2450-5Y-EW [TEKTRONIX]

2450-EC Graphical Potentiostat;
2450-5Y-EW
型号: 2450-5Y-EW
厂家: TEKTRONIX, INC.    TEKTRONIX, INC.
描述:

2450-EC Graphical Potentiostat

文件: 总15页 (文件大小:1262K)
中文:  中文翻译
下载:  下载PDF数据表文档文件
2450-EC Graphical Potentiostat  
Datasheet  
Key Features  
• Perform Cyclic, Squarewave, or Galvanic Voltammetry,  
Chronoamperometry, and Chronopotentiometry  
• Simplified user interface for faster test setup and  
analysis of results  
• Real-time plotting of voltammograms on the  
front panel  
• Analytical graph cursors for immediate analysis of  
results without the need for a PC  
• Create libraries of reusable, customizable  
experimental software with built-in scripting  
A Tektronix Company  
• Screen capture function allows copying test results  
from the display to reports  
The 2450-EC Graphical Potentiostat brings speed,  
flexibility, and simplicity right to your fingertips. Its  
• 10 nV and 10 fA measurement sensitivity  
innovative graphical user interface (GUI) and advanced,  
capacitive touchscreen technology allow intuitive usage  
and minimize the learning curve to enable researchers,  
scientists, and students to learn faster, work smarter,  
and invent easier. The 2450-EC is a versatile instrument,  
particularly well-suited for research and development in  
fundamental electrochemical lab research, characterizing  
the next generation of materials and electrolytes, new  
energy storage devices, and faster, smaller sensors.  
• Front panel input banana jacks; rear panel input  
triaxial connections  
• Context-sensitive help function minimizes learning  
• Front-panel USB 2.0 memory I/O port for transferring  
data, test scripts, or test configurations  
While potentiostats are excellent instruments for  
electrochemistry applications, they typically lack any front  
panel display and control knobs, often are 2-quadrant  
systems only, and must be completely controlled by a  
computer with software that is not always open for users  
to customize tests beyond what the software can do.  
The 2450-EC includes a wide range of voltages and  
currents for sourcing or measuring, nV / fA sensitivities,  
and high impedance sense leads with a typical input  
resistance of 50 G ohms and only 1pA of input bias  
current, typically acceptable with a wide variety of  
reference electrodes. The 2450-EC can run internal  
application test scripts so electrochemistry meaurements  
can be run without the use of an external computer.  
Results are immediately displayed right on the instrument  
front panel touchscreen. Connecting the 2450-EC to  
a 2-, 3-, or 4-electrode cell is simple with the included  
translation cable.  
Datasheet  
Electrochemical Cell  
Working  
Electrode  
Reference  
Electrode  
Counter  
Electrode  
The 2450-EC can be easily connected to a 3-electrode cell.  
Learn Faster; Work Smarter;  
Invent Easier  
Unlike traditional potentiostats that lack a user-interface  
front panel to interact with, the 2450-EC features a five-  
inch, full-color, high resolution touchscreen that facilitates  
ease of use, and optimizes overall speed and productivity.  
Built-in, context-sensitive help enables intuitive operation  
and minimizes the need to review a separate manual.  
These capabilities combined with its application versatility  
make the 2450-EC inherently easy to use for basic and  
advanced measurement applications, regardless of your  
experience level with electrochemistry instruments.  
2450-EC Cyclic Voltammetry Menu  
Convert Raw Data into Information  
A full graphical plotting window converts raw data and  
displays it immediately as useful information, such as  
cyclic voltammograms. The touch screen interface  
makes it easy to observe, interact with, and explore  
measurements with “zoom and pinch” simplicity. By using  
the built-in graphing cursors, you can immediately analyze  
your data without a computer. All graphic screens can be  
saved to a USB thumb drive for incorporation into reports  
and journals. Using the graphical sheet view, test data  
can also be displayed in tabular form. The instrument  
supports exporting data to a spreadsheet for further  
analysis, dramatically improving productivity for research  
and development. This combination of high performance  
and high ease of use offers unparalleled insight into your  
test results.  
2450-EC Cyclic Voltammetry Test Configuration Screen  
Graph view of results  
2
TEK.COM  
2450-EC Graphical Potentiostat  
Built-in real-time graphing, charting, scope-like cursors, and data display spreadsheet for export simplifies converting test results into  
useful information.  
Test Applications  
Typical Applications  
The 2450-EC’s built-in scripting language enables  
electrochemists, chemists, and materials scientists to  
create libraries of reusable, customizable experimental  
software for running tests including cyclic voltammetry,  
chronoamperometry, chronopotentiometry, and more. The  
following electrochemistry test scripts are loaded in the  
internal memory of the 2450-EC.  
Ideal for electrochemical research  
and development in a wide variety  
of applications studies, including:  
• Basic Analytical Research  
– Electrochemical cells  
– Electrode studies  
– Solid electrolytes  
Cyclic Voltammetry: Potential is swept at a user  
programmable scan rate between two to four  
defined vertices while current is measured. Current is  
measured using an analog integration method. This  
application can also be used to perform Linear Sweep  
Voltammetry.  
• Materials Research  
Electrode compositions  
– Electrolyte solutions  
– Ceramics, polymers, ferro/  
piezoelectrics  
Open Circuit Potential: Measures the cell potential  
difference between two electrodes with high input  
impedance as a function of time.  
– Organic semiconductors  
– Low-k dielectrics  
– Biomaterials  
Potential Pulse and Square Wave with Current  
Measure: The 2450-EC sources potential at  
programmable peak and base levels and the resulting  
current is recorded at the pulse peak level.  
– Nanomaterials  
– Electrodesposition  
• Energy Systems and Storage  
– Dye-sensitized solar cells  
– Batteries  
Current Pulse and Square Wave with Voltage  
Measure: The 2450-EC sources current at  
programmable peak and base levels and the resulting  
potential is recorded at the pulse peak level.  
– Fuel cells, flow batteries  
– Supercapacitors  
Chronoamperometry: The potential is stepped to  
a programmed value and the resulting current is  
measured as a function of time.  
• Sensors  
– Environmental monitoring  
– Industrial process control  
– Healthcare/medical  
Chronopotentiometry: The current is stepped to  
a programmed value and the resulting potential is  
measured as a function of time.  
In addition to pre-loaded test scripts, the built-in scripting  
language enables the user to create their own library of  
electrochemistry test scripts that can be modified as the  
test and measurements evolve.  
TEK.COM  
3
Datasheet  
All-in-One Instrument  
Quad. II  
Quad. I  
+1A  
The 2450-EC offers a highly flexible, four-quadrant voltage  
and current source/load coupled with precision voltage  
and current meters. When not used in potentiostat type  
applications, this all-in one instrument can be repurposed  
as a general lab instrument, including use as a:  
+100mA  
–200V  
–20V  
+20V  
+200V  
• Precision power supply with V and I readback  
• True current source  
–100mA  
–1A  
• Digital multimeter (DCV, DCI, ohms, and power with  
6½-digit resolution)  
• Precision electronic load  
• Trigger controller  
Quad. III  
Quad. IV  
2450-EC power envelope.  
Ease of Use Beyond the Touchscreen  
In addition to its five-inch, color touchscreen, the 2450-EC front panel has many features that supplement its speed, user-  
friendliness, and learnability, including a USB 2.0 memory I/O port, a HELP key, a rotary navigation/control knob, a front/  
rear input selector button, and banana jacks for basic bench applications. The USB 2.0 memory port supports easy data  
storing, saving instrument configurations, loading test scripts, and system upgrades. Plus, all front panel buttons are backlit  
to enhance visibility in low-light environments.  
5˝ color graphical touchscreen display  
Rotary  
navigation/  
control knob  
Online  
HELP key  
USB 2.0  
memory I/O  
Front/rear  
input selector  
2450-EC front panel with high-resolution, capacitive touchscreen.  
Comprehensive Built-in Connectivity  
Rear panel access to rear-input triax connectors, remote control interfaces (GPIB, USB 2.0, and LXI/Ethernet), D-sub 9-pin  
digital I/O port (for internal/external trigger signals and handler control), instrument interlock control, and TSP-Link® jacks  
enables easy configuration of multiple instrument test solutions and eliminates the need to invest in additional adapter  
accessories.  
4
TEK.COM  
2450-EC Graphical Potentiostat  
Triax inputs  
Interlock  
ernet  
USB Digital I/O TSP-Link  
GPIB  
Rear panel connections are optimized for signal integrity.  
• Native X-Y graphing, panning, and zooming  
Free Instrument Control  
Start-up Software  
• Spreadsheet/tabular viewing of data  
• Saving and exporting data for further analysis  
• Saving of test setups  
The 2450-EC can be repurposed for applications beyond  
electrochemistry as a general purpose lab tool, e.g.  
I-V testing, leakage testing, battery charge/discharge  
profiling, etc. KickStart, Keithley’s instrument control  
non-programming start-up software, lets users start  
taking measurements in minutes for typical current versus  
voltage applications. In most cases, users merely need to  
make quick measurements, graph the data, and store the  
data to disk to perform analysis in software environments  
such as Excel.  
• Screenshot capturing of graph  
• Annotation of tests  
• Command line dialog for sending and receiving data  
• HTML help  
• GPIB, USB 2.0, Ethernet compliant  
Simplified Programming with  
Ready-to-Use Instrument Drivers  
KickStart offers the following functionality:  
For those who prefer to create their own customized  
application software, native National Instruments  
LabVIEW® drivers, as well as IVI-C and IVI-COM drivers  
are available at www.tektronix.com.  
• Instrument configuration control to perform I-V  
characterization  
With KickStart start-up software, users are ready to take measurements in minutes.  
TEK.COM  
5
Datasheet  
Test Script Specifications  
2450-EC Cyclic Voltammetry Specifications  
Potential Range  
–20 V E +20 V  
Current Measure Ranges  
Source Limit (Compliance)  
Scan Rate  
10 μA, 100 μA, 1 mA, 10 mA, 100 mA, 1 A  
100% of selected current range  
0.1 mV/s to 3500 mV/s  
Potential Step Size During Scanning  
100 μV (0.1 mV/s scan rate < 35 mV/s)  
1 mV (35 mV/s scan rate < 350 mV/s )  
10 mV (350 mV/s scan rate 3500 mV/s)  
Number of Cycles  
1 to 100  
User Selectable Sampling Intervals Points/Test (10 to 10000)  
Points/Cycle (10 to 10000)  
Seconds/Point (0.01 to 100)  
Points/Second (0.01 to 100)  
Maximum Total Number of Samples 100,000  
2450-EC Open Circuit Potential Specifications  
Potential Measure Ranges  
20 mV, 200 mV, 2 V, 20 V  
0.75 s sample interval 100 s  
1 to 100,000  
Sample Interval  
Number of Samples  
2450-EC Potential Pulse and Square Wave Specifications  
Peak Potential  
–20 V Epeak +20 V  
Base Potential  
–20 V Ebase +20 V  
Current Measure Ranges  
Source Limit (Compliance)  
Pulse Period and Width  
Current Measure Range = 1 μA  
1 μA, 10 μA, 100 μA, 1 mA, 10 mA, 100 mA, 1 A  
105% of selected current range  
200 ms period 3600 s  
100 ms pulse width (0.99 × period)  
Current Measure Range > 1 μA  
4 ms period 3600 s  
2ms pulse width (0.99 × period)  
Sample Time  
Minimum is 166.667 μs.  
Maximum is lesser of 166.667 ms and (pulse width – 1 ms)  
Number of Cycles  
Program Time  
1 to 100,000  
(1 × period) program time (100,000 × period)  
6
TEK.COM  
2450-EC Graphical Potentiostat  
2450-EC Current Pulse and Square Wave Specifications  
Peak Current  
–1.05 A Ipeak +1.05 A  
–1.05 A Ibase +1.05 A  
20 mV, 200 mV, 2 V, 20 V  
105% of selected potential range  
Base Current  
Potential Measure Ranges  
Source Limit (Compliance)  
Pulse Period and Width  
|Ipeak| and |Ibase| 1.05 μA 200 ms period 3600 s  
100 ms pulse width (0.99 × period)  
|Ipeak| or |Ibase| > 1.05 μA  
4 ms period 3600 s  
2 ms pulse width (0.99 × period)  
Sample Time  
Minimum is 166.667 μs.  
Maximum is lesser of 166.667 ms and (pulse width – 1 ms)  
Number of Cycles  
Program Time  
1 to 100,000  
(1 × period) program time (100,000 × period)  
2450-EC Chronoamperometry Specifications  
Step Potential  
–20 V Estep +20 V  
Current Measure Ranges  
Source Limit (Compliance)  
Number of Steps  
Step Duration  
10 nA, 100 nA, 1 μA, 10 μA, 100 μA, 1 mA, 10 mA, 100 mA, 1 A  
100% of selected current range  
1 to 10  
10 ms step duration 99,999 s  
10 ms sample interval 100 s  
Minimum is 166.667 μs.  
Sample Interval  
Sample Time  
Maximum is lesser of 166.667 ms and (sample interval – 5 ms) and (step duration – 5 ms)  
Maximum Number of Samples  
100,000 total for all steps  
2450-EC Chronopotentiometry Specifications  
Step Current  
–1.05 A Istep +1.05 A  
20 mV, 200 mV, 2 V, 20 V  
100% of selected potential range  
1 to 10  
Potential Measure Ranges  
Source Limit (Compliance)  
Number of Steps  
Step Duration  
10 ms step duration 99,999 s  
10 ms sample interval 100 s  
Sample Interval  
Sample Time  
Minimum is 166.667 μs.  
Maximum is lesser of 166.667 ms and (sample interval – 5 ms) and (step duration – 5 ms)  
Maximum Number of Samples  
100,000 total for all steps  
TEK.COM  
7
Datasheet  
Specifications  
Voltage Specifications 1, 2  
Source  
Measure 3  
Accuracy (23° 5°C)  
1 Year  
(% setting + volts)  
Accuracy (23° 5°C)  
1 Year  
Noise (RMS)  
(<10 Hz)  
Input  
Range  
Resolution  
500 nV  
5 μV  
Resolution  
10 nV  
Resistance  
>10 GΩ  
>10 GΩ  
>10 GΩ  
>10 GΩ  
>10 GΩ  
(% rdg. + volts)  
20.00000 mV  
200.0000 mV  
2.000000 V  
20.00000 V  
200.0000 V  
0.100% + 200 μV  
0.015% + 200 μV  
0.020% + 300 μV  
0.015% + 2.4 mV  
0.015% + 24 mV  
1 μV  
1 μV  
0.100% + 150 μV  
0.012% + 200 μV  
0.012% + 300 μV  
0.015% + 1 mV  
0.015% + 10 mV  
100 nV  
1 μV  
50 μV  
10 μV  
100 μV  
1 mV  
500 μV  
5 mV  
10 μV  
100 μV  
Current Specifications 1, 2  
Source  
Measure3  
Accuracy (23° 5°C)4  
1 Year  
(% setting + amps)  
Accuracy (23° 5°C)  
1 Year  
Noise (RMS)  
(<10 Hz)  
Voltage  
Burden  
Range  
Resolution  
500 fA  
5 pA  
Resolution  
10 fA  
(% rdg. + amps)  
10.00000 nA5  
100.0000 nA5  
1.000000 μA  
10.00000 μA  
100.0000 μA  
1.000000mA  
10.00000mA  
100.0000mA  
1.000000 A  
0.100% + 100 pA  
0.060% + 150 pA  
0.025% + 400 pA  
0.025% + 1.5 nA  
0.020% + 15 nA  
0.020% + 150 nA  
0.020% + 1.5 μA  
0.025% + 15 μA  
0.067% + 900 μA  
500 fA  
500 fA  
5 pA  
<100 μV  
<100 μV  
<100 μV  
<100 μV  
<100 μV  
<100 μV  
<100 μV  
<100 μV  
<100 μV  
0.100% + 50 pA  
0.060% + 100 pA  
0.025% + 300 pA  
0.025% + 700 pA  
0.020% + 6 nA  
100 fA  
1 pA  
50 pA  
500 pA  
5 nA  
40 pA  
400 pA  
5 nA  
10 pA  
100 pA  
1 nA  
50 nA  
500 nA  
5 μA  
0.020% + 60 nA  
0.020% + 600 nA  
0.025% + 6 μA  
40 nA  
100 nA  
3 μA  
10 nA  
100 nA  
1 μA  
50 μA  
0.030% + 500 μA  
Temperature Coefficient (0°–18°C and 28°–50°C)  
(0.15 × accuracy specification)/°C.  
Notes  
1. Speed = 1 PLC.  
2. All specifications are guaranteed with output ON.  
3. Accuracies apply to 2- and 4-wire mode when properly zeroed.  
4. For sink mode, 1 μA to 100 mA range accuracy is (0.15% + offset × 4). For 1A range, accuracy is (1.5% + offset × 8).  
5. Rear panel triax connections only.  
8
TEK.COM  
2450-EC Graphical Potentiostat  
Resistance Measurement Accuracy (Local or Remote Sense) 1, 2  
Normal Accuracy  
(23°C 5°C)  
1 Year, (% rdg. + ohms)  
Enhanced Accuracy 4  
(23°C 5°C)  
1 Year, (% rdg. + ohms)  
Default  
Default  
Test Current  
Range  
Resolution 3  
<2.000000 Ω 5  
20.00000 Ω  
1 μΩ  
10 μΩ  
100 μΩ  
1 mΩ  
10 mΩ  
100 mΩ  
1 Ω  
User defined  
100 mA  
10 mA  
Source IACC + Meas. VACC  
0.098% + 0.003 Ω  
0.077% + 0.03 Ω  
0.066% + 0.3 Ω  
Meas. IACC + Meas. VACC  
0.073% + 0.001 Ω  
0.053% + 0.01 Ω  
0.045% + 0.1 Ω  
200.0000 Ω  
2.000000 kΩ  
20.00000 kΩ  
200.0000 kΩ  
2.000000 MΩ  
20.00000 MΩ  
200.0000 MΩ  
>200.0000 MΩ 5  
1 mA  
100 μA  
10 μA  
0.063% + 3 Ω  
0.043% + 1 Ω  
0.065% + 30 Ω  
0.046% + 10 Ω  
1 μA  
0.110% + 300 Ω  
0.110% + 1 kΩ  
0.049% + 100 Ω  
0.052% + 500 Ω  
0.349% + 5k Ω  
10 Ω  
1 μA  
100 Ω  
100 nA  
User defined  
0.655% + 10 kΩ  
Source IACC + Meas. VACC  
Meas. IACC + Meas. VACC  
Temperature Coefficient, (0°–18°C and 28°–50°C)  
(0.15 × accuracy specification)/°C.  
Source Current, Measure Resistance Mode  
Total uncertainty = Isource accuracy + Vmeasure accuracy (4-wire remote sense).  
Source Voltage, Measure Resistance Mode  
Total uncertainty = Vsource accuracy + Imeasure accuracy (4-wire remote sense).  
Guard Output Impedance  
0.5 Ω (DC) in ohms mode.  
Notes  
1. All specifications are guaranteed with output ON.  
2. Accuracies apply to 2- and 4-wire mode when properly zeroed.  
3. 6.5 digit measure resolution  
4. Source readback enabled. Offset compensation ON.  
5. Source Current, Measure Resistance or Source Voltage, Measure Resistance only. For best measurements, use rear panel triax connectors.  
Operating Characteristics  
Max. Output Power  
20 W, four-quadrant source or sink operation.  
Source Limits  
Vsource: 21 V (1 A range), 210 V (100 mA range)  
Isource: 1.05 A (20 V range), 105 mA (200 V range).  
Overrange  
Regulation  
105% of range, source and measure.  
Voltage: Line: 0.01% of range. Load: 0.01% of range + 100 μV.  
Current: Line: 0.01% of range. Load: 0.01% of range + 100 pA.  
Source Limits  
Voltage Source Current Limit: Bipolar current limit set with single value. Min. 10% of range.  
Current Source Voltage Limit: Bipolar voltage limit set with single value. Min. 10% of range.  
V-Limit / I-Limit Accuracy  
Overshoot  
Add 0.3% of setting and 0.02% of reading to base specification.  
Voltage Source: <0.1% typical (full scale step, resistive load, 20 V range, 10 mA I-Limit.  
Current Source: <0.1% typical (1 mA step, RLoad = 10 kΩ, 20 V range).  
Range Change Overshoot  
Output Settling Time  
Maximum Slew Rate  
Overshoot into a fully resistive 100 kΩ load, 10 Hz to 20 MHz BW, adjacent ranges: 250 mV typical  
Time required to reach 0.1% of final value, 20 V range, 100 mA I-Limit: <200 μs typical.  
0.2 V/μs, 200 V range, 100 mA limit into a 2 kΩ load (typical).  
User selectable values, 5% tolerance. Factory default = none.  
10 Hz–1 MHz (RMS): 2 mV typical into a resistive load.  
Over Voltage Protection  
Voltage Source Noise  
Common Mode Voltage  
250 V DC.  
TEK.COM  
9
Datasheet  
Common Mode Isolation  
Noise Rejection (typical)  
>1 GΩ, <1000 pF.  
NPLC  
0.01  
0.1  
NMRR  
CMRR  
60 dB  
60 dB  
1
60 dB  
100 dB*  
* Except lowest two current ranges ~90dB.  
Load Impedance  
20 nF typical (standard). Stable into 50 μF typical (High-C mode).  
High-C mode valid for 100 μA range s, 200 mV ranges.  
Max. Voltage Drop Between Force and Sense Terminals  
5 V.  
Max. Sense Lead Resistance  
Sense Input Impedance  
Guard Offset Voltage  
1 MΩ for rated accuracy.  
>10 GΩ.  
<300 μV, typical  
System Measurement Speeds 1  
Reading Rates (readings/second) typical for 60 Hz (50 Hz):  
Script (TSP) Programmed  
Measure  
Source-Measure Sweep  
To GPIB To USB  
NPLC/Trigger Origin  
0.01 / Internal  
0.01 / External  
0.1 / Internal  
To Mem.  
3130 (2800)  
2170 (2050)  
540 (460)  
500 (430)  
59 (49)  
To GPIB  
To USB  
To LAN  
To Mem.  
To LAN  
1620 (1540)  
1580 (1510)  
470 (400)  
460 (400)  
58 (48)  
2830 (2570) 2825 (2600) 2790 (2530) 1710 (1620) 1620 (1540) 1630 (1540)  
2150 (2030) 2170 (2040) 2160 (1990) 1670 (1590) 1580 (1500) 1590 (1510)  
530 (450)  
490 (420)  
58 (49)  
530 (450)  
500 (430)  
59 (49)  
530 (450)  
500 (420)  
59 (49)  
470 (410)  
470 (400)  
58 (48)  
460 (400)  
460 (390)  
58 (48)  
470 (400)  
460 (400)  
58 (48)  
0.1 / External  
1.00 / Internal  
1.00 / External  
58 (48)  
57 (48)  
58 (48)  
58 (48)  
57 (48)  
57 (47)  
57 (48)  
57 (48)  
SCPI Programmed 2  
Measure  
Source-Measure Sweep  
To GPIB To USB  
NPLC/Trigger Origin  
0.01 / Internal  
0.01 / External  
0.1 / Internal  
To Mem.  
To GPIB  
To USB  
To LAN  
To Mem.  
To LAN  
3130 (2800) 3060 (2760) 3000 (2790) 3010 (2710) 1710 (1630) 1610 (1600) 1440 (1380) 1690 (1590)  
2350 (2200)  
540 (460)  
510 (440)  
59 (49)  
2320 (2170)  
540 (450)  
510 (430)  
59 (49)  
2340 (2190) 2320 (2130) 1680 (1590) 1560 (1570)  
1410 (1360) 1660 (1560)  
540 (460)  
510 (440)  
59 (49)  
540 (450)  
510 (430)  
59 (49)  
470 (410)  
470 (400)  
58 (48)  
470 (410)  
470 (400)  
58 (48)  
450 (390)  
450 (390)  
57 (48)  
470 (410)  
470 (400)  
58 (48)  
0.1 / External  
1.00 / Internal  
1.00 / External  
58 (49)  
58 (49)  
58 (49)  
58 (49)  
58 (48)  
58 (48)  
57 (47)  
58 (48)  
Notes  
1. Reading rates applicable for voltage or current measurements, autozero off, autorange off, filter off, binary reading format, and source readback off.  
2. SCPI programming mode. Speeds do not apply to SCPI 2400 mode.  
10 TEK.COM  
2450-EC Graphical Potentiostat  
General Characteristics (default mode unless specified)  
Factory Default Standard  
Power-Up  
SCPI Mode.  
Source Output Modes  
Source Memory List  
Memory Buffer  
Real-Time Clock  
Remote Interfaces  
GPIB  
Fixed DC Level, Memory/Configuration List (mixed function), Stair (linear and log).  
100 points max. (SCPI 2400 Mode only).  
>250,000 readings. Includes selected measured value(s) and time stamp.  
Lithium battery backup (3 yr. + battery life).  
IEEE-488.1 compliant. Supports IEEE-488.2 common commands and status model topology.  
2.0 Full Speed USBTMC.  
USB Device  
(rear panel, type B)  
USB Host  
(front panel, type A)  
USB 2.0, support for flash drives, FAT32.  
RJ-45 (10/100BT).  
Ethernet  
Digital I/O Interface:  
Lines  
6 Input/Output user defined for digital I/O or triggering.  
9-pin female D.  
Connector  
Input Signal Levels  
Input Voltage Limits  
Maximum Source Current  
Maximum Sink Current  
5V Power Supply Pin  
Handler  
0.7 V (maximum logic low), 3.7 V (minimum logic high).  
–0.25 V (Abs. minimum), +5.25 V (Abs. maximum).  
+2.0 mA @ >2.7 V (per pin).  
–50 mA @ 0.7 V (per pin, solid-state fuse protected).  
Limited to 500 mA @ >4 V (solid-state fuse protected).  
User definable Start of Test, End of Test, 4 category bits.  
SCPI or TSP command sets.  
Programmability  
TSP Mode  
Embedded Test Script Processor (TSP) accessible from any host interface.  
Static or DHCP.  
IP Configuration  
Expansion Interface  
The TSP-Link expansion interface allows TSP enabled instruments to trigger and communicate with  
each other.  
LXI Compliance  
Display  
1.4 LXI Core 2011.  
5 inch capacitive touch, color TFT WVGA (800×480) with LED backlight.  
Front: Banana. Rear: Triaxial (3-lug).  
Input Signal Connections  
Interlock  
Active High Input.  
Cooling  
Forced air, variable speed.  
Over Temperature Protection  
Power Supply  
VA Rating  
Internally sensed temperature overload puts unit in standby mode.  
100 V to 240 V RMS, 50–60 Hz (automatically detected at power up).  
190 volt-amps max.  
Altitude  
Maximum 2000 meters above sea level.  
EMC  
Conforms to European Union EMC Directive.  
NRTL listed to UL61010-1 and UL61010-2-30. Conforms with European Union Low Voltage Directive.  
MIL-PRF-28800F Class 3 Random.  
Safety  
Vibration  
TEK.COM 11  
Datasheet  
Warm-Up  
1 hour to rated accuracies.  
Dimensions  
With bumpers and handle: 106 mm high × 255 mm wide × 425 mm deep (4.18 in × 10.05 in × 16.75 in).  
Without bumpers and handle: 88 mm high × 213 mm wide × 403 mm deep (3.46 in × 8.39 in × 15.87 in.)  
Weight  
With bumpers and handle: 4.04 kg (8.9 lbs.).  
Without bumpers and handle: 3.58 kg (7.9 lbs.).  
Environment  
Operating: 0°–50°C, 70% R.H. up to 35°C. Derate 3% R.H./°C, 35°–50°C.  
Storage: –25°C to 65°C.  
Supplied Accessories  
Electrochemistry Translation Cable Accessory Kit  
8608  
High Performance Test Leads  
CA-180-3A  
CS-1616-3  
USB-B-1  
TSP-Link/Ethernet Cable  
Safety Interlock Mating Connector  
USB Cable, Type A to Type B, 1m (3.3 ft)  
Documentation CD  
Application Test Scripts and Documentation on USB Flash Drive  
SMU Potentiostats and EC-Upgrade Kit  
Quick Start Guide  
Test Script Builder Software  
KickStart Startup Software  
LabVIEW and IVI Drivers  
(available at www.tektronix.com)  
(available at www.tektronix.com)  
(available at www.tektronix.com)  
Available Accessories  
Test Leads and Probes  
1754  
5804  
5805  
5806  
5808  
5809  
8605  
8606  
8608  
2-wire Universal 10-Piece Test Lead Kit  
Kelvin (4-Wire) Universal 10-Piece Test Lead Kit  
Kelvin (4-Wire) Spring-Loaded Probes  
Kelvin Clip Lead Set  
Low Cost Single-pin Kelvin Probe Set  
Low Cost Kelvin Clip Lead Set  
High Performance Modular Test Leads  
High Performance Modular Probe Kit  
High Performance Clip Lead Set  
12 TEK.COM  
2450-EC Graphical Potentiostat  
Cables, Connectors, Adapters  
237-ALG-2  
237-BAN-3A  
2450-TRX-BAN  
7078-TRX-*  
7078-TRX-GND  
8607  
3-slot Male Triax Connector to 3 Alligator Clips  
Triax to Banana Plug  
Triax to Banana Adapter. Converts the 4 Triax adapters on the rear panel to 5 banana jacks  
3-slot, Low Noise Triax Cable  
3-slot Male Triax To BNC Adapter (guard removed)  
2-wire, 1000V Banana Cables, 1 m (3.3 ft)  
CA-18-1  
Shielded Dual Banana Cable, 1.2 m (4 ft)  
CAP-31  
Protective Shield/Cap for 3-lug Triax Connectors  
Triax 3-lug Special Shorting Plug. Shorts center pin to outer shield  
Safety Interlock Mating Connector  
CS-1546  
CS-1616-3  
Communication Interfaces and Cables  
7007-1  
Shielded GPIB Cable, 1 m (3.3 ft)  
7007-2  
Shielded GPIB Cable, 1 m (6.6 ft)  
CA-180-3A  
KPCI-488LPA  
KUSB-488B  
USB-B-1  
CAT5 Crossover Cable for TSP-Link/Ethernet  
IEEE-488 Interface for PCI Bus  
IEEE-488 USB-to-GPIB Interface Adapter  
USB Cable, Type A to Type B, 1 m (3.3 ft)  
Triggering and Control  
2450-TLINK  
DB-9 to Trigger Link Connector Adapter.  
Trigger Link Cable, DIN-to-DIN, 1 m (3.3 ft)  
Trigger Link Cable, DIN-to-DIN, 2 m (6.6 ft)  
8501-1  
8501-2  
Rack Mount Kits  
2450-BenchKit  
4299-8  
Ears and Handle for 2450-NFP-RACK and 2450-RACK models  
Single Fixed Rack Mount Kit  
4299-9  
Dual Fixed Rack Mount Kit  
4299-10  
Dual Fixed Rack Mount Kit. Mount one 2450 and one Series 26xxB  
Dual Fixed Rack Mount Kit. Mount one 2450 and one Series 2400, Series 2000, etc.  
4299-11  
Test Fixtures  
8101-PIV  
DC Test Fixture  
Software Options  
Kickstart  
Instrument Control Software  
ACS Basic Edition  
Semiconductor Parametric Test Software for Component and Discrete Devices  
TEK.COM 13  
Datasheet  
Available Services  
2450-3Y-EW  
1 Year Factory Warranty extended to 3 years from date of shipment  
1 Year Factory Warranty extended to 5 years from date of shipment  
KeithleyCare® 3 Year ISO 17025 Calibration Plan  
KeithleyCare 3 Year Calibration w/Data Plan  
2450-5Y-EW  
C/2450-3Y-17025  
C/2450-3Y-DATA  
C/2450-3Y-STD  
C/2450-5Y-17025  
C/2450-5Y-DATA  
C/2450-5Y-STD  
KeithleyCare 3 Year Std. Calibration Plan  
KeithleyCare 5 Year ISO 17025 Calibration Plan  
KeithleyCare 5 Year Calibration w/Data Plan  
KeithleyCare 5 Year Std. Calibration Plan  
Ordering Information  
2450-EC  
Graphical Potentiostat, 200 V, 1 A, 20 W Instrument  
Warranty Information  
Warranty Summary  
This section summarizes the warranties of the 2450. For complete warranty information, refer to the 2450  
Reference Manual. Any portion of the product that is not manufactured by Keithley is not covered by this  
warranty and Keithley will have no duty to enforce any other manufacturer’s warranties.  
Hardware Warranty  
Software Warranty  
Keithley Instruments, Inc. warrants the Keithley manufactured portion of the hardware for a period of one  
year from defects in materials or workmanship; provided that such defect has not been caused by use of  
the Keithley hardware which is not in accordance with the hardware instructions. The warranty does not  
apply upon any modification of Keithley hardware made by the customer or operation of the hardware  
outside the environmental specifications.  
Keithley warrants for the Keithley produced portion of the software or firmware will conform in all material  
respects with the published specifications for a period of ninety (90) days; provided the software is used on  
the product for which it is intended in accordance with the software instructions. Keithley does not warrant  
that operation of the software will be uninterrupted or error-free, or that the software will be adequate for  
the customer’s intended application. The warranty does not apply upon any modification of the software  
made by the customer.  
14 TEK.COM  
Contact Information:  
Australia* 1 800 709 465  
Austria 00800 2255 4835  
Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777  
Belgium* 00800 2255 4835  
Brazil +55 (11) 3759 7627  
Canada 1 800 833 9200  
Central East Europe / Baltics +41 52 675 3777  
Central Europe / Greece +41 52 675 3777  
Denmark +45 80 88 1401  
Finland +41 52 675 3777  
France* 00800 2255 4835  
Germany* 00800 2255 4835  
Hong Kong 400 820 5835  
India 000 800 650 1835  
Indonesia 007 803 601 5249  
Italy 00800 2255 4835  
Japan 81 (3) 6714 3010  
Luxembourg +41 52 675 3777  
Malaysia 1 800 22 55835  
Mexico, Central/South America and Caribbean 52 (55) 56 04 50 90  
Middle East, Asia, and North Africa +41 52 675 3777  
The Netherlands* 00800 2255 4835  
New Zealand 0800 800 238  
Norway 800 16098  
People’s Republic of China 400 820 5835  
Philippines 1 800 1601 0077  
Poland +41 52 675 3777  
Portugal 80 08 12370  
Republic of Korea +82 2 6917 5000  
Russia / CIS +7 (495) 6647564  
Singapore 800 6011 473  
South Africa +41 52 675 3777  
Spain* 00800 2255 4835  
Sweden* 00800 2255 4835  
Switzerland* 00800 2255 4835  
Taiwan 886 (2) 2656 6688  
Thailand 1 800 011 931  
United Kingdom / Ireland* 00800 2255 4835  
USA 1 800 833 9200  
Vietnam 12060128  
* European toll-free number. If not  
accessible, call: +41 52 675 3777  
Find more valuable resources at TEK.COM  
Copyright © Tektronix. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that  
in all previously published material. Specification and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. All other trade names  
referenced are the service marks, trademarks or registered trademarks of their respective companies.  
020817.SBG 1KW-60118-2  

相关型号:

2450-6122TB

50 CONTACT(S), MALE, STRAIGHT TWO PART BOARD CONNECTOR, SOLDER
3M

2450-EC

2450-EC Graphical Potentiostat
TEKTRONIX

24500-10-100B

1.00mm (.039") Pitch Round Conductor Flat Cable
MOLEX

24500-20-100B

1.00mm (.039") Pitch Round Conductor Flat Cable
MOLEX

24500-24-100B

1.00mm (.039") Pitch Round Conductor Flat Cable
MOLEX

24500-26-100B

1.00mm (.050") Pitch Round Conductor Flat Cable - 28 AWG (7 x 36) Stranded Tinned, 26 Circuits
MOLEX

24500-30-100B

1.00mm (.050") Pitch Round Conductor Flat Cable
MOLEX

24500-34-100B

1.00mm (.050") Pitch Round Conductor Flat Cable
MOLEX

24500-40-100B

1.00mm (.039") Pitch Round Conductor Flat Cable - 28 AWG (7 x 36) Stranded, Tinned, 40 Circuits
MOLEX

24500-44-100B

1.00mm (.039") Pitch Round Conductor Flat Cable - 28 AWG (7 x 36) Stranded, Tinned, 44 Circuits
MOLEX

24500-50-100B

1.00mm (.039") Pitch Round Conductor Flat Cable
MOLEX

245001

Shock-Safe Panel Mount Type
LITTELFUSE