMC and RoHS). Designed to meet IEC60335-1
Immunity
Test
ESD
Standard
EN61000-4-2
Test Level
Criteria
Notes
4
A
A
EN61000-4-3
EN60601-1-2 table 9
EN61000-4-4
3
80-2700MHz
Radiated Susceptibility
-
A
Pass
Electrical FastTransient Burst
Surge
3
A
(AC Port,100kHz)
EN61000-4-5
3
A
-
EN61000-4-6
3
-
A
10V (61000-6-2)
6V (60601-1-2)
-
Conducted Susceptibility
Magnetic fields
EN60601-1-2 table 6
EN61000-4-8
A
N/A
N/A
A
EN61000-4-11 Class 2
Industrial, including
EN60601-1-2
0% for 1/2 cycle
0% for 1 cycle
70% for 25/30 cycles
0% for 250/300 cycles
Voltage Dips and
Input Interruptions
A
A
(100Vac)
B
2
DTM36-C8 Series
Specifications
Model
DTM36-C8
Output
Output Voltage Tolerance
Output Voltage Adjustment
Switching Frequency (Typical)
Line Regulation
%
±2% at 60% load
-
kHz
%
None
65
±1% of rated output voltage from 100 to 240Vac
5V: ±6%, 9V: ±4%, 12-48V: ±2% of rated output voltage
See model selector table
Load Regulation (3)
External Load Capacitance
Ripple & Noise (4)
Temperature Coefficient
Minimum Load
%
uF
mV
%/°C
-
5V: 100, 9V & 12V: 120, 13.5V: 130, 15V: 150, 18V: 180, 24V: 240, 36V: 360, 48V: 480
±0.05
No minimum load required
Overcurrent Protection
Overvoltage Protection
Indicators
-
Hiccup mode with auto recovery
V
Diode clamp, auto recovery. See model selector table
LED indicates DC output is on
-
Parallel Operation
Environmental
Operating Temperature
Storage Temperature
Humidity (non condensing)
Cooling
-
Not possible
°C
-30 to 60, derate linearly from 100% to 60% load from 40 to 60
°C
-30 to +85
%RH
5 - 93%RH
-
m
Vac
MΩ
-
Convection
Altitude
5,000
Withstand Voltage (For 1 minute)
Isolation Resistance
IP Rating
Input to Output 4,000 (2xMOPP)
>100 at 25°C
IP21
Vibration (Operating)
Shock
-
Meets MIL-STD-810F Table 516.5, Table 516.5-1, 10ms, each axis 3 times (± X, Y, Z axis)
Meets MIL-STD-810F Table 514.5C-VIII, 15~200Hz, X, Y, Z axis, 1hr each axis, total 3hrs
-
Other
Weight (Typ)
g
150
100 x 45 x 22.5
Size (LxWxH)
mm
Size (LxWxH)
Inches
3.937 x 1.771 x 0.886
IEC 320-C8 (2 prong)
See model selector (options)
1,000,000 max
AC Input Connector
DC Output Connector
-
-
MTBF-MIL-HDBK-217F,GB,25°C,115Vacinput hrs
Warranty yrs
3
Notes
See website for detailed specifications, test methods and installation manual
(1) Measured at 25%, 50%, 75% and 100% load conditions
(2) Derate linearly from 100% to 75% load from 90Vac input to 80Vac input
(3) Measured from 60 - 100% and 60 - 20% load
(4) Ripple and noise is measured at 20MHz BW. Add 0.1uF ceramic capacitor and 10uF Electrolytic capacitor in parallel across the output
3
DTM36-C8 Series
Outline Drawing (Straight connector with 20AWG 1,800mm cable shown)
4
DTM36-C8 Series
For additional information, please visit
https://product.tdk.com/en/power/
DTM36-C8_Datasheet_260660 August 5th, 2020 v1.5
5
DTM36-C8 Series
相关型号: